Implantable Cardioverter Defibrillators: Technical Aspects

Published on 20/06/2015 by admin

Filed under Cardiovascular

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 4111 times

115

Implantable Cardioverter Defibrillators

Technical Aspects

Implantable cardioverter defibrillators (ICDs) have revolutionized the treatment of malignant ventricular arrhyhtmias. Their basic tasks include tachycardia detection and termination. To achieve these tasks, the ICD relies on complex integral steps inlcuding sensing of myocardial potentials, delivering these signals to the ICD circuitary board to be filtered and analyzed, and delivering life-saving therapies back to the heart. This chapter outlines the technical aspects of these life-saving devices, including those of the programmer, diagnostics, and telemetry.

System Elements

The physical components of the implanted system consist of the ICD generator, the pacing and sensing electrodes, and one or more high-energy electrodes. The titanium casing of the ICD generator usually constitutes one of the high-energy electrodes. The electrodes, or leads, attach to the generator header through sealed connectors. Until recently, the ICD leads all divided into one bipolar IS-1 (bradycardia) and one or two defibrillation (DF)-1 (high energy) connectors that are inserted into the ICD generator header. A fully approved (March 15, 2010) International Organization for Standardization (ISO) standard (ISO 27186) implemented by several manufacturers is the DF-4 connector standard that is also 3.2 mm, but combines the connection into a single connector for both low-energy (pacing and sensing) and high-energy (shocking) electrode function. The older IS-1/DF-1 lead design is bulky in the device pocket and adds to the length of the lead. In addition, the trifurcation or bifurcation of the lead also creates the potential for errors in making connections to the header. A similarly constructed (IS-4) connection standard is also implemented for quadripolar low-voltage leads. This standard is implemented only for left ventricular cardiac venous leads and permits noninvasive programming of the pacing vectors after the incision is closed. The IS-4 for LV leads and DF-4 for ICD leads are similar but distinct enough not to allow connection errors in the header (Figure 115-1).

The DF-4 connector design provides a single setscrew that secures on the tip of the lead with spring contacts for the ring and DF electrodes. Inside the DF-4 connector, there are double-sealing rings between electrodes to secure good isolation between the high- and low-voltage electrode insulation. These sealing rings have been moved from the electrode to the connector block to prevent any damage that occurs during lead implantation.

The advantages of DF-4 design include quicker connections to the ICD because of the single terminal pin, improved patient comfort and satisfaction, and easier reoperations with decreased debridement because of shorter and less complex lead body in the pocket. The risk of procedural errors, such as set screw stripping, untightenned set screws, and port mismatch (switching RV/SVC DF-1 electrodes), should decrease. Subcutaneous tunneling (submammary implant) may be easier as well.

The potential limitations to this lead design are initially the lack of adapters for including additional high- or low-voltage leads, including subclavian, subcutaneous, azygous vein, or coronary sinus coils for high defibrillation threshold (DFT) patients. As with any change, there are potential unidentified reliability risks, but this development was carefully designed and tested over a decade of research.

Implantable Cardioverter Defibrillator Generator

Implantable cardioverter defibrillator generators have decreased significantly in size because of significant advancements in battery and microprocessor technologies. Most of the volume of current ICDs is occupied by its battery and capacitors. The newest generations of devices have added device-initiated long-range telemetry device interrogation, automatic alert notifications, bioimpedance measurement capability, and other advanced features.

Current devices provide for all these options with a generator volume of approximately 30 mL. The ICD generator casing is made of electrically active titanium, considered to be the preferred material, because of its conductivity, strength, biocompatability, corrosion resistance, and light weight. The casing serves mainly to protect the circuitry from the corrosive effects of body fluids; it also serves as an active high-voltage electrode in many current ICD models. The header is generally made of polymethylmethacrylate, so that the connections with the leads can be visually confirmed during implantation and can be inspected if ICD system troubleshooting is required for component malfunction or suspicion of failure. The lead connections gain access to the ICD circuitry through feed through wires, which penetrate the casing through sealed openings.

The interior of the ICD consists of one or more batteries, capacitors, a direct current (DC)-DC converter, hybrid with microprocessor, telemetry communication coil, and their connections. The sensed ventricular signals, generally 5 to 25 mV in amplitude, enter the generator through the leads, are filtered, and then are analyzed by the algorithms programmed into the hybrid. The hybrid consists of electronic circuitry embedded in a silicon wafer, specifically designed for analysis of these signals and identification of either tachycardia or fibrillation. Once the detection criteria are met, the specific programmed pacing and high-energy shock therapies are then delivered to the patient.

Batteries

Unlike other batteries, ICD batteries have many performance requirements. Besides being compact, the battery must be capable of charging the capacitors by delivering high-energy currents in the range of 75 A to charge the capacitors within seconds, and it must have a low current drain on the order of mA. Factors affecting the current drain include pacing and defibrillation needs and the quiescent current needed for ongoing tasks such as powering the hybrid, monitoring intrinsic rhythm, and logging the data in the memory. In addition, the battery performance over time must be predictable to allow for adequate warnings before it is depleted.

The battery used by most ICD generators is a lithium silver vanadium oxide cell (Li/SVO). There are two kinds of Li/SVO batteries: anode-limited (Li) and cathode-limited (SVO) batteries. Anode-limited limited batteries have two voltage plateaus: an early short-lived one followed later by a long-lived one. Most current ICD batteries are cathode-limited (SVO) batteries and carry a charge of 1000 to 2000 mA-hr and at the beginning of life. The battery generates approximately 3.2 V at full charge. At middle of life, these batteries suffer from inherent internal impedance buildup resulting from the accumulation of a film on the Li electrode. Although this impedance can result in prolonged charge time, pulsing the battery frequently by charging the capacitors is usually sufficient in preventing this phenomenon.

The battery status is generally estimated by its measured output voltage. With significant decline in open circuit voltage and a rise in internal resistance, the ability to deliver adequate current to charge the capacitors becomes impaired causing significant prolongation in charge time; this is called the elective replacement time (ERI) or recommended replacement time. Once the ERI is reached, generator replacement can be scheduled electively, usually within 3 months, depending on the frequency of therapy (which determines the depletion rate). A later indicator, end of life, reflects a significantly lower voltage and indicates a more urgent need for generator replacement because of the associated long capacitor charge times required to achieve appropriate shock energy. In addition to voltage criteria, the time required to charge the capacitors to full energy is also used as a measurement of the battery status, and may activate the elective replacement indicator (ERI). In certain devices, charging the capacitors stops after 20 seconds, and delivers the stored energy as shock therapy. When these devices reach ERI, the stored energy may be significantly less than the programmed energy.

Newer ICDs have encorporated newer battery technologies to improve battery performance and longevity. Balancing the cell to an appropriate electron reduction slows the progression of internal battery impedance overtime. In addition, the development of hybrid cathode batteries (lithium/silver vanadium oxide blended with carbon monofluoride [Li/CFx-SVO]) as well as manganese oxide (lithium manganese dioxide [LiMnO2]) has improved service life and resulted in a stable charge time throughout the lifetime of the battery. In addition, LiMnO2 batteries have no midlife impedance rise and stable voltage during the lifetime of the device, with a slow gradual decay toward an ERI independent of the rate of energy use.

Capacitors

The rate of energy delivery for cardiac defibrillation is much greater than can be delivered directly from ICD batteries. Therefore, capacitors are used to store the energy over longer period of time (seconds) and deliver it over a shorter period of time (milliseconds). Capacitors are measured by their capacitance (C), which is a measure of the amount of electrical charge that the capacitor can store for a given voltage. Multiple capacitors are charged in the parallel configuration from the battery through a DC-DC converter and then connected in series to be able to deliver the stored energy (30 to 80 J) but at high voltages (up to 850 V) within 10 to 20 ms. The rate with which the capacitor is being charged depends on its capacitance (C) and the internal impedance of the battery and the circuitry; however, the rate with which the capacitor delivers its charge to the patient depends on its capacitance (C) and the lead–body impedance.

To achieve high stored charge, current ICDs incorporate new designs such as the use of multiple capacitors in series and the use of capacitors with convoluted surfaces to increase surface area. Most capacitors used in the ICD industry are aluminum and tantalum electrolytic capacitors because of their ability to charge within the prescribed time limitations. With time, the dielectric layer in these capacitors becomes deformed, causing current to flow through it (“capacitor leak”). This leak will result in suboptimal charge and prolonged charge time. Capacitor reformation, in which the capacitors are discharged through a high-impedance circuitry to allow for longer discharge time, will usually regenerate the dielectric layer and fix the leak.

Electrodes

There are three essential functions of defibrillator systems: (1) detection of the tachycardia, (2) pacing stimulation of the heart, and (3) shock stimulation of the heart. The electrodes are the noninsulated segments of the leads that deliver these functions. The technology used for detection (sensing) and pacing is similar to that used by pacemakers; however, the bipole for sensing and pacing is sometimes from the tip of the ventricular lead to the distal shocking electrode instead of from the tip to the ring electrode. This system is an integrated bipolar instead of a true bipolar system. In cardiac resynchronization therapy devices, tachycardia detection is usually restricted to the right ventricular electrodes.

High-voltage coil electrodes are ideal for high-energy shocks. The lower impedance of the coils and the conductor cables to the coils permit a high-current charge discharge from the capacitors. In addition, their wide surface area creates a broader electric field that can enable the defibrillation of more myocardial mass and reduce the current density near the coil.

Nonthoracotomy or Transvenous Leads

Nonthoracotomy ICD leads (NTLs) were designed to carry high defibrillation energy to the inside of the heart. These leads can have a dedicated proximal sensing/pacing ring electrode (true bipolar) or use the distal high-voltage shocking coil as the proximal sensing/pacing electrode (integrated bipolar). True bipolar leads usually offer better discrimination for sensing, being less susceptable to far-field oversensing and postshock undersensing. On the other hand, integrated bipolar leads offer better defibrillation performance because of the shorter tip-to–distal coil distance. NTL leads can have one right ventricular (single coil) or two (dual coil) high-voltage shocking electrodes. The distal coil is usually placed in the right ventricular apex, and the proximal coil is placed in the superior vena cava. Although the dual-coil ICD lead system have predominated in the United States and Europe, its clinical superiority over the single-coil lead system is not well established. Defibrillation efficiancy is slightly improved, usually by 1 to 3 J.1 The effect of the dual-coil ICD lead on defibrillation threshold is multifactorial: altered defibrillation electrical field vector, lowered shock impedance, and an on shortening the shock waveform duration. However, this small benefit needs to be weighed against the added complexity of lead construction and its potential effect on lead reliability, as well as the potential for more complex extraction when needed. In addition, the lower right atrial position of the proximal coil, in the case of severely enlarged right ventricle, can increase DFT by a negative current vector effect. In other patients, there is a need to implant other coils (coronary sinus, middle cardiac vein, subcutaneous coil, or azygous coil) when maximal shocks are ineffective; this is almost always more effective in reducing the required defibrillation energy by 5 to 15 J.2

ICD leads are designed with either coaxial or multilumen constructions. Coaxial design in which layered conductors are separated by layers of polyurethane insulation material—used in some original designs—is no longer used. This design was associated with insulation failure because of metal ion oxidation of the middle polyuerethane insulation layer in the Medtronic Transvene (Minneapolis, MN) family of leads and was manifest by low stimulation impedance and undersensing and oversensing. Current ICD leads use multilumen construction designs in which conductors run in parallel through a single insulating body. However, some variations of this design have been the subject of other failure mechinisms, such as in the case of Sprint Fidelis (Medtronic) and Riata leads (St. Jude Medical [Sylmar, CA]).

Currently available leads either (1) have silicone insulation back filling in between and behind the metal defibrillation coils to decrease tissue ingrowth or (2) have been covered with expanded polytetrafluoroethylene (WL GORE & Associates, (Flagstaff, AZ) ePTFE). This ePTFE coating is slippery, permits the defibrillation current to be transmitted, and prevents tissue ingrowth.

Invariably, NTLs have a high-energy coil located near the distal end and lying within the right ventricular cavity. Manufacturers have released NTLs with similar construction, although some details differ (Table 115-1). The physics of DC flow, however, requires at least one other electrode to complete the shocking circuit. The development of smaller ICD generators has allowed for pectoral implantation, which has enabled the use of the generator casing as the second electrode (i.e., “hot can”). An animal study compared the defibrillation efficacy of a hot can ICD system placed in the left pectoral or subaxillary location with a right pectoral location, and left or right abdominal locations. The left pectoral and axillary subcutaneous positions were superior to all other locations. The right pectoral location was superior to either abdominal location. These results imply that alternative ICD implantation sites are feasible in the event of an inability to implant a left prepectoral device: left subclavian venous occlusion, history of left mastectomy or radiation, left sided arteriovenous fistula, or other reasons to avoid the left prepectoral area.

Tachyarrhythmia Detection

The recognition of tachyarrhythmias by an ICD is a complex interaction of several dependent variables; however, the task of the ICD is more complicated because it must also recognize the lack of tachyarrhythmias. This central insight is crucial, because the patient will spend all but a small fraction of his or her life in a nontachycardia rhythm. Therefore, assuming the efficacy of the pacing and shock therapies, rhythm recognition arbitrates between quality and length of life.

Not all of the factors required for accurate rhythm recognition are potentially affected by the ICD technology. Most notably, the rate and mechanism of the arrhythmia and the programming of the ICD are major determinants of rhythm recognition and are factors that are almost completely independent of the technologic solution. However, by understanding the nature of the signals presented to the ICD, allowing the ICD to adapt to these signals, and limiting programming options, appropriate ICD function is frequently achieved.

Sensing

All current ICDs use ventricular heart rate as the cornerstone variable in tachycardia recognition. To determine heart rate, the interval between each depolarization of the ventricle must be measured. It is not interval recognition but the detection of individual electrogram events that becomes the basic building block in this process. The process begins with the placement of the sensing lead. The sensed electrogram depends on the health of the myocardium in close proximity to the lead, the far-field structures of the diaphragm, anterior chest wall and right atrium, and other electrical devices such as pacemakers, cellular phones, and other sources of electromagnetic interference. Detection of the electrogram events is completely dependent on the quality of the signal, and the quality of the signal is determined primarilly at the time of the lead placement. Additional aspects that are potentially dependent on the position of the lead are measures of electrogram morphology, such as electrogram event width.

In dual-chamber devices, the accurate recognition of arrhythmias adds another layer of complexity and hopefully specificity with the inclusion of atrial lead data input into the generator. The intracardiac location of the atrial lead (far away from the annulus to minimize the far field right ventricular signal) and a short interelectrode distance between the distal and proximal electrodes improve the signal-to-noise ratio of the sensed atrial signal and can thus improve the accuracy of the data used for tachcyardia discrimination.

Band-Pass Filtering

The sense amplifier processes signals presented to the pulse generator by the sensing electrodes and allows signals of certain frequencies to be presented to the detection logic, whereas others are filtered out. This band-pass filter consists of a high-frequency cuttoff to filter out myopotentials signals and a low-frequency cutoff to filter out repolarization T wave signals. The mid range is intended to represent a band of frequency containing true signal events. The intent is to prevent extraneous signals from fooling the device into falsely detecting tachyarrhythmias. Unfortunately, there is some frequency overlap between repolarization and depolarization waves, atrial and ventricular events, postpacing polarization and depolarization of the ventricles, myopotentials and cardiac depolarizations, and environmental signals and cardiac events.

Autogain and Autothreshold

One of the larger challenges presented to ICD detection algorithms is the marked variability in the amplitudes of the signals presented to the ventricular lead: naturally conducted beats through the His-Purkinje system of 5 to 25 mV, paced amplitudes of 5000 mV with polarization voltage after each pacer spike, premature ventricular contractions and ventricular tachycardia amplitudes from 2 to 25 mV, asystolic electrogram amplitudes of 0 to 0.15 mV, and ventricular fibrillation amplitudes from 0.2 to 20 mV. The transitions between each of the rhythms must be identified accurately.

The most difficult problems are distinguishing frequent excursions from low to high amplitude and back to low-amplitude electrogram events. The two approaches are the use of autogain and autothreshold. The autogain technique uses fixed amplitude voltage thresholds and amplifies the signal at continually varying gains according to various algorithms to single count both small and large signals. The autothreshold technique uses one amplifier gain and a continuously varying threshold to accomplish the same end. Usually the threshold will vary within a single cycle, decaying after each sensed event after an adjustment proportional to the amplitude of the signal. Sometimes the floor or minimal voltage limit can be programmed to distinguish between noise on the signal and low-amplitude VF signals. After a paced beat, the autogain is set to maximum and the autothreshold is set to minumum.

Detection Algorithms

Once an electrogram event is marked or detected, the detection algorithm interprets the pattern of the intervals between the events. The algorithms attempt to develop a hierarchy of detected rhythms that require different therapies such as bradycardia, ventricular tachycardia requiring antitachycardia pacing, ventricular tachycardia requiring low-energy shocks, and VF requiring higher-energy shocks. The exclusion of sinus tachycardia and atrial fibrillation with overlapping rates with ventricular arrhythmias is particularly difficult. Additional criteria to distinguish these rhythms include the ventricular rate criterion the duration of the arrhythmia, the acuteness of the increase in rate, and the beat-to-beat variation of the cycle length. Derivatives of the electrogram morphology including templates, duration, and vector correlations, and with the addition of an atrial electrode, the pattern of the relation of the atrial and ventricular events have more recently improved the specificity of arrhythmia detection.

Rate and Duration

The primary characteristic of ventricular arrhythmias requiring treatment is the rate and duration of the arrhythmia. Slow or nonsustained arrhythmias do not require therapy; however, the negative consequence of waiting to see whether an arrhythmia is going to end without an intervention from the ICD increases with rapid rates. The possibility that failure to detect low-amplitude electrogram events will cause a delay or prevent detection is also a serious issue. Consequently, the faster the rhythm, the shorter the duration that should be required for detection. Arrhtyhmia rates are measured by measuring the beat-to-beat interval and are then classified as ventricular tachycardia or ventricular fibrillation. The duration of the arrhythmia is then measured by the number of interval to detect (NID) or the length of time to detect (seconds). For ventricular tachycardia detection, consecutive intervals faster than a programmed rate provide an effective algorithm. However, for VF, a probabilistic NID counter where X short cycle length intervals out of a larger number of Y consecutive intervals is more sensitive because of the marked variability of electrogram amplitudes and intervals during VF.

Sudden Onset

Reentrant ventricular tachycardia and some not clearly reentrant arrhythmias start at a rapid rate, which is usually distinct from the preceding nontachycardia rhythm. The sudden shortening of the cycle length is a factor that is somewhat useful in differentiating sinus tachycardia from slow ventricular tachycardia. The best criterion for preventing ventricular tachycardia detection during sinus tachycardia is programming a more rapid ventricular tachycardia detection rate; however, when that is not possible there is some, but limited, value in using sudden onset. Unfortunately, using this approach increases the risk that the arrhythmia will not be treated if the ventricular tachycardia arises in the setting of sinus tachycardia or exercise. In addition, this discriminator might inhibit therapies for ventricular tachycardia (VT) that started below the programmed detection rate, even when it accelerates beyond the programmed detection rate.

Atrial Lead Criteria

With placement of an atrial lead, the relation of atrial activity to the ventricular activity is available for analysis. If the data are accurate, with the caveats mentioned earlier concerning far-field signals, then AV dissociation during a rapid ventricular rhythm makes the diagnosis of ventricular tachycardia certain. If there is a 1 : 1 relation, the chance that the rhythm is ventricular tachycardia or supraventricular tachycardia is dependent on the ratio of the durations of the AV to the VA interval. Unfortunately, it is still possible to have atrial tachyarrhythmia and ventricular tachycardia at the same time. Atrial leads reduce the incidence but will not prevent inappropriate therapy because all algorithms are biased so that VF is rarely, if ever, missed. In one study, the addition of an atrial lead significantly decrease the rate of inappropriate supraventricular tachycardia (SVT) detection (as VT) and the rate of inappropriate therapy delivery, but it did not decrease the rate of ICD shock; however, this is not a consistent finding.3

Redetection and Reconfirmation

Detection of tachycardia, like any other test, follows bayesian statistics. The probability of a truly positive result is directly proportional to the prevalence of the condition in the population being treated. Once a rhythm has been detected with rigorous criteria and treated, the likelihood is that a rhythm that meets the rate criteria will need to be treated again, and quickly, to prevent syncope. Redetection is the process with which the ICD examines whether the programmed therapy was effective in terminating the presenting arrhyhtmia. In most ICDs, the SVT-VT discriminators are not applied during this process, and redetection is based solely on the programmed rate cut off. The quickness and sensitivity of arrhythmias algorithms are enhanced during redetection by reducing the number of intervals required and by eliminating additional criteria. Conversely, it takes time to charge the ICD capacitors, during which time the arrhythmia may have terminated. For this reason, before the delivery of therapy, reconfirmation of the tachycardia is required to prevent shocks when treating patients with nonsustained tachycardia during sinus rhythm. The danger is that the autogain or autothreshold algorithm has failed to detect low-amplitude VF during progressive hemodynamic collapse. Therefore, reconfirmation of rhythms in the VF zone should be used with care. Some devices permit the allocation of committed and noncommitted shocks in the VF or in the ventricular tachycardia and VF rate zones.

Tachyarrhythmia Therapy

Pacing Therapy

All transvenous ICDs provide for bradycardia support with single-, dual-, or triple-chamber stimulation; however, the subcutaneous defibrillator only provides temporary postshock cardiac stimulation. The pacing mode, rate, and stimulation output is often separately programmable after a high-energy shock for a period to provide overdrive suppression of arrhythmias. Although many patients with ICDs receive dual-chambered rate-adaptive pacemakers, a small minority of these patients have an indication or need for pacing. The Dual Chamber and VVI Implantable Defibrillator (DAVID) trial examined the effect of pacing in the DDDR mode compared with ventricular backup pacing in patients who had ICD but no pacing indications. In these patients, DDDR pacing increased the combined end-point of hospitalization for congestive heart failure (CHF) or death.4 In the DAVID-II trial, atrial support pacing at AAI-70 was tested and showed no significant improvement in quality-of-life outcomes compared with VVI-40.5 As a result, patients without clear indications for antibradycardia pacing should be set to ventricular backup pacing. In patients with wide QRS duration and New York Heart Association functional class III or intravenous CHF, biventricular stimulation is appropriate. This therapy could also be applied in patients with less severe CHF or need for antibradycardia pacing.6

In addition, antitachycardia pacing is available in all the current transvnous ICDs, but not in the subcutaneous defibrillator. This therapy consists of the delivery of a short burst of ventricular pacing, at a programmable rate faster than the detected ventricular tachycardia rate. A variety of antitachycardia pacing techniques are available, including simple bursts at a fixed rate and ramp pacing, in which each subsequent paced interval is incrementally shorter (i.e., faster) than the preceding interval. Effectiveness has been demonstrated to be 90% and 72% in terminating slower and fast ventricular tachycardia using antitachycardia pacing programmed empirically. Its effectiveness was 95% when the testing of the antitachycardia pacing modality was performed before discharge. Antitachycardia pacing was also effective in terminating supraventricular tachcyardias for which patients would have received a shock.

For some patients, a drawback to antitachycardia pacing is the potential for acceleration of the tachycardia to greater rates or to VF, so that high-energy shocks are required. Although 4% of patients experience acceleration of the tachycardia by antitachycardia pacing, this was not associated with any increase in mortality or incidence of syncopal events.7,8 Compared to shock therapy, antitachycardia pacing is delivered with no delay from the detection of the tachycardia (no capacitor to charge), with less morbidity (pain and ventricular stunning) and less battery usage. Once antitachycardia pacing therapy fails to terminate the tachyarrhythmia, shock therapy is delivered. In fact, in the EMPIRIC trial, standardized prespecified ICD programming with antitachycardia pacing during slow and fast ventricular tachycardia was at least as effective as physician-tailored programming without an increase in mortality or shock-related morbidity.9

Shock Therapy

The electrical shocks delivered by ICDs arise from the discharge of the capacitors through the heart via the high-voltage electrodes. Capacitor discharge follows an exponential decay (Figure 115-2), dependent on two variables: the capacitance, and the impedance of the electrode/patient body system. As either of these variables is increased, the voltage and current discharge decayed more slowly, producing a flatter waveform. One to three capacitors are commonly used in each generator, each rated at 106 to 480 µF capacitance and capable of a maximum voltage of 350 to 850 V. These capacitors are usually charged in parallel, to a programmable voltage up to the maximum. When discharged, however, these capacitors are configured in series, so that the total voltage is doubled or tripled depending on the number of capacitors to a maximum voltage of 750 to 850 V. Although the voltage is multiplied in the series configuration, the capacitance is reduced by the number of capacitors used; therefore, the resultant system capacitance is 120 to 150 µF for most currently available clinical devices. Lowering the capacitance of the system will have favorable clinical outcome by decreasing the discharge time. Because most lead impedance values are between 30 and 70 Ω, voltage decreases by 60% to 90% in 20 ms. The shock waveform (discussed later) is generated by the discharge of the capacitors through the electrode system to the patient. Unlike with antitachycardia pacing, ICD shocks can be associated with increased mortality and morbidity. In the Primary Prevention Parameters Evaluation trial, programming strategies to decrease ICD shocks by avoiding therapy for slow and nonsuatained ventricular tachcyardia and applying antitachycardia even to faster VT were associated with less morbidity including mortality.10

Defibrillation Threshold and Safety Margin

The main determinant of success of defibrillation is the magnitude of the electric field generated across the heart, which is proportional to the spatial derivative of the stored voltage. However, because device companies report their device outputs in energy units (Joules), it has been a trend to refer to effective defibrillation energy in the Joule unit of measure. The simplest measure of defibrillation effectiveness is usually termed the DFT, defined as the lowest delivered shock strength required for defibrillation; however, the clinical predictive value of defibrillation testing is limited by the probabilistic nature of defibrillation. The DFT is often determined in a step-down manner, in which progressively lower-intensity shocks are delivered after VF is induced, and the lowest successful shock strength is defined as the DFT. Alternatively, shocks can be delivered in a step-up fashion, beginning at low energies. However, the step-up method will have more failed shocks for patients with high energy requirements. Actually, the time of circulatory arrest is shorter for most patients with the step-up method because the charge times are much shorter with lower energy, but longer for the few patients with high energy requirements. A third method of estimating defibrillation efficacy has been validated, in which test shocks are delivered to the vulnerable period of the cardiac cycle, near the peak of the surface T wave. Theoretically, low-energy shocks delivered during this phase of repolarization will produce VF caused by the resultant dispersion of repolarization times. If a test shock delivered during this period fails to result in VF, it is deemed to be above the DFT; subsequently, lower shock intensities can then be tested until VF occurs. This concept is termed the upper limit of vulnerability and has the advantage of inducing VF only once during the testing period. The shock of the lowest energy that does not induce VF is considered a similar indicator of defibrillation efficacy. The disadvantage of this method is that the ability of the ICD to detect VF correctly and the subsequent decision-making algorithm to delivery VF therapy will not have been tested.

The concept of the DFT is overly simplistic, because electrical defibrillation is best described as a sigmoid-shaped probability function (Figure 115-3). Nonetheless, the step-down DFT is a convenient clinical measurement to obtain during implantation and generally correlates with a probability of success of approximately 70% to 80%. The probability of success with this approach has significant clinical relevance, because appropriate implantation of ICDs requires a high confidence that the first shock will be successful. Historically, using epicardial leads and monophasic waveforms, a safety margin (a margin between the DFT and the maximal output of the ICD) of 10 J was considered minimal implantation criteria. More recent studies have suggested that, with biphasic shocks and NTLs, a margin of 1.9-fold the DFT energy provides a 95% probability of success. It is important to note that DFT testing is device- and lead-specific because the delivered energy is not the only parameter required to ensure successful defibrillation. Other parameters include lead/body impedance, waveform tilt, truncation, pulse duration, and positive and negative-phase peak voltage; these can all affect defibrillation success and are usually device- or lead-specific.11,12

Monophasic Waveforms

The simplest defibrillator waveform is the monophasic waveform, in which the capacitor discharge is truncated before complete discharge of the capacitor, because the terminal voltage tail may prove proarrhythmic. Monophasic waveforms were used in early ICD generators and were effective in epicardial systems. The magnitude of a monophasic (or biphasic) shock is generally described by both the amplitude (peak voltage or current) and tilt of the waveform (see Figure 115-2). For example, a waveform whose amplitude is reduced to half the initial value has a 50% tilt.

When the NTL was first developed and monophasic waveforms were used, the right ventricular coil electrode was configured as the cathode (negative), whereas the superior vena cava coil or subcutaneous patch electrode was configured as the anode (positive). When subsequent animal and clinical studies were performed comparing the right ventricular cathodal configuration with the right ventricular anodal configuration, the anodal right ventricular electrode resulted in a significantly lower DFT in patients and a lower need for supplementary electrodes.

Biphasic Waveforms

The most beneficial ICD innovation, from a viewpoint of defibrillation efficacy, was the development of the biphasic defibrillation waveform (Figure 115-4). This waveform consists of the capacitor discharge divided into two phases of opposite polarity. The first phase is identical to a monophasic waveform (although usually of a shorter duration) before the capacitor discharge is truncated. The electrical connection within the generator is then quickly reversed (usually within 2 to 3 µs), and the second phase is discharged in the opposite polarity for an additional period (usually 3 to 6 ms). Table 115-2 lists the biphasic waveform characteristics currently available in clinical ICDs.

The basic physiology of defibrillation and its relationship to best or optimized biphasic waveform is well studied. Biphasic waveform characteristics in animals identified two factors that appear to maximize defibrillation efficacy as follows: (1) the first phase should be longer than the second phase; and (2) the polarity of the first phase appears to be important only at phase 1 durations greater than 10 ms or when the second phase duration is greater than the first phase.

Additional studies have examined the effectiveness of other phase characteristics. One interesting modification involves doubling the second phase initial voltage by switching the capacitance from series to parallel between the first and second phases. Doing so also reduces the capacitance of the second phase by half, resulting in an increased tilt and a more rapid decay of voltage; animal studies have suggested that this improves the defibrillation efficacy. Multiple recent studies have reported improved defibrillation thresholds with modification of the tilt percentages. The results of these studies were then used to modify the preset tilts in the current available devices. Some devices also have the capacity to specify the duration of the first and second phase of the biphasic shock, which determines the tilt. Although this specificity can make a difference in a few patients, in general it is not required.

System Malfunction

ICDs and their lead systems are complex medical device systems. System integrity is challenged by several factors, including high temperature, high osmolarity, high salt environment, and continuous mechanical stress from the beating heart and body movement. Despite these challenges, the “failure prediction” is small—less than 1% over 5 years. ICD safety alerts or recalls are usually generated in response to a higher than normal projected risk of failure or if the failure would subject the patient to a serious adverse consequence. System failure could occur in the device battery, device electronics, or lead design and interact with the implantation techniques and patient characteristics.

Fortunately, primary battery failure is extremely rare. More often, battery depletion is caused by excessive power drain from pacing, capacitor charge, and electrogram storage. High pacing thresholds and lead insulation failure are additional causes of excessive current drain. On the other hand, repetitive capacitor charges owing to nonsustained ventricular tachycardia or temporary ventricular oversensing can cause important reductions in battery longevity. In 2012, Medtronic identified a rare (0.04% to 0.15%) dysfunction resulting in a greater-than-expected drop in the Entrust ICD battery voltage. This drop resulted in shortened predicted longevity of the ICD and quicker transition from ERI to end of life. Usually these high-current drain issues can be managed with increased frequency of battery assessments.

In 2005, the U.S. Food and Drug Administration (FDA) issued a class I recall (reasonable probability that the use of these devices will cause serious adverse health consequences or death with failure) regarding the use of Guidant PRIZM 2 DR CONTAK Renewal and CONTAK Renewal 2 devices because of failure deliver a life-saving therapy as a result of short circuiting that occurred across the space between the anode (+) backfill tube and the cathode (−) DF feedthrough wire. In 2005, there was also a voluntary class II recall (risk of adverse health consequences or death is remote) for the Medtronic Marquis regarding a battery short circuit.

Failure of transvenous high-voltage ICD leads could be the result of a combination of multiple factors: patient-related, operator-related, and engineering design–related factors. Patient-related factors include thoracic inlet syndrome, Twiddler syndrome, and certain activities like bench pressing. Operator methods that have contributed to lead failure include subclavian or second rib approach, venous access, poorly applied suture sleeves, excessively tight lead coiling within the pocket, over-torquing of the lead, and iatrogenic damage to the leads during implantation. Lead-related factors that can contribute to lead failure include lead engineering and suture design flaws.

Sprint Fidelis ICD leads (Medtronic) were the subject of class I recall because of higher-than-expected conductor fracture rate. There was also evidence of increased risk of high-voltage conductor fracture if a pace-sense conductor fracture has previously occurred; therefore, it is recommended to implant a new high-voltage lead instead of a pace/sense lead if a lead fracture of any type has occurred. Use of a lead integrity algorithm is recommended: extending arrhythmia detection, activating audible alarms, and using quick alert notification to monitor fully functional leads. A lead integrity algorithm could provide most patients with a 3-day warning before inappropriate shocks.

St. Jude Medical Riata and Riata ST leads were the subjects of class I recall because of higher-than-expected erosion in the silicone coating around its conductors. Insulation breaks seen in these leads could be caused by an inside-out or an outside-in abrasion (see Figure 115-5). The latter often occur close to the pocket and are often the result of lead contact with another part of the lead or the pulse generator. Although inside-out externalizations of the inner cables are not uncommon and can be visualized easily using fluoroscopy, they do not seem to place patients at significant risk of electrical dysfunction. However, outside-in abrasions are difficult to be identified using fluoroscopy and can result in shorting of the high voltage; this is not unique to the Riata lead family, nor is it a new failure mechanism.

Not all recalled elements needs to be replaced. Multiple issues should be addressed before replacement, including the risk of keeping the recalled unit, risk of replacement (infection and extraction), risk of failure of the new element to be replaced, and the age of the unit to be replaced.

Recent and Future Directions

Remote Monitoring

The development of the device monitoring systems with or without wandless communication has permitted the early diagnosis of arrhythmias and system malfunction from the patient’s home and with few limits to geographic location or distance from the monitoring physician. It has the potential to safely decrease clinic visits by 40% as seen in the TRUST trial.13 The CONNECT trial showed that remote monitoring with automatic clinician alerts can reduce the time between clinical events and clinical decisions, as well as the length of hospitalization stay. The ECOST trial showed that remote monitoring allows for early and reliable dectection of ICD lead failure and can reduce inappropriate ICD shocks and save battery life. However, other trials like the Tele-HF and EVATEL, did not show any significant benefit of remote monitoring on hospital readmission, all-cause mortality, and major clinical events. Additional studies are needed to evaluate whether monitoring will result in improvements in outcomes or reduction in heath care expendature.

Long-Distance Telemetric Communication

Long-distance telemetric communication (wandless telemetry) in these devices is sometimes accompished through Medical Implant Communications Service, which is an ultra-low-power mobile radio frequency service designed for transmitting diagnostic and therapeutic information from implanted medical devices. Its band operates between 402 and 405 MHz. This bandwidth does not need any licensing and is shared by different manufacturers (currently Medtronic, St. Jude Medical, and Biotronik, Portland, Oregon). Each manufacturer has applied mechanisms to avoid interference and cross communication with other devices. There is another frequency spectrum in use for long-range telemetry by Boston Scientific (St. Paul, MN), the Industrial, Scientific, and Medical band (902 to 928 MHz). A frequency of 914 MHz is used for communication from the ICD to the home monitor, which is also coupled with Bluetooth communication with a weight scale and blood pressure cuff.

Investigational Shock Waveform

Multiple triphasic waveform patterns were tested in animals and failed to improve shock efficacy in terminating ventricular fibrillation over biphasic waves.14 However, multiple serial monophasic shocks were shown to terminate monomorphic ventricular tachycardia with less total energy requirement and peak voltage compared to a single monophasic or biphasic shock in a canine myocardial infarction model.15 If these findings could be replicated in humans, it could translate into prolonged battery life, less pain, and less myocardial stunning and necrosis from the ICD shocks.

Subcutaneous Implantable Cardioverter Defibrillators

Subcutaneous ICDs were developed to address some of the problems and limitations of transvenous lead system ICDs (see Figure 115-6). These systems include but are not limited to: myocardial perforation and tamponade, vascular thrombosis and dissection, pulmonary embolism, brachial plexus injury, and percutaneous lead extraction. The Q-TRAK subcutaneous ICD lead (Boston Scientific) has two sensing electrodes separated by an 8-cm defibrillation coil producing three sensing vectors when connected to the defibrillation can. The lead is tunneled subcutaneously parallel to the left sternal border and is extended to connect to the subcutaneous ICD (SQ-RX Pulse Generator, Boston Scientific) that is placed in the anterior or midaxillary region around the fifth or sixth intercostal spaces. The ICD has a lithium-manganese oxide battery that can deliver an 80-J shock with a fixed 50% tilt. Cardiac signal sensing is automatically selected from the three available sensing vectors (distal-proximal electrodes, distal electrode-can, and proximal electrode-can) based on the analysis of the signal amplitude and signal-to-noise ratio. Arrhythmia detection algorithm uses morphology, beat-to-beat morphology, and QRS width analysis to define treatable arrhythmias. This algorithm is highly sensitive for the detection of ventricular arrhythmias similar to algorithms found in transvenous ICDs.16

Because there is no venous access, this approach is attractive for patients with compromised vascular access (AV fistulas, postmastectomy, mechanical tricuspid valve, congenital heart disease) and could be useful for traditional ICD indications as long as no bradycardia, CRT, or antitachycardia pacing is required. Extraction of this system should also be safer because of the subcutaneous location of the lead. The limitations of this system stem from its inability to provide pacing therapy. It is not indicated in patients who require bradycardia therapy or biventricular pacing. In addition, its role could be limited in patients who require frequent therapies for ventricular tachycardias that are amenable to antitachycardia pacing.

Magnetic Resonance Imaging–Conditional Implantable Cardioverter Defibrillators

Several interactions can be seen during magnetic resonance imaging scanning of patients with ICDs: displacement of ferromagnetic material, reed switch interaction, magnetic saturation of the ICD transformer, heat-induced myocardial necrosis, myocardial stimulation, device reset, and oversensing and undersensing of ventricular arrhythmias. There is no magnetic resonance imaging–conditional ICD (FDA designation) available, although CE Mark for use in the European Union has been designated for one unit. Displacement was addressed by reducing the ferromagnetic components of the ICD. The undesirable interactions seen with the reed switch were avoided by replacing it with a Hall sensor. The internal circuitry was isolated to prevent unintentional ICD reset or reprogramming. Finally, the lead design was modified to decrease polarization, improve evoked response sensing, and improve heat dispersion, thus preventing lead tip heating.

Adjunctive Sensor Technology

Newer devices have incorporated special adjunctive features like intrathoracic impedance monitoring, left atrial pressure monitoring, and right ventricular pressure monitoring. Variation of intrathoracic impedance, measured by the subthreshold electrical impulse between the can and the right ventricular coil with or without a left ventricle lead, correlates with changes in pulmonary fluid accumulation. Because fluid is a good conductor, decompensated heart failure often leads to a decrease in intrathoracic impedance. However, decreases in impedance are also seen in other clinical scenarios, such as pleural or pericardiac effusion, pneumonia, and increased intraabdominal pressure, shortly after device implantation. Studies are currently underway to evaluate the effects on heart failure mortality and hospitalization.

Implantable right ventricular or left atrial pressure sensors, which allow for continuous hemodynamic monitoring, have also been used in ICDs. These pressure measurements can be used to adjust medical therapy. The hope is to prevent acute heart failure decompensation with pulmonary edema or complications related to overtreatment.

References

1. Gold, M, Val-Mejias, J, Lennan, RB, et al. Optimization of superior vena cava coil position and usage for transvenous defibrillation. Heart Rhythm. 2008; 5:394–399.

2. Cesario, D, Bhargava, M, Valderrábano, M, et al. Azygos vein lead implantation: a novel adjunctive technique for implantable cardioverter defibrillator placement. J Cardiovasc Electrophysiol. 2004; 15(7):780–783.

3. Friedman, PA, McClelland, RL, Bamlet, WR, et al. Dual-chamber versus single-chamber detection enhancements for implantable defibrillator rhythm diagnosis. The Detect Supraventricular Tachycardia. Study. Circulation. 2006; 113(25):2871–2879.

4. The DAVID Investigators. Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator. JAMA. 2002; 288:3115–3123.

5. Wilkoff, BL, Kudenchuk, PJ, Buxton, AE, et al. The DAVID (Dual Chamber and VVI Implantable Defibrillator) II trial. J Am Coll Cardiol. 2009; 53:872–880.

6. Moss, AJ, Hall, WJ, Cannom, DS, et al. Cardiac-resynchronization therapy for the prevention of heart-failure events. N Engl J Med. 2009; 361(14):1329–1338.

7. Wathen, MS, DeGroot, PJ, Sweeney, MO, et al. Prospective randomized multicenter trial of empirical antitachycardia pacing versus shocks for spontaneous rapid ventricular tachycardia in patients with implantable cardioverter-defibrillators. Pacing Fast Ventricular Tachycardia Reduces Shock Therapies (PainFREE Rx II) trial results. for the PainFREE Rx II Investigators. Circulation. 2004; 110:2591–2596.

8. Wathen, MS, Sweeney, MO, DeGroot, PJ, et al. Shock reduction using antitachycardia pacing for spontaneous rapid ventricular tachycardia in patients with coronary artery disease. Circulation. 2001; 104:786–801.

9. Wilkoff, BL, Ousdigian, KT, Sterns, LD, et al. A comparison of empiric to physician-tailored programming of implantable cardioverter-defibrillators: results from the prospective randomized multicenter EMPIRIC trial. J Am Coll Cardiol. 2006; 48(2):330–339.

10. Wilkoff, BL, Williamson, BD, Stern, RS, et al. Strategic programming of detection and therapy parameters in implantable cardioverter-defibrillators reduces shocks in primary prevention patients: results from the PREPARE (Primary Prevention Parameters Evaluation) Study. J Am Coll Cardiol. 2008; 52:541–550.

11. Sweeney, MO, Natale, A, Volosin, KJ, et al. Prospective randomized comparison of 50/50% versus 65/65% tilt biphasic waveform on defibrillation in humans. Pacing Clin Electrophysiol. 2001; 24:60–65.

12. Schauerte, PN, Ziegert, K, Waldmann, M, et al. Effect of biphasic shock duration on defibrillation threshold with different electrode configurations and phase 2 capacitances: prediction by upper-limit-of-vulnerability determination. Circulation. 1999; 99:1516–1522.

13. Varma, N, Epstein, AE, Irimpen, A, et al. Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator follow-up: the Lumos-T Safely Reduces Routine Office Device Follow-up (TRUST) trial. Circulation. 2010; 122(4):325–332.

14. Davis, R, Malkin, R. Simultaneous comparison of many triphasic defibrillation waveforms. Open Biomed Eng J. 2012; 6:1–4.

15. Janardhan, A, Li, W, Gutbrod, S, et al. Low-energy three-stage electrotherapy delivered through implantable leads significantly reduces the cardioversion threshold in a canine model of persistent atrial fibrillation. Circulation. 2011; 124:A15917.

16. Gold, MR, Theuns, DA, Knight, BP, et al. Head-to-head comparison of arrhythmia discrimination performance of subcutaneous and transvenous ICD arrhythmia detection algorithms: the START study. J Cardiovasc Electrophysiol. 2012; 23(4):359–366.