Closed-circuit anesthesia

Published on 07/02/2015 by admin

Filed under Anesthesiology

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 2170 times

Closed-circuit anesthesia

Allen B. Shoham, MD

Closed-circuit anesthesia refers to a technique in which, during the delivery of an inhalation anesthetic agent using an anesthesia work station (machine) with a circle system, the adjustable pressure relief valve is closed and the fresh gas inflow is adjusted such that it is just sufficient to match the amount of inhaled anesthetic gas and O2 that is taken up by the patient and the amount of gas removed by any side-stream monitors (for measurement of anesthetic gas concentration, CO2, and O2) (Box 13-1). To prevent hypercapnia, a CO2 absorber must be used; all exhaled gases except CO2 are rebreathed when this technique is used. One must ensure that the circuit is free of any leaks (more than 20 potential sites exist for gases to leak from the circuit/system) (Box 13-2).

Whereas, during administration of most inhalation anesthetics, one monitors only the fractional inspired O2 concentration (FIO2), when using a closed-circuit anesthetic technique, one should monitor both the FIO2 and the expired oxygen concentration (FEO2). Normally, the FEO2 is not significantly different from the FIO2; however, during closed-circuit anesthesia, the FEO2 may be significantly less if the amount of O2 delivered in the fresh-gas inflow is inadequate to meet the patient’s O2 consumption. Oxygen consumption determines the amount of O2 one must deliver through the fresh-gas inflow. Anesthetic uptake, on the other hand, is proportional to blood/gas solubility and the alveolar-venous inhalation anesthetic concentration gradient and cardiac output.

During the use of a closed-circuit anesthetic, it is important that flowmeters and vaporizers have been calibrated to be accurate at low flows and varying circuit pressures. It may be difficult to deliver the required amount of inhalation anesthetic using this technique with certain vaporizers that are inaccurate at low flow rates (see Box 13-2). If this is the case, one can inject the inhalation anesthetic directly into the expiratory limb of the circle system, but this is rarely done.

Some major obstacles must be overcome if anesthesia is induced using a closed-circuit anesthesia system (see Box 13-2). Before tissue uptake of the anesthetic can occur, an adequate amount of the inhalation anesthetic must be in the breathing circuit, the patient’s alveoli, and the arterial circulation. During conventional inhalation anesthetic induction, uptake of the inhalation anesthetic agent is greatest during the initial phases of induction; uptake decreases as venous concentration approaches alveolar concentration of the inhalation anesthetic. During induction with a closed-circuit anesthetic, it is difficult to denitrogenate the patient; the N2 that the patient exhales dilutes the breathing circuit, thereby slowing induction and increasing the risk for awareness and, possibly, resulting in delivery of a hypoxic gas mixture.

During closed-circuit anesthesia, O2 is the preferred carrier gas because its uptake is relatively constant. The administration of N2O makes induction more complicated because its uptake follows a power function and requires constant titration during induction and further increases the risk of delivery of a hypoxic gas mixture. To avoid these difficulties, most clinicians use high flows on induction to denitrogenate the breathing circuit and the patient’s functional residual capacity and to establish adequate anesthetic concentrations prior to switching to a closed system (Figure 13-1).

Maintenance of anesthesia

If N2O is used as one of the carrier gases for the maintenance of anesthesia using a closed-circuit system, one must be particularly vigilant in monitoring O2 concentration. As N2O uptake by the patient decreases, alveolar N2O concentration increases, thereby displacing O2 with a subsequent decrease in O2 concentration within the alveolus; therefore, constant titration of O2 and N2O inflow is essential. Once the inhalation anesthetic gas concentration is at a steady state, the requirement for additional inhalation anesthetic decreases, remaining relatively constant.

Each milliliter of liquid inhalation anesthetic (isoflurane, sevoflurane, desflurane) produces approximately 200 mL of vapor (±10%). Closed-circuit anesthesia can be maintained with a vaporizer or by intermittent injections of inhalation anesthetic at appropriate time intervals.

During the maintenance phase, a constant circuit volume must be maintained by (1) adjusting flows to maintain a constant reservoir-bag size; (2) with ascending bellows, adjusting the flow to be just below the top of the housing of the bellows at end exhalation; and (3) with descending bellows, adjusting the fresh-gas flow to allow the bellows to just reach the bottom of its housing at end exhalation. It is crucial that there are no gas leaks in the circuit or a negative pressure transmitted to the bellows at any time, as this may entrain room air and decrease O2 and the concentration of inhalation anesthetic gas.

Nitrogen accumulates in the breathing circuit over time; therefore, the system should be flushed for a few minutes every hour to denitrogenate the system and prevent production of a hypoxic gas mixture. Accumulation of carbon monoxide, compound A, methane, ethanol, and acetone are other reasons for flushing the system. The clinical consequences of these compounds are unknown; however, they have been proposed to cause an increased incidence of postoperative nausea and vomiting. The clinical significance of compound A is still controversial; therefore, it is not recommended to use sevoflurane in closed-circuit anesthesia.