Pulmonary Function Testing

Published on 01/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 4727 times

Pulmonary Function Testing

F. Herbert Douce

The most important function of the lungs is gas exchange. As mixed venous blood passes through the pulmonary circulation, the lungs add oxygen (O2) and remove excess carbon dioxide (CO2). The ability of the lungs to perform gas exchange depends on the following four general physiologic functions:

Pulmonary function tests can provide valuable information about these important individual processes that support gas exchange. Various measurements are available to aid in the diagnosis and assessment of pulmonary diseases, to determine the need for therapy, and to evaluate the effectiveness of respiratory care. For respiratory therapists (RTs), knowledge of these tests and the ability to interpret the measurements are essential for assessing patients objectively and for planning and implementing effective patient care. The key terms used in this chapter are terms adopted and defined by the pulmonary medical community and should become the standard vocabulary of all RTs.1

Pulmonary Function Testing

A complete evaluation of the respiratory system includes a patient history, physical examination, chest x-ray examination, arterial blood gas analysis, and tests of pulmonary function. Test results become most meaningful when considered in the context of a complete evaluation. Although diagnostic pulmonary function testing is performed in a laboratory setting and usually only on patients in a stable condition, RTs also perform many of these tests at the bedside on patients who are acutely ill. There are three categories of pulmonary function tests, measuring (1) dynamic flow rates of gases through the airways, (2) lung volumes and capacities, and (3) the ability of the lungs to diffuse gases. A combination of these measurements provides a quantitative picture of lung function. Although pulmonary function tests do not diagnose specific pulmonary diseases, these tests identify the presence and type of pulmonary impairments and the degree of pulmonary disease present. Some basic tests of pulmonary function are often performed at the bedside to provide immediate information about the need for respiratory therapy and its effectiveness.

Purposes

Generally, the primary purposes of pulmonary function testing are to identify pulmonary impairment and to quantify the severity of pulmonary impairment if present.2 Pulmonary function testing has diagnostic and therapeutic roles and helps clinicians answer some general questions about patients with lung disease (Box 19-1).

The indications for pulmonary function testing are as follows:

• To identify and quantify changes in pulmonary function. The most common purposes of pulmonary function testing are to detect the presence or absence of pulmonary disease, to classify the type of disease as either obstructive or restrictive, and to quantify the severity of pulmonary impairment as mild, moderate, severe, or very severe. Over time, pulmonary function tests help quantify the progression or the reversibility of the disease.3

• To evaluate need and quantify therapeutic effectiveness.4 Pulmonary function tests may aid clinicians in selecting or modifying a specific therapeutic regimen or technique (e.g., bronchodilator medication, airway clearance therapy, rehabilitation exercise protocol). Clinicians and researchers use pulmonary function tests to measure changes in lung function objectively before and after treatment.

• To perform epidemiologic surveillance for pulmonary disease. Screening programs may detect pulmonary abnormalities caused by disease or environmental factors in general populations, in people in occupational settings, in smokers, or in other high-risk groups. In addition, researchers have determined what normal pulmonary function is by measuring the pulmonary function of healthy people.5

• To assess patients for risk of postoperative pulmonary complications. Preoperative testing can identify patients who may have an increased risk of pulmonary complications after surgery.6 Sometimes the risk of complications can be reduced by preoperative respiratory care, and sometimes the risk may be significant enough to rule out surgery.

• To determine pulmonary disability.7 Pulmonary function tests can also determine the degree of disability caused by lung diseases, including occupational diseases such as pneumoconiosis of coal workers. Some federal entitlement programs and insurance policies rely on pulmonary function tests to confirm claims for financial compensation.

There are also contraindications to pulmonary function testing.4 Patients with acute, unstable cardiopulmonary problems, such as hemoptysis, pneumothorax, myocardial infarction, and pulmonary embolism, and patients with acute chest or abdominal pain should not be tested. Testing could be harmful if needed treatment would be delayed. Patients who have nausea and who are vomiting should not be tested because there is a risk of aspiration. Testing for patients who have had recent cataract removal surgery should be delayed because changes in ocular pressure may be harmful to the eye. Pulmonary function testing requires patient effort and cooperation. Patients with dementia or confusion may not achieve optimal or repeatable results. Pulmonary function testing should not be performed if valid and reliable results cannot be predicted. In patients who are acutely ill or who have recently smoked a cigarette, the test validity of measuring the forced vital capacity (FVC) may be hindered.

Pathophysiologic Patterns

Pulmonary function testing provides the basis for classifying pulmonary diseases into two major categories, obstructive pulmonary disease and restrictive pulmonary disease. These two types of lung diseases sometimes occur together as a mixed impairment. Obstructive and restrictive types of lung diseases differ in several important ways. Figure 19-1 shows normal lungs with the pathophysiologic aspects of obstructive lung diseases and restrictive lung diseases, and the differences are summarized in Table 19-1. The primary problem in obstructive pulmonary disease is an increased airway resistance (Raw). Raw is the difference in pressure between the ends of the airways divided by the flow rate of gas moving through the airway, according to the following formula: Raw = ΔP/image.

TABLE 19-1

Comparison of Obstructive and Restrictive Types of Pulmonary Diseases

Characteristic Obstructive Disease Restrictive Disease
Anatomy affected Airways Lung parenchyma, thoracic pump
Breathing phase difficulty Expiration Inspiration
Pathophysiology Increased airway resistance Decreased lung or thoracic compliance
Useful measurements Flow rates Volumes or capacities

There is an inverse relationship between Raw and flow rates (image). If the pressure difference is constant, a reduced flow rate indicates an increase in Raw. Because the radius of the airways normally lessens slightly during expiration, flow rates are usually measured during expiration. By rearranging the symbols in Poiseuille’s law (see Chapter 6), Raw is inversely related to the radius of the airways according to the following formula: Raw = ΔP/image = η8l/r4.

When airway radius (r) decreases, Raw increases, while the flow rate of gas through the airways (image) decreases. Airway radius can be reduced by excessive contraction of the bronchial and bronchiolar muscles (bronchospasm), excessive secretions in the airways, swelling of the airway mucosa, airway tumors, collapse of the bronchioles, and other causes. By measuring flow rates, pulmonary function tests measure indirectly the size of the airways, Raw, and the presence of obstructive disease.

The primary problem in restrictive lung disease is reduced lung compliance, thoracic compliance, or both. Compliance is the volume of gas inspired per the amount of inspiratory effort; effort is measured as the amount of pressure created in the lung or in the pleural space when the inspiratory muscles contract. Compliance is calculated according to the following formula: C = ΔV/ΔP.

There is a direct relationship between compliance (C) and volume (V). If the pressure difference is constant, a reduced inspiratory volume indicates a reduction in compliance. Reduced lung compliance is usually the result of alveolar inflammation, pulmonary fibrosis, or neoplasms in the alveoli; a reduced thoracic compliance may be the result of thoracic wall abnormalities such as kyphoscoliosis. Neuromuscular diseases also can result in reduced lung volumes and restrictive-type pulmonary impairments, mainly by affecting the function of the inspiratory muscles. In these circumstances, lung compliance and thoracic compliance may be normal, but the patient is unable to generate enough subatmospheric pressure to take a full, deep breath.

Some obstructive diseases and some restrictive diseases also may affect the ability of the lung to diffuse gases. In some diseases, there is damage to the alveolar-capillary membrane, or less alveolar surface area is accessible for diffusion. Measuring the diffusing capacity of the lung for carbon monoxide (DLCO) can identify the destruction of alveolar tissue or the loss of functioning alveolar surface area.

For each measurement of pulmonary function, there is a normal value and a lower limit of normal (LLN). Measurements less than the LLN indicate the presence of an abnormality. The severity of pulmonary impairment is based on a comparison of each patient’s measurement with the predicted normal value for the patient. Several methods are used for comparison with the normal value. A common method of comparison is to compute a percentage of the predicted normal value according to the following equation:

< ?xml:namespace prefix = "mml" />% Predicted=Measured valuePredicted normal value×100

image

Determining if the patient’s value is within 1 or 2 standard deviations of the predicted normal value is an alternative method used in some laboratories. The predicted percentage or the number of standard deviations from the predicted normal value can be used to quantify severity of impairment. Typical degrees of severity are listed in Table 19-2.

TABLE 19-2

Severity of Pulmonary Impairments Based on a Percentage of Predicted Normal Values

Degree of Impairment Obstruction based on FEV1 Restriction or Obstruction Based on TLC, FRC, RV Gas Exchange Based on DLCO
Normal 80%-120% 80%-120% 80%-120%
Mild 70%-79% 70%-79% or 121%-130% 61%-79%
Moderate 60%-69% 60%-69% or 131%-140% 40%-60%
Moderately severe 50%-59% 50%-59% or 141%-150%  
Severe 35%-49% 35%-49% or 151%-165% <40%
Very severe <35% <35% or >165%  

image

Infection Control

Pulmonary function testing is considered safe, but there is potential to transmit infective microorganisms to patients and technologists.8 Transmission can occur by direct or indirect contact. Standard precautions should be applied because of the potential exposure to saliva or mucus, which could possibly contain blood or other potentially hazardous microorganisms. Patients with oral lesions pose the greatest potential hazard, and patients with compromised immune systems are at the greatest risk. Practitioners should wear gloves when handling potentially contaminated mouthpieces, valves, tubing, and equipment surfaces. When performing procedures on patients with potentially infectious airborne diseases, practitioners should wear a personal respirator or a close-fitting surgical mask, especially if the testing induces coughing. Practitioners should always wash their hands between testing patients and after contact with testing equipment. Although it is unnecessary to clean the interior surfaces of the testing instruments routinely between patients,9 the mouthpiece, nose clips, tubing, and any parts of the instrument that come into direct contact with a patient should be disposed, sterilized, or disinfected between patients. Any equipment surface showing visible condensation from exhaled air should be discarded, disinfected, or sterilized before reuse. When testing instruments are disassembled for cleaning and disinfecting, manufacturer recommendations should be considered, and recalibration may be necessary before testing resumes. The routine use of low-resistance, in-line barrier filters is controversial.1012 Filters may be appropriate when internal surfaces of manifolds and valves proximal to mouthpieces are inaccessible or difficult to disassemble for cleaning and disinfecting. Filters provide visible evidence to reassure patients that their protection has been considered.

Equipment

Pulmonary function testing requires measurement of gas volume or flow, and various instruments and measurement principles are used to make these measurements. There are two general types of measuring instruments: instruments that measure gas volume and instruments that measure gas flow. Both types of instruments simultaneously measure time, and both compute various volumes and flow rates used in pulmonary function testing. The term spirometer is sometimes used as a generic term for all volume-measuring and flow-measuring devices.

Volume-measuring devices are specifically called spirometers and include water-sealed, bellows, and dry rolling seal types. These devices expand as they collect gas volumes. The magnitude of the expansion is the volume measured, and the speed of expansion represents the flow rate. In the absence of leaks and with low momentum forces, volume-measuring devices can be extremely accurate for measuring volumes, and with low inertia and friction forces, volume-measuring devices can be extremely accurate when computing flow rates.

Flow-measuring devices are commonly called pneumotachometers, although some practitioners reserve this term for only the device originally designed by Fleisch. These devices measure flow using a variety of unique principles. The Fleisch-type pneumotachometer measures the change in pressure as gas flows through a minimal, constant resistance according to the formula: image = ΔP ÷ R. Different manufacturers have used several materials to provide the resistance, including screens, capillary tubes, and fiber sheets made of silk, nylon, or filter paper. With multiple uses, condensation from exhaled air can collect in these devices and alter the resistance and accuracy of the device; some devices are heated to body temperature to prevent condensation. Known as thermistors or mass flowmeters, another type of flow-measuring device measures the temperature change created by gas flowing through it. There are also tubinometers, which use rotation of a fan or blades similar to a windmill. The number of rotations indicates volume, and the speed of the rotations indicates flow. How gas flow affects the transmission of sound waves and the force of flow stretching a spring have also been used to measure flow. Detailed descriptions and examples of each type of device are beyond the scope of this chapter and are available elsewhere.13

Regardless of the type of device or the principle of measurement used, several important characteristics are common to all volume-measuring and flow-measuring devices. Having an understanding of these common characteristics provides RTs the ability to select and use these devices properly. Every measuring instrument has capacity, accuracy, error, resolution, precision, linearity, and output.14,15 The ideal instrument would have unlimited capacity to measure every pulmonary parameter, and it would have perfect accuracy and precision over its entire measurement range; there are no ideal instruments.

The capacity of an instrument refers to the range or limits of how much it can measure. Most instruments are designed with capacities to measure volumes and flow rates of all adults. The accuracy of a measuring instrument is how well it measures a known reference value. For volume measurements, standard reference values are provided by a graduated 3.0-L calibration syringe.16 No measuring instrument is perfect, and there usually is an arithmetic difference between reference values and measured values. This difference is called the error. Accuracy and error are opposing terms; the greater the accuracy, the smaller is the error. Accuracy and error are commonly expressed as percentages, with their sum always equaling 100%. To determine percent accuracy and percent error, several reference values are measured, and the mean of the measured values is computed and compared with the reference values according to the following equations:

% Accuracy=Mean measured valueReference value×100

image

or

% Error=Mean measured valueReference valueReference value×100

image

Resolution is the smallest detectable measurement; instruments with high resolution can measure the smallest volumes, flows, and times. Precision is synonymous with reliability of measurements and the opposite of variability. When multiple known reference values are measured, the standard deviation of the mean measured reference value is the statistic that indicates the precision of an instrument. A small standard deviation indicates low variability and high precision. Linearity refers to the accuracy of the instrument over its entire range of measurement, or its capacity. Some devices may accurately measure large volumes or high flow rates but may be less accurate when measuring small volumes or low flow rates. To determine linearity, accuracy and precision are calculated at different points over the range (capacity) of the device.

Output includes the specific measurements made or computed by the instrument. Most volume-measuring and flow-measuring devices measure the FVC and forced expiratory volume in 1 second (FEV1). Others calculate various forced expiratory flow (FEF) rates, and some measure tidal volume (VT) and minute ventilation (image). Diagnostic spirometers usually measure and calculate vital capacity (VC), FVC, FEV1, peak expiratory flow (PEF) rate, and FEF rates. Some measure and calculate maximal voluntary ventilation (MVV). Some of these instruments may be a component of a laboratory system providing the volume-measuring or flow-measuring capability for other diagnostic tests of pulmonary function. For example, they may be used with gas analyzers to measure functional residual capacity (FRC) and total lung capacity (TLC) or the inspiratory VC during single-breath diffusing capacity (DLCOSB). Whether a spirometer or pneumotachometer is used in a diagnostic laboratory, a physician’s office, or at the bedside in an intensive care unit, it should meet or exceed the national performance standards for volume-measuring and flow-measuring devices.

In 1978, the American Thoracic Society (ATS) adopted the initial standards for diagnostic spirometers. These standards have been adopted by other medical organizations and government agencies. Updated most recently in 2005 in collaboration with the European Respiratory Society (ERS), the standards are now recognized internationally as the standards for the industry.17 Some instruments have been independently evaluated against the standards or compared with instruments that meet those standards. Regardless of the measuring principle used by the instrument or the purpose of the patient testing, RTs should use only devices that meet or exceed current ATS/ERS performance standards. According to the ATS/ERS standards, when measuring a slow VC, the spirometer should be able to measure for up to 30 seconds, and for the FVC, the time capacity should be at least 15 seconds. When measuring the VC, FVC, and forced expiratory volumes, a volume-measuring spirometer should have a capacity of at least 8 L and should measure volumes with less than a 3% error or within 50 ml of a reference value, whichever is greater. These standards, including the 8-L standard for capacity, also apply to children. A diagnostic spirometer that measures flow should be at least 95% accurate (or within 0.2 L/sec, whichever is greater) over the entire 0 to 14 L/sec range of gas flow. The standards are summarized in Table 19-3.

TABLE 19-3

2005 Spirometer Performance Standards of the American Thoracic Society/European Respiratory Society (ATS/ERS) Task Force

Test Volume (L) Flow (L/sec) Accuracy Time (sec) Back Pressure (cm H2O/L/sec)
VC 0.5-8 L 0-14 ≤3% or 0.05 L* 30  
FVC 0.5-8 L 0-14 ≤3% or 0.05 L* 15 <1.5%scm H2O/L/sec at 14 L/sec
FEV1 0.5-8 L 0-14 ≤3% or 0.05 L* 1 <1.5%scm H2O/L/sec at 14 L/sec
PEF   0-14 ≤10% or 0.3 L/sec*   <1.5%scm H2O/L/sec at 14 L/sec
FEF   ±14 ≤5% or 0.2 L/sec*   <1.5%scm H2O/L/sec at 14 L/sec
MVV 250 L/min at 2 L/breath   ±10% or 15 L/min* 12-15 <1.5%scm H2O/L/sec at 14 L/sec

image

*Whichever is greater.

From Miller MR, Hankinson J, Brusasco V, et al: Standardisation of spirometry. Eur Respir J 26:319–338, 2005.

The spirometer standards also require spirometers to have a thermometer or to produce values corrected for body temperature, ambient pressure, and fully saturated with water vapor (BTPS). Standards also require that graphic outputs be of sufficient size and scale of display and recording to allow for visual inspection during testing, validation, and hand measurements. For visual display on a computer monitor, the resolution required is 0.050 L, and the scale of the volume axis must be 5 mm/L; the scale for the time axis must be at least 10 mm/sec. For validation and hand measurement functions from graph paper, the resolution must be 0.025 L, and the scale of the volume axis must be 10 mm/L; the scale for the time axis must be at least 20 mm/sec. Most manufacturers have designed their spirometers to meet or exceed the validation and hand measurement standards.

For quality control, the standards include verifying volume accuracy with a 3.0-L calibration syringe at least daily, although best practice in many laboratories is to verify accuracy before each test subject. The 2005 standards recognize that 3.0-L calibration syringes may have up to 0.5% error, and error may be acceptable if in the ±3.5% range. Volume linearity should be verified quarterly using 1.0-L increments over the entire volume range; flow linearity should be checked weekly using at least three different flow ranges. Recorder speed should be checked with a stopwatch quarterly. When new versions of software are installed, testing known subjects and comparing results is recommended. For comprehensive quality assurance of pulmonary function testing, there are three equally important aspects to consider: verifying the accuracy and precision of the measuring instruments, the performance of the technologist, and the test results when measuring a standard.

Most modern pulmonary function laboratories use computers for data acquisition and reduction. Computer-assisted testing decreases the time necessary to complete the tests and enhances the effectiveness of pulmonary function testing by increasing accuracy, increasing patient acceptance, and monitoring patient performance. Although computer-assisted testing and interpretations of test results are often applied by a computer, pulmonary function testing always requires a trained and competent RT to administer the tests, and computer analysis should not replace human analysis.

Principles of Measurement and Significance

For tests of pulmonary function, four important general principles should be considered: test specificity, sensitivity, validity, and reliability. Most tests of pulmonary function are not specific because several different diseases may cause the test result to be abnormal. This limitation of many pulmonary function tests explains why these tests identify a pattern of impairment rather than diagnose specific diseases. Some tests are extremely sensitive, and apparently healthy individuals may have an abnormal test result. However, some tests are not sensitive; individuals must be extremely sick to have an abnormal test result. To be meaningful, each test must be valid, or the test is not measuring what it is intended to measure. When performing pulmonary function testing, strictly following testing procedures, ensuring patient effort and performance, and ensuring equipment accuracy and calibration establish test validity. Test reliability is the consistency of the test results. A reliable test produces consistent test results with minimal variability. To be reliable, each test must be performed more than once. Ensuring test validity and reliability is the most important role of the RT. Test results that are invalid or unreliable can lead to misdiagnosis, mistreatment, and poor outcomes.

In most pulmonary function laboratories, there are three components to pulmonary function testing: (1) performing spirometry for measuring airway mechanics, (2) measuring lung volumes and capacities, and (3) measuring the diffusing capacity of the lung (DL). For each component, there are various techniques and different types of equipment that make the measurements. When the purpose of the testing is to identify the presence and the degree of pulmonary impairment and the type of pulmonary disease, all three testing components are required. When the purpose of the testing is more limited, such as to assess postoperative pulmonary risk or to evaluate and quantify therapeutic effectiveness, the scope of measurement also is limited. Many pulmonary function laboratories also perform arterial blood gas analysis (see Chapter 18), and some laboratories provide more specialized and advanced tests, such as bronchial challenge tests and exercise stress tests.

Spirometry

Spirometry includes the tests of pulmonary mechanics—the measurements of FVC, FEV1, several FEF values, forced inspiratory flow rates, and MVV. Measuring pulmonary mechanics is assessing the ability of the lungs to move large volumes of air quickly through the airways to identify airway obstruction. Some measurements are aimed at large intrathoracic airways, some are aimed at small airways, and some assess obstruction throughout the lungs. Measuring flow rates is a surrogate for measuring airways resistance according to the formula: Raw = ΔP ÷ image. A decrease in flow rate signifies an increase in airways resistance and the presence of airway obstruction when patient effort creating the difference between mouth pressure and lung pressure is constant (see Clinical Practice Guideline 19-1).17

19-1   Spirometry

AARC Clinical Practice Guideline (Excerpts)*

Contraindications

Circumstances listed here could affect the reliability of spirometry measurements. In addition, forced expiratory maneuvers may aggravate these conditions, which may make test postponement necessary until the medical condition resolves. The following are some relative contraindications to performing spirometry:

Quality Control

• Volume verification (i.e., calibration): At least daily before testing, use a calibrated known-volume syringe with a volume of at least 3 L to ascertain that the spirometer reads a known volume accurately. The known volume should be injected or withdrawn at least three times, at flows that vary between 2 L/sec and 12 L/sec (3-L injection times of approximately 1 second, 6 seconds, and between 1 seconds and 6 seconds). The tolerance limits for an acceptable calibration are ±3% of the known volume. For a 3-L calibration syringe, the acceptable recovered range is 2.91 to 3.09 L. The practitioner is encouraged to exceed this guideline whenever possible (i.e., reduce the tolerance limits to less than ±3%).

• Leak test: Volume-displacement spirometers must be evaluated for leaks daily. One recommendation is that any volume change of more than 10 ml/min while the spirometer is under at least 3 cm H2O pressure be considered excessive.

• A spirometry procedure manual should be maintained.

• A log that documents daily instrument calibration, problems encountered, corrective action required, and system hardware or software changes should be maintained.

• Computer software for measurement and computer calculations should be checked against manual calculations if possible. In addition, biologic laboratory standards (i.e., healthy, nonsmoking individuals) can be tested periodically to ensure historic reproducibility, to verify software upgrades, and to evaluate new or replacement spirometers.

• The known-volume syringe should be checked for accuracy at least quarterly using a second known-volume syringe, with the spirometer in the patient-test mode; this validates the calibration and ensures that the patient-test mode operates properly.

• For water-seal spirometers, water level and paper tracing speed should be checked daily. The entire range of volume displacement should be checked quarterly.


*For the complete guideline, see American Association for Respiratory Care: Clinical practice guideline: spirometry. 1996 update, Respir Care 41:629, 1996.

Although performing tests of pulmonary mechanics is considered safe, some adverse reactions have occurred, including pneumothorax,18 syncope, chest pain, paroxysmal coughing, and bronchospasm associated with exercise-induced asthma.19 The contraindications for pulmonary function testing are primarily for testing mechanics. The ATS/ERS 2005 standards for spirometry17 specify the validity and reliability criteria of the measurements and accuracy and precision limits of the measuring equipment. These standards have been incorporated into the clinical practice guidelines of the AARC, medical societies, and government agencies.4,2022

Forced Vital Capacity

FVC is the most commonly performed test of pulmonary mechanics, and many measurements are made while the patient is performing the FVC maneuver (Figure 19-2). Measuring FVC often occurs under baseline or untreated conditions. For baseline testing, patients should temporarily abstain from bronchodilator medications. Short-acting bronchodilators (e.g., β-agonist albuterol, anticholinergic agent ipratropium bromide) should not be used for 4 hours before baseline spirometry, whereas long-acting β-agonist bronchodilators and oral therapy with aminophylline should be stopped for 12 hours. When a patient’s baseline results show airway obstruction, performing FVC after treatment (e.g., albuterol bronchodilator aerosol or metered dose inhaler) can help determine if the treatment is effective. The FVC maneuver is also performed repeatedly during bronchial provocation testing.

FVC may be measured on a spirometer that measures volumes or flows, that presents a graph of volume and time or flow and volume, that is mechanical or electronic, and that has a calculator or computer. The forced expiratory VC sometimes is followed by a forced inspiratory VC to produce a complete image of forced breathing called a flow-volume loop.23

FVC is an effort-dependent maneuver that requires careful patient instruction, understanding, coordination, and cooperation. Spirometry standards for FVC specify that patients must be instructed in the FVC maneuver, that the appropriate technique be demonstrated, and that enthusiastic coaching occur. When measuring FVC, the RT needs to coach the preceding inspiratory capacity (IC) as enthusiastically as the FVC. According to the standards, nose clips are encouraged, but not required, and patients may be tested in the sitting or standing position. Although standing usually produces a larger FVC compared with sitting, sitting is considered safer in case of lightheadedness. It is recommended that the position be consistent for repeat testing of the same patient. FVC should be converted to body temperature conditions and reported as liters under BTPS conditions.

To ensure validity, each patient must perform a minimum of three acceptable FVC maneuvers. To ensure reliability, the largest FVC and second largest FVC from the acceptable trials should not vary by more than 0.150 L. To perform an FVC trial, the patient should inhale rapidly and completely to TLC from the resting FRC level. The forced exhalation of an acceptable FVC trial should begin abruptly and without hesitation. A satisfactory start of expiration is defined as an extrapolated volume at the zero time point less than 5% of FVC or 0.150 L, whichever is greater (Figure 19-3). The volume exhaled before the zero time point is called the extrapolated volume. To be valid, no more than 5% of the VC or 0.150 L is allowed to be exhaled before the zero time point. An acceptable FVC trial also is smooth, continuous, and complete. A cough, an inspiration, a Valsalva maneuver, a leak, or an obstructed mouthpiece while an FVC maneuver is being performed disqualifies the trial. FVC must be completely exhaled or an exhalation time of at least 6 seconds must occur for adults and children older than 10 years (longer times are commonly needed for patients with airway obstruction). A 3-second exhalation is acceptable for children younger than 10 years old. An end expiratory plateau must be obvious in the volume-time curve; the objective standard is less than 0.025 L exhaled during the final second of exhalation. Consistent with its definition, the largest acceptable FVC (BTPS) measured from the set of three acceptable trials is the patient’s FVC.

Forced Expiratory Volume in 1 Second

During FVC testing, several other measurements are also made. FEV1 is a measurement of the volume exhaled in the first second of FVC (see Figure 19-2, A). To ensure validity of FEV1, the measurement must originate from a set of three acceptable FVC trials. The first second of forced exhalation begins at the zero time point (see Figure 19-3). To ensure reliability of FEV1, the largest FEV1 and second largest FEV1 from the acceptable trials should not vary by more than 0.150 L. Consistent with its definition, the largest FEV1 (BTPS) measured is the patient’s FEV1. The largest FEV1 sometimes comes from a different trial than the largest FVC.

The %FEV1/FVC, also called the forced expiratory volume in 1 second-to-vital capacity ratio (FEV1/FVC), is calculated by dividing the patient’s largest FEV1 by the patient’s largest VC and converting it to a percentage (by multiplying by 100). The two values do not have to come from the same trial; the VC should be the largest VC measured, even if measured as a slow VC or during inspiration.

Except for PEF rate, all other measurements that originate from FVC come from the “best curve”—these include forced expiratory flow between 200 ml and 1200 ml of FVC (FEF200-1200); forced expiratory flow between 25% and 75% of FVC (FEF25%-75%); forced expiratory flow between 75% and 85% of FVC (FEF75%-85%); and instantaneous FEF25%, FEF50%, and FEF75%. The best test curve is defined as the trial that meets the acceptability criteria and gives the largest sum of FVC plus FEV1. The validity and reliability of these other measurements of pulmonary mechanics are based on their origin from a valid and reliable FVC.

Forced Expiratory Flow Between 200 ml and 1200 ml of Forced Vital Capacity and Forced Expiratory Flow Between 25% and 75% of Forced Vital Capacity

FEF200-1200 and FEF25%-75% represent average flow rates that occur during specific intervals of FVC. Both measurements can be made on a volume-time spirogram as the slope of a line connecting the two points in their subscripts. For FEF200-1200, the 200-ml point and the 1200-ml point are identified. A straight line is drawn connecting these points, and the line is extended to intersect two vertical time lines 1 second apart on the graph (Figure 19-4). The volume of air measured between the two time lines is FEF200-1200 in liters per second. The volume measured must be corrected to BTPS.

FEF25%-75% is a measure of the flow during the middle portion of FVC, or the time necessary to exhale the middle 50%. For FEF25%-75%, the VC of the best curve is multiplied by 25% and 75%, and the points are identified on the tracing. A straight line is drawn connecting these points, and the line is extended to intersect two vertical time lines 1 second apart on the graph. The volume of air measured between the two time lines is FEF25%-75% in liters per second. The volume measured must be corrected to BTPS (Figure 19-5).

Peak Expiratory Flow

PEF is difficult to identify on a volume-time graph of FVC. The peak flow is the slope of the tangent to the steepest portion of the FVC curve. PEF is easy to identify on a flow-volume graph as the highest point on the graph (see Figure 19-2, B).23 PEF is sometimes measured independently of FVC with a peak flowmeter. These devices are designed to indicate only the greatest expiratory flow rate. The validity of PEF rate is based on a preceding inspiration to TLC and a maximal effort. The FVC principles of ensuring reliability should apply to measurements of PEF rate. The two largest repeated measurements should agree within 5%.

In addition to PEF rate, the other instantaneous flow rates, such as forced expiratory flow at 25% (FEF25%) of FVC, forced expiratory flow at 50% (FEF50%) of FVC, and forced expiratory flow at 75% (FEF75%) of FVC, during FVC are graphed on a flow-volume curve. When FVC is followed by a forced inspiratory VC, a flow-volume loop is produced (see Figure 19-2, B). On the flow-volume loop, the maximal forced inspiratory flow rate at 50% (FIF50%) of VC can be measured and compared with FEF50%.

Maximal Voluntary Ventilation

Another measurement of pulmonary mechanics is MVV. MVV is another effort-dependent test for which the patient is asked to breathe as deeply and as rapidly as possible for at least 12 seconds. MVV is a test that reflects patient cooperation and effort, the ability of the diaphragm and thoracic muscles to expand the thorax and lungs, and airway patency. Because of the potential for acute hyperventilation and fainting or coughing, the patient should be seated. Measuring systems that incorporate rebreathing may minimize hyperventilation. After a demonstration of the expected breathing pattern is performed, the patient should be instructed to breathe as rapidly and as deeply as possible for at least 12 seconds. The patient’s breathing is measured on a spirogram (Figure 19-6) or electronically for the specific number of seconds (t) and the volume (V) breathed when the MVV is converted to liters per minute. As with all volumes measured on a spirometer, the recorded values should be in BTPS conditions. The validity of MVV depends on the duration of the maneuver, which should be at least 12 seconds; the breathing frequency, which should be at least 90/min; and the average volume, which should be at least 50% of FVC. Patients should perform at least two MVV trials when the first trial does not exceed 80% of the subject’s FEV1 × 40, which may indicate less than maximal effort, or 80% of the predicted normal value, which may indicate disease. Reliability is shown when there is less than 20% variability between the two largest trials. The largest MVV (BTPS) should be reported.

Quality Assurance

For comprehensive quality assurance of spirometry, the accuracy and precision of the volume or flow measuring device must be verified with a 3.0-L syringe using multiple full strokes at various injection speeds to mimic fast and slow flow rates. The average volume should meet the ±3% standard; the standard deviation should be small, and the 95% confidence interval or expected performance range should be determined. Subsequent individual checks of accuracy with the 3.0-L syringe should be compared with the ±3% standard and the expected performance range to ensure that the device is measuring consistently. The performance of the technologist to coach subjects during spirometric testing should be observed and reviewed periodically. Technologists successfully achieving testing acceptability and repeatability criteria should be monitored and feedback to the technologist should be provided periodically.

Significance

The normal values for the spirometric measurements of pulmonary mechanics are based on height, age, gender, and ethnicity. Table 19-4 provides common regression equations to predict normal values for the measurements of pulmonary mechanics for individuals of specific height (in centimeters), age (in years), and gender.2426 A positive correlation exists between measurements of pulmonary mechanics and height, and a negative correlation exists between measurements of pulmonary mechanics and age for patients older than 20 years. Male values are larger than female values when height and age are equal. The populations that were studied to determine the normal values of pulmonary mechanics were predominantly white. To account for ethnic differences of nonwhites, the predicted normal values for whites commonly are reduced by 12% to 15% when applied to nonwhites. Ethnic-specific equations for special populations, such as African-Americans and Mexican-Americans, have also been developed.24

TABLE 19-4

Examples of Regression Equations for Predicted Normal Pulmonary Mechanics in White Adults

Parameters Equations R2 Reference
Men ≥20 years old      
FVC (L) 0.00018642 (Ht)2 + 0.00064 (A) − 0.000269 (A)2 − 0.1933 0.8668 24
FEV6 (L) 0.00018188 (Ht)2 − 0.00842 (A) − 0.000223 (A)2 + 0.1102 0.8692 24
FEV1 (L) 0.00014098 (Ht)2 − 0.01303 (A) − 0.000172 (A)2 + 0.5536 0.8510 24
% FEV1/FVC 88.066 − 0.2066 (A) 0.3448 24
FEF200-1200 (L/sec) 0.0429 (Ht) − 0.047 (A) + 2.010 0.440 25
FEF25%-75% (L/sec) 0.00010345 (Ht)2 − 0.04995 (A) + 2.7006 0.5601 24
PEF (L/sec) 0.00024962 (Ht)2 + 0.08272 (A) − 0.001301 (A)2 + 1.0523 0.7808 24
FEF25% (L/sec) 0.088 (Ht) − 0.035 (A) − 5.618   23
FEF50% (L/sec) 0.069 (Ht) − 0.015 (A) − 5.4   23
FEF75% (L/sec) 0.44 (Ht) − 0.012 (A) − 4.143   23
MVV (L/min) 1.19 (Ht) − 0.816 (A) − 37.9   26
Women ≥18 years old      
FVC (L) 0.00014815 (Ht)2 + 0.01870 (A) − 0.000382 (A)2 − 0.3560 0.7344 24
FEV6 (L) 0.00014395 (Ht)2 + 0.01317 (A) − 0.000352 (A)2 − 0.1373 0.7457 24
FEV1 (L) 0.00011496 (Ht)2 − 0.00361 (A) − 0.000194 (A)2 + 0.4333 0.7494 24
% FEV1/FVC 90.809 − 0.2125 (A) 0.3955 24
FEF200-1200 (L/sec) 0.0570 (Ht) − 0.036 (A) − 2.532 0.530 25
FEV25%-75% (L/sec) 0.00006982 (Ht)2 − 0.01904 (A) − 0.000200 (A)2 + 2.3670 0.5005 24
PEF (L/sec) 0.00018623 (Ht)2 + 0.06929 (A) − 0.001031 (A)2 + 0.9267 0.5559 24
FEF25% (L/sec) 0.043 (Ht) − 0.025 (A) − 0.132   23
FEF50% (L/sec) 0.035 (Ht) − 0.013 (A) − 0.444   23
FEF75% (L/sec) 3.042 − 0.014 (A)   23
MVV (L/min) 0.84 (Ht) − 0.685 (A) − 4.87   26

image

A, Years; Ht, cm; L, liters at BTPS.

Although traditional textbooks suggest the typical normal VC is 4.80 L, the predicted normal FVC for a 20-year-old, 180-cm man approaches 5.60 L. A reduced FVC may occur with obstructive or restrictive impairments. Figure 19-7 shows FVC from volume-time spirometer tracings for normal, obstructive, and restrictive conditions. The FVC values in both the obstructed and the restricted curves are shown as reduced volumes compared with the normal curve. The primary difference between the curve in the restricted patient compared with the curve in the obstructed patient is the slope of the tracing; obstructive diseases produce flattened slopes and smaller FEV1.

Figure 19-8 displays the FVC from flow-volume tracings for obstructive and restrictive conditions. The shapes of these tracings are different; obstructive diseases produce lower peaks and lower flow rates at all lung volumes. Forced inspiratory flow rates sometimes are useful for identifying extrathoracic airway obstructions. In moderate and severe obstructive lung diseases, the FVC is reduced if weakened bronchioles collapse and trap air in the lungs, creating an increase in RV. Some laboratories compare the volumes of slow vital capacity (SVC) and FVC to identify air trapping. VC is reduced in restrictive lung diseases because the patient’s inhaled volume is reduced.

Forced expiratory volume in half of a second (FEV0.5) is an indicator of patient effort during the initial phase of the FVC maneuver. With good effort, a patient should exhale at least 50% of his or her VC in the initial half of a second.

Although FEV1 is measured as a volume, FEV1 is considered a flow rate. The predicted normal FEV1 for a 20-year-old, 180-cm man approaches 4.70 L. FEV1 may be reduced with obstructive or restrictive impairments. For patients with airway obstruction, FEV1 measures the general severity of airway obstruction. For patients with restrictive impairment, FEV1 may be reduced when the patient’s VC is smaller than the predicted FEV1.

The FEV1/FVC ratio separates patients with airway obstruction from individuals with normal pulmonary function and from patients with restrictive impairment. The LLN can be determined or the predicted normal %FEV1/FVC can be calculated by dividing the predicted normal FEV1 by the predicted normal VC. Generally, individuals without airway obstruction are able to exhale at least 70% of their VC in the first second, and individuals with airway obstruction exhale less than 70% of their VC in the first second.

To interpret other flow rates, a generalization may be helpful. Gas exhaled during the early portion of the FVC reflects the resistance in the larger airways, and gas exhaled during the later portion of the FVC reflects the resistance in the smaller airways. As exhalation of FVC proceeds, flow decreases, and the airways reflected in the measurements get smaller. Any flow measured in the first half of the FVC reflects on the bronchi; any flow measured beyond 50% of the VC reflects on the bronchioles.

PEF, FEF200-1200, and FEF25% occur near the onset of FVC. Typical normal values are similar; PEF is 9.5 L/sec, FEF200-1200 is 8.5 L/sec, and FEF25% is 9.0 L/sec. A reduced PEF rate, FEF200-1200, or FEF25% may occur as a result of a large airway obstruction and from lack of sufficient effort to inhale maximally and exhale forcibly. FEF25%-75% and FEF50% occur in the middle of FVC. Because FEF25%-75% is an average of half the VC and FEF50% is an instantaneous flow, the typical normal values are less similar; FEF25%-75% is 4.5 L/sec, and FEF50% is 6.5 L/sec. Reduced FEF25%-75% or FEF50% may occur because of small airway obstruction and from lack of effort to sustain a maximal exhalation. FEF75% and FEF75%-85% occur late in FVC and reflect on the smallest airways. Typical values are 3.5 L/sec for FEF75% and 1.5 L/sec for FEF75%-85%. Sometimes patients who are asymptomatic for cough, sputum production, or dyspnea may have reduced flow in the small airways. A singular reduction in small airway flow may indicate nothing at all or may be an early indicator of obstruction.

The shape of the flow-volume loop and the FEF50%/FIF50% ratio provide additional information about upper airway obstruction. Compared with the normal flow-volume loop, a fixed upper airway obstruction produces a curve that appears box-shaped. In Figure 19-9, both expiratory and inspiratory flows are decreased and limited by the solid obstruction; the FEF50%/FIF50% ratio remains normal. Variable upper airway obstructions produce two different shapes depending on the site of the obstruction. Because the intraairway pressure during a forced inspiration is less than atmospheric outside the thorax, a variable extrathoracic upper airway obstruction limits inspiratory flow, and the FEF50%/FIF50% ratio is greater than 1.0. Because the intraairway pressure during a forced inspiration is greater than atmospheric pressure inside the thorax, a variable intrathoracic upper airway obstruction limits expiratory flow, and the FEF50%/FIF50% ratio is less than 1.0.

Similar to other spirometric measurements of pulmonary mechanics, normal values of MVV are based on gender, age, and height. MVV is reduced in patients with moderate and severe airway obstruction. A measured value less than 75% of predicted is significant. The normal for men is approximately 160 to 180 L/min; it is slightly lower in women. In restrictive lung disease, MVV may be normal or only slightly reduced. Respiratory muscle strength is a primary determinant of MVV in patients with interstitial lung disease and an important determinant in patients with chronic obstructive pulmonary disease (COPD). Undernourished patients also may have a reduced MVV.26

Reversibility

Based on the initial results of baseline spirometry, additional testing of pulmonary mechanics is often desirable. If the baseline test indicates airway obstruction, determining the reversibility of the obstruction is indicated. RTs also use the concept of reversibility when evaluating routine therapy by performing spirometry before and after therapy. In the laboratory, the FVC maneuver is often repeated after the patient has received a bronchodilator administered by small volume nebulizer or metered dose inhaler. This laboratory protocol is commonly known as spirometry before and after bronchodilator. Reversibility of the airway obstruction indicates effective therapy. Although improvement in other measurements of pulmonary function is sometimes used, reversibility is defined as a 15% or greater improvement in FEV1 and at least a 200-ml increase in FEV1. Improvement is determined using the percent change formula:

% Improvement=Post-FEV1Pre-FEV1Pre-FEV1×100

image

Bronchial Challenge

When the patient’s history suggests episodic symptoms of hyperreactive airways and airway obstruction, such as seasonal or exercise-induced wheezing, and the results of baseline spirometry are normal, performing a bronchial provocation may be indicated.27 Bronchial provocation testing uses an agent to stimulate a hyperreactive airway response and to create airway obstruction. Although several types of provocations are possible, such as inhaling histamine or cold air or exercising, provoking a hyperreactive airway response by inhaling methacholine is the most popular technique with the most predictable results. The procedure usually begins with the patient inhaling a normal saline aerosol and then repeating the FVC maneuver. Some very sensitive patients exhibit hyperreactive airways with saline alone; a positive response to saline is defined as a decrease in FEV1 of 10% or greater. The methacholine provocation protocol systematically exposes the patient to increasing dosages of methacholine. Usually starting with a low dose of 0.03 mg/ml, patients inhale methacholine aerosol and then repeat the FVC maneuver. A positive response to methacholine is defined as a decrease in FEV1 of 20% or greater (another example of percent change). If a positive response does not occur, the methacholine dose is doubled to 0.06 mg/ml, and the FVC maneuver is repeated. The process of “double-dosing” and performing FVC maneuvers continues until there is a positive response or until the full dose, 16 mg/ml, is given. If a positive response occurs, treatment with a fast-acting bronchodilator is indicated, and sometimes administering O2 is helpful. The final test report should include the concentration of methacholine that caused the 20% decrease in FEV1 in the form of PD%FEV1. For example, PD22FEV1 = 4 mg/ml indicates that the provocation dose of 4 mg/ml resulted in a 22% decrease in FEV1.

Lung Volumes and Capacities

There are four lung volumes and four lung capacities. A lung capacity consists of two or more lung volumes. The lung volumes are tidal volume (VT), inspiratory reserve volume (IRV), expiratory reserve volume (ERV), and residual volume (RV). The four lung capacities are TLC, IC, FRC, and VC. These volumes and capacities are shown in Figure 19-10. The lung volumes that can be measured directly with a spirometer or pneumotachometer include VT, IC, IRV, ERV, and VC. Because the RV cannot be exhaled, the RV, FRC, and TLC must be measured using indirect methods.

Knowing TLC is necessary to identify patients with a restrictive pattern of pulmonary impairment. Measuring FRC is necessary to quantify hyperinflation, which may be associated with obstructive impairment. Calculating RV is necessary to gauge any air trapping present. Measurements of VT, IC, ERV, IRV, and VC may be used in calculations of TLC or be useful to clinicians considering weaning parameters such as the rapid-shallow-breathing index (f/VT) or inspiration goals of hyperinflation therapy. Standards for measuring lung volumes and capacities were initially published in 2005; these standards focus primarily on the techniques to measure FRC.28 Following the measurement of FRC, measurements of ERV and VC enable calculation of TLC according to the formulas: TLC = (FRC − ERV) + VC and TLC = FRC + IC.

The VT is measured directly from a spirogram (see Figure 19-10). For the purposes of ensuring test validity and standardization, the patient should be in a sitting or reclining position and wearing a nose clip. It sometimes takes the patient 1 to 2 minutes to be at rest and become accustomed to the nose clip and mouthpiece. The patient breathes through a tight-fitting mouthpiece until a normal, rhythmic breathing pattern is established. Because VT varies normally from breath to breath, an average VT is a more reliable measurement. In the laboratory, an average VT sometimes is measured during 3 minutes of quiet breathing while the spirometer records volumes and graphs volume and time. At the bedside, an average VT usually is measured over 1 minute; the patient breathes normally into a spirometer that stores in memory each volume exhaled for 1 minute and computes an average. An alternative approach is to measure the total volume of air exhaled for 1 minute (image) and divide by the breathing frequency (f) counted during the same period. The following formula can be used to calculate the VT: VT = image ÷ f.

The inspiratory capacity (IC) is also measured directly from a spirogram. The patient is asked to inhale maximally from the resting FRC at the end of a normal effortless exhalation. To ensure validity, a consistent resting expiratory level should be obvious on the spirogram before inhaling. To ensure reliability, the IC should be measured at least twice, and the two largest measurements should agree within 5%. Because the definition of IC is the maximal volume inhaled, the largest measurement is the patient’s IC. (See Clinical Practice Guidelines 19-2 and 19-3.)

19-2   Methacholine Challenge Testing

AARC Clinical Practice Guideline (Excerpts)*

Contraindications

The following are absolute contraindications to methacholine challenge testing:

Relative contraindications include the following:


*For the complete guideline, see American Association for Respiratory Care: Clinical practice guideline: methacholine challenge testing: 2001 revisions and updates, Respir Care 46:523–530, 2001.

19-3   Static Lung Volumes

AARC Clinical Practice Guideline (Excerpts)*


*For the complete guideline, see American Association for Respiratory Care: Clinical practice guideline: static lung volumes: 2001 revisions and updates, Respir Care 46:531–539, 2001.

The expiratory reserve volume (ERV) is measured directly from the spirogram (see Figure 19-10). The patient is asked to breathe normally for a few breaths and then exhale maximally. The ERV is the volume of air exhaled between the resting expiratory level and the maximal exhalation level on the spirogram. To ensure validity, a consistent resting expiratory level should be obvious on the spirogram before exhaling maximally. To ensure reliability, the ERV should be measured at least twice and the two largest measurements should agree within 5%. Because the definition of ERV is the maximal volume exhaled, the largest measurement is the patient’s ERV.

The VC is the most commonly measured lung volume. There are several methods of measuring the VC. The VC can be measured during inspiration or during a slow prolonged expiration when air trapping is of concern. To measure the VC during inspiration, the patient exhales maximally and then inhales as deeply as possible. The volume of the maximal inspiration is the inspiratory VC. To measure the VC during expiration, the patient inhales maximally and then exhales maximally, taking all the time necessary to exhale completely. The exhaled volume is the slow VC. An alternative method is to measure the IC and the ERV and add these volumes together for a “combined” VC, but this method should be reserved only for patients who cannot otherwise execute the VC. The VC also is measured when it is exhaled forcefully and as rapidly as possible. This technique is called the FVC, and it is used to assess pulmonary mechanics under the section spirometry.

Because the RV cannot be exhaled, RV, FRC, and TLC cannot be measured directly with a spirometer or pneumotachometer. There are three indirect techniques to measure these lung volumes: helium dilution, nitrogen washout, and body plethysmography. The helium dilution and nitrogen washout techniques measure whatever gas is in the lungs at the beginning of the test, if the gas is in contact with unobstructed airways. The body plethysmographic technique measures all the gas in the thorax at the resting expiratory volume. Because the plethysmographic technique measures all gas in the thorax, including gas that is trapped distal to obstructed airways or gas in the pleural space, the lung volume measured by this technique is called the thoracic gas volume (TGV) (VTG, or FRCPleth). In patients with obstructive lung disease with gas trapping, TGV is often larger than FRC measured by helium dilution or nitrogen washout. In healthy individuals, TGV is identical to FRC measured by both the gas dilution and the washout techniques. When FRC is known, RV can be calculated as the difference between FRC and ERV. TLC also can be calculated by adding RV to VC.

Helium Dilution

The helium dilution technique for measuring lung volumes uses a closed, rebreathing circuit (Figure 19-11).29 This technique is based on the assumptions that a known volume and concentration of helium in air begin in the closed spirometer, that the patient has no helium in his or her lungs, and that an equilibration of helium can occur between the spirometer and the lungs. For the helium dilution procedure to be performed, a measurable volume of helium is added into the spirometer circuit, and the initial concentration of helium (FiHe) is measured. Next, the valve is turned to connect the patient to the breathing circuit usually at the resting expiratory level of the FRC. Starting the test at RV requires maximal expiratory effort by the patient and is not considered a reliable starting point. Although starting the test at the TLC level requires a maximal inspiration, TLC may be a reliable alternative beginning point.

The patient is connected to the helium-air mixture, and the concentration of helium is diluted slowly by the patient’s lung volume. Wearing nose clips, the patient breathes normally in the closed circuit. Exhaled CO2 is absorbed with soda lime, and O2 is added at a rate equal to the patient’s O2 consumption. A constant volume is maintained to ensure accurate helium concentration measurements. The patient rebreathes the gas in the system until equilibrium of helium concentration is established. In healthy patients and patients with a small FRC, equilibration occurs in 2 to 5 minutes. Patients with obstructive lung disease may require 20 minutes to equilibrate because of slow gas mixing in the lungs. The helium dilution time or the duration of the test is a gross index of the distribution of ventilation.

For FRC to be calculated using the helium dilution technique, several observations must be made: the volume of helium added (vol He) to the closed spirometer, the initial helium concentration (FiHe) before the patient is connected to the breathing circuit, the final helium concentration (FfHe) after helium equilibrium between the spirometer and patient is established, the spirometer temperature, and the time necessary for helium equilibration to occur. If the patient is connected to the circuit at the resting level, the FRC can be calculated with the following equation:

FRC=(vol He÷FiHe)×[(FiHeFfHe)×FfHe]

image

Corrections for temperature and helium absorption are normally applied. All lung volumes and capacities must be reported under BTPS conditions. Volumes measured by spirometers are at ambient temperature, pressure, and saturated (ATPS) conditions and must be adjusted for the temperature difference between the spirometer and the patient’s body temperature. This ATPS to BTPS adjustment can increase volumes 5% to 10%, and the difference is large enough to invalidate the test results, unless the correction is made.

Although helium is an inert gas with a negligible solubility in plasma, another correction is sometimes applied. A small amount of helium is thought to diffuse across the alveolar-capillary membrane and is lost in the measurement of final helium concentration. To account for the loss, 30 ml of BTPS-corrected volume is subtracted for each minute of helium breathing, up to 200 ml for a 7-minute test.30 Once these corrections are made, RV can be calculated by subtracting ERV from FRC according to the equation: RV = FRC − ERV.

Nitrogen Washout

The nitrogen washout technique uses a nonrebreathing or open circuit (Figure 19-12).31 The technique is based on the assumptions that the nitrogen concentration in the lungs is 78% and in equilibrium with the atmosphere, that the patient inhales 100% O2, and that the O2 replaces all of the nitrogen in the lungs. Similar to the helium dilution technique, the patient is connected to the system at either the resting expiratory level or the TLC. The patient’s exhaled gas is monitored, and its volume and nitrogen percentage are measured.

Generally, two types of circuits are used to measure lung volumes with this technique. In one type of circuit, all of the exhaled gases are collected in a large container, where the volume and concentration of nitrogen are measured. In the second type of circuit, the volume and concentration of each exhaled breath are measured separately and stored in a memory; the sum of the volumes and the weighted average of the nitrogen concentration are calculated by a computer.

Wearing nose clips, the patient breathes 100% O2 until nearly all of the nitrogen has been washed out of the lungs, leaving less than 1.5% nitrogen in the lungs. When the peak exhaled concentration of nitrogen is less than 1.5%, the patient exhales completely, and the fractional concentration of alveolar nitrogen (FAN2) is noted. Similar to the helium technique, the time it takes to wash out the nitrogen is approximately 2 to 5 minutes in healthy individuals and longer in patients with obstructive lung disease. The test must occur in a leakproof circuit because the presence of any air increases the measured nitrogen percentages and results in grossly elevated measurements of lung volume.

For FRC or TLC to be calculated by the nitrogen washout technique, several measurements must be made: the total volume of gas exhaled during the test (VE), the fractional concentration of exhaled nitrogen in the total gas volume (FEN2), the fractional concentration of nitrogen in the alveoli at the end of the test (FAN2), and the spirometer temperature. FRC (or TLC, if the test began at TLC) can be calculated with the following equation:

FRC=(VE×FEN2)÷(0.78FAN2)

image

The calculated FRC (or TLC, if connection to the breathing circuit occurred after inspiring maximally) must be adjusted for the temperature difference between the spirometer and the patient’s body temperature using the BTPS correction factor. During the test, some nitrogen from the plasma and body tissues is usually excreted and exhaled with lung nitrogen. For this reason, another correction is needed. The volume of tissue nitrogen excreted (Vtis in milliliters) is directly related to the duration (t in minutes) of the test and weight (W in kilograms) of the patient. A correction for this extra nitrogen should be made according to the following formula: Vtis (ml) = (0.1209 image − 0.0665) × (W ÷ 70). Vtis (ml) is subtracted from the BTPS-corrected lung volume. RV is the difference between ERV and FRC.

The validity of the helium dilution and nitrogen washout techniques can be ensured by measuring known volumes accurately, such as a 3.0-L syringe, while recognizing that this method would not include O2 consumption and O2 titration for the helium method or tissue nitrogen excretion for the nitrogen method. The reliability of these techniques can be established by repeated measurements of known volumes or healthy patients that agree within 5%. Establishing reliability of these measurements on patients requires repeating tests after reestablishing baseline lung gases. For patients with severe obstructive lung disease, waiting up to 1 hour may be necessary. When the FRC is measured by either He dilution or N2 washout, any leak anywhere in the patient, tubing, or gas analyzers would increase the measured FRC and may be falsely interpreted as hyperinflation.

Plethysmography

The plethysmography technique applies Boyle’s law and uses measurements of volume and pressure changes to determine lung volume, assuming temperature is constant.32 The plethysmography technique measures the volume of all compressible gas in the thorax, including gas trapped behind airway obstructions or in the pleural space. Gas in the abdomen may also be included in the measurement. The whole-body plethysmograph consists of a sealed chamber in which the patient sits (Figure 19-13). Pressure transducers (electronic manometers) measure pressure at the mouth and in the chamber. An electronically controlled shutter near the mouthpiece allows the airway to be occluded periodically, measuring airway pressure changes under conditions of no airflow. Without air flow, pressure changes measured at the mouth are pressure changes in the alveoli. According to Boyle’s law (V × P = k), when temperature is constant, volume changes in the thorax create volume changes in the chamber, which are reflected by pressure changes in the chamber. The pressure and volume changes are included in the equation:

VTG1×Palv1=VTG2×Palv2

image

When measurement of TGV is being done, the patient sits in the chamber and initially breathes normal tidal volumes through the mouthpiece. When the patient is near FRC, the shutter is closed at end expiration for 2 to 3 seconds. The patient holds his or her cheeks and performs gentle panting at 1 Hz or one pant per second.33 During panting, changes in airway pressure (ΔP) and changes in chamber volume (ΔV) are measured. Because the panting maneuver occurs with small pressure changes around barometric pressure, the simplified equation used to calculate TGV is VTG = PB × (ΔV ÷ ΔP), where PB is the barometric pressure in cm H2O. A series of three to five panting maneuvers should be performed. After panting, the patient should exhale completely to record ERV and then inhale maximally to record the inspiratory vital capacity.

Because the body plethysmographic method of measuring FRC actually measures TGV, the value obtained for some patients may be larger than values resulting from either the helium dilution or nitrogen washout techniques. Such a difference occurs whenever there is gas in the thorax that is not in communication with patent airways, as might be the case in patients with pneumothorax, pneumomediastinum, or emphysema. The RV is the difference between TGV and ERV, and the TLC is the sum of RV and VC, or the sum of TGV and IC.

Quality Assurance

Comprehensive quality assurance for measuring lung volumes depends on the measuring technique. For helium dilution and nitrogen washout techniques, the accuracy and precision of the volume or flow measuring device must be ensured as in spirometry; the accuracy and linearity of the gas analyzer must be verified, and the leak test must be acceptable while monitoring change in volume and gas concentrations over at least 1 minute. Technologists successfully identifying the subject’s breath-to-breath resting level and complying with retest waiting periods should be monitored. Correctly measuring the volume of the 3.0-L syringe provides a quality control standard. For the plethysmographic technique, the box and mouth pressure transducers must be calibrated and accurate. The technologist’s choice of the best-fit line of the relationship between box and mouth pressures should be monitored. Measuring a known volume using a flask with squeeze bulb connected to the mouthpiece or using a consistent known subject provides a quality control standard. For all techniques, achieving testing acceptability and repeatability criteria should be monitored, and feedback to the technologist should be provided periodically.

Significance

Changes in lung volumes and capacities are generally consistent with the pattern of impairment. TLC, FRC, and RV increase with obstructive lung diseases and decrease with restrictive impairment. Some lung volumes provide valuable diagnostic information. For example, TLC is always reduced in restrictive lung disease, unless obstruction and restriction occur together. When obstruction and restriction occur together, the TLC may be a less sensitive measure of the restrictive impairment. Other volumes and capacities may remain normal with mild obstructive or restrictive disease. The pattern of lung volume changes and the proportion of FRC and RV to TLC are also important.

The normal VT is approximately 500 to 700 ml for an average healthy adult. In the normal population, great variation of tidal volumes and measurements beyond the normal range are not indicative of a disease process. Normal VT is often observed in both restrictive and obstructive lung diseases. VT alone is not a valid indicator of the type of lung disease.

The normal IC is approximately 3600 ml, with a significant variation in the normal population. IC may be normal or reduced in restrictive and obstructive lung diseases. A reduction of IC occurs in restrictive lung diseases because the patient’s inhaled volume is reduced, and there is a reduction in TLC. In mild obstructive lung diseases, IC is usually normal. In moderate and severe obstructive diseases, IC can be reduced because the resting expiratory level of FRC has increased owing to hyperinflation of the lungs. An increase in IC may occur when the patient inhales from below the resting expiratory level when the measurement is performed; athletes and musicians who play wind instruments may also have increased inspiratory capacities. RTs use the measurement of IC in clinical protocols to decide between methods of lung expansion therapy (see Chapter 39).

IRV is not commonly measured. Similar to VT and IC, IRV can be normal in both restrictive and obstructive diseases and is not a useful diagnostic measurement. The normal value for IRV is 3.10 L.

The normal ERV is approximately 1.20 L and represents approximately 20% to 25% of the VC. It can be either normal or reduced in obstructive and restrictive lung diseases. ERV is subtracted from FRC to calculate RV.

The normal value of the VC is 4.80 L and represents approximately 80% of TLC. Normal values for VC can vary significantly depending on age, gender, height, and ethnicity. A reduction of VC occurs in restrictive lung diseases because the patient’s inhaled volume is reduced and there is a reduction in TLC. In mild obstructive lung diseases, the slow VC is usually normal if the patient exhales leisurely and has had enough time to exhale completely or if the VC is measured during inspiration. Measurements made from FVC provide valuable data for pulmonary mechanics.

RV, FRC, and TLC are the most important measurements of lung volumes. Age, height, gender, ethnicity, and sometimes weight or body surface area correlate with normal values for these lung volumes.34 Table 19-5 provides common regression equations to predict the lung volumes for individuals of specific height (in centimeters), age (in years), and gender. A positive correlation exists between lung volumes and height, and a negative correlation exists between lung volumes and age for patients older than 20 years. Male values are larger than female values when height and age are equal.

TABLE 19-5

Examples of Regression Equations for Predicting Normal Lung Volumes and Capacities in Adults

Lung Volumes Equations
Men  
FRC (L) 0.0234 (Ht) + 0.01 (A) − 1.09
RV (L) 0.0131 (Ht) + 0.022 (A) − 1.23
TLC (L) 0.0799 (Ht) − 7.08
FRC/TLC% 43.8 + 0.21 (A)
RV/TLC% 14.0 + 0.39 (A)
Women  
FRC (L) 0.0224 (Ht) + 0.001 (A) − 1.00
RV (L) 0.0181 (Ht) + 0.016 (A) − 2.00
TLC (L) 0.0660 (Ht) − 5.79
FRC/TLC% 45.1 + 0.16 (A)
RV/TLC% 19.0 + 0.34 (A)

A, Years; Ht, cm; L, liters at BTPS.

Stocks J, Quanjer PH: Reference values for residual volume, functional residual capacity and total lung capacity. Eur Respir J 8:492–497, 1995.

The typical normal TLC is 6.00 L. The normal RV is approximately 1.20 L and represents approximately 20% of TLC. FRC is approximately 2.40 L, which represents approximately 40% of the TLC. RV and FRC are usually enlarged in acute and chronic obstructive lung diseases because of hyperinflation and air trapping (Figure 19-14).

TLC may also be enlarged in COPD. TLC is always reduced in restrictive lung diseases because of a loss of lung volume; RV and FRC are often reduced proportionately. Certain acute disorders, such as pulmonary edema, atelectasis, and consolidation, also cause a reduction of TLC and FRC.

Diffusing Capacity

The third major category of pulmonary function testing is measuring the ability of the lungs to transfer gases across the alveolar-capillary membrane. The diffusing capacity of the lung (DL) is sometimes called the transfer factor. Carbon monoxide (CO) is the gas normally used to measure the DL. The diffusing capacity of the lung for carbon monoxide (DLCO) is expressed in ml/min/mm Hg under standard temperature and pressure and dry conditions (STPD). Generally, DLCO is the difference between the volume of CO inhaled and the volume of CO exhaled, considering the partial pressure of CO in the lung at the time of measurement. The general equation used to explain the diffusing capacity is DLCO = VE (FiCO − FeCO) ÷ (PaCO). The difference between CO inhaled and exhaled is attributed to CO diffusing into the pulmonary capillary blood.

CO is used as the transfer gas because CO is similar to O2 in important ways. CO and O2 have similar molecular weights and solubility coefficients. Similar to O2, CO also chemically combines with hemoglobin (Hb). CO has a very high affinity for Hb and diffuses rapidly into the pulmonary blood. CO has an affinity for Hb nearly 210 times greater than O2, and the high affinity keeps the pulmonary capillary partial pressure of CO (PcCO) near zero. Consequently, the diffusion of CO across the alveolar-capillary membrane is membrane limited and not limited as much by the partial pressure gradient.

Factors known to affect test results should be controlled or standardized; these include body position, activity, PaO2, Hb and carboxyhemoglobin (COHb) levels, and pulmonary blood volume. To focus the test on diffusion through the alveolar-capillary membrane, the patient should be tested at rest in a seated position, should not breathe supplemental O2 for 10 minutes before testing, and should not have an abnormal level of COHb before the test. Mathematical corrections can be applied for patients who cannot abstain from O2. Performing the diffusing capacity on patients who have recently smoked a cigarette or who have been exposed to environmental CO may hinder test validity.35 Patients should refrain from smoking on the day of the test. Hb of all patients undergoing diffusing capacity should be measured, and a mathematical correction should be applied if Hb level is abnormal.

Single-Breath Technique

There are several techniques to measure the diffusing capacity of the lung for CO, including steady-state, intrabreath, and rebreathing techniques, but the single-breath method (DLCOSB) is the most common measurement technique because it is quick and reproducible. Standards for measuring diffusing capacity of the lung were initially published in 1995 and updated in 2005; these standards focus primarily on the DLCOSB.36,37 The entire test can be performed in just slightly longer than 10 seconds. The patient exhales completely to RV, rapidly inspires a VC of a gas mixture containing 0.3% CO and an inert tracer gas such as He in air, maintains breath holding for 10 seconds, and then exhales rapidly at least 1.0 L. Inspiring at least 85% of VC measured during spirometry is expected. After a predesignated volume of 0.75 to 1.0 L is exhaled, a sample of alveolar gas is collected and analyzed for expired CO (FeCOt) and helium (FeHe). Because the breath holding period (t) begins when inspiration of the gas mixture begins and the period ends when the alveolar sample is collected, inspiration and expiration should be rapid, and this period should not exceed 11 seconds. To regulate the breath holding period, some measuring systems close the mouthpiece with a timed shutter. The suitable breathing pattern requires some patient cooperation and coordination; some patients benefit from a timer as a visual aid.

The single-breath method (DLCOSB) is based on the diffusion decay curve described by Forster and colleagues38 (Figure 19-15). When a bolus of CO gas is inhaled, the rate of gas diffusion declines logarithmically. The diffusing capacity of the lung is a function of lung volume (VA) in STPD conditions exposed to the test gas,39 the duration (60 ÷ t) that the test gas is in contact with the lung, the initial concentration of test gas in the lung (FaCO0), and the final concentration of test gas in the lung (FaCOt), according to the following equation:

image

where (PB − 47) is ambient barometric pressure corrected for water vapor pressure at 37° C.

Helium or sometimes neon is in the gas mixture as a tracer gas for the dilution of the inspired CO concentration by RV and to measure the effective TLC by a single-breath helium dilution. The inspired CO concentration of 0.3% is not the concentration of CO received by the lungs because the inspired volume is diluted by RV of the patient. The dilution of CO is reflected by the dilution of helium, and the FaCO0 can be calculated according to the following equation:

image

The FaCO0 is the concentration of CO in the lung at “zero time” before any diffusion occurs. The single-breath technique distributes the gas mixture through unobstructed airways to an alveolar volume that is also called the effective total lung capacity. The effective total lung capacity (VA) can be calculated according to the following equation:

VA=VC×(FiHe÷FeHe)

image

The VA is necessary to calculate the DLCO, and it is used in the determination of the diffusing capacity of the lung-to-alveolar volume ratio (DLCO/VA).

The reliability of the DLCO is based on repeatability of the test. At least 4 minutes should be allowed between tests to allow an adequate elimination of CO from the lungs. In patients with obstructive airway disease, a longer period (e.g., 10 minutes) may be necessary. The actual DLCO reported should be the mean of 2 acceptable tests. An acceptable test is defined as one that is reproducible to within 10% or 3 ml of the CO/min/mm Hg value, whichever is greater. (See Clinical Practice Guideline 19-4.)

19-4   Single-Breath Carbon Monoxide Diffusing Capacity

AARC Clinical Practice Guideline (Excerpts)*

Indications

Tests of diffusing capacity may be indicated in the following situations:

• Evaluation and follow-up of parenchymal lung diseases associated with dusts (e.g., asbestos) or drug reactions (e.g., amiodarone) or related to sarcoidosis

• Evaluation and follow-up of emphysema and cystic fibrosis

• Differentiation between chronic bronchitis, emphysema, and asthma in patients with obstructive patterns

• Evaluation of pulmonary involvement in systemic diseases (e.g., rheumatoid arthritis, lupus erythematosus)

• Evaluation of cardiovascular diseases (e.g., pulmonary hypertension, pulmonary edema, thromboembolism)

• Prediction of arterial desaturation during exercise in COPD

• Evaluation and quantification of disability associated with interstitial lung disease

• Evaluation of the effects of chemotherapy agents or other drugs known to induce pulmonary dysfunction

• Evaluation of hemorrhagic disorders

Assessment of Test Quality

Individual test maneuvers and results should be evaluated according to the American Thoracic Society recommendations. In particular, the following recommendations are pertinent:

• The inspiratory volume should exceed 90% of the largest previously measured vital capacity (FVC or VC).

• Breath hold time should be between 9 seconds and 11 seconds, with a rapid inspiration.

• The washout volume (dead space) should be 0.75 to 1 L, or 0.50 L if the subject’s VC is <2 L. If a washout volume other than 0.75 to 1 L is used, it should be noted.

• Two or more acceptable tests should be averaged. The maneuvers should be reproducible to within 10% or 3 ml of CO/min/mm Hg, whichever is greater.

• The subject should have refrained from smoking for 24 hours before the test.

• Corrections for Hb and COHb should be included; correction for tests performed at high altitude is recommended.

• If Hb correction is made, both the corrected and the uncorrected DLCO values should be reported.

• Equipment calibration and quality control measures specific to measuring diffusing capacity should be applied and documented.


*For complete guideline, see American Association for Respiratory Care: Clinical practice guideline: single-breath carbon monoxide diffusing capacity, 1999 update, Respir Care 44:539–546, 1999.

Significance

Normal values for the DLCO using the single-breath technique are based primarily on a patient’s age, height, and gender (Table 19-6). A typical normal value for a 20-year-old healthy man is 40 ml/min/mm Hg.40 Before interpretation of the test result, corrections for abnormal Hb level, breathing supplemental O2 or testing at altitude, and elevated COHb levels may be necessary, if these conditions apply. Initially, all Hb levels were corrected to 15 g/dl.41 More recent and more specific corrections have been advocated. For men and adolescents, if the Hb level varies from 14.6 g/dl, the predicted normal value should be adjusted using the following equation:42

TABLE 19-6

Examples of Regression Equations for Predicting Normal Diffusing Capacity in Adults

Parameter Regression Equations
Men  
DLCOSB (ml/min/mm Hg) 0.0416 (Ht) − 0.219 (A) − 26.34
DLCOSB/VA (ml/min/mm Hg/L) 6.61 − 0.034 (A)
Women  
DLCOSB (ml/min/mm Hg) 0.0256 (Ht) − 0.144 (A) − 8.36
DLCOSB/VA (ml/min/mm Hg/L) 7.34 − 0.032 (A)

A, Years; Ht, cm; L, Liters at BTPS; ml, milliliters at STPD.

Crapo RO, Morris AM: Standardized single breath normal values for carbon monoxide diffusing capacity. Am Rev Respir Dis 123:185, 1981.

Predicted DLCO for Hb=Predicted DLCO×(1.7Hb/10.22+Hb)

image

For women and children, the normal Hb level is 13.4 g/dl, and the equation is42:

Predicted DLCO for Hb=Predicted DLCO×(1.7 Hb/9.38+Hb)

image

If the patient’s PaO2 differs from 100 mm Hg because of breathing supplemental O2 or performing the test at altitude, the DLCO would be affected by approximately 0.31% to 35% per mm Hg difference from 100 mm Hg.43,44 The predicted normal value should be adjusted using the following equations:

image

Predicted DLCO for altitude=Predicted DLCO/[1.0+0.0031(PiO2150)]

image

The DLCO may be reduced from the predicted normal in patients with obstructive or restrictive lung diseases. With destruction of alveoli in pulmonary emphysema, with small lung volumes, and with fibrosis of alveoli in asbestosis, the DLCO may be less than normal. Pulmonary embolism also may decrease the DLCO. The DLCO may be useful in identifying which patients with obstructive impairment are likely to desaturate during exercise and which may benefit from O2 therapy. The DLCO may be increased in patients with polycythemia, congestive heart failure (resulting from an increase in pulmonary vascular blood volume), and elevated cardiac output. Factors that can alter the DLCO above or below the normal value are summarized in Box 19-2.

The diffusing capacity of the lung-to-effective total lung capacity ratio (DLCO/VA) differentiates between diffusion abnormalities caused by having a small lung volume compared with diffusion abnormalities caused by alveolar-capillary membrane pathologies. Patients whose only problem is small lungs would have a decreased DLCO, but their DLCO/VA ratio would be normal. Patients with pulmonary emphysema or fibrosis would have a decreased DLCO and a decreased DLCO/VA ratio.

Interpretation of the Pulmonary Function Report

Interpretive strategies for pulmonary function testing abound. Most computer-based pulmonary function testing systems have algorithms in their software programs for computer-assisted interpretations of the pulmonary function report.45 A consensus for interpreting test results is growing.46,47 Table 19-7 summarizes pulmonary function changes that may occur in advanced obstructive and restrictive patterns of lung diseases, and Figure 19-16 presents a simple algorithm to assess pulmonary function test results in clinical practice.47

TABLE 19-7

Pulmonary Function Changes in Advanced Lung Diseases

Measurement Normal* Obstructive Restrictive
VT 500%smL N or ↑ N or ↓
IRV 3.10 L N or ↓
ERV 1.20 L N or ↓
RV 1.20 L
IC 3.60 L N or ↓
FRC 2.40 L
TLC 6.00 L N or ↑
FVC 4.80 L
FEV1 4.20 L N or ↓
FEV1/FVC >70% N or ↑
FEF200-1200 8.5 L/sec N
FEF25%-75% 4.5 L/sec N
PEF 9.5 L/sec N
FEF25% 9.0 L/sec N
FEF50% 6.5 L/sec N
FEF75% 3.5 L/sec N
MVV 160 L/min N or ↓
DLCO 40%smL/min/mm Hg N or ↓ N or ↓
DLCO/VA 6.6%smL/min/mm Hg/L N or ↓ N or ↓

image

N, No change.

*Values for 20-year-old, 70-kg man.

When considering a pulmonary function report, the %FEV1/VC ratio is a good place to start because it provides an initial focus as normal, restrictive, or obstructive impairment. When the %FEV1/FVC is less than the LLN, there is airway obstruction. When the %FEV1/FVC is greater than the LLN, there is no airway obstruction. The LLN %FEV1/FVC can be determined directly for various populations using regression equations in Table 19-8 or simply estimated at 70%. If the %FEV1/FVC ratio is greater than the LLN or 70% and if the TLC is less than the LLN, often defined as less than 80% predicted normal, the patient has a restrictive impairment according to this algorithm. The severity of the restriction is based on the percent predicted or on the number of standard deviations below the LLN TLC according to Table 19-2. If the %FEV1/FVC ratio is less than 70%, the patient likely has an obstructive impairment; the severity of the obstruction is based on the percent predicted normal FEV1 according to Table 19-2. If the percent predicted normal DLCO is less than 80%, the patient has a diffusion impairment. Some laboratories also report the DLCO/VA ratio, which indexes the DLCO for lung volume measured during the single-breath test. If the DLCO/VA ratio is also less than 80% of the indexed value, the cause of the diffusion impairment is considered to be within the lung, and if the DLCO/VA ratio is greater than 80% of the indexed value, the cause of the diffusion impairment is considered to be due to small lung volume.

TABLE 19-8

Examples of Regression Equations for Determining Lower Limit of Normal of Forced Expiratory Volume in 1 Second-to-Vital Capacity Ratio (%FEV1/FVC) in Adults

Population Equations R2
Men    
White 78.388 − 0.2066 (A) 0.3448
African-American 78.822 − 0.1828 (A) 0.1538
Mexican-American 80.925 − 0.2186 (A) 0.2713
Women    
White 81.015 − 0.2125 (A) 0.3955
African-American 80.978 − 0.2039 (A) 0.2284
Mexican-American 83.044 − 0.2248 (A) 0.3352

A, Years.

Hankinson JL, Odencratz JR, Fedan KB: Spirometric reference values from a sample of the general U.S. population. Am J Respir Crit Care Med 159:179, 1999.

Mini Clini

Identifying Patterns of Pulmonary Impairment

Pulmonary Function Report 1

Pulmonary Measurements Predicted Normal Value Measured Baseline Conditions Percent Predicted Baseline Measured After Bronchodilator Treatment Percent Predicted After Treatment
SVC (L) 5.00 3.00 60%    
FVC (L) 5.00 3.00 60%    
FEV1 (L) 4.00 2.80 70%    
%FEV1/FVC 80% 94%    
FEF25%-75% (L/sec) 4.00 3.75 94%    
PEF (L/sec) 8.00 8.25 103%    
TLC (L) 7.20 3.96 55%    
FRC (L) 4.10 2.00 49%    
RV (L) 2.20 1.40 60%    
DLCO (ml/min/mm Hg) 34.0 15.9 47%    
DLCO/VA (ml/min/mm Hg/L) 7.20 3.50 49%    

image

Pulmonary Function Report 2

Pulmonary Measurements Predicted Normal Value Measured Baseline Conditions Percent Predicted Baseline Measured After Bronchodilator Treatment Percent Predicted After Treatment
SVC (L) 5.00 3.50 70% 4.25 85%
FVC (L) 5.00 3.30 66% 4.00 80%
FEV1 (L) 4.00 2.00 50% 2.50 62%
%FEV1/FVC 80% 57% 62%
FEF25%-75% (L/sec) 4.00 1.00 25% 2.00 50%
PEF (L/sec) 8.00 6.00 75% 6.50 81%
TLC (L) 5.27 5.51 105% 5.36 102%
FRC (L) 3.11 4.55 146% 3.60 116%
RV (L) 1.67 2.60 156% 2.00 120%
DLCO (ml/min/mm Hg) 28.7 25.25 88%
DLCO/VA (ml/min/mm Hg/L) 5.45 5.17 96%

image

Solution Report 2

The FEV1/FVC is less than 70%; there is airway obstruction. FEV1 is 50% of predicted; the obstruction is moderate. Because the FEF25%-75% is 25% of predicted, the major site of obstruction is in the bronchioles. After bronchodilator inhalation, FEV1 improved by 24% (remember to compute percent change), showing effective treatment and partial reversibility of the obstruction. The large FRC and RV show hyperinflation and air trapping, which also improved after bronchodilator therapy. Diffusing capacity is in the normal range, indicating no diffusion impairment and no alveolar problems. Overall, this report shows a moderate obstructive pattern with hyperinflation and air trapping responsive to bronchodilators and consistent with acute hyperreactive airways disease, such as asthma.

Pulmonary Function Report 3

Pulmonary Measurements Predicted Normal Value Measured Baseline Conditions Percent Predicted Baseline Measured After Bronchodilator Treatment Percent Predicted After Treatment
SVC (L) 5.00 4.00 80% 4.25 85%
FVC (L) 5.00 3.50 70% 4.00 80%
FEV1 (L) 4.00 2.00 50% 2.20 55%
%FEV1/FVC 80% 57% 55%
FEF25%-75% (L/sec) 4.00 1.75 50% 2.00 50%
PEF (L/sec) 8.00 6.00 75% 6.50 80%
TLC (L) 5.27 5.51 105% 5.36 102%
FRC (L) 3.11 4.55 146% 3.79 122%
RV (L) 1.67 2.60 156% 2.24 134%
DLCO (ml/min/mm Hg) 28.7 14.25 56%
DLCO/VA (ml/min/mm Hg/L) 5.45 3.17 58%

image

Solution Report 3

This case is similar to case 2, but there are some important differences. The FEV1/FVC is less than 70%; there is airway obstruction. FEV1 is 50% of predicted; the obstruction is moderate. After a single bronchodilator treatment, FEV1 improved by 10% (remember to compute percent change)—not enough to show that bronchodilator therapy was immediately effective. The large FRC and RV show hyperinflation and air trapping, which did improve after bronchodilator therapy. DLCO and DLCO/VA are reduced, suggesting alveolar involvement. This report shows a moderate obstructive pattern with hyperinflation and air trapping not responsive to bronchodilators. There is diffusion impairment and alveolar disease. Overall, this report is consistent with COPD, the combination of chronic bronchitis and pulmonary emphysema.

Pulmonary Function Report 4

Pulmonary Measurements Predicted Normal Value Measured Baseline Conditions Percent Predicted Baseline Measured After Bronchodilator Treatment Percent Predicted After Treatment
SVC (L) 5.38 4.84 90%    
FVC (L) 5.38 4.92 92% 5.16 109%
FEV1 (L) 4.33 2.95 68% 3.24 75%
%FEV1/FVC 80% 54% 63%  
FEF25%-75% (L/sec) 5.23 1.20 23% 1.08 21%
PEF (L/sec) 9.96 6.32 63% 7.53 76%
TLC (L) 7.51 6.38 85%    
FRC (L) 4.10 3.51 86%    
RV (L) 2.10 1.58 75%    
DLCO (ml/min/mm Hg) 37.22 29.60 89%    
DLCO/VA (ml/min/mm Hg/L) 4.96 4.06 82%    

image

Pulmonary Function Report 5

Pulmonary Measurements Predicted Normal Value Measured Baseline Conditions Percent Predicted Baseline Measured After Bronchodilator Treatment Percent Predicted After Treatment
SVC (L) 3.85 1.93 50%    
FVC (L) 3.85 2.01 52%    
FEV1 (L) 3.01 1.66 55%    
%FEV1/FVC 78% 86%    
FEF25%-75% (L/sec) 3.40 1.85 55%    
PEF (L/sec) 6.50 4.55 70%    
TLC (L) 5.65 3.39 60%    
FRC (L) 3.01 2.11 70%    
RV (L) 1.80 1.35 75%    
DLCO (ml/min/mm Hg) 22.13 13.28 60%    
DLCO/VA (ml/min/mm Hg/L) 3.91 3.60 92%    

image

Share this: