Quality assurance

Published on 04/03/2015 by admin

Filed under Hematology, Oncology and Palliative Medicine

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 2521 times

Chapter 25 Quality assurance

In the haematology laboratory it is essential to ensure that the right test is carried out on the right specimen and that the correct results are delivered to the appropriate recipient without delay. Quality assurance (QA) is defined as the overall programme for achieving these objectives. It must also ensure adequate control of the pre-analytical and post-analytical stages, i.e. from specimen collection to the timely despatch of an informative report. A QA programme should also include standardization of tests and of instrumentation in order to achieve acceptable levels of precision and accuracy. These objectives represent good laboratory practice (GLP); the mechanism for achieving GLP is encompassed in Total Quality Management (TQM) (Table 25.1).

Table 25.1 Definitions used in total quality management (TQM)

1. Quality assurance (QA) Overall programme that ensures that the final results reported are correct. QA includes an adequate control of the pre-analytical and post-analytical stages from specimen collection to the timely dispatch of an informative report. A QA programme also includes standardization of procedures and instrumentation.
2. Quality control (QC) Measures that must be included during each assay run to verify that the test is working properly.
3. Proficiency testing (PT) Procedure to determine the quality of the results generated by the laboratory. Accordingly, PT is a challenge to the effectiveness of the QA and QC programmes and it can be internal or external (also termed ‘quality assessment’).
4. Internal quality control (IQC) Monitoring the haematology test procedures to ensure continual evaluation of the reliability of the daily work of the laboratory with validation of tests before reports are released.
5. External quality assessment (EQA) Evaluation by an outside agency of the between-laboratory and between-method comparability. It can be organized nationally, regionally or internationally.
6. Continuous quality improvement (CQI) Philosophy and attitude for analysing capabilities and processes and improving them repeatedly to achieve the objective of customer satisfaction.
7. Accreditation Certification by a duly recognized authority of the facilities, capability, objectivity, competence and integrity of an agency.
8. Non-compliance A process which does not comply with a quality system.

Standardization

Clinical laboratory errors lead to adverse effects on patient diagnosis, therapy and outcomes, also resulting in the inappropriate use of funds. Standardization plays an important role in patient care because it contributes to a decrease in the number of errors, thus improving the harmonization of procedures and comparability between different laboratories. The introduction of the International Standardization Organization (ISO) standards BS EN ISO 9001 (Quality management systems – requirements), BS EN ISO 17025 (General requirements), BS EN ISO 15189 (Medical laboratories – particular requirements for quality and competence) and BS EN ISO 22870 (Point-of-care testing) places quality at the heart of all activities (see also Table 24.8). In the UK, the focus for laboratory practice has been on Clinical Pathology Accreditation (UK) Ltd (CPA), which has evolved to reflect the way service delivery is perceived within pathology and is now linked to UK Accreditation Service. Quality management systems may also include requirements arising from regulations related to blood and tissues and the Good Manufacturing Practice (GMP) guide, while through the ISO standards, there is a move towards international harmonization.

Standardization of haematology laboratory practice plays a pivotal role in patient care and is essential for certification and accreditation. It should include the following topics:

Standardization of procedures and devices used in the haematology laboratory are the concern of the international professional organizations, especially the International Council for Standardization in Haematology (ICSH). The International Organization of Standardization (ISO) and Comité Européen de Normalization (CEN) have also established standards for medical laboratory practice and for the use of in vitro diagnostic medical devices. At a national level, the British Committee for Standards in Haematology (BCSH) publishes guidelines in books, on websites or as journal articles, and in the USA, a wide range of practice guidelines have been published by the Clinical and Laboratory Standards Institute (CLSI) (formerly the National Committee for Clinical Laboratory Standards, NCCLS). Lists of published documents and catalogues from these various organizations can be found on their various websites (see Table 25.2; see also Tables 24.5 and 24.8).

Table 25.2 Some organizations involved in standardization and quality management systems

Abbreviation Organization Website
AFNOR Association Française de Normalization www.afnor.org
AENOR Asociación Española de Normalización y Certificación www.aenor.es
INSTAND Gesellschaft zur Förderung der Qualitätssicherung in medizinischen Laboratorien www.instandev.de
WHO WHO EQAS for Haematology www.who.int/diagnostics_laboratory/quality/haematology/en/index.html
AMREF African Medical and Research Foundation, Nairobi, Kenya www.amref.org
ANCLSH Asian Network for Clinical Laboratory Standardization and Harmonization www.ancls.org
ASQ American Society of Quality www.asq.org
BCSH British Committee for Standards in Haematology www.bcshguidelines.com
CAP College of American Pathologists www.cap.org
CEN The European Committee for Standardization www.cen.eu
CLSI Clinical and Laboratory Standards Institute www.clsi.org
ICSH International Council for Standardization in Haematology www.islh.org
IRMM Institute of Reference Materials and Measurements Irmm.jrc.ec.europa.eu
ISO International Organization for Standardization www.iso.org
JCAHO Joint Commission for the Accreditation of Healthcare Organizations www.jointcommission.org
PPTC Pacific Paramedical Training Centre New Zealand www.pptc.org.nz
RCPA Royal College of Pathologists Australia www.rcpaqap.com.au
UK NEQAS United Kingdom National External Quality Assessment Scheme www.ukneqas.org.uk
EQALM European Committee of External Quality Assurance Programmes in Laboratory Medicine www.eqalm.org

External quality-assessment schemes have a role in identifying unsatisfactory performance by devices. Some of the terms and definitions generally used in laboratory practice are summarized in Table 25.3.

Table 25.3 Terms relating to quality control in clinical laboratory practice

Precision Can be controlled by replicate tests and by repeated tests on previously measured specimens
Accuracy Can be checked only by the use of reference materials that have been assayed by reference methods
Standardization Encompasses both materials and methods. Material standard or reference preparations are used to calibrate analytical instruments and to assign quantitative values to calibrators. Where possible they must be traceable to defined physical or chemical measurement based on the metrological units of length (metre), mass (kilogram), amount of substance (mole) and time (seconds)
Reference method Is an exactly defined technique that is used in association with a reference preparation, when available, to provide sufficiently precise and accurate data for scientific purposes and for assessing the validity of other methods
Selected method Is one that is directly comparable with and traceable to the international reference method; it serves as an alternative to the reference method when an international reference material is not available; it should be used for evaluation and validation of a proposed routine (working) method
Working (or recommended) method Is intended for use in routine practice, taking account of economy of labour and materials and ease of performance and having been shown by a validation study with a reference method to be sufficiently reliable for its intended purpose
Controls Are preparations that are used for either internal quality control or external quality assessment. Some control preparations have assigned values but they should not be used as standards because the assigned values are usually only approximations and they are often stable for a limited time only
True value This is an ideal concept which in general cannot be achieved
Accepted true value The value approximating the true value, the difference between the two values is negligible
Error The discrepancy between the result of a measurement and the true (or accepted true) value
Random error An error which varies in an unpredictable manner, in magnitude and sign, when a large number of measurements of the same quantity are made under effectively identical conditions
Systematic error An error which, in the course of a number of measurements of the same value of a given quantity, remains constant when measurements are made under the same conditions or varies according to a definite law when conditions change

Control materials and reference standards

The main international authority concerned with material standards (reference preparations) for laboratory medicine is the World Health Organization (WHO). In the European Union, the Institute for Reference Materials and Measurements (IRMM) has established numerous ‘certified reference materials’ for haematology and clinical chemistry. International standards are not intended for routine use but serve as stable standards for assigning values to commercial (or laboratory-produced) ‘secondary standards’ or calibrators.

Control materials are available commercially and they can also be made locally, although there may be technical difficulties in preparing such ‘homemade’ materials. For example, stored plasma may become turbid; chemical or serological analysis may be affected by instability of enzymes; immunological reactions may be interfered with by added preservatives. With the blood count, there are especially difficult problems because of the need to ensure homogeneity in aliquot samples and due to the instability of blood cells, while procedures that enhance the stability of blood samples may also affect the behaviour of the cells. Thus, control material is not strictly analogous to fresh blood. Nonetheless, provided attention is paid to these difficulties, preserved or stabilized blood provides suitable material for internal quality control procedures for haemoglobin concentration, red cell counts, platelet counts and leucocyte counts (see p. 599: Preparation of extended-life material for QA).

Reference Standards

International reference materials relevant to haematology are held at designated institutions (Table 25.4), the relevant websites should be checked for availability of any particular one.

The accessibility of an international reference preparation of haemiglobincyanide (HiCN), first developed by the International Council for Standardization in Haematology (ICSH), has provided improved accuracy of haemoglobin measurement. In some countries, preparations that conform to the international standard are certified by the appropriate national authorities. An important feature of this material is that it is stable for at least several years. A limited quantity of the international standard can be obtained from WHO; a comparable certified reference material is available from IRMM (Table 25.4) and ICSH has recently produced a new preparation with similar specifications.1 Where the use of cyanide reagent for routine haemoglobinometry is prohibited, the haemiglobincyanide standard can still be used to assign a haemoglobin value to a lysate or a whole blood preparation, which is then used as the local secondary standard after appropriate dilution. However, many laboratories are unable to make use of this reference preparation as they no longer have suitable instruments on which to use it. Undiluted lysate is usually stable for up to 6 months, or frozen for several years. Whole blood is stable for about 3 weeks, but for only a few days after dilution. Both whole blood and lysates are useful for quality assurance of haemoglobinometry; whole blood reference samples should be introduced into batches of blood samples and all the samples should be assayed together. This applies to both automated and manual methods.

Table 25.4 Standard materials available internationally

General haematology Erythropoietin, human, urinary
Erythropoietin, recombinant DNA-derived
Ferritin, human, recombinant
Hb A2
Hb F
Folate, whole blood
C-reactive protein
Human serum for immunoassay
WHO/NIBSC
Haemiglobincyanide
Haemoglobin Reference Standard
WHO/NIBSC
BCR/IRMM
ICSH
Immunohaematology Anti-A blood typing serum, anti-B blood typing serum, anti-RhD incomplete blood typing serum, RhD complete blood typing serum, E complete blood typing serum WHO/CLB
Immunology Human serum immunoglobulin
Immunoglobulin G (IgG), A, M, E
Antinuclear factor, homogeneous
Horseradish peroxidase-conjugate sheep antihuman IgG
Human serum complement components C1q, C4, C5, factor B and functional CH50
WHO/NIBSC
Coagulation Ancrod, antithrombin plasma, antithrombin concentrate, factor VIII and von Willebrand factor plasma, II, VII, IX, X, VIII, IXa concentrate, plasma fibrinogen, plasmin, plasminogen activator inhibitor, tissue plasminogen activator recombinant, streptokinase, alpha-thrombin, beta-thromboglobulin, antihuman platelet antigen 1a, platelet factor 4, proteins S and C plasma, urokinase high molecular weight, von Willebrand factor concentrate, von Willebrand factor antigen WHO/NIBSC
Thromboplastin bovine combined, thromboplastin human recombinant, thromboplastin rabbit plain BCR/IRMM
WHO/CLB
WHO/NIBSC National Institute for Biological Standards and Control, South Mimms EN6 3QH, UK. E-mail: standards@nibsc.ac.uk; www.NIBSC/Catalogue list/Reference standards. See also: World Health Organization/biological standards  
BCR/IRMM Institute for Reference Materials and Measurements, Retieseweg B2440, Geel, Belgium. E-mail: bcr.sales@irmm.jrc.be; www.irmm/reference materials  
WHO/CLB Central Laboratory of Netherlands Red Cross Blood Transfusion Service, 125 Plesmanlaan, 1066 AD Amsterdam, Netherlands. Netherlands Blood Transfusion Service/Central Laboratory  

Quality assurance procedures

The procedures that should be included in a quality assurance programme vary with the tests undertaken, the instruments used and (especially if these include a fully automatic counting system) the size of the laboratory and the numbers of specimens handled. Also, the computer facilities available and the amount of time that can be devoted to the quality control assurance must be taken into account. At least some form of internal quality control must be undertaken and there must be participation in an external quality assessment scheme where one is available. Some control procedures should be performed daily and other performance checks should be done at appropriate intervals. The latter is particularly important when there is a change in staff and after maintenance service or repair has been carried out on equipment. A comprehensive protocol is summarized in Table 25.5.

Table 25.5 Schedule for quality control procedures

1. Calibration with reference standards
  Instruments 6-month intervals or more frequently if control chart or EQA indicates bias or fluctuation in results and after any repair/service
  Others: pipettes and scales Calibration of equipment used for manual techniques should not be overlooked; these should be done annually
  Diluting systems Initially and at 1–2-week intervals
2. Control chart with control material
  Daily or more frequently with each batch of specimens
  Duplicate tests on two or three patients’ samples: if control chart or delta check shows discrepancies
3. Analysis of patients’ results
  Daily to check constancy of mean values for MCV, MCH, MCHC: correlation assessment of test report
  Cumulative results: following previous tests and if changes in clinical state
  Blood film examination if unusual test results and/or counter flags appear
4. EQAS performance
  Assessment monthly

During the last 20 years, automation, standardization and technological advances have significantly improved the analytical reliability of laboratory results with a high level of accuracy and decreased error rates in the blood count assay.

Quality design must begin with analytical quality, as it is the essential quality characteristic of any laboratory test. Analytical variations may arise from unsuspected abnormal binding protein(s) in patients, such as heterophile antibodies, anti-animal antibodies and anti-idiotypic antibodies. The exact effect will depend on the site of the interaction with the reaction, leading to falsely raised or lowered measurements.2 More recent data underline the importance of analytical accuracy due to the frequent calibration error that leads to analytical bias affecting the number of patients passing decision thresholds in practice guidelines. The effects of this medically and economically have been demonstrated.3 To ensure reliability in the analytical phase, procedures are required for internal quality control, external quality assessment and standardization. All laboratory staff require training in these various aspects of quality assurance. A useful training manual from WHO* describes the principles and methods, together with practical exercises to illustrate these.4 Another good teaching source is J.O. Westgard’s website: www.westgard.com; this includes a ‘Lesson of the Month’ and other current topics that are regularly updated. The quality control of the analytical stage includes Internal Quality Control (IQC) and External Quality Assessment (EQA).

Internal Quality Control

IQC is based on monitoring the haematology test procedures that are performed in the laboratory and includes measurements on specially prepared materials and repeated measurements on routine specimens, together with daily statistical analysis of the data. IQC is primarily a demonstration of precision. It ensures continual checks that the established reliability of the laboratory’s work does not fluctuate and that reports are validated before they are released. It is based on monitoring the procedures that are actually used for the tests in the laboratory.

IQC includes:

Control Charts

These were first applied in clinical chemistry by Levey and Jennings.5 They are now widely used in haematology for both automated and manual procedures. Samples of the control specimen are included in every batch of patients’ specimens and the results are checked on a control chart. To check precision, it is not necessary to know the exact value of the control specimen. If, however, its value has been determined reliably by a reference method, the same material can also be used to check accuracy or to calibrate an instrument. If possible, controls with high, low and normal values should be used. It is advisable to use at least one control sample per batch, even if the batch is very small, also at set intervals during a large run, at least once for every 50 patient specimens. Because the controls are intended to simulate random sampling, they must be treated exactly like the patients’ specimens. The results obtained with the control samples can be plotted on a chart as described below.

The mean value and standard deviation (SD) of the control specimen should first be established in the laboratory where the tests are performed. Using arithmetic graph paper, a horizontal line is drawn to represent the mean (as a base), and on an appropriate scale of quantity and unit, lines representing +2SD and −2SD are drawn above and below the mean. The results of successive control sample measurements are plotted. If the test is satisfactory, sequential results oscillate about the mean value and <5% of the results fall outside 2SD. Figure 25.1 illustrates a control chart from an automated system; a similar principle can be used for simple methods where the data are plotted manually (Fig. 25.2).

Any of the following indicates a fault in technique or in the instrument or reagent:

The fault may be in the reagents or the laboratory-ware or it may be caused by incorrect adjustment/calibration of the instrument or other equipment, e.g. pipettes, technical error or even clerical error in transcribing the results. Before an intensive investigation, the test should be repeated with another sample and the possibility must also be considered that the inconsistency may be the result of deterioration or infection of the batch of reagent material or insufficient mixing of the sample. This control process is unlikely to detect an error in an individual specimen, which can only be detected by correlation checks. For haemoglobinometry, it may be useful to use both whole blood and lysate in a quality control check because differences in results obtained with these two preparations help to identify errors resulting from incorrect dilution, inadequate mixing or failure of a reagent to bring about complete lysis. If the control specimen is included with each batch of tests during the course of a day, their measurements should not differ by more than the established CV. A trend of sequentially increasing or decreasing values with the repeated measurements is indicative of drift.

Duplicate Tests on Patients’ Specimens

Duplicate tests on patients’ specimens provide another way of checking the precision of routine work.6

MCHC is used to identify any drift of the three indices and used to identify instrument faults; an increased SD signifies loss of precision. To ensure that each batch is representative, the samples should be randomized before analysis and, if possible, within any batch of 20, no more than seven should come from one clinical source or from patients with the same clinical condition. In laboratories still using manual methods, a simple adaptation of the same principle can be applied, confined to MCHC and excluding results from any special clinic that are likely to be specifically biased. From the daily means for all measurements on 10 consecutive working days, an overall daily mean and SD are established. The mean MCHC is then calculated at the end of each day. If the test does not vary by more than ±2SD, it is considered satisfactory but this may be misleading if there is an error in the same direction in both haemoglobin and PCV. The results may be displayed graphically, as illustrated in Figure 25.3. It is useful in validating successive batches of calibrators. The method is now incorporated in many automated blood counters (Fig. 25.4). A similar instrument-specific procedure is used by some manufacturers who gather the data submitted through network links by users of their instruments. This enables them to maintain a constant check of performance of these instruments overall and to detect any that require recalibration or investigation of faults.

Correlation Check

Correlation check implies that any unexpected result of a test must be checked to see whether it can be explained on clinical grounds or whether it correlates with other tests. Thus, for example, unexpectedly higher or lower haemoglobin might be explained, e.g. by a blood transfusion or by a haemorrhage, respectively. A low MCHC should be confirmed by demonstrating hypochromic red cells on a Romanowsky-stained blood film; a high MCV must correlate with macrocytosis. Similarly, the blood films should be examined to confirm marked leucocytosis or leucopenia or thrombocytosis or thrombocytopenia, to distinguish between platelets and red cell fragments or conversely between giant platelets and normal-sized red cells or to check an erroneously high leucocyte count as a result of incompletely lysed red cells in haemoglobinopathies or liver disease. This morphological procedure can only be used if the blood smear is correctly made and stained; otherwise, blood cell morphology may be misleading.

Recording blood count data on cumulative report forms or charts is good clinical practice and provides an inbuilt quality control system by making it easy to detect an aberrant result when compared with a previously determined baseline. This is especially useful in detecting the occasional wild errors caused by incorrect labelling of the specimen, inadequate suspension of the blood before sampling, partial clotting of a blood sample or deterioration on storage. A discrepant result without apparent clinical reason must be suspect until confirmed by a repeat test on a fresh specimen. The occurrence of a contrasting discrepancy in two different specimens on the same day would suggest that two specimens have been mixed up. It is worthwhile emphasizing the importance of the blood film for quality control, especially as it may tend to be considered obsolete with the increasingly automated blood count systems. Any obvious discrepancy between the count obtained by the analyser and subjective impression of morphology should always be checked. A blood film will confirm or refute an abnormally high or low leucocyte or platelet count. An important cause of artefactual leucopenia and thrombocytopenia is partial clotting of the specimen – this may be revealed by the presence, in the film, of fibrin strands with a mass of aggregated platelets. The film will also identify anomalies due to the presence of cold agglutinins, a high leucocyte count due to incomplete lysis of red cells in haemoglobinopathies, etc.

External Quality Assessment

EQA is based on the evaluation by an outside agency of the analytical performance on specially supplied samples by a number of laboratories. The objective is to achieve between-laboratory and between-method comparability and it does not necessarily guarantee accuracy unless the specimens have been assayed by a reference laboratory alongside a reference preparation of known value.

EQA complements internal quality control and is a basic requirement for clinical laboratory certification and accreditation. Certification is a procedure by which a third party gives written assurance that a product, process or service conforms to specific requirements, whereas accreditation is a procedure by which an authoritative body gives formal recognition that a body or person is competent to carry out specific tasks. Accreditation standards related to clinical laboratories place emphasis on having an effective quality assurance system in place, with a commitment to meeting the needs of patients and their doctors as users of laboratory services and a need for continuous cycle of quality improvement at the centre of all policy-making operational decisions (see also Chapter 24, p. 579)

The term EQA, also known as proficiency testing (PT), was adopted in 1979 by a WHO working group on Quality Assurance of Health Laboratories. It is defined as: ‘a system whereby a set of reagents and techniques are assessed by an external source and the results of the testing laboratory are compared with those of an approved reference laboratory or agency’.7 It allows an individual laboratory to compare its performance for one or more tests or techniques against that of other laboratories. Thus, even when all precautions are taken to achieve accuracy and precision in the laboratory, errors arise that are only detectable by objective EQA of the performance of a number of laboratories on material which has been supplied specially for the purpose. EQAS are usually organized, nationally or regionally, as one or PT programmes. National schemes are frequently termed NEQAS (National External Quality Assessment Scheme).

EQA analysis of performance is retrospective and, in addition to inter-laboratory comparability, its main objective is to achieve the harmonization of methods and procedures or educational purposes. EQAS is not necessarily intended to measure accuracy except when the control material is assayed by a reference laboratory alongside a standard reference. The principle is that the same material is sent from a national or regional centre to at least 20 participating laboratories. Results are returned to the EQAS organizer, statistically analysed and a target value with its range of variability (SD) calculated. This provides evaluation of the performance of each individual participant and identifies outliers. It is important that surveys should be performed at regular intervals, although their frequency may vary, depending on the diagnostic importance of the particular test, how frequently they are requested and their technical reliability. It is recommended that at least two specimens should be distributed together for each survey, a minimum of four times each year. Availability of survey material will always influence distribution timings, frequency and quality due to this (e.g. bone marrow films lacking particles).

The main purposes of EQA are to ensure continuous reliable performance by individual laboratories and to achieve harmonization or concordance between laboratories. However, some analysers handle preserved blood differently from routine specimens and, even if correctly calibrated, different types of counter may differ in their responses to EQA samples. It may thus be necessary to analyse results separately for different groups of instruments. When there are unexplained differences in counts on EQA samples with different instruments in the same laboratory, counts should be made on fresh EDTA blood samples with the different instruments to ascertain their true comparability and, if necessary, recalibration of one should be undertaken to achieve concordance. Where appropriate, results from commercial kits and in-house methods should also be analysed in complementary groups. EQAS also have other additional complementary functions:

In summary, EQA is primarily intended to check the technical competence of individual laboratories, but it also provides an overview for assessing the state of the art and for identifying problems that have occurred in instruments, reagents or kits that may be affecting an entire group of users through no fault of their own. This provides a means for verification of manufacturers’ conformity to their claimed specifications and for monitoring product performance in the laboratory, a requirement in Europe in the context of the Directive on In-Vitro Diagnostic Medical Devices.8

More information of this type would be useful for determining the priorities for national or international agencies engaged in developing standards. EQA should be used for highlighting shortcomings and for evaluating methods, technologies and reference/standard preparations. It is important for a clinical laboratory to participate in EQAS for accreditation or certification, as clearly stated in ISO 9000, ISO Guides 25 and 58 and also in ISO 15189.9

Standardization of EQA Schemes

There are many such schemes for haematology based on the different laboratory diagnostic procedures. Many of these schemes are officially promoted or sponsored by national governments or local heath authorities. It is essential that participants in EQAS should have confidence in their efficiency and their effectiveness. ICSH has prepared guidelines for the organization and management of EQAS using proficiency testing. These guidelines are intended to help maintain a meaningful standard in the organization of EQAS and to harmonize the way in which they function. They include the following important principles and technical criteria:

2. Blood counts should be distributed at least monthly, other tests quarterly or more frequently depending on their clinical importance and reliability of analytical methods:

To analyse results on EQA specimens, it is first necessary to establish the target values. These might be ‘truth’ as determined by one or more reference centres or ‘consensus’ from the results of the participants. Referees should use reference methods which are traceable to a primary reference standard and the material should be tested on a minimum of five aliquots, five times on three different occasions and a mean value calculated from all the results. Consensus is based on the use of routine methods. As consensus may be biased by the most commonly used method or instrument group, it might be necessary to establish a different target value for each method or group. Consensus values should only be calculated when the number of results available is sufficient to allow statistically meaningful results. These are more convenient and practical to use than referees. Moreover, because of the absence of absolute metrological standards for most quantitative tests in haematology (haemoglobin concentration is an outstanding exception), the consensus mean or median is, in general, more likely to give a closer approximation of the true value. For qualitative tests, the correct result may be assumed as either that obtained as a consensus by 80% of the participants (as long as there is no clear division of results between methods) or by using results from a ‘Gold Standard’ (e.g. PCR).

Assessment of Participant Performance

Participant’s performance evaluation with EQAS can be carried out by several procedures: quantitative tests, semi-quantitative tests and interpretative tests.

Quantitative Tests

Deviation index

From the results returned by the participants, the median or mean and SD are calculated. The deviation index (DI; also termed ‘z-score’) is used by many EQA schemes for assessing performance in quantitative tests. This is the amount of deviation from the mean (image) or median (m) relative to a unit of 1 standard deviation (SD) (see below). The median is used when there is a non-Gaussian distribution of data with a wide range of results.10 The SD is ‘trimmed’ to exclude any results outside ± 3SD. An individual laboratory can then compare its performance in the survey with that of other laboratories and with its own previous performance from the deviation index (DI) or z-score. This is calculated as the difference between the individual laboratory’s result and the median or mean relative to the SD. Thus,

image

When results have a Gaussian distribution, the mean and SD are adjusted in a preliminary calculation by excluding all results in the highest and lowest 5%.

Instead of using a trimmed SD, as described earlier, the SD may be calculated from a constant or historical CV, which takes account of technical variance of the method, clinical utility of the test and the critical range of measurement for diagnostic discrimination in this method. A DI score of <0.5 denotes excellent performance; a score between 0.5 and 1.0 is satisfactory; and one between 1.0 and 2.0 is still acceptable. However, a score >2.0 suggests that the analyser calibration should be checked, whereas a DI >3.0 indicates a serious defect requiring urgent attention.

The DI provides a simple method for judging performance in a survey and it also indicates whether there have been changes in sequential surveys, thus distinguishing between casual errors and persistent unsatisfactory performance. A limitation of DI is that it is purely statistical. As the state of the art improves, some blood count parameters will have a CV of only 1–2%. Thus, the DI will indicate poor performance with unrealistically small deviations from the median. It may be better to use clinical relevance when determining the acceptable limits of percentage deviation from the target value. There are some differences in these limits as established in the USA for CLIA ′88 requirements, those proposed by ICSH and those by an ISO panel.911

Semi-Quantitative Tests

Reactions such as lysis, agglutination or colour change are recorded as 0, 1+, 2+, 3+ or 4+. Assessment of performance should be based on extent of divergence from the target value, which might be a consensus of participant results or a referee’s results. Account must be taken of the diagnostic and clinical significance of an incorrect or confusing result.

EQA schemes should assess not only technical reliability but also professional competence in interpretation of the measurements. Thus, participants should be required to report on the technical significance of their results (i.e. whether within normal reference values for the specified method) and also on the clinical significance, taking account of any clinical information provided. Incorrect interpretation of a correct quantitative result is often due to lack of understanding of the concept of reference values and to the use of inappropriate reference ranges. It is important in the post-analytical phase of quality management for each laboratory to establish its own reference values for normal and for specific groups (e.g. smokers, pregnancy). With qualitative tests, too, account must be taken of the clinical significance of the answer. For each test it is necessary to decide if it is as serious to err by reporting a feature which does not exist as to miss an abnormality which is present and to penalize the score accordingly. Scoring may be either as a credit for correct answers or as a penalty for an incorrect answer. Thus, using the penalty method in G6PD screening, a correct answer scores zero, while an error will be graded depending on the implications of misclassification for the subsequent management of the patient: it is more serious to report a low level as normal (score 5) than a normal value as low (score 3), and when the true value is intermediate, it is more serious to report it as normal in a female (score 3) than in a male (score 1).

Interpretive Tests

In assessing qualitative or interpretive tests (e.g. stained blood smears for morphology), participant results are compared with the consensus obtained from a panel of referees or by concordance of at least 80% of the participants. Some schemes may use a panel of experts to produce a ‘Comments’ section in the report. The features reported are graded on their clinical significance, taking account of the specific medical condition, as follows:

All correct observations are given a positive score and false-positive observations are subtracted in accordance with the grading. The result is expressed as a percentage of the total possible score established by the referees. The more stringent limits have been proposed to provide greater sensitivity in diagnostic discrimination, but they may be impractical, resulting in an overestimate of the numbers of poor performers.12,13

Similarly, scores on parasite films will depend on clinical significance; a negative result on a film containing Plasmodium falciparum or a microfilaria, will be penalized more than a Plasmodium ovale, reported as a Plasmodium malariae.

While few EQA schemes provide a bone marrow slide on a regular basis, due to the difficulties of producing large numbers of films from the same donor, the introduction of digital morphology EQAS may be useful in the future. Preparation and reporting on these films should be standardized from conception using guidelines.14

Internal audit for total quality management

Internal audit provides evidence that the quality management system (QMS) has been effectively established, implemented and maintained. Systems of internal audit should be established and audits conducted according to a schedule and against agreed criteria. Audits are performed by individuals with appropriate training and, if possible, by those who are independent of the work being audited. A record of internal audit activities is logged within the audit document, which includes details of any non-compliances, corrective action(s) and/or preventative action(s). The results of internal audit should be regularly evaluated and any decisions taken to alter/improve the QMS should be documented and should be subject to monitoring and review.

Preparation of extended-life material for use in quality assessment

Commercial products are available but, with appropriate expertise, control materials can also be made locally. All such preparations will require to have in-house values assigned for the relevant parameters.

Preparation of Haemolysate

Method

Stored at 4°C, the product should retain its assigned value for at least several months or for 1–2 years if kept at −20°C.

Preparation of Stabilized Whole Blood Control Material

Preparation of Stable Control Material for EQAS

Preparation of QC Material for Platelet Count

References

1 Davis B.H., Jungerius B. International Council for Standardization in Haematology technical report 1–2009: new reference material for haemiglobincyanide for use in standardization of blood haemoglobin measurements. International Journal of Laboratory Hematology. 2010;32:139-141.

2 Kazmierczak S.C., Catrou P.G. Analytical interference. More than just a laboratory problem. Am J Clin Pathol. 2000;113:9-11.

3 Klee G.G., Schryver P.G., Kisabeth R.M. Analytical bias specifications based on the analysis of effects on performance of medical guidelines. Scand J Clin Lab Invest. 1999;59:509-512.

4 Lewis S.M. Quality Assurance in Haematology Document LAB/98.4. Geneva: World Health Organization; 1998.

5 Levey S., Jennings E.R. The use of control charts in the clinical laboratory. Am J Clin Pathol. 1950;20:1059-1066.

6 Cembrowski G.S., Lunetsky E.S., Patrick C.C., et al. An optimized quality control procedure for hematology analyzers with the use of retained patient specimens. Am J Clin Pathol. 1988;89:203-210.

7 World Health Organization. EURO report 36: External quality assessment of health laboratories. Copenhagen: WHO Regional Office for Europe; 1981.

8 Directive 98/79/EC of the European Parliament and of the Council on in vitro Diagnostic Medical Devices, October 1998.

9 International Standard ISO 15189 2007 Medical Laboratories. Particular Requirements for Quality and Competence

10 International Council for Standardization in Haematology. Guidelines for organization and management of external quality assessment using proficiency testing. International Journal of Haematology. 1998;68:45-52.

11 International Organization for Standardization. Guide 43–41 proficiency testing by interlaboratory comparisons. In Development and Operation of Proficiency Testing Schemes. Geneva: ISO; 1997.

12 Vives Corrons J.L., Albarede S., Flandrin G., et al. Guidelines for blood smear preparation and staining procedure for setting up an external quality assessment scheme for blood smear interpretation. Part I: Control Material. Clin Chem Lab Med. 2004;42(8):922-926.

13 Vives Corrons J.L., Blerk M Albarede S., et al. Guidelines for setting up an External Quality Assessment Scheme for blood smear interpretation. Part II: Survey Preparation, Statistical Evaluation and Reporting. Clin Chem Lab Med. 2006;44(8):1039-1043.

14 Lee H., Erbert W.N., Porvit A., et al. ICSH guidelines for standardization of bone marrow specimens and reports: International Council for Standardization in Haematology. International Journal of Laboratory Hematology. 2008;30:349-363.

15 Deom A., El Aouad R., Heick C.C., et al. Requirements and Guidance for External Quality Assurance Programmes for Health Laboratories: Document LAB/2000. Geneva: World Health Organization; 2000.

16 Reardon D.M., Mack D., Warner B., et al. A whole blood control for blood count analysers and source material for an external quality assessment scheme. Med Lab Sci. 1991;48:19-26.

17 Zhang Z., Tatsumi N., Tsuda I., et al. Long-term preservation of platelet count in blood for external quality control surveillance using prostaglandin E1. Clin Lab Haematol. 1999;21:71.