Mechanical Ventilators

Published on 01/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 3 (1 votes)

This article have been viewed 7574 times

Mechanical Ventilators

Robert L. Chatburn and Teresa A. Volsko

To initiate and manage a mechanical ventilator safely and effectively, the respiratory therapist (RT) must thoroughly understand (1) ventilator design, classification, and operation; (2) appropriate clinical application of ventilatory modes (i.e., the proper matching of ventilator capability with physiologic need); and (3) the physiologic effects of mechanical ventilation, including gas exchange and pulmonary mechanics. This chapter focuses on the first and second of these. This chapter explains classification terminology and outlines a framework for understanding current and future ventilatory support devices.13 For the application of ventilators, the specific indications and clinical use of the various modes of full and partial ventilatory support are outlined.

How Ventilators Work

To understand how ventilators work, one must have some knowledge of basic mechanics. A ventilator is simply a machine, which is a system designed to alter, transmit, and direct applied energy in a predetermined manner to perform useful work.4 Ventilators are provided with energy in the form of either electricity or compressed gas. The energy is transmitted or transformed (by the drive mechanism of the ventilator) in a predetermined manner (by the control circuit) to augment or replace the patient’s muscles in performing the work of breathing (the desired output). To understand mechanical ventilators, the following four basic functions of ventilators must be understood:

This simple outline format can be expanded to add as much detail about a given ventilator as desired.

Input Power

The power source for a ventilator is either electrical energy (energy = volts × amperes × time) or compressed gas (energy = pressure × volume).5

Pneumatic Power

A pneumatically powered ventilator uses compressed gas as its power source. Most modern intensive care unit (ICU) ventilators are pneumatically powered. Ventilators powered by compressed gas usually have internal pressure-reducing valves so that the normal operating pressure is lower than the source pressure. Ventilators can operate without interruption from hospital-piped gas sources, which are usually regulated to 50 psi (pounds per square inch) but are subject to periodic fluctuations.

Most pneumatically powered ICU ventilators still require electrical power to support their control functions (see the following section on control mechanisms). However, a few pneumatically powered ventilators can function without electrical power. These devices are ideal in situations where electrical power is unavailable (e.g., during certain types of patient transport) or as a backup to electrically powered ventilators in case of power failures. They are also particularly useful where electrical power is undesirable, such as near magnetic resonance imaging (MRI) equipment.

Power Transmission and Conversion

The power transmission and conversion system consists of the drive and output control mechanisms. The drive mechanism generates the actual force needed to deliver gas under pressure. The output control consists of one or more valves that regulate gas flow to the patient.

Control System

Knowledge of the mechanics of breathing provides a good foundation for understanding how ventilators work. Specifically, the pressure needed to drive gas into the airway and inflate the lungs is important. The formula that relates these variables is known as the equation of motion for the respiratory system (Figure 42-1):7

< ?xml:namespace prefix = "mml" />PVENT+PMUS=(E×V)+(R×V˙)+auto-PEEP Equation 42-1, A

image Equation 42-1, A

where PVENT is the pressure generated by the ventilator above positive end expiratory pressure (PEEP), PMUS is the pressure generated by the ventilatory muscles to expand the lungs and chest wall, E is respiratory system elastance, V is the change in lung volume above functional residual capacity (FRC), R is respiratory system resistance, and image is flow (usually zero unless auto-PEEP is present). Auto-PEEP is the difference between end expiratory airway pressure and end expiratory lung pressure. The combined muscle and ventilator pressures cause gas to flow into the lungs. Elastance (elastance = Δpressure/Δvolume) and resistance (resistance = Δpressure/Δflow) together constitute the impedance or load against which the muscles and ventilator do work. The equation of motion is sometimes expressed in terms of compliance (compliance = Δvolume/Δpressure) instead of elastance.

The auto-PEEP term in the equation of motion indicates that if auto-PEEP is present, more force is required by the ventilator or the muscles or both to generate a given tidal volume and flow. Under passive conditions (PMUS = 0) for volume control ventilation, where the terms E × V and E × image are constant because of the preset values of tidal volume and inspiratory flow, as auto-PEEP increases, peak inspiratory pressure increases. For pressure control ventilation, the airway pressure waveform is constant because of the preset value of inspiratory pressure. As auto-PEEP increases, both inspired tidal volume and peak inspiratory flow decrease. If inspiration is unassisted (PVENT = 0), the equation shows that PMUS must exceed auto-PEEP before inspiratory flow can begin. If the patient is connected to a ventilator, PMUS must be greater than auto-PEEP to trigger an assisted breath. In these cases, auto-PEEP increases the patient’s work of breathing. However, auto-PEEP is not always undesirable, as in the application of airway pressure release ventilation where expiratory time is intentionally shortened to create gas trapping in the absence of a set PEEP. Auto-PEEP is also present as an unavoidable and perhaps desirable factor during high-frequency ventilation.

Pressure, volume, and flow all are changeable variables measured relative to their baseline or end expiratory values. When pressure, volume, and flow are plotted as functions of time, characteristic waveforms for volume-controlled ventilation and pressure-controlled ventilation are produced (Figure 42-2). The conventional order of presentation is pressure, volume, and flow from top to bottom. Convention also dictates that positive flow values (above the horizontal axis) correspond to inspiration and that negative flow values (below the horizontal axis) correspond to expiration. The vertical axes are in units of the measured variables (usually cm H2O for pressure, L or ml for volume, and L/min or L/sec for flow). The horizontal axis of these graphs is time. Many ventilators have monitors that display pressure, volume, and flow waveforms, providing the clinician with information to evaluate ventilator-patient interaction.

Figure 42-2 shows that the expiratory lung pressure curves are the same shape for both volume and pressure control. This shape is called an exponential decay waveform (often called a decelerating waveform), and it is characteristic of passive emptying of the lungs (exhalation). Solving the equation of motion for lung pressure (assuming no auto-PEEP) provides the following expression:

PL=VTCet/RC Equation 42-2

image Equation 42-2

where PL is lung pressure during passive exhalation, VT is tidal volume, e is the base of the natural logarithms (approximately 2.72), t is time (in this case, the time allowed for exhalation), R is respiratory system resistance, and C is respiratory system compliance. The product of R and C has units of time and is called the time constant. It is referred to as a “constant” because after a period of time equal to RC, the lung pressure changes by 63%. In the next period equal to RC, lung pressure changes another 63% and so on to infinity. When the expiratory time is equal to the time constant, the patient will have passively exhaled 63% of his or her tidal volume. This can be shown using Equation 42-2 by setting t equal to RC:

PL=VTCeRC/RC=VTC2.721=VTC0.37 Equation 42-3

image Equation 42-3

This expression shows that after an expiratory time equal to one time constant, only 37% of the lung pressure is left. Because volume is equal to pressure multiplied by compliance, multiplication of both sides of Equation 42-3 by compliance shows that only 37% of the tidal volume remains in the lungs. The implication is that 63% of the tidal volume has been exhaled. After two time constants (i.e., t = 2RC), exhalation is 86% completed, and after three time constants, exhalation is 95% complete. After five time constants, exhalation is considered to be 100% complete for practical purposes (Figure 42-3).

A similar expression can be derived for passive inhalation:

PL=VTC(1et/RC) Equation 42-4

image Equation 42-4

When this equation is graphed, it looks like the curved line showing lung pressure and volume during inhalation for pressure-controlled ventilation in Figure 42-2. These same equations govern the passive flow curves for volume-controlled and pressure-controlled ventilation. It is important to understand the concept of time constants to make appropriate ventilator setting adjustments. In any mode of ventilation, the expiratory time should be at least three time constants long to avoid clinically important gas trapping. Similarly, in pressure-controlled modes, inspiratory time (for passive inspiration) should be at least five time constants long to get the maximum tidal volume from the set pressure gradient (i.e., peak inspiratory pressure [PIP] − end expiratory pressure).

Control Circuit

To manipulate pressure, volume, and flow, a ventilator must have a control circuit. A control circuit is a system of components that measures and directs the output of the ventilator to replace or assist the breathing efforts of the patient. A ventilator control circuit may include mechanical, pneumatic, electrical, electronic, or fluidic components. Most modern ventilators combine two or more of these subsystems to provide user control.

Mechanical control circuits use devices such as levers, pulleys, and cams. These types of circuits were used in the early manually operated ventilators illustrated in history books.8 Pneumatic control is provided using gas-powered pressure regulators, needle valves, jet entrainment devices, and balloon-valves. Some transport ventilators use pneumatic control systems.

Electrical control circuits use only simple switches, rheostats (or potentiometers), and magnets to control ventilator operation. Electronic control circuits use devices such as resistors, capacitors, diodes, and transistors and combinations of these components in the form of integrated circuits. The most sophisticated electronic systems use preprogrammed microprocessors to control ventilator function.

Fluidic logic-controlled ventilators, such as the Bio-Med MVP-10 (Bio-Med Devices, Stanford, Connecticut) and Sechrist IV-100B (Sechrist, Anaheim, California), also use pressurized gas to regulate the parameters of ventilation. However, instead of simple pressurized valves and timers, these ventilators use fluidic logic circuits that function similar to electrical circuit boards.9 Fluidic control mechanisms have no moving parts. In addition, fluidic circuits are immune to failure from surrounding electromagnetic interference, as can occur around MRI equipment.

Control Variables

A control variable is the primary variable that the ventilator manipulates to cause inspiration. There are only three explicit variables in the equation of motion that a ventilator can control: pressure, volume, and flow. Because only one of these variables can be the independent variable, the others are dependent variables. In other words, only one variable can be directly controlled at a time, and a ventilator must function as a pressure, volume, or flow controller. Time is implicit in the equation of motion and in some cases serves as a control variable. Figure 42-4 illustrates the criteria for determining what variable the ventilator controls at any given time.

Figure 42-5 illustrates the important variables for volume-controlled modes. It shows that the primary variable we wish to control is the patient’s minute ventilation. A particular ventilator may allow the operator to set minute ventilation directly. More frequently, minute ventilation is adjusted by means of a set tidal volume and frequency. Tidal volume is a function of the set inspiratory flow and the set inspiratory time. Inspiratory time is affected by the set frequency and, if applicable, the set inspiratory-to-expiratory (I : E) ratio. The mathematical relationships among all these variables are shown in Table 42-1.

TABLE 42-1

Equations Relating the Important Parameters for Volume-Controlled and Pressure-Controlled Ventilation

Mode Parameter Symbol Equation
Volume-controlled Tidal volume (L) VT image
image
Mean inspiratory flow (L/min) image image
image
Pressure-controlled Tidal volume (L) VT VT = ΔP × C × (1 − e−t/τ)
Instantaneous inspiratory flow (L/min) image image
Pressure gradient (cm H2O) ΔP ΔP = PIP − PEEP
Both modes Exhaled minute ventilation (L/min) image image
Total cycle time or ventilatory period (seconds) TCT TCT = TI + TE = 60 ÷ f
I : E ratio I : E image
Time constant (seconds) τ τ = R × C
Resistance (cm H2O/L/sec) R image
Compliance (L/cm H2O) C image
Elastance E image
Mean airway pressure (cm H2O) image image
Primary variables Pressure (cm H2O) P  
Volume (L) V  
Flow (cm H2O/L/sec) image  
Time (sec) τ  
Inspiratory time (sec) TI  
Expiratory time (sec) TE  
Frequency (breaths/min) f  
Base of natural logarithm (≈2.72) e  

image

With pressure-controlled modes, the goal is also to maintain adequate minute ventilation. However (as the equation of motion shows), when pressure is controlled, tidal volume and minute ventilation are determined not only by the ventilator’s pressure settings but also by the elastance and resistance of the patient’s respiratory system. This additional variable makes minute ventilation (and gas exchange) less stable in pressure-controlled modes than in volume-controlled modes. Figure 42-6 shows the important variables for pressure-controlled modes. Tidal volume is not set on the ventilator. It is the result of the pressure settings and the patient’s lung mechanics and the inspiratory time. On some ventilators, the speed with which the PIP is achieved (i.e., the pressure rise time) is adjustable. That adjustment affects the shape of the pressure waveform and the mean airway pressure. Mean airway pressure is important because, within reasonable limits, as the mean airway pressure increases, arterial O2 tension increases. Mean airway pressure is higher for pressure-controlled modes than for volume-controlled modes (at the same tidal volume) owing to the differences in the shapes of the airway pressure waveforms.

image
FIGURE 42-6 Influence diagram for pressure-controlled ventilation. Variables are connected by straight lines such that if any two are known, the third can be calculated (see Table 42-2). Arrows represent relationships that are more complex. Purple circles represent variables that are directly controlled by ventilator settings. Gray circles show indirectly controlled variables.

Phase Variables

A complete ventilatory cycle or breath consists of four phases: (1) the initiation of inspiration, (2) inspiration itself, (3) the end of inspiration, and (4) expiration. To understand a breath cycle, the clinician must know how the ventilator starts, sustains, and stops inspiration and must know what occurs between breaths.

The phase variable is a variable that is measured and used by the ventilator to initiate some phase of the breath cycle. The variable causing a breath to begin is the trigger variable. The variable limiting the magnitude of any parameter during inspiration is the target variable. The variable causing a breath to end is the cycle variable. To describe what happens during expiration, the baseline variable that is in effect must be known. Figure 42-7 shows the criteria for determining phase variables.

Trigger Variable

On most modern ventilators, either the machine or the patient can initiate a breath. If the machine initiates the breath, the trigger variable is time. If the patient initiates the breath, pressure, flow, or volume may serve as the trigger variable. Manual or operator-initiated triggering is also available on most ventilators.

Time Triggering

When triggering by time, a ventilator initiates a breath according to a predetermined time interval, without regard to patient effort. In the past, time triggering was the only method available to initiate a ventilator cycle. At the present time, time triggering is most commonly seen when using the intermittent mandatory ventilation (IMV) mode.

Specific systems for setting a breathing rate vary from ventilator to ventilator. A rate control is the most common approach, which divides each minute into equal time segments, allotting one time segment for each full breath. When a rate control is used, inspiratory and expiratory times vary according to other control settings, such as flow and volume. An alternative approach is to provide separate timers for inspiration and expiration. Changing either or both of these timers alters the breathing rate.

Pressure Triggering

Pressure triggering occurs when a patient’s inspiratory effort causes a decrease in pressure within the breathing circuit. When this pressure decrease reaches the pressure-sensing mechanism, the ventilator starts gas delivery. On most ventilators, the pressure decrease needed to trigger a breath can be adjusted. The trigger level is often called the sensitivity. Typically, the trigger level is set 0.5 to 1.5 cm H2O below the baseline expiratory pressure. Setting the trigger level to a higher number, such as 3 cm H2O, makes the ventilator less sensitive and requires the patient to work harder to initiate inspiration. Conversely, setting the trigger level lower makes the ventilator more sensitive to patient effort. How fast the ventilator mechanism responds to patient effort is called the response time. It is important for the ventilator to have a short response time to maintain optimal synchrony with the patient’s inspiratory efforts. Either a large pressure decrease (i.e., low sensitivity setting) or a delay in flow delivery can increase a patient’s work of breathing.

Flow Triggering

Using flow as the trigger variable is more complex. A ventilator that uses flow triggering typically provides a continuous low flow of gas through its circuit. The ventilator measures the flow coming out of the main flow control valve and the flow through the exhalation valve. Between breaths, these two flows are equal (assuming no leaks in the patient circuit). When the patient makes an inspiratory effort, the flow through the exhalation valve falls below the flow from the output valve. The difference between these two flows is the flow trigger variable.

Flow Trigger Variable

To adjust the sensitivity of a flow-triggered system, the clinician usually sets both a base continuous flow and a trigger flow level. Typically, the trigger flow level is set to 1 to 3 L/min (below baseline). If the base continuous flow is set at 10 L/min and the trigger is set at 2 L/min, the ventilator triggers a breath when the output flow decreases to 8 L/min or less. An alternative approach used with some ventilators is simply to measure the flow at the “wye” connector and start breath delivery on that signal.

Compared with pressure, using flow as the trigger variable decreases a patient’s work of breathing.11,12 However, ventilators that use a flow-triggering mechanism tend to be highly susceptible to circuit leaks or movement caused by turbulence or gas flow through condensed water. Either of these conditions can cause spurious breaths, which can disrupt patient-ventilatory synchrony and increase the work of breathing. The perception exists, with regard to the most recent generation of ICU ventilators, that pressure and flow triggering are equally effective. Using volume as the trigger can help overcome synchrony problems caused by circuit leaks, but at the present time only the Draeger Babylog (Draeger Medical, Telford, Pennsylvania) uses true volume triggering.

Other Trigger Variables

Other variables can be used to trigger inspiration, such as a decrease in total minute ventilation below a preset threshold (e.g., mandatory minute ventilation mode on the Draeger Evita XL ventilator) or the electrical signal generated by diaphragmatic contraction (e.g., neurally adjusted ventilatory assist mode on the Maquet SERVO-i ventilator).

Target Variable

A target variable is one that can reach and maintain a preset level before inspiration ends but does not terminate inspiration. Pressure, flow, or volume can serve as a target variable.

Clinicians often confuse target variables with cycle variables. A cycle variable always ends inspiration. A target variable does not terminate inspiration—it sets an upper bound only for pressure, volume, or flow. Figure 42-8 illustrates the importance of distinguishing between target and cycle variables.

Cycle Variable

The inspiratory phase always ends when some variable reaches a preset value. The variable that is measured and used to end inspiration is called the cycle variable. The cycle variable can be pressure, volume, flow, or time. Manual cycling is also available on some modern ventilators.

Volume Cycling

When a ventilator is set to volume cycle, it delivers flow until a preselected volume has been expelled from the device. As soon as the set volume is met, inspiratory flow stops, and expiratory flow begins. The volume that passes through the ventilator’s output control valve is never exactly equal to the volume delivered to the patient because of the volume compressed in the patient circuit. Some ventilators use a sensor at the “wye” connector (e.g., the Hamilton Galileo, Hamilton Medical, Reno, NV) for accurate tidal volume measurement. Others measure volume at some point inside the ventilator and calculate the tubing compliance during operational checks, before connecting the patient to the ventilator circuit. It is imperative for the operator to know whether the ventilator compensates for compressed gas in its tidal volume readout.

Time Cycling

Time cycling means that expiratory flow starts because a preset time interval has elapsed. There are several time intervals of interest during inspiration. One is the inspiratory flow time. As the name implies, this is the time during which inspiratory flow is delivered to the patient. Another interval is the inspiratory hold time, during which inspiratory flow has ceased but expiratory flow is not yet allowed. The sum of these two intervals is the inspiratory time. Time cycling occurs when the inspiratory time has elapsed. An inspiratory hold time may not be used. If it is used, it may be set directly, or it may occur indirectly if the set inspiratory time is longer than the inspiratory flow time (determined by the set tidal volume and flow; time = volume/flow).

Mini Clini

Calculate Inspiratory Hold Time Given Set Inspiratory Time, Tidal Volume, and Flow When Using the Siemens Servo

Discussion

1. Calculate the inspiratory flow time (TIF) using appropriate unit conversions:

TIF=tidalvolumeinspiratoryflow=500ml60L/minute×1L1000ml×60seconds1minute=0.5L1L/second=0.5second

image

2. Compare the set inspiratory time (0.8 second) with the flow time resulting from the tidal volume and flow settings (0.5 second). Inspiratory time lasts longer than inspiratory flow. Because inspiration is time cycled, this means that there is an inspiratory hold of duration equal to 0.8 − 0.5 = 0.3 second.

3. You could eliminate the inspiratory hold either by decreasing the inspiratory flow or by decreasing the inspiratory time. The goal is to minimize the mean inspiratory pressure. You choose to decrease inspiratory time for two reasons: (1) It decreases the I : E ratio and may allow more time for spontaneous breaths to occur, lowering mean intrathoracic pressure; (2) decreasing inspiratory flow may make tidal volume delivery slower than the patient demands, decreasing patient-ventilator synchrony.

Patient versus Machine Triggering and Cycling

Trigger and cycle signals can be grouped into two categories: patient-generated and machine-generated. The determination of whether the patient or the machine generated the signal is based on the equation of motion. However, this time the equation is expressed in another way:

Pinsp=PE+PR Equation 42-5

image Equation 42-5

where Pinsp reflects the pressure required for inspiration, PE is the pressure resisting inspiration by the elastance of the respiratory system, and PR is the pressure resisting the inspiration by the resistance of the respiratory system. This equation can be expanded as follows:

Pvent+Pmus=(PE,vent+PR,vent)+(PR,mus+PR,mus) Equation 42-6

image Equation 42-6

where Pvent is the pressure generated by the ventilator, independent of the patient, Pmus is the pressure generated by the patient’s ventilatory muscles independent of the ventilator, PE,vent is the elastic load supported by the ventilator, PE,mus is the elastic load supported by the patient’s muscles, PR,vent is the resistive load supported by the ventilator, and PR,mus is the resistive load supported by the patient’s muscles. In this form of the equation, we can easily differentiate sources of patient versus machine trigger and cycle signals.

For the definition of patient trigger and cycle signals, the patient’s ventilatory efforts are represented by Pmus, PE,mus, and PR,mus. Because PE = EV and PR = RV, it follows that patient trigger and cycle signals are based on three variables (Pmus, Vmus, and image) and two parameters (E and R). The ventilator can be triggered or cycled by a signal representing Pmus (e.g., the electrical signal of the diaphragm as with neurally adjusted ventilatory assist or a calculated estimate of Pmus) or the volume or flow generated by the patient’s ventilatory muscles (Vmus and image [e.g., volume triggering in the Draeger Babylog or flow triggering on many other ventilators]). The ventilator can be triggered or cycled by the patient’s passive respiratory system mechanics (E and R). Examples are flow triggering with the AutoRelease feature on the Draeger Evita Infinity V500 ventilator and flow cycling during pressure support on any ventilator. Manufacturers may have specific marketing terms for flow cycling. For example, flow cycling on the Puritan Bennett 840 ventilator (Pleasanton, California) is termed E-sensitivity. Given the definition of patient trigger and cycle signals, we can define machine trigger and cycle signals as being anything else that starts or ends inspiration (excluding operator-generated manual signals).

Baseline Variable

The baseline variable is the parameter controlled during expiration. Although pressure, volume, or flow could serve as the baseline variable, pressure control is the most practical and is implemented by all modern ventilators.

Baseline or expiratory pressure is always measured and set relative to atmospheric pressure. For baseline pressure to equal atmospheric pressure, it is at zero. For baseline pressure to exceed atmospheric pressure, it is set at a positive value, called positive end expiratory pressure (PEEP). Although seldom used, the baseline pressure could be set below atmospheric pressure, a technique called negative end expiratory pressure (NEEP).

Zero end expiratory pressure (ZEEP) is the default baseline value during positive pressure ventilation, meaning that it is normally in effect unless purposely changed. Regardless of the mechanism by which gas is delivered to the lungs, it must leave before the next inspiration. Exhalation normally occurs by virtue of the stored pressure in the expanded lungs and thorax. With ZEEP in effect, when exhalation begins, the ventilator’s expiratory valve simply opens to the atmosphere, exposing the patient’s airway to a relative pressure of zero. At this point, alveolar pressure exceeds airway pressure, gas moves from alveoli out to the atmosphere, and the lungs and thorax passively recoil down to their resting volume, or functional residual capacity (FRC).

PEEP is the application of pressure above atmospheric pressure at the airway throughout expiration. PEEP elevates a patient’s FRC and can help improve oxygenation by preventing collapse of alveolar units that are made unstable by lack of surfactant or disease.

NEEP is the application of subatmospheric pressure to the airway during expiration. NEEP was originally developed to overcome the harmful cardiovascular effects of positive pressure ventilation. The assumption was that NEEP could offset the impedance to venous return created by the positive pressure during inspiration. NEEP has occasionally been promoted as a way to help patients overcome expiratory airway resistance. In this approach, negative pressure is applied only to help return airway pressure to baseline. Because NEEP can cause airway collapse and decrease the FRC if misapplied, great care must be taken in applying this technique.

Modes of Ventilation

The objective of mechanical ventilation is to ensure that the patient receives the minute volume of appropriate gases required to satisfy respiratory needs, while not damaging the lungs, impairing circulation, or increasing the patient’s discomfort. Mode of ventilation is the manner in which a ventilator achieves this objective. A mode can be identified by specifying a combination of the following:

An example of how this mode classification system can be applied is shown in Table 42-2. The key to understanding modes of ventilation is to link simple, defined terms in a way that allows descriptions of varying complexity to be built; this is much more practical than trying to memorize arbitrary names for every new feature a manufacturer wishes to promote. It is analogous to using an alphabet of several dozen letters to build words rather than memorizing tens of thousands of separate ideographs that each represents a word. One need only compare the English written language with the Chinese written language to appreciate the analogy.3

TABLE 42-2

Specifications for Some Modes Found on the Draeger Evita XL Ventilator

Draeger Mode Name Breathing Pattern Mandatory Breaths* Spontaneous Breaths
Targeting Scheme Trigger Target Cycle Targeting Scheme Trigger Target Cycle
CMV VC-CMV Set-point T F, V T NA NA NA NA
Operational Logic: Every breath is volume controlled and mandatory. Every breath is machine triggered and cycled
CMV + AutoFlow PC-CMV Adaptive T, F P T NA NA NA NA
Operational Logic: Mandatory breaths are pressure controlled, but patient may trigger breath. If target tidal volume is not met, pressure target is automatically adjusted
CMV + Pressure Limited Ventilation PC-CMV Dual T, F F, V, P T NA NA NA NA
Operational Logic: Mandatory breath starts out in volume control but switches to pressure control if airway pressure reaches set Pmax
SIMV VC-IMV Set-point T, F F, V T Set-point P P P
Operational Logic: Mandatory breaths are volume-controlled. Spontaneous breaths may occur within window determined by set rate and are not assisted (i.e., inspiratory pressure stays at baseline)
PC+ PC-IMV Set-point T, F P T Set-point F P F
Operational Logic: Mandatory breaths are pressure-controlled. Spontaneous breaths may occur within window determined by set rate and are not assisted (i.e., inspiratory pressure stays at baseline)
SIMV + AutoFlow PC-IMV Adaptive T, F P T Set-point P P P
Operational Logic: Mandatory breaths are pressure-controlled and pressure target is automatically adjusted if target tidal volume is not met. Spontaneous breaths may occur with window determined by set rate and are not assisted (i.e., inspiratory pressure stays at baseline)
CPAP PC-CSV NA NA NA NA Set-point P P P
Operational Logic: Spontaneous breaths are unassisted
Pressure Support PC-CSV NA NA NA NA Set-point F P F
Operational Logic: Spontaneous breaths are assisted (i.e., inspiratory pressure increases above baseline)
SmartCare PC-CSV NA NA NA NA Intelligent F P F
Operational Logic: Spontaneous breaths are assisted (i.e., inspiratory pressure increases above baseline). Pressure support level automatically adjusted by rule-based expert system

image

CMV, Continuous mandatory ventilation (all breaths are mandatory); CSV, continuous spontaneous ventilation (all breaths are spontaneous); F, flow; IMV, intermittent mandatory ventilation (spontaneous breaths between mandatory breaths); NA, not available; P, pressure; PC, pressure-controlled; T, time; V, volume; VC, volume-controlled.

*Patient can take spontaneous breaths during mandatory breaths with PC and AutoFlow but not with Pressure Limited Ventilation.

Flow trigger may be turned off. When off, mandatory breaths cannot be triggered, but spontaneous breaths are automatically pressure triggered with factory set sensitivity.

Volume target achieved if inspiratory time set longer than tidal volume/flow.

Control Variable

In the equation of motion, the control variable is the variable that is predetermined for a given inspiration (i.e., pressure or volume). Predetermined means that the operator presets the parameters of the variable’s waveform independent of the patient’s mechanics. In volume control, the operator may preset the inspiratory flow and tidal volume. In pressure control, the operator may preset the inspiratory pressure and inspiratory time. In a more exotic form of pressure control, such as proportional assist ventilation, the operator presets the parameters of elastance and resistance to be supported, and the ventilator delivers pressure according to the equation of motion.

There are clinical advantages and disadvantages to volume versus pressure control, which are discussed in Chapter 43. Briefly, volume control results in a more stable minute ventilation (and more stable blood gases) than pressure control if lung mechanics are unstable. Pressure control allows better patient-ventilatory synchrony with the patient because inspiratory flow is not constrained to a preset value. Although it is possible to control only one variable at a time, a ventilator can automatically switch between pressure control and volume control in an attempt to guarantee minute ventilation while maximizing patient synchrony. This feature is called dual control13 and is discussed in the section on Targeting Schemes.

Volume Control

If the ventilator controls volume, the volume and flow waveforms remain consistent, but pressure varies with changes in respiratory mechanics. To qualify as a true volume controller, a ventilator must measure volume and use this signal to control the volume waveform. Volume can be controlled directly by the displacement of a device such as a piston or bellows. Volume can be controlled indirectly by controlling flow; this follows from the fact that volume and flow are inverse functions of time (i.e., volume is the integral of flow, and flow is the derivative of volume).

If the ventilator controls flow, the flow and volume waveforms remain consistent, but pressure varies with changes in respiratory mechanics. Flow can be controlled directly using something as simple as a flow meter or as complex as a proportional solenoid valve. Flow can be controlled indirectly by controlling volume.

Infant ventilators, such as the Sechrist, are the simplest examples of flow controllers. In this ventilator, flow is controlled directly by a flowmeter and an exhalation valve. As long as the airway pressure does not reach the set pressure limit, the resulting volume waveform remains constant.10 Current-generation adult ICU ventilators typically function as flow controllers. However, these systems are much more complex than ventilators such as the Sechrist; flow is measured and adjusted hundreds of times per second through sophisticated computerized output control valves.

For simplicity in classifying modes of ventilation, we consider flow control to be volume control. It has been observed that direct control of flow is indirect control of volume because volume is the mathematical inverse of flow (i.e., volume is the integral of flow with respect to time).

Breath Sequence

Specifying only the control variable for a mode, the clinician can distinguish only among pressure and volume control modes; this is often all that is needed to communicate. At the bedside, the clinician might simply have to indicate that the patient has become asynchronous with the ventilator, and the mode has been changed from “volume control” to “pressure control.”

The second component of the mode classification scheme is the breath sequence. A breath is defined as a positive change in airway flow (inspiration) paired with a negative change in airway flow (expiration), both relative to baseline flow and associated with ventilation of the lungs. This definition excludes flow changes caused by hiccups or cardiogenic oscillations, but it allows the superimposition of a spontaneous breath on a mandatory breath or vice versa (these breath types are defined subsequently). Traditionally, the flow baseline is taken as flow = 0 L/min (i.e., end expiration). However, because “breaths” can be superimposed on existing flow in various circumstances (e.g., high-frequency ventilation), the inspiratory and expiratory movements of gas must be judged relative to the level of flow existing when these movements occur.

Typically, expiration immediately follows inspiration. However, with some modes, such as airway pressure release ventilation, there can be a large inspiration, followed by several small inspirations and expirations, followed by a large expiration.

The classification of modes requires the definition of two basic categories of breaths: spontaneous and mandatory. A spontaneous breath is a breath for which the start and end of inspiration may be determined by the patient, independent of any machine settings for inspiratory time and expiratory time. In other words, the patient both triggers and cycles the breath. As a consequence, the patient retains substantial, if not complete, control over the timing (frequency and inspiratory time) and size (tidal volume) of the breath. A spontaneous breath may occur during a mandatory breath (Figure 42-9). A spontaneous breath may be assisted or unassisted. An assisted breath is a breath during which all or part of inspiratory or expiratory flow is generated by a change in transrespiratory pressure (i.e., airway pressure minus body surface pressure) owing to an external agent (e.g., manual or automatic resuscitator, mechanical ventilator).

A mandatory breath is a breath for which the start or end of inspiration (or both) is determined by the ventilator, independent of the patient; the machine triggers or cycles the breath. As a consequence, the patient substantially, if not completely, loses control over the timing (frequency and inspiratory time) and size (tidal volume) of the breath. Mandatory breaths are generally assisted. Box 42-1 shows an algorithm defining spontaneous and mandatory breaths.

Having defined spontaneous and mandatory breaths, there are three possible sequences of breaths, designated as follows:

The abbreviation CMV has been used to mean a variety of things by ventilator manufacturers. The most logical use in this classification system is to represent continuous mandatory ventilation as part of a continuum from full ventilatory support to unassisted breathing. The abbreviation IMV has a long history of consistent use to mean intermittent mandatory ventilation (i.e., a combination of mandatory and spontaneous breaths). However, the development of the active exhalation valve and other innovations has made it possible for the patient to breathe spontaneously during a mandatory breath. This is primarily a feature used to help ensure synchrony between the ventilator and patient in the event that the mandatory breath parameters (e.g., preset inspiratory time, pressure, volume, or flow) do not match the patient’s inspiratory demands. This feature blurs the historical distinction between CMV and IMV.

The key difference now between CMV and IMV is that with CMV, the clinical intent is to make every inspiration a mandatory breath, whereas with IMV, the clinical intent is to partition ventilatory support between mandatory and spontaneous breaths. This means that during CMV, if the operator decreases the ventilatory rate (often considered to be a safety “backup” rate in the event of apnea), the level of ventilatory support is unaffected, provided that the patient continues triggering mandatory breaths at the same rate (i.e., each breath is assisted to the same degree). With IMV, the rate setting directly affects the number of mandatory breaths and the level of ventilatory support (assuming that spontaneous breaths are not assisted to the same degree as mandatory breaths). CMV is normally considered a method of “full” ventilatory support, whereas IMV is usually viewed as a method of partial ventilatory support. For classification purposes, if spontaneous breaths are not allowed between mandatory breaths, the breath sequence is CMV; otherwise, the sequence is IMV (Figure 42-10). Given that almost every ventilator has a mechanism to synchronize breath delivery with patient effort, it is no longer necessary to add an S to designate synchronized IMV (SIMV). The term SIMV was important in the early days of mechanical ventilation but is an anachronism now. Patient triggering can be specified in the description of mode phase variables.

There has been no consistent abbreviation to signify a breathing pattern composed of all spontaneous breaths. However, the logical progression would be from CMV to IMV to CSV.

When a breath sequence is added to the control variable in classifying a mode, the result is a breathing pattern. A breathing pattern provides a greater ability to discriminate between similar modes. For example, it is possible to distinguish between pressure-control IMV (PC-IMV) and pressure-control CSV (PC-CSV). By classifying modes based solely on the breathing pattern, there are only five possibilities in two groups (Table 42-3). The utility of this system is immediately obvious. A new mode, such as airway pressure release ventilation (APRV), can be explained as simply a form of PC-IMV. Assuming the clinician already understands the concept of PC-IMV, it takes little effort to understand the additional nuances of APRV (e.g., different labels for control settings, alarms). This level of description avoids the cumbersome verbal ad hoc definition for APRV, such as “a mode that allows spontaneously breathing patients to breathe at a positive-pressure level but drops briefly to a reduced pressure level for carbon dioxide (CO2) elimination during each breathing cycle,” or the misleading description of some authors who explain APRV as two levels of continuous positive airway pressure.

TABLE 42-3

All Modes of Ventilation Can Be Characterized by One of Five Breathing Patterns

Breath Control Variable Breath Sequence Abbreviation
Volume (control) Continuous mandatory ventilation VC-CMV
Intermittent mandatory ventilation VC-IMV
Pressure (control) Continuous mandatory ventilation PC-CMV
Intermittent mandatory ventilation PC-IMV
Continuous spontaneous ventilation PC-CSV

image

PC-IMV and PC-CSV can also be used to clarify what bilevel positive airway pressure ventilation means. For example, on the Respironics BiPAP S/T or BiPAP AVAPS (Phillips Healthcare, Andover, Massachusetts) noninvasive home ventilators, the “timed” mode is PC-IMV, whereas the “spontaneous” mode is PC-CSV. BiPAP ventilation and bilevel ventilation are particularly ambiguous terms. To make matters even more confusing, the Puritan Bennett 840 ventilator has a “bilevel” mode that allows for additional pressure support during a pressure-limited, time-cycled, mandatory breath. With PEEP, the mandatory pressure limit, and the pressure support limit, the mode actually provides “trilevel” ventilation.

The specification of breathing pattern is also useful for grouping together modes that function the same way but are given different names. Both “CMV plus pressure limited ventilation” (Draeger Evita 4) and “volume-assured pressure support” (Bird 8400ST) are PC-IMV. It is also possible to group ventilators in terms of the number of breathing patterns they offer; some offer only one or two, whereas others offer all five. This grouping might be useful as an initial screening tool when planning ventilator purchases. Although the Bird 8400ST and 8400 STi may still be in use, this particular ventilator is not actively supported by the manufacturer (CareFusion, San Diego, California).

Targeting Schemes

Control variables and the differences between pressure and volume control have been discussed, but what is meant by control has not been explained. There are two general ways to control a variable: open loop control and closed loop control.14

Closed Loop Control

Closed loop control means that the delivered pressure, volume, and flow can be measured and used as feedback information to control the driving mechanism (similar to speed control in an automobile). Inspiratory volumes, flows, and pressures can be made to match specified input values despite disturbances such as changes in patient load and minor leaks in the system.

The basic concept of closed loop control has evolved into at least six different ventilator control systems or targeting schemes14 (set-point, dual, servo, adaptive, optimal, and intelligent). These targeting schemes are the foundation that makes possible several dozen apparently different modes of ventilation. Once it is understood how these targeting schemes work, many of the apparent differences are seen to be similarities. A lot of the confusion surrounding ventilator marketing hype is avoided, and the true clinical capabilities of different ventilators are appreciated.

Dual

Dual targeting is a more advanced version of set-point targeting. It gives the ventilator the decision of whether the breath will be volume controlled or pressure controlled according to the operator-set priorities. The breath may start out in pressure control and automatically switch to volume control, as in the Bird VAPS mode or, the reverse, as in the Draeger Pmax mode. The Maquet SERVO-i ventilator has a mode called volume control, and the operator presets both inspiratory time and tidal volume as would be expected with any conventional volume control mode. However, if the patient makes an inspiratory effort that decreases inspiratory pressure by 3 cm H2O, the ventilator switches to pressure control and, if the effort lasts long enough, flow cycles the breath. If the tidal volume and inspiratory time are set relatively low and the inspiratory effort is relatively large, the resultant breath delivery is indistinguishable from pressure support. As a result, the tidal volume may be much larger than the expected, preset value. This occurrence highlights the need to understand dual targeting. Because both pressure and volume may be the control variables during dual targeting, by convention we designate the control variable as the one with which the breath initiates; the alternative control variable may never be implemented during the breath, depending on the other factors in the targeting scheme.

Servo

Set-point targeting attempts to maintain a constant output to match a constant input, whereas servo targeting is designed to track a moving input, similar to power steering on an automobile. Servo targeting was developed during World War II to aim ships’ guns and radar equipment. It makes the proportional assist mode possible.15 In this mode, the output of the ventilator follows and amplifies the patient’s own flow pattern. The ventilator can support the abnormal load imposed by disease, while the patient’s own muscles handle a normal load secondary to the respiratory system’s natural resistance and compliance.

Adaptive

Adaptive targeting means automatic adjustment of one set-point to maintain a different operator-selected set-point. One of the first examples of a mode using adaptive control was pressure-regulated volume control on the Siemens Servo 300 ventilator (Siemens Corp, New York City, New York). Adaptive targeting is an evolutionary step because it gives the ventilator the capability to determine a set-point level independent of the operator. While set-point targeting operates within breaths, adaptive control introduces another feedback loop that operates between breaths (i.e., pressure targeting within breaths and volume targeting between breaths). Using feedback of volume allows the ventilator to adapt to changes in the patient’s lung mechanics. Despite having various names for the specific modes it allows, adaptive targeting to date has been implemented most commonly as a way for the ventilator to adjust the inspiratory pressure of a breath automatically to meet an operator set volume target over several breaths. Another example would be adjustment of the mandatory breath frequency to achieve a minute ventilation target.

Optimal

Optimal targeting takes adaptive control a step further by allowing the ventilator to set both volume and pressure set-points. Optimal control takes its name from the fact that a mathematical model is used to find the best (e.g., highest or lowest) value of some performance function. Hamilton Medical makes the only commercially available ventilators with this feature. Optimal targeting allows the ventilator to make all subsequent adjustments after the operator sets the target minute ventilation. Hamilton Medical (and the authors who did the basic research published in the literature) refer to the optimal control mode as adaptive support, which confuses their more highly evolved targeting scheme, optimal control, with the simpler adaptive control described earlier.

Other forms of optimal control give the ventilator even more authority using exhaled CO2 as a feedback signal.16 In Europe, Hamilton Medical has introduced IntelliVent, a system for complete closed loop control of minute ventilation, fractional inspired oxygen (FiO2), and PEEP using both CO2 and SpO2 as feedback signals.

Intelligent

The term intelligent refers to automatic targeting strategies that make use of artificial intelligence methods, such as “fuzzy logic,” rule-based expert systems, and even neural networks.14 This type of automatic targeting is yet another evolutionary step because it gives the ventilator more information than what may be contained in a simple, static, mathematical model. Knowledge-based targeting schemes attempt to capture the experience of human experts and expand the scope of control to potentially all parameters of the ventilatory mode. An experimental application of this type of control has been described for automatic adjustment of pressure support.17 That system has now been commercialized as the Draeger SmartCare mode.

An even more sophisticated approach coupled a knowledge base with fuzzy logic.18 In this case, the ventilator used both instantaneous measurements of physiologic values such as respiratory rate and saturation and their rates of change. Fuzzy logic19 was used as a way to integrate the measurements with predefined ranges of values representing the patient status. When the patient’s status was determined, appropriate expert rules were selected from a lookup table and used to adjust the ventilator. Although this was a limited application, it proved the concept.

The most convincing proof of concept was presented by East and colleagues.20 These investigators used a rule-based expert system for ventilator management in a large, multicenter prospective randomized trial. Although survival and length of stay were not different between human and computer management, computer control resulted in a significant reduction in multiorgan dysfunction and lower incidence and severity of lung overdistention injury. However, the most important finding was that expert knowledge can be encoded and successfully shared with institutions that had no input into the model. The expert system did not directly control the ventilator but rather made suggestions for the human operator. Theoretically, the operator could be eliminated.

Perhaps the most exotic targeting scheme to date is the artificial neural network.21 This experimental system (not yet commercially available) did not directly control the ventilator but acted as a decision support system. Snowden and coworkers21 commented that the neural network was capable of learning, which offers significant advantages over static rule-based systems.

Neural networks are essentially data modeling tools used to capture and represent complex input-output relationships. A neural network learns by experience the same way a human brain does, by storing knowledge in the strengths of internode connections. As data modeling tools, neural networks have been used in many business and medical applications for both diagnosis and forecasting.22 A neural network, similar to an animal brain, is composed of individual neurons. Signals (action potentials) appear at the unit’s inputs (synapses). The effect that each signal has may be approximated by multiplying the signal by some number or weight to indicate the strength of the synapse. The weighted signals are summed to produce overall unit activation. If this activation exceeds a certain threshold, the unit produces an output response. As the network learns, the weights change, affecting the final output.

Specifying the targeting scheme in a mode description can help to distinguish between modes that look nearly identical on a graphics monitor and present conceptual or verbal problems when trying to differentiate them. It might be difficult to appreciate the difference between pressure support and volume support on a Maquet SERVO-i ventilator, but consider these simple descriptions: Pressure support is PC-CSV with set-point targeting of inspiratory pressure. Volume support is PC-CSV with adaptive targeting of inspiratory pressure. If a clinician knows the definitions of these words and acronyms, he or she can immediately understand how different the modes are. Attention would also be directed to the clinical implications for the patient (e.g., what settings are required). Knowledge of targeting schemes also allows the clinician to see that something like Draeger’s AutoFlow feature is not just a “supplement” or “extra setting” as the operator’s manual indicates but creates a whole new mode. Operating the Draeger Evita 4 in CMV yields VC-CMV with set-point targeting of inspiratory volume and flow. However, activating AutoFlow when CMV is set yields PC-CMV with adaptive targeting of inspiratory pressure and vastly different clinical ramifications for the patient. These two modes are about as different as any two modes can be.

If the breath sequence is IMV, a complete description of the mode includes targeting schemes for both mandatory and spontaneous breaths. On the Puritan Bennett 840 ventilator, the mode called synchronized intermittent mandatory ventilation would be described as VC-IMV with set-point targeting of volume for mandatory breaths and set-point targeting of pressure for spontaneous breaths.

At the highest level of detail, a mode can be fully characterized by adding the phase variables, followed by detailing the operational logic programmed into the software control system. The specification of the breathing pattern that the mode can produce, the type of targeting, and the specific strategy (phase variable and operational logic) it uses for both mandatory and spontaneous breaths make up a complete classification for any mode of ventilation (see Table 42-2).

The AutoMode feature on the Draeger Evita XL ventilator illustrates one example of why it is important to separate mandatory from spontaneous breath descriptions. AutoMode is a form of IMV, and there are three different types, as follows:

Two important facts about current ventilators should be apparent. First, ventilators offer far more complex modes than in the past, so understanding how they work is no trivial task. Manufacturers are throwing together combinations of features that not only are difficult to comprehend but also strain rational justification, often with no clinical data to support their efficacy. Second, there is no agreement among ventilator manufacturers when it comes to nomenclature. Also, the names they create for modes are baffling.

Output Waveforms

Electrocardiograms and blood pressure waveforms are studied to understand heart physiology. In the same way, to understand ventilator-patient interaction, output waveforms must be examined. The output waveforms of interest during ventilatory support are pressure, volume, and flow. For each control variable, current ventilators produce a limited number of waveforms. Basic waveforms are shown in Figure 42-11.

Because the waveforms in Figure 42-11 are models, they do not show the minor deviations, or “noise,” often seen during actual ventilator use. Such noise can be caused by many factors, including vibration and turbulence. These waveforms also do not show the effect of expiratory circuit resistance because this varies depending on the ventilator and type of circuit. The waveforms do not show the various indicators of problems with ventilator-patient synchrony (e.g., improper sensitivity setting and gas trapping), which are beyond the scope of this chapter. Finally, waveform appearances change when the time scale is altered. A faster sweep (shorter time scale) tends to widen a given waveform, whereas a slower sweep speed (longer time scale) compresses the waveform.

Most ventilator waveforms are rectangular, exponential, ramp, or sinusoidal in shape. Although various subtypes are possible, we describe only the most common here. Waveforms are listed according to the shape of the control variable waveform. Any new waveforms produced by future ventilators can easily be accommodated by this system.

Pressure

Rectangular

Mathematically, a rectangular waveform is referred to as a step or instantaneous change in transrespiratory pressure from one value to another (see Figure 42-11, A). In response, volume increases exponentially from zero to a steady-state value equal to compliance multiplied by the change in airway pressure (i.e., PIP − PEEP). Inspiratory flow decreases exponentially from a peak value (at the start of inspiration) equal to (PIP − PEEP) divided by resistance.

Exponential

Exponential pressure waveforms are common outputs of neonatal ventilators and can be produced by adjusting the pressure rise control on some newer adult ventilators. The resulting pressure and volume waveforms can take on various shapes ranging from an exponential rise (same shape as the volume waveform in Figure 42-11, A) to a linear rise (same shape as the volume waveform in Figure 42-11, B). Generally, the flow waveform is similar to that seen in Figure 42-11, A except that peak inspiratory flow is reached gradually rather than instantaneously (resulting in a rounded rather than peaked waveform), and peak flow is lower than with a rectangular pressure waveform.

Volume

Ramp

Volume controllers that produce an ascending ramp waveform (i.e., the Bennett MA-1) produce a linear rise in volume from zero at the start of inspiration to the peak value, or set tidal volume, at end-inspiration (see Figure 42-11, B). In response, the flow waveform is rectangular. The pressure waveform rises instantaneously from zero to a value equal to resistance multiplied by flow at the start of inspiration. From here, it rises linearly to its peak value (PIP) equal to elastance multiplied by tidal volume plus resistance multiplied by flow.

Flow

Rectangular

A rectangular flow waveform is perhaps the most common output (see Figure 42-11, B). When the flow waveform is rectangular, volume is a ramp waveform, and pressure is a step followed by a ramp as described for the ramp volume waveform.

Ramp

Many respiratory care practitioners (and ventilator manufacturers) refer to ramp waveforms as either accelerating or decelerating flow patterns. The use of either of these terms is usually inappropriate. If a car slows, we do not say that its velocity decelerates; we say that the car decelerates. We do not say that a cyclotron is a velocity accelerator but that it is a particle accelerator. The rate of change of position of an object is the velocity of the object; analogously, the rate of change of volume is flow. The rate of change of velocity of an object is the acceleration of the object; likewise, the rate of change of flow is the acceleration of volume, not the acceleration of flow. So if we want to say that flow changes, we should simply talk about an increasing flow or a decreasing flow (or an accelerating volume or a decelerating volume), not an accelerating flow or a decelerating flow.

Ascending Ramp

A true ascending ramp waveform starts at zero and increases linearly to the peak value (see Figure 42-11, C). Actual ventilator flow waveforms may be truncated; inspiration starts with an initial instantaneous flow. The Bear-5 starts inspiration at 50% of the set peak flow. Flow increases linearly to the set peak flow rate. In response to an ascending ramp flow waveform, the pressure and volume waveforms are exponential with a concave upward shape. As previously described with the 8400 St and STi ventilators, this particular ventilator may still be in use, however, the current manufacturer does not actively produce or support this product.

Descending Ramp

A true descending ramp waveform starts at the peak value and decreases linearly to zero (see Figure 42-11, D). Ventilator flow waveforms are usually truncated; inspiratory flow rate decreases linearly from the set peak flow until it reaches some arbitrary threshold where flow drops immediately to zero (e.g., the Puritan Bennett 7200a ends inspiration when the flow rate decreases to 5 L/min). In response to a descending ramp flow waveform, the pressure and volume waveforms are exponential with a concave downward shape.

Sinusoidal

Some ventilators offer a mode in which the inspiratory flow waveform approximates the shape of the first half of a sine wave (see Figure 42-11, E). As with the ramp waveform, ventilators often truncate the sine waveform by starting and ending flow at some percentage of the set peak flow rather than start and end at zero flow. In response to a sinusoidal flow waveform, the pressure and volume waveforms are also sinusoidal but out of phase with each other.

Effects of Calibration Errors and the Patient Circuit

The pressure, volume, and flow the patient receives are never precisely the same as what the clinician sets on the ventilator. These differences sometimes are caused by instrument inaccuracies or calibration error. In addition, the patient delivery circuit contributes to discrepancies between the desired and actual patient values because the patient circuit has its own compliance and resistance. The pressure measured on the inspiratory side of a ventilator always is higher than the pressure at the airway opening owing to patient circuit resistance. In addition, the volume and flow coming out of the ventilator exceeds the volume and flow delivered to the patient because of the compliance of the patient circuit.

Using an analogy to electrical circuits, compliance of the delivery circuit can be shown to be connected in parallel with the compliance of the respiratory system (i.e., both elements sharing the same driving pressure). Consequently, the total compliance of the ventilator-patient system is simply the sum of the two compliances. The resistance of the delivery circuit is connected in series with the respiratory system resistance (i.e., both elements sharing the same flow) so that the total resistance is the sum of the two. Based on these assumptions, the relationship between the volume input to the patient (at the point of connection to the patient’s airway opening) and the volume output from the ventilator (at the point of connection to the patient circuit) can be described by the following equation:

Volume input to patient=Volume output from ventilator1+Cpc/Crs Equation 42-7

image Equation 42-7

where Cpc is the compliance of the patient circuit, and Crs is the total compliance of the patient’s respiratory system. The equation shows that the larger the patient circuit compliance compared with the patient’s respiratory system, the larger the denominator on the right-hand side of the equation, and the smaller the delivered tidal volume is compared with the volume coming from the ventilator’s drive mechanism.

Assuming that the volume exiting the ventilator is the set tidal volume, the patient circuit compliance (Cpc) is calculated as follows:

Cpc=Set tidal volumePplatPEEP Equation 42-8

image Equation 42-8

where Pplat is the pressure measured during an inspiratory hold maneuver with the Y-piece of the patient circuit occluded (patient not connected), and PEEP is end expiratory pressure (i.e., baseline pressure). Most authors recommend the use of PIP for Pplat in this equation, which is acceptable but may lead to a slight underestimation of patient circuit compliance. Pplat is slightly lower than PIP because of the flow-resistive pressure decrease of the patient circuit if pressure is not measured at the Y-piece. This difference is greatest in small-bore, corrugated patient circuit tubing but is probably insignificant.

The effects of patient circuit compliance are most troublesome during volume-controlled ventilation. In neonatal ventilation, the patient circuit compliance can be three times that of the respiratory system, even with small-bore tubing and a small-volume humidifier. In an attempt to deliver a preset tidal volume, the volume delivered to the patient may be only 25% of that exiting the ventilator, whereas 75% is compressed in the patient circuit.

During pressure-controlled ventilation, the compliance of the patient circuit has the effect of rounding the leading edge of a rectangular pressure waveform (see Figure 39-9), which could reduce the volume delivered to the patient. This effect is prevented if the pressure limit is maintained for at least five time constants of the respiratory system.

For both pressure-controlled and volume-controlled ventilation, the patient circuit compliance and resistance, along with the resistance of the exhalation valve (in series with the patient circuit and respiratory system resistance) increase the expiratory time constant. A large circuit compliance coupled with a short expiratory time can lead to inadvertent PEEP or auto-PEEP. The set values for pressure, volume, and flow may be different from the output (from ventilator) values because of calibration errors and different from the input (to the patient) because of the effects of the patient circuit. These two general sources of error cause discrepancies between the desired and actual patient values.

Operator Interface

The ventilator’s operator interface, or ventilator display, has undergone extensive evolution over the last 30 years. Originally, the displays on ventilators were “analog.” Operator inputs, or settings, were accomplished with hard-wired knobs, buttons, and dials. The ventilator outputs, such as alarm conditions and ventilating pressure, were displayed with bulbs, light emitting diodes (LEDs), and meters. Some older home care ventilators still use analog displays (Figure 42-12). The development of inexpensive microprocessors has led manufacturers to use “digital” displays almost exclusively on all types of ventilators. Digital interfaces use LED or LCD screens for visual display of ventilator data along with some multipurpose hard-wired buttons. The simplest example would be the display of a bilevel positive airway pressure machine used for treating sleep apnea at home (Figure 42-13). A more advanced digital display uses dedicated special-purpose buttons and dials (Figure 42-14). The most advanced interfaces use the concept of the “virtual” instrument, meaning that knobs, buttons, dials, and meters are simulated on a computer screen (sometimes a touch screen) and often incorporating a single mechanical dial that is used to set multiple parameters (Figure 42-15). Computer screens allow graphic displays of alarm settings as bar graphs along with pressure, volume, and flow waveforms as scalars or loops.

Operator Inputs

The operator inputs include the parameters of the ventilatory pattern (i.e., mode and phase variables) and any desired alarm settings. We discuss some specific examples of trigger, target, and cycle variables and an assortment of alarm parameters. Each ventilator has its own unique layout, so a complete description of all available devices is beyond the scope of this chapter.

Mode Settings

There is no standardization among ventilator manufacturers regarding the names of modes, their classification, or how the operator identifies and sets them on a given device. This requires the operator to learn the specific terminology and layout of each individual brand of ventilator for every function it provides. For clinical settings where many different ventilator brands are used, this can be a daunting and perhaps unachievable task. For an example of how different and confusing ventilator displays can be regarding mode selection, compare Figures 42-16 and 42-17.

Trigger, Target, and Cycle Variable Settings

There was a time when all the input settings and output displays of a modern ICU ventilator could fit on the face of the ventilator, even in digital format (Figure 42-18). However, with the proliferation of computer screens for displays and the increasing complexity of modes, the user is now often faced with the need to switch between multiple screens to access all of the ventilator’s capabilities (Figure 42-19). This is a relatively new challenge for the ventilator industry, and more work needs to be done to perfect the operator interface. More recent research on the topic suggests that the design of the user interface is relevant to the occurrence of operational failures and that ventilator designers could optimize the user-interface design to reduce operational failures.23

Trigger variables include selection of either pressure or flow triggering and the trigger thresholds (in cm H2O or L/min). Sometimes these variables are explicitly described, and sometimes they are just referred to as a sensitivity setting. Frequency is also a trigger variable, setting the time delay between time-triggered inspirations. A preset minute ventilation can also be a trigger variable in modes such as Draeger’s mandatory minute ventilation mode and Hamilton’s adaptive support mode; if the total minute ventilation from mandatory and spontaneous breaths falls below the preset value, the ventilator triggers mandatory breaths. Draeger has introduced a novel trigger mechanism (called AutoRelease on its Evita Infinity V500 ventilator) based on the percentage of peak expiratory flow. Mandatory pressure-controlled breaths are flow triggered when expiratory flow decays to the preset threshold, usually expressed as a percentage of the peak expiratory flow for the breath. Setting the cycle threshold at a lower value (e.g., 25%) increases the average inspiratory time, whereas setting the threshold higher (e.g., 75%) decreases the average inspiratory time. The actual inspiratory time for any given breath is determined by the patient’s mechanics (i.e., resistance, compliance, and inspiratory effort). The trigger mechanism is designed for use during the airway pressure release mode, and target variables for volume control typically include inspiratory flow and inspiratory flow waveform (e.g., square, descending ramp, or sinusoidal).

Target variables for pressure control include inspiratory pressure and sometimes pressure rise time. The latter variable sets the speed with which inspiratory pressure increases to the target value and controls peak inspiratory flow. Inspiratory pressure is another term that causes confusion because of nonstandard usage. On some ventilators, it is measured relative to atmospheric pressure (and should be called peak inspiratory pressure), and on others it is measured relative to PEEP and should be called simply inspiratory pressure.24 The problem is that when a patient is changed from one ventilator (or mode) using one convention to another ventilator (or mode) with the other convention, the risk of inadvertently setting the wrong inspiratory pressure increases and could lead to adverse events.

For example, imagine a ventilated patient with respiratory system resistance of 10 cm H2O/L/sec and compliance of 0.035 L/cm H2O. This patient is transported on a ventilator with set inspiratory pressure of 25 cm H2O (relative to atmospheric pressure), PEEP of 10 cm H2O, and inspiratory time of 1.0 second. Under the assumptions of passive ventilation and no intrinsic PEEP, the measured peak airway pressured with this particular ventilator is 25 cm H2O, the ΔP is 25 − 10 = 15 cm H2O, and the tidal volume would be 495 ml. On arrival to the ICU, pressure-control ventilation is initiated with a ventilator in which inspiratory pressure is set relative to PEEP. If the same settings were used (inspiratory pressure 25 cm H2O and PEEP 10 cm H2O) the new peak airway pressure would be 25 + 10 = 35 cm H2O, ΔP would be 25 cm H2O, and the tidal volume would 825 ml. In this example, the patient is put at risk of ventilator-induced lung injury and cardiac compromise because of the inadvertent increase in ΔP and mean airway pressure. If the transport had been in the opposite direction, the patient would have been at risk of hypoventilation because of a lower ΔP and loss of oxygenation owing to decreased mean airway pressure. One could argue that a knowledgeable clinician would not make such a mistake because of familiarity with the two ventilators. What if the patient was being manually ventilated during intrahospital transport to the ICU and the only data available were the inspiratory pressure and PEEP? The risk of confusion, if not actual harm, is inherent in the way ventilator settings are documented, owing to lack of standardization.

Cycle variables for volume control are usually tidal volume or inspiratory time. Some ventilators allow the operator to preset minute ventilation instead of tidal volume. This arrangement sets tidal volume indirectly as a function of frequency and minute ventilation (i.e., tidal volume = minute ventilation/frequency). Inspiratory pause time is a time-based cycle variable that may be an option. Some ventilators allow the operator to set an inspiratory pause time directly, and others allow setting both inspiratory flow and inspiratory time. In the latter case, inspiratory flow time is equal to the tidal volume divided by inspiratory flow, and inspiratory pause time is inspiratory time minus inspiratory flow time.

For example, if the tidal volume is 500 ml (0.5 L) and the inspiratory flow is 60 L/min (1 L/sec), inspiratory flow time is 0.5 L/(1 L/sec), or 0.5 second. If the inspiratory time is set to 1 second, an inspiratory pause is created lasting 1.0 − 0.5 = 0.5 second. Cycle variables for pressure control generally include inspiratory time (or I : E ratio) for mandatory breaths and possibly flow cycle threshold for spontaneous (pressure support) breaths. Pressure support breaths are flow cycled when inspiratory flow decays to the preset threshold, usually expressed as a percentage of the peak inspiratory flow for the breath. Setting the cycle threshold at a lower value (e.g., 25%) increases the average inspiratory time, whereas setting the threshold higher (e.g., 75%) decreases the average inspiratory time. The actual inspiratory time for any given breath is determined by the patient’s mechanics (i.e., resistance, compliance, and inspiratory effort). All ventilators except for some home care devices, allow adjustment of the baseline pressure, or PEEP.

Alarm Settings

The purpose of ventilator alarms is to bring events to the attention of the clinician. Events are conditions or occurrences that require clinician awareness or intervention. Events can be classified according to four levels of priority.25 Level 1 events are immediately life-threatening. These include insufficient or excessive gas delivery to the patient, exhalation valve failure, control circuit failure, or loss of power. Alarm indicators in this category should be mandatory (cannot be turned off by the operator), redundant, and noncanceling. Level 2 events range from mild irregularities in machine function to dangerous situations that could threaten patient safety if left unattended. These include failure of the air-O2 blending system, inadequate or excessive PEEP, autotriggering, circuit leak, circuit occlusion, inappropriate I : E ratio, and failure of the humidification system. Alarms in this category are not redundant and may be self-canceling (i.e., automatically turned off if the event ceases). Level 3 events reflect changes in the level of ventilatory support. Examples include changes in the patient’s ventilatory drive or respiratory system mechanics and the presence of auto-PEEP. Level 3 events often trigger the same alarms as levels 1 and 2. Level 4 events are focused entirely on the patient. These include changes in gas exchange, dead space, oxygenation, and cardiovascular functions. Many ventilators do not warn of these events, and external monitors are required for surveillance.

Ventilators do not display alarm settings in terms of levels of priority. Instead, they tend to lump them all together on the screen (Figures 42-20 and 42-21; see Figure 42-15). The setting of alarm thresholds is a complicated topic that has been studied but for which little information is available regarding mechanical ventilation. The basic problem is to maximize true alarms and minimize false alarms. A high false alarm rate leads to habituation and clinicians ignoring warnings. False alarms can also lead to inappropriate responses. In one 200-hour study of an ICU, 1214 alarms occurred, and 2344 tasks were performed. On average, alarms occurred six times per hour; 23% were effective, 36% were ineffective, and 41% were ignored.26 In another ICU study, during 982 hours of observation, 5934 alarms occurred, corresponding to six per hour. About 40% of the alarms did not correctly describe the patient condition and were classified as technically false; 68% of those were caused by manipulation. Only 885 (15%) of all alarms were considered clinically relevant.27

Although these studies did not address mechanical ventilator alarms specifically, it is not hard to imagine similar results for such a study. Ventilator alarms are usually set by the operator (or as default values by the ventilator) as either a set value or a set percentage of the current value. Examples would be low and peak airway pressure alarms set at the current value ±5 cm H2O or low and peak tidal volume and minute ventilation set at ±25% of the current value.25 The problem is that the parameters we want to set alarms for, in particular, airway pressure, tidal volume, and minute ventilation, are highly variable, with significant portions at extreme values (Figure 42-22).28 Limits set as absolute values or percentages may reduce safety for some extreme values, while increasing nuisance events for other values. An alternative approach might be to reference the alarm limits to the current value of the parameter such that extreme values have tighter limits. Further research is needed to identify optimization algorithms (i.e., minimize both harmful and nuisance events) for intelligent targeting schemes to set alarms automatically during mechanical ventilation.

Ventilator Output Displays

Ventilator output displays are essentially the values of monitored parameters that result from the operator settings. There are four basic ways to present the monitored data: as numbers, as waveforms, as trend lines, and in the form of abstract graphic symbols.

Waveforms and Loops

Most ICU ventilators display waveforms (sometimes called scalars) of airway pressure, volume, and flow as functions of time. Such displays are useful for identifying the effects of changes in settings or mechanics on the level of ventilation. They are also very useful for identifying sources of patient-ventilator asynchrony, such as missed triggers, flow asynchrony, and delayed or premature cycling.29 They can also display one variable against another as an x-y or “loop” display. The most common loop displays show pressure on the horizontal axis and volume on the vertical axis or volume on the horizontal axis and flow on the vertical axis. Pressure-volume loop displays are useful for identifying optimal PEEP levels (quasistatic loops only) and overdistention. Flow-volume loops are useful for identifying the response to bronchodilators. An example of a composite display showing numeric values, waveforms, and loops is shown in Figure 42-23.

Picture Graphics

A new development in ventilator displays involves the use of picture graphics to represent useful information about the patient-ventilator system. A study30 showed that subjects with graphic rather than conventional displays of obstructed endotracheal tubes and auto-PEEP problems were detected and treated faster. The investigators also reported significantly lower subjective workloads using the graphic display.

Hamilton Medical was the first manufacturer to use innovative picture graphics on their G5 ventilator. In particular, they created a graphic representation of the lungs, called a dynamic lung panel, which visually displayed information about resistance and compliance by the shape and color of the lungs and airways (Figure 42-25). In addition, Hamilton Medial created a unique graphic representation called the vent status panel that displays key parameters (e.g., oxygenation, ventilation, and spontaneous breathing activity) and shows when each item is in or out of an acceptable zone and for how long. This graphic display makes weaning status easy to identify. Draeger Medical followed with a similar graphic display called Smart Pulmonary View, which is a graphic display of respiratory system compliance and resistance and of the spontaneous and mandatory minute volume (Figure 42-26).

Types of Ventilators

Ventilators may be divided into categories according to type and the setting in which the ventilator is to be used. The two types or classes ventilators may be divided into are conventional and nonconventional. Conventional ventilators produce breathing patterns that are at or near physiologic normal values for the intended population (e.g., adult, pediatric, and infant). There are also manufacturing limits on the maximum breath rate a conventional ventilator may deliver. The U.S. Food and Drug Administration (FDA) places a maximum breath rate limit of 150 breaths/min for conventional ventilators. Tidal volumes that are either operator-set or delivered to the patient as a result of a preset pressure through a conventional ventilator are sufficient or large enough to clear anatomic dead space. Conversely, high-frequency ventilators typically produce respiratory frequencies or breathing rates that are much higher than physiologically possible and tidal volumes that are less than anatomic dead space.

Ventilators may also be categorized by the setting in which they are used along the continuum of care—specifically critical care, subacute, home care and long-term care, and transport. The designs of the ventilators match the needs of the patient population and the unique characteristics of the setting in which they are used. Ventilator manufacturers have paid particular attention to economic constraints health care facilities are facing. Innovations in ventilator design have broadened their use across settings. An example is the use of ventilators across a spectrum of ages. This innovation does not negate the need for specific ventilators for use with infants and children but provides a mechanism through which respiratory care departments can maximize the use of capital expenditures. Although ventilators are not subcategorized as adult or pediatric in this chapter, it is crucial for the practitioner to be aware of factors such as minimal tidal volume limits, trigger sensitivity, response time, and availability of leak compensation when selecting a ventilator for use with pediatric patients. A comprehensive listing of all ventilators available for use in the United States is not provided. Rather, a small sample representing some commercially available products and their specifications is provided. Comprehensive listings may be found in textbooks dedicated to respiratory care equipment.

Critical Care Ventilators

Critical care ventilators provide clinicians with sophisticated methods for breath delivery. This class of ventilators also provides advanced monitoring capabilities and tools that enable clinicians to assess the patient-ventilator interaction easily. Calculations of lung mechanics parameters, including auto-PEEP, static and dynamic compliance, inspiratory/expiratory resistance, rapid shallow breathing index, time constant, and work of breathing, are integrated into ventilators used in this environment. Integrated physiologic noninvasive and invasive assessment tools, such as esophageal pressure monitoring and end-tidal CO2 monitoring, are also commercially available. The availability of these features equips bedside caregivers with the tools needed to assess patient-ventilator interaction and match ventilator capability with physiologic need. Both high-frequency and conventional ventilators may be used in the critical care setting.

High-Frequency Ventilators

The availability of sophisticated devices, such as jet ventilators and high-frequency oscillators, facilitates ventilatory management of infants, children, and adults in the PC-IMV mode who fail to maintain adequate oxygenation and acid-base balance with conventional ventilatory support. High-frequency jet ventilators, such as the Bunnell Life Pulse (Bunnell Inc, Salt Lake City, Utah), deliver short bursts, or jet pulses, of mixed gas through a special adapter for endotracheal tubes or a specially designed endotracheal tube. The LifePort Endotracheal Tube Adapter (Bunnell Inc, Salt Lake City, Utah) is designed to replace the 15-mm outer diameter adapter on the proximal portion of a traditional endotracheal tube. This adapter has a high-frequency ventilator circuit connection port. The use of this adapter minimizes the need to reintubate the patient with a specially designed endotracheal tube. The connection port may be capped off or occluded when there is no longer a need for high-frequency ventilation. An endotracheal tube specially designed for use with high-frequency ventilation is similar in appearance to a conventional tube, with two additional small chambers molded into the wall. This triple-lumen design enables tracheal pressures to be monitored during the delivery of jet pulses.

High-frequency jet ventilators require a high pressure source (20 to 50 psig) to function. This type of ventilator consists of a system for regulating inlet pressure (psig) and a cycling mechanism and a device such as an air-O2 blender for controlling FiO2. The small volumes of gas are delivered at rapid rates (4 to 250 times the normal respiratory rate).31 The benefits of rescue and elective use of high-frequency jet ventilation as a treatment for acute lung disease and acute respiratory distress syndrome (ARDS) are unclear.31 However, the literature supports its effectiveness in maintaining alveolar ventilation and reducing morbidity during surgical repair of tracheal and airway anomalies.32,33

High-frequency oscillation also allows very small tidal volumes to be delivered at rapid frequencies (180 to 1200 cycles/minute). High-frequency oscillators use a piston-driven diaphragm to produce the airflow oscillations. Tidal volume depends on the force and distance the piston moves from baseline. A special endotracheal tube is not required to implement this form of PC-IMV. The SensorMedics 3100A and 3100B high-frequency oscillators (CareFusion) are approved and commercially available for use in neonatal/pediatric (<35 kg) and pediatric/adult (>35 kg) populations. There are independent reports of improved morbidity and mortality with the use of high-frequency oscillation in infants and children with diffuse alveolar and small airways disease.34 Combined evidence from observational and clinical trials appears to be favorable with respect to improved mortality rates in patients with heterogeneous lung disease, such as ARDS.35,36 In contrast to conventional ventilators, convention plays a minor role in gas transport with high-frequency jet ventilation and high-frequency oscillation. Rather, the mixing and diffusion of gases in the airways enhances fresh gas delivery to and elimination of CO2 from the alveoli.

Conventional Ventilators

Conventional ventilators may be used with various interfaces in the critical care setting. Options are available on this type of ventilator for use with an artificial airway (endotracheal or tracheostomy tube) or noninvasively with assorted interfaces (e.g., nasal, oronasal, or full-face masks). The availability of the noninvasive option eliminates the need for a standalone noninvasive ventilator. However, standalone noninvasive ventilators do have application and are also used in the critical care setting. This type of ventilator is discussed in detail subsequently. Ventilators used in the critical care environment have the capability to assess and monitor complex ventilator-patient interactions. Work of breathing or the intricacies of breath delivery may be examined by evaluating numerically or graphically displayed data. Many ventilators also apportion automatic adjustments to breath delivery in response to changes in lung mechanics and parameters preset by the operator.

In addition to universal patient population applications, some critical care ventilators are manufactured with an internal battery. The SERVO-i has a plug-in battery module. This ventilator can provide at least 3 hours of uninterrupted power supply when six rechargeable 12-V batteries are used in the module. Uninterrupted ventilatory assistance may be provided not only in the critical care setting but also during interhospital transport. Minimizing circuit disconnections can enhance safety by reducing the risks for derecruitment,37 hemodynamic instability,38 and factors contributing to nosocomial infections.39 The aforementioned features enable clinicians to optimize the patient-ventilator interaction.

Subacute Care Ventilators

Patients who have an acute illness, injury, or exacerbation of disease process receive subacute care. Generally, mechanically ventilated patients in this setting have a stable cardiopulmonary status. Their condition is such that the care provided in this setting does not depend heavily on high-technology monitoring or complex diagnostic procedures. Rather, the focus is on coordinated services aimed at managing complex medical conditions, liberation from ventilatory support, and rehabilitation services. Ventilators used in this care venue have less sophisticated monitoring systems and mode options. Subacute care may be rendered in freestanding facilities or within a specialized unit within a hospital. Consequently, the ventilators used in this environment bridge the gap between ventilators designed specifically for critical care and ventilators for home care or long-term care. The Savina (Draeger Medical) offers features germane to critical care ventilators, such as graphic display of pressure, volume, and flow waveforms and the availability of a noninvasive ventilation mode of operation. The availability of a low-pressure O2 option enables O2 delivery independent of a central gas supply. An O2 concentrator can be used to supply O2 to the patient breathing circuit; this is analogous to O2 delivery methods used by ventilators in the home care or long-term care environment.

Home Care Ventilators

Patients with chronic respiratory failure from primary pulmonary disease, trauma, or neuromuscular disease may require ventilatory assistance to augment or replace spontaneous breathing and maintain life. Ventilators designed for use in the home or long-term care institutions facilitate the transition of patients from an acute or subacute care environment to one focused on enhancing the individual’s quality of life and providing services to sustain or improve physical or physiologic function in a cost-efficient manner.40 Ventilators must be able to support the ventilatory needs of the patient and provide supplemental O2 in a venue where compressed gas resources are limited, power supply interruptions may occur, and patient mobility needs must be met. The interface on this type of ventilator is much simpler than the interface found on a ventilator used in critical or subacute care. The availability of a low-pressure input port is an essential feature that allows supplemental O2 to be delivered by stationary or portable devices commonly used in the home, such as O2 concentrators, small high-pressure tanks, or portable liquid O2 reservoirs. Machine dimensions are also an important consideration, and this type of ventilator is generally compact in nature. The option to lock operator-set parameters minimizes the occurrence of inadvertent setting changes and concomitant alterations in alveolar ventilation and acid-base balance.

Ventilators used in home care and long-term care facilities require not only an internal battery for brief power interruptions but also connections for an external battery when the power supply is interrupted for extended periods as a result of natural disasters; man-made occurrences; or participation in academic, employment, or recreational activities. The Carina home ventilator (Draeger Medical) can provide invasive or noninvasive ventilation. Although the ventilator’s primary power source is 100-V or 240-V AC, the internal battery provides patients with approximately 2 hours of power. There is also an external battery pack that offers an additional 10-hour power supply when fully charged.

Transport Ventilators

Transport ventilators share attributes common to ventilators used in the home and critical care environments. It is necessary for this ventilator type to be lightweight, compact, durable, and maintained on a reliable power supply and to have low compressed gas consumption. The ventilator interface should be easy to navigate, allowing the clinician to set or change parameters before or during movement to and from a prescribed destination. Operator-set and monitored data should be visible under optimal conditions or conditions of low ambient light. These characteristics enhance patient safety and minimize the potential for complications or adverse effects to occur during air or ground transport. Monitoring is also an important consideration. Clinical practice guidelines recommend the level of monitoring during patient transport be analogous with that provided to the patient during stationary care.41,42 Modern transport ventilators provide the ability to display scalar waveforms and numerical data.

The BioMed Crossvent 4+ (BioMed Devices, Inc, Guilford, Connecticut) ventilator can be used to transport critically ill patients of any age, from infant to adult. This ventilator may be configured with a blender to allow for the delivery of a range of O2 concentrations from 21% to 100% or with an air-entrainment unit that delivers either 50% or 100% O2 without the need for an external air supply source. Pressure-controlled and volume-controlled ventilation may be provided in the CMV, IMV, or CSV mode. Tidal volumes may be adjusted from 5 to 2500 ml, and flow may be delivered at rates up to120 L/min. The internal battery offers 6 hours of uninterrupted power when fully charged. This unit is small (28 cm × 25.4 cm × 14 cm) and weighs less than 5 kg. The units main display screen is color and backlit for enhanced visibility.

Noninvasive Ventilators

Noninvasive ventilation is used across the continuum of care—from critical care to home care with individuals of any age. As previously mentioned, a noninvasive ventilation feature may be incorporated in critical care ventilators (e.g., SERVO-i and Puritan Bennett 840), subacute ventilators (Savina), and home care ventilators (Carina home). However, standalone noninvasive ventilators exist and are used extensively in various settings from the hospital to the home. In the acute and critical care setting, noninvasive ventilators have been used to reduce complications associated with diagnostic procedures, such as bronchoscopy,43 and in the treatment of acute respiratory insufficiency and respiratory failure44 and prevention of postextubation failure.45 This technology has also been associated with positive outcomes in the outpatient setting. The literature reports the use of noninvasive ventilation to restore and maintain adequate alveolar ventilation with individuals compromised by neuromuscular disorders, congestive heart failure, chronic obstructive lung disease, and sleep-disordered breathing.46

Features common to home and hospital grade units enhance patient comfort and promote adherence to therapy. Ramp time allows the clinician to program a delay in the initiation of a delivered inspiratory pressure. Ramp time is usually adjustable (e.g., 0 to 45 minutes), during which the patient breathes at a preset or operator-set expiratory pressure (e.g., 4 cm H2O). Likewise, rise time can be altered to reduce pressure overshoot and enhance breath delivery. An additional helpful tool is the ability to detect and quantify interface leak, estimated tidal volume, and minute ventilation delivery. Hospital-grade units have the capability to display patient data in graphic and numeric form. Clinicians are able to view pressure, flow, and scalar waveforms, such as on the BiPAP Vision (Respironics Inc, Murrysville, Pennsylvania). As with critical care ventilators, careful analysis of waveforms may assist clinicians in the identification and correction of patient-ventilator synchrony.

Summary Checklist

• Ventilators can be described in terms of their input power requirements (e.g., electrical or pneumatic) and how the input power is transformed into desired outputs of pressure, volume and flow.

• A key feature of a ventilator is the variety of modes of ventilation it offers.

• A mode of ventilation is a predetermined pattern of interaction with the patient. Modes are given many confusing names but they can be understood using a simple classification system.

• Mode classification is based on identifying 3 main components: the primary control variable, the breath sequence, and the targeting schemes used for mandatory and spontaneous breaths.

• Pressure control means that pressure delivery is predetermined by a targeting scheme such that inspiratory pressure is either proportional to patient effort or has a particular waveform regardless of respiratory system mechanics. Volume control means that inspiratory flow and volume delivery are predetermined by a targeting scheme to have particular waveforms independent of respiratory system mechanics.

• A spontaneous breath is one for which the timing and size of the breath is determined by the patient, i.e., inspiration is patient triggered and patient cycled. A mandatory breath is one for which the patient cannot determine the timing and/or size of the breath, i.e., inspiration is machine triggered and/or machine cycled.

• Spontaneous breaths may be assisted (meaning that the ventilator provides some portion of the work of breathing) or unassisted. Mandatory breaths are generally assisted.

• The breath sequence of a mode is the pattern of mandatory vs spontaneous breaths. Continuous spontaneous ventilation (CSV) means that all breath delivered by the ventilator are spontaneous. Intermittent mandatory ventilation (IMV) means that spontaneous breaths can occur between mandatory breaths. Continuous mandatory ventilation means that spontaneous breaths cannot occur between mandatory breaths.

• The targeting scheme is a description of the relation between operator settings and ventilator outputs for a mode. Currently, all modes can be classified as having one of 6 targeting schemes (set-point, dual, servo, adaptive, optimal, and intelligent).

• Ventilators can also be categorized by the settings in which they are used. Examples include: critical care, sub-acute care, home care, transport and noninvasive ventilators.