Equipment for paediatric intensive care

Published on 27/02/2015 by admin

Filed under Anesthesiology

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 1 (1 votes)

This article have been viewed 10253 times

Chapter 104 Equipment for paediatric intensive care

This chapter discusses equipment and its performance required in paediatric and neonatal intensive care. How to use equipment is not discussed. Examples of equipment are given, but exhaustive lists are not intended. Only essential equipment is considered.

Not discussed includes equipment for aerosol therapy, anaesthesia, bronchoscopy, cardiovascular monitoring, cardiac output measurement, echocardiography, ultrasonography, electroencephalographic monitoring, extracorporeal membrane oxygenation, ventricular assistance, home mechanical ventilation, hyperbaric treatment, negative pressure ventilation, nitric oxide therapy, phrenic nerve pacing, mechanical devices for treatment of shock, suction devices and tracheostomy–cricothyrotomy.

All equipment should be purchased, used and maintained in accordance with recognised local standards and manufacturers’ instructions. Medical and nursing staff should read the manuals accompanying all items of equipment and receive instruction on safe use.

Numerous items including laryngoscopes, resuscitation bags and ventilator circuits are available as disposable items in response to a market demand for infection safety and reduction of sterilisation costs. The user should subject such items to measures of quality control.

AIRWAY MANAGEMENT

ENDOTRACHEAL TUBES

TUBE LENGTH (DEPTH)

The endotracheal tube tip should be located in the mid-trachea so that the risks of accidental extubation and endobronchial intubation are minimised. Correct positioning of the tube should always be checked by auscultation in the axillae, by chest X-ray immediately after intubation and daily thereafter.1 At intubation, visualisation of the depth of the tube passed through the cords should be a guide only, since the tube advances when the head is released from extension as the laryngoscope is removed.2

The lengths (depths) at which orotracheal and nasotracheal tubes can be chosen initially on the basis of weight or gestation in premature neonates are given in Table 104.1, and on the basis of age for infants and children in Table 104.2. The lengths are at the lips for oral tubes and at the exit from the nose for nasal tubes. For oral tubes, a guide to correct depth for 2 years and over is given by the formula: (age in years/2 + 12 cm), while for nasal tubes the formula is: (age in years/2 + 15 cm). The depth of insertion of any tube should be documented or marked on the tube.

SUCTION CATHETERS

Sizes 5, 6, 8, 10 and 12 FG should be available. The suction catheter should be large enough to remove secretions but not too large to occlude the endotracheal/tracheostomy tube lumen (Table 104.3). Many suitable types are available – differing in number and location of suction ports (end and side ports are needed) and whether the tip is angled or straight. Angled tips facilitate entry into the left main bronchus. Satin-finish types with smooth port edges are needed. Single use suction catheters may be reused up to 24 hours without adding to the risk of pneumonia.3 Closed suction systems permit suction without disconnection from ventilator as may be indicated during nitric oxide therapy, severe lung disease or frequent suction.

Table 104.3 Recommended suction catheters

Internal diameter (mm) of ETT or tracheostomy Recommended suction catheter (FG)
2.5 5
3.0 6
3.5–5.5 8
6.0–6.5 10
7.0 and larger 12

ETT, endotracheal tube.

MECHANICAL VENTILATION

RESUSCITATOR ‘BAGGING’ CIRCUITS

Manual pulmonary inflation is often necessary via a mask or endotracheal/tracheostomy tube. Two types of device are available: flow-inflated and self-inflated bags.

SELF-INFLATED BAGS

These bags are designed to give positive-pressure ventilation attached to either a mask or endotracheal/tracheostomy tube. Rebreathing is prevented by one-way duck-bill valves, spring disk/ball valves or diaphragm/leaf valves.

The Laerdal bag series (infant, child, adult) typify these devices. A pressure-relief valve (infant and child size) opens at 35 cmH2O (3.5 kPa). A pressure monitor can be incorporated in the circuit. Supplemental oxygen is added to the ventilation bag, with or without attachment of a reservoir bag, whose movement may serve as a visual monitor of tidal volume during spontaneous ventilation when intubated. However, the valve may offer too much resistance for spontaneous ventilation and they should not be used to provide supplemental oxygen to a spontaneously breathing patient via a mask placed loosely over the face.

With Laerdal and Partner bags, negligible amounts of oxygen (0.1–0.3 l/min) issue from the patient valve when 5–15 l/min of oxygen is introduced into bags unconnected to patients.4 The valve is unlikely to open unless the mask is sealed well on the face. The delivered oxygen concentration is dependent on the flow rate of oxygen, use of the reservoir bag, and the state of the pressure relief valve (whether open or closed).

Other self-inflated bags include the Combibag, Ambu, AirViva and Hudson RCI.

MECHANICAL VENTILATORS

CONVENTIONAL VENTILATORS

Some ventilators are designed specifically for neonates and infants. A few adult ventilators are also satisfactory. In addition to general requirements of ventilators, the paediatric ventilator should have:

A generic classification of ventilators, their operational characteristics and their modes of ventilation has been proposed.5 It is described here briefly to enable functional description and comparison of some ventilators.

Ventilators may be operationally classified as pressure, volume, flow or time controllers. These are the independent or control variables. Since only one variable can be controlled at a given time, i.e. predetermined, the remaining features of the waveform are dependent variables. If pressure, volume or flow is not predetermined, the ventilator is presumed to be a time controller in which only the timing of the inspiratory and expiratory phases is controlled.

These concepts allow understanding of any mode of ventilation and interpretation of bedside pulmonary mechanics, i.e. resistance, compliance, time constants, etc. The manner of controlling the independent variable may be an open-loop system, in which no feedback occurs during inspiration to achieve the target goal, or a closed-loop or feedback or servo-controlled system in which the independent variable is modified according to progress during inspiration.

During inspiration, dual control may be used and may be altered from one independent variable to another. For example, in the VAPS mode (volume assured pressure support mode) of the Bird VIP Gold ventilator, control is switched from pressure to flow control if a preset tidal volume has not been achieved. In another example using the Pmax feature of the Drager Evita 4, if in flow control a pressure limit has been reached, control is passed to pressure while the tidal volume is monitored and the inspiratory time increased until the preset tidal volume has been achieved.

PHASE VARIABLES

The manner of initiation of inspiration (triggering), sustaining inspiration (limit variables), initiation of expiration (cycling) and sustaining expiration are the phase variables. Initiation of inspiration, or triggering, may be dictated by machine or patient. Usual triggers are the elapse of time or generation of pressure or flow by the patient. Flow triggering during spontaneous ventilation is more energy efficient and better tolerated than other variables, particularly when the flow sensor is located at the endotracheal tube. Mandatory breaths are all those triggered by a preset frequency or minimum minute ventilation, or terminated by a preset frequency or tidal volume. Spontaneous breaths are those initiated by the patient and terminated by lung mechanics or ventilatory drive. The quantification of the inspiratory phase is by time, pressure or volume. Initiation of expiration (termination of inspiration) or cycling is achieved by elapse of time, pressure, volume or flow. Expiration is usually sustained by pressure alone.

Modes of ventilation may be described on the following generic basis. It is derived from the controller variables. It should not be assumed that a manufacturer’s mode terminology accurately defines its functional attributes or is intuitively obvious. Terminology is not necessarily interchangeable between different manufacturers. Although some new modes are genuine inventions, another purpose appears to be differentiation of product from competition. The capabilities of any ventilator can be defined by the following:

If conditional variables exist, these act to change the pattern of ventilation – for example, to give a sigh breath based on elapse of time, or switching from patient-triggered breaths to machine-triggered breaths in the SIMV mode or mandatory minute ventilation modes.

Volume/flow-control modes are suitable for conditions in which lung compliance and resistance are variable. In these circumstances, the peak inspiratory pressure attained is a valuable observation. However, when a pressure limit is imposed, such information is lost. A mode consisting of flow control, time triggering, pressure limitation and time cycling is commonly used because of the relative ease of operation. However, relatively small leaks may severely compromise tidal volume, mean airway pressure and oxygenation. Large leaks may also delay attainment of preset pressure until late in the inspiratory cycle. Thus, it is helpful to display the volume and pressure waveforms. Some ventilators do, however, provide leak compensation (e.g. Bird VIP Gold). Gas flow should be set above 2 l/kg per min to achieve the preset pressure early in the inspiratory period. This flow rate is also suitable for peak inspiratory flow rate during spontaneous respiratory effort with intermittent mandatory ventilation.

Pressure in the circuit should be monitored in close proximity to the endotracheal tube. Changes in endotracheal tube resistance from encrustation or kinking have a major adverse effect on tidal volume.6

Neonates and infants have small tidal volumes (5–50 ml), and are therefore prone to alveolar hypoventilation when variable air leaks occur around the endotracheal tube and from the ventilator–humidifier circuit. Moreover, the internal compliance and compressible volume of the circuit may exceed the tidal volume. Small tidal volumes are difficult to set and maintain when the lung compliance and resistance are abnormal.7

Most neonatal intensive care units (ICUs) use pressure-control modes which in hyaline membrane disease achieves adequate oxygenation with minimal barotrauma.8 The pressure-control mode is more appropriate than the volume-control mode for neonates and infants 0.5–5.0 kg body weight. Ventilators, which are pressure controllers, deliver a variable flow dependent on lung and airway characteristics.

Improvements in ventilatory management of newborns are observed with a new generation of ventilators, which utilise changes in airway pressure, airway gas flow or abdominal movement to trigger inspiration, and which have ultrashort trigger delays.9 Previously, ventilators were unable to cope with the very short inspiratory times (mean 0.31, SD 0.06 seconds) of newborns with respiratory distress syndrome.10 Patient-triggered ventilation (PTV) permits synchronous ventilation with better gas exchange, prevents the patient fighting the ventilator and reduces the need for neuromuscular paralysis. 11,12 Some ventilators suitable for paediatric use are the Bear Cub 2001, Bear Cub 750vs, Bird T Bird, Bird VIP Gold, Drager Babylog 8000 (plus), Drager Evita 4, Evita XL, Infant star 950, Sechrist IV-100B, Sechrist IV-200 SAVI, Siemens Servo 900C, Siemens Servo 300, Siemens Servo-i, Viasys Avea, and SLE 5000.

HIGH-FREQUENCY VENTILATORS

These provide ventilation at at least four times normal rate and may be classified into five types:5

High-frequency oscillatory ventilation (HFO) at 15 Hz has been shown to achieve better gas exchange with less barotrauma than conventional mechanical ventilation in infants with respiratory distress.13,14 Examples of high-frequency oscillators are the Sensor Medics and 3100A and 3100B (adult) and the Humming series models, of which the latest is Atom Humming V. The latter also offers a synchronised intermittent mandatory ventilation mode. An example of a high-frequency jet ventilator is the Bunnell Life Pulse jet ventilator which is used in conjunction with a conventional ventilator to provide positive end-expiratory pressure (PEEP) and humidified gas for spontaneous ventilation. An example of a high-frequency positive-pressure ventilator is the Nellcor Puritan-Bennett Infant Star 950, which operates as a flow interrupter.

NON-INVASIVE VENTILATION (NIV)

The application of ventilation ‘BiPAP’ (bilevel positive airways pressure) or CPAP by means other than intubation (NIV) has become an essential part of paediatric intensive care practice. Although a proprietary name (Respironics), ‘BiPAP’ has become synonymous with non-invasive ventilation. It is distinct from BIPAP (biphasic positive airway pressure), a mode available in some conventional ventilators in which spontaneous ventilation is permitted during all phases of mechanical ventilation. CPAP or BiPAP may be given by an interface of a nasal, oronasal (full face) mask with harnesses with or without a chin strap (headgear). Other interface devices such as mini-masks, nasal pillows, nasal cushions, mouthpiece/lipseal and helmets are less common in paediatrics.

CPAP is commonly used to relieve upper-airway obstruction or improve oxygenation in lung disease. BiPAP is commonly used to provide ventilation in chronic hypoventilatory states such as congenital central hypoventilation syndrome, neuromuscular diseases, pretransplantation cystic fibrosis and in acute conditions such as asthma, pneumonia and pulmonary oedema. Both CPAP and BiPAP are used before resorting to invasive ventilation and after extubation. Although stand-alone, BiPAP machines are primarily intended for use in adult patients with obstructive sleep apnoea and are not generally intended for use as life support. Consequently, many machines are not licensed to provide ventilation via endotracheal tube or tracheostomy but this capability is a distinct advantage and may be provided as an optional mode in mechanical ventilators.

The defining characteristics of BiPAP are the automatic breath-by-breath compensation for air leaks and the ability to trigger and cycle by flow or time in synchronisation with patient effort. The devices can be set to provide a spontaneous (assisted) mode, timed (mandatory) mode or a combined mode. When the inspiratory positive airway pressure (IPAP) is set the same as the expiratory positive airway pressure (EPAP), CPAP is provided. Examples of CPAP machines are Tranquility (Healthdyne) and CPAP S6 series (Resmed). Examples of BiPAP machines are the BiPAP S/T, S/T-D, S/T-D 30, Harmony and Synchrony series (Respironics), the KnightStar 335 (Nellcor-Puritan-Bennett) but not suitable for children < 30 kg, and the Sullivan VPAP II and VPAP II ST (Resmed).

OXYGEN THERAPY AND MONITORING

TRANSCUTANEOUS GAS ANALYSERS

Transcutaneous gas tensions in paediatric patients equate well with arterial gas tensions. However, PaO2 is underestimated by PtcO2 during mild hyperoxaemia in neonates.17 Transcutaneous gas values are dependent on both arterial gas tensions and on blood flow. During normovolaemia with adequate flow, PtcO2 reflects PaO2 and PtcCO2 reflects PaCO2. However, during circulatory failure, a reduction in PtcO2 reflects a reduction in flow, provided PaO2 is normal.18 Likewise, during circulatory failure, PtcCO2 exceeds PaCO2 and is flow dependent.19 Typical transcutaneous dual function single electrodes are Hewlett Packard Virida tcpO2/tcCO2, Sensormedics FasTrac SaO2/CO2 monitor.

Continuous pulse oximetry (SpO2) has evolved to be an integral part of monitoring in neonatal and paediatric intensive care. Transcutaneous pulse oximetry is more convenient than transcutaneous partial pressure measurement. Each patient should have a pulse oximeter. The value of pulse oximetry is detection of hypoxaemia, not hyperoxaemia. They cannot be used accurately to titrate oxygen therapy, which requires blood or transcutaneous partial pressure measurement. The accuracy of pulse oximetry at low SaO2 (< 75%) is poor.20 Numerous devices are available.

APNOEA MONITORS

Infant apnoea is often due to upper-airway obstruction, seizures or gastro-oesophageal reflux,21 although it may be idiopathic and is commonly associated with prematurity. Short apnoeic spells (< 15 seconds) are normal during sleep and feeding. Significant apnoea is associated with bradycardia, cyanosis and hypotonia. The detection and treatment of apnoea in intensive care are dependent on cardiovascular monitoring and constant nursing observation. In the home or ward, where observation is less than that available in the ICU, apnoea may be detected by the use of a pressure sensor in a pad or mattress under the infant, or by impedance pneumography, in which diaphragmatic movement causes a change in resistance between two electrodes on the skin. Apnoea of central origin is detected early by these devices but obstructive apnoea, resulting in paradoxical respiratory effort, is detected late. Some devices incorporate electrocardiogram monitoring and rely upon bradycardia as evidence of hypoxaemia, which is a useful but late sign in obstructive apnoea, when both hypoxaemia and movement of the diaphragm are present. False-positive alarms are not uncommon.

THERMOREGULATION

HUMIDIFIERS

Adequate humidification of ventilator circuits is of utmost importance in the paediatric patient. Unless adequately humidified, the relatively narrow endotracheal and tracheostomy tubes are prone to encrustation and blockage. Insensible water and heat loss are also prevented with adequate humidification. High-flow heated humidifiers provide near 100% humidification (44 mg H2O/l gas, 47 mmHg) at body temperature and are preferred. They are easily adapted for use into ventilator circuits, face and tracheostomy masks, T-piece circuits and oxygen head boxes. When used in ventilator circuits, it is important to realise that they contribute to internal compliance and a compressible volume, which produce errors in both volume-preset and pressure-preset ventilators. Ideally, the compliance should not exceed 1.0 ml/cmH2O for infants less than 10 kg body weight.24

Several methods exist to maintain the water level including an airlock system, parallel-fill system, float-system and a pinch-clamp system.5 Systems that automatically maintain a constant level are preferred for the aforementioned reasons. Incorporation of servo-control is also important. A thermistor probe incorporated into the proximal inspiratory circuit and heating of the inspiratory limb ensure that gas is delivered to the patient at body temperature. However, because some heat loss may occur along the inspiratory limb and much more along the expiratory limb (which is not heated), significant condensation (rain-out) occurs. A water trap must be incorporated to gather ‘rain-out’ and the circuit, particularly the expiratory limb, positioned in a dependent manner to prevent aspiration by the patient. Such water, although potentially contaminated by airway bacteria, does not pose nosocomial infection risk and may be allowed to drain into the humidifier where high temperature inhibits bacterial growth. Circuits may be changed infrequently. Examples of servo-controlled heated humidifiers are Puritan-Bennett Cascade II and Fisher and Paykel MR600, 700, 720, 850 series.

Condenser-type humidifiers or ‘artificial noses’ have a role in intensive care. They are passive humidifiers capturing exhaled water vapour and heat, permitting their reuse on inhalation (heat and moisture exchangers, HME). Numerous devices exist, some incorporating filters and/or hydroscopic materials leading to subclassifications of heat and moisture exchanging filter (HMEF), hydroscopic heat and moisture exchanger (HHME) and hydroscopic heat and moisture exchanging filter (HHMEF). They are often added to tracheostomy or endotracheal tubes of spontaneously breathing patients or inserted into the inspiratory limb of a ventilator circuit for short periods, e.g. for transport. They cannot replace full humidification. HMEs provide approximately one-quarter to one-third relative humidity whereas HHME/HHMEFs provide approximately one-half to three-quarters relative humidity.5 Problems with these devices include deterioration of performance with time, resistance and dead space. The appropriate size for the patient (adult, paediatric, neonatal) must be used. They are not suitable for use if the patient is hypothermic, has thick or bloody secretions or requires frequent changes. They should never be used in conjunction with heated water humidifiers because of potential circuit blockage. They should not be moistened and the addition of oxygen, although sanctioned by some manufacturers, will compromise performance.

Unheated humidifiers of the passover, bubble, bubble diffuser and jet types are inefficient and have limited applicability in the intensive care setting. Other types of humidifiers such as jet nebulisers and ultrasonic nebulisers have little role in intensive care.

MISCELLANEOUS EQUIPMENT

REFERENCES

1 Greenbaum DM, Marshall KE. The value of routine daily chest X-rays in intubated patients in the medical intensive care unit. Crit Care Med. 1982;10:29-30.

2 Bosman YK, Foster PA. Endotracheal intubation and head posture in infants. South Afr Med J. 1977;52:71-78.

3 Scoble MK, Copnell B, Taylor A, et al. Effect of reusing suction catheters on the occurrence of pneumonia in children. Heart Lung. 2001;30:225-233.

4 Carter BG, Fairbank B, Tibballs J, et al. Oxygen delivery using self-inflating resuscitation bags. Pediatr Crit Care Med. 2005;6:125-128.

5 Branson RD, Hess DR, Chatburn RL. Respiratory Care Equipment. Philadelphia: Lippincott Williams and Wilkins, 1999.

6 Abrahams N, Fisk GC, Vonwiller JB, et al. Evaluation of infant ventilators. Anaesth Intensive Care. 1975;3:6-11.

7 Simbruner G, Gregory GA. Performance of neonatal ventilators: the effects of changes in resistance and compliance. Crit Care Med. 1981;9:509-514.

8 Reynolds EOR, Taghizadeh A. Improved prognosis of infants mechanically ventilated for hyaline membrane disease. Arch Dis Child. 1974;49:505.

9 Greenough A, Milner AD. Respiratory support using patient triggered ventilation in the neonatal period. Arch Dis Child. 1992;67:69-71.

10 South M, Morley CJ. Respiratory timing in intubated neonates with respiratory distress syndrome. Arch Dis Child. 1992;67:446-448.

11 Hird MF, Greenough A. Patient triggered ventilation using a flow triggered system. Arch Dis Child. 1991;66:1140-1142.

12 Cleary JP, Bernstein G, Mannino FL, et al. Improved oxygenation during synchronized intermittent mandatory ventilation in neonates with respiratory distress syndrome: a randomized, crossover study. J Pediatr. 1995;126:407-411.

13 Ogawa Y, Miyasaka K, Kawano T, et al. A multicenter randomized trial of high frequency oscillatory ventilation as compared with conventional mechanical ventilation in preterm infants with respiratory failure. Early Human Dev. 1993;32:1-10.

14 HiFO study group. Randomized study of high-frequency oscillatory ventilation in infants with severe respiratory distress syndrome. J Pediatr. 1993;122:609-619.

15 Shann F, Gatachalian S, Hutchinson R. Nasopharyngeal oxygen in children. Lancet. 1988;2:1238-1240.

16 Cole AGH. Small oxygen analysers. Br J Hosp Med. 1983;29:469-471.

17 Martin RJ, Robertson SS, Hopple MM. Relationship between transcutaneous and arterial oxygen tension in sick neonates during mild hyperoxaemia. Crit Care Med. 1982;10:670-672.

18 Tremper KK, Shoemaker WC, Shippy CR, et al. Transcutaneous oxygen monitoring of critically ill adults with and without low flow shock. Crit Care Med. 1981;9:706-709.

19 Tremper KK, Shoemaker WC, Shippy CR, et al. Transcutaneous PCO2 monitoring on adult patients in the ICU and operating room. Crit Care Med. 1981;9:752-755.

20 Carter BG, Carlin JB, Tibballs J, et al. Accuracy of two pulse oximeters at low arterial hemoglobin–oxygen saturation. Crit Care Med. 1998;26:1128-1133.

21 Brown LW. Home monitoring of the high-risk infant. Clin Pediatr. 1984;11:85-100.

22 Hull D, Chellappah G. On keeping babies warm. In: Chiswick ML, editor. Recent Advances in Perinatal Medicine. Edinburgh: Churchill Livingstone; 1983:153-168.

23 Bell EF, Rios GR. Performance characteristics of two double-walled infant incubators. Crit Care Med. 1983;11:663-667.

24 Holbrook PR, Taylor G, Pollak MM, et al. Adult respiratory distress syndrome in children. Pediatr Clin North Am. 1980;27:677-685.

25 Shyam VSR, Tinker J. Recent developments in infusion devices. Br J Hosp Med. 1981;25:69-75.

26 Dickenson JR. Syringe pumps. Br J Hosp Med. 1983;29:187-191.

27 Weiss M, Banziger O, Neff T, et al. Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med. 2000;26:776-779.

28 Lui K, Doyle PE, Buchanan N. Oscillometric and intra-arterial blood pressure measurements in the neonates: evaluation and comparison of methods. Aust Paediatr J. 1982;18:32-34.

29 Savage JM, Dillon MJ, Taylor JFN. Clinical evaluation and comparison of the infrasonde, arteriosonde and mercury sphygmomanometer in measurements of blood pressure in children. Arch Dis Child. 1979;54:184-189.

30 Showman A, Betts EK. Hazard of automatic noninvasive blood pressure monitoring. Anesthesiology. 1981;55:717-719.