Analysis and Monitoring of Gas Exchange

Published on 01/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 5 (2 votes)

This article have been viewed 5201 times

Analysis and Monitoring of Gas Exchange

Mark S. Siobal

Ultimately, gas exchange takes place inside each of the body’s cells, where complex metabolic pathways use oxygen (O2) to create energy, while producing carbon dioxide (CO2) as a waste product. Although it is possible to analyze gas exchange at the cellular level, clinical focus normally is on gas exchange between the lungs and blood or between the blood and tissues. Gas exchange between the lungs and blood is usually analyzed by measuring O2 and CO2 levels in the arterial blood. Clinicians also can measure CO2 levels in the expired air to monitor ventilation. The most common approach to analyzing gas exchange between the blood and tissues is to measure O2 levels in the mixed venous (pulmonary artery [PA]) blood. This chapter focuses on these important concepts and the parameters that reflect gas exchange.

Invasive Versus Noninvasive Procedures

Invasive procedures require insertion of a sensor or collection device into the body, whereas noninvasive monitoring is a means of gathering data externally.1 Because laboratory analysis of gas exchange requires blood samples, it is usually considered invasive. Monitoring can be either invasive or noninvasive. Generally, invasive procedures tend to provide more accurate data than noninvasive methods, but they carry greater risk.

When both approaches are available, the need for measurement accuracy should dictate which is chosen. However, clinicians can sometimes combine the two approaches—using the invasive approach to establish accurate baseline information, while applying the noninvasive method for ongoing monitoring of a stable patient. After the gradient between the invasive and noninvasive method is established, trends in the change of the noninvasive method can be useful in making clinical decisions.

Measuring Fractional Inspired Oxygen

Analysis of gas exchange begins with knowledge of the system inputs—the inspired O2 and CO2 concentrations. Healthy individuals breathe air that contains a fixed O2 concentration (21%) and negligible amounts of CO2. Patients who are ill often have hypoxemia and are given supplemental O2. O2 analyzers are used to measure the fractional inspired O2 concentration (FiO2).

Instrumentation

Although many methods exist for measuring O2 concentrations, most bedside systems apply electrochemical principles. There are two common types of electrochemical O2 analyzers: (1) the polarographic (Clark) electrode and (2) the galvanic fuel cell. Under ideal conditions of temperature, pressure, and relative humidity, both types are accurate to within ± 2% of the actual concentration.1

The Clark electrode is similar to electrodes used in blood gas analyzers and transcutaneous monitors (see later section on Transcutaneous Blood Gas Monitoring). This system typically consists of a platinum cathode and a silver–silver chloride anode (Figure 18-1). O2 molecules diffuse through the sensor membrane into the electrolyte, where a polarizing voltage causes electron flow between the anode and cathode. While silver is oxidized at the anode, the flow of electrons reduces O2 (and water) to hydroxyl ions (OH) at the cathode. The more O2 molecules that are reduced, the greater is the electron flow across the poles (current). The resulting change in current is proportional to the PO2, with its value displayed on a galvanometer, calibrated in percent O2. Response times for Clark electrode O2 analyzers range from 10 to 30 seconds.

Most galvanic fuel cells use a gold anode and a lead cathode. In contrast to the Clark electrode, current flow across these poles is generated by the chemical reaction itself. Unless accessories such as alarms are included, a galvanic cell needs no external power; this means that galvanic cells respond more slowly than Clark electrodes, sometimes taking 60 seconds.

The Clark electrode and galvanic cell are suitable for basic FiO2 monitoring. When greater accuracy or faster response times are needed (e.g., when performing indirect calorimetry), a paramagnetic, zirconium cell, Raman scattering, or mass spectroscopy analyzer should be selected.

Problem Solving and Troubleshooting

Because O2 analyzers include replaceable components that deteriorate over time (batteries, electrodes, membranes, electrolytes), the best way to avoid problems is through preventive maintenance, which should include both scheduled parts replacement and routine operational testing by biomedical engineering personnel. As with any preventive maintenance program, it is essential that detailed records be kept on each piece of equipment.

Even with the best preventive maintenance, O2 analyzers sometimes malfunction. The clinician would know that an analyzer is not working if it fails to calibrate or gives an inconsistent reading during use. The most common causes of analyzer malfunction are low batteries (Clark electrode systems), sensor depletion, and electronic failure. Because a low battery condition is so common with Clark electrode systems, the first step in troubleshooting is to replace the batteries. If the analyzer still does not calibrate on fresh batteries, the problem is probably a depleted sensor. With most analyzers, a depleted sensor must be replaced (some Clark electrodes can be recharged). If an analyzer still fails to calibrate after battery and sensor replacement, the most likely problem is an internal failure of its electrical system. In this case, the device should be taken out of service and repaired.

Inaccurate readings also can occur with electrochemical analyzers, resulting from either condensed water vapor or pressure fluctuations. Galvanic cells are particularly sensitive to condensation. To avoid this problem during continuous use in humidified circuits, the clinician should place the analyzer sensor proximal to any humidification device.

Fuel cell and Clark electrode readings also are affected by ambient pressure changes. Under conditions of low pressure (high altitude), these devices read lower than the actual O2 concentration. Conversely, higher pressures, such as pressures that occur during positive pressure ventilation, cause these devices to read higher than the actual FiO2. These observations are consistent with the fact that both devices measure the PO2 but report a percent concentration scale.

Sampling and Analyzing Blood Gases

In the clinical setting, it is common for the collection of blood specimens (sampling) to be performed separately from their analysis. Each procedure involves different knowledge and skill. For these reasons, these topics are covered separately.

Sampling

Clinicians have been using blood samples to assess gas exchange parameters for more than 50 years.2 The definition of respiratory failure still is based largely on blood gas measurements. Depending on the need, blood gas samples can be obtained by percutaneous puncture of a peripheral artery, from an indwelling catheter (arterial, central venous, or PA), or by capillary sampling.

Arterial Puncture and Interpretation

Results obtained from sampling arterial blood gas (ABG) are the cornerstone in the diagnosis and management of oxygenation and acid-base disturbances. ABGs are considered the “gold standard” of gas exchange analysis, against which all other methods are compared.

Arterial puncture involves drawing blood from a peripheral artery (radial, brachial, femoral, or dorsalis pedis) through a single percutaneous needle puncture (Figure 18-2). The radial artery is the preferred site for arterial blood sampling for the following reasons:

Other sites (brachial, femoral, and dorsalis pedis) are riskier and should be used only by clinicians specifically trained in their use. Likewise, arterial puncture in infants (through either the radial or the temporal artery) requires advanced training. Arterial cannulation sites with indwelling catheters include radial, brachial, femoral, dorsalis pedis, umbilical (in neonates), and axillary arteries. The focus here is on radial artery puncture.

To guide practitioners in providing quality care, the American Association for Respiratory Care (AARC) has published Clinical Practice Guideline: Sampling for Arterial Blood Gas Analysis.3 Complementary recommendations have been published by the National Committee for Clinical Laboratory Standards.4 Modified excerpts from the AARC guideline appear in Clinical Practice Guideline 18-1.

18-1   Sampling for Arterial Blood Gas Analysis

AARC Clinical Practice Guidelines (Excerpts)*

Contraindications

• Abnormal results of a modified Allen test (lack of collateral circulation) may be indicative of inadequate blood supply to the hand and suggest the need to select another puncture site.

• Arterial puncture should not be performed through a lesion or distal to a surgical shunt. For example, arterial puncture should not be performed on a patient undergoing dialysis. If there is evidence of infection or peripheral vascular disease involving the selected limb, an alternative site should be selected.

• Because of the need for monitoring the femoral puncture site for an extended period, femoral punctures should not be performed outside the hospital.

• Coagulopathy or medium-dose to high-dose anticoagulation therapy, such as heparin or warfarin (Coumadin), streptokinase, and tissue plasminogen activator (but not aspirin), may be a relative contraindication.


*For the complete guideline, see American Association for Respiratory Care: Clinical practice guideline: sampling for arterial blood gas analysis. Respir Care 37:891, 1992.

Equipment

Box 18-1 lists the equipment needed to perform an arterial puncture. Commercial vendors provide kits containing most of the equipment listed. If provided, the needle capping device serves two purposes. First, it isolates the sample from air exposure (to ensure accurate results). Second, it helps prevent inadvertent needlestick injuries. There are many different capping device designs; devices that allow single-handed recapping are preferred. If a capping safety device is not provided, the clinician should use the single-handed “scoop” method to cap the needle before removing it and plugging the syringe.

Procedure

Box 18-2 outlines the basic procedure for radial artery puncture of adults. Before radial artery puncture is performed, a modified Allen test (Figure 18-3) is recommended. The test is normal (indicating adequate collateral circulation) if the palm, fingers, and thumb flush pink within 5 to 10 seconds after pressure on the ulnar artery is released. A normal test result indicates the presence of collateral circulation but may not predict the development of complications after radial artery puncture or cannulation.

Box 18-2

Procedure for Radial Artery Puncture

• Check the medical record to (1) confirm the order and indications and (2) determine the patient’s primary diagnosis, history (especially bleeding disorders or blood-borne infections), current status, respiratory care orders (especially oxygen therapy or mechanical ventilation), and anticoagulant or thrombolytic therapy.

• Confirm steady-state conditions (20-30 minutes after changes).

• Obtain and assemble necessary equipment and supplies.

• Wash hands and don barrier protection (e.g., gloves, eyewear).

• Identify the patient using current patient safety standards.

• Explain the procedure to the patient.

• Position the patient, extending the patient’s wrist to approximately 30 degrees.

• Perform a modified Allen test, and confirm collateral circulation.

• Clean site thoroughly with 70% isopropyl alcohol or an equivalent antiseptic.

• Inject a local anesthetic subcutaneously and periarterially (wait 2 minutes for effect).*

• Use a preheparinized blood gas kit syringe, or heparinize a syringe and expel the excess (fill dead space only).

• Palpate and secure the artery with one hand.

• Insert the needle, bevel up, through the skin at a 45-degree angle until blood pulsates into the syringe.

• Allow 1 ml of blood to fill syringe (the need to aspirate indicates a venous puncture).

• Apply firm pressure to puncture site with sterile gauze until the bleeding stops.

• Expel any air bubbles from the sample, and cap or plug the syringe.

• Mix the sample by rolling and inverting the syringe.

• Place the sample in a transport container (ice slush) if specimen is not to be analyzed within 10-30 minutes.

• Dispose of waste materials and sharps properly.

• Document the procedure and patient status in the chart and on the specimen label.

• Check the site after 20 minutes for hematoma and adequacy of distal circulation.


*Optional.

From Malley WJ: Clinical blood gases: application and noninvasive alternatives, Philadelphia, 1990, Saunders; Shapiro BA, Peruzzi WT, Kozelowski-Templin R: Clinical application of blood gases, ed 5, St Louis, 2005, Mosby.

The modified Allen test has been a widely used clinical method to assess adequacy of ulnar artery collateral blood flow despite the lack of evidence that it can predict ischemic complications in the setting of complete radial artery occlusion.5 The criteria for an abnormal test result are not agreed on, which renders the significance of an abnormal test unclear. The test result may be inaccurate in predicting postcannulation hand ischemia, has poor interrater reliability, and is known to yield a high incidence of false normal and abnormal results. The modified Allen test cannot be performed on most critically ill patients who are either uncooperative or unconscious. In addition, prior radial artery cannulation, severe circulatory insufficiency, wrist or hand burns, or jaundice makes interpreting the results difficult. Performance of a modified Allen test before radial artery puncture or cannulation should not be considered a “standard of care,” but the need for its use and appropriate application should be well recognized.6

In patients who have undergone previous radial artery cannulation, the modified Allen test can provide documentation of possible arterial thrombosis and should be used to direct catheter placement. In that circumstance, it is imprudent to ignore totally the utility of the modified Allen test, especially if another arterial site is available for cannulation.7

In most cases, a sample volume of 0.5 to 1 ml of blood is adequate. The actual sample volume needed depends on the following: (1) the anticoagulant used, (2) the requirements of the specific analyzer used, and (3) whether other tests are to be performed on the sample.

The following rules for careful handling of the needle help avoid transmission of blood-borne diseases:

Problem Solving and Troubleshooting

There are two major problem areas associated with arterial puncture. The first problem area involves difficulties in getting a good sample. The second problem area involves preanalytic error.

Getting a Good Sample

Problems with getting a good sample include an inaccessible artery, absent pulse, deficient sample return, and alteration of test results caused by the patient’s response. If the selected artery cannot be located, another site should be considered. Likewise, if an adequate pulse cannot be palpated at the chosen site, another site should be selected, or an acceptable noninvasive approach should be considered as an alternative (e.g., pulse oximetry).

If the clinician gets only a small spurt of blood, the needle has probably passed through the artery. In this situation, the needle is slowly withdrawn until a pulsatile flow fills the syringe. The tip of the needle is never redirected without it first being withdrawn to the subcutaneous tissue. If the needle must be withdrawn completely and the clinician does not have an adequate sample, the procedure is repeated with a fresh blood gas kit.

Small sample volumes or the need to apply syringe suction also may indicate that venous blood has been obtained. However, when drawing arterial blood from hypotensive patients or when using small needles (<23-gauge), the clinician may need to pull gently on the syringe barrel. Excessive suction can alter the blood gas results. If the clinician suspects that pain or anxiety during the procedure may have altered the results (most typically causing hyperventilation), he or she should consider using a local anesthetic for subsequent sampling attempts.

Preanalytic Error

Preanalytic errors are problems occurring before sample analysis that can alter the accuracy of the blood gas results. Table 18-1 summarizes the most common errors associated with arterial blood sampling, including recommendations on how to recognize and avoid these problems.8,9 Clinicians can avoid most preanalytic errors by ensuring that the sample is obtained anaerobically, is properly anticoagulated (with immediate expulsion of air bubbles), and is analyzed within 10 to 30 minutes.

TABLE 18-1

Preanalytic Errors Associated With Arterial Blood

Error Effect on Parameters How to Recognize How to Avoid
Air in sample ↓ PCO2 Visible bubbles or froth Discard frothy samples
  ↑ pH Low PCO2 inconsistent with patient status Fully expel bubbles
  ↑ low PO2   Mix only after air is expelled
  ↓ high PO2   Cap syringe quickly
Venous admixture ↑ PCO2 Failure of syringe to fill by pulsations Avoid brachial and femoral sites
  ↓ pH Patient has no symptoms of hypoxemia Do not aspirate sample
  Can greatly lower PO2   Use short-bevel needles
      Avoid artery “overshoot”
      Cross-check with SpO2
Excess anticoagulant (dilution) ↓ PCO2
↑ pH
↑ low PO2
↓ high PO2
Visible heparin remains in syringe before sampling Use premade lyophilized (dry) heparin blood gas kits
Fill dead space only
Collect >2 ml (adults) and >0.6 ml (infants)
Metabolic effects ↑ PCO2
↓ pH
↓ PO2
Excessive time lag since sample collection
Values inconsistent with patient status
Analyze within 15 min
Place sample in ice slush

image

The traditional method used to avoid preanalytic errors caused by blood cell metabolism is to chill the sample quickly by placing it in an ice slush. Chilling is needed if the sample is not to be analyzed within 10 to 30 minutes.3 Chilled samples should be discarded if they are not analyzed within 60 minutes. PaO2 of samples drawn from subjects with elevated white blood cell counts may decrease rapidly, and immediate chilling is recommended. Chilled samples can result in potassium transport between blood cells and plasma and can result in erroneous elevation in potassium measured from a blood gas sample. Use of a glass syringe or a plastic syringe with low diffusibility minimizes the risk of room air gases contaminating the sample. Pneumatic tube transport of samples containing small air bubbles can have a noticeable effect on increasing PaO2.9

Interpretation of Arterial Blood Gases

Given that gas exchange is a dynamic process, looking at results from a single blood sample is akin to looking at a single frame in a feature-length movie. If the scene is changing rapidly, the single frame can be misleading. Conversely, if the scene is relatively stable, a single frame can provide useful information. Blood gas results must be interpreted in light of the patient status at the time the sample was obtained.

Any major change in either patient condition or therapy disrupts the patient’s steady state. However, over time, a steady state normally returns. The time needed to restore a steady state varies with the patient’s pulmonary status. Patients with healthy lungs achieve a steady state in only 5 minutes after changes, whereas patients with chronic obstructive pulmonary disease (COPD) may require 20 to 30 minutes. If a patient’s FiO2 is changed, the measured PaO2 would accurately reflect the patient’s gas exchange status within 5 minutes in healthy individuals but may require 20 to 30 minutes in patients with COPD.

To document the patient’s status, the following need to be recorded: (1) date, time, and site of sampling; (2) results of the modified Allen test, when performed; (3) patient’s body temperature, position, activity level, and respiratory rate; and (4) FiO2 concentration or flow and all applicable ventilatory support settings. Noting such information may prove useful in interpretation of the results.

In the first step of interpretation of the results, the clinician must ensure he or she is looking at the results of the correct patient. The name and patient identification number from the blood gas report must match the patient. Interpretation of the results can be divided into two basic steps: (1) interpretation of the oxygenation status and (2) interpretation of the acid-base status.

The oxygenation status is determined by examination of the PaO2, arterial O2 saturation (SaO2), and arterial O2 content (CaO2) The PaO2 represents the partial pressure of O2 in the plasma of the arterial blood and is the result of gas exchange between the lung and blood. The PaO2 is reduced in various settings but most often when lung disease is present. PaO2 of less than 40 mm Hg is called severe hypoxemia, PaO2 of 40 to 59 mm Hg is called moderate hypoxemia, and PaO2 of 60 mm Hg to the predicted normal is called mild hypoxemia.

SaO2 represents the degree to which the hemoglobin (Hb) is saturated with O2 (see Chapter 11). Normally, the Hb saturation with O2 is 95% to 100% with healthy lungs. When the lungs cannot transfer O2 into the blood at normal levels, the SaO2 decreases in most cases in proportion to the degree of lung disease present. Blood gas analyzers report a calculated SaO2. Measurement of SaO2 by hemoximetry and Hb content is required for accurate determination of CaO2.

CaO2 represents the content of O2 in 100 ml of arterial blood and is a function of the amount of Hb present and the degree to which it is saturated. A normal CaO2 is 18 to 20 ml of O2 per 100 ml of arterial blood. A reduced CaO2 is often the result of low PaO2 and SaO2, reduced Hb level, or both.

The acid-base status of the patient is determined by evaluating the pH, PaCO2, and plasma HCO3. The steps for interpreting the acid-base status of the ABG results are described in Chapter 13.

Indwelling Catheters (Arterial and Central Venous Pressure and Pulmonary Artery Lines)

Indwelling catheters provide ready access for blood sampling and allow continuous monitoring of vascular pressures, without the traumatic risks associated with repetitive percutaneous punctures. However, infection and thrombosis are more likely with indwelling catheters than they are with intermittent punctures.

The most common routes for indwelling vascular lines are a peripheral artery (usually radial, brachial, or less commonly dorsalis pedis and axillary) or femoral artery, a central vein (usually the vena cava), and the PA. In neonates, the umbilical artery is cannulated for arterial blood sampling. Table 18-2 summarizes the usefulness of these various sites in providing relevant clinical information. Chapter 46 provides details on the use of these systems for hemodynamic pressure and flow monitoring.

TABLE 18-2

Common Sites for Indwelling Vascular Catheters and the Information They Provide

Location Blood Collection Pressure Monitoring
Sample Reflects Pressure Reflects
Peripheral, umbilical artery Arterial blood Pulmonary gas exchange (O2 uptake/CO2 removal) Systemic arterial pressure LV afterload; vascular tone; blood volume
Central vein Venous blood (unmixed) Not useful for assessing gas exchange; can be used for some other laboratory tests CVP Fluid volume; vascular tone; RV preload
Pulmonary artery Mixed venous blood (balloon deflated) Gas exchange at tissues (O2 consumption/CO2 production) PAP; PCWP RV afterload; vascular tone; blood volume; LV preload

image

CVP, Central venous pressure; LV, left ventricular; PAP, pulmonary artery pressure; PCWP, pulmonary capillary wedge pressure; RV, right ventricular.

Equipment

Figure 18-4 shows the basic setup used for an indwelling vascular line, in this case, a brachial artery catheter. The catheter connects to a disposable continuous-flush device (Delta-flow; Utah Medical Products, Midvale, UT). This device keeps the line open by providing a continuous low rate of flow (2 to 4 ml/hr) of intravenous saline solution through the system.

Heparinized saline flush solution has been commonly used with indwelling vascular catheters. However, it has been shown that heparin does not significantly improve arterial catheter function, extend the duration of use, or decrease the number of manipulations required. Additionally, results of coagulation studies can be affected by heparinized flush solution, and unnecessary exposure to heparin may increase the risk of heparin-induced thrombocytopenia.10 Because arterial pressures are much higher than venous pressures, the intravenous bag supplying these systems must be pressurized, usually by using a hand bulb pump. A strain-gauge pressure transducer connected to the flush device provides an electrical signal to an amplifier or monitor, which displays the corresponding pressure waveform.

Procedure

Access for sampling blood from most intravascular lines is provided by a three-way stopcock (Figure 18-5). Equipment and supplies are the same as specified for arterial puncture, with the addition of a second “waste” syringe. Box 18-4 outlines the proper procedure for taking an arterial blood sample from a three-way stopcock system.

Box 18-4   Procedure for Sampling Arterial Blood from an Indwelling Catheter

• Check the medical record to affirm order (as per arterial puncture).

• Confirm steady-state conditions (20-30 minutes after changes).

• Obtain and assemble needed equipment and supplies.

• Wash hands and don barrier protection (e.g., gloves, eyewear).

• Identify the patient using current patient safety standards.

• Explain the procedure to the patient.

• Attach the waste syringe to the stopcock port.

• Position the stopcock so that blood flows into the syringe and the IV bag port is closed.

• Aspirate at least 1-2 ml, or five to six times the tubing volume, of fluid or blood.

• Reposition the stopcock handle to close off all ports.

• Disconnect and properly discard waste syringe.

• Attach new heparinized syringe to the sampling port.

• Position the stopcock so that blood flows into the sample syringe and the IV bag port is closed.

• Fill syringe with 1 ml of blood.

• Reposition the stopcock handle to close off the sampling port and open the IV bag port.

• Disconnect the syringe, expel air bubbles from sample, and cap or plug the syringe.

• Flush the line and stopcock with the IV solution.

• Mix the sample by rolling and inverting the syringe.

• Confirm that the stopcock port is open to the IV bag solution and catheter.

• Confirm undampened pulse pressure waveform on the monitor graphic display.

• Place the sample in a transport container (ice slush) if specimen is not to be analyzed within 10-30 minutes.

• Dispose of waste materials properly.

• Document the procedure and patient status in the chart and on the specimen label.

IV, Intravenous.

The procedure is slightly different when obtaining mixed venous blood samples from PA catheters because PA catheters have separate sampling and intravenous infusion ports and a balloon at the tip is used to measure pulmonary capillary wedge pressure. The clinician must ensure that the balloon is deflated and withdraw the sample slowly (e.g., approximately 3 ml/min or 1 ml in 20 seconds). If the clinician fails to deflate the balloon or withdraws the sample too quickly, the venous blood may be “contaminated” with blood from the pulmonary capillaries. The result is always a falsely high O2 level. In addition, close attention must be paid to the infusion rate through the catheter. Rapid flow of IV fluid can dilute the blood sample and affect O2 content measurements.

Capillary Blood Gases

Capillary blood gas sampling is used as an alternative to direct arterial access in infants and small children. Properly obtained capillary blood from a well-perfused patient can accurately reflect and provide clinically useful estimates of arterial pH and PCO2 levels.6 However, capillary PO2 is of no value in estimating arterial oxygenation, and O2 saturation by pulse oximetry must also be evaluated when a capillary blood gas sample is obtained. Respiratory therapists (RTs) must exercise extreme caution when using capillary blood gases to guide clinical decisions. Direct arterial access is still the preferred approach for assessing gas exchange in infants and small children with severe acute respiratory failure.

Capillary blood values are meaningful only if the sample site is properly warmed. Warming the skin (to approximately 42° C) causes dilation of the underlying blood vessels, which increases capillary flow well above tissue needs. Blood gas values resemble the values in the arterial circulation; this is why a sample obtained from a warmed capillary site is often referred to as arterialized blood. It has been shown that capillary blood samples from the earlobe reflect arterial PCO2 and PO2 better than samples drawn from a finger stick.11 The posterior medial or lateral curvature of the heel is the recommended site for capillary puncture specimens in infants less than 1 month old to avoid nerve and bone damage.

To guide practitioners in providing quality care, the AARC has published Clinical Practice Guideline: Capillary Blood Gas Sampling for Neonatal and Pediatric Patients.12 Modified excerpts from the AARC guideline appear in Clinical Practice Guideline 18-2.

18-2   Capillary Blood Gas Sampling for Neonatal and Pediatric Patients

AARC Clinical Practice Guideline (Excerpts)*

Contraindications

Capillary punctures should not be performed at or through the following:

Capillary punctures should not be performed:

Relative contraindications include:


*For the complete guideline, see American Association for Respiratory Care: Clinical practice guideline: capillary blood gas sampling for neonatal and pediatric patients. Respir Care 46:506, 2001.

Procedure

Box 18-5 outlines the basic procedure for capillary blood sampling. The most common site for sampling is the heel, specifically the lateral aspect of the plantar surface.

Analyzing

The primary analytes or parameters of pH, PCO2, and PO2 in a blood sample are measured with a blood gas analyzer. Typically, analyzers use these measures to compute several secondary values, such as plasma bicarbonate, base excess or deficit, and Hb saturation. If actual measurement of total Hb saturation (oxyhemoglobin [HbO2], methemoglobin [metHb], and carboxyhemoglobin [HbCO]) is required, the sample usually must be analyzed separately using hemoximetry (see p. 401). Some analyzers combine the blood gas and hemoximetry measurements, which may require a larger sample size (usually 100 mcl).

Blood gas analysis and hemoximetry are moderately complex laboratory procedures. Clinicians performing these tests must have documented training and must demonstrate proficiency in performing the procedures, preventive maintenance, troubleshooting, and instrument calibration. In addition, clinicians must be skilled in validating test results using rigorous quality control methods.14

To guide practitioners in providing quality care, the AARC has published Clinical Practice Guideline: Blood Gas Analysis and Hemoximetry.15 Related recommendations have been published by the National Committee for Clinical Laboratory Standards.9 Modified excerpts from the AARC guideline appear in Clinical Practice Guideline 18-3.

18-3   Blood Gas Analysis and Hemoximetry

AARC Clinical Practice Guideline (Excerpts)*

Monitoring

The following aspects of analysis should be monitored, and corrective action should be taken as indicated:

• Presence of air bubbles or clots in the specimen, with evacuation before mixing and sealing the syringe

• Assurance that a continuous sample is aspirated (or injected) into the analyzer and that all the electrodes are covered by the sample (confirmed by direct visualization if possible)

• Assurance that 8-hour quality control and calibration procedures have been completed and that instrumentation is functioning properly before patient sample analysis

• Assurance that the specimen was properly labeled, stored, and analyzed within an acceptable period

• Participation in an accredited (recognized) proficiency testing program

• As part of any quality assurance program, indicators must be developed to monitor potential sources of error:


*For the complete guideline, see American Association for Respiratory Care: Clinical practice guideline: in vitro pH and blood gas analysis and hemoximetry. Respir Care 46:498, 2001.

Instrumentation

Many instrumentation companies manufacture laboratory blood gas analyzers. Although available in a range of designs, these devices typically share the following key components:

Measurement of the three primary parameters—pH, PCO2, and PO2—is accomplished using three separate electrodes. To measure PO2, blood gas analyzers use the Clark polarographic electrode (see Figure 18-1).

The pH electrode consists of two electrodes or half-cells (Figure 18-6). The measuring half-cell contains a silver–silver chloride rod surrounded by a solution of constant pH and enclosed by a pH-sensitive glass membrane. As the sample passes this membrane, the difference in H+ concentration on either side of the glass changes the potential of the measuring electrode. The reference half-cell (here mercury–mercurous chloride) produces a constant potential, regardless of sample pH. The difference in potential between the two electrodes is proportional to the H+ concentration of the sample, which is displayed on a voltmeter calibrated in pH units.

To measure PCO2, blood gas analyzers use the Severinghaus electrode, which is essentially a pH electrode exposed to an electrolyte solution that is in equilibrium with the sample through a CO2-permeable membrane. As CO2 diffuses through this membrane into the electrolyte solution, it undergoes the following hydration reaction:

< ?xml:namespace prefix = "mml" />CO2+H2OH2CO3H++HCO3

image

The greater the partial pressure of CO2, the more H+ produced by this reaction, and the more pH of the electrolyte solution changes. The measuring electrode detects the pH change as a change in electrical potential, which is proportional to the PCO2 of the sample.

In addition to electrochemical electrodes, a sensor technology based on optical fluorescence and the process of photoluminescence is available to measure pH, PCO2, and PO2. Optical fluorescence sensors use fluorescent dyes that are illuminated with light of a specific wavelength. The illuminating light is transmitted, reflected, absorbed, and reemitted to varying degrees depending on the concentration of O2, CO2, and hydrogen ions. These photochemical alterations in light illumination are used to calculate pH, PCO2, and PO2 values. Compared with electrochemical sensors, optical sensors do not require electrical connections, making them less affected by electrical interference and drift. The sensor reactions are reversible so that the analyte is not consumed. These features make them easier to miniaturize for intravascular measurement of blood gases.16,17

Procedure

To provide accurate and clinically useful data, blood gas analysis must be performed as follows:

Prior discussion addressed how to avoid preanalytic errors. Subsequent discussion focuses on blood gas quality control and key elements involved in the analysis procedure.

Box 18-6 outlines the steps commonly used in most established procedures for laboratory blood gas analysis. One should always refer to the manufacturer’s literature for the particular steps to use with a specific analyzer.

Rigorous application of U.S. Centers for Disease Control and Prevention (CDC) standard precautions is essential. In addition, the Occupational Safety and Health Administration (OSHA) requires personnel to wear gloves when handling all laboratory specimens. Although splashes are rare during analysis, some manufacturers provide extra protection by mounting splash shields on their instruments. If splashes are anticipated, the operator can wear a face shield.

Waste fluids are potentially infectious and should be handled as if they were blood samples. In addition, the National Committee for Clinical Laboratory Standards recommends adding a strong disinfectant, such as 2% glutaraldehyde or a 1 : 4 solution of sodium hypochlorite, to the waste container of the instrument either during use or before disposal.

Quality Assurance

Quality patient care depends on consistently accurate blood gas results. Modern laboratory analyzers are often automated, computer-controlled, self-calibrating systems. This sophistication has led to the false assumption that accurate results are “automatic,” with clinicians needing only to input the sample properly and record the results. Nothing could be further from the truth. As with all diagnostic laboratory procedures, the accuracy of blood gas testing depends on rigorous quality control.

The Clinical Laboratory Standards Institute (CLSI), formerly the National Committee for Clinical Laboratory Standards (NCCLS), establishes guidelines and standards for blood gas analysis and quality assurance. Government regulatory agencies collaborate to update the Clinical Laboratory Improvement Amendments (CLIA) that establish proficiency testing requirements.18 Although an in-depth review of laboratory quality control is beyond the scope of this text, all RTs must understand the key elements.19

Figure 18-7 depicts the key components of laboratory quality control. A brief description of each element follows.

Automated Calibration

Calibration is the only fully automated element of blood gas quality control. Blood gas analyzers regularly calibrate themselves by adjusting the output signal of each electrode when exposed to media having known values. In most units, the media used to calibrate the gas electrodes are precision mixtures of O2 and CO2. For the pH electrode, standard pH buffer solutions are used. Calibration media must meet the requirements set by nationally recognized standards organizations. Users are responsible for ensuring that calibration media are properly stored and that in-use life and expiration dates are strictly enforced.

Calibration is performed to ensure that the output of the analyzer is both accurate and linear across the range of measured values. Parameters must be measured with known input values representing at least two points, usually a low and a high value. Figure 18-8 shows a typical two-point calibration procedure. In this example, the instrument’s initial precalibration response indicates that the output readings are consistently higher than the actual input, with this positive bias worsening at higher levels. Calibration is performed first by adjusting the offset (or balance) of the instrument so that the low output equals the low input (in this case zero). Next, the gain (or slope) of the device is adjusted to ensure that the high output equals the high input. When both offset and gain are adjusted against known inputs, the instrument is properly calibrated and can undergo calibration verification with control samples.

Calibration Verification by Control Media

Calibration verification establishes and periodically confirms the validity of blood gas analyzer results. Calibration verification requires analysis of at least three materials with known values spanning the entire range of values expected for clinical samples. Ideally, these materials, called controls, should mimic real blood samples chemically and physically. Because requirements for use of control media currently vary among regulatory agencies, users should consult the applicable regulations directly. As a general recommendation, at least two levels of control media should be analyzed during every 8-hour shift. Rotation among the three levels should ensure that all three levels are analyzed at least once every 24 hours.

Internal Statistical Quality Control

Internal quality control takes calibration verification a step further by applying statistical and rule-based procedures (Westgard rules)20,21 to help detect, respond to, and correct instrument error. In one common approach, the results of control media analyses are plotted on a graph and compared with statistically derived limits, usually ±2 standard deviation (SD) ranges (Figure 18-9). Control results that fall outside these limits indicate analytic error.

There are two categories of analytic error: (1) random error and (2) systematic error. Random error is observed when sporadic, out-of-range data points occur (see Figure 18-9, point A). Random errors are errors of precision or, more precisely, imprecision. Conversely, either a trending or an abrupt shift in data points outside the statistical limits (see Figure 18-9, point B) is sometimes observed. This phenomenon is called systematic error or sometimes bias. Bias plus imprecision equals total instrument error, or inaccuracy. Table 18-3 outlines the major factors causing these two types of error and suggests some common corrective actions.

TABLE 18-3

Correction of Analytic Errors

Error Type Common Contributing Factors Common Corrective Actions
Imprecision (random) errors Statistical probability Rerun control
  Sample contamination Repeat analysis on different instrument
  Sample mishandling  
Bias (systematic) errors Contaminated buffers Perform function check of suspected problem area
  Incorrect gas concentrations Repair or replace failed components
  Incorrect procedures  
  Component failure  

External Quality Control (Proficiency Testing)

The federal government mandated a rigorous program of external quality control for analytic laboratories. CLIA were established in 1988. To meet these standards, analytic laboratories must undergo regular proficiency testing designed to evaluate their operating procedures and the competence of their personnel.22 Proficiency testing requires analysis and reporting on externally provided control media with unknown values, usually three times per year, with five samples per test. There are many CLIA-approved proficiency testing providers.23 A commonly used provider is the College of American Pathologists (CAP) proficiency testing survey. Proficiency testing survey analyses must be performed along with the regular workload by the personnel routinely responsible for testing, following the laboratory’s standard testing practices.

Criteria for acceptable performance specify a range around a target value, such as ±0.04 for pH. A single incidence of unsatisfactory performance requires documentation of remedial action. Multiple or recurring incidences of poor performance can result in severe sanctions, including suspension of Medicare and Medicaid reimbursement or the loss of the laboratory’s operating license and accreditation.

Point-of-Care Testing

Point-of-care testing takes blood gas analysis from the specialized laboratory to the patient’s bedside.24 Point-of-care testing reduces turnaround time, which should improve care and reduce costs. Theoretically, cost savings can be accrued by eliminating delays in therapy, decreasing patient length of stay in the hospital and emergency department.25 Additional cost savings may occur if point-of-care testing decreases the need for specialized laboratory personnel. Point-of-care testing is used increasingly in the hospital and physician office settings.26

Instrumentation

Figure 18-10 shows a typical point-of-care blood gas analyzer (GEM 4000; Instrumentation Laboratory, Bedford, MA). In addition to blood gas analysis, such devices can be used to measure several chemistry and hematology parameters, including serum electrolytes, blood glucose levels, blood urea nitrogen, hematocrit, hemoximetry, lactate, bilirubin, and prothrombin and partial thromboplastin times.

These devices are portable, and some can perform 900 tests on a single set of batteries. They typically include a display screen for accessing menu functions and viewing results. Most devices include a simple keypad or touch screen for data and command entry. Analysis occurs using disposable cartridges or inside a chamber in the body of the unit.

Some devices employ single-use sample cartridges that differ according to the array of tests being performed. Each cartridge contains the necessary calibration solution, a sample handling system, a waste chamber, and miniaturized electrochemical or photochemical sensors. The cartridge system requires no operator oversight because it is self-calibrating and disposable after a single use. After self-calibration and introduction of the sample into the cartridge, the sensors measure the concentration of the analytes and conduct their output signal through conductive contact pads to the analyzer microprocessor. Test results usually are ready within 90 seconds. Waste management involves simple removal and proper disposal of the analysis cartridge.

Other devices use self-contained multiuse cartridge packs that include all testing components, are maintenance-free, and incorporate automated quality control management systems. Multiuse cartridges are typically replaced every 30 days or when testing components are used up.

Blood Gas Monitoring

A blood gas monitor is a bedside tool (usually dedicated to a single patient) that can provide measurements either continuously or at appropriate intervals without permanently removing blood from the patient. Four systems are in current clinical use: (1) transcutaneous blood gas monitor, (2) intraarterial (in vivo) blood gas monitor, (3) extraarterial (ex vivo) blood gas monitor, and (4) tissue O2 monitor.

Transcutaneous Blood Gas Monitoring

Transcutaneous blood gas monitoring provides continuous, noninvasive estimates of arterial PO2 and PCO2 through a surface skin sensor. As with capillary sampling, the device arterializes the underlying blood by heating the skin. Warming also increases the permeability of the skin to O2 and CO2, which allows them to diffuse more readily from the capillaries to the sensor, where they are measured as transcutaneous partial pressures (PtcO2 and PtcCO2).

Numerous factors influence the agreement between arterial blood and transcutaneous gas measurements, with O2 levels being affected most. The two most important factors are age and perfusion status. Table 18-4 summarizes these relationships using the ratio of PtcO2 to PaO2. A ratio of 1 : 1 indicates “perfect” agreement between PtcO2 and PaO2. As can be seen, this level of agreement occurs only in neonates.

TABLE 18-4

Ratios Correlating PtcO2 With PaO2

Age Group PtcO2/PaO2 Ratio Perfusion Status PtcO2/PaO2
Premature infants 1.14 : 1 Stable 0.79 : 1
Neonates 1.00 : 1 Moderate shock 0.48 : 1
Children 0.84 : 1 Severe shock 0.12 : 1
Adults 0.79 : 1    
Older adults 0.68 : 1    

image

From Tobin MJ: Respiratory monitoring. JAMA 264:244–251, 1990.

In terms of age, the younger the patient, the better is the agreement between PaO2 and PtcO2; this is mainly because of age-related differences in skin composition. With regard to perfusion status, PaO2 and PtcO2 are similar only in patients with normal cardiac output and fluid balance because accurate transcutaneous measures require adequate skin perfusion. Low cardiac output, shock, and dehydration all cause peripheral vasoconstriction and impair capillary flow, which decreases the PtcO2 level. Some clinicians use PtcO2 not to monitor oxygenation as a surrogate for PaO2 but to assess blood flow changes during procedures such as vascular surgery and resuscitation.

Agreement between PaCO2 and PtcCO2 is better because CO2 is more diffusible. PaCO2 changes of 5 mm Hg can be monitored or “trended” by transcutaneous blood gas analysis.

Based on these factors, PtcCO2 monitoring is a reasonable choice when there is a need for continuous, noninvasive analysis of trends in ventilation and PaCO2. In hemodynamically stable infants and children, PaO2 can be “correlated” against PtcO2, decreasing the need for repeated arterial samples. Because pulse oximetry cannot provide accurate estimates of excessive blood O2, the transcutaneous monitor may be useful for monitoring hyperoxia in neonates. However, prevention of hyperoxia in premature neonates is more often achieved by maintaining pulse oximetry saturation between 85% and 93%.31

Transcutaneous blood gas monitoring of PtcCO2 can also be useful in adult patients during deep sedation and mechanical ventilation in the emergency department and intensive care unit and during surgery.3237 PtcCO2 is a more accurate reflection PaCO2 than both PetCO2 and nasal etCO2 in intubated and spontaneously breathing adult patients. The use of PtcCO2 in conjunction with pulse oximetry reduces the need for repeated arterial blood gas sampling.38

To guide practitioners in providing quality care, the AARC has published Clinical Practice Guideline: Transcutaneous Blood Gas Monitoring for Neonatal and Pediatric Patients.39 Modified excerpts from the AARC guideline appear in Clinical Practice Guidelines 18-4.

18-4   Transcutaneous Blood Gas Monitoring for Neonatal and Pediatric Patients

AARC Clinical Practice Guideline (Excerpts)*

Monitoring

The schedule of patient and equipment during transcutaneous monitoring should be integrated into assessment of the patient and determination of vital signs. Results should be documented in the patient’s medical record and should detail the following conditions:


*For the complete guideline, see American Association for Respiratory Care: Clinical practice guideline: transcutaneous blood gas monitoring for neonatal and pediatric patients. Respir Care 49:1070, 2004.

Instrumentation

Figure 18-11 shows a simplified diagram of a transcutaneous blood gas monitor sensor. Included are a heating element and two electrodes, one for O2 and one for CO2. These electrodes are similar in design to the electrodes found in bench-top analyzers. However, instead of measuring gas tensions in a blood sample, transcutaneous electrodes measure PO2 and PCO2 in an electrolyte gel between the sensor and the skin. When properly set up, the response time for these electrodes is 20 to 30 seconds, a bit slower than the response time for pulse oximetry.

Figure 18-12 shows a transcutaneous monitor with digital signal processing. A Severinghaus-type PtcCO2 electrode and two-wavelength reflectance SpO2 are combined into a single sensor. The sensor can be applied to the skin surface in neonates and infants or to the earlobe of pediatric and adult patients for combined noninvasive monitoring of ventilation and oxygenation.

Procedure

Box 18-7 outlines the basic procedure for setting up a transcutaneous blood gas monitor. The most common sites for electrode placement in infants and children are the abdomen, chest, and lower back. Once the electrodes are properly set up, the clinician should compare the monitor readings with the readings obtained with a concurrent ABG. Consistency between values validates monitor performance under the existing conditions. This validation should be repeated anytime the patient’s status undergoes a major change. During validation studies of patients with anatomic shunts, the electrode site and arterial sampling site should be on the same “side” of the shunt.

Problem Solving and Troubleshooting

Transcutaneous blood gas monitoring is a complex and labor-intensive activity that requires ongoing training and careful quality control. Table 18-5 lists the major factors that can affect the accuracy or limit the performance of a transcutaneous monitor. In terms of technical limitations, the lengthy stabilization time needed by transcutaneous monitors precludes their use during short procedures or in emergencies. In such cases, the pulse oximeter is a better choice.

TABLE 18-5

Factors Affecting Transcutaneous Blood Gas Monitors

Technical Factors Clinical Factors
Labor-intensive, high-skill procedure Poor perfusion
Lengthy stabilization time Hyperoxemia
Improper calibration can be difficult to detect Improper sensor application or placement
Heating required to obtain valid PO2 results Use of vasoactive drugs
Proper sensor-electrolyte contact is essential Variation in skin characteristics

Sensors must be calibrated and maintained using methods similar to the methods described for bench-top analyzers. Improper calibration yields erroneous patient information. Improper calibration can be difficult to detect on some systems. Meticulous care of the sensor membranes is also essential for proper maintenance.

Because the sensor is heated, clinicians must take care to avoid thermal injury to the patient’s skin. Thermal injury can be avoided by (1) careful monitoring of sensor temperature (the safe upper limit is approximately 42° C) and (2) regularly rotating the sensor site.

Proper sensor-electrolyte contact is essential, as is proper application to the skin surface. A loosely applied sensor may have air leaks or may become dislodged. In either case, the resulting measurements would approach those in room air:

PO2=159mm Hg

image

PCO2=0mm Hg

image

Conversely, excessive pressure on the sensor compresses the underlying capillaries and produces a falsely low PtcO2. Even with proper application and placement, PtcO2 measures can vary by 10% at different sites, with values from the extremities generally being lower than values obtained from the chest or abdomen.39

When arterial and transcutaneous blood gas values are inconsistent with each other or with the clinical status of the patient, the clinician should explore possible causes before reporting any results. Often, discrepancies can be reduced by switching the monitoring site or recalibrating the instrument. If these steps fail to resolve the inconsistencies, the clinician should recommend an alternative method for assessing gas exchange, such as pulse oximetry or more frequent ABG analysis.

Intraarterial (In Vivo) Blood Gas Monitoring

Over the past 20 years, the desire for continuous in vivo blood gas analysis has led to remarkable strides in technology. However, the clinical requirements for such systems (Box 18-8) are extremely demanding and have yet to be fully met. Potential benefits of continuous blood gas analysis include real-time monitoring and a reduction in therapeutic decision-making time, less blood loss and the need for transfusion (especially in pediatric patients), lower infection risk to the patient and blood exposure to the health care provider, improved accuracy by reducing preanalytic errors, and elimination of specimen transport.

Early research focused on miniaturized versions of standard blood gas analyzer electrodes. However, problems with instrument drift, fouling of electrode surfaces, wire breakage, current leakage, and corrosion have limited clinical application of these systems. Better success has been achieved using indwelling fiberoptic photochemical sensors, or optodes.

Instrumentation

Rather than using electrochemical electrodes, miniaturized optical fluorescence sensors, called optodes, measure blood gas parameters by photochemical reactions and changes in the intensity of light transmission through optical fibers. Figure 18-13 shows how optodes are combined (with a thermocouple) at the tip of a flexible fiberoptic catheter that is inserted into a peripheral artery. Several optode-based systems have been developed and validated for commercial use for continuous intraarterial blood gas assessment and monitoring of PaO2 during cardiopulmonary bypass.40,41

Clinical Performance and Usefulness

Despite technical improvements, the actual performance of in vivo blood gas monitors falls short of the clinical requirements previously specified. Compared with standard blood gas analysis, accuracy is improved.42,43 However, concerns about bias in PO2 measurement,44 sensor accuracy over extended periods in divergent patient groups,45 the need for femoral artery insertion because of blood pressure waveform dampening in the radial artery,46 high acquisition cost of the monitor and sensor catheters, dedication of a monitor to a single patient,47 and lack of evidence showing an impact on patient care have prevented widespread adoption and use of this technology.

Extraarterial (Ex Vivo) Blood Gas Monitoring

Extraarterial (ex vivo) on-demand blood gas monitoring systems are a logical compromise between bench-top and in vivo blood gas analysis. Ex vivo systems eliminate all the problems associated with indwelling sensors, while still providing quick results. In concept, the only major shortcoming of ex vivo systems is their inability to provide real-time continuous data.

Instrumentation and Procedure

Both optode-based and electrochemical-based systems have been developed. Figure 18-14 depicts an optode-based, ex vivo blood gas monitoring system. The optodes are located in a sensor cassette inserted in-line with the arterial catheter near the patient’s wrist. To measure pH, PCO2, and PO2, the system is closed to the intravenous fluid source at the stopcock (Figure 18-14, A). Subatmospheric pressure is created in the syringe attached to the stopcock (see Figure 18-14, A), which functions as an aspirating reservoir (Figure 18-14, B); this causes arterial blood to flow into the sensor cassette for analysis. During analysis, the stopcock (see Figure 18-14, A) is returned to its original position (off to the aspirating syringe). The connection through the line to the pressure transducer is restored, and blood pressure monitoring is able to continue. Blood gas parameter results are displayed in approximately 1 to 2 minutes. When analysis is complete, the blood sample is returned to the patient by emptying the aspirating reservoir (see Figure 18-14, B) and flushing the system through the flow valve (Figure 18-14, D).

Clinical Performance and Usefulness

In clinical trials, ex vivo on-demand systems performed as well as laboratory blood gas analyzers in adults, neonates, and infants.4850 Measurements could be obtained every 3 to 5 minutes from a peripheral or umbilical artery, and the intermittent errors commonly associated with in vivo systems were not observed. A commercially available system for blood gas monitoring in neonates and infants also incorporates electrolyte, hematocrit, and Hb measurements.50

Ex vivo blood gas monitoring systems share many of the requirements, advantages, and disadvantages related to in vivo systems stated in the previous section. Further justification of the costs associated with monitoring weighed against potential patient benefit are needed before widespread use of this technology occurs.

Tissue Oxygen

Tissue O2 (PtO2) can be measured by probes inserted directly into organs, tissue, and body fluids. Ease of probe placement and the sensitivity of PtO2 as an indicator of tissue perfusion make tissue O2 monitoring attractive for clinical research applications. Clinical indications for measuring PtO2 include monitoring brain tissue O2 as an early sign of ischemia, assessing brain blood flow autoregulation, and monitoring the adequacy of brain perfusion in patients with traumatic brain injury.51 In patients with traumatic brain injury, brain PtO2 values when intracranial pressure and cerebral perfusion are normal are between 25 mm Hg and 30 mm Hg. The critical threshold for ischemic brain damage and poor outcome is suspected to be around a brain PtO2 of 10 to 15 mm Hg.51

Oximetry

Oximetry is the measurement of blood Hb saturations using spectrophotometry. According to the principles of spectrophotometry, every substance has a unique pattern of light absorption, similar to a fingerprint. The pattern of light absorption of a substance varies predictably with the amount present; this is known as the Lambert-Beer law. By measuring the light absorbed and transmitted by a substance, scientists can identify its presence and determine its concentration.

The particular pattern of light absorption exhibited by a substance at different wavelengths is called its absorption spectrum. As shown in Figure 18-16, each form of Hb (e.g., Hb, HbO2, HbCO, metHb) has its own unique pattern. By comparison of the amount of light transmitted through (or reflected from) a blood sample at two or more specific wavelengths, the relative concentrations of two or more forms of Hb can be measured. For example, oxygenated Hb absorbs less red light (600 to 750 nm) and more infrared light (850 to 1000 nm) than deoxygenated or reduced Hb does. Comparing a blood sample’s light absorption with red and infrared light yields the %HbO2 and %Hb. For measurement of the concentration of additional forms of Hb, additional (more than two) wavelengths of light need to be used.

Several types of oximetry are used in clinical practice, including hemoximetry (also called cooximetry), pulse oximetry, venous oximetry, and tissue oximetry. Hemoximetry is a laboratory analytic procedure requiring invasive sampling of arterial blood. Pulse oximetry is a noninvasive monitoring technique performed at the bedside. Venous oximetry requires invasive monitoring through a fiberoptic catheter placed in the vena cava or pulmonary artery. Tissue oximetry is a noninvasive method of measuring the saturation of Hb at the tissue level.

Hemoximetry

Hemoximetry is an analytic method of oximetry and is covered in the AARC Clinical Practice Guideline: Blood Gas Analysis and Hemoximetry15 (see Clinical Practice Guideline 18-3). Related recommendations have been published by CLSI).9

Instrumentation

Figure 18-17 is a simplified diagram showing the key components of a laboratory hemoximeter. Light generated by a thallium cathode lamp passes through a series of lenses and filters, yielding the specific wavelengths needed for analysis. A beam splitter divides the light into two portions, directing one through a reference solution and the other through a sample chamber, or cuvette. Photodetection sensors measure the amount of light transmitted through these two sources. By comparing the difference in light transmission through the reference and sample solutions, a microprocessor computes the relative amount of Hb present, with its output sent to the calibrated device meter or display. Because a laboratory hemoximeter uses four or more different wavelengths of light, it can simultaneously compute the relative concentrations of multiple forms of Hb, such as Hb, HbO2, HbCO, and metHb.

Procedure and Quality Assurance

Similar to modern blood gas analyzers, laboratory hemoximeters are highly automated and simple to use. Some devices now combine both technologies into a single instrument. However, the caveats remain the same. Accurate and clinically useful hemoximetry results can be expected only if an error-free sample is assessed on a calibrated analyzer, using the manufacturer’s protocol.

Although variations exist among devices, the basic procedure is similar. First, the blood is introduced into the sampling port of the analyzer, usually either by aspiration or injection. Required sample sizes vary from approximately 200 µL to 40 µL (microanalysis). Once introduced, erythrocyte Hb is released into the solution by hemolysis (incomplete hemolysis can cause erroneous results). After hemolysis, the sample is transported to the cuvette for analysis. On completion of the analysis, the sampling system (cuvette and tubing) is flushed and cleaned. As with blood gas analysis, operators must follow CDC Standard Precautions and ensure proper disposal of syringes and waste materials.

Quality assurance procedures for hemoximetry are essentially the same as the procedures used for blood gas analysis, differing only with regard to the control materials used. In addition, careful cleaning and maintenance of the cuvette chamber is essential because clouding of its walls decreases absorbance and can cause falsely elevated values.53

Problem Solving and Troubleshooting

A major assumption underlying hemoximetry is that the measured changes in light absorbance result only from variations in the relative concentrations of various hemoglobins. In practice, this assumption does not always hold true. Table 18-6 outlines some of the potential problems and resulting errors that can occur with hemoximetry.

TABLE 18-6

Problems Causing Measurement Errors With Hemoximeters

Problem Potential Error
Incomplete hemolysis Falsely low total Hb, HbO2
Sickle cell anemia (caused by incomplete hemolysis) Falsely low HbO2
Presence of vascular dyes (e.g., methylene blue) Falsely low total Hb, HbO2
High lipid levels (e.g., from parenteral nutrition) Falsely low total Hb, HbO2
Presence of high levels of fetal hemoglobin Falsely high HbCO
Elevated bilirubin levels (>20 mg/dl) Falsely high total Hb, HbO2, metHb
Dirty cuvette chamber Falsely high total Hb, HbO2

Pulse Oximetry

A pulse oximeter is a portable noninvasive monitoring device that provides estimates of arterial blood oxyhemoglobin saturation levels. So as not to confuse these estimates with actual SaO2 measures obtained by hemoximetry, the abbreviation SpO2 is used to refer to pulse oximetry readings.

No other device in recent medical history has been so widely and quickly adopted into clinical practice. With this widespread use have come equally widespread misconceptions regarding the appropriate applications and limitations of this technology.54 In addition, the true benefit of pulse oximetry related to patient outcomes is unknown.55

To guide practitioners in providing quality care, the AARC has published Clinical Practice Guideline: Pulse Oximetry.56 Modified excerpts from the AARC guideline appear in Clinical Practice Guideline 18-5.

18-5   Pulse Oximetry

AARC Clinical Practice Guideline (Excerpts)*


*For the complete guideline, see American Association for Respiratory Care: Clinical practice guideline: sampling for arterial blood gas analysis. Respir Care 37:891, 1992.

Instrumentation

The pulse oximeter combines the principle of spectrophotometry, as used by hemoximeters, with photoplethysmography. Photoplethysmography uses light to detect the tiny volume changes that occur in living tissue during pulsatile blood flow. However, compared with a hemoximeter, the pulse oximeter usually uses only two wavelengths of light, one red (approximately 660 nm) and one infrared (approximately 940 nm) (see Figure 18-16). In addition, rather than measuring light transmission through a blood sample in a glass cuvette, the pulse oximeter measures transmission through living tissue, such as a finger or earlobe, or reflectance through the skin surface.

Figure 18-18, A provides a schematic block diagram of a pulse oximeter, consisting of a transmission sensor, processor, and display unit. The sensor has two sides. From one side, separate red and infrared LEDs alternately transmit light through the tissue. The transmitted light intensity is measured by a photodetector on the other side. The resulting output signal is filtered and amplified by instrument electronics, with processing and display functions controlled by a microprocessor.

Figure 18-18, B shows a schematic of a reflectance pulse oximeter sensor. This type of sensor has only one side, which contains both the LED light sources and the photodetector. The principle of operation is identical to a transmission sensor except that the sensor is placed on the skin surface, usually the forehead, and reflected light from the tissue back to the sensor is used to calculate SpO2.

Figure 18-19 shows a typical output signal generated by the photodetector (the pulsatile component can be observed on instruments that have a plethysmographic display). A baseline component represents the stable absorbance of the tissue bed, which mainly is the result of venous and capillary blood. At the top is the pulsatile component, caused by intermittent arterial flow through the tissues. By comparing light absorbance during the pulsatile phase with the baseline value at each wavelength, a pulse-added measure is obtained that is independent of incident light. Arterial oxyhemoglobin saturation is computed as the ratio of the pulse-added absorbances at the two different wavelengths.

In terms of accuracy, the pulse oximetry readings of sick patients usually fall within ±3% to 5% of the readings obtained with invasive hemoximetry.56,57 Generally, the lower the actual SaO2, the less accurate and reliable is the SpO2 measurement. Most clinicians consider pulse oximeter readings unreliable at saturations less than 80%. Instrument response times vary by manufacturer, sensor location, and the patient’s hemodynamic status from 10 seconds to 1 minute or longer.

Procedure

The actual procedure used to measure SpO2 varies according to the device used, sensor site selected, and whether a spot check or continuous monitoring is required. Box 18-9 lists key points to be considered when performing pulse oximetry.

Given the limits of this technology, meticulous documentation is a must. Specifically, all SpO2 results should be recorded in the patient’s medical record. The following details should be documented:

Problem Solving and Troubleshooting

Problems with pulse oximetry fall into the following two categories: (1) problems inherent in the technology itself and (2) problems associated with clinical interpretation and use of data. Dozens of technical factors may affect the readings, limit the precision, or alter the performance of pulse oximeters. Table 18-7 summarizes the most important of these factors and the types of errors they cause.

TABLE 18-7

Factors Affecting Accuracy or Precision of Pulse Oximeters

Factor Potential Error
Presence of HbCO Falsely high %HbO2
Presence of high levels of metHb Falsely low %HbO2 if SaO2 >85%
  Falsely high %HbO2 if SaO2 <85%
Presence of fetal hemoglobin No effect
Anemia (very low hematocrit, <10%) Falsely low CaO2 and high % HbO2
Vascular dyes (e.g., methylene blue) Falsely low %HbO2
Elevated bilirubin levels No effect
Dark skin pigmentation Falsely high %HbO2 (3%-5%)
Nail polish (especially black) Falsely high %HbO2
Ambient light Varies (e.g., falsely high %HbO2 in sunlight); also may cause falsely high pulse reading
Poor perfusion (vasoconstriction) Inadequate signal; unpredictable results
Motion artifact Unpredictable, spurious readings
Electrocautery Falsely low HbO2
Magnetic resonance imaging Falsely low HbO2

Motion artifact probably is the most common source of error and false alarms. Although new technologies promise to reduce motion artifact, relocation of the sensor to the earlobe, toe, or forehead can minimize the problem. Falsely elevated readings can occur with dark skin pigmentation at low saturation levels; this can be compensated for by setting oximeter low alarms 3% to 5% higher in applicable cases.

Early studies of the effect of nail polish found significant differences in lowering SpO2 readings. More recent studies have found no effect or small differences that are not considered to be clinically relevant; this may be due to improvements in LED light sources.58 It has been suggested that the effect of nail polish could be minimized either by using a different site or by rotating the sensor so that the light path does not cross the fingernails.

If ambient light interference is creating problems, the sensor can be loosely covered with an opaque towel or cloth. Problems that occur during procedures producing electromagnetic interference (e.g., electrocautery, magnetic resonance imaging) need only be recognized. Careful monitoring of the patient during episodes of false low alarms is essential.

Regarding problems with the use and interpretation of pulse oximetry data, rule number one is to treat the patient, not the monitor. The clinician should never interpret or act on monitoring data without first assessing the patient and verifying proper sensor placement and signal quality. A related problem is simple confusion over the relationship between oxyhemoglobin saturation and PO2. Many clinicians rely solely on PaO2 readings to assess oxygenation and do not understand oxyhemoglobin saturation. To these clinicians, an SpO2 reading of 80% might be confused easily with PaO2 of 80 mm Hg. The latter measure of partial pressure is normal, whereas a saturation of 80% indicates moderate to severe hypoxemia, equivalent to PaO2 of approximately 50 mm Hg.

A similar interpretation error (PaO2 vs. SpO2) occurs because of the limited accuracy of most pulse oximeters. It is common practice to set the low alarm of a monitoring oximeter to 90%. In theory, this practice makes sense because an SaO2 reading of 90% normally corresponds to a PO2 reading of approximately 60 mm Hg, which is the lower limit of clinically acceptable oxygenation. However, with the accuracy of some oximeters being only ±4%, an SpO2 reading of 90% could mean an actual SaO2 reading of 86%, corresponding to a PO2 level of 55 mm Hg or less.

At the high end, oximetry data can be even less meaningful. Because of the characteristics of the oxyhemoglobin dissociation curve (see Chapter 11), a patient with an SpO2 reading of 100% could represent a PaO2 level anywhere between 100 mm Hg and 600 mm Hg. The lesson here is not to use the pulse oximeter for monitoring hyperoxia (as may be important for neonates).

The pulse oximeter does not measure PCO2. A patient breathing an elevated FiO2 can have normal SpO2 readings despite severe hypercarbia. ABG analysis is needed when acute ventilatory failure may be present.

SpO2 can read falsely high when carbon monoxide poisoning or methemoglobinemia is present. This false reading is due to the fact that the two-wavelength pulse oximeter measures only saturation of the Hb and not specifically saturation with O2. HbCO and metHb cannot be distinguished from HbO2 with a pulse oximeter. A falsely high SpO2 reading occurs when significant HbCO is present. When metHb is elevated, the SpO2 reading is higher than the actual measured SaO2. As metHb level increases, SpO2 decreases and plateaus at approximately 85% when the metHb level reaches 30%.59

To address these limitations, pulse oximeters using seven or more wavelengths of light have been developed. Use of multiwavelength pulse oximeters capable of measuring Hb, HbO2, HbCO, and metHb has been referred to as pulse cooximetry. The accuracy of measurements does not equal the accuracy of conventional hemoximetry, but pulse cooximetry may be useful for trend monitoring in some clinical situations.60

As with transcutaneous monitoring, if pulse oximetry and blood gas values are inconsistent with each other or the clinical status of the patient, the RT should explore possible causes before reporting, interpreting, or acting on results. Often, discrepancies can be reduced by switching sites or replacing the sensor probe. If these steps fail to resolve the inconsistencies, the RT should document the problem and recommend obtaining an ABG with hemoximetry if indicated.

Venous Oximetry

Continuous central venous (vena cava) and mixed venous (pulmonary artery) O2 saturation monitoring (SimageO2) is performed to assess the balance between O2 delivery and use as an indirect index of global tissue oxygenation and perfusion. Decreased SimageO2 has been found to be indicative of cardiac failure in patients with myocardial infarction and to predict poor prognosis in patients after cardiovascular surgery, in patients with severe cardiopulmonary disease, and in patients with septic or cardiogenic shock.61 Regional and organ-specific SimageO2 monitoring has been performed via catheters placed in the coronary sinus, hepatic vein, and cranial jugular venous bulb for cerebral perfusion monitoring. Normal values for SimageO2 range from 60% to 80%.

Instrumentation

Figure 18-20 shows a diagram of an SimageO2 monitoring system. Venous oximetry is measured through a fiberoptic catheter by reflectance spectrophotometry. Two or three wavelengths of light are emitted from LEDs through fiberoptic filaments into the venous blood. Some of this light is reflected back and received through another fiberoptic channel, which is read by a photodetector. The amount of light that is absorbed by the venous blood and reflected back is determined by the amount of O2 that is saturated or bound to Hb. This information is processed by the monitor, updated, and displayed as SimageO2.

Tissue Oximetry

O2 saturation at the tissue level (StO2) assesses the adequacy of circulation and O2 delivery. Early detection of low StO2 can be used as method of early detection of tissue hypoperfusion in patients with traumatic injuries.64,65 Cerebral StO2 monitoring can also be used to monitor brain oxygenation and detect cerebral ischemia during neurosurgical or cardiovascular procedures.6668

Capnometry and Capnography

Capnometry is the measurement of CO2 in respiratory gases. A capnometer is the device that measures CO2. Capnography is the graphic display of CO2 levels as they change during breathing.

Although capnography can be applied to any patient, its primary clinical use is for monitoring during either general anesthesia (where it is a standard of care) or mechanical ventilation. The remainder of this section assumes application during mechanical ventilation. To guide practitioners in providing quality care, the AARC has published Clinical Practice Guideline: Capnography/Capnometry During Mechanical Ventilation.69 Modified excerpts from the AARC guideline appear in Clinical Practice Guidelines 18-6.

18-6   Capnography and Capnometry During Mechanical Ventilation

AARC Clinical Practice Guideline (Excerpts)*

Indications

Based on available evidence, capnography may be indicated for the following:

• Evaluation of exhaled CO2, especially end-tidal CO2 levels (PetCO2)

• Monitoring severity of pulmonary disease and evaluating the patient’s response to therapy, especially that intended to do the following:

• Determining that tracheal, rather than esophageal, intubation has taken place

• Continued monitoring of the integrity of the ventilatory circuit, including the artificial airway

• Evaluation of the efficiency of mechanical ventilatory support (by [PaCO2 − PetCO2])

• Monitoring adequacy of pulmonary, systemic, and coronary blood flow

• Monitoring inspired CO2 when CO2 gas is being therapeutically administered

• Graphic evaluation of ventilator-patient interface

• Measurement of the volume of CO2 elimination to assess metabolic rate or alveolar ventilation


*For the complete guideline, see American Association for Respiratory Care: Clinical practice guideline: capnography/capnometry during mechanical ventilation. Respir Care 48:534, 2003.

Instrumentation

The key component in a capnograph is a rapidly responding CO2 analyzer. Rapid CO2 analysis can be achieved using infrared absorption, Raman scattering, mass spectroscopy, or photoacoustic technology, with the infrared capnometer being the most common.

Figure 18-21 provides a simple schematic of a double-beam infrared capnometer. A filtered infrared light source passes through a sample chamber. (Because glass absorbs infrared radiation, the chamber “windows” usually are constructed of sodium chloride or sodium bromide.) After the infrared light passes through the sample chamber, a lens focuses the remaining, unabsorbed radiation onto an electrical photodetector. Because CO2 absorbs infrared radiation, the greater the concentration of CO2 in the sample, the less infrared light arrives at the detector. Variations in the concentration of CO2 alter the electrical output signal of the detector. This signal is used either to display the CO2 concentration with LEDs (capnometer) or to generate a real-time graphic display (capnogram).70

Capnometers use two different methods to sample the respiratory gases: mainstream sampling and sidestream sampling (Figure 18-22). The mainstream analyzer places an in-line analysis chamber between the patient’s airway and the ventilator circuit. The sidestream analyzer uses a sampling tube to pump a small volume of gas continually from the ventilator circuit into the analysis chamber within the device. Table 18-8 lists the advantages and disadvantages of these two approaches. Differences notwithstanding, clinicians should use the method best suited to the patient’s needs and the device with which they are most experienced and familiar.

TABLE 18-8

Advantages and Disadvantages of Mainstream and Sidestream Capnometers

  Mainstream Sidestream
Advantages Sensor at patient airway No bulky sensors or heaters at airway
  Fast response (crisp waveform) Ability to measure N2O
  Short lag time (real-time readings) Disposable sample line
  No sample flow to reduce tidal volume Ability to use with nonintubated patients
Disadvantages Secretions and humidity can block sensor window Secretions block sample tubing
  Sensor requires heating to prevent condensation Trap required to remove water from sample
  Requires frequent calibration Frequent calibration required
  Bulky sensor at patient airway Slow response to CO2 changes
  Does not measure N2O Lag time between CO2 change and measurement
  Difficult to use with nonintubated patients Sample flow may decrease tidal volume
  Reusable adapters require cleaning and sterilization  

From Kacmarek RM, Hess D, Stoller J, editors: Monitoring in respiratory care, St Louis, 1993, Mosby.

Interpretation

Interpretation of the capnogram can be useful in assessing trends in alveolar ventilation and detecting ventilation/perfusion ratio (image) imbalance caused by either pulmonary disease or cardiovascular disorders. Capnometry also has been used to estimate physiologic dead space, to detect esophageal intubation, to assess blood flow during cardiac arrest, and to determine positive end expiratory pressure levels. To interpret abnormal events, clinicians first must understand the normal capnogram.

Normal Capnogram

Figure 18-23 shows a typical normal single-breath capnogram. Initially, the expired PCO2 is 0 mm Hg, indicating exhalation of pure dead space gas (A, phase I). Soon after, alveolar gas begins mixing with dead space gas, causing a rapid increase in expired PCO2 (A to B, phase II). Later in expiration, the CO2 concentration begins leveling off. This plateau indicates exhalation of gas coming mainly from ventilated alveoli (B to C, phase III). Gas sampled at the end of exhalation is called end-tidal gas, with its partial pressure of CO2 abbreviated as PetCO2. In healthy individuals, PetCO2 averages 3 to 5 mm Hg less than PaCO2, or 35 to 43 mm Hg (approximately 5% to 6% CO2). The sharp downstroke and return to baseline that normally occurs after the end-tidal point indicates inhalation of fresh gas with zero CO2.

Abnormal Capnogram

The first step in assessing the capnogram is to determine the actual PetCO2 and whether it has changed over time. Table 18-9 differentiates between the causes of high and low PetCO2 readings by the suddenness of the change. A PetCO2 of zero usually indicates a system leak, esophageal intubation, or cardiac arrest.

TABLE 18-9

Conditions Associated With Changes in PetCO2

Change High PetCO2 Low PetCO2
Sudden Sudden increase in cardiac output Sudden hyperventilation
  Sudden release of a tourniquet Sudden decrease in cardiac output
  Injection of sodium bicarbonate Massive pulmonary embolism
    Air embolism
    Disconnection of ventilator
    Obstruction of endotracheal tube
    Leakage in the circuit
Gradual Hypoventilation Hyperventilation
  Increase in CO2 production Decrease in oxygen consumption
    Decreased pulmonary perfusion

Note: An absent PetCO2 means that a system leak, esophageal intubation, or cardiac arrest has occurred.

After the capnogram has been assessed for changes in PetCO2, the waveform and its pattern should be analyzed. A normal capnogram starts with a sharp upstroke, followed by a plateau and then a rapid downstroke. As indicated in Figure 18-24, changes in this normal contour may indicate a ventilation/perfusion abnormality. Such patterns, although not diagnostic, can indicate the severity of the image disturbance and can warn of developing problems, such as acute pulmonary emboli.

Waveform changes also may occur with equipment malfunction. Because the normal inspired CO2 level is zero, the capnogram baseline also should be at zero. An elevated baseline (>0 mm Hg) indicates rebreathing. However, an expired CO2 level of zero might indicate patient disconnect. (For more information on the use of capnography during mechanical ventilation, see Chapter 46.)

Problem Solving and Troubleshooting

Monitoring a patient with a capnograph that is properly calibrated and operating according to the manufacturer’s specifications presents few major problems. The most significant error is assuming that the end-expired CO2 levels can substitute for actual PaCO2 measurements. The most common problem is contamination or obstruction of the sampling system or monitor by secretions or condensate.27 Proper use of water traps and regular changing of sample tubing or chambers can help prevent this problem. Other potential problems include the following:

Summary Checklist

• To measure the inspired O2 concentration, a properly calibrated electrochemical O2 analyzer should be used.

• The most common causes of O2 analyzer malfunction are low batteries, sensor depletion, and electronic failure.

• As the “gold standard” of gas exchange analysis, ABG results help the clinician assess ventilation, acid-base balance, oxygenation, and the O2-carrying capacity of blood.

• The radial artery is the preferred site for adult arterial blood sampling. Before radial puncture, a modified Allen test to confirm collateral circulation is performed.

• For critically ill patients, the clinician waits 20 to 30 minutes after a change in treatment before sampling arterial blood.

• Most preanalytic blood gas errors can be avoided by ensuring that the sample was obtained anaerobically, is properly anticoagulated, and is analyzed within 15 minutes.

• Indwelling peripheral artery, central venous, and pulmonary artery catheters give ready access for blood sampling and allow continuous pressure monitoring but with increased risk of infection and thrombosis.

• Capillary blood pH and PCO2 are sometimes used to assess acid-base status in infants and children. Capillary PO2 is of little value in estimating arterial oxygenation.

• To perform blood gas analysis and hemoximetry, the clinician must be proficient in performing procedures, preventive maintenance, troubleshooting, instrument calibration, and quality control.

• A blood gas analyzer measures pH, PCO2, and PO2 using three separate electrodes.

• To obtain accurate blood gas results, the clinician ensures that the sample is free of preanalytic error and follows the manufacturer’s recommended analysis protocol.

• Blood gas analysis quality control involves a cycle of performance validation for new instruments, preventive maintenance and function checks, automated calibration, calibration verification with control media, internal statistical quality control, external proficiency testing, and thorough recordkeeping of all processes.

• Portable point-of-care blood gas analyzers can achieve accuracy and precision levels comparable with laboratory-based analyzers.

• A blood gas monitor provides bedside measurements either continuously or at appropriate intervals, without permanently removing blood from the patient; this may be accomplished transcutaneously or with either in vivo or ex vivo blood analysis.

• Transcutaneous blood gas monitoring provides continuous noninvasive analysis of gas exchange, but it is useful only for hemodynamically stable infants or children.

• Current in vivo blood gas monitors are not yet reliable enough to replace traditional blood gas analysis, but they may provide information at a level of accuracy and reliability sufficient for trend analysis.

• Oximetry is the measurement of blood Hb saturations using spectrophotometry. Hemoximetry is a laboratory procedure that requires an arterial blood sample. Pulse oximetry combines spectrophotometry with photoplethysmography to obtain a noninvasive measure of blood Hb saturations.

• At best, pulse oximetry readings fall within ±3% to 5% of readings obtained by hemoximetry.

• Dozens of technical factors affect the readings, limit the precision, or alter the performance of pulse oximeters. To interpret test results properly, clinicians must have in-depth knowledge of these factors.

• Capnometry is the measurement of CO2 in respiratory gases. A capnometer is the device that measures the CO2. Capnography is the graphic display of CO2 levels as they change during breathing.

• A capnogram may be used to assess trends in alveolar ventilation, to identify image imbalance caused by cardiopulmonary disorders, to estimate physiologic dead space, to detect esophageal intubation, and to determine the amount of blood flow during cardiac arrest.