Pacemakers and defibrillators

Published on 27/02/2015 by admin

Filed under Anesthesiology

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 1995 times

Chapter 25 Pacemakers and defibrillators

Basic cardiac electrophysiology

While pacemakers and defibrillators are not strictly anaesthetic equipment, they are frequently encountered both in the elective and the emergency situation. An understanding of their hardware and software is important, not least because of their interaction with the strong electromagnetic fields that may be encountered in anaesthetic practice.

To understand pacemaker and defibrillator functioning, an understanding of normal cardiac electrophysiology is necessary. The myocardium consists of an interconnected network of myocytes. At rest, the myocyte interior is maintained at a negative potential (−80 mV) in relation to the extracellular fluid. This is due to the relative impermeability of the plasma membrane, differential intra- and extracellular ion concentrations and active ion transporters. Either because of the spontaneous inward leak of positively charged ions, in the sinus node for example, or because of an external electrical stimulus, the potential difference across the membrane decreases. At a threshold potential (approximately −70 mV in ventricular myocytes) various ion channels become activated producing further ion influx. This leads to myocyte depolarization and the action potential (Fig. 25.1). Calcium influx leads to the release of additional intracellular calcium stores activating the contractile apparatus, a process described as excitation-contraction coupling.

Further ion movements, particularly potassium efflux, restore the myocyte’s potential to its resting value. During the early part of the action potential the myocyte cannot be induced to depolarize again – the absolute refractory period. Later on, a stimulus of sufficient strength can induce further depolarization – the relative refractory period.

Myocytes in different areas of the heart have different ion channels and, therefore, different action potentials. Some cells spontaneously discharge due to a resting inward leak of positively charged ions, with some cells, for example in the sinus node, discharging at a faster rate than others; this is known as automaticity. Due to the inter-linked nature of myocytes, the cells with the fastest rate of depolarization set the heart rate.

Pacemakers

The NASPE/BPEG code

With the increasing complexity of pacemakers, a three letter code was designed in 1974 by the Inter-Society Commission for Heart Disease Resources to define a pacemaker’s characteristics. This was refined in 1981 and subsequently led to the 1987 North American Society of Pacing and Electrophysiology (NASPE)/British Pacing and Electrophysiology Group (BPEG) five letter generic code.

The first letter of this code denotes the chamber(s) that are paced, the second letter the chamber(s) that are sensed, the third letter the response to sensing and the fourth letter programmability and rate response. The fifth letter, rarely used, denotes anti-tachycardia functions. NASPE/BPEG defibrillator (1993) and lead (1996) codes have also been designed.

In general terms, five forms of pacing are likely to be encountered clinically, the first four being available for both temporary and permanent pacing:

1. VVI. A single lead is placed in the ventricle, for pacing (VVI) and sensing (VVI). If the ventricular rate is above the minimum set (the base rate), pacing is inhibited (VVI). Otherwise pacing occurs.

2. AAI. A single lead is placed in the atrium. It functions in a similar way to the VVI and is used for sick sinus syndrome where AV nodal conduction is normal.

3. DDD. Leads are present in the atrium and ventricle. Both chambers (hence dual) can be paced (DDD) and sensed (DDD). If activity in either chamber is sensed a pacing stimulus will be inhibited. If no ventricular activity follows atrial activity, ventricular pacing will be triggered (DDD representing the dual responses of inhibition or triggering).

4. VOO/AOO/DOO. These are the simplest pacing modes. No sensing occurs (_O_) and therefore there can be no response to sensing (__O); the chamber(s) is paced at a fixed rate. This mode is used temporarily where the sensing of external noise could lead to the inappropriate inhibition of pacing, e.g. diathermy. It is also known as asynchronous pacing.

5. VDD. Only one lead is used which paces the ventricle (VDD). An electrode further up the lead is positioned in the atrium allowing dual chamber sensing (VDD). This allows physiological pacing, i.e. ventricular activity follows atrial at a physiological PR interval as long as atrial pacing is unnecessary. This mode is uncommon and is used as an alternative to DDD pacing in order to reduce system complexity.

6. The presence of an (R) as the fourth letter denotes a pacemaker able to increase the paced heart rate in response to the patient’s activity (see below, Software).

Pacing terminology

A number of pacing terms require explanation:

• Threshold (Fig. 25.2) – the minimum stimulus needed to capture the heart. It is given as current (amperes) or voltage (volts) over a time period (pulse width, ms).

• Unipolar/bipolar – refers to the arrangement (Fig. 25.3) of the negatively charged cathode and the positively charged anode. In the bipolar circuit, the cathode and the anode are separated by a short distance at the pacing lead tip. The circuit is localized to the immediate myocardium. In the unipolar circuit, the cathode is at the pacing lead tip with the anode at a distance; in the permanent pacing system it is found in the pacing box. The unipolar pacing circuit is larger and, therefore, produces a larger pacing artefact on the ECG. While there is little difference between the pacing characteristics of these two conformations, bipolar sensing is less susceptible to external noise.

• Impedance – this represents the total resistance to current flow in the pacing circuit. This resistance occurs in the leads, at the lead–myocardial interface for endocardial leads and in the tissue between the cathode and the anode. Fractures in the leads increase impedance, while insulation breaks reduce it.

Temporary pacing

Temporary pacing is used in emergency situations of life-threatening bradycardia or where a bradyarrhythmia could occur temporarily.

Transvenous pacing

This is the usual method for temporary pacing. A thin, semi-flexible, shaped, bipolar pacing lead is normally used. The central circulation is accessed via the internal jugular, subclavian or femoral veins, the first of these being preferable as there is a lower risk of complication (e.g. pneumothorax with the subclavian route, infection and deep vein thrombosis with the femoral). Furthermore, with the internal jugular approach, the veins used for permanent pacing are not directly punctured. The lead is manipulated into position at the right ventricular apex under X-ray guidance. Some leads may have a central shapeable stylet to allow more accurate positioning. Others have a tip-mounted flotation balloon to allow non-radiographic positioning. Active fixation (screw tipped) leads are also available to prevent lead displacement. In most cases one ventricular lead is used. Atrial pacing and dual chamber pacing are also possible. For temporary transvenous pacing, due consideration must be given to the possible need for permanent pacemaker insertion and the impact thereon of infection and local complications at vascular access. It is hence advisable to avoid the subclavian route and to use ultrasound guidance for vascular puncture. Various guidelines are listed in the bibliography section at the end of this chapter.

The pacing lead is connected to an external box which allows various programming options:

Transoesophageal and transgastric pacing

An electrode (on a device similar to an oesophageal stethoscope or nasogastric tube),1 placed in the oesophagus, can capture the left atrium, while in the stomach ventricular pacing may be possible. Transoesophageal atrial pacing (TAP) is a simple and safe method for temporary treatment of bradyarrhythmias and is particularly applicable to use during anaesthesia.2 Because voltages of up to 20V or more may be required to achieve capture (Fig. 25.2), a signal amplifier is needed if using an ordinary external (transvenous) pacing box (signal generator). TAP, like transgastric pacing, is rarely used because of unfamiliarity with the technique and scarcity of equipment.

Transcutaneous pacing

In the peri-arrest situation it may be difficult to position a transvenous temporary wire. It is possible to capture the ventricle by passing a large enough current through the chest using specially designed electrodes (Fig. 25.4). These electrodes may also be used for monitoring and defibrillation.

There is significantly greater impedance in the transcutaneous pacing circuit than with endocardial stimulation, due to the additional significant impedances of the electrode–skin interface, the lung and the pericardium. High currents are required, usually 50–90 mA, at pulse widths of 10–20 ms. Pacing thresholds may be further increased by poor electrode–skin contact, metabolic disturbance (hypoxia, acidosis) and pericardial effusion. Unpleasant cutaneous nerve and skeletal muscle stimulation occur at currents as low as 10 mA making this method of pacing, without sedation, short-lived by necessity. Pulse widths of the order of 10 ms produce optimal pacing thresholds and reduce patient discomfort. Significant pacing artefacts occur on the ECG, making myocardial capture difficult to assess without pulse or blood pressure monitoring.

Another form of transcutaneous pacing is also used by cardiothoracic surgeons. One or more wires are implanted directly into the atrial and/or ventricular myocardium at the time of operation and brought out through the skin. Two wires can be used per chamber to allow bipolar pacing, or a second wire may be stitched into the praecordial skin as and when pacing is actually required. These can be connected to a temporary pacing box. Similar to temporary wires they are prone to infection and thresholds can rise without warning. They are usually used to await recovery of the normal conducting system. Atrial pacing may also reduce the incidence of postoperative atrial fibrillation.

Permanent pacing

The principles of permanent pacing remain the same as temporary pacing; the indications continue to expand and have been recently updated. Following significant improvements in battery and lead technology, devices are now small enough to be placed subcutaneously in the pre-pectoral region, the leads passed via the subclavian or cephalic veins. Other routes including the femoral veins and epicardial systems may also be used where subclavian access is impossible or due to previous pacemaker infection.

Hardware

The pacemaker box consists of:

A number of different power sources have been used, including nuclear devices. However, the lithium-iodine battery is currently the industry standard.

In contrast to temporary leads, permanent pacing leads are expensive, highly engineered multi-component devices, required to be in situ for several decades, without failing or producing significant local damage. Their components include:

1. Electrode. Most newly implanted leads are bipolar with a cathodal tip and a larger anode further back. The cathode is surprisingly complex – it is small, producing: a high charge-density, thus improving myocardial capture; and a high pacing impedance which reduces battery drain. Platinum, titanium or activated carbon is used for their good conducting properties and durability. A small reservoir of steroid is also found in the lead tip to reduce the inflammatory reaction at the lead–myocardial interface, maintaining lower stimulation thresholds.

2. Fixation mechanism. It is necessary to secure the lead in the myocardium thus maintaining capture. Passive fixation is achieved with small flexible protrusions from the lead tip, tines, which hook into the trabeculated muscle. For less stable positions active fixation mechanisms are available, often a small retractable screw (Fig. 25.5D).

3. Lead conductor. This is the wire, often made of a complex alloy, which transmits the electrical current. Bipolar leads require two, one each for the cathode and the anode. Coiling the wires produces great lead flexibility, the anodal conductor being wound inside the cathode, separated by an insulator (coaxial wire). A recent advance involves coating each individual conductor with insulation (coated wire technology), which allows the anodal and cathodal conductors to run together producing leads with even smaller diameters.

4. Lead insulation. Silicon was used initially; polyurethane is used today; it allows flexibility, resistance to damage, biocompatibility and good handling characteristics.

5. Lead connector. The metal connectors which link the conductor to the pacemaker. There is now an industry standard, IS-1.

6. Central lumen. A central lumen runs the length of the lead allowing the passage of a stylet that can be shaped to allow accurate positioning. A number of leads are also shaped to allow positioning in certain positions, e.g. the coronary sinus or the right atrial appendage.

Software

Pacemakers are becoming increasingly complex with multiple programmable parameters and functions, including:

The evolving complexity of the programmable parameters (the most important of which are summarized above) has made it increasingly difficult to diagnose pacemaker malfunction from the surface ECG without knowledge of the pacemaker set-up. In general terms, ‘pacemaker malfunction’ due to software or hardware failure is a rare event.

Vagal nerve stimulators

Another application of pacemaker technology that is becoming increasingly prevalent is the implantable vagal nerve stimulator (VNS). This is currently approved as adjunctive therapy for medically refractive epilepsy and major depression. Possible indications in the future may eventually include obesity, chronic pain syndromes and various neuropsychiatric disorders, such as obsessive compulsive and panic disorders.

The system consists of a constant current pulse generator placed subcutaneously, as with cardiac pacemakers, in the pectoral region and connected onto the left vagus with a tunnelled lead culminating in spiral platinum electrodes embedded in silicone rubber (Fig. 25.6). The left nerve is used as the right vagus nerve carries a higher proportion of cardiac efferents. Again, as with cardiac pacemakers, subsequent programming is via an external computer with radiofrequency signals. Stimulation patterns are episodic (e.g. up to 90 s on and 5–10 min off) and patients carry a magnet whose transient application results in an additional pre-programmed burst of stimulation which may abate or prevent a seizure. The magnet can be held or fixed over the signal generator to pause stimulation, in order to limit some of the problems below.

Patients usually also carry a copy of the manufacturer’s instructions for reference.

In addition to the usual issues regarding electromagnetic interference (see later), there are a number of specific problems that are noteworthy for anaesthetists. They are mostly related to ‘on time’ of the device, when stimulation is taking place:

Defibrillators

Introduction

To understand defibrillator design and function and why defibrillation may fail requires some understanding of the pathophysiology of fibrillation and defibrillation.

Ventricular fibrillation (VF) is a complex arrhythmia consisting of random, disorganized, three-dimensional waves of depolarization, thus producing the loss of cardiac output and allowing arrhythmia persistence.

The mechanisms through which a shock of sufficient strength allows the return of spontaneous, co-ordinated electrical activity remain incompletely understood. A shock can have three effects depending on its timing in the action potential. Early on, during the absolute refractory period, no effect occurs. Later on, action potential prolongation is seen. Later still, a new action potential is induced. It is hypothesized that a shock of sufficient strength and appropriate timing will extend the refractory period in enough of the myocardium to allow the waves of depolarization to die out. Subsequently, cardiac automaticity allows the return of normal electrical activity, depending upon the underlying cardiac condition.

The amount of current required to produce defibrillation is known as the defibrillation threshold (DFT). For historical reasons, this is given in terms of energy (joules). Its determinants explain the success or failure of a shock. As with the pacing threshold, there is a minimum current below which defibrillation will not occur. Above a certain level, detrimental effects may occur, reducing the likelihood of success. Between these values success is most likely. However, as the waves of depolarization in VF are entirely random, an element of chance exists that the shock is delivered at the optimal time to defibrillate a critical mass of myocardium. This concept is known as the probabilistic nature of DFTs.

Factors affecting DFTs include:

1. Charge characteristics. A monophasic shock is one where the polarity of the shock remains constant throughout its delivery (Fig. 25.7A). In biphasic shocks (Fig. 25.7B) the polarity is reversed during delivery. In general terms, DFTs are lower and there is less post-shock myocardial depression with biphasic shocks. The shape and time-course of the shock can also be varied to produce optimal effects.

2. Electrode position. In implantable systems shocks can occur in a number of configurations that may affect the DFT (see Hardware, below). Altering the position of the endocardial coils can also have some effect. External paddle positions may also have an effect.

3. Shock polarity. Which electrode is used as the anode and which the cathode can affect DFTs, particularly with monophasic shocks.

4. Underlying cardiac condition. The more severe the cardiac condition, as assessed using heart size, ejection fraction, QRS width or heart failure symptoms, the higher the DFT. Furthermore, the time spent in VF prior to defibrillation adversely affects DFTs.

5. Metabolic disturbance. DFTs are higher in hypoxic and acidotic patients.

6. Medication. DFTs can be affected by numerous medications. Of note, intravenous amiodarone reduces DFTs, whereas its chronic administration increases them. Fentanyl reduces DFTs. Common inhalational anaesthetic agents do not appear to have significant effects.

7. Shock impedance. Shock impedance affects current delivery to the myocardium. The transthoracic impedance for external defibrillation is dependent on lung volume, skin contact and tissue thickness.

The implantable cardioverter defibrillator

Implantable cardioverter defibrillators (ICDs) have been commercially available since 1985, enabling those at risk of recurrent life-threatening ventricular tachyarrhythmia to experience an improved quality of life. Over recent years technological improvements have allowed improved defibrillation success, easier implantation, greater longevity and increased programmability. Initially the devices were large and required abdominal placement and epicardial patches. In 2011, devices as small as 33 cm3 are available and can be implanted in the same manner as pacemakers.

Hardware

The ICD system (Fig. 25.5) consists of a box containing:

However, it is also required to produce a large current over a short period to produce the shock, something the lithium-iodine battery is incapable of. The lithium silver vanadium oxide battery is used instead. Of note, because of its differing electrochemistry, ICD batteries are much less efficient for pacing, a feature reducing their longevity.

Defibrillator leads are constructed in a very similar way to pacing leads with some added complexity:

Electromagnetic interference

Pacemakers are required to detect low amplitude electrical signals and communicate with pacemaker programmers with radiofrequency transmissions. They are, therefore, susceptible to interference from external sources of electromagnetic radiation. As pacemaker and ICD design has progressed and with the increased use of bipolar leads, electromagnetic interference is now less of an issue.

Frequencies of between 0 and 1011 Hz, representing radiofrequency and microwave energies, can affect pacing systems. Pacemakers respond to electromagnetic interference in different ways. If the device recognizes the signal as noise it can react by turning off its sensing circuits during the noise (VOO/DOO), resetting permanently to a default mode (often VVI) or by triggering a paced beat to prevent noise inhibiting its output. Clearly if the signals are not recognized as noise, pacemaker inhibition, reprogramming or ICD shock therapy may be triggered.

There are many sources of electromagnetic interference which can affect pacemakers and ICDs. However, those that are likely to influence anaesthetic practice include the following:

1. Diathermy – this can affect pacemakers in a number of ways, as discussed above. In addition, the diathermy current can activate the rate-response circuitry, damage the pacing circuits or pass via the lead damaging the lead-myocardial interface thus affecting pacing thresholds. Bipolar pacemakers are much less susceptible to this. Suggested management includes:

2. Cardioversion/defibrillation – like diathermy, the considerable current used can overwhelm the protective circuitry and reprogramme or damage the device or the lead–myocardial interface. Management includes device interrogation pre- and post-procedure. Paddles should be placed at least 10 cm from the box and at 90° to the axis of box to lead tip (preferably AP).

3. MRI – the scanner can interact in a number of ways with pacing systems, not least because of the ferromagnetic components of old devices. Strong electrical signals are also produced that can either inhibit or induce rapid pacing. In general terms, an MRI is contraindicated in pacemaker/ICD patients, although case reports exist of patients undergoing this investigation without ill-effect. There is now an MRI-safe pacemaker available from Medtronic with the other pacing companies soon to follow suit. The model is safe to use in an MRI scanner six weeks after implant. MRI-safe ICDs are also likely to come at some point.

4. Lithotripsy – the shock can trigger an arrhythmia, cause mode switch or pacemaker inhibition or damage the box and, in particular, the piezoelectric crystals used for activity sensing. The manufacturer should be consulted particularly for abdominally placed systems, devices should be programmed to an asynchronous mode and ICDs should be disabled.

5. Radiotherapy – if directed at the pacing box this may destroy the circuitry. Pacing systems require repositioning.

6. Electroconvulsive therapy – adverse effects are unlikely, but pacemaker checks and asynchronous pacing are suggested.

In general terms, consultation in advance with the pacing technicians for up-to-date advice is strongly recommended. Magnet placement may not be sufficient and may even be detrimental.