Medical suction apparatus

Published on 27/02/2015 by admin

Filed under Anesthesiology

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 3 (1 votes)

This article have been viewed 13419 times

Chapter 20 Medical suction apparatus

Suction apparatus is vital to safe medical practice, especially in anaesthesia, resuscitation and intensive care. It is used for the clearance of mucus, blood and debris from the pharynx, trachea and main bronchi. During surgery, suction is used to provide a clear operating field for the surgeon. Specially adapted suction apparatus can also be used for other procedures, such as gastrointestinal, wound and pleural drainage.

Main components

The main components of a medical suction system are:

The schematic drawings in Fig. 20.1 outline the methods used in suction apparatus, using internationally agreed symbols.

Vacuum source

The sub-atmospheric pressure required may be generated by:

Pump types

Fig. 20.2A shows a piston pump, which is capable of creating high vacuum but, in transportable models, has a relatively low displacement (i.e. can only sustain low flow rates). Fig. 20.2B shows a diaphragm pump, which is a variation of the piston pump. It is mechanically simpler but is also frequently much noisier. One reason for the increased noise is that often, rather than using a conventional rotating electric motor, a much simpler large electromagnet displaces the diaphragm. Fig. 20.2C shows a form of rotary pump that can produce a high vacuum without conventional one-way valves. Fig. 20.2D shows a rotary pump capable of producing very high flows, as would be required in a dental surgery. It works in the same way as a vacuum cleaner and has the disadvantage of being extremely noisy. Fig. 20.2E shows a pump that works on the principle of the Archimedean screw. This type of pump can produce a high vacuum for a comparatively small size of machine. Fig. 20.2F shows the principle of a Venturi pump which makes use of the Bernoulli effect. Compressed fluid (gas or liquid), passing through a narrow orifice, creates a region of negative pressure beyond that orifice, which can be used to entrain adjacent air/debris. The main disadvantage of this simple affair is that it uses and is thus wasteful of large volumes of driving fluid, which is usually oxygen from cylinders. However, it does have the virtue of being extremely portable (Fig. 20.3). Finally, Fig. 20.2G shows a simple bellows mechanism with a pair of one-way valves, as would be used in manually operated suction apparatus.

The pump may be:

There are international standards for each type of suction apparatus and these are listed at the end of this chapter.

Other components of suction apparatus

The suction nozzle, catheter or hand-piece

The design of what is referred to in International Standards as the ‘applied part’ depends upon the application. The commonest examples of hand-held suction nozzles are shown in Fig. 20.5. The key requirement of any suction ‘applied part’ is that the smallest internal diameter is at the very tip. If there are smaller diameters between the tip and the collection vessel, then blockages are likely to occur. The shape of the tip should be smooth so as to prevent damage to delicate tissues. The practice of allowing the tip to be occluded by any tissue, to reduce the noise of suction when not actually aspirating, must be deprecated as it causes tissue damage. Similarly, the noise of suction when using transportable, electrically powered suction machines should be reduced by switching off the motor rather than by occluding the suction tubing, as this may overload the motor. Hand-held suction nozzles of the Yankauer type (Fig. 20.5) often have a hole on the handle for fine control of suction with a finger.

Bronchial suction catheters should have smooth tips to avoid damaging delicate tissues and usually have several holes around the tip.

To protect staff, closed suction catheter systems are now commonly used (Fig. 20.6). The catheter is enclosed in a flexible sheath, which is permanently attached to a special 15 mm-taper adapter, which is left in circuit between the catheter mount and the endotracheal or tracheostomy tube. These systems are replaced every 24 h or as specified by the manufacturer.

Local vacuum units

Piped vacuum systems are now installed in most hospitals. The ‘behind the wall’ equipment and terminal outlets are described in Chapter 1. Medical suction apparatus may be connected into this system, wherever there is a terminal outlet.

There are two main types of ‘local’ apparatus:

1. Free-standing floor units, mains electricity powered and often with two collection vessels used for surgical purposes in the operating theatre (Fig. 20.7). Small, ‘low-suction’, low displacement units are also available for bedside use for intracavitary and continuous wound drainage. Generating typically 5–50 mmHg sub-atmospheric pressure, these devices are usually mains and rechargeable battery-powered piston pumps.

2. Wall-mounted units are local suction controllers for connection to central piped vacuum source (see Chapter 1). These have a single collection vessel as shown in Fig. 20.8. There are two types of controller: conventional ‘high-suction’ and also ‘low-suction’. Low-suction controllers are deliberately limited to provide safe suction for intra-pleural drainage or nasogastric suction. Confusion between the two types of suction controller can have disastrous consequences, hence the resurgence of free-standing ‘low-suction’ units.

Choice of suction apparatus

When selecting a suction apparatus for a particular purpose, the following points need to be considered:

• Must it be portable? If so, should it be hand/foot operated (Fig. 20.9)? If not, should it be powered by electricity (mains or battery) or is pipe-line vacuum available? Could it be powered from a gas cylinder using a Venturi injector?

• Is a high displacement needed?

• Is a high vacuum needed?

• What size should the collection vessel be?

It is important to ascertain, for an electrically driven suction machine, whether it is rated for continuous or intermittent use.

High-volume aspirators, as used in dental surgery, use a pump similar to that used in a vacuum cleaner (Fig. 20.10A). They usually have several suction tubes of different diameters for different applications (Fig. 20.10B). These tubes and their nozzles should never be obstructed, as a high flow of air is required to cool the motor, which would otherwise overheat. This type of high-flow suction is needed in dental surgery to aspirate the spray of water used to cool the tooth during high-speed drilling.