Monitoring of the Heart and Vascular System

Published on 07/02/2015 by admin

Filed under Anesthesiology

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 2742 times

Chapter 9 Monitoring of the Heart and Vascular System

HEMODYNAMIC MONITORING

For patients with severe cardiovascular disease and those undergoing surgery associated with rapid hemodynamic changes, adequate hemodynamic monitoring should be available at all times. With the ability to measure and record almost all vital physiologic parameters, the development of acute hemodynamic changes may be observed and corrective action may be taken in an attempt to correct adverse hemodynamics and improve outcome. Although outcome changes are difficult to prove, it is a reasonable assumption that appropriate hemodynamic monitoring should reduce the incidence of major cardiovascular complications. This is based on the presumption that the data obtained from these monitors are interpreted correctly and that therapeutic decisions are implemented in a timely fashion.

Many devices are available to monitor the cardiovascular system. These devices range from those that are completely noninvasive, such as the blood pressure (BP) cuff and ECG, to those that are extremely invasive, such as the pulmonary artery (PA) catheter. To make the best use of invasive monitors, the potential benefits to be gained from the information must outweigh the potential complications. In many critically ill patients, the benefit obtained does seem to outweigh the risks, which explains the widespread use of invasive monitoring. Transesophageal echocardiography (TEE), a minimally invasive technology, has gained in popularity as an alternative to the PA catheter and is considered a standard-of-care monitoring device in the perioperative management for certain procedures, such as mitral valvuloplasty or surgery for congenital heart defects.

Standard monitoring for cardiac surgical patients includes BP, ECG, central venous pressure (CVP), urine output, temperature, capnometry, pulse oximetry, and intermittent arterial blood gas analysis (Box 9-1). The next tier of monitoring includes PA catheters, left atrial pressure (LAP) catheters, thermodilution cardiac output (CO) measurements, TEE, and indices of tissue oxygen transport (Box 9-2). All of these measurements and their derivatives can be obtained and recorded. The interpretation of these complex data, however, requires an astute clinician who is aware of the patient’s overall condition and the limitations of the monitors.1

ARTERIAL PRESSURE MONITORING

Blood pressure monitoring is the most commonly used method of assessing the cardiovascular system. The magnitude of the BP is directly related to the CO and the systemic vascular resistance (SVR). This is analogous to Ohm’s law of electricity (voltage = current × resistance), in which BP is analogous to voltage, CO to flow, and SVR to resistance. An increase in the BP may reflect an increase in CO or SVR, or both. Although BP is one of the easiest cardiovascular variables to measure, it gives only indirect information about the patient’s cardiovascular status.

Mean arterial pressure (MAP) is probably the most useful parameter to measure in assessing organ perfusion, except for the heart, in which the diastolic BP is the most important. MAP is measured directly by integrating the arterial waveform tracing over time, or using the formula:

image

or

image

where SBP is systolic blood pressure and DBP is diastolic blood pressure. The pulse pressure is the difference between SBP and DBP.

Anesthesia for cardiac surgery is frequently complicated by rapid and sudden lability of the BP because of several factors, including direct compression of the heart, impaired venous return due to retraction and cannulation of the venae cavae and aorta, arrhythmias from mechanical stimulation of the heart, and manipulations that may impair right ventricular (RV) outflow and pulmonary venous return. Sudden losses of significant amounts of blood may induce hypovolemia at almost any time. The cardiac surgical population also includes many patients with labile hypertension and atherosclerotic heart disease. A safe and reliable method of measuring acute changes in the BP is required during cardiac surgery with cardiopulmonary bypass (CPB).

Continuous BP monitoring with noninvasive devices have not proven to be suitable for cardiac surgery. Intra-arterial monitoring provides a continuous, beat-to-beat indication of the arterial pressure and waveform, and having an indwelling arterial catheter enables frequent sampling of arterial blood for laboratory analyses. Direct intra-arterial monitoring remains the gold standard for cardiac surgical procedures.

Arterial Cannulation Sites

Factors that influence the site of arterial cannulation include the location of surgery, the possible compromise of arterial flow due to patient positioning or surgical manipulations, and any history of ischemia of or prior surgery on the limb to be cannulated. Another factor that may influence the cannulation site is the presence of a proximal arterial cutdown. The proximal cutdown may cause damped waveforms or falsely low BP readings due to stenosis or vascular thrombosis.

The radial artery is the most commonly used artery for continuous BP monitoring because it is easy to cannulate with a short (20-gauge) catheter. It is readily accessible during surgery, and the collateral circulation is usually adequate and easy to check. It is advisable to assess the adequacy of the collateral circulation and the absence of proximal obstructions before cannulating the radial artery for monitoring purposes.

The ulnar artery provides most blood flow to the hand in about 90% of patients. The radial and ulnar arteries are connected by a palmar arch, which provides collateral flow to the hand in the event of radial artery occlusion. It has been shown that if there is adequate ulnar collateral flow, circulatory perfusion pressure to the fingers is adequate after radial arterial catheterization. Many clinicians routinely perform Allen’s test before radial artery cannulation to assess the adequacy of collateral circulation to the hand.

Allen’s test is performed by compressing the radial and ulnar arteries and by exercising the hand until it is pale. The ulnar artery is then released (with the hand open loosely), and the time until the hand regains its normal color is noted. With a normal collateral circulation, the color returns to the hand in about 5 seconds. If, however, the hand takes longer than 15 seconds to return to its normal color, cannulation of the radial artery on that side is controversial. The hand may remain pale if the fingers are hyperextended or widely spread apart, even in the presence of a normal collateral circulation. Variations on Allen’s test include using a Doppler probe or pulse oximeter to document collateral flow. If Allen’s test demonstrates that the hand depends on the radial artery for adequate filling, and other cannulation sites are not available, the ulnar artery may be selected.2

Chest wall retractors, such as the Favaloro retractor, may impede radial arterial pressure monitoring in cardiothoracic procedures in some patients. The arm on the affected side may have diminished perfusion during extreme retraction of the chest wall. If the left internal mammary artery is used during myocardial revascularization, the right radial artery could be monitored to avoid this problem. Alternatively, a noninvasive BP cuff on the right side could be used to confirm the accuracy of the radial artery tracing during extreme chest wall retraction.

Monitoring of the radial artery distal to a brachial arterial cutdown site is not recommended. Acute thrombosis or residual stenosis of the brachial artery will lead to falsely low radial arterial pressure readings. Other considerations related to the choice of a radial arterial monitoring site include prior surgery of the hand, selection of the nondominant hand, and the preference of the surgeon, the anesthesiologist, or both.

The brachial artery lies medial to the bicipital tendon in the antecubital fossa, in close proximity to the median nerve. Brachial artery pressure tracings resemble those in the femoral artery, with less systolic augmentation than radial artery tracings. Brachial arterial pressures were found to more accurately reflect central aortic pressures than radial arterial pressures before and after CPB. The complications from percutaneous brachial artery catheter monitoring are lower than those after brachial artery cutdown for cardiac catheterization.3 A few series of perioperative brachial arterial monitoring have documented the relative safety of this technique.

The femoral artery may be cannulated for monitoring purposes but is usually reserved for situations in which other sites are unable to be cannulated or it is specifically indicated (e.g., descending thoracic aortic aneurysm surgery for distal pressure monitoring). Peripheral artery cannulation for hemodynamic monitoring, including 3899 femoral artery cannulations, has been studied. Temporary occlusion was found in 10 patients (1.45%), whereas serious ischemic complications requiring extremity amputation were reported in 3 patients (0.18%). Other complications were pseudoaneurysm formation in 6 patients (0.3%), sepsis in 13 patients (0.44%), local infection (0.78%), bleeding (1.58%), and hematoma (6.1%). The femoral artery for hemodynamic monitoring purposes was as safe as radial artery cannulation.

In patients undergoing thoracic aortic surgery, distal aortic perfusion (using partial CPB, left-heart bypass, or a heparinized shunt) may be performed during aortic cross-clamping to preserve spinal cord and visceral organ blood flow. In these situations, it is useful to measure the distal aortic pressure at the femoral artery (or, dorsalis pedis or posterior tibial artery) to optimize the distal perfusion pressure. In repairs of aortic coarctation, simultaneous femoral and radial arterial monitoring may help determine the adequacy of the surgical repair by documenting the pressure gradient after the repair. It is necessary to consult with the surgeon before cannulating the femoral vessels because these vessels may be used for extracorporeal perfusion or placement of an intra-aortic balloon pump during the surgical procedure.

Insertion Techniques

Direct Cannulation

Proper technique is helpful in obtaining a high degree of success in arterial catheterization. The wrist should be placed in a dorsiflexed position on an armboard and immobilized in a supinated position. It is helpful to draw the course of the artery for 1 inch and to be comfortably seated. Doppler devices and ultrasonic vessel finders may also be of value. Local anesthetic is injected intradermally over the artery, and a small skin nick may be made to allow passage of the catheter-over-needle assembly into the subcutaneous tissue without crimping secondary to penetration of the unit through the skin. A 20-gauge or smaller, 3- to 5-cm, nontapered Teflon catheter over needle is used, without a syringe attached, to make the puncture. If a syringe is used, the plunger should be removed to allow free flow of blood to detect when the artery has been punctured. The angle between the needle and the skin should be shallow (30 degrees or less), and the needle should be advanced parallel to the course of the artery. When the artery is entered, the angle between the needle and skin is reduced to 10 degrees, the needle is advanced another 1 to 2 mm to ensure that the tip of the catheter also lies within the lumen of the vessel, and the outer catheter is then threaded off the needle while watching that blood continues to flow out of the needle hub (Fig. 9-1). After insertion of the catheter, the wrist should be taken out of the dorsiflexed position, because continued extreme dorsiflexion can lead to median nerve damage by stretching of the nerve over the wrist. An armboard may still be used to prevent the wrist from flexing, which causes kinking of the catheter and damping of the arterial waveform.

CENTRAL VENOUS PRESSURE MONITORING

Central venous pressure catheters are used to measure the filling pressure of the right ventricle, give an estimate of the intravascular volume status, and assess RV function. For accurate pressure measurement, the distal end of the catheter must lie within one of the large intrathoracic veins or the right atrium. Although water manometers have been used in the past, an electronic system is preferred because it allows the observation of the right atrial (RA) waveform, which provides additional information. In any pressure monitoring system, it is necessary to have a reproducible landmark (e.g., the midaxillary line) as a zero reference. This is especially important in monitoring venous pressures, because small changes in the height of the zero reference point produce proportionately larger errors compared with arterial pressure monitoring.

The normal CVP waveform consists of three upward deflections (A, C, and V waves) and two downward deflections (X and Y descents) (Fig. 9-2). The A wave is produced by right atrial contraction and occurs just after the P wave on the ECG. The C wave occurs because of the isovolumic ventricular contraction forcing the tricuspid valve to bulge upward into the right atrium. The pressure within the right atrium then decreases as the tricuspid valve is pulled away from the atrium during RV ejection, forming the X descent. RA filling continues during late ventricular systole, forming the V wave. The Y descent occurs when the tricuspid valve opens and blood from the right atrium empties rapidly into the right ventricle during early diastole.

The CVP is a useful monitor if the factors affecting it are recognized and its limitations are understood. The CVP reflects the patient’s blood volume, venous tone, and RV performance. Following serial measurements (trends) is more useful than individual numbers. The response of the CVP to a volume infusion is a useful test.

The CVP does not give a direct indication of left-heart filling pressure, but it may be used as an estimate of left-sided pressures in patients with good LV function. A good correlation has been shown between the CVP and left-sided filling pressures during a change in volume status in patients with coronary artery disease and left ventricular ejection fraction (LVEF) greater than 0.4.

Internal Jugular Vein

Cannulation of the internal jugular vein (IJV) was first described by English and coworkers in 1969. Its popularity among anesthesiologists has steadily increased since that time. Advantages of this technique include the high success rate as a result of the relatively predictable relationship of the anatomic structures; a short, straight course to the right atrium that almost always assures RA or superior vena cava (SVC) localization of the catheter tip; easy access from the head of the operating room table; and fewer complications than with subclavian vein catheterization. The IJV is located under the medial border of the lateral head of the sternocleidomastoid (SCM) muscle (Fig. 9-3). The carotid artery is usually deep and medial to the IJV. The right IJV is preferred, because this vein takes the straightest course into the SVC, the right cupola of the lung may be lower than the left, and the thoracic duct is on the left side.

The preferred middle approach to the right IJV is shown in Figure 9-4. With the patient supine or in Trendelenburg position and the head turned toward the contralateral side, the fingers of the left hand are used to palpate the two heads of the SCM muscle and the carotid pulse. These fingers then hold the skin stable over the underlying structures while local anesthetic is infiltrated into the skin and subcutaneous tissues. A 22-gauge “finder” needle is placed at the apex of the triangle formed by the two heads of the SCM muscle at a 45-degree angle to the skin and directed toward the ipsilateral nipple. If venous blood return is not obtained, the needle is withdrawn to the subcutaneous tissue and then passed in a more lateral or medial direction until the vein is located. This needle reduces the risk of consequences related to inadvertent carotid arterial puncture and tissue trauma if localization of the vein is difficult. When venous blood is aspirated through the “finder” needle, the syringe and needle are withdrawn, leaving a small trail of blood on the drape to indicate the direction of the vein. Alternatively, the needle and syringe can be fixated and used as an identifying needle. Then, a syringe attached to an 18-gauge intravenous catheter-over-needle is inserted in an identical fashion. When venous return is present, the whole assembly is lowered to prevent the needle from going through the posterior wall of the central vein and advanced an additional 1 to 2 mm until the tip of the catheter is within the lumen of the vein. The catheter is then threaded into the vein.

Once the catheter is advanced into the vein, the needle is removed, and an empty syringe is attached to the cannula to withdraw a sample of blood. To confirm that an artery has not been inadvertently cannulated, comparison of the color of the blood sample to an arterial sample drawn simultaneously is recommended. If this is inconclusive or there is no arterial catheter in place, the cannula may be attached to a transducer by sterile tubing to observe the pressure waveform. Another option is to attach the cannula to sterile tubing and allow blood to flow retrograde into the tubing. The tubing is then held upright as a venous manometer, and the height of the blood column is observed. If the catheter is in a vein, it will stop rising at a level consistent with the CVP and demonstrate respiratory variation. A guidewire is then passed through the 18-gauge catheter, and the catheter is exchanged over the wire for a CVP catheter. The use of more than one technique to confirm the venous location of the catheter may provide additional reassurance of correct placement before cannulation of the vein with a larger cannula.

ultrasonic guidance of internal jugular vein cannulation

Ultrasound has been increasingly used to define the anatomic variations of the IJV. A review and meta-analysis of randomized controlled trials looking at ultrasound-guided central venous cannulation found that real-time two-dimensional ultrasound for IJV cannulation had a significantly higher success rate overall and on the first attempt compared with the landmark method in adults.4 Most studies have demonstrated that two-dimensional ultrasonic guidance of IJV cannulation is helpful in locating the vein, permits more rapid cannulation, and decreases the incidence of arterial puncture.5 Circumstances in which ultrasonic guidance of IJV cannulation can be advantageous include patients with difficult neck anatomy (e.g., short neck, obesity), prior neck surgery, anticoagulated patients, and infants.

Ultrasound has provided more precise data regarding the structural relationship between the IJV and the carotid artery (Fig. 9-5). Troianos and associates found that in more than 54% of patients, more than 75% of the IJV overlies the carotid artery. Patients who were older than 60 years were more likely to have this type of anatomy.6 There was greater overlap of the IJV and the carotid artery when the head is rotated 80 degrees compared with head rotation of only 0 to 40 degrees. The data from 2 and 4 cm above the clavicle did not differ, and the percentage overlap was larger on the left side of the neck compared with the right. Excessive rotation of the head of the patient toward the contralateral side may distort the normal anatomy in a manner that increases the risk of inadvertent carotid artery puncture.7 Doppler ultrasonography has also been used to demonstrate that the Valsalva maneuver increases IJV cross-sectional area by approximately 25% and that the Trendelenburg position increases it by approximately 37%.

Indications

Central venous pressure monitoring is often performed to obtain an indication of intravascular volume status. The accuracy and reliability of CVP monitoring depend on many factors, including the functional status of the right and left ventricles, the presence of pulmonary disease, and ventilatory factors, such as positive end-expiratory pressure (PEEP). The CVP may reflect left-sided heart filling pressures, but only in patients with good LV function. Elderly patients have a high incidence of coronary artery and pulmonary disease, and the CVP is therefore less likely to accurately reflect left-sided filling pressures in this population. Perioperative indications for the insertion of a central venous catheter are listed in Box 9-4.

The CVP should be monitored in all patients during CPB. When the catheter tip is in the SVC, it indicates RA pressure and cerebral venous pressure. Significant increases in CVP can produce critical decreases in cerebral perfusion pressure. This is occasionally caused by a malpositioned SVC cannula during CPB and must be corrected immediately by the surgeon to avoid cerebral edema and poor cerebral perfusion.

Complications

The complications of central venous cannulation can be roughly divided into three categories: complications of vascular access, complications of catheter insertion, or complications of catheter presence. These are summarized in Box 9-5.

Inadvertent arterial puncture during central venous cannulation is not uncommon. The two main reasons why this phenomenon occurs are that all veins commonly used for cannulation lie in close proximity to arteries and that the venous anatomy is quite variable. Localized hematoma formation is the usual consequence. This may be minimized if a small-gauge needle is initially used to localize the vein or ultrasonic guidance is employed.

If RA or RV perforation occurs during central venous cannulation, pericardial effusion or tamponade may result. The likelihood of this complication is increased when inflexible guidewires, long dilators, or catheter are used.

The dilators used in many of the central catheters kits may be a major cause of vessel perforation. The dilator may bend the guidewire, creating its own path, causing it to perforate a vessel wall. Several kits have dilators that are much longer than the catheters, and they constitute a further risk factor for possible perforation of the heart or vessels.

The physiology of fluid accumulation in the pericardial sac is such that sudden cardiovascular collapse occurs once a critical volume has been reached. This is explained by the compliance curve of the normal pericardium. The curve is flat until the critical volume is reached and then rises steeply with any further increment in volume. If pericardial tamponade is imminent, immediate pericardiocentesis is indicated. Withdrawal of small volumes of blood results in marked hemodynamic improvement because of the nature of pericardial compliance.

Transient atrial and ventricular arrhythmias commonly occur as the guidewire is passed into the right atrium or right ventricle during central venous cannulation using the Seldinger technique. This most likely results from the relatively inflexible guidewire causing extrasystoles as it contacts the endocardium. Ventricular fibrillation during guidewire insertion has been reported. The same investigators reported a 70% reduction in the incidence of arrhythmias when guidewire insertion was limited to 22 cm.

There are also reports of complete heart block due to guidewire insertion during central venous cannulation. These cases can be successfully managed using a temporary transvenous or external pacemaker. This complication has previously been reported with PA catheterization. The problem most likely resulted from excessive insertion of the guidewire, with impingement of the wire in the region of the right bundle branch. It is recommended that the length of guidewire insertion be limited to the length necessary to reach the SVC-RA junction to avoid these complications. It is also imperative to monitor the patient appropriately (e.g., ECG and pulse monitoring) and to have resuscitative drugs and equipment immediately available when performing central venous catheterization.

Strict aseptic technique is required to minimize catheter-related infections. Full barrier precautions during insertion of central venous catheters have been shown to decrease the incidence of catheter-related infections.8 Subcutaneous tunneling of central venous catheters inserted into the internal jugular and femoral veins, antiseptic barrier-protected hub for central venous catheters, and antiseptic/antibiotic-impregnated short-term central venous catheters have been shown to reduce catheter-related infections. Hospital policies differ with respect to the permissible duration of catheterization at particular sites, but routine replacement of central venous catheters to prevent catheter-related infections is not recommended.

PULMONARY ARTERIAL PRESSURE MONITORING

The introduction of the flow-directed PA catheter was a quantum advance in the monitoring of patients in the perioperative period. Since the 1970s, its use has increased the amount of diagnostic information that can be obtained at the bedside in critically ill patients. It is impressive to observe large changes in the PAP and PCWP with almost no reflection in the CVP. Connors and coworkers prospectively analyzed 62 consecutive PA catheterizations. They found that fewer than one half of a group of clinicians correctly predicted the PCWP or CO and that more than 50% made at least one change in therapy based on data from the PA catheterization. Waller and Kaplan demonstrated that a group of experienced cardiac anesthesiologists and surgeons who were blinded to the information from PA catheterization during CABG surgery were unaware of any problem during 65% of severe hemodynamic abnormalities. Similarly, Iberti and Fisher showed that ICU physicians were unable to accurately predict hemodynamic data on clinical grounds and that 60% made at least one change in therapy and 33% changed their diagnosis based on PA catheter data*. The clinical significance of these changes has been questioned because the weight of evidence-based medicine on the subject does not support improvements in outcome related to PA catheter monitoring, and the overall use of the PA catheter has decreased 60–80% over the past decade. Nevertheless, with increasing numbers of patients with multisystem organ dysfunction undergoing cardiac surgical procedures, PA catheter monitoring is prevalent in cardiac surgical settings. An understanding of the potential benefits and pitfalls of PA catheterization is therefore essential for anesthesiologists.

Specific information that can be gathered with the PA catheter and the quantitative measurements of cardiovascular and pulmonary function that can be derived from this information are listed in Tables 9-1 and 9-2.

Table 9-1 Normal Intracardiac Pressures

Location/Pressure Mean (mm Hg) Range (mm Hg)
Right atrial 5 1-10
Right ventricular 25/5 15-30/0-8
Pulmonary arterial systolic/diastolic 23/9 15-30/5-15
Mean pulmonary arterial 15 10-20
Pulmonary capillary wedge 10 5-15
Left atrial 8 4-12
Left ventricular end-diastolic 8 4-12
Left ventricular systolic 130 90-140

Table 9-2 Derived Hemodynamic Parameters

Parameter/Formula Normal Values
Cardiac index: CI = CO/BSA 2.8-4.2 L/min/m2
Stroke volume: SV = CO × 1000/HR 50-110 mL/beat
Stroke index: SI = SV/BSA 30-65 mL/beat/m2
Left ventricular stroke work index: LVSWI = 1.36 × (MAP − PCWP) × SI/100 45-60 g•m/m2
Right ventricular stroke work index: RVSWI = 1.36 × (MPAP − CVP) × SI/100 5-10 g•m/m2
Systemic vascular resistance: SVR = (MAP − CVP) × 80/CO 900-1400 dynes•s•cm−5
Systemic vascular resistance index: SVRI = (MAP − CVP) × 80/CI 1500-2400 dynes•s•cm−5/m2
Pulmonary vascular resistance: PVR = (MPAP − PCWP) × 80/CO 150-250 dynes•s•cm−5
Pulmonary vascular resistance index: PVRI = (MPAP − PCWP) × 80/CI 250-400 dynes•s•cm−5/m2

BSA = body surface area; CI = cardiac index; CO = cardiac output; CVP = central venous pressure; HR = heart rate; LVSWI = left ventricular stroke work index; MAP = mean arterial pressure; PAP = pulmonary arterial pressure; PCWP = pulmonary capillary wedge pressure; PVR = pulmonary vascular resistance; PVRI = pulmonary vascular resistance index; RVSWI = right ventricular stroke work index; SI = stroke index; SV = stroke volume; SVR = systemic vascular resistance; SVRI = systemic vascular resistance index.

One of the main reasons that clinicians measure PCWP and PA diastolic pressure (PADP) is that these parameters are estimates of LAP, which is an estimate of left ventricular end-diastolic pressure (LVEDP). LVEDP is an index of left ventricular end-diastolic volume (LVEDV), which correlates well with left ventricular preload. The relationship between LVEDP and LVEDV is described by the left ventricular compliance curve. This nonlinear curve is affected by many factors, such as ventricular hypertrophy and myocardial ischemia. The PCWP and PADP do not directly measure LV preload. The relationship of these parameters is diagrammed in Figure 9-6.

The PCWP and PADP pressures will not accurately reflect LVEDP in the presence of incorrect position of the PA catheter tip, pulmonary vascular disease, high levels of PEEP, or mitral valvular disease. The patency of vascular channels between the distal port of the PA catheter and the LA is necessary to ensure a close relationship between the PCWP and LAP. This condition is met only in the dependent portions of the lung (West’s zone III), in which the pulmonary venous pressure exceeds the alveolar pressure. Otherwise, the PCWP will reflect the alveolar pressure, not the LAP. Because PEEP decreases the size of West’s zone III, it has been shown to adversely affect the correlation between the PCWP and LAP, especially in the hypovolemic patient. Nevertheless, the correlation of PCWP and LAP could be maintained in the presence of PEEP by placing the catheter tip below the left atrium. The acute respiratory distress syndrome (ARDS) seems to prevent the transmission of increased alveolar pressure to the pulmonary interstitium. This preserves the relationship between the PCWP and LAP during the application of PEEP. It is not considered prudent, however, to temporarily disconnect patients from PEEP to measure preload.

The presence of large V waves in the PCWP tracing of patients with mitral regurgitation leads to an overestimation of the LVEDP. In patients with mitral stenosis, using the PCWP instead of the LAP to assess the transmitral gradient has been shown to overestimate the severity of mitral stenosis. However, when the PCWP was adjusted for the time delay through the pulmonary vasculature, the mean LAP and mean PCWP correlated well. It has been demonstrated that there is a significant positive gradient between the PCWP and the LAP in the initial hour after CPB. Box 9-6 is a summary of conditions that may alter the relationship between the PCWP and the LVEDP.

Placement of the Pulmonary Artery Catheter

The considerations for the insertion site of a PA catheter are the same as for CVP catheters. The right IJV approach remains the technique of choice because of the direct path between this vessel and the right atrium. The placement of PA catheters through subclavian vein introducers may be complicated by kinking of the catheter when the sternum is retracted during cardiothoracic surgery.

Passage of the PA catheter from the vessel introducer to the PA can be accomplished by monitoring the pressure waveform from the distal port of the catheter or under fluoroscopic guidance. Waveform monitoring is the more common technique for perioperative right-sided heart catheterization. First, the catheter must be advanced through the vessel introducer (15 to 20 cm) before inflating the balloon. The inflation of the balloon facilitates further advancement of the catheter through the right atrium and right ventricle into the PA. Normal intracardiac pressures are shown in Table 9-1. The pressure waveforms seen during advancement of the PA catheter are illustrated in Figure 9-7. The RA waveform is seen until the catheter tip crosses the tricuspid valve and enters the right ventricle. In the right ventricle, there is a sudden increase in SBP but little change in DBP compared with the RA tracing. Arrhythmias, particularly premature ventricular complexes, usually occur at this point but almost always resolve without treatment once the catheter tip has crossed the pulmonary valve. The catheter is rapidly advanced through the right ventricle toward the PA.

As the catheter crosses the pulmonary valve, a dicrotic notch appears in the pressure waveform and there is a sudden increase in diastolic pressure. The PCWP tracing is obtained by advancing the catheter 3 to 5 cm farther until there is a change in the waveform associated with a drop in the measured mean pressure. Deflation of the balloon results in reappearance of the PA waveform and an increase in the mean pressure value. Using the right IJV approach, the right atrium is entered at 25 to 35 cm, the right ventricle at 35 to 45 cm, the PA at 45 to 55 cm, and the PCWP at 50 to 60 cm in most patients.

If the catheter does not enter the PA by 60 cm, the balloon should be deflated. It should be withdrawn into the right atrium, and another attempt should be made to advance the catheter into proper position. Excessive coiling of the catheter in the right ventricle should be avoided to prevent catheter knotting. The balloon should be inflated only for short periods to measure the PCWP. The PA waveform should be continually monitored to be certain that the catheter does not float out into a constant wedge position because this may lead to PA rupture or pulmonary infarction. The PA catheter is covered by a sterile sheath that must be secured at both ends to prevent contamination of the external portion of the catheter. Not infrequently, the PA catheter must be withdrawn a short distance because the extra catheter in the right ventricle floats out more peripherally into the PA over time as the catheter softens.

The time for the entire PA catheter insertion procedure is 10 to 15 minutes in experienced hands. Many clinicians believe that the PA catheter should be inserted before the induction of anesthesia. Studies showed that PA catheter insertion in the awake patient did not result in myocardial ischemia or deleterious hemodynamic changes after preanesthetic medication and with continuation of all preoperative cardiac medications. A chest radiograph should be obtained postoperatively in all patients to check the position of the PA catheter. Most catheters pass into the right middle or lower lobes. The location of the PA catheter can also be checked by TEE imagine of the catheter in the RA, RV, and pulmonary artery.

Indications

The indications for PA catheterization are given in Box 9-7. The ability of PA catheters to positively influence patient outcome has never been conclusively proved in large-scale, prospective studies. There remains considerable controversy regarding the risk/benefit ratio of PA catheters. Many studies have reported no change or even worse outcome in patients who were monitored with PA catheters. Randomized trials on patients with myocardial infarction seemed to confirm these data, whereas earlier prospective studies on surgical patients showed improved outcome.911

Major problems with PA catheter outcome studies include flaws in study design and insufficient statistical power. The most common design flaws were a lack of therapeutic protocols or treatment algorithms and inadequate randomization, which introduce observer bias.12 Physician knowledge is another confounding variable, as demonstrated in multicenter studies that indicated competency in interpreting PA catheter–derived data was lacking in many individuals and depended on such factors as the level of training and the frequency of use. As many as 47% of physicians could not correctly determine the PCWP to within 5 mm Hg.

Another problem with PA catheter outcome studies is the clinical setting, specifically operating room versus ICU. Patients in the ICU might have disease too far advanced to make invasive hemodynamic monitoring useful. Studies that have reported improved outcome used invasive hemodynamic monitoring to optimize oxygen delivery in the perioperative period.

The operative procedures and medical conditions that are cited as indications for PA catheterization in the perioperative period remain controversial and vary by institution. In a global sense, the indications for using a PA catheter are assessing volume status, measuring CO, measuring image, and deriving hemodynamic parameters. In 2003, the American Society of Anesthesiologists (ASA) Task Force on Pulmonary Artery Catheterization published updated practice guidelines for PA catheterization (http://www.asahq.org/publicationsAndServices/pulm_artery.pdf).13 These guidelines emphasized that patient, surgery, and practice setting had to be considered. Generally, the routine use of PA catheters is indicated in high-risk patients (e.g., ASA 4 or 5) and high-risk procedures (e.g., where large fluid changes or hemodynamic disturbances are expected). The practice setting is important, because there is evidence that inadequate training or experience may increase the risk for perioperative complications associated with the use of a PA catheter. It is recommended that the routine use of a PA catheter should be confined to centers with adequate training and experience in the perioperative management of patients with the PA catheter (Box 9-8). A summary of procedural indications that is relatively aggressive is shown in Box 9-9.

BOX 9-8 American Society of Anesthesiologists’ Practice Guidelines for Pulmonary Artery Catheter Use

The use of the PA catheter has significantly contributed to the understanding and care of patients with cardiac disease. Nevertheless, further large-scale randomized controlled trials are needed to clearly define which, if any, patient populations benefit from PA catheter monitoring. The body of evidence is inconclusive. The risks associated with perioperative PA catheter monitoring may outweigh the benefits in low-to-moderate risk patients, whereas high-risk patients undergoing major surgery probably benefit from right-sided heart catheterization. One major caveat is that the data derived from PA catheter monitoring must be interpreted correctly and treatment protocols should be followed to derive maximal benefit. In summary, using an evidence-based medicine approach, PA catheter monitoring does not improve patient outcome in most patient populations and may actually be harmful in certain circumstances.

Complications

The complications associated with PA catheter placement include almost all of those detailed in the section on CVP placement. The ASA Task Force on Pulmonary Artery Catheterization concluded that serious complications due to PA catheterization occur in 0.1% to 0.5% of patients monitored with a PA catheter. Higher estimates are found in the literature and probably represent different patient populations, hospital settings, level of experience with PA catheter management, and other factors.

Endobronchial Hemorrhage

Iatrogenic rupture of the PA has become more common since the advent of PA catheter monitoring in the ICU and operating room.14 Several risk factors have emerged: advanced age, female sex, pulmonary hypertension, mitral stenosis, coagulopathy, distal placement of the catheter, and balloon hyperinflation. Balloon inflation in distal pulmonary arteries is probably accountable for most episodes of PA rupture because of the high pressures generated by the balloon. Hypothermic CPB may also increase risk due to distal migration of the catheter tip with movement of the heart and hardening of the PA catheter. It is now common practice to pull the PA catheter back 3 to 5 cm when CPB is instituted.

It is important to consider the cause of the hemorrhage when forming a therapeutic plan. If the hemorrhage is minimal and a coagulopathy coexists, correction of the coagulopathy may be the only necessary therapy. Protection of the uninvolved lung is of prime importance. Tilting the patient toward the affected side, placement of a double-lumen endotracheal tube, and other lung-separation maneuvers should protect the contralateral lung. Strategies proposed to stop the hemorrhage include the application of PEEP, placement of bronchial blockers, and pulmonary resection. The clinician is obviously at a disadvantage unless the site of hemorrhage is known. A chest radiograph will usually indicate the general location of the lesion. Although the cause of endobronchial hemorrhage may be unclear, the bleeding site must be unequivocally located before surgical treatment is attempted. A small amount of radiographic contrast dye may help to pinpoint the lesion if active hemorrhage is present. In severe hemorrhage and with recurrent bleeding, transcatheter coil embolization has been used. This may emerge as the preferred treatment method.

Pacing Catheters

The possible indications for placement of a pacing PA catheter are shown in Box 9-10. The actual use in a group of 600 patients undergoing cardiac surgery is shown in Table 9-3.15

A multipurpose PA catheter contains five electrodes for bipolar atrial, ventricular, or atrioventricular (AV) sequential pacing. The intraoperative success rates for atrial, ventricular, and AV sequential capture have been reported as 80%, 93%, and 73%, respectively.

The Paceport and A-V Paceport catheters have lumina for the introduction of a ventricular wire (Paceport) or for atrial and ventricular wires (A-V Paceport) for temporary transvenous pacing. The success rate for ventricular pacing capture was 96% for the Paceport. The success rates for atrial and ventricular pacing capture before CPB were 98% and 100%, respectively, in a study of the A-V Paceport.

Mixed Venous Oxygen Saturation Catheters

Monitoring the image is a means of providing a global estimation of the adequacy of oxygen delivery relative to the needs of the various tissues. The formula for image calculation can be derived by modifying the Fick equation and assuming that the effect of dissolved oxygen in the blood is negligible:

image

A decrease in the image can indicate one of the following situations: decreased CO; increased oxygen consumption; decreased arterial oxygen saturation; or decreased hemoglobin (Hb) concentration. To measure image, blood is aspirated from the distal port of the PA catheter slowly, so as not to contaminate the sample with oxygenated alveolar blood.

The addition of fiberoptic bundles to PA catheters has enabled the continuous monitoring of image using reflectance spectrophotometry. The catheter is connected to a device that includes a light-emitting diode and a sensor to detect the light returning from the PA. image is calculated from the differential absorption of various wavelengths of light by the saturated and desaturated hemoglobin.

If it is assumed that there is constant oxygen consumption and arterial oxygen content, changes in image should reflect changes in CO. Several investigators have come to the conclusion that it provides a valuable measure of CO during surgery. The image has been shown to correlate with cardiac index during CABG surgery when oxygen consumption is constant, but not with the cardiac index when oxygen consumption is changing, such as during shivering after anesthesia. The usefulness of the catheter may primarily be its ability to continuously monitor the balance between oxygen delivery and consumption,16 and image may also help predict survival after acute myocardial infarction.

CARDIAC OUTPUT MONITORING

The CO is the amount of blood pumped to the peripheral circulation by the heart each minute. It is a measurement that reflects the status of the entire circulatory system, not just the heart, because it is governed by autoregulation from the tissues. The CO is equal to the product of the SV and the heart rate. Preload, afterload, heart rate, and contractility are the major determinants of the CO. The measurement of CO is of particular interest in patients with cardiac disease.

Indicator Dilution

The indicator dilution method is based on the observation that, for a known amount of indicator introduced at one point in the circulation, the same amount of indicator should be detectable at a downstream point. The amount of indicator detected at the downstream point is equal to the product of CO and the change in indicator concentration over time. CO is calculated using the Stewart-Hamilton equation:

image

in which CO is cardiac output, I is amount of indicator injected, and ∫C dt is the integral of indicator concentration over time (60 converts seconds to minutes).

Cold saline (i.e., thermodilution) or lithium ions are used as the indicator, whereas dye (e.g., indocyanine green) or radioisotopes are rarely used in current practice. Blood flow is directly proportional to the amount of the indicator delivered and inversely proportional to the amount of indicator that is present at a sampling site distal to the injection site.

Thermodilution

intermittent thermodilution cardiac output

The thermodilution method, using the PA catheter, is the most commonly used method at present for invasively measuring CO in the clinical setting. With this technique, multiple COs can be obtained at frequent intervals using an inert indicator, and without blood withdrawal. A bolus of cold fluid is injected into the right atrium, and the resulting temperature change is detected by the thermistor in the PA. When a thermal indicator is used, the modified Stewart-Hamilton equation is used to calculate CO:

image

in which CO is the cardiac output (L/min), V is the volume of injectate (mL), TB is the initial blood temperature (°C), TI is the initial injectate temperature (°C), K1 is the density factor, K2 is the computation constant, and 0 ΔTB(t)dt is the integral of blood temperature change over time.

Solution of this equation is performed by a computer that integrates the area under the temperature versus time curve. CO is inversely proportional to the area under the curve.

The temperature-versus-time curve is the crux of this technique, and any circumstances that affect it have consequences for the accuracy of the CO measurement. Specifically, anything that results in less “cold” reaching the thermistor, more “cold” reaching the thermistor, or an unstable temperature baseline will adversely affect the accuracy of the technique. Less “cold” reaching the thermistor would result in overestimation of the CO. This could be caused by a smaller amount of indicator, an indicator that is too warm, a thrombus on the thermistor, or partial “wedging” of the catheter. Conversely, underestimation of the CO will occur if excessive volume of injectate, or injectate that is too cold, is used to perform the measurement. Intracardiac shunts have unpredictable effects that depend on the anatomy and physiology of individual patients. Variations of up to 80% in measured CO occur when the rate of administration of intravenous crystalloid infusions caused fluctuations in baseline blood temperature. The rapid temperature decrease seen after hypothermic CPB has been shown to result in the underestimation of CO by 0.6 to 2.0 L/min. The normal changes in the PA that occur with each respiratory cycle appear to be exaggerated in the early phase after hypothermic CPB. This may cause peak-to-peak errors in estimation of intermittent CO of up to 50% if initiated at different times during the ventilatory cycle. This effect was significantly decreased with thermal equilibration, approximately 30 minutes after CPB. This problem is less prevalent currently, because hypothermic CPB is used less commonly.

The precision of the thermodilution CO technique is not very good but can be improved by ensuring that, for each determination, the rate and duration of the injection are kept as constant as possible. Whenever possible, 10-mL volumes of injectate should be used and the timing of the injection in the respiratory cycle should be the same.

ANALYSIS AND INTERPRETATION OF HEMODYNAMIC DATA

The information provided by hemodynamic monitoring permits the calculation of various derived parameters that assist in evaluating patients clinically. The formulas, normal values, and units for the calculation of various hemodynamic parameters are presented in Table 9-2. These parameters include the SVR, pulmonary vascular resistance (PVR), SV, left ventricular stroke work (LVSW), and right ventricular stroke work (RVSW). As an example of information that may be obtained, graphs of PCWP versus SV can be constructed for individual patients; these “Starling curves” provide insight into the contractile state of the heart. Although these parameters are easily derived using the standard formulas, many modern monitors perform these calculations. To compare data among patients of different body weights and types, the various hemodynamic parameters may be normalized by indexing them to body surface area.

Systemic and Pulmonary Vascular Resistances

Systemic vascular resistance represents an estimation of the afterload of the left ventricle. Afterload is roughly defined as the force that impedes or opposes ventricular contraction. Higher SVR results in increased LV systolic wall stress. This has clinical significance because LV wall stress is one of the major determinants of myocardial oxygen consumption. Elevations in wall stress have been observed in patients with LV enlargement due to systemic hypertension, aortic stenosis, and aortic regurgitation.

Clinically, calculations of SVR are used to assess the response to inotropic, vasodilatory, and vasoconstrictive agents.19 For example, a patient who is hypotensive despite a high normal CO has a low SVR. The SVR is calculated, and then therapy is instituted (e.g., a vasoconstrictor). A repeat calculation of the SVR enables the clinician to titrate the therapy to the appropriate endpoint. Despite this common use in the operating room and ICU setting, there is good evidence that SVR is not an accurate indicator of true afterload. Nevertheless, SVR remains the clinical technique for measuring afterload at the present time.

PVR remains the traditional measure of afterload of the right ventricle. Systolic PAP may provide a better estimation of RV afterload. PVR and PAP do provide some clinically useful information regarding the pulmonary vasculature and are readily available in patients with PA catheters. The PVR should be used in conjunction with other hemodynamic data to assess the response of the pulmonary vasculature to pharmacologic therapy and physiologic changes.

Frank-Starling Relationships

Myocardial function depends on the contractile state and the preload of the ventricle (sarcomere length at end-diastole). The relationship between the ventricular preload and myocardial work (ventricular stroke work) is the Frank-Starling relationship. The slope of the curve indicates the contractile state of the myocardium (Fig. 9-8). For clinical purposes, it is usually not feasible to measure actual end-diastolic volumes (can be estimated with TEE), and approximations of end-diastolic pressure, such as the PCWP or LAP, are often substituted. This introduces error, because the relationship between end-diastolic pressure and volume is usually nonlinear (as described by the diastolic ventricular compliance curve) and is dynamic. Unfortunately, the Frank-Starling relationship is extremely sensitive to changes in afterload. Patients with LV or RV dysfunction may have severe decrements in SV with increased SVR or PVR, respectively.

ELECTROCARDIOGRAPHY

Interpretation of the ECG is often considered the domain of the cardiologist, but clinicians providing perioperative and critical care also derive important information from it by the standard 12-lead tracing or as a continuous “monitoring” modality.

The practicing anesthesiologist relies on the ECG to make critical decisions at many phases of the perioperative period in patients undergoing cardiac or noncardiac (particularly vascular) surgery.20

Lead Systems

Einthoven established electrocardiography using three extremities as references: the left arm, right arm, and left leg. He recorded the difference in potential between the left arm and right arm (lead I), between the left leg and right arm (lead II), and between the left leg and left arm (lead III). Because the signals recorded were differences between two electrodes, these leads were called bipolar. The positive or negative polarity of each of the limbs was chosen by Einthoven to result in positive deflections of most of the waveforms and has no innate physiologic significance. He postulated that the three limbs defined an imaginary equilateral triangle with the heart at its center.

Wilson refined and introduced the unipolar precordial leads into clinical practice. To implement these leads, he postulated a mechanism whereby the absolute level of electrical potential could be measured at the site of the exploring precordial electrode (the positive electrode). A negative pole with zero potential was formed by joining the three limb electrodes in a resistive network in which equally weighted signals cancel each other out. He called this the central terminal. He described three additional limb leads (VL, VR, and VF). These leads measured new vectors of activation, and in this way the hexaxial reference system for determination of electrical axis was established. He subsequently introduced the six unipolar precordial V leads in 1935 (Fig. 9-9).

Detection of Myocardial Ischemia

Pathophysiology of ST-Segment Responses

The ST segment is the most important portion of the QRS complex for evaluating ischemia (Box 9-11). The origin of this segment, at the J point, is easy to locate. Its end, which is generally accepted as the beginning of any change of slope of the T wave, is more difficult to determine. In normal individuals there may be no discernible ST segment as the T wave starts with a steady slope from the J point, especially at rapid heart rates. The TP segment has been used as the isoelectric baseline from which changes in the ST segment are evaluated, but with tachycardia this segment is eliminated, and during exercise testing the PR segment is used. The PR segment is used in all ST-segment analyzers.

Repolarization of the ventricle proceeds from the epicardium to the endocardium, opposite to the vector of depolarization. The ST segment reflects the midportion, or phase 2, of repolarization during which there is little change in electrical potential. It is usually isoelectric. Ischemia causes a loss of intracellular potassium, resulting in a current of injury. The electrophysiologic mechanism accounting for ST-segment shifts (elevation or depression) remains controversial. The two major theories are based on a loss of resting potential as current flows from the uninjured to the injured area (i.e., diastolic current) and on a true change in phase 2 potential as current flows from the injured to the uninjured area (i.e., systolic current). With subendocardial injury, the ST segment is depressed in the surface leads. With epicardial or transmural injury, the ST segment is elevated. 21,22

Electrocardiographic Manifestations of Ischemia

With myocardial ischemia, repolarization is affected, resulting in downsloping or horizontal ST-segment depression. Various local effects and differences in vectors during repolarization result in different ST morphologies that are recorded by the different leads. It is generally accepted that ST changes in multiple leads are associated with more severe degrees of coronary artery disease.

The classic criterion for ischemia is 0.1 mV (1 mm) of ST-segment depression measured 60 to 80 ms after the J point.2325 The slope of the segment must be horizontal or downsloping. Downsloping depression may be associated with a greater number of diseased vessels and a worse prognosis than horizontal depression. Slowly upsloping depression with a slope of 1 mV/s or less is also used but is considered less sensitive and specific (and difficult to assess clinically). Nonspecific ST-segment depression can be related to drug use, particularly digoxin. Interpretation of ST-segment changes in patients with LV hypertrophy is particularly controversial given the tall R-wave baseline, J-point depression, and steep slope of the ST segment. Although a number of studies have excluded such patients, others (including those using other modalities or epidemiologic studies) observed that LV hypertrophy is a highly significant predictor of adverse cardiac outcome.

The criteria for ischemia with ST-segment elevation (≥0.1 mV in ≥2 contiguous leads) are used in conjunction with clinical symptoms or elevation of biochemical markers to diagnose acute coronary syndromes. It usually results from transmural ischemia, but it may potentially represent a reciprocal change in a lead oriented opposite to the primary vector with subendocardial ischemia. Perioperative ambulatory monitoring studies have also included more than 0.2 mV in any single lead as a criterion, but ST-segment elevation is rarely reported in the setting of noncardiac surgery. It is commonly observed, however, during weaning from CPB in cardiac surgery and during CABG surgery (on and off pump) with interruption of coronary flow in a native or graft vessel. ST-segment elevation in a Q-wave lead should not be analyzed for acute ischemia, although it may indicate the presence of a ventricular aneurysm.

Intraoperative Lead Systems

Detection of perioperative myocardial ischemia has received considerable attention over the past several decades and more recently with publication of several studies of clinical monitoring and therapy (e.g., perioperative β-blockade). Many of these studies demonstrated associations of perioperative ischemia with adverse cardiac outcomes in adults undergoing a variety of cardiac and noncardiac surgical procedures. The ease of use of new ST-segment trending software in operating room monitors has resulted in its routine use.

The recommended leads for intraoperative monitoring, based on several clinical studies, do not differ substantially from those used during exercise testing.26 Clinical studies using continuous 12-lead ECG analysis reported that almost 90% of responses involved ST-segment depression alone (75% in V5 and 61% in V4). In approximately 70% of patients, significant changes were observed in multiple leads. The sensitivity of each of the 12 leads is shown in Figure 9-10. When considered in combination (as occurs clinically), the use of leads V4 and V5 increased sensitivity to 90%, whereas sensitivity for the standard clinical combination of leads II and V5 was 80%. Use of leads V2 through V5 and lead II captured all episodes (Table 9-4).

Table 9-4 Sensitivity for Different Electrocardiographic Lead Combinations

Number of Leads Combination Sensitivity (%)
1 lead II 33
  V4 61
  V5 75
2 leads II/V5 80
  II/V4 82
  V4/V5 90
3 leads V3/V4/V5 94
  II/V4/V5 96
4 leads II/V2-V5 100

Data from London MJ, Hollenberg M, Wong MG, et al: Intraoperative myocardial ischemia: Localization by continuous 12-lead electrocardiography. Anesthesiology 69:232, 1988.

A larger clinical study of patients undergoing vascular surgery using a longer period of monitoring (up to 72 hours) with more specific criteria for ischemia (>10 minute duration of episode) extended these observations. It was reported that V3 was most sensitive for ischemia (87%) followed by V4 (79%), whereas V5 alone was only 66% sensitive. In the subgroup of patients in whom prolonged ischemic episodes ultimately culminated in infarction, V4 was most sensitive (83%). In this study, all myocardial infarctions were non–Q-wave events detected by troponin elevation. Use of two precordial leads detected 97% to 100% of changes. Based on analysis of the resting isoelectric levels of each of the 12 leads (a unique component of this study), it was recommended that V4 was the best single choice for monitoring of a single precordial lead, because it was most likely to be isoelectric relative to the resting 12-lead preoperative ECG. In contrast, the baseline ST segment was more likely above isoelectric in V1 through V3 and below isoelectric in V5 and V6. Surprisingly, no episodes of ST elevation occurred in this study.

A cohort of vascular patients monitored in the ICU for the first postoperative day with continuous 12-lead monitoring used a threshold of 20 minutes for an ischemic episode. Eleven percent of 149 patients met the criteria, with ST-segment depression in 71% and ST-segment elevation alone in 18% (12% had both). Most changes were detected in V2 (53%) and V3 (65%). Using the standard two-lead system (II and V5), only 41% of episodes would have been detected.

The use of multiple precordial leads, although appealing, is not likely to become common clinical practice, owing to the limitations of existing monitors (and cables). Even if such equipment were available, it is likely that considerable resistance would occur from practitioners due to the extra effort associated with this approach. Perhaps, in the future, when lower cost wireless technologies are perfected, this approach may become a clinical reality; in the meantime, a combination of lens v5 and II remains the clinical choice of most practitioners.

REFERENCES

1. Dorje P., Tremper K. Systolic pressure variation: A dynamic measure of the adequacy of intravascular volume. Semin Anesth Periop Med Pain. 2005;24:147.

2. Barbeau G.R., Arsenault F., Dugas L., et al. Evaluation of the ulnopalmar arterial arches with pulse oximetry and plethysmography: Comparison with the Allen’s test in 1010 patients. Am Heart J. 2004;147:489.

3. Armstrong P.J., Han D.C., Baxter J.A., et al. Complication rates of percutaneous brachial artery access in peripheral vascular angiography. Ann Vasc Surg. 2003;17:107.

4. Hind D., Calvert N., McWilliams R., et al. Ultrasonic locating devices for central venous cannulation: Meta-analysis. BMJ. 2003;327:361.

5. Riopelle J., Ruiz D., Hunt J., et al. Circumferential adjustment of ultrasound probe position to determine the optimal approach to the internal jugular vein: A noninvasive geometric study in adults. Anesth Analg. 2005;100:512.

6. Troianos C.A., Kuwik R., Pasqual J., et al. Internal jugular vein and carotid artery anatomic relation as determined by ultrasonography. Anesthesiology. 1996;85:43.

7. Parry G. Trendelenburg position, head elevation and a midline position optimize right internal jugular vein diameter. Can J Anaesth. 2004;51:379.

8. Pawar M., Mehta Y., Kapoor P., et al. Central venous catheter-related bloodstream infections: Incidence, risk factors, outcome, and associated pathogens. J Cardiothorac Vasc Anesth. 2004;18:304.

9. Sandham J.D. Pulmonary artery catheter use—refining the question. Crit Care Med. 2004;32:1070.

10. Chittock D.R., Dhingra V.K., Ronco J.J., et al. Severity of illness and risk of death associated with pulmonary artery catheter use. Crit Care Med. 2004;32:911.

11. Sandham J.D., Hull R.D., Brant R.F., et al. A randomized, controlled trial of the use of pulmonary artery catheters in high-risk surgical patients. N Engl J Med. 2003;348:5.

12. Vender J. Pulmonary artery catheter utilization: The use, misuse, or abuse. J Cardiothorac Vasc Anesth. 2006;20:295-299.

13. American Society of Anesthesiologists Task Force on Pulmonary Artery Catheterization: Practice guidelines for pulmonary artery catheterization: An updated report by the American Society of Anesthesiologists Task Force on Pulmonary Artery Catheterization. Anesthesiology. 2003;99:988.

14. Abreu A.R., Campos M.A., Krieger B.P. Pulmonary artery rupture induced by a pulmonary artery catheter: A case report and review of the literature. J Intensive Care Med. 2004;19:291.

15. Risk S.C., Brandon D., D’Ambra M.N., et al. Indications for the use of pacing pulmonary artery catheters in cardiac surgery. J Cardiothorac Vasc Anesth. 1992;6:275.

16. Reinhart K., Kuhn H.J., Hartog C., Bredle D.L. Continuous central venous and pulmonary artery oxygen saturation monitoring in the critically ill. Intensive Care Med. 2004;30:1572.

17. Zollner C., Goetz A.E., Weis M., et al. Continuous cardiac output measurements do not agree with conventional bolus thermodilution cardiac output determination. Can J Anaesth. 2001;48:1143.

18. Manecke G.R., et al. Cardiac output from the arterial catheter: Deceptively simple. J Cardiothorac Vasc Anesth. 2007;21:629.

19. Bashore T. Afterload reduction in chronic aortic regurgitation: It sure seems like a good idea. J Am Coll Cardiol. 2005;45:1031.

20. Weinfurt P.T. Electrocardiographic monitoring: An overview. J Clin Monit. 1990;6:132.

21. Tsuda H., Tobata H., Watanabe S., et al. QRS complex changes in the V5 ECG lead during cardiac surgery. J Cardiothorac Vasc Anesth. 1992;6:658.

22. Crescenzi G., Scandroglio A.M., Pappalardo F., et al. ECG changes after CABG: The role of the surgical technique. J Cardiothorac Vasc Anesth. 2004;18:38.

23. Horacek B.M., Wagner G.S. Electrocardiographic ST-segment changes during acute myocardial ischemia. Card Electrophysiol Rev. 2002;6:196.

24. Carley S.D. Beyond the 12 lead: Review of the use of additional leads for the early electrocardiographic diagnosis of acute myocardial infarction. Emerg Med (Fremantle). 2003;15:143.

25. Zimetbaum P.J., Josephson M.E. Use of the electrocardiogram in acute myocardial infarction. N Engl J Med. 2003;348:933.

26. London M.J. Multilead precordial ST-segment monitoring: “The next generation?”. Anesthesiology. 2002;96:259.