Aerosol Drug Therapy

Published on 01/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 4 (2 votes)

This article have been viewed 11775 times

Aerosol Drug Therapy

Jim Fink

An aerosol is a suspension of solid or liquid particles in gas. Aerosols occur in nature as pollens, spores, dust, smoke, smog, fog, and mist.1 A primary function of the upper airway and respiratory tract is to protect the lungs from invasion by these aerosols. In the clinical setting, medical aerosols are generated with atomizers, nebulizers, and inhalers—devices that physically disperse matter into small particles and suspend them into a gas. Aerosols can be used to deliver bland water solutions to the respiratory tract (see Chapter 35) or to administer drugs to the lungs, throat, or nose for local and systemic effect. This chapter focuses on the principles of aerosol drug therapy.

The aim of medical aerosol therapy is to deliver a therapeutic dose of the selected agent (drug) to the desired site of action. The indication for any specific aerosol is based on the need for the specific drug and the targeted site of delivery.1 For patients with pulmonary disorders, administration of drugs by aerosol offers higher local drug concentrations in the lung with lower systemic levels compared with other forms of administration. Improved therapeutic action with fewer systemic side effects provides a higher therapeutic index.2

Characteristics of Therapeutic Aerosols

Effective use of medical aerosols requires an understanding of the characteristics of aerosols and their effect on drug delivery to the desired site of action. Key concepts include aerosol output, particle size, deposition, and a phenomenon known as aging.

Aerosol Output

The rate that aerosol is generated is a key parameter in aerosol administration. Aerosol output is defined as the mass of fluid or drug contained in aerosol produced by a nebulizer. Output is expressed as either a unit of mass leaving the nebulizer or as a proportion of the dose placed in the nebulizer. Output rate is the mass of aerosol generated per unit of time. Output varies greatly among different nebulizers and inhalers. For drug delivery systems, emitted dose describes the mass of drug leaving the mouthpiece of a nebulizer or inhaler as aerosol.

Aerosol output can be measured by collecting the aerosol that leaves the nebulizer on filters and measuring either their weight (gravimetric analysis) or quantity of drug (assay). Gravimetric measurements of aerosols are less reliable than drug assay techniques because weight changes resulting from water evaporation cannot be differentiated from changes in drug mass. A drug assay provides the most reliable measure of aerosol output.

A substantial proportion of particles that leave a nebulizer never reach the lungs. The ability of aerosols to travel through the air, enter the airways, and become deposited in the lungs is based on numerous variables ranging from particle size to breathing pattern. Understanding and skillful manipulation of these variables can greatly improve pulmonary delivery of aerosols.

Particle Size

Aerosol particle size depends on the substance for nebulization, the method used to generate the aerosol, and the environmental conditions surrounding the particle.3 It is impossible to determine visually whether a nebulizer is producing an optimal particle size. The unaided human eye cannot see particles less than 50 to 100 µm in diameter (equivalent to a small grain of sand). The only reliable way to determine the characteristics of an aerosol suspension is laboratory measurement. The two most common laboratory methods used to measure medical aerosol particle size distribution are cascade impaction and laser diffraction. Cascade impactors are designed to collect aerosols of different size ranges on a series of stages or plates. The mass of aerosol deposited on each plate is quantified by drug assay, and a distribution of drug mass across particle sizes is calculated. In laser diffraction, a computer is used to estimate the range and frequency of droplet volumes crossing the laser beam.

Because medical aerosols contain particles of many different sizes (heterodisperse), the average particle size is expressed with a measure of central tendency, such as mass median aerodynamic diameter (MMAD) for cascade impaction or volume median diameter (VMD) for laser diffraction. These measurement techniques of the same aerosol may report different sizes, so it is important to know which measurement is used. The MMAD and VMD both describe the particle diameter in micrometers (µm). In an aerosol distribution with a specific MMAD, 50% of the particles are smaller and have less mass, and 50% are larger and have greater mass.

The geometric standard deviation (GSD) describes the variability of particle sizes in an aerosol distribution set at 1 standard deviation above or below the median (15.8% and 84.13%). Most aerosols found in nature and used in respiratory care are composed of particles of different sizes, described as heterodisperse. The greater the GSD, the wider the range of particle sizes, and the more heterodisperse the aerosol. Aerosols consisting of particles of similar size (GSD ≤ 1.2) are referred to as monodisperse. Nebulizers that produce monodisperse aerosols are used mainly in laboratory research and in nonmedical industries.

Deposition

When aerosol particles leave suspension in gas, they deposit on (attach to) a surface. Only a portion of the aerosol generated and emitted from a nebulizer (emitted dose) may be inhaled (inhaled dose). A fraction of the inhaled dose is deposited in the lungs (respirable dose). Inhaled mass is the amount of drug inhaled. The proportion of the drug mass in particles that are small enough (fine-particle fraction) to reach the lower respiratory tract is the respirable mass. Not all aerosol delivered to the lung is retained, or deposited. A small percentage (1% to 5%) of inhaled drug may be exhaled. Whether aerosol particles that are inhaled into the lung are deposited in the respiratory tract depends on the size, shape, and motion of the particles and on the physical characteristics of the airways and breathing pattern. Key mechanisms of aerosol deposition include inertial impaction, gravimetric sedimentation, and brownian diffusion.1,3

Inertial Impaction

Inertial impaction occurs when suspended particles in motion collide with and are deposited on a surface; this is the primary deposition mechanism for particles larger than 5 µm. The greater the mass and velocity of a moving object, the greater its inertia, and the greater the tendency of that object to continue moving along its set path (Figure 36-1). When a particle of sufficient (large) mass is moving in a gas stream and that stream changes direction, the particle tends to remain on its initial path and collide with the airway surface.

Because inertia involves both mass and velocity, the higher the flow of a gas stream, the greater the tendency for particles to impact and be deposited in the airways. Turbulent flow patterns, obstructed or tortuous pathways, and inspiratory flow rates greater than 30 L/min are associated with increased inertial impaction. Turbulent flow and convoluted passageways in the nose cause most particles larger than 10 µm to impact and become deposited. This process produces an effective filter that protects the lower airway from particulates such as dust and pollen. However, particles 5 to 10 µm tend to become deposited in the oropharynx and hypopharynx, especially with the turbulence created by the transition of air as it passes around the tongue and into the larynx.

Sedimentation

Sedimentation occurs when aerosol particles settle out of suspension and are deposited owing to gravity. The greater the mass of the particle, the faster it settles (Figure 36-2). During normal breathing, sedimentation is the primary mechanism for deposition of particles 1 to 5 µm. Sedimentation occurs mostly in the central airways and increases with time, affecting particles 1 µm in diameter. Breath holding after inhalation of an aerosol increases the residence time for the particles in the lung and enhances distribution across the lungs and sedimentation. A 10-second breath hold can increase aerosol deposition 10% and increase the ratio of aerosol deposited in lung parenchyma to central airway by fourfold.4

Diffusion

Brownian diffusion is the primary mechanism for deposition of small particles (<3 µm), mainly in the respiratory region where bulk gas flow ceases and most aerosol particles reach the alveoli by diffusion. These aerosol particles have very low mass and are easily bounced around by collisions with carrier gas molecules. These random molecular collisions cause some particles to contact and become deposited on surrounding surfaces. Particles 1 to 0.5 µm are so stable that most remain in suspension and are cleared with the exhaled gas, whereas particles smaller than 0.5 µm have a greater retention rate in the lungs.

Figure 36-3 summarizes the relationships between particle size and aerosol deposition in the respiratory tract. The depth of penetration and deposition of a particle in the respiratory tract tend to vary with size and tidal volume (VT).5 With this knowledge, it may be possible to target aerosol deposition to specific areas of the lung by using the proper particle size and breathing pattern.

Aging

Aerosols are dynamic suspensions. Individual particles constantly grow, shrink, coalesce, and fall out of suspension. The process by which an aerosol suspension changes over time is called aging. How an aerosol ages depends on the composition of the aerosol, the initial size of its particles, the time in suspension, and the ambient conditions to which it is exposed.

Aerosol particles can change size as a result of either evaporation or hygroscopic water absorption. The relative rate of particle size change is inversely proportional to the size of a particle, so small particles grow or shrink faster than large particles. Small water-based particles shrink when exposed to relatively dry gas. Aerosols of water-soluble materials, especially salts, tend to be hygroscopic, absorbing water and growing when introduced into a high-humidity environment.5

Particle size is not the only determinant of deposition. Inspiratory flow rate, flow pattern, respiratory rate, inhaled volume, ratio of inspiratory time to expiratory time (I : E ratio), and breath holding all influence where a particle of any specific size is deposited. The presence of airway obstruction is one of the greatest factors influencing aerosol deposition. It has been shown that total pulmonary deposition is greater in smokers and patients with obstructive airway disease than in healthy persons (Figure 36-4). Similarly, when inspiratory flow rates are constant, the deposition fraction of monodisperse aerosols increases with increased VT, length of respiratory (inspiratory) period, and particle size (Figure 36-5).

These dynamic variables make it difficult to predict exactly what occurs to aerosol particles when they enter a gas stream and are inhaled. For this reason, prediction of actual aerosol deposition for an individual patient is difficult.

Quantifying Aerosol Delivery

As mentioned in the preceding sections, many characteristics of aerosols can account for the variances in quantity of aerosolized medication delivered to the patient. Although the precise amount of drug delivered to the patient’s airways can be difficult to determine, it can be measured in terms of the patient’s clinical response to aerosol drug therapy, including the desired therapeutic effects and any unwanted adverse effects. The amount of aerosol deposited to a patient’s airways can be quantified using specialized equipment and tests.

One approach used to quantify aerosol deposition to the human body (in vivo) involves scintigraphy, in which a drug is “tagged” with a radioactive substance (e.g., technetium), aerosolized, and inhaled. A scanner (similar to scanners used in nuclear medicine) measures the distribution and intensity of radiation across the device and the patient’s head and thorax. The result is a radiation map of aerosol deposition in the upper airway, the lungs (central and peripheral airways), and the stomach. This information is used to calculate the percentage of drug retained by the device and delivered to various areas in the patient.6

A less direct approach relates the systemic pharmacokinetic profile of a drug delivered by aerosol to an assay of the drug in a patient’s blood or urine over time. This method does not estimate actual lung delivery, but it provides insight into systemic drug levels achieved after aerosol administration. Care must be taken to differentiate drug absorbed through the lungs from drug absorbed through the gastrointestinal tract. Simple laboratory, or in vitro, models, which simulate a range of VT values, inspiratory flow rates, I : E ratios, and respiratory rates, have been useful in predicting inhaled mass of drug and relative performance of nebulizers.7

Hazards of Aerosol Therapy

The primary hazard of aerosol drug therapy is an adverse reaction to the medication being administered (see Chapter 32). Other hazards to the patient include infection, airway reactivity, systemic effects of bland aerosols, drug concentration, and eye irritation. Care providers and bystanders risk these hazards as a result of exposure to secondhand aerosol drugs.

Infection

Aerosol generators can contribute to nosocomial infections by spreading bacteria by the airborne route.8 The most common sources of bacteria are patient secretions, contaminated solutions (i.e., multiple-dose drug vials), and caregivers’ hands. Offending organisms are primarily gram-negative bacilli, in particular, Pseudomonas aeruginosa and Legionella pneumophila (the cause of the highly virulent legionnaires’ disease).9

Various procedures can help reduce contamination and infection associated with respiratory care equipment. Guidelines from the U.S. Centers for Disease Control and Prevention (CDC) state that nebulizers should be sterilized between patients, frequently replaced with disinfected or sterile units, or rinsed with sterile water (not tap water) and air dried every 24 hours (see Chapter 4).

Airway Reactivity

Cold air and high-density aerosols can cause reactive bronchospasm and increased airway resistance, especially in patients with preexisting respiratory disease.10 Medications such as acetylcysteine, antibiotics, steroids, cromolyn sodium, ribavirin, and distilled water have been associated with increased airway resistance and wheezing during aerosol therapy. Administration of bronchodilators before or with administration of these agents may reduce the risk or duration of increased airway resistance.

The risk of inducing bronchospasm always should be considered when aerosols are administered. Monitoring for reactive bronchospasm should include peak flow measurements or percentage forced expiratory volume in 1 second (%FEV1) before and after therapy; auscultation for adventitious breath sounds; observation of the patient’s breathing pattern and overall appearance; and, most essential, communicating with the patient during therapy to determine the perceived work of breathing.11

Pulmonary and Systemic Effects

Pulmonary and systemic effects are associated with the site of delivery and the drug being administered. However, even bland aerosols present risk. Excess water can cause overhydration, and excess saline solution can cause hypernatremia. Animal data indicate that long-term, continuous administration of bland aerosols can cause localized inflammation and tissue damage, atelectasis, and pulmonary edema.

Preliminary assessment should balance the need versus the risk of aerosol therapy, especially among patients at high risk, such as infants, patients who are prone to fluid and electrolyte imbalances, and patients with atelectasis or pulmonary edema. For patients unable to clear their own secretions, suctioning or other airway clearance techniques may be indicated as an adjunct to aerosol therapy. Care must be taken to ensure that patients are capable of clearing secretions when the secretions are mobilized by aerosol therapy. Appropriate airway clearance techniques should accompany any aerosol therapy designed to help mobilize secretions (see Chapter 40).

Drug Concentration

During nebulization, the evaporation, heating, baffling, and recycling of drug solutions undergoing jet or ultrasonic nebulization increase solute concentrations.12 This process can expose the patient to increasingly higher concentrations of the drug over the course of therapy and result in a larger concentration of drug remaining in the nebulizer at the end of therapy. This increase in concentration usually is time-dependent; the greatest effect occurs when nebulization of medications occurs over extended periods, as in continuous aerosol drug delivery.

Eye Irritation

Aerosol administration via a face mask may deposit drug in the eyes and cause eye irritation. In very rare cases, anticholinergic medications (see Chapter 32) have been suspected to worsen preexisting eye conditions, such as forms of glaucoma. Caution should be exercised when a face mask is used during aerosol drug therapy. In addition, special mask designs that have been shown to reduce drug deposition in the eyes or mouthpieces should be considered for at-risk patients.13,14

Secondhand Exposure to Aerosol Drugs

Workplace exposure to aerosol may be detectable in the plasma of bystanders and health care providers. Repeated secondhand exposure to bronchodilators is associated with increased risk of occupational asthma. Institutions should develop and implement an occupational health and safety policy to minimize the risk of secondhand aerosol exposure for care providers and bystanders.1517 Unless filters are placed in the expiratory limb, 40% of aerosol produced during mechanical ventilation is exhausted to the air of the intensive care unit.18 Implementation of an occupational health and safety policy could include using systems that introduce less aerosol to the atmosphere (pressurized metered dose inhalers [pMDIs], dry powder inhalers [DPIs], and breath-actuated nebulizers), filtering exhalation to contain aerosol, and using environmental controls.

Aerosol drug Delivery Systems

Effective aerosol therapy requires a device that quickly delivers sufficient drug to the desired site of action with minimal waste and at a low cost.19 Aerosol generators in use include pMDIs with or without spacers or holding chambers, DPIs, small and large volume (jet) nebulizers, hand-bulb atomizers (including nasal spray pumps), ultrasonic nebulizers (USNs), and vibrating mesh (VM) nebulizers as well as numerous emerging technologies.2

Clinicians are often exposed to competing and sometimes conflicting claims about the different delivery systems and may not be provided the information they need to select the correct system for a given situation. Because device selection can make the difference between successful and unsuccessful therapy, clinicians must have in-depth knowledge of the operating principles and performance characteristics of these various systems and how best to select and apply them.20

Metered Dose Inhalers

The pMDI is the most commonly prescribed method of aerosol delivery in the United States. The pMDI is portable, compact, and easy to use and provides multidose convenience. A uniform dose of drug is dispensed within a fraction of a second after actuation and is reproducible throughout the canister life. The pMDI and actuator are designed for the specific drug formulation and dose volume to be delivered. Although the pMDI appears to be a simple device, it represents sophisticated technology and engineering. Although relatively easy to use, it is commonly misused by patients.

The pMDI is used to administer bronchodilators, anticholinergics, and steroids. More formulations of these drugs are available for use by pMDIs than for use with nebulizers. Properly used, pMDIs are at least as effective as other nebulizers for drug delivery. For this reason, pMDIs often are the preferred method for delivering bronchodilators to spontaneously breathing patients and patients who are intubated and undergoing mechanical ventilation.21

Most pMDIs are “press and breathe,” but there is increasing presence of a variation known as breath-actuated pMDIs. The basic components of pMDI are similar regardless of type, manufacturer, or active ingredient; commonly used pMDIs are shown in Figure 36-6.

A pMDI is a pressurized canister that contains the prescribed drug (a micronized powder or aqueous solution) in a volatile propellant combined with a surfactant and dispersing agent (Figure 36-7). When the canister is inverted (nozzle down) and placed in its actuator, or “boot,” the volatile suspension fills a metering chamber that controls the amount of drug delivered. Pressing down on the canister aligns a hole in the metering valve with the metering chamber. The high propellant vapor pressure quickly forces the metered dose out through this hole and through the actuator nozzle.

Aerosol production takes approximately 20 msec. As the liquid suspension is forced out of the pMDI, it forms a plume, within which the propellants vaporize, or “flash.” Initially, the velocity of this plume is high (approximately 15 m/sec). However, within 0.1 second, the plume velocity decreases to less than half its maximum as the plume moves away from the actuator nozzle. At the same time, propellant evaporation causes the initially large particles (35 µm) generated at the actuator orifice to decrease rapidly in size.

The output volume of pMDIs ranges from 30 to 100 mcl. Approximately 60% to 80% by weight of this spray consists of the propellant, with only approximately 1% being active drug (50 mcg to 5 mg, depending on the drug formulation). For a chlorofluorocarbon (CFC) pMDI used in a standard actuator, loss of drug in the valve stem housing and on the actuator mouthpiece amounts to 10% to 15% of the nominal dose from the metering valve.

From their inception in the mid-1950s to the beginning of the twenty-first century, chlorofluorocarbons (CFCs) such as Freon were the propellants used in pMDIs. Manufacture of CFCs for most applications was prohibited because of the effect of these compounds on global warming, with a period of transition provided for pMDIs. A consortium of eight pharmaceutical companies developed hydrofluoroalkane (HFA)-134a to be more environment-friendly and possibly clinically safer than CFCs.22 Redesign of key components of the pMDI has resulted in improved performance.23

In addition to the propellant, pMDIs use dispersal agents to improve drug delivery by keeping the drug in suspension. The most common dispersal agents are surfactants, such as soy lecithin, sorbitan trioleate, and oleic acid. These agents help keep the drug suspended in the propellant and lubricate the valve mechanism but may also cause adverse responses (coughing or wheezing) in some patients.

Before initial use and after storage, every pMDI should be primed by shaking and actuating the device to atmosphere one to four times (see label for the specific device). Without priming, the initial dose actuated from a new pMDI canister contains less active substance than subsequent actuations.20 This “loss of dose” from a pMDI occurs when drug particles rise to the top of the canister over time (“cream”). A reduction in emitted dose with the first actuation commonly occurs with a pMDI after storage, particularly with the valve pointed in the downward position. Loss of prime is related to valve design and occurs when propellant leaks out of the metering chamber during periods of nonuse (e.g., 4 hours). The result is reduced pressure and drug released with the next actuation.20 Improved designs of metering valves developed for use with HFA propellants reduce these losses. It is recommended that a single dose be wasted before the next dose is inhaled when a CFC pMDI has not been used for 4 to 6 hours. An HFA pMDI requires no wasting of dose for periods exceeding 2 days.

Breath-Actuated Pressurized Metered Dose Inhaler

A variation of a pMDI is a breath-actuated model, which incorporates a trigger that is activated during inhalation. This trigger theoretically reduces the need for the patient or caregiver to coordinate MDI actuation with inhalation.24 However, patients may stop breathing when the MDI is actuated (“cold Freon effect”) or have suboptimal inspiration. Evaluation of the efficacy of breath-actuated pMDIs in children younger than 6 years is limited, and their use should be restricted to older children and adults. Oropharyngeal deposition of steroids using these devices is still very high.

The Aerocount Autohaler is a flow-triggered pMDI developed and marketed by the 3M Corporation (St. Paul, MN) (Figure 36-8). The device is designed to eliminate the need for hand-breath coordination by automatically triggering in response to the patient’s inspiratory effort.24 To use the Autohaler, the patient cocks a lever on the top of the unit, which sets in motion a downward spring force. Using the closed-mouth technique, the patient draws through the mouthpiece. When the patient’s flow rate exceeds 30 L/min, a vane releases the spring, which forces the canister down and triggers the pMDI. In the United States, the Autohaler is available only with pirbuterol, a bronchodilator similar to albuterol. Current data indicate that the device reduces pharyngeal impaction and enhances lung deposition. A possible limitation of the device is that it can be breath actuated only. Patients experiencing an acute exacerbation of bronchospasm may be unable to generate sufficient flows to trigger the Autohaler. This theoretical concern has not been widely observed in clinical studies of patients with severe exacerbation of asthma receiving treatment in emergency departments. Nevertheless, caution may be appropriate in ordering breath-triggered pMDIs for small children and patients prone to severe levels of airway obstruction.

The Easihaler (GlaxoSmithKline, Philadelphia) is a breath-actuated pMDI that has been developed with a range of medications and is currently available in Europe and Canada. Release in the United States is anticipated in the near future. A new generation of pMDIs such as the Tempo (MAP Pharmaceuticals, Mountain View, CA) have been designed to be breath actuated with lower force of the plume exiting the mouthpiece, reducing oropharyngeal deposition and increasing lung dose (Figure 36-9).

Dose Counters

A serious limitation of pMDIs is the lack of a “counter” to indicate the number of doses remaining in the canister. After the number of label doses have been administered, the pMDI may seem to give another 20 to 60 doses, which may deliver little or no medications as the doses “tail-off.” Tail-off effect refers to variability in the amount of drug dispensed toward the end of the life of the canister. The result of tail-off is swings from normal to almost no dose emitted from one breath to the next with no reliable indicator to the user. Without a dose counter, there is no viable method to determine remaining drug in a pMDI other than manually keeping a log of every dose taken. The U.S. Food and Drug Administration (FDA) is requiring all new pMDIs to have a counter technology to track pMDI actuations remaining. Third-party dose counters may be added to older pMDI models but may not have the accuracy of built-in technology (Figure 36-10).

Factors Affecting Pressurized Metered Dose Inhaler Performance and Drug Delivery

Aerosol Delivery Characteristics

Although pMDIs can produce particles in the respirable range (MMAD 2 to 6 µm),20 the initial velocity and dispersion of the aerosol plume generate larger particles that decrease in size as they leave the pMDI, resulting in approximately 80% of the dose leaving the actuator to impact and become deposited in the oropharynx. A significant proportion of this oropharyngeal deposition is swallowed and may be a factor in systemic absorption of some drugs. Pulmonary deposition ranges from 10% to 20% in adults and larger children (less in infants).26 The exact amount of drug delivered to an individual patient is unpredictable because of high variability between patients and because pMDI drug administration is technique-dependent.

Technique

The successful administration of aerosol drugs by pMDI is highly technique-dependent. Two-thirds of patients and health care professionals who should teach pMDI use do not perform the procedure properly.27 Box 36-1 outlines the recommended steps for self-administering a bronchodilator by simple pMDI. Thorough preliminary patient instruction can last 10 to 30 minutes and should include demonstration, practice, and confirmation of patient performance (demonstration pMDIs with placebo are available from manufacturers for this purpose). Repeated instruction improves performance; repeat instruction is done most appropriately with follow-up clinic or home visits. Demonstration and return demonstration must occur several times for best patient adherence to device use.

For best effect, the pMDI should be actuated once at the beginning of inspiration. Common hand-breath coordination problems include actuating the pMDI before or after the breath. Some patients, especially infants, young children, elderly adults, and patients in acute distress, may be unable to coordinate actuation of the pMDI with inspiration. Some patients exhibit a “cold Freon effect,” which occurs when the cold aerosol plume reaches the back of the mouth and the patient stops inhaling. All of these problems reduce aerosol delivery to the lung to the point that the patient does not benefit from the medication, but they can be corrected entirely or in part by use of the proper pMDI accessory device.

Most pMDI labels call for placing the mouthpiece between the lips. However, research has shown that positioning the outlet of the pMDI approximately 4 cm (two fingerbreadths) in front of the mouth improves lung deposition by decreasing oropharyngeal impaction.28 Holding the canister outside the open mouth (at two fingerbreadths) provides a space for the particles to decelerate while evaporating, allowing particle size to reduce to respirable size. Use of the open-mouth technique with a low inspiratory flow rate can result in a doubling of the dose delivered to the lower respiratory tract of an adult from approximately 7% to 10% to 14% to 20%. However, this technique is more difficult for patients to perform reliably than the closed-mouth technique. Although it may reduce oropharyngeal deposition, the technique has not been shown to improve the clinical response to pMDI bronchodilators.

Concerns have been raised about use of the open-mouth technique with ipratropium bromide because poor coordination can result in drug being sprayed into the eyes. Use of anticholinergic agents has been associated with increased ocular pressure, which could be dangerous for patients with glaucoma. For avoidance of ocular exposure, the drug manufacturer recommends patients use the closed-mouth technique with ipratropium.

The high percentage of oropharyngeal drug deposition with use of steroid pMDIs can increase the incidence of opportunistic oral yeast infection (thrush) and changes in the voice (dysphonia). Rinsing the mouth after steroid use can help avoid this problem, but most pMDI steroid aerosol impaction occurs deep in the hypopharynx, which cannot be easily rinsed with gargling. For this reason, steroid pMDIs should not be used alone but always in combination with a spacer or valved holding chamber. See Box 36-2 for instructions for determining dosage left in the pMDI.

Box 36-2   Determining Dose Left in Pressurized Metered Dose Inhaler

Tracking the number of actuations (puffs) remaining in a pMDI can be done with or without dose counters (see Figure 36-10).

Without Dose Counters

The user should29:

1. Read the label to determine how many puffs of drug the pMDI has when full.

2. Calculate how long the pMDI will last by dividing the total number of puffs in the pMDI by the total puffs used per day. If the pMDI is used more often than planned, it will run out sooner.

3. Identify the date that the medication will run out, and mark it on the canister or on a calendar.

4. For drugs that are prescribed to be taken as needed, track the number of puffs of drug administered on a daily log sheet and subtract them from the remaining puffs to determine the amount of medication left in the pMDI.

5. Keep the daily log sheet in a convenient place, such as taped to the bathroom mirror.

6. Refill the pMDI prescription when there are a few days of use remaining in the pMDI.

7. Dispose of the pMDI properly when the last dose is dispensed.

Pressurized Metered Dose Inhaler Accessory Devices

Various pMDI accessory devices have been developed to overcome the two primary limitations of these systems: hand-breath coordination problems and high oropharyngeal deposition. Accessory devices include breath-actuated pMDIs, spacers, and holding chambers.

Spacers and Holding Chambers

Spacers and valved holding chambers are pMDI accessory devices designed to reduce both oropharyngeal deposition and the need for hand-breath coordination. Despite differences in design, all spacers add distance between the pMDI and the mouth, reducing the initial forward velocity of the pMDI droplets, which occurs with partial evaporation of propellant in the time the aerosol traverses the length of the spacer. The reduction in initial forward velocity decreases the number of nonrespirable particles reaching the airway. With retention of the larger droplets in the spacer or holding chamber and evaporation of propellants before entering the airway, the “cold Freon effect,” which causes many children to stop inhaling, is reduced, as is the foul taste associated with some of the drug aerosols. The same drug used with different accessory devices may produce differences in MMAD, GSD, and fine-particle fraction. The quantity of respirable drug available at the spacer or valved holding chamber exit depends on spacer volume and design and on formulation factors. The placement of a valve between the pMDI and the chamber and the mouthpiece works like a baffle reducing the size of particles inhaled. A simple tube spacer may reduce oral deposition by 90%, whereas a valved holding chamber can reduce oral deposition by 99%.

Valved holding chambers protect the patient from poor hand-breath coordination, with exhaled gas venting to the atmosphere, allowing aerosol to remain in the chamber available to be inhaled with the next breath. Valved holding chambers allow infants, small children, and adults who cannot control their breathing pattern to be treated effectively with pMDIs.

It is increasingly common practice to provide asthmatic patients an accessory device to use with the pMDI and to teach them to use the pMDI with and without the accessory device. The patients are instructed to use the device with the pMDI whenever they feel short of breath. Many of these patients find that they get much better relief from the pMDI with an accessory device than with the pMDI alone.

Basic concepts for spacer devices include (1) small volume adapters, (2) open tube designs, (3) bag reservoirs, and (4) valved holding chambers (Figure 36-11). More than a dozen different devices with volumes ranging from 15 to 750 ml have been developed over the past 30 years.

A spacer is a simple valveless extension device that adds distance between the pMDI outlet and the patient’s mouth. This distance allows the aerosol plume to expand and the propellants to evaporate before the medication reaches the oropharynx. Larger particles leaving the pMDI tend to impact on the spacer walls. In combination, this phenomenon reduces oropharyngeal impaction and increases pulmonary deposition. Proper use of a simple open-tube spacer still requires some hand-breath coordination because a momentary delay between triggering and inhaling the discharged spray results in a substantial loss of drug and reduced lung delivery. Exhalation into a simple spacer after pMDI actuation clears the aerosol from the device and wastes most of the dose to the atmosphere. This reduction in dose also occurs with small volume reverse-flow design spacers if there is no provision for “holding” the aerosol in the device.30

Similar to spacers, holding chambers allow the aerosol plume to develop and reduce oropharyngeal deposition. A holding chamber also incorporates one or more valves that prevent aerosol in the chamber from being cleared on exhalation. This allows patients with a small VT to empty the aerosol from the chamber over two or more successive breaths. Generally, holding chambers provide less oropharyngeal deposition, higher respirable drug dosages, and better protection from poor hand-breath coordination than simple spacers.

The MMAD of the aerosol emitted from the pMDI exiting a spacer decreases approximately 25%, whereas the fraction containing particles less than 5 µm in diameter increases. This change is largely due to rapid evaporation of propellant in the spacer. With valved holding chambers, in addition to evaporation of the plume, the valves act as baffles of larger particles, increasing the respirable fraction further.

Holding chambers produce a finer, slower moving, more “respirable” aerosol with less impaction of drug in the oropharyngeal area (1% of dose) than simple spacers (10%) or a pMDI alone (80% of dose). Deposition after inhalation of a radiolabeled pMDI solution aerosol from the AeroChamber, compared with deposition from the same pMDI inhaled with the open-mouth technique, showed a 10-fold to 17-fold decrease in the amount of radioactivity deposited in the oropharyngeal-laryngeal area while a similar lung dose was maintained. This finding was true for both healthy subjects and patients with chronic obstructive pulmonary disease (COPD).31 The advantage of reduced oropharyngeal deposition is fewer side effects from steroid aerosols, as shown in numerous published clinical trials. If multiple actuations of one or more drugs are placed into a spacer, both the total dose and the respirable dose of drug available for inhalation are reduced. The extent of these losses may vary for different drugs and spacer designs.20

Holding chambers with masks are available for use in the care of infants, children, and adults. These units allow effective administration of aerosol from a pMDI to patients who are unable to use a mouthpiece device (because of their size, age, coordination, or mentation). Holding chambers are helpful in administration of pMDI steroids because deposition of the drug in the mouth is largely eliminated, and systemic side effects can be minimized.

Even with a holding chamber, respirable particles containing drug settle out and become deposited within the device, causing a whitish buildup on the inner chamber walls. This residual drug poses no risk to the patient but may be rinsed out periodically. Drug output from plastic spacers has been shown to decrease owing to the presence of an electrostatic charge. With these devices, a buildup of material can be seen on the walls of the chamber. As more material builds up on the wall of the chamber, the charge is dissipated, and more drug is inhaled by the patient. Washing the chamber with water (without soap) causes the electrostatic charge to be reestablished, making the device less effective for the next few puffs, until the static charge in the chamber (which attracts small particles) is again reduced.32 Optimal technique is outlined in Box 36-3.

Use of conductive metal or nonelectrostatic plastic chambers or washing the plastic chamber periodically with deionizing detergent (liquid dishwashing soap) can overcome the loss of fine-particle mass owing to electrostatic charge and increase the inhaled mass from 20% to 50% of the emitted dose of the pMDI, even in children (Figure 36-12).32 The effect of washing the chamber with conventional dishwashing soap reduces this static charge for up to 30 days. All valved holding chambers and spacers should be cleaned regularly, typically monthly, as recommended by the manufacturer. Use of dilute liquid dishwashing soap, with or without rinsing, and allowing to air dry are recommended.

The addition of a one-way valve to convert an open tube into a reservoir for the aerosol, the incorporation of the actuator in the pMDI, the shape of the device, flow of air through the device, edge effects, masks, and manufacturing materials all affect aerosol characteristics and yield. The inhalation valve, which is used to contain the aerosol, also acts as a baffle to reduce oropharyngeal deposition. This valve must be able to withstand the initial pressure from the pMDI when the device is triggered to retain aerosol and have sufficiently low resistance to open readily when the user inhales, in particular, when the user is a child or an infant. Exhalation valves in a face mask attached to a spacer device must also provide low resistance. Issues of spacer volume, VT, frequency of breathing, and mechanical dead space between the spacer and mouth are of particular concern when these devices are used by children.33 Differences of twofold to threefold in the amount of drug available at the mouth have been measured among spacers used at the present time to treat infants. Clinicians should determine the delivery efficiencies of spacer devices before using the devices in the care of a particular population.

Accessory devices are used with either the manufacturer-designed boot that comes with the pMDI or with a “universal adapter” that triggers the pMDI canister. Different formulations of pMDI drugs operate at different pressures and have a different-sized orifice in the boot that is specifically designed by the manufacturer for use exclusively with that pMDI. The output characteristics of a pMDI change when an adapter with a different-sized orifice is used. With HFA pMDIs, the diameter of the actuator orifice is smaller, and the spray is predictably finer. When the HFA pMDI is used in an actuator designed for use with CFC pMDIs, output is reduced. When these HFA formulations are used with any particular spacer, it is important to know how comparable the available dose and particle size distribution are to the dose and particle size from an existing CFC pMDI.20

Dry Powder Inhalers

A DPI is typically a breath-actuated dosing system. With a DPI, the patient creates the aerosol by drawing air though a dose of finely milled drug powder with sufficient force to disperse and suspend the powder in the air. DPIs are inexpensive, do not need propellants, and do not require the hand-breath coordination needed for pMDIs. However, dispersion of the powder into respirable particles depends on the creation of turbulent flow in the inhaler. Turbulent flow is a function of the ability of the patient to inhale the powder with a sufficiently high inspiratory flow rate (Figure 36-13). In terms of both lung deposition and drug response, DPIs are as effective as pMDIs.34

Equipment Design and Function

Most passive dry powder–dispensing systems require the use of a carrier substance (lactose or glucose) mixed into the drug to enable the drug powder to deaggregate more readily and flow out of the device. Reactions to lactose or glucose seem to be fewer than reactions to the surfactants and propellants used in pMDIs, even though the amount of these substances is substantially greater than the amount of the drug and can represent 98% or more of the weight per inhaled dose in some formulations.

As shown in Figure 36-14, there are numerous DPIs on the market, which can be divided into three categories based on the design of their dose containers: (1) unit-dose DPI, (2) multiple unit-dose DPI, and (3) multiple dose drug reservoir DPI.

Unit-dose DPIs, such as the Aerolizer (Schering-Plough, Kenilworth, NJ) and the HandiHaler (Boehringer Ingleheim, Ingelheim am Rhein, Germany), dispense individual doses of drug from punctured gelatin capsules. Multiple unit-dose DPIs (Diskhaler; GlaxoSmithKline, Philadelphia) contain a case of four or eight individual blister packets of medication on a disk inserted into the inhaler. Multiple dose DPIs include the Twisthaler (Schering-Plough), Flexhaler (AstraZeneka, London), and the Diskus (GlaxoSmithKline). The Twisthaler and Flexhaler have a multidose reservoir powder system preloaded with a quantity of pure drug sufficient for dispensing 120 doses of medication, and the Diskus incorporates a tape system that contains up to 60 sealed single doses (Figure 36-15).

The particle size of the dry powder particles of drug ranges from 1 to 3 µm. However, the size of the lactose or glucose particles can range from approximately 20 to 65 µm, so most of the carrier (≤80%) is deposited in the oropharynx.

Factors Affecting Dry Powder Inhaler Performance and Drug Delivery

Performance of DPIs can be affected by the materials used in production and manufacturing.

Patient’s Inspiratory Flow Ability

The high peak inspiratory flow rates (>60 L/min) required to dispense the drug powder from most current DPI designs result in a pharyngeal dose comparable to the dose received from a typical pMDI without an add-on device. If inhalation is not performed at the optimal inspiratory flow rate for a particular device, delivery to the lung decreases as the dose of drug dispensed decreases and the particle size of the powder aerosol increases (Figure 36-16).34

Despite the foregoing issues, DPIs generally are convenient and easy to use. Newer designs are being developed that provide aerosols with higher fine-particle fractions and more reproducible dosing, independent of inspiratory flow rate.

Passive, or patient-driven, DPIs rely on the patient’s inspiratory effort to dispense the dose. The result is differences in lung delivery and clinical response. Active or powered DPI devices, which deaggregate the powder before inhalation, are independent of patient effort. Active DPIs use an energy source to deaggregate the powder and suspend the powder into an aerosol, allowing the dose to be suspended independent of patient inspiratory flow rates. One example of an active DPI is the Exubera powder insulin delivery system (Pfizer, New York). This device is manually powered using pneumatic pressure to disperse the powder into a reservoir chamber from which the aerosol is inhaled.

Technique

As with pMDIs, to derive the maximum benefit from a DPI, proper technique is essential. Box 36-4 outlines the basic steps for ensuring optimal drug delivery. The most critical factor in using a passive DPI is the need for high inspiratory flow. Patients must generate an inspiratory flow rate of at least 40 to 60 L/min to produce a respirable powder aerosol. Because infants, small children (<5 years old) (Figure 36-17), and patients who are unable to follow instructions cannot develop flow this high, these patients cannot use DPIs. Also, patients with severe airway obstruction may be unable to achieve the required flow; DPIs should not be used in the management of acute bronchospasm.

Although hand-breath coordination is not as important with DPIs as it is with pMDIs, exhalation into the device before inspiration can result in loss of drug delivery to the lung. Some devices also require assembly, which can be cumbersome or difficult for some patients, especially in an emergency. It is important that patients receive demonstrations with their inhalers and have the opportunity to assemble and use the DPI (return demonstration) before self-administration. Although the DPI may require cleaning in accordance with the product label, the device should never be submerged in water. Moisture in the device dramatically reduces available dose. Based on the different types of DPIs and the various drug container closure systems, Table 36-1 provides methods to determine the dose in the DPI.

TABLE 36-1

Determining Doses Left in the Dry Powder Inhaler

    Drug Container Doses Type Indicator Meaning of Dose Indicator
Unit-Dose DPI Aerolizer or HandiHaler Single capsule 1 None Check capsule to ensure full dose was inhaled. Repeat to empty capsule
Multiple Unit-Dose DPI Diskhaler Dose blisters 4 or 8 None Inspect visually to confirm use of all blisters
  Diskus Blister strip 60 Red numbers Red numbers indicate that ≤5 doses are left in DPI
Multiple Dose DPI Flexhaler Reservoir 60 or 120 “0” Marked in intervals of 10 doses; “0” indicates empty
Twisthaler Reservoir 30 “01” “01” indicates last dose

image

Nebulizers

Nebulizers generate aerosols from solutions and suspensions. The three categories of nebulizers include (1) pneumatic jet nebulizers, (2) USNs, and (3) VM nebulizers. Nebulizers are also described in terms of their reservoir size. Small volume nebulizers (SVNs) most commonly used for medical aerosol therapy hold 5 to 20 ml of medication. Large volume nebulizers, also known as jet nebulizers, hold up to 200 ml and may be used for either bland aerosol therapy (see Chapter 35) or continuous drug administration.

Pneumatic (Jet) Nebulizers

Gas-powered jet nebulizers (Figure 36-18, A) have been in clinical use for longer than 100 years. Most modern jet nebulizers are powered by high-pressure air or oxygen (O2) provided by a portable compressor, compressed gas cylinder, or 50-psi wall outlet.

Factors Affecting Nebulizer Performance

Nebulizer design, gas pressure, gas density, and medication characteristics affect SVN performance (Box 36-5).

Nebulizer Design

As shown in Figure 36-18, B, a typical SVN is powered by a high-pressure stream of gas directed through a restricted orifice (the jet). The gas stream leaving the jet passes by the opening of a capillary tube immersed in solution. Because it produces low lateral pressure at the outlet, the high jet velocity draws the liquid up the capillary tube and into the gas stream, where it is sheared into filaments of liquid that break up into droplets. This primary spray produces a heterodisperse aerosol with droplets ranging from 0.1 to 500 µm.35

This spray is directed against one or more baffles. A baffle is a surface on which large particles impact and fall out of suspension, whereas smaller particles remain in suspension, reducing the size of particles remaining in the aerosol. A sphere or plate placed in line with the jet flow can serve as a baffle, as can the internal walls of the nebulizer, the surface of the solution used for nebulization, or the internal walls of the delivery system. In many designs, droplets that impact baffles in the SVN return to the medication reservoir for nebulization again.

Baffles are key elements in nebulizers; well-designed baffling systems decrease both the MMAD (size) and the GSD (range of sizes) of the generated aerosol. Atomizers operate with the same basic principles as nebulizers without baffling and produce aerosols with larger MMAD and GSD. Unintentional baffles are created by the angles within delivery tubing, by interfaces with other devices outside the aerosol generator, and by the surfaces of the upper airway itself.36

Residual drug volume, or dead volume, is the medication that remains in the SVN after the device stops generating aerosol and “runs dry.”37 The residual volume of a 3-ml dose can range from 0.5 to more than 2.2 ml, which can be more than two-thirds of the total dose. The greater the residual drug volume, the more drug that is unavailable as aerosol, and the less efficient the delivery system. Residual volume also depends on the position of the SVN. Some SVNs stop producing aerosol when tilted 30 degrees from vertical. Increasing the fill volume allows a greater proportion of active medication to be used for nebulization. In a nebulizer with a residual volume of 1.5 ml, a fill of 3 ml would leave only 50% of the nebulizer charge (nominal dose) available for nebulization. In contrast, a fill of 5 ml would make 3.5 ml, or more than 70% of the medication, available to be inhaled. The unit-dose volumes of drugs were based on clinical response of patients using nebulizers with substantial residual drug volumes. Although increasing dose volume may increase available dose, it should be considered off-label administration, and no significant difference in clinical response has been shown to date with varying diluent volumes and flow rates.

Gas Source (Hospital versus Home)

Gas pressure and flow through the nebulizer affect particle size distribution and output. Within operating limits, the higher the pressure or flow, the smaller the particle size, the greater the output, and the shorter the treatment time. A nebulizer that produces an MMAD of 2.5 µm when driven by a gas source of 50 psi at 6 to 10 L/min may produce an MMAD of more than 5 µm when operated on a home compressor (or ventilator) developing 10 psi. Too low a gas pressure or flow can result in negligible nebulizer output. Consequently, nebulizers used for home care should be matched to the compressor according to data supplied by the manufacturer so that the combination of specific equipment performs efficient nebulization of the desired medications prescribed for the patient. In Europe, equipment manufacturers are required to characterize the performance of their nebulizer and compressor combinations using standardized methods so that consumers can compare performance. Until such time that similar standards are required in the United States, clinicians should ascertain whether the system prescribed meets these criteria.

Other concerns in the use of disposable nebulizers with compressors at home involve possible degradation of performance of the plastic device over multiple uses. One study showed that repeated use did not alter MMAD or output as long as the nebulizer was cleaned properly. Failure to clean the nebulizer properly resulted in degradation of performance because of clogging of the Venturi orifice, reducing the output flow, and buildup of electrostatic charge in the device.

Density

Gas density affects both aerosol generation and delivery to the lungs. The lower the density of a carrier gas, the less turbulent the flow (i.e., the lower the Reynolds number), resulting in less aerosol impaction. This phenomenon has been shown with low-density helium-O2 mixtures (heliox). The lower the density of a carrier gas, the less aerosol impaction occurs as gas passes through the airways, and the greater the deposition of aerosol in the lungs.38 However, when heliox is used to drive a jet nebulizer at standard flow rates, aerosol output is substantially less than with air or O2, and aerosol particles are considerably smaller. When driving a nebulizer with heliox, twofold to threefold greater flow is required to produce a comparable aerosol output. Heliox concentrations of 40% or greater have been shown to improve aerosol deposition.39

Humidity and Temperature

Humidity and temperature can affect particle size and the concentration of drug remaining in the nebulizer. Evaporation of water and adiabatic expansion of gas can reduce the temperature of the aerosol to 10° C less than ambient temperature. This cooling may increase solution viscosity and reduce the nebulizer output, while decreasing particle MMAD.40 Aerosol particles entrained into a warm and fully saturated gas stream increase in size. These particles also can coalesce (stick together), increasing the MMAD further and, in the case of a DPI, can severely compromise the output of respirable particles. How much these particles enlarge depends primarily on the tonicity of the solution. Aerosols generated from isotonic solutions probably maintain their size as they enter the respiratory tract. Hypertonic solutions tend to enlarge, whereas evaporation can cause hypotonic droplets to evaporate and shrink.

Small Volume Nebulizers

Four categories of jet SVNs include (1) continuous nebulizer with simple reservoir, (2) continuous nebulizer with collection reservoir bag, (3) breath-enhanced nebulizer, and (4) breath-actuated nebulizer (Figure 36-19). The most commonly used SVN is the constant output design. Supplemental air is entrained across the top of the device and dilutes the aerosol produced within the nebulizer as it exits toward the patient. Aerosol is generated continuously, with 30% to 60% of the nominal dose being trapped as residual volume in the nebulizer, and more than 60% of the emitted dose is wasted to the atmosphere. Continuous nebulization wastes medication because the aerosol is produced throughout the respiratory cycle and is largely lost to the atmosphere, as shown in Figure 36-20. Patients with an I : E ratio of 40 : 60 (or 1 : 1.5) lose 60% of the aerosol generated to the atmosphere. If 50% of the total dose is emitted from the nebulizer, and 50% of that aerosol is in the respiratory range and 40% of that is inhaled by the patient, less than 10% deposition is commonly measured in adults receiving continuous nebulizer therapy. In neonates and infants, given the small minute volumes and small airways with increased impaction and reduced sedimentation, deposition can be only 0.5%.

Aerosolized medication can also be conserved with reservoirs.41 A reservoir on the expiratory limb of the nebulizer conserves drug aerosol.

Small Volume Nebulizer With a Reservoir

Many types of disposable SVNs are packaged with a 6-inch (15-cm) piece of aerosol tubing to be used as a reservoir (see Figure 36-18, A). This may increase inhaled dose by 5% to 10% or increase the inhaled dose from 10% to approximately 11% with the reservoir tube.

Continuous Small Volume Nebulizer With Collection Bag

Bag reservoirs hold the aerosol generated during exhalation and allow the small particles to remain in suspension for inhalation with the next breath, while larger particles rain out, attributed to a 30% to 50% increase in inhaled dose.41 A collection bag is attached on the expiratory side of the nebulizer “T,” which collects aerosol leaving the SVN when the patient is not actively inhaling. Some of the aerosol in the bag is inhaled with the next inspiration, increasing total dose efficiency. The patient inhales aerosol from the SVN and reservoir through a one-way valve with exhalation through a second valve to the atmosphere.

Breath-Enhanced Nebulizers

Breath-enhanced nebulizers generate aerosol continuously, using a system of vents and one-way valves to minimize aerosol waste.42 In the Pari LC Sprint (Pari, Midlothian, VA) breath-enhanced nebulizer (Figure 36-21), an inspiratory vent allows the patient to draw in air through the nebulization chamber generating and containing aerosolized drug. On exhalation, the inlet vent closes, and aerosol exits by a one-way valve near the mouthpiece; this process can increase inhaled mass by 50% over standard continuous nebulizers and reduces aerosol waste to the atmosphere.

Breath-Actuated Nebulizers

Breath-actuated nebulization can increase inhaled aerosol mass by threefold to fourfold over conventional continuous nebulization. Historically, this increase was accomplished with a patient-controlled finger port that directed gas to the nebulizer only during inspiration. Although this system wastes less aerosol, it can increase treatment time fourfold. This approach requires good hand-breath coordination, something not all patients possess, especially when they are distressed.

Breath-actuated SVNs have been introduced that synchronize aerosol generation based on the patient’s breathing pattern. Breath-actuated nebulizers generate aerosol only during inspiration. This feature eliminates waste of aerosol during exhalation and increases the delivered dose threefold or more over continuous and breath-enhanced nebulizers. Dosimeters, used in pulmonary function laboratories, sense inspiration and pulse airflow to the jet orifice and transform a conventional nebulizer into a breath-actuated system.

AeroEclipse (Trudell Medical International, London, Ontario, Canada) is a breath-actuated SVN. A unique, spring-loaded, one-way valve design draws the jet to the capillary tube during inspiration and causes nebulization to cease when the patient’s inspiratory flow decreases below the threshold or the patient exhales into the device (Figure 36-22). Expiratory pressure on the valve at the initiation of exhalation moves the nebulizer baffle away from its position directly above the jet orifice, reduces the pressure, and stops aerosolization. Because aerosol is generated only during inhalation, exhaled aerosol and contamination of the environment during the expiratory phase of the breathing cycle are largely eliminated.

It can be difficult to determine when a nebulizer treatment is complete. Malone and colleagues43 found that with three different fill volumes, albuterol delivery from the nebulizer ceased after the onset of inconsistent nebulization (sputtering) (Figure 36-23). Aerosol output declined one-half within 20 seconds of the onset of sputtering. The concentration of albuterol in the nebulizer cup increased significantly when the aerosol output declined, and further weight loss in the nebulizer was caused primarily by evaporation. The authors concluded that aerosolization past the point of initial nebulizer sputter is ineffective.

Table 36-2 summarizes some of these key factors for many commercially available SVNs.44 Numerous SVNs are on the market, and they vary widely in design and performance. SVNs of the same design and lot number can exhibit variable performance, even to the point that some nebulizers of the same model number do not work at all.45 Managers and clinicians always must evaluate SVNs carefully before purchasing or using them. Manufacturers should provide data on the performance of their nebulizers under common use conditions.

TABLE 36-2

Comparison of Different Nebulizers

  Jet Ultrasonic Vibrating Mesh
Features      
Power source Compressed gas or electrical mains Electrical mains Batteries or electrical mains
Portability Restricted Restricted Portable
Treatment time Long Intermediate Short
Output rate Low Higher Highest
Residual volume 0.8-2.0 ml Variable but low ≤0.2 ml
Environmental Contamination      
Continuous use High High High
Breath-activated Low Low Low
Performance variability High Intermediate Low
Formulation Characteristics      
Temperature Decreases* Increases Minimum change
Concentration Increases Variable Minimum change
Suspensions Low efficiency Poor efficiency Variable efficiency
Denaturation Possible Probable Possible
Cleaning Required, after single use Required, after multiple use Required after single use
Cost Very low High High

image

*For jet nebulizers, the temperature of the reservoir fluid decreases about 15° C during nebulization because of evaporation.

For ultrasonic nebulizers, vibration of the reservoir fluid causes a temperature increase during aerosol generation, which can be 10° C to 15° C.

Denaturation of DNA occurs with all the nebulizers.

Modified from Dolovich MB, Dhand R: Aerosol drug delivery: developments in device design and clinical use. Lancet 377:1032, 2011.

Technique

Box 36-6 outlines the optimal technique for using an SVN for aerosol drug delivery. Use of an SVN is less technique-dependent and device-dependent than use of a pMDI or DPI delivery system. Slow inspiratory flow optimizes SVN aerosol deposition. However, deep breathing and breath holding during SVN therapy do little to enhance deposition over normal tidal breathing.46 Because the nose is an efficient filter of particles larger than 5 mm, many clinicians prefer not to use a mask for SVN therapy. As long as the patient is mouth breathing, there is little difference in clinical response between therapy given by mouthpiece and therapy given by mask. The selection of delivery method (mask or mouthpiece) should be based on patient ability, preference, and comfort.

Infection Control Issues

The CDC recommends that nebulizers be cleaned and disinfected, or rinsed with sterile water, and air dried between uses. Oie and Kamiya,47 studying microbial contamination of antibiotic aerosol solutions, found that after 7 days, five of six solutions were contaminated. The contamination appeared to have been caused by storage of multiple dose solutions at room temperature instead of in a refrigerator and reuse of syringes for measuring the solution. Refrigerating solutions and discarding syringes every 24 hours eliminated bacterial contamination.

Large Volume Jet Nebulizers

Large volume jet nebulizers are also used to deliver aerosolized drugs to the lung. A large volume nebulizer is particularly useful when traditional dosing strategies are ineffective in the management of severe bronchospasm. When a patient with airway obstruction does not respond to a standard dosage of bronchodilator, it is common to repeat the treatment every 15 minutes. An alternative approach is to provide continuous nebulization with a specialized large volume nebulizer.

The high-output extended aerosol respiratory therapy nebulizers HEART (Cardinal Health, Dublin, OH), and HOPE (B & B Medical Technologies, Carlsbad, CA) are examples of devices designed for this purpose. These nebulizers have a reservoir greater than 200 ml that produces an aerosol with an MMAD of 2.2 to 3.5 µm. Actual output and particle size vary with the pressure and flow at which the nebulizer operates. A potential problem with continuous bronchodilator therapy (CBT) is increase in drug concentration. Patients receiving CBT need close monitoring for signs of drug toxicity (e.g., tachycardia and tremor). An additional strategy is to use an intravenous infusion pump to drip premixed bronchodilator solution into a standard SVN. Although an equipment-intensive approach, this technique can provide dosing equivalent to every 15 minutes.48

Another special-purpose large volume nebulizer is a small particle aerosol generator (SPAG) (Figure 36-24). The SPAG was manufactured by ICN Pharmaceuticals specifically for administration of ribavirin (Virazole) to infants with respiratory syncytial virus infection. The device is unique in clinical respiratory care practice in that it incorporates a drying chamber with its own flow control to produce a stable aerosol. The SPAG reduces medical gas source from the normal 50 pounds per square inch gauge (psig) line pressure to 26 psig with an adjustable regulator. The regulator is connected to two flowmeters that separately control flow to the nebulizer and drying chamber. The nebulizer is located within the glass medication reservoir, the fluid surface and wall of which serve as primary baffles. As it leaves the medication reservoir, the aerosol enters a long, cylindrical drying chamber. Here the second (separate) flow of dry gas is entrained, reducing particle size by evaporation, creating a monodisperse aerosol with an MMAD of 1.2 to 1.4 µm. Nebulizer flow should be maintained at approximately 7 L/min with total flow from both flowmeters not less than 15 L/min. The latest model operates consistently even with back pressure and can be used with masks, hoods, tents, or ventilator circuits.

Two specific problems are associated with SPAG use to deliver ribavirin. The first is caregiver exposure to the drug aerosol. Approaches to limit caregiver exposure are discussed later (see the section on Controlling Environmental Contamination). The other problem occurs only when the SPAG is used to deliver ribavirin through a mechanical ventilator circuit. Drug precipitation can jam breathing valves or occlude the ventilator circuit. This problem can be overcome by (1) placing a one-way valve between the SPAG and the circuit and (2) filtering out the excess aerosol particles before they reach the exhalation valve, changing filters frequently to avoid increasing expiratory resistance.46

Hand-Bulb Atomizers and Spray Pumps

Hand-bulb atomizers and nasal spray pumps are used to administer sympathomimetic, anticholinergic, antiinflammatory, and anesthetic aerosols to the upper airway, including nasal passages, pharynx, and larynx (see also Chapter 32). These agents are used to manage upper airway inflammation and rhinitis, to provide local anesthesia, and to achieve systemic effects. Guidelines for the delivery of drugs to the upper airway have been developed by the American Association for Respiratory Care (AARC).49

Because the spray pump generates relatively low pressure and does not have baffles, it produces an aerosol suspension with large particle size (high MMAD and GSD), which are ideal for upper airway deposition. (Nasopharyngeal deposition is greatest for particles 5 to 20 µm.) Deposition with the hand-bulb atomizer applied to the nose occurs mostly in the anterior nasal passages with clearance to the nasopharynx. The 100-mcl puffs appear to deposit more medication than 50-mcl puffs, and deposition to a greater surface area occurs with a 35-degree spray angle than with a 60-degree angle.

Ultrasonic Nebulizers

The USN uses a piezoelectric crystal to generate an aerosol. The crystal transducer converts an electrical signal into high-frequency (1.2- to 2.4-MHz) acoustic vibrations. These vibrations are focused in the liquid above the transducer, where they disrupt the surface and create oscillation waves (Figure 36-25). If the frequency of the signal is high enough and its amplitude strong enough, the oscillation waves form a standing wave that generates a geyser of droplets that break free as fine aerosol particles.

USNs are capable of higher aerosol outputs (0.2 to 1.0 ml/min) and higher aerosol densities than conventional jet nebulizers. Output is determined by the amplitude setting (sometimes user-selected); the greater the signal amplitude, the greater the nebulizer output. Particle size is inversely proportional to the frequency of vibrations. Frequency is device-specific and is not user-adjustable. For example, the DeVilbiss (Somerset, PA) Portasonic nebulizer operating at a frequency of 2.25 MHz produces particles with an MMAD of 2.5 µm, whereas the DeVilbiss Pulmosonic nebulizer operating at 1.25 MHz produces particles in the 4- to 6-µm range. Particle size and aerosol density also depend on the source and flow of gas conducting the aerosol to the patient.

Large Volume Ultrasonic Nebulizers

Large volume USNs (used mainly for bland aerosol therapy or sputum induction) incorporate air blowers to carry the mist to the patient (see Chapter 35). Low flow through the USN is associated with smaller particles and higher mist density. High flow yields larger particles and less density. In contrast to jet nebulizers, the temperature of the solution placed in a USN increases during use. As the temperature increases, the drug concentration increases, as does the likelihood of undesired side effects.

Small Volume Ultrasonic Nebulizers

Many small volume USNs have been marketed for aerosol drug delivery (see Figure 36-25). In contrast to the larger units, some of these systems do not use a couplant compartment; the medication is placed directly into the manifold on top of the transducer. The transducer is connected by a cable to a power source, often battery-powered to increase portability. These devices have no blower; the patient’s inspiratory flow draws the aerosol from the nebulizer into the lung.

Small volume USNs have been promoted for administration of a wide variety of formulations ranging from bronchodilators to antiinflammatory agents and antibiotics.50 Use of a small volume USN may increase available respirable mass for designs with less residual drug volume than SVNs; this may reduce the need for a large quantity of diluent to ensure delivery of the drugs. The contained portable power source adds a great deal of convenience in mobility. Both theoretical advantages of the ultrasonic devices are outweighed by relatively high purchase costs and poor reliability.

Small volume USNs have been used to administer undiluted bronchodilators to patients with severe bronchospasm.50 Because the nebulizers have minimal residual drug volume, the treatment time is reduced with smaller volumes; however, it may be increased with standard dosing volumes. Use of undiluted bronchodilators has been described in the literature, but this is not included at the present time in the manufacturer’s label on product dosing information. Some ventilator manufacturers (e.g., Maquet, Rastatt, Germany) have promoted the use of USNs for administration of aerosols during mechanical ventilation. In contrast to SVNs, USNs do not add extra gas flow to the ventilator circuit during use. This feature reduces the need to change and reset ventilator and alarm settings during aerosol administration.51

Vibrating Mesh Nebulizers

Two types of VM nebulizers, active and passive, are available commercially.52 Active VM nebulizers use a dome-shaped aperture plate, containing more than 1000 funnel-shaped apertures. This dome is attached to a plate that is also attached to a piezoceramic element that surrounds the aperture plate. Electricity applied to the piezoceramic element causes the aperture plate to be vibrated at a frequency of approximately 130 kHz (or one-tenth that of a USN), moving the aperture plate up and down by 1 µm or 2 µm, creating an electronic micropump. The plate actively pumps the liquid through the apertures, where it is broken into fine droplets. The exit velocity of the aerosol is low (<4 m/sec), and the particle size can range from 2 to 3 µm (MMAD), varying with the exit diameter of the apertures (Figure 36-26). Examples of an active VM nebulizer include the Aeroneb Go, Pro, and Solo nebulizers (Aerogen, Inc, Galway, Ireland) and the eFlow (Pari, Midlothian, VA). An active VM nebulizer can provide nebulization with single drops 15 mcl of formulations containing small and large molecules, suspensions, microsuspensions, and liposomes.

Passive VM nebulizers use a mesh separated from an ultrasonic horn by the liquid for nebulization. A piezoelectric transducer vibrates the ultrasonic horn, which pushes fluid through the mesh. Passive VM nebulizers include the NEU-22 (Omron, Kyoto, Japan) and the I-Neb (Philips Respironics, Murraysville, PA).

The residual drug volumes with either type of VM nebulizer range from 0.1 to 0.4 ml, in contrast to other types of liquid aerosol generators with residual drug volumes of 0.8 to 1.5 ml. Because a greater percentage of standard unit doses are emitted as aerosol, care should be exercised when transitioning to these devices to ensure that the higher dose does not create adverse effects.

New-Generation Nebulizers

Low-velocity (soft mist) aerosol, smaller particle size distribution, and systems that minimize residual volume of medication left in the nebulizer substantially improve aerosol device efficiency. Along with improved performance, some “smart” nebulizers have the capability to monitor patient compliance and aid in managing the patient’s treatment schedule.

With pulmonary deposition increased from the old standard of approximately 10% to more than 60% of the nominal dose, these recent device improvements may be accompanied by greater systemic side effects, unless the delivered dose is reduced. The key is to be able to target an effective delivered dose to the lungs.

New Nebulizer Designs for Liquids

New nebulizer designs are available for delivery of liquids.53

AERx

The AERx device (Aradigm Corp, Hayward, CA) uses a drug solution in a unit-dose, sterile, preservative-free blister pack containing 25 to 50 mcl of fluid. The drug is extruded under pressure through a nozzle containing many small, precision-drilled holes that produce a fine, respirable spray on inhalation. The aerosolization nozzle is part of the disposable blister and is not reused. The dose from a single blister is metered in approximately 1.5 seconds. The emitted dose is more than 70% of the dose contained in the blister with an inspiratory flow rate range of 30 to 85 L/min. The AERx device is being tested for use with numerous drugs in liquid form for both topical and systemic therapy. The AERx device has built-in electronic monitoring capabilities for measuring inspiratory flow rate (IFR) during dosing and for triggering and dispensing the dose at the appropriate inspiratory flow rate for optimal delivery. The dose administered is logged to provide a record of treatments and an indication of patient compliance with therapy. The AERx device is in clinical trials in the United States.

Respimat

The Respimat soft mist inhaler (Boehringer, Ingelheim am Rhein, Germany) is a small hand-held inhaler that uses mechanical energy to create an aerosol from liquid solutions to produce a low-velocity spray (10 mm/sec) that delivers a unit dose of drug in a single actuation. To operate the device, patients twist the body of the device to load an internal spring, place the mouthpiece of the Respimat between the lips, and press a button to release the drug through a uniblock to create spray, which is released over 1.1 to 1.4 seconds, depending on the formulation configuration. The Respimat device requires hand-breath coordination on the part of the patient, as does a pMDI, but because of the longer spray time, it seems more likely to get a greater percent of emitted dose despite coordination issues. Because of the small particle size and low-velocity spray, pulmonary deposition of 40% is independent of inspiratory flows with oral deposition (40%) half the oral dose used with most pMDIs and DPIs (80%). The Respimat is currently available with several drugs in Europe and is slated for introduction with tiotropium in the United States.54

Smart Nebulizers

The I-Neb (Phillips Respironics, Murrysville, PA) is a breath-actuated passive VM nebulizer with adaptive aerosol delivery that monitors pressure changes and inspiratory time for the patient’s first three consecutive breaths (Figure 36-27).55 Drug is then aerosolized over 50% of the inspiratory maneuver during the fourth and all subsequent breaths. Targeted inhalation mode guides the patient to take serially longer inspirations to achieve optimal inhalation duration, reducing the time for administration. When the prescribed emitted dose has been aerosolized, the system provides an audible signal indicating the treatment should be stopped and the remaining medication discarded. Built-in electronics monitor patient treatment schedules and delivered doses with the goal to improve compliance with therapy. The I-Neb has been released for delivery of prostacyclin.

The Akita (Activaero, Gemuenden/Wohra, Germany) allows controlled inhalation of aerosol produced by either a jet or VM nebulizer. The Akita controls inspiratory flow to keep it slow (12 to 15 L/min) reducing impaction loss of aerosols in the upper airways. Patient pulmonary function is stored on a smart card programmed to tell the device when to generate aerosol during inspiration. Aerosol generated early targets distal airways, whereas aerosol generated later in the breath targets larger, more central airways.56 Smart nebulizers can track the actual time, duration, and dose administered for each treatment and provide logs of use that can be downloaded for the medical or research record.

Advantages and Disadvantages of Aerosol Systems

Knowledge of the advantages and disadvantages of various aerosol drug delivery systems is crucial for proper selection and application. Table 36-3 compares pMDI, DPI, SVN, and USN delivery systems.

TABLE 36-3

Advantages and Disadvantages of Aerosol Drug Delivery Systems

Advantages Disadvantages
MDI
Convenient
Inexpensive
Portable
No drug preparation required
Difficult to contaminate
Patient coordination required
Patient activation required
High percentage of pharyngeal deposition
Risk of abuse
Difficult to deliver high doses
Not all medications are available
Most units still use ozone-depleting CFCs
MDI With Accessory Device
Less patient coordination required
Less pharyngeal deposition
No drug preparation required
More complex for some patients
More expensive than MDI alone
Less portable than MDI alone
Not all medications available
DPI
Less patient coordination required
Breath-activated
Breath hold not required
Can provide accurate dose counts
No CFCs
Requires high inspiratory flow
Most units are single dose
Risk of pharyngeal deposition
Not all medications are available
Difficult to deliver high doses
SVN
Inexpensive
Less patient coordination required
High doses possible (even continuous)
No CFC release
Wasteful
Drug preparation required
Contamination possible if device not cleaned carefully
Not all medications available
Pressurized gas source required
Long treatment times
USN
Moderate residual volume
Quiet
Smaller residual drug volume than SVN
Aerosol accumulates during exhalation
Expensive
Prone to electrical or mechanical breakdown
Not all medications available
Drug preparation required
VM Nebulizer  
Low residual volume
Quiet
Does not require gas or propellant
Flow-independent delivery
Shorter treatment times
Expensive
Not all medications available
Drug preparation required

image

Modified from Hess D: Aerosol delivery. Respir Care Clin N Am 1:235, 1995.

Special Medication Delivery Issues for Infants and Children

Children and infants have a smaller airway diameter than adults. In addition, their breathing rate is faster, nose breathing filters out large particles and deposits more medication in the upper airway, and mouthpiece administration often cannot be used before 3 years of age. Patient cooperation and ability vary with age and developmental ability. Finally, infants and small children have lower minute volumes than adults and so inhale a smaller proportion of the output of a continuous nebulizer than adults.7

Normal tidal breathing is the most effective method for administering aerosols to an infant. Mouth breathing enhances medication delivery to the airways of adults, but there is little evidence to show that this is true for infants, who are preferential nose breathers up to 1 year of age. Aerosols should never be administered to a crying child. Crying greatly reduces lower airway deposition of aerosol medication (Figure 36-28).

For infants and children who can tolerate a mask, a medication nebulizer can be fitted to an appropriately sized aerosol mask. There is no difference in clinical response between mouthpiece and close-fitting mask treatment, so patient tolerance, compliance, and preference should guide selection of the device. There is evidence that the aerosol available to the patient is substantially less when a loosely fitting mask (>1-cm leak) is used rather than a snug mask or mouthpiece with either a nebulizer or a pMDI with a holder chamber.57,58 If a patient cannot tolerate mask treatment (e.g., will not wear a close-fitting mask), a commonly used strategy is “blow-by” technique, in which the practitioner directs the aerosol from the nebulizer toward the patient’s nose and mouth from a distance of several inches from the face. There are no peer-reviewed published data supporting the use of the blow-by technique. Studies suggest that almost no drug enters the airway with this method. Rather than “blow-by,” it may be more efficient to take the time to condition the infant or child to tolerate the mask without crying or to deliver medication with a close-fitting mask when the patient is asleep.59

Given the cognitive and functional limitations of very young patients, not all delivery devices are suitable for these patients. To help guide clinicians, the accompanying Rule of Thumb outlines age-specific guidelines for using aerosol devices in pediatric and neonatal patients.

Spontaneous breathing in all patients, including pediatric and neonatal patients, results in greater deposition of aerosol from an SVN than occurs with positive pressure breaths (e.g., intermittent positive pressure ventilation). This mode of ventilation reduces aerosol deposition more than 30% compared with the effect of spontaneously inhaled aerosols.60

Selecting an Aerosol Drug Delivery System

The American College of Chest Physicians commissioned an extensive evidence-based review of the literature to determine which type of aerosol delivery system is superior. It was concluded that pMDIs, DPIs, and nebulizers all work with comparable clinical results, as long as they are prescribed for the appropriate patients and are used properly.61 Consequently, clinicians need to know the strengths and limitations of each type of device, match the device to each patient, and ensure that the patient or caregiver is trained to use the device properly.62

To guide practitioners in selecting the best aerosol delivery system for a given clinical situation, the AARC has published relevant clinical practice guidelines for aerosol delivery to the upper airway,49 to the lung parenchyma,63 and to neonatal and pediatric patients.64 Figure 36-29 is a selection algorithm that provides guidance regarding device selection.

Regardless of the device used, the clinician must be aware of the limitations of aerosol drug therapy. First, depending on the device and patient, 10% or less of drug emitted from an aerosol device may be deposited in the lungs (Figure 36-30). As indicated in Box 36-7, additional reductions in lung deposition can occur in many clinical situations that sometimes necessitate the use of higher dosages. Clinical efficacy varies according to both patient technique and device design. For these reasons, the best approach to aerosol drug therapy is to use an assessment-based protocol that emphasizes individually tailored therapy modified according to patient response.

Assessment-Based Bronchodilator Therapy Protocols

Although the choice of delivery system affects how well an aerosolized drug works, it is ultimately the patient’s response that determines the therapeutic outcome. Because patients vary markedly in response to the dose and route of drug administration, it makes sense to tailor aerosol drug therapy to each patient. This approach is best determined with an assessment-based protocol.

Sample Protocol

Figure 36-31 is an algorithm underlying a bronchodilator therapy protocol for acutely ill adults or children admitted to an emergency department.65 The protocol relies heavily on bedside assessment of the severity of airway obstruction based on the patient’s response to varying drug dosages.

According to the algorithm, a patient with acute airway obstruction (wheezing, cough, dyspnea, and peak expiratory flow rate [PEFR] <60% of predicted value) would receive up to three SVN treatments with a standard dose of albuterol, repeated at 20-minute intervals, or 4 puffs of pMDI albuterol with a holding chamber (up to 12 puffs). Each treatment is followed by a dose-response assessment to determine the “best” dose. Once determined, this best dose, with the pMDI or SVN, is repeated 1 hour later, then every 4 hours as needed, supplemented with patient education. If use of the SVN or pMDI with holding chamber fails to relieve the symptoms, CBT with 15 mg/hr albuterol is generally started.

Assessing Patient Response

Careful, ongoing patient assessment is key to an effective bronchodilator therapy protocol. To guide practitioners in implementing effective bedside assessment, the AARC has published Clinical Practice Guideline: Assessing Response to Bronchodilator Therapy at Point of Care.66

Use and Limitations of Peak Flow Monitoring

Because the peak flow measurement is effort-dependent and volume-dependent, evaluation of patient performance is subjective, and there are no good acceptability criteria. In addition, agreement between conventional spirometry values, such as forced vital capacity (FVC) and FEV1, and bedside PEFR values may be poor for individual patients. Although peak flow measurement can be used at the bedside to assess treatment effectiveness and to monitor trends, conventional spirometry remains the standard for determining bronchodilator response.66

Some peak flowmeters are more accurate and reliable than others. Even different units of the same model may give variable results. For this reason, the AARC recommends that when monitoring trends, the same unit be used for a given patient and that the patient’s range be reestablished if a different flowmeter is used.66

Other Components of Patient Assessment

Sole dependence on tests of expiratory airflow for assessing patient response to therapy is unwise because not all patients can perform these maneuvers. Other components of patient assessment useful in evaluating bronchodilator therapy include patient interviewing and observation, measurement of vital signs, auscultation, blood gas analysis, and oximetry.

When possible, the patient should be interviewed to determine the pertinent respiratory history and current level of dyspnea. A validated dyspnea rating scale may be useful for this purpose. Initial determination of patient age and level of consciousness is helpful in selecting both delivery device and starting drug dosage. Observing the patient for signs of increased work of breathing (e.g., tachypnea, accessory muscle use) provides a baseline for assessing status as therapy progresses. Restlessness, diaphoresis, and tachycardia also may indicate severity of airway obstruction but must not be confused with bronchodilator overdose.

Increased cough has been associated with the onset of asthma. The frequency, severity, and effectiveness of cough should be assessed before and after therapy.

In terms of breath sounds, a decrease in wheezing accompanied by an overall decrease in the intensity of breath sounds indicates worsening airway obstruction or patient fatigue. Improvement is indicated when wheezing decreases and the overall intensity of breath sounds increases.

All patients with acute airway obstruction should be monitored for oxygenation status with pulse oximetry. This value can be used in conjunction with observational assessment to titrate the level of inspired O2 given to the patient (see Chapter 35). Arterial blood gases are not essential for determining patient response to bronchodilator therapy but may be needed for patients in severe distress to assess for hypercapnic respiratory failure.

Frequency of Patient Assessment

How frequently patients should undergo assessment for bronchodilator therapy depends primarily on the acuity of the condition. A patient in unstable condition and in acute distress should undergo closer and more frequent scrutiny than a patient in stable condition. Box 36-8 provides guidance regarding the frequency of assessment according to acuity.

Box 36-8   Frequency of Assessment of Bronchodilator Therapy

Patient Education

The desired outcome of all bronchodilator protocols is restoration of normal airflow and cessation of therapy. For patients who need ongoing maintenance therapy after the acute phase of illness, the goal should be effective self-administration. An effective program of aerosol drug self-administration depends on thorough patient education.

The patient’s ability to understand the therapy and its goals significantly affects the therapeutic efficacy of any treatment. Whenever possible, patients should be taught to understand the basic administration techniques, to keep track of dosing requirements, to recognize undesirable side effects, and to understand the options and actions required to reduce or eliminate these effects. In addition, patients should be able to demonstrate good technique regarding the use of each aerosol device that they are expected to use in self-care. Practitioner demonstration followed by repeated patient return demonstration is a must and should be done frequently, such as with each office or clinic visit.

Special Considerations

Acute Care and Off-Label Use

Every drug approved for inhalation to date has been designed for and tested in populations of ambulatory patients with moderate disease. As patients with lung disease become acutely and critically ill, the approved label doses, frequency of administration, and devices may not be practical or effective, especially for treatment of patients requiring ventilatory support. In such cases, clinicians may explore and consider nonstandard methods (doses, frequency, and devices) for administration of approved inhaled drugs to patients in the acute care environment, known as off-label use. Another type of off-label use involves drugs that have not been approved for inhalation, ranging from heparin to certain antibiotics. Although physicians may order such drugs via inhalation, the risk to the patient and institution is greater when the administration of such drugs via inhalation has not been thoroughly studied. All forms of off-label use should be avoided when approved and viable alternatives exist. Likewise, off-label administration should always be backed by appropriate departmental or institutional policies and procedures.

Continuous Nebulization for Refractory Bronchospasm

Patients in the emergency department with severe exacerbation of asthma or acute bronchospasm often have been taking standard doses of their bronchodilators for 24 to 36 hours before admission without response. Giving nebulizer treatments with standard bronchodilator doses and repeating the treatments until the symptoms are relieved can require hours of staff time. Administering higher doses of albuterol in short time frames can be accomplished by nebulization of undiluted albuterol (8 to 20 breaths) or by protocol titration with a pMDI and holding chamber (up to 12 puffs). If these strategies fail to provide relief, CBT with albuterol nebulization doses ranging from 5 to 20 mg/hr have proved safe and effective for adult and pediatric patients (Figure 36-32).

Figure 36-33 is a treatment algorithm for high-dose therapy and CBT for pediatric patients with status asthmaticus who are unable to perform peak flow maneuvers.67 Candidates for this protocol are children who, despite frequent beta agonist treatments, remain in extremis with bronchospasm, dyspnea, cough, chest tightness, and diminished breath sounds.

According to this protocol, children older than 6 years with tachypnea, hypoxemia, increased work of breathing, and restlessness who do not respond to standard therapy are given CBT with a large volume nebulizer or SVN at a dose rate of 15 mg/hr (see the accompanying Mini Clini “CBT Dosage Computations” for dosage computations). A standardized asthma score is used to evaluate children younger than 6 years for the severity of the condition (Table 36-4). Patients with an asthma score of 4 or higher are given CBT.

TABLE 36-4

Pediatric Asthma Score

  Score
Indicator 0 1 2
PaO2 >70 mm Hg (air) <70 mm Hg (air) <70 mm Hg (40% O2)
SpO2 >94% (air) <94% (air) <94% (40% O2)
Cyanosis No Yes Yes
Breath sounds Equal Unequal Absent
Wheezing None Moderate Marked
Accessory muscle use None Moderate Marked
Level of consciousness Alert Agitated or depressed Comatose

image

Modified from Volpe J: Therapist-driven protocols for pediatric patients. Respir Care Clin N Am 2:117, 1996.

After CBT is started, the patient is carefully assessed every 30 minutes for the first 2 hours and thereafter every hour. A positive response is indicated by an increase in PEFR of at least 10% after the first hour of therapy. The goal is at least 50% of the predicted value. For small children, improved oxygenation (oxygen saturation by pulse oximeter [SpO2] >92% on room air) with evidence of decreased work of breathing indicates a favorable response. Once the patient “opens up,” intermittent SVN administration is resumed, or a pMDI dose-response assessment is conducted.

The patient has responded poorly to CBT if any of the indicators listed in Table 36-4 worsens. The patient must be observed for adverse drug responses, including worsening tachycardia, palpitations, and vomiting. In these situations, the attending physician must be contacted immediately.

As an alternative to large volume drug nebulizers, some protocols are based on high-dose pMDI therapy (12 to 24 puffs per hour).68 To provide an extra margin of safety, some clinicians recommend that patients receiving CBT undergo continuous electrocardiogram monitoring and measurement of serum potassium level every 4 hours.

Aerosol Administration to Mechanically Ventilated Patients

Since the advent of modern mechanical ventilation, clinicians have administered aerosols to patients with the sickest of lungs. Four primary forms of aerosol generator are used to deliver aerosols during mechanical ventilation: SVN, USN, VM nebulizer, and pMDI with third-party adapter. Table 36-5 summarizes the factors affecting aerosol drug delivery to mechanically ventilated patients. Techniques to optimize delivery to patients receiving ventilatory support are described.69

TABLE 36-5

Factors Affecting Aerosol Drug Delivery During Mechanical Ventilation

Category Factor
Ventilator-related Mode of ventilation
  VT
Respiratory rate
Duty cycle
Inspiratory waveform
Breath-triggering mechanism
Circuit-related Size of endotracheal tube
  Type of humidifier
Relative humidity
Density and viscosity of inhaled gas
Device-related MDI Type of spacer or adapter used
  Position of spacer in circuit
Timing of MDI actuation
SVN Type of nebulizer used
  Fill volume
Gas flow
Cycling: inspiration vs. continuous
Duration of nebulization
Position in circuit
Patient-related Severity of airway obstruction
  Mechanism of airway obstruction
Presence of dynamic hyperinflation
Spontaneous ventilation
Disease process
Drug-related Dose
  Aerosol particle size
Targeted site for delivery
Duration of action

image

Regarding doses, the amount of drug required to achieve the same therapeutic end point is substantially similar for medications delivered by pMDI to intubated patients (8%) and patients who are not intubated (8% to 10%). In stable patients with COPD receiving ventilatory support, 4 puffs of albuterol via pMDI with chamber and 2.5 mg via SVN were shown to produce maximum bronchodilation with effects lasting for 4 hours. However, some differences in response were noted that may have been due to the level of airway obstruction and the techniques used for assessing response.

Techniques for assessing the response to a bronchodilator in intubated patients undergoing mechanical ventilation differ from techniques used in the care of spontaneously breathing patients because expiration is passive during mechanical ventilation, and forced expiratory values (PEFR, FVC, FEV1) cannot normally be obtained. Additional techniques can be used for mechanically ventilated patients because (1) a change in the differences between peak and plateau pressures (the most reliable indicator of a change in airway resistance during continuous mechanical ventilation) can be measured, (2) automatic positive end expiratory pressure levels may decrease in response to bronchodilators (see Chapter 41), and (3) breath-to-breath variations make measurements more reliable when the patient is not actively breathing with the ventilator.70

Techniques for aerosol administration vary by type of aerosol generator and device used. The optimal technique for drug delivery to mechanically ventilated patients with each type of aerosol generator is described in Box 36-9.

Box 36-9   Optimal Technique for Aerosolized Drug Delivery to Mechanically Ventilated Patients

1. Review order, identify the patient, gather equipment, and assess the need for bronchodilators.

2. Clear the airways as needed, by suctioning the patient as needed.

3. If using a circuit with heat and moisture exchanger (HME), remove HME from between the aerosol generator and the patient.

4. If using heated humidifier, do not turn off or disconnect before or during treatment.

5. Assemble equipment (tubing, nebulizer, circuit adapter).

6. Fill the nebulizer with recommended volume and medication per physician order and label.

7. Place adapter in the inspiratory limb, 6 inches from the “wye,” and connect aerosol generator.

8. Turn off or minimize bias flow during treatment.

9. Connect the nebulizer to a gas or power source, as appropriate.

10a. For jet nebulizer (including SVN): Use gas source on ventilator to synchronize nebulization with inspiration, if available; otherwise, set gas flow 2 to 10 L/min as recommended on nebulizer label, and adjust ventilator volume or pressure limit and alarms to compensate for added flow and volume.

10b. For USN and VM nebulizer: Attach power source and cable from controller.

10c. For pMDI: Shake canister and connect to spacer or adapter; actuate at beginning of inspiration.

11. Observe aerosol cloud for adequate aerosol generation during nebulization.

12. After appropriate dose is administered, remove aerosol generator from the ventilator circuit.

13. Reconnect HME, as appropriate.

14. Return ventilator settings and alarms to previous values.

15. Ensure there is no leak in the ventilator circuit.

16. Rinse the nebulizer with sterile or distilled water, shake off excess water, and allow to air dry.

17. Store aerosol device in a clean, dry place.

18. Monitor heart rate, SpO2, blood pressure, and patient-ventilator synchronization.

19. Monitor the patient for adverse response.

20. Assess the airway, and suction as needed; document findings.

Use of a Small Volume Nebulizer During Mechanical Ventilation

The aerosol administered by SVN to intubated patients receiving mechanical ventilation tends to be deposited mainly in the tubing of the ventilator circuit and expiratory filter. Under normal conditions with heated humidification and standard jet nebulizers, pulmonary deposition ranges from 1.5% to 3.0%.70,71 When nebulizer output, humidity level, VT, flow, and I : E ratio are optimized, deposition can increase to 15%.

There are several disadvantages with SVN use during mechanical ventilation. Although in vitro models showed 40% higher aerosol delivery compared with heated humidity, these effects have not been shown in patients, whereas the risks associated with administering cold and dry gas through an endotracheal tube have. A heat and moisture exchanger should be considered a barrier to aerosol administration and should always be removed if placed between the nebulizer and the patient airway. When available with the specific ventilator being used, breath actuation can increase aerosol delivery by 30%, but it may extend administration time by more than threefold. Introducing additional flow into the ventilator circuit may change parameters of flow and delivered volumes and require changes to alarm settings during and after nebulization. The smaller the patient, the greater the impact of added flow into the ventilator circuit, where 6 L/min of additional gas flow can more than double VT and inspiratory pressure, placing the patient at risk. Risk is high for not changing ventilator parameters and not returning parameters to pretreatment levels after administration. There is also a tendency for condensate and secretions to drain into the nebulizer reservoir, contaminating medication being delivered to the lungs.

Use of a Vibrating Mesh Nebulizer During Mechanical Ventilation

Aerosol administration by a VM nebulizer has been estimated to deliver greater than 10% deposition in adults and infants without the addition of gas into the ventilator circuit. The low residual drug volume and small particle size are associated with higher efficiency. Similar to the USN, the VM nebulizer does not add gas flow into the ventilator circuit, so ventilator parameters and alarms do not need to be adjusted before, during, or after nebulization. In contrast to jet SVNs and USNs, the medication reservoir of the VM nebulizer is above the circuit and separated from the ventilator tubing by the mesh, reducing the risk of retrograde contamination of medication in the reservoir from the ventilator circuit. Because of the nature of the mesh, the reservoir can be opened and medication can be added to the nebulizer without creating a perceptible leak during ventilation.

Use of a Pressurized Metered Dose Inhaler During Mechanical Ventilation

Results of in vitro studies show that effective aerosol delivery by pMDIs during mechanical ventilation can range from 2% to 30%. Direct pMDI actuation by simple elbow adapters typically results in the least pulmonary deposition, with most of the aerosol impacting in either the ventilator circuit or the tracheal airway. Higher aerosol delivery percentages occur only when an actuator or spacer is placed in-line in the ventilator circuit. These spacers allow an aerosol “plume” to develop before the bulk of the particles impact on the surface of the circuit or endotracheal tube. The result is a more stable aerosol mass that can penetrate beyond the artificial airway and be deposited mainly in the lung. This situation leads to a better clinical response at lower doses.52

Aerosol Generator Placement

Placement of aerosol generators in the ventilator circuit can have a substantial impact on the available lung dose of drug. During adult ventilation without bias flow, placement of aerosol generators distal to the patient in the inspiratory limb may increase inhaled dose for jet nebulizers, where continuous gas flow acts to charge the inspiratory limb of the ventilator circuit with aerosol increasing the inhaled dose. In contrast, pMDI, USN, and VM nebulizer devices were more efficient when placed proximal to the patient.72 With continuous or bias flow through the ventilator circuit, the delivery is reduced as flow increases, whereas placement of a VM nebulizer near the ventilator increases delivery (Figure 36-34).73

Placement During High-Flow Nasal Oxygen

Researchers used a VM nebulizer to simulate the delivery of aerosol via a high-flow nasal O2 setup using infant, pediatric, and adult cannulas, with inhaled dose ranging from 8% to 28%.76 Figure 36-35 shows such a setup, including the location of the VM nebulizer. In addition to the type and location of the nebulizer used with high-flow nasal O2, the inhaled dose seems to vary based on cannula size, respiratory pattern, and O2 flow. Heliox (80 : 20) appears to improve aerosol delivery at higher flow rates with these setups.77

Placement During Intrapulmonary Percussive Ventilation

Intrapulmonary percussive ventilation provides high-frequency oscillation of the airway while administering aerosol particles. During intrapulmonary percussive ventilation, the aerosol generator should be placed in the circuit as close to the patient’s airway as practical. Aerosol administration during intrapulmonary percussive ventilation has been compared with a standard jet nebulizer. The MMAD was smaller with intrapulmonary percussive ventilation than with the jet (0.2 µm vs. 1.89 µm), and the fine-particle fraction was lower (16.2% vs. 67.5%). However, lung dose was similar (2.49% with intrapulmonary percussive ventilation vs. 4.2% with the jet nebulizer). It was concluded that intrapulmonary percussive ventilation was too variable and too unpredictable to recommend for drug delivery to the lung.78

Controlling Environmental Contamination

Drugs for nebulization that escape from the nebulizer into the atmosphere or are exhaled by the patient can be inhaled by anyone in the vicinity of the treatment. The risk imposed by this environmental exposure is clear and is associated with a range of drugs and patients with infectious disease. Pentamidine and ribavirin were associated with health risks to health care providers even when used in conjunction with filters on exhalation ports of nebulizers, containment and scavenger systems, and high-efficiency particulate air (HEPA) filter hoods and ventilation systems (Figure 36-36).

Continuous pneumatic nebulizers produce the greatest amount of secondhand aerosol, with most (60%) of the aerosol produced passing directly into the environment. The Respirgard II (Vital Signs, Totowa, NJ) nebulizer was developed for administration of pentamidine, adding one-way valves and an expiratory filter to contain aerosol that is exhaled and not inhaled. Breath-actuated nebulizers, DPIs, and pMDIs tend to generate less secondhand aerosol.

A survey found that RTs were more than twice as likely as physical therapists to develop asthma-like symptoms during the course of their careers. The authors associated this with administration of ribavirin and exposure to gluteraldehyde.17 There have been anecdotal reports of respiratory care clinicians who have developed a sensitivity to secondhand aerosol from bronchodilators. Further research is required to understand more thoroughly the hazards of secondhand exposure to aerosols in the clinical setting. Most of nebulizer therapy currently delivered does not include filtering systems.

Patients with infectious and resistant organisms, such as tuberculosis, severe acute respiratory syndrome, and H1N1 virus, require respiratory isolation, and caregivers require protection. RTs have a duty to take appropriate steps to protect themselves and their patients. In these cases, the aerosols generated by the patient from coughing, speaking, or laughing can transmit disease. These aerosols can travel substantial distances between rooms and even floors in institutions. Although exposure to secondhand aerosol generated by a nebulizer is undesirable, it poses less risk than the infected aerosols produced by mucosa of patients.82 In essence, it is unlikely that any medical aerosol that is inhaled by the patient would be contaminated. Nonetheless, during the severe acute respiratory syndrome outbreak, some centers outlawed use of medical aerosols to reduce exposure. Efforts should be taken to reduce transmission of both nebulizer aerosols and patient-generated aerosols to the environment.

Various techniques are available for protecting patients and caregivers from environmental exposure during aerosol drug therapy. The greatest occupational risk for RTs has been associated with the administration of ribavirin and pentamidine. Conjunctivitis, headaches, bronchospasm, shortness of breath, and rashes have been reported among individuals administering these drugs.83 Patients given aerosolized ribavirin or pentamidine must be treated in a private room, booth, or tent or at a special station designed to minimize environmental contamination.

Booths and Stations

Booths or stations should be used for sputum induction and aerosolized medication treatments given in any area where more than one patient is treated. The area should be designed to provide adequate airflow to draw aerosol and droplet nuclei from the patient into an appropriate filtration system or an exhaust system directly to the outside. Booths and stations should be adequately cleaned between patients.

A variety of booths and specially designed stations are available for delivery of pentamidine or ribavirin. The Emerson containment booth (Figure 36-37) is an example of a system that completely isolates the patient during aerosol administration. The AeroStar Aerosol Protection Cart (Respiratory Safety Systems, San Diego, CA) is a portable patient isolation station for administration of hazardous aerosolized medication. It has been used during sputum induction and for pentamidine treatment. The patient compartment is collapsible with a swing-out counter and three polycarbonate walls. Captured aerosols are removed with a HEPA filter. A prefilter is used to retain larger dust particles and to prevent early loading of the more expensive HEPA filter.

Filters and nebulizers used in treatments with pentamidine and ribavirin should be treated as hazardous wastes and disposed of accordingly. Goggles, gloves, and gowns should be used as splatter shields and to reduce exposure to medication residues and body substances. Staff members should be screened for adverse effects of exposure to the aerosol medication. The risks and safety procedures should be reviewed regularly.

In addition to the risks associated with administration of aerosol medication, risk of tuberculosis transmission has become a great concern because of an increase in case numbers and the development of multidrug-resistant strains of the organism. Tuberculosis is transmitted in the form of droplet nuclei (0.3 to 0.6 µm) that carry tuberculosis bacilli. Patients with known or suspected tuberculosis need private rooms with negative pressure ventilation that exhausts to the outside. If environmental isolation is impossible or the health care worker must enter the patient’s room, personal protective equipment should be used.

Personal Protective Equipment

Personal protective equipment is recommended when caring for any patient with a disease that can be spread by the airborne route.84 The greatest risk is communication of tuberculosis or chickenpox. Although environmental controls should be instituted in the care of these patients, standard and airborne precautions should also be implemented. Various masks and respirators have been recommended for use when caring for a patient with tuberculosis or other respiration-transmitted diseases. Traditional surgical masks, particulate respirators, disposable and reusable HEPA filters, and powered air-purifying respirators have been used. No data are available for determining the most effective and most clinically useful device to protect health care workers and others, although the U.S. Occupational Safety and Health Administration requires specific levels of protection (HEPA filters and powered air-purifying respirators). Guidelines from the World Health Organization recommend surgical masks for all patient care with the exception of N95 masks for aerosol-generating procedures such as sputum induction. Evidence from laboratory studies of potential airborne spread of influenza from contagious patients indicates that guidelines related to the current 1-m respiratory zone may need to be extended to a larger respiratory zone and include eye protection.85

Summary Checklist

• An aerosol is a suspension of solid or liquid particles in gas. In the clinical setting, therapeutic aerosols are made with atomizers or nebulizers.

• The general aim of aerosol drug therapy is delivery of a therapeutic dose of the selected agent to the desired site of action.

• Where aerosol particles are deposited in the respiratory tract depends on their size, shape, and motion and on the physical characteristics of the airways. Key mechanisms causing aerosol deposition include inertial impaction, sedimentation, and brownian diffusion.

• For targeting aerosols for delivery to the upper airway (nose, larynx, trachea), particles in the 5- to 20-µm MMAD range are used; for the lower airways, 2- to 5-µm particles are used; and for the lung parenchyma (alveolar region), 1- to 3-µm particles are used.

• The primary hazard of aerosol drug therapy is an adverse reaction to the medication being administered. Other hazards include infection, airway reactivity, systemic effects of bland aerosols, and drug reconcentration.

• Drug aerosol delivery systems include pMDIs, DPIs, SVNs, large volume jet nebulizers, hand-bulb atomizers (nasal spray pumps), USNs, and VM nebulizers.

• MDIs are the preferred method for maintenance delivery of bronchodilators and steroids to spontaneously breathing patients. The effectiveness of this therapy is highly technique-dependent.

• Accessory devices, spacers, and holding chambers are used with pMDIs to reduce oropharyngeal deposition of a drug and to overcome problems with poor hand-breath coordination.

• Effective use of DPIs does not require hand-breath coordination, but it does require high inspiratory flows. Some patients in stable condition prefer DPI delivery systems.

• Compared with pMDI and DPI delivery systems, use of an SVN is less technique-dependent and is more commonly used in acute care.

• Large volume drug nebulizers can be used to provide continuous aerosol delivery when traditional dosing strategies are ineffective in controlling severe bronchospasm.

• Small volume USNs can be used to administer bronchodilators, antiinflammatory agents, and antibiotics.

• Because patients vary greatly in their response to a particular drug dose and route of administration, aerosol drug therapy should be tailored to each patient with an assessment-based protocol.

• Careful, ongoing patient assessment is the key to an effective bronchodilator therapy protocol. Components of the assessment include a patient interview, observation, expiratory airflow tests, vital sign measurements, auscultation, blood gas analysis, and oximetry.

• Protocols for CBT have proved safe and effective in the management of refractory bronchospasm in both adults and children.

• Many factors affect the efficiency of aerosol drug delivery during mechanical ventilation. Proper selection of aerosol generator type, position in the circuit, dose, and accessory equipment is needed to optimize deposition and achieve the desired clinical outcome.

• Various techniques are available to protect patients and caregivers from environmental exposure during aerosol drug therapy.