Storage and Delivery of Medical Gases

Published on 01/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 9840 times

Storage and Delivery of Medical Gases

David L. Vines

The hospital “oxygen service” is the origin from which the current technology-laden field of respiratory care evolved. Although respiratory therapists (RTs) have assumed many more challenging duties, ensuring the safe and uninterrupted supply of medical gases is still a key responsibility.

There are many commercially produced gases, but only a few are used medically (Table 37-1). Medical gases are classified as laboratory gases, therapeutic gases, or anesthetic gases. Laboratory gases are used for equipment calibration and diagnostic testing. Therapeutic gases are used to relieve symptoms and improve oxygenation of patients with hypoxemia. Anesthetic gases are combined with oxygen (O2) to provide anesthesia during surgery. It is important for RTs to be familiar with all aspects of gases used in the clinical setting, especially the chemical symbols, physical characteristics, ability to support life, and fire risk. In regard to fire risk, medical compressed gases are classified as either nonflammable (do not burn), nonflammable but supportive of combustion (also termed oxidizing), or flammable (burns readily, potentially explosive).1 Of the gases listed in Table 37-1, the focus of this chapter is on the therapeutic gases.

TABLE 37-1

Physical Characteristics of Medical Gases

Gas Chemical Symbol Color Taste Odor Can Support Life Flammability
Laboratory Gases
Nitrogen N Colorless Tasteless Odorless No Nonflammable
Helium He Colorless Tasteless Odorless No Nonflammable
Carbon dioxide CO2 Colorless Slightly acidic Odorless No Nonflammable
Therapeutic Gases
Air AIR Colorless Tasteless Odorless Yes Supports combustion
Oxygen O2 Colorless Tasteless Odorless Yes Supports combustion
Helium/oxygen (heliox) He/O2 Colorless Tasteless Odorless Yes Supports combustion
Carbon dioxide/oxygen CO2/O2 Colorless Slightly acidic Odorless No Supports combustion
Nitric oxide NO Colorless Tasteless Metallic No Supports combustion
Anesthetic Gas
Nitrous oxide N2O Colorless Slightly sweet Slightly sweet No Supports combustion

image

Characteristics of Medical Gases

Oxygen

Characteristics

O2 is a colorless, odorless, transparent, and tasteless gas.1 It exists naturally as free molecular O2 and as a component of a host of chemical compounds. O2 constitutes almost 50% by weight of the earth’s crust and occurs in all living matter in combination with hydrogen as water. At standard temperature, pressure, and dry (STPD), O2 has a density of 1.429 g/L, being slightly heavier than air (1.29 g/L). O2 is not very soluble in water. At room temperature and 1 atm pressure, only 3.3 ml of O2 dissolves in 100 ml of water.

O2 is nonflammable, but it greatly accelerates combustion. Burning speed increases with either (1) an increase in O2 percentage at a fixed total pressure or (2) an increase in total pressure of O2 at a constant gas concentration. Both O2 concentration and partial pressure influence the rate of burning.2

Production

O2 is produced through one of several methods. Chemical methods for producing small quantities of O2 include electrolysis of water and decomposition of sodium chlorate (NaClO3). Most large quantities of medical O2 are produced by fractional distillation of atmospheric air.1 Small quantities of concentrated O2 are produced by physical separation of O2 from air.

Fractional Distillation

Fractional distillation is the most common and least expensive method for producing O2. The process involves several related steps. First, atmospheric air is filtered to remove pollutants, water, and carbon dioxide (CO2). The purified air is liquefied by compression and cooled by rapid expansion (Joule-Thompson effect).

The resulting mixture of liquid O2 and nitrogen (N, N2) is heated slowly in a distillation tower. N2, with its boiling point of 195.8° C (320.5° F), escapes first, followed by the trace gases of argon, krypton, and xenon. The remaining liquid O2 is transferred to specially insulated cryogenic (low-temperature) storage cylinders. An alternative procedure is to convert O2 directly to gas for storage in high-pressure metal cylinders. These methods produce O2 that is approximately 99.5% pure. The remaining 0.5% is mostly N2 and trace argon. U.S. Food and Drug Administration (FDA) standards require an O2 purity of at least 99.0%.3

Physical Separation

Two methods are used to separate O2 from air.4 The first method entails use of molecular “sieves” composed of inorganic sodium aluminum silicate pellets. These pellets absorb N2, “trace” gases, and water vapor from the air, providing a concentrated mixture of more than 90% O2 for patient use. The second method entails use of a vacuum to pull ambient air through a semipermeable plastic membrane. The membrane allows O2 and water vapor to pass through at a faster rate than N2 from ambient air. This system can produce an O2 mixture of approximately 40%. These devices, called oxygen concentrators, are used primarily for supplying low-flow O2 in the home care setting. For this reason, details about the principles of operation and appropriate use are discussed in Chapter 51.

Air

Atmospheric air is a colorless, odorless, naturally occurring gas mixture that consists of 20.95% O2, 78.1% N2, and approximately 1% “trace” gases, mainly argon. At STPD, the density of air is 1.29 g/L, which is used as the standard for measuring specific gravity of other gases. O2 and N2 can be mixed to produce a gas with an O2 concentration equivalent to that of air. Medical-grade air usually is produced by filtering and compressing atmospheric air.1,5

Figure 37-1 shows a typical large medical air compressor system. In these systems, an electrical motor is used to power a piston in a compression cylinder. On its downstroke, the piston draws air through a filter system with an inlet valve. On its upstroke, the piston compresses the air in the cylinder (closing the inlet valve) and delivers it through an outlet valve to a reservoir tank. Air from the reservoir tank is reduced to the desired working pressure by a pressure-reducing valve before being delivered to the piping system.

For medical gas use, air must be dry and free of oil or particulate contamination.5 The most common method used for drying air is cooling to produce condensation. For avoidance of oil or particulate contamination, medical air compressors have air inlet filters and polytetrafluoroethylene (Teflon) piston rings as opposed to oil lubrication. Large medical air compressors must provide high flow (at least 100 L/min) at the standard working pressure of 50 pounds per square inch gauge (psig) for all equipment in use.

Smaller compressors (Figure 37-2) are available for bedside or home use. These compressors have a diaphragm or turbine that compresses the air and generally do not have a reservoir. This design limits the pressure and flow capabilities of these devices. For this reason, small compressors must never be used to power equipment that needs unrestricted flow at 50 psig, such as pneumatically powered ventilators (see Chapter 42). However, small diaphragm or turbine compressors are ideal for powering devices such as small-volume medication nebulizers (see Chapter 36).

Carbon Dioxide

At STPD, CO2 is a colorless and odorless gas with a specific gravity of 1.52 (approximately 1.5 times heavier than air).1 CO2 does not support combustion or maintain animal life. For medical use, CO2 usually is produced by heating limestone in contact with water. The gas is recovered from this process and liquefied by compression and cooling. The FDA purity standard for CO2 is 99%.3

Mixtures of O2 and 5% to 10% CO2 are occasionally used for therapeutic purposes as noted in Chapter 38. Therapeutic uses include the management of singultus (hiccups), prevention of the complete washout of CO2 during cardiopulmonary bypass, and regulation of pulmonary vascular pressures in some congenital heart disorders. However, CO2 mixtures are more commonly used for the calibration of blood gas analyzers (see Chapter 18) and for diagnostic purposes in the clinical laboratory.

Helium

Helium (He) is second only to hydrogen as the lightest of all gases; it has a density at STPD of 0.1785 g/L. He is odorless, tasteless, nonflammable, and chemically and physiologically inert. It is a good conductor of heat, sound, and electricity but is poorly soluble in water. Although He is present in small quantities in the atmosphere, it is commercially produced from natural gas through liquefaction to purity standards of at least 99%.3

He cannot support life, so breathing 100% He would cause suffocation and death. For therapeutic use, He must always be mixed with at least 20% O2. Heliox (a gas mixture of O2 and He) may be used clinically to manage severe cases of airway obstruction. Its low density decreases the work of breathing by making gas flow more laminar. He is discussed in more detail in Chapter 38.

Nitric Oxide

Nitric oxide (NO) is a colorless, nonflammable, toxic gas that supports combustion. It is produced by oxidation of ammonia at high temperatures in the presence of a catalyst. In combination with air, NO forms brown fumes of nitrogen dioxide (NO2). Together, NO and NO2 are strong respiratory irritants that can cause chemical pneumonitis and a fatal form of pulmonary edema. Exposure to high concentrations of NO alone can cause methemoglobinemia (see Chapter 11). High levels of methemoglobin can cause tissue hypoxia.

As discussed in Chapter 38, NO is approved by the FDA for use in the treatment of term and near-term infants for hypoxic respiratory failure. The American Academy of Pediatrics (AAP) has published a policy statement recommending the use of NO in the care of term and near-term infants when mechanical ventilation is failing because of hypoxic respiratory failure. The AAP suggests that NO be used before extracorporeal membrane oxygenation.6 A systemic review from the Cochrane database supports the recommendation that inhaled NO at 20 ppm may be beneficial in term and near-term infants who do not have a diaphragmatic hernia (see Chapter 31).7 The use of inhaled NO in the treatment of premature neonates with hypoxic respiratory failure does not improve outcomes and may increase the risk of intracranial hemorrhage.8

Nitrous Oxide

Nitrous oxide (N2O) is a colorless gas with a slightly sweet odor and taste that is used clinically as an anesthetic agent. Similar to O2, N2O can support combustion. However, N2O cannot support life and causes death if inhaled in pure form. For this reason, inhaled N2O must always be mixed with at least 20% O2. N2O is produced by thermal decomposition of ammonium nitrate.1

The use of N2O as an anesthetic agent is based on its central nervous system depressant effect. However, only dangerously high levels of N2O provide true anesthesia. N2O/O2 mixtures are almost always used in combination with other anesthetic agents.

Long-term human exposure to N2O has been associated with a form of neuropathy. In addition, epidemiologic studies have linked chronic N2O exposure with an increased risk of fetal disorders and spontaneous abortion.1 On the basis of this knowledge, the National Institute for Occupational Safety and Health (a division of the Occupational Safety and Health Administration) has set an upper exposure limit for hospital operating rooms of 25 ppm N2O.1

Storage of Medical Gases

Medical gases are stored either in portable high-pressure cylinders or in large bulk reservoirs. Bulk reservoirs require a separate distribution system to deliver the gas to the patient.

Gas Cylinders

The containers used to store and ship compressed or liquid medical gases are high-pressure cylinders. The design, manufacture, transport, and use of these cylinders are carefully controlled by both industrial standards and federal regulations. Gas cylinders are made of seamless steel and are classified by the U.S. Department of Transportation (DOT) according to their fabrication method. DOT type 3A cylinders are made from carbon steel, and DOT type 3AA containers are manufactured with a steel alloy tempered for higher strength.1

Markings and Identification

Medical gas cylinders are marked with metal stamping on the shoulders that supplies specific information (Figure 37-3).1,9 Although the exact location and order of these markings vary, the practitioner should be able to identify several key items of information.

The letters DOT or ICC (Interstate Commerce Commission) are followed by the cylinder classification (3A or 3AA) and the normal filling pressure in pounds per square inch (psi). Below this information usually is the letter size of the cylinder (E, G, and so on) followed by the cylinder serial number. A third line provides a mark of ownership, often followed by the manufacturer’s stamp or a mark identifying the inspecting authority. An abbreviation indicating the method of cylinder manufacturer is usually on the opposite side of the cylinder. Also in this area is information about the original safety test and dates of all subsequent tests.

Safety tests are conducted on each cylinder every 5 or 10 years, as specified in DOT regulations.1,9 During these tests, cylinders are pressurized to five thirds of their service pressure. While the cylinder is under pressure, technicians measure cylinder leakage, expansion, and wall stress. The notation EE followed by a number indicates the elastic expansion of the cylinder in cubic centimeters under the test conditions. An asterisk (*) next to the test date indicates DOT approval for 10-year testing. A plus sign (+) means the cylinder is approved for filling to 10% greater than its service pressure. An approved cylinder with a service pressure of 2015 psi can be filled to approximately 2200 psi. After hydrostatic testing, cylinders are subjected to internal inspection and cleaning.

In addition to these permanent marks, all cylinders are color-coded and labeled for identification of their contents.1,10 Table 37-2 lists the color codes for medical gases as adopted by the Bureau of Standards of the U.S. Department of Commerce.11 For comparison, the color codes adopted by the Canadian Standards Association also are included. Color codes are not standardized internationally. For this reason, cylinder color should be used only as a guide. As with any drug agent, the cylinder contents always must be identified through careful inspection of the label. To be absolutely sure about the O2 concentration provided by a cylinder, the user must analyze the gas before administering it (see Chapter 18).12

TABLE 37-2

Color Codes for Medical Gas Cylinders

Gas United States Canada
O2 Green White*
CO2 Gray Gray
N2O Blue Blue
Cyclopropane Orange Orange
He Brown Brown
C2H4 Red Red
CO2-O2 Gray/green Gray/white
He-O2 Brown/green Brown/white
N2 Black Black
Air Yellow* Black/white
N2-O2 Black/green Pink

C2H4, Ethylene.

*Vacuum systems historically are identified as white in the United States and yellow in Canada. For this reason, the CGA recommends that white not be used for any cylinders in the United States and that yellow not be used in Canada.

Cylinder Sizes and Contents

Letter designations are used for different sizes of cylinders (Figure 37-4). Sizes E through AA are referred to as “small cylinders” and are used most often for transporting patients and anesthetic gases. These small cylinders are easily identified because of their unique valves and connecting mechanisms. Small cylinders have a post valve and yoke connector. Large cylinders (F through H and K) have a threaded valve outlet (Figure 37-5) (discussed later).

Cylinder Safety Relief Valves

In a closed cylinder, any increase in gas temperature increases gas pressure. Should the temperature increase too much (as in a fire), the high gas pressure could rupture and explode the cylinder. To prevent this type of accident, all cylinders have high-pressure relief valves. These relief valves are of three basic designs: frangible disk, fusible plug, and spring-loaded. The frangible metal disk ruptures at a specific pressure. The fusible plug melts at a specific temperature. The spring-loaded valve opens and vents gas at a set high pressure. In each case, the activated valve vents gas from the cylinder and prevents pressure from becoming too high.

Most small cylinders have a fusible plug relief valve. Most large cylinders have a spring-loaded relief valve. These safety relief valves are always located in the cylinder valve stems.

Filling (Charging) Cylinders

How a cylinder is filled depends on whether its contents will be gaseous or liquid. Some gases stored in liquid form can remain at room temperature, but others must be maintained in a cryogenic (low-temperature) state. Cryogenic storage is discussed later.

Liquefied Gases

Gases with critical temperatures greater than room temperature can be stored as liquids at room temperature (see Chapter 6). These gases include CO2 and N2O. Rather than being filled to filling pressure, cylinders of these gases are filled according to a specified filling density. The filling density is the ratio between the weight of liquid gas put into the cylinder and the weight of water the cylinder could contain if full. The filling density for CO2 is 68%. This system allows the manufacturer to fill a cylinder with liquid CO2 up to 68% of the weight of water that a full cylinder could hold. The filling density of N2O is 55%.

Cylinder pressures for gases stored in the liquid phase are much lower than for gases stored in the gas phase. Because the liquid does not fill the entire volume of a cylinder, the space above the liquid surface contains gas in equilibrium with the liquid. The pressure in a liquid-filled cylinder equals the pressure of the vapor at any given temperature.

Pressure in a cylinder depends on the state of its contents. In a gas-filled cylinder, the pressure represents the force required to compress the gas into its smaller volume. In contrast, the pressure in a liquid-filled cylinder is the vapor pressure needed to keep the gas liquefied at the current temperature.

Measuring Cylinder Contents

Because of the previously described differences in the physical state of matter of compressed and liquid gases, different methods are needed to measure the contents of the cylinder.

Liquid Gas Cylinders

In a liquid gas cylinder or container, the measured pressure is the vapor pressure above the liquid. This pressure bears no relationship to the amount of liquid remaining in the cylinder. As long as some liquid remains (and the temperature remains constant), the vapor pressure and the gauge pressure remain constant. When all the liquid is gone and the cylinder contains only gas, the pressure decreases in proportion to a reduction in volume. Monitoring the gauge pressure of liquid gas cylinders is useful only after all the liquid vaporizes. Weighing a liquid-filled cylinder is the only accurate method for determining the contents.

Figure 37-6 compares the behavior of compressed gas and liquid gas cylinders during use. The vapor pressure of liquid gas cylinders varies with the temperature of the contents. The pressure in an N2O cylinder at 21.1° C (70° F) is 745 psig; at 15.6° C (60° F), the pressure decreases to 660 psig. As the temperature increases toward the critical point, more liquid vaporizes, and the cylinder pressure increases. If a cylinder of N2O warms to 36.4° C (97.5° F) (its critical temperature), all the contents convert to gas. Only at this temperature and higher does the cylinder gauge pressure accurately reflect cylinder contents.

Estimating Duration of Cylinder Gas Flow

When a cylinder of therapeutic gas is used, it often is necessary to predict how long the contents will last at a given flow. The duration of flow of a cylinder can be estimated if the following are known: (1) the gas flow, (2) the cylinder size, and (3) the cylinder pressure at the start of therapy. For a given flow, the more gas a cylinder holds, the longer it lasts. Conversely, the higher the flow, the shorter the emptying time. The duration of flow of a cylinder is directly proportional to the contents and inversely proportional to flow, as expressed in the following formula:

< ?xml:namespace prefix = "mml" />Duration of flow=ContentsFlow

image

The units commonly used in the United States for measurement of these quantities are not the same. Cylinder contents are generally specified in cubic feet or gallons, whereas gas flow normally is measured in liters. Table 37-3 provides the factors needed to convert these units.

TABLE 37-3

Gas Volume Conversion Factors

Liters Cubic Feet Gallons
28.316 1 7.481
1 0.03531 0.2642
3.785 0.1337 1

image

Rather than memorizing various cylinder contents and constantly converting metric and English units, the user can quickly calculate duration of flow by using cylinder factors. Cylinder factors are derived for each common gas and cylinder size with the following formula:

Cylinder factor(L/psig)=Cubic feet(full cylinder)×28.3Pressure(full cylinder)in psig

image

In the numerator of the previous equation, the English-metric conversion constant (28.3) is used to convert cubic feet to liters. Dividing the resulting volume by the pressure in a full cylinder yields the cylinder factor. The derived factor represents the volume of gas leaving a given cylinder for every 1-psig decrease in pressure. Table 37-4 provides cylinder factors for the therapeutic medical gases and common cylinder sizes.

TABLE 37-4

Factors for Calculation of Cylinder Duration of Flow (Minutes)

Gas Cylinder Size
D E G H and K
O2, O2/N2, air 0.16 0.28 2.41 3.14
O2/CO2 0.20 0.35 2.94 3.84
He/O2 0.14 0.23 1.93 2.50

image

When the factor for a given gas and cylinder is known, calculating the duration of flow is a simple matter of applying the following equation:

Duration of flow(min)=Pressure(psig)×Cylinder factorFlow(L/min)

image

A wide margin of safety must be allowed in estimation of cylinder duration of flow. This principle is especially important if the RT cannot be present during use and must return with a full cylinder. Some clinicians return 30 to 40 minutes before the calculated time; others compute duration of flow to a level of 300 to 500 psig rather than 0 psig (empty). Assuming the calculations are correct and there is no change in flow, both methods ensure an uninterrupted supply. The accompanying Rule of Thumb presents a shortcut for estimating cylinder duration of flow.

Estimating Duration of Liquid Oxygen Cylinder Gas Flow

The only accurate method for determining the volume of gas in a liquid-filled cylinder is by weight. Because 1 L of liquid O2 weighs 2.5 lb and produces 860 L of O2 in its gaseous state, the amount of gas in a liquid O2 cylinder can be calculated with the following formula:

Amount of gas in cylinder=Liquid O2 weight(lb)×8602.5 lb/L

image

After the amount of O2 remaining in the cylinder is determined, the duration of the gas in minutes can be calculated with the following formula:

Duration of gas(min)=Amount of gas in cylinder(L)Flow(L/min)

image

As with gaseous O2 cylinders, a wide margin of safety is needed for estimation of cylinder duration. This margin of safety varies with the size of the portable O2 unit or large storage container.

Gas Cylinder Safety

The following guidelines for cylinder safety are from the recommendations of the National Fire Protection Agency (NFPA)2 and the Compressed Gas Association (CGA).1 For ease of use, these safety guidelines are divided into cylinder storage, transport, and use.

Cylinder Storage

The following guidelines apply to cylinder storage:

• Store gas cylinders in racks or chain cylinders to the wall to prevent them from falling or becoming damaged.

• Other than the wooden racks used to store the cylinders, store no other combustible material in the vicinity of cylinders or gas supply systems.

• Store gas cylinders away from sources of heat. Keep the cylinder temperature less than 125° F (<51.7° C).

• Store flammable gases separately from gases that support combustion, such as air, O2, and N2O.

• If a cylinder is not in use, keep the protective cylinder cap in place.

• Do not store air compressors and gas cylinders together. A fire involving one or the other can damage both gas delivery systems.

• Contain and store cylinder supply systems in an enclosure constructed of a material with at least a 1-hour fire resistive rating that is well ventilated and well drained.

• Segregate full and empty cylinders; store them separately if possible.

• Place on each door or gate of the enclosure a sign that cautions the presence of an oxidizing gas and alerts against smoking. This sign must be readable from a distance of at least 5 ft (1.5 m).

• Store liquid O2 containers in a cool, well-ventilated area because of the venting of small amounts of O2 from these low-pressure containers. The venting of O2 prevents these containers from overpressurizing because liquid O2 is continuously converting to gaseous O2.

Cylinder Use

The following guidelines apply to cylinder use:

• Secure gas cylinders at the patient’s bedside in a way that prevents them from falling. Secure cylinders to the wall with a chain, bind or chain them to a suitable cart, or support the cylinder with a stand.

• Do not use flammable materials, especially oil or grease, on regulators, cylinders, fittings, or valves. This restriction includes oily hands, rags, and gloves.

• Never cover a cylinder with any material, including bed linens or hospital gowns.

• Open the cylinder valve slightly to remove dust and dirt before attaching the regulator. When slightly opening the valve, ensure no one is in front of the valve. “Crack” the cylinder before bringing it to the patient’s bedside.

• Never use cylinder valves or regulators that need repair.

• Do not alter or deface cylinder markings or color.

• Never place cylinders near sources of heat.

• Never secure cylinders to movable objects unless the object has an apparatus that can contain the cylinder safely.

• Ensure that the connection between the regulator and the cylinder valve is an American standard safety system (ASSS) for H and G cylinders and a pin-index safety system (PISS) for E cylinders.

• When O2 is in use, post a “No Smoking” sign unless signs in the entrances are posted that prohibit smoking in the facility.

Bulk Oxygen

Large acute care facilities use large volumes of O2 every day. To meet these needs, a centralized bulk storage and delivery system is required. By definition, bulk O2 storage systems hold at least 20,000 cubic ft of gas, including the unconnected reserves that are on site.2 Bulk O2 may be stored in either gaseous or liquid form, but liquid storage is most common. When needed, the O2 flows from this central source throughout the facility through a piping system with outlets conveniently located.

A bulk O2 system has several advantages over portable cylinders. Although initially expensive to construct, bulk O2 systems are far less expensive over the long-term. Bulk O2 systems are less prone to interruption. These systems eliminate the inconvenience and hazard of transporting and storing numerous cylinders. Bulk O2 systems regulate delivery pressures centrally, eliminating the need for separate pressure-reducing valves at each outlet. These systems also operate at low pressures, making them much safer than high-pressure cylinders.

Safety standards for bulk O2 systems are set by the NFPA and are subject to further control by local fire and building codes.2 RTs should be familiar with both bulk units in general and the specific gas supply systems used in their facilities.

Gas Supply Systems

There are three types of centrally located gas supply systems: an alternating supply system or cylinder manifold system, a cylinder supply system with reserve supply, and a bulk gas system with a reserve.2 The alternating supply or cylinder manifold system consists of large (normally H or K size) cylinders of compressed O2 banked together in series (Figure 37-7). This alternating supply system has two sides: a primary bank and a reserve bank. When the pressure in the primary bank decreases to a set level, a control valve automatically switches over to the reserve bank. When this occurs, the primary bank is taken off-line, and the empty cylinders are replaced with full ones. The replenished primary bank becomes the reserve bank. Some large alternating supply systems are permanently fixed and are refilled on site by a supply truck. These cylinder manifold systems have pressure-reducing valves for regulation of delivered pressure and normally have low-pressure alarms. These alarms sound when reserve switchover occurs, and they warn of impending depletion or malfunction. Cylinder manifolds or alternating supply systems are used to supply O2 from a central location in small facilities or to supply specialty gases, such as N2O, to operating rooms (Figure 37-8).

A cylinder supply system with a reserve consists of a primary supply, a secondary supply, and a reserve supply. When the primary gas supply is depleted by the demand, this supply system automatically switches to the secondary supply. Master signal panels indicate that the changeover has occurred. This supply system operates in a manner similar to the alternating system except that this system has a reserve supply if primary and secondary supplies become depleted. Liquid containers may be used as the primary and secondary gas sources, but the reserve supply usually is high-pressure gas cylinders. Gas cylinders are used as the reserve supply because low-pressure liquid containers lose approximately 3% of the supply per day.2

For economy, safety, and convenience, most large health care facilities use a liquid bulk O2 system. A small volume of liquid O2 provides a very large amount of gaseous O2 and minimizes space requirements. However, along with this advantage comes a major problem. O2 has a critical temperature well below room temperature (−118.6° C [−181.4° F]).1 Liquid O2 must continually be stored below this temperature, or it reverts to its gaseous state.

To stay in liquid form, O2 is stored in large stand tanks (Figure 37-9) at relatively low pressure (<250 psig). These stand tanks are similar to giant thermos bottles, consisting of inner and outer steel shells separated by an insulated vacuum chamber (Figure 37-10). Because it eliminates heat conduction, the vacuum keeps the liquid O2 below its critical temperature without refrigeration. When it flows through vaporizer coils exposed to ambient temperature, the liquid O2 quickly converts back to a gas. With the O2 in its gaseous form, the pressure is decreased to the standard working pressure of 50 psi by a pressure-reducing valve. A safety vent allows vaporized liquid O2 to escape if warming causes cylinder pressure to increase above a set limit.

Smaller liquid cylinders are used for home O2 supply. These cylinders come in several sizes and hold between image cubic ft and image cubic ft of liquid O2. Small liquid O2 cylinders are refilled on site by means of transfer of liquid O2 from a large cylinder. Chapter 51 describes the use of these small liquid O2 cylinders in the home.

Bulk Oxygen Safety Precautions

The NFPA sets standards for the design, construction, placement, and use of bulk O2 systems.2 A key provision in these standards is the requirement for a reserve or backup gas supply to equal the average daily gas usage of the hospital. To meet this requirement, most large facilities have a second, smaller liquid stand tank. Smaller facilities may use a cylinder gas manifold as the backup.

Failure of bulk O2 supply systems has been reported with resultant major problems.1315 Failure of a bulk O2 supply can be life-threatening to any patient receiving O2 or gas-powered ventilatory support. For this reason, the respiratory care staff must be prepared. Adherence to an established protocol is a quick way to identify and prioritize all affected patients. When affected patients are identified, staff members move appropriate backup equipment to the bedside (e.g., portable cylinders, bag-valve-mask resuscitators). Trained personnel bypass the failed system and provide needed patient support, while engineers determine the cause of the failure and correct it.

Distribution and Regulation of Medical Gases

Before it can be administered to a patient, a medical gas must be delivered to the bedside and the pressure reduced to a workable level. This is the primary function of gas distribution and regulation systems. Modern hospital gas distribution systems deliver bulk O2 and compressed air to patient rooms and special care areas through an elaborate piping network. This network may include a vacuum source and, for surgical areas, N2O. Patient transport still requires the use of portable cylinders. Whether delivery occurs by central bulk supply or cylinder, patient safety is always the primary aim. For this reason, RTs must be proficient in the use of both delivery systems.

Central Piping Systems

Structural standards for piping systems are established by the NFPA and are described in more detail elsewhere.2 Figure 37-11 shows a simple central piping gas system. The gas pressure in a central piping system normally is reduced to the standard working pressure of 50 psi at the bulk storage location. A main alarm warns of decreases in pressure or interruptions in flow from the source. Zone valves (Figure 37-12) throughout the system can be closed for system maintenance or in case of fire. Wall or station outlets at the delivery sites allow connection of various types of equipment to the gas distribution system. Because most delivery outlets include O2, air, vacuum, and possibly N2O, special safety connectors are used to help prevent accidental misconnections.

Safety Indexed Connector Systems

One of the greatest risks in medical gas therapy is giving the wrong gas to a patient. Carefully reading the cylinder or outlet labels is the best way to avoid these accidents. However, human error does occur. For this reason, the industry has developed indexed safety systems for gas delivery and regulation equipment. These safety systems make misconnection between pieces of equipment nearly impossible. For example, an indexed safety system normally prevents connecting a cylinder of N2O to an O2 delivery system. Three basic indexed safety systems are used in the delivery and regulation of medical gases: (1) the American National Standard/Compressed Gas Association Standard for Compressed Gas Cylinder Valve Outlet and Inlet Connections, or the ASSS; (2) the diameter-index safety system (DISS); and (3) the PISS.16,17

American Standard Safety System

Adopted in the United States and Canada, the ASSS provides standards for threaded high-pressure connections between large compressed gas cylinders (sizes F through H/K) and their attachments.16 Specifications exist for more than 60 gases and gas mixtures. Figure 37-13 shows a typical ASSS connection between a threaded cylinder outlet and a pressure-reducing valve nipple. Use of the ASSS standards makes misconnections difficult because the size (bore) of the cylinder outlet and its threading differ based on the type of gas in the cylinder.

Because there are only 26 connections for the 62 listed gases and mixtures, each gas may not have a unique connection. Some gases have identical connections. Catalogs of cylinder equipment show the connection specifications for each type of cylinder and gas. A typical description for a large cylinder of O2 is as follows: CGA-540 0.903-14NGO-RH-Ext. The connection for the threaded outlet of this cylinder is listed by the CGA as connection number 540. The outlet has a thread diameter (bore) of 0.903 inch; there are 14 threads per inch; and the threads are right-handed (RH) and external (Ext). It generally is necessary to use only one or two outlet connections because most of the gases that are used by RTs are grouped within a few connector sizes. However, practitioners should be familiar with the classification scheme in general because expanding instrumentation and scope of services may bring RTs in contact with other gases and gas systems.

Pin-Index Safety System

Pin indexing is part of the ASSS but applies only to the valve outlets of small cylinders, up to and including size E. These cylinders have a yoke type of connection. Figure 37-14 illustrates the general structure of the pin-indexed yoke connection. The upper yoke fits over the lower valve stem. Two pins, projecting from the inner surface of the yoke connector, mate with two pinholes bored into the valve stem. Proper pin position aligns the small receiving nipple of the yoke with the recessed cylinder valve outlet. Tightening the hand screw on the yoke firmly seats the receiving nipple into the valve outlet. A nylon washer or bushing typically is used to ensure a leak-free connection.

Similar to the ASSS, the PISS helps prevent accidental misconnections between pieces of equipment. The exact positions of pins and pinholes vary for each gas. Unless the pins and holes align perfectly, the yoke nipple cannot seat in the recessed valve outlet. Six holes and pin positions constitute the total system. Because overlapping holes cannot be used, there are 10 possible pin combinations. Figure 37-15 is a diagram of the location of all six possible holes and their index numbers. Table 37-5 lists the gases included in the PISS system, including their index positions.

TABLE 37-5

Pin Index Hole Positions*

Gas Pin Positions
O2 2-5
O2/CO2 (CO2 not >7%) 2-6
He/O2 (He not >80%) 2-4
C2H4 1-3
N2O 3-5
C3H6 3-6
He/O2 (He >80%) 4-6
O2/CO2 (CO2 >7%) 1-6
Air 1-5

image

C2H4, Ethylene; C3H6, cyclopropane.

*See Figure 37-15.

image
FIGURE 37-15 Location of the pin-index holes in the cylinder valve face for different gases. See Table 37-5 for pin-index hole locations for various gases.

Diameter-Index Safety System

The ASSS and the PISS provide standards for high-pressure connections between cylinders and equipment; the DISS was established to prevent accidental interchange of low-pressure (<200 psig) medical gas connectors.17 RTs typically find DISS connections (1) at the outlets of pressure-reducing valves attached to cylinders; (2) at the station outlets of central piping systems; and (3) at the inlets of blenders, flowmeters, ventilators, and other pneumatic equipment.

As shown in Figure 37-16, the DISS connection consists of an externally threaded body and a mated nipple with a nut. As the two parts are joined, the shoulders of the nipple and the bores of the body mate, with the union held together by a hand-tightened hex nut. Indexing is achieved by varying the dimensions of the borings and shoulders. There are 11 indexed DISS connections and 1 connection for O2, for a total of 12.17 The standard threaded O2 connector (0.5625 inch in diameter and 18 threads per inch) preceded adoption of this safety system. Nonetheless, it has been assigned a DISS number of 1240.

Although O2 and air are generally used from a central outlet, it may be necessary to administer other gases that have different DISS connections. To avoid stocking a large variety of pressure regulators, flowmeters, and connectors for special gas use, adapters can be used to convert various DISS connections so that they can be used for different purposes. Using adapters to bypass a safety system carries the increased risk of misconnection. For this reason, RTs should exercise extreme caution when adapting equipment connections. Misconnections have occurred, with negative patient consequences.13,18

Quick-Connect Systems

Station outlets at the patient’s bedside allow quick access to a bulk supply of O2 and air or a vacuum source. Station outlets have DISS connections or quick-connect systems that are gas-specific or vacuum-specific. Various manufacturers have designed specially shaped connectors for each gas (Figure 37-17). Because each connector has a distinct shape, it does not fit into an outlet for another gas, and each manufacturer has its own unique design. For this reason, connectors from different manufacturers are not interchangeable. As long as a facility is standardized for a single quick-connect system, this incompatibility is seldom a problem.

A variety of safety systems help prevent inadvertent misconnections between medical delivery systems and equipment. Figure 37-18 summarizes the use of and relationships between the ASSS, PISS, and DISS systems as applied to cylinder gases. Proficiency in the proper use of these systems is a basic skill of RTs.

Regulating Gas Pressure and Flow

Whatever the source of medical gas, for safe administration to a patient, the pressure and flow must be regulated. If the goal is solely a reduction in gas pressure, a reducing valve is used. For control of gas flow to a patient, a flowmeter is used. If control of both pressure and flow is needed, a regulator is used.

Cylinder gases such as O2 and air exert a pressure that is much too high for use with respiratory care equipment. For use at the bedside, these high pressures must be reduced to a lower “working” level. In the United States, this working pressure is 50 psig. For bulk delivery systems with individual station outlets, built-in reducing valves decrease the delivered pressure to 50 psig. This standard pressure can be directly applied to power devices such as ventilators (see Chapter 42). However, if the goal is to control gas delivery to a patient for O2 therapy or nebulized medication (see Chapters 36 and 38), a flowmeter must also be used.

High-Pressure Reducing Valves

There are two basic types of high-pressure reducing valves: single-stage and multiple-stage. Reducing valves are available as preset or adjustable. Although all of these valves function on the same principle, the design, features, and use are different. This section differentiates preset reducing valves and adjustable reducing valves and discusses multiple-stage reducing valves.

Preset Reducing Valve

Figure 37-19 shows the basic design of a high-pressure preset reducing valve. High-pressure gas (2200 psig for O2) enters through the valve (A), with the inlet pressure displayed on the pressure gauge (B). The body of the valve is divided into a high-pressure chamber (C) and an ambient-pressure chamber (D) by a flexible diaphragm (E). Attached to the diaphragm in the ambient-pressure chamber is a spring (F), which is fixed to the other side of the chamber. Also attached to the diaphragm, but in the high-pressure chamber, is a valve stem (G) that sits on the high-pressure inlet (H). Gas flows through the valve inlet (H) into the high-pressure chamber and on to the gas outlet (I). The pressure chamber is supplied with a safety vent (L) preset to 200 psig to release pressure in the event of malfunction.

The spring tension is calibrated to give when the pressure on the diaphragm exceeds 50 psig. When this happens, the valve stem is pushed forward and closes the high-pressure inlet, preventing further entry of gas into the reducing valve. However, as long as gas is allowed to escape from the pressure chamber through the outlet (I), the inlet valve remains open and allows gas flow. The regulator maintains a balance between outlet flow and inlet pressure. Automatic adjustment of the diaphragm-spring combination keeps the pressure in the high-pressure chamber at a near-constant 50 psig—hence the name preset. Preset reducing valves are normally used in conjunction with high-pressure gas cylinders to decrease the pressure to the standard 50 psig used with most respiratory care equipment.

Adjustable Reducing Valve

Although most respiratory care equipment works at the standard 50 psig, some devices need variable pressures. To provide variable outlet pressures from a high-pressure gas source, an adjustable reducing valve is needed. Figure 37-20 shows the basic design of a high-pressure adjustable reducing valve. As with the preset design, the inlet valve (H) remains open until the gas pressure exceeds the spring tension, displacing the diaphragm and blocking further gas entry. However, while the preset reducing valve provides a fixed pressure, the adjustable reducing valve allows a change in outlet pressure. Outlet pressure can be changed with a threaded hand control (K) attached to the end of the diaphragm spring. Changing the tension on the valve spring varies pressure over a wide range, usually between 0 psig and 100 psig.

The adjustable reducing valve commonly is used in combination with a Bourdon-type flow gauge (discussed later). The combination of a flowmeter with a reducing valve is called a regulator.

Multiple-Stage Reducing Valve

As the name suggests, a multiple-stage reducing valve reduces pressure in two or more steps. Multiple-stage reducing valves can be either preset or adjustable and can be combined with a flowmeter device as a true regulator. Two-stage reducing valves are used occasionally, and three-stage units are rarely needed. A two-stage reducing valve functions as two single-stage reducing valves working in series. Gas enters the first stage, where the pressure is reduced to an intermediate level (usually 200 to 700 psig). Gas then enters the second stage, where the pressure is decreased to working level (usually 50 psig). Because each pressure chamber has one safety relief vent, the user usually can determine the number of stages in a reducing valve by noting the number of relief vents present. Because they reduce pressure in multiple steps, these valves provide more precise and smooth flow control. However, they are larger and more expensive than single-stage reducing valves. For this reason, a multiple-stage reducing valve should be considered only if minimal fluctuations in pressure or flow are critical factors, as in research activities. For routine hospital work, single-stage reducing valves are satisfactory.

Proper Use of High-Pressure Reducing Valves

When a cylinder attached to a high-pressure reducing valve is open, gas undergoes rapid decompression followed by rapid recompression. Because the recompression is adiabatic (see Chapter 6), the gas temperature quickly increases. These rapid pressure and temperature changes are potentially hazardous. Rapid pressure swings can cause failure of reducing valve components.19 Failed components can become high-velocity projectiles, endangering the practitioner and the patient. Rapid temperature changes can ignite combustible materials.20 Ignition of combustible materials in the presence of 100% O2 can cause an explosion. Box 37-1 provides guidelines for minimizing the risk associated with setting up O2 cylinders with a high-pressure reducing valve or regulator.1

Box 37-1   Safe Procedure for Setup of an Oxygen Cylinder and Reducing Valve or Regulator

1. Secure the cylinder according to the CGA guidelines. Verify the contents from the label that matches the color code and valve indexing.

2. Remove the protective cap or wrap, and inspect the cylinder valve to ensure that it is free of dirt, debris, and oil.

3. Warn any persons present that the cylinder valve is about to be “cracked” and that it will make some noise. Turn the cylinder valve away from persons present, stand to the side, and quickly open and close the valve. This removes any dust or small debris from the cylinder valve outlet.

4. Inspect the valve or regulator inlet for debris, dirt, and oil. Check the device label, and confirm that it is intended for high-pressure service and for use with the gas to be administered. O2-reducing valves and regulators should have a label stating: Oxygen: Use No Oil.

5. After the valve or regulator inlet is confirmed to be free of contaminants, securely tighten (but do not force) the device onto the cylinder outlet. When making connections to the cylinder, use appropriate wrenches that are free of oil and grease. Never use pipe wrenches. Use only cylinder valve connections that conform to the ASSS and the PISS. Low-pressure connections must comply with the DISS or be noninterchangeable, low-pressure quick connects. Never connect fixed or adjustable orifices or metering devices directly to a cylinder without a pressure-reducing valve.

6. Confirm that the regulator or reducing valve is in the off or closed position, and slowly open the cylinder valve to pressurize the attached reducing valve or regulator. After pressurization has occurred, open the cylinder valve completely and turn it back one-fourth to one-half turn (this maneuver prevents a condition known as “valve freeze,” in which the valve cannot be turned).

Low-Pressure Gas Flowmeters

As with drugs, giving a medical gas to a patient requires knowledge of the dosage being delivered. Physicians often prescribe O2 dosage as a flow, in liters per minute. In addition, certain gas-mixing equipment requires accurate knowledge of input flows, sometimes involving two or more gases. Flowmeters allow the rate of gas flow to a patient to be set and controlled. When the gas source is a high-pressure gas cylinder, a regulator (reducing valve plus flowmeter) is required. However, when the source is a bulk central supply system, the pressure has already been reduced to 50 psig by the time it reaches the outlet stations; this eliminates the need for pressure reduction and requires only a flowmeter.

Three categories of flowmeters are used in respiratory care: the flow restrictor, the Bourdon gauge, and the Thorpe tube. The Thorpe tube has two different designs: pressure compensated or not pressure compensated (uncompensated). Although uncompensated Thorpe tubes are rare, they may still be used at some institutions. For this reason, the principles underlying each of the four types of flow metering devices are compared and contrasted.

Flow Restrictor

The flow restrictor is the simplest and least expensive flowmeter device. As shown in Figure 37-21, a flow restrictor consists solely of a fixed orifice calibrated to deliver a specific flow at a constant pressure (50 psig). The operation of the flow restrictor is based on the principle of flow resistance, as described in Chapter 6. Specifically, the flow of gas through a tube can be quantified with the following equation:

image
FIGURE 37-21 Flow restrictor.

R=P1P2V

image

Rearranging the equation to solve for flow (V) yields the following:

V=P1P2R

image

where V is the volumetric flow per unit time, P1 is the pressure at the upstream point (point 1), P2 is the pressure at the downstream point (point 2), and R is the total resistance to gas flow.

By design, a flow restrictor requires a source of constant pressure (usually 50 psig). As long as the source pressure remains fixed, P1 − P2 should stay constant. With a fixed-size orifice, the flow resistance (R) also remains constant. The rate of gas flow through a flow restrictor can be increased by increasing P1 (upstream pressure) or by selecting a larger orifice size. Both fixed and adjustable orifice flow restrictors are used clinically. Commercially produced flow restrictors are calibrated at 50 psig. Table 37-6 summarizes the advantages and disadvantages of flow restrictors.

TABLE 37-6

Advantages and Disadvantages of Flow Restrictors

Advantages Disadvantages
Low-cost, simple, reliable (no moving parts) Different versions required for different flows
Cannot be set to incorrect flow Accuracy varies with changes in source and downstream pressures
Can be used in any position (gravity-independent) Cannot be used with high-resistance equipment

Bourdon Gauge

A Bourdon gauge (Figure 37-22) is a flowmeter device that is always used in combination with an adjustable pressure-reducing valve. Similar to the flow restrictor, the Bourdon gauge uses a fixed orifice. In contrast to the flow restrictor, the Bourdon gauge operates under variable pressures, as adjusted with the pressure-reducing valve. The Bourdon gauge is a fixed-orifice, variable-pressure flowmeter, so increasing the upstream pressure increases gas flow out of the device unless downstream pressure also increases.

As shown in Figure 37-23, a Bourdon gauge has a calibrated fixed orifice (A), which creates outflow resistance. The gauge itself is attached to the flow stream with a connector (B) located proximal to the orifice. Inside the gauge is a curved, hollow, closed tube (C) that responds to pressure changes by changing shape. The force of gas pressure tends to straighten the tube, causing its distal end to move. This motion is transmitted to a gear assembly and indicator needle (D). A numbered scale is calibrated to read the needle movement in units of flow (liters per minute).

As with the flow restrictor with fixed orifice, the output flow of the Bourdon gauge is proportional to the driving pressure. However, the Bourdon gauge provides a continuous range of flow, which the user adjusts by altering the driving pressure. Although the gauge actually measures pressure changes, it displays the corresponding flow.

As with a flow restrictor, gravity does not affect a Bourdon gauge. The Bourdon gauge is the best choice when a flowmeter cannot be maintained in an upright position. This situation is common when a patient is being transported with a portable O2 source. In these instances, keeping the E cylinder upright is seldom easy, and movement of both the O2 supply and the patient is common. Combined with its continuous range of flows, this feature makes the Bourdon gauge the metering device of choice for patient transport.

The main disadvantage of the Bourdon gauge is its inaccuracy when pressure distal to the orifice (downstream pressures) changes. Specifically, if downstream pressure increases (as when high-resistance equipment is used), the pressure difference across the orifice and actual output flow decrease. However, the Bourdon gauge flow reading depends on upstream pressure, which stays constant. In this situation, the gauge reading is falsely higher than the actual delivered flow. Because it measures upstream pressure, the gauge registers flow even when the outlet is completely blocked (Figure 37-24). A user who needs accurate flow when using a device that creates high resistance should not select a Bourdon gauge. A compensated Thorpe tube should be used instead.

Integrated O2 cylinders (Figure 37-25), including the Grab ’n Go System (Praxair, Inc, Danbury, Connecticut), have combined the O2 cylinder with a pressure regulator and an adjustable flow restrictor to meter O2 flow. These portable O2 systems eliminate the need for separate O2 tanks, Bourdon gauge regulators, and O2 keys or wrenchs (needed to turn on standard E-cylinders). These integrated systems virtually eliminate problems and delays associated with incorrectly mounted regulators. The practitioner simply selects the flow on the flow-adjusting knob and connects the O2 tubing to the system connection and the patient.

Thorpe Tube

The Thorpe tube flowmeter (Figure 37-26) is always attached to a 50-psig source, either a preset pressure-reducing valve or a bedside station outlet. Compared with the flow restrictor and the Bourdon gauge, the Thorpe tube functions as a variable-orifice, constant-pressure flowmeter, so increasing the size of the orifice increases the gas flow. Figure 37-27 shows how a Thorpe tube works. The key component in this device is a tapered transparent tube that contains a float. The diameter of the tube increases from bottom to top. Gas flow suspends the float against the force of gravity. To read the flow, one simply compares the float position with an adjacent calibrated scale, normally calibrated in liters per minute.

image
FIGURE 37-26 Thorpe tube flowmeter.

Although the Bourdon gauge measures pressure, the Thorpe tube is used to measure true flow. Flow measurement involves the complex interaction of gravity and fluid dynamics. When gas begins to flow into a Thorpe tube, the initial pressure difference lifts the float. As the needle valve is opened, the float rises in the widening tube, the space available for flow around it increases, and resistance to flow decreases. The float ultimately stabilizes when the pressure difference across the float (an upward force) equals the opposing downward force of gravity.

As the needle valve of the flowmeter is opened, the increase in flow initially disrupts this balance, causing an increase in the pressure difference across the float. With the upward pressure difference greater than the downward force of gravity, the float rises. However, as the float rises, the available “orifice” increases in diameter. Flow resistance around the float decreases, and the pressure difference again equilibrates with gravity. The float position stabilizes at a higher level, proportionate to the greater flow around it.

Thorpe tubes come in two basic designs: pressure compensated and pressure uncompensated. The term pressure compensation refers to a design that prevents changes in downstream resistance, or back pressure, from affecting meter accuracy. All manufacturers now supply only pressure-compensated Thorpe tubes for administration of medical gas. However, some ventilators and anesthesia machines still use uncompensated Thorpe tubes. For this reason, clinicians using these devices must understand the effect of back pressure on the accuracy of these devices. Downstream resistance increases when the user connects a flowmeter to certain types of equipment. Almost all therapy gas equipment produces some flow restriction. Devices such as jet nebulizers produce very high downstream resistance. Depending on their design, Thorpe tube flowmeters respond to resistance in one of two ways.

The uncompensated Thorpe tube flowmeter is calibrated in liters per minute but at atmospheric pressure (without restriction). Gas from a 50-psig source flows into the meter at a rate controlled by a needle valve located before the flow tube (Figure 37-28, A). When the user attaches flow-restricting equipment to the meter, downstream resistance increases, increasing pressure in the flow tube. As long as this pressure does not exceed 50 psig, gas continues to flow through the tube. However, the added downstream resistance increases the pressure in the flow tube above atmospheric pressure. At higher pressures, a greater amount of gas flows through a given restriction than at atmospheric pressure. With the float at a given height, more gas flows through the tube than is indicated on the scale. Under these conditions, an uncompensated Thorpe tube falsely shows a flow lower than that actually delivered to the patient.4

In contrast, the scale of the compensated Thorpe tube flowmeter is calibrated at 50 psig instead of at atmospheric pressure. Its flow control needle valve is placed after (distal to) the flow tube (see Figure 37-28, B). The entire meter operates at constant 50-psig pressure. Knowing that the compensated Thorpe tube operates at 50 psig helps identify it. When a compensated Thorpe tube is connected to a 50-psig gas source with the needle valve closed, the float “jumps” and then returns to zero as the Thorpe tube is pressurized. Because the entire meter operates at constant pressure, an increase in downstream resistance increases pressure distal to the needle valve only. As long as the downstream pressure does not exceed 50 psig (in which case flow ceases), the position of the float accurately reflects actual outlet flow. For this reason, the pressure-compensated Thorpe tube is the preferred instrument in most clinical situations.

The only factor limiting the use of a pressure-compensated Thorpe tube is gravity. Because it is accurate only in an upright position, a Thorpe tube is not the ideal choice for patient transport. In these cases, the gravity-independent Bourdon gauge is a satisfactory alternative. Figure 37-29 summarizes the effects of downstream resistance, or back pressure, on the Bourdon gauge and pressure-compensated and uncompensated Thorpe tube flow metering devices.

Summary Checklist

• All therapy gases must contain at least 20% O2; all such gases support combustion.

• Medical gases are stored either in portable high-pressure cylinders or in large centralized bulk reservoirs.

• For positive identification of the contents of a medical gas cylinder, the label must be carefully read.

• The pressure in a gas-filled cylinder indicates its contents; the pressure in a liquid-filled cylinder does not.

• To compute duration of flow (minutes) of a medical gas cylinder, multiply the cylinder pressure (pounds per square inch) by the cylinder factor, and divide the result by the set flow (liters per minute).

• Gas supply systems provide gas at 50 psig to outlets throughout a facility through a network of pipes. Such a system must include both zone valves for repairs or fire and alarms to warn of failure.

• Failure of a bulk gas supply system can threaten the lives of patients receiving O2 therapy or being supported with pneumatically powered devices. A protocol must exist to deal with this emergency.

• Indexed safety systems help prevent misconnections between equipment. The ASSS provides high-pressure connections with large cylinders; the PISS does the same for small cylinders; and DISS connections are for low-pressure outlets, typically 50 psig.

• A reducing valve is used for reduction of gas pressure. A flowmeter is used for control of gas flow. A regulator is used for control of both pressure and flow.

• A flow restrictor is used to provide fixed low flows of O2. A Bourdon gauge is used to meter flow during patient transport. A compensated Thorpe tube is used when accurate flows are needed with high-resistance equipment.