Blood Gas Sampling, Analysis, Monitoring, and Interpretation

Published on 12/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 4268 times

3 Blood Gas Sampling, Analysis, Monitoring, and Interpretation

Note 1: This book is written to cover every item listed as testable on the Entry Level Examination (ELE), Written Registry Examination (WRE), and Clinical Simulation Examination (CSE).

The listed code for each item is taken from the National Board for Respiratory Care’s (NBRC) Summary Content Outline for CRT (Certified Respiratory Therapist) and Written RRT (Registered Respiratory Therapist) Examinations (http://evolve.elsevier.com/Sills/resptherapist/). For example, if an item is testable on both the ELE and the WRE, it will simply be shown as: (Code: …). If an item is only testable on the ELE, it will be shown as: (ELE code: …). If an item is only testable on the WRE, it will be shown as: (WRE code: …).

Following each item’s code will be the difficulty level of the questions on that item on the ELE and WRE. (See the Introduction for a full explanation of the three question difficulty levels.) Recall [R] level questions typically expect the exam taker to recall factual information. Application [Ap] level questions are harder because the exam taker may have to apply factual information to a clinical situation. Analysis [An] level questions are the most challenging because the exam taker may have to use critical thinking to evaluate patient data to make a clinical decision.

Note 2: A review of the most recent Entry Level Examinations (ELE) has shown an average of 10 questions (out of 140), or 7% of the exam, that cover blood gas sampling, analysis, monitoring, and interpretation. A review of the most recent Written Registry Examinations (WRE) has shown an average of 5 questions (out of 100), or 5% of the exam, that cover blood gas sampling, analysis, monitoring, and interpretation. The Clinical Simulation Examination is comprehensive and may include everything that should be known by an advanced level respiratory therapist.

MODULE A

Blood can be sampled from a systemic artery, pulmonary artery, or “arterialized” capillary to determine a patient’s oxygen and carbon dioxide pressures, acid-base status (pH), and related values. As is discussed later, a pulmonary artery sample is taken to learn a patient’s mixed venous values, and an “arterialized” capillary sample is taken from a neonate when an arterial sample cannot be obtained. Arterial blood gas (ABG) is the term commonly used when discussing drawing a sample of blood from a patient’s systemic artery.

There are three broad, general indications for this recommendation:

Some specific indications follow.

MODULE B

1. Blood gas sampling device selection and preparation

A properly prepared sampling device is essential to obtaining a blood sample that accurately reflects the patient’s physiologic values.

a. Get the appropriate blood gas sampler

Obtaining an ABG sample generally involves selecting a prepackaged, sterile blood gas kit that contains the following:

If a blood gas kit is not available, an appropriate individual needle, 3-mL syringe, and liquid sodium or lithium heparin must be obtained. These should be available at any nursing station or from the respiratory care department.

A capillary blood gas (CBG) sample requires the following:

In addition, if the blood sample cannot be analyzed within 30 minutes, place it into an ice-water mix in a cup to chill the blood.

2. Manipulate arterial catheters by order or protocol (WRE code: IIA20b) [Difficulty: WRE: R, Ap, An]

a. Select the appropriate catheter

An arterial catheter (also called an arterial line) is a flexible catheter that is placed into a peripheral artery for the purposes of sampling blood or continuously monitoring the patient’s blood pressure, or both. The blood-sampling procedure is explained in the following discussion. Chapter 5 contains a discussion on blood pressure monitoring and illustrations of how the monitoring system is assembled. The radial artery is the most common site for catheter insertion. Alternate arterial sites may be used if needed.

An adult’s radial artery is usually catheterized. To do so, a short needle covered with a flexible plastic catheter (an angiocatheter) is selected. Usually the needle is 23 or 24 gauge. After the angiocatheter is inserted into the artery, the needle is removed, and the catheter is left in the artery.

A newborn with a severe cardiopulmonary problem should have the catheter inserted into either of the umbilical arteries. Usually, a long, flexible umbilical artery catheter (UAC) is placed into the patient and advanced into the aorta (see Figure 1-7). If indicated, this should be done as quickly as possible after birth, before arterial spasm prevents the catheter from being advanced. If the patient weighs more than 1250 g, a 6-French (6 Fr) catheter is used; a 3.5 Fr catheter is used if the neonate weighs less than 1250 g. Besides obtaining blood samples and monitoring the blood pressure, one can give the newborn glucose or a blood transfusion through the catheter. A neonatal patient also may have the umbilical vein catheterized.

3. Get a blood sample from an arterial catheter (Code: IB9f and IIIE2b) [Difficulty: ELE: R, Ap; WRE: An]

The steps for obtaining an arterial blood sample include the following:

4. Perform an arterial puncture to obtain a blood sample for analysis (Code: IB9f and IIIE2a) [Difficulty: ELE: R, Ap; WRE: An]

A number of possible variations on the technique exist. The following is a general but thorough listing of the steps and any important related information.

1. Radial artery site

See Figure 3-2 for the basic procedure. Circulation to the hand is stopped by pressing both the radial and the ulnar arteries closed. Releasing the pressure over the ulnar artery should result in the hand flushing within 5 to 15 seconds. This is a positive test result and proves that the ulnar artery has adequate circulation to the hand. If the hand does not flush within 15 seconds of the release of the ulnar artery, the circulation is inadequate, and the radial artery of that wrist must not be punctured. Another site must be evaluated for puncture.

2. Dorsalis pedis artery site

Press down on the dorsalis pedis artery to occlude it. Press on the nail of the great toe so that it blanches. Release the pressure on the nail, and watch for a rapid return of color. This normal test finding confirms that a good blood flow exists through the posterior tibial and lateral plantar arteries. It is safe to draw a sample from the site. A slow return of blood flow indicates poor circulation; another site must be chosen.

h. Draw the blood sample.

5. Perform arterialized capillary blood gas sampling (Code: IB9g and IIIE2c) [Difficulty: ELE: R, Ap; WRE: An]

Occasionally, a sample of arterialized capillary blood from a patient must be obtained for blood gas analysis. The usual clinical situation is a neonate who has a pulmonary problem that warrants evaluation. However, because of the neonate’s small arteries, a sample cannot be drawn. The steps and key points to keep in mind during the sampling procedure follow.

d. Use a pediatric lance to deeply puncture the outer edge of the heel (Figure 3-4). Blood should flow freely without squeezing the area. Wipe away the first blood drop with a sterile gauze pad. The blood will be “dearterialized” if it is squeezed out with venous blood, and the sample will be useless for blood gas analysis.

MODULE C

1. Manipulate a point-of-care blood gas analyzer by order or protocol (Code: IIA10) [Difficulty: ELE: R, Ap; WRE: An]

b. Put the equipment together and make sure that it works properly (CSE code: IIA10)

POC analyzers can be powered through a self-contained battery or by plugging the unit into a standard alternating current electrical outlet. The units have single-use disposable cartridges that include the electrodes and calibration reagents and will hold waste samples. The following discussion covers the basic principles of operation of the electrodes in POC and standard centrally located analyzers.

1. pH electrode

The modern pH electrode has existed since the mid-1950s and is usually referred to as the Sanz electrode after its principal inventor. The basic principle behind the pH analyzer is its ability to measure the voltage (potential for electrical flow) between two different solutions. This is based on the different hydrogen ion (H+) concentrations between the solutions that reflect their relative pH values. The reference electrode is immersed in a solution with a pH of 6.840 that fills a glass or plastic chamber. The blood sample, of unknown pH, is placed in a separate measuring chamber called a cuvette. These two chambers are separated by a special glass membrane that contains metals and sodium ions (Na+), thus making it pH sensitive. Both chambers are kept at a stable 37° C temperature. See Figure 3-5 for a graphic representation of the pH electrode. When blood or a quality control (QC) material is introduced into the cuvette, the potential exists for hydrogen ions to replace the sodium ions in the pH-sensitive glass if the two pHs are different. The replacement is proportional to the difference in the two pHs.

3. Po2 electrode

This unit is completely different from the others mentioned. It was developed in the late 1950s by Clark and thus is usually called a Clark electrode. It is also sometimes known as a polarographic electrode because of the basis of its operation. Figure 3-7 is a drawing of key features of the unit. A phosphate-KCl buffer solution surrounds the silver anode. A thin membrane separates the blood-filled cuvette from direct contact with the electrode but allows oxygen molecules to diffuse slowly through to contact the platinum wire cathode. The whole unit is heated to 37° C. The term polarographic comes from the addition of about 0.7 volt to the cathode to make it slightly “polarized” or negative compared with the anode. This is needed to ensure that oxygen is rapidly chemically reduced (that it gains electrons) at the cathode. This creates an electrical current directly proportional to the number of reduced oxygen molecules.

image

Figure 3-7 Schematic illustration of the Clark electrode for measuring PO2.

(From Harrison BA, Shapiro C: Clinical application of blood gases, ed 4, St Louis, 1989, Mosby.)

It must be understood that the partial pressure of oxygen (Po2) being measured is derived from oxygen that is dissolved in the plasma. It does not come from the hemoglobin found in the erythrocytes (red blood cells). The reported value for the saturation of oxygen in the hemoglobin (Sao2) is calculated by using a mathematical table. Under normal conditions, the calculated Sao2 value is the same as or close to the true Sao2 value. Carbon monoxide (CO) poisoning is the only commonly seen clinical situation during which a calculated saturation can be incorrectly high. If carbon monoxide poisoning is suspected or known, the patient’s blood sample should be analyzed on a CO-oximeter unit.

d. Perform quality control procedures for a blood gas analyzer (Code: IIC1) [Difficulty: ELE: R, Ap; WRE: An]

4. Quality control materials

A variety of QC materials are available to calibrate the electrodes for Po2, Pco2, and pH measurements. Their uses vary. Each of the following materials has its advantages, disadvantages, and limitations. The manufacturer of a particular brand or model of blood gas analyzer may require that a specific type of material be used in its units.

Aqueous buffers are water based and are used to check pH and Pco2 measurements; they cannot be used to check Po2 measurements. Commercially prepared gases are used to check Po2 and Pco2 measurements; they cannot be used to check pH measurements. The following CO2 mixes may be used: 0, 5%, 10%, and 12%. The following O2 mixes may be used: 0, 12%, 20%, 20.95% from room air, 21%, and 100%.

Tonometered liquids are exposed in the laboratory to known oxygen and carbon dioxide gas mixes until the liquids are saturated and have the same partial pressures as the gases. Three types of tonometered liquids are used. First, using human or animal serum or whole blood is the most accurate method available and is used mainly to obtain Po2 and Pco2 levels. Although whole human blood cannot be used for pH measurements, a bovine blood product can be used to obtain all three values. Second, assayed liquids are non–water-based liquids that are pretonometered by the manufacturer and are available in sealed glass vials. They can be used to obtain Po2, Pco2, and pH measurements. These are very popular because of the advantages of speed and simplicity. Third, oxygenated fluorocarbon-based emulsions (perfluorinated compounds) can be used to obtain Po2, Pco2, and pH measurements. They are considered as accurate as whole blood without its associated risks.

5. Levey-Jennings charts

The Levey-Jennings charts (also known as Shewhart/Levey-Jennings or quality control [QC] charts) are used to record the results of each calibration procedure. They are similarly designed with time plotted on the horizontal scale and the analyte (Po2, Pco2, or pH) on the vertical scale. The vertical scale for each analyte has a central value for what is normally expected. On both sides of this normal value are standard deviation (SD) points, showing movement away from what is expected. When an analyte electrode is operating within the acceptable limits, it is said to be in control. In general, if an analyte is within 2 SDs of the normal value, it is considered to be in control.

An out-of-control situation exists whenever a single calibration value or a series of calibration values is outside established limits. A random error is an unpredictable aberration in precision that occurs when the QC material is sampled. A systematic error shows an accuracy problem and is much more serious. It must be investigated, corrected, and documented. Figure 3-8 shows an example of both random error and systematic error. Rules have been established for determining whether the error is random or systematic (Table 3-1).

TABLE 3-1 Westgard’s Rules for Determining When an Aalyzer Is Not Functioning Properly

Rule name Levey-Jennings chart
RANDOM ERROR
1-2 SDs The measurement is more than 2 SDs but not more than 3 SDs from the mean.
1-3 SDs The measurement is more than 3 SDs from the mean.
R*-4 SDs Two consecutive measurements are 4 SDs or more apart.
SYSTEMATIC ERROR
2-2 SDs Two consecutive measurements are either 2 SDs above or 2 SDs below the mean.
4-1 SDs Four consecutive measurements are either 1 SD above or 1 SD below the mean.
7-trend Seven consecutive measurements are on only one side of the mean; each measurement is progressively more out of control.
10-mean Ten consecutive measurements are on only one side of the mean.

SDs, standard deviations.

Modified from Lane EE, Walker JF: Clinical arterial blood gas analysis, St Louis, 1987, Mosby.

7. Pco2 electrode

8. Po2 electrode

e. Perform quality control procedures for a point-of-care blood gas analyzer (Code: IIC3) [Difficulty: ELE: R, Ap; WRE: An]

The general principles of QC and QA discussed with standard blood gas analyzers apply to point-of-care (POC) analyzers as well. Remember to flush the electrode membrane after each use, if possible, to prevent protein buildup. If protein buildup occurs, the response time is longer than normal. Follow the manufacturer’s guidelines to change an electrode membrane as needed. Make sure that no air bubbles are under the membrane or within the tubing through which the blood travels. Rerun the calibration for any of the electrodes and reanalyze the sample if you are suspicious of the result. If the electrode does not calibrate close to the reference buffer solutions or gases, it should not be used.

In a random-error situation, the practitioner likely made a simple error when introducing the material or running the analyzer. Common problems include an air bubble injected into the unit or incomplete flushing of the previous sample. Usually flushing out any residual blood and then carefully injecting more of the current patient blood samples will correct the problem. Run the analyzer again to obtain new patient values or run the same patient sample through another analyzer and compare the two sets of results.

A systematic error usually indicates a problem with the analyzer, QC materials, or processes. Examples of systematic errors include misanalyzed CO2 or O2 standards for calibration; contaminated QC materials; or deteriorated oxygen, carbon dioxide, or pH electrode function. Each of these must be investigated until the problem is found and corrected. The unit cannot be used again until after it is proven to work properly and to give accurate results.

f. Perform blood gas analysis (Code: IB9j and IIIE3c) [Difficulty: ELE: R, Ap; WRE: An]

Modern blood gas analyzers are simple to operate. Follow the manufacturer’s guidelines for insertion of the blood sample. Perform the specified steps in the analysis. Print out the results. Many current units run a self-diagnosis if any problems appear. Correct any identified problems as discussed above.

1. CO-oximeter/hemoximeter

a. Perform quality control procedures on the CO-oximeter/hemoximeter (Code: IIC1) [Difficulty: ELE: R, Ap; WRE: An].

The unit is preassembled by the manufacturer. Practical experience with a unit is recommended to understand how to add a patient blood sample and perform calibration duties. (See Figure 3-9 for a schematic drawing of a CO-oximeter.) A thallium-neon hollow cathode lamp emits light in the infrared-visible range. A device called a monochromator contains four filters and rotates through the light beam. Each filter allows only one specific wavelength to pass through it. These four monochromatic wavelengths correspond to the three isosbestic points, discussed later (shown in Figure 3-10) and 626.6 nm. This last wavelength is poorly absorbed by all four hemoglobin moieties. It is used to find the maximal difference in absorption so that the relative amounts of the hemoglobin species can be determined.

When a blood sample is placed into the cuvette, the same four wavelengths are passed through it. The amount of absorbance at each wavelength is measured and compared with the absorbance at each wavelength by a reference sample solution (see the following discussion). The computer integrates the data and calculates the total hemoglobin concentration and the amounts of the four hemoglobin moieties.

Total hemoglobin (THb) concentration should be calibrated when the unit is installed, at regular intervals suggested by the manufacturer, after the sample tubing is changed, after the cuvette is disassembled or changed, and whenever a suspicious reading is seen. Calibration is done by filling the cuvette with a special dye produced by the manufacturer and analyzing it by following the prescribed steps.

Routine calibration is done every 30 minutes. The unit obtains and stores absorbance readings at the four different wavelengths from a “blank” solution in the reference detector. When the same blank solution is added to the sample cuvette, the same four wavelengths are measured. Their absorption levels are normally identical. The same procedure is performed after every patient sample is analyzed.

The following are examples of common problems and their solutions:

b. Perform CO-oximetry/hemoximetry (Code: IB9j and IIIE3c) [Difficulty: ELE: R, Ap; WRE: An].

Most hospitals have a CO-oximeter in addition to a blood gas analyzer. A CO-oximeter should be used to analyze a blood gas sample whenever carbon monoxide poisoning is known or suspected. In addition, a CO-oximeter gives a complete analysis of the relative amounts of the different types of patient hemoglobin.

CO-oximeters also are called spectrophotometric oximeters and are the most accurate method available to measure the four different hemoglobin moieties (species or variations in the hemoglobin molecule). These hemoglobin species include the following:

Each of these hemoglobin moieties has a unique spectroscopic “fingerprint” of absorbed light-wave frequencies. See Figure 3-10 for the spectral analysis of the various forms of hemoglobin.

Follow the manufacturer’s guidelines on rewarming the blood sample to body temperature, hemolyzing the sample, and inserting it into the measurement cuvette. Failure to do so may result in incorrect patient values.

The principle of operation of a CO-oximeter is the comparison of the relative absorbances of four wavelengths of light by oxyhemoglobin, reduced hemoglobin, and carboxyhemoglobin. This is done by comparing the absorptions at the three isosbestic points (at which the moieties being compared have equal absorption) and a wavelength point at which the greatest difference in absorption between the two moieties is found. By computer integration of the data, the relative proportions of oxyhemoglobin, reduced hemoglobin, and carboxyhemoglobin are determined. If the total is less than 100% of the hemoglobin present, the difference has to be methemoglobin (or, rarely, sulfhemoglobin). The unit then provides data on total hemoglobin; percentages for oxyhemoglobin, reduced hemoglobin, carboxyhemoglobin, and methemoglobin; and total amounts for them in grams per deciliter of blood. Some units also calculate O2 content.

MODULE D

2. Interpret the results of arterial blood gas analysis (Code: IB10j, IB10f, and IIIE4a) [Difficulty: ELE: R, Ap; WRE: An]

If the blood sampling equipment preparation, the sampling procedure, and blood gas analyzer operation were all done properly, the respiratory therapist and physician can believe that the blood gas results are accurate and reliable. However, if there was a problem with any aspect of obtaining and analyzing the blood gas sample, that sample must be discarded and a new sample obtained and analyzed.

A number of authors have written extensively on how to interpret ABGs. The system proposed by Shapiro and associates (1994) has been found to be both practical and relatively easy to understand. Most of the following discussion and tables are based on this system. This does not mean that if one has learned another system, he or she is at any disadvantage for taking the National Board for Respiratory Care (NBRC) examination.

a. Assessment of oxygenation

Hypoxemia or hypoxia can rapidly become life threatening. This will often be first seen when the patient develops cardiac arrhythmias, unstable vital signs, and mental confusion or unconsciousness. Table 3-2 shows the normal Pao2 values for the newborn, child to adult, and the elderly when room air (almost 21% oxygen) is inhaled at sea level. These values decrease progressively as altitude increases. However, under most clinical conditions, this is not a factor unless working in a high-altitude setting.

TABLE 3-2 Age-Based Acceptable Levels of Partial Pressure of Oxygen in Arterial Blood (Pao2) When Breathing Room Air (21% Oxygen) at Sea Level

Age Pao2
NEWBORN  
Acceptable range 40 to 70 torr
CHILD TO ADULT  
Normal 97 torr
Acceptable range >80 torr
Hypoxemia <80 torr
OLDER ADULT  
60-year-old >80 torr
70-year-old >70 torr
80-year-old >60 torr
90-year-old >50 torr

Modified from Shapiro BA, Peruzzi WT, Kozelowski-Templin R: Clinical application of blood gases, ed 5, Chicago, 1994, Mosby-Year Book.

A general rule is that any patient is seriously hypoxemic if the Pao2 is less than 60 torr on room air. See Table 3-3 for guidelines on judging the seriousness of hypoxemia. Once hypoxemia is recognized, it must be corrected. The most obvious way to correct hypoxemia is to give supplemental oxygen. The clinician must realize that oxygen alone will not correct the hypoxemia if the patient is hypoventilating (increased Paco2), has heart failure, or is unable to carry or make use of the oxygen. In general, try to keep the patient’s Pao2 level between 60 and 100 torr.

TABLE 3-3 Evaluation of Hypoxemia

CONDITIONS: ROOM AIR IS INSPIRED; THE PATIENT IS YOUNGER THAN 60 YEARS*
Hypoxemia Pao2 Sao2
Mild 60-79 torr 90-94%
Moderate 40-59 torr 75-89%
Severe <40 torr <75%
CONDITIONS: SUPPLEMENTAL O2 IS INSPIRED; THE PATIENT IS YOUNGER THAN 60 YEARS
Hypoxemia Pao2
Uncorrected Less than room air acceptable limit
Corrected Within room air acceptable limit (<100 torr)
Excessively corrected >100 torr

* Subtract 1 torr of O2 from limits of mild and moderate hypoxemia for each year older than 60. A Pao2 value of <40 torr indicates severe hypoxemia in any patient at any age.

Modified from Shapiro BA, Peruzzi WT, Kozelowski-Templin R: Clinical application of blood gases, ed 5, Chicago, 1994, Mosby-Year Book.

Shapiro and associates (1994) suggested the following formula, in which FiO2 is the fraction of inspired oxygen, for determining whether the patient will be hypoxemic on room air: “If Pao2 is less than FiO2 ×5, the patient can be assumed to be hypoxemic on room air.”

Figure 3-11 shows a normal oxyhemoglobin dissociation curve. The saturation value is important to know because it shows how much hemoglobin is saturated with oxygen. Several important points of correlation exist between the Sao2 and the Pao2. (Calculated saturation values can be misleadingly high if the patient has inhaled carbon monoxide. In this situation, it is best to measure the saturation directly on a CO-oximeter–type blood gas analyzer.)

Figure 3-12 shows a number of factors that can influence the oxyhemoglobin dissociation curve and how oxygen associates and dissociates from hemoglobin. In a patient with normal oxygenation, these factors are not clinically significant. However, when the Pao2 level is less than 60 torr and the Sao2 level is less than 90%, these factors can become an important consideration. As can be seen, a left-shifted oxyhemoglobin dissociation curve results in a lower Pao2 value at any given saturation. This results in even less oxygen being delivered to the tissues.

b. Assessment of carbon dioxide and pH

The pH is the next important value to interpret because extreme acidemia/acidosis and alkalemia/alkalosis can be life threatening. The carbon dioxide value is important to interpret because it has a direct effect on the pH and indirectly affects the oxygen level. A high or low Paco2 level, by itself, is not life threatening.

Table 3-4 shows normal values for pH and Paco2 and the acceptable ranges around the mean or average. Table 3-5 shows the most widely acceptable therapeutic ranges for pH and Paco2. Values that fall outside of these ranges present a progressively greater risk to the patient.

TABLE 3-5 Acceptable Clinical Ranges for Partial Pressure of Carbon Dioxide in Arterial Blood (Paco2) and pH

Paco2 30-50 torr*
pH 7.30-7.50

* This is the range for patients with an acute change. It does not apply to patients with long-standing disease, such as chronic obstructive pulmonary disease. These patients may have Paco2 values greater than 50 torr.

Modified from Shapiro BA, Peruzzi WT, Kozelowski-Templin R: Clinical application of blood gases, ed 5, Chicago, 1994, Mosby-Year Book.

Table 3-6 shows the definitions of Shapiro and associates for alkalemia and acidemia from a respiratory cause. An acute change in the patient’s ventilation causes the following when starting from a Paco2 of 40 torr:

TABLE 3-6 Naming Unacceptable Values for Partial Pressure of CO2 in Arterial Blood (Paco2) and pH

Paco2 >45 torr Respiratory acidosis/alveolar hypoventilation/ventilatory failure
pH <7.35 Acidemia
Paco2 <35 torr Respiratory alkalosis/alveolar hyperventilation
pH >7.45 Alkalemia

Modified from Shapiro BA, Peruzzi WT, Kozelowski-Templin R: Clinical application of blood gases, ed 5, Chicago, 1994, Mosby-Year Book.

Thus the body can be seen as better able to compensate with metabolic buffers for a respiratory acidosis than a respiratory alkalosis.

Metabolic effects are evaluated by interpreting either the bicarbonate (HCO3) value or the base excess/base deficit (BE/BD) value. Both reveal whether any metabolic effect on the pH exists. Normal values are as follows:

Values indicating metabolic alkalosis of a primary or secondary nature are as follows:

Values indicating metabolic acidosis of a primary or secondary nature are as follows:

Tables 3-7 to 3-9 show definitions of terms and classifications of the various acid-base states. As stated earlier, there are other systems for interpreting blood gases. All are probably satisfactory for interpretation purposes and preparing for the NBRC examinations.

TABLE 3-7 Clinical Terminology for Arterial Blood Gas Measurements

Clinical Terminology Clinical Findings
Respiratory acidosis/alveolar hypoventilation/ventilatory failure Paco2 >45 torr
Acute ventilatory failure Paco2 >45 torr; pH <7.35
Chronic ventilatory failure Paco2 >45 torr; pH 7.36-7.40
Respiratory alkalosis/alveolar hyperventilation Paco2 <35 torr
Acute alveolar hyperventilation Paco2 <35 torr; pH >7.45
Chronic alveolar hyperventilation Paco2 <35 torr; pH 7.41-7.45
Acidemia pH <7.35
Acidosis Pathophysiologic condition in which the patient has a significant base deficit (plasma bicarbonate level below normal)
Alkalemia pH >7.45
Alkalosis Pathophysiologic condition in which the patient has a significant base excess (plasma bicarbonate level above normal)

Modified from Shapiro BA, Peruzzi WT, Kozelowski-Templin R: Clinical application of blood gases, ed 5, Chicago, 1994, Mosby-Year Book.

TABLE 3-8 Evaluation of Ventilatory and Metabolic Effects on Acid-Base Status

EVALUATION OF Paco2
Paco2 >45 torr Respiratory acidosis/alveolar hypoventilation/ventilatory failure
Paco2 35-45 torr Acceptable alveolar ventilation
Paco2 <35 torr Respiratory alkalosis/alveolar hyperventilation
EVALUATION OF Paco2 IN CONJUNCTION WITH pH*
Acceptable Alveolar Ventilation (Paco2 from 35 to 45 torr)
pH >7.50 Metabolic alkalosis
pH 7.30-7.50 Acceptable pH
pH <7.30 Metabolic acidosis
Alveolar Hypoventilation (Paco2 >45 torr)
pH >7.50 Partially compensated metabolic alkalosis
pH 7.30-7.40 Chronic ventilatory failure
pH <7.30 Acute ventilatory failure
Alveolar Hyperventilation (Paco2 <35 torr)
pH >7.50 Acute alveolar hyperventilation
pH 7.40-7.50 Chronic alveolar hyperventilation
pH 7.30-7.40 Compensated metabolic acidosis
pH <7.30 Partially compensated metabolic acidosis

Paco2, partial pressure of CO2 in arterial blood.

* Some authors use a narrower pH range for these classifications.

Modified from Shapiro BA, Peruzzi WT, Kozelowski-Templin R: Clinical application of blood gases, ed 5, Chicago, 1994, Mosby-Year Book.

3. Interpret the results of CO-oximetry/hemoximetry blood gas analysis (Code: IB10j and IIIE4b) [Difficulty: ELE: R, Ap; WRE: An]

If the blood sampling equipment preparation, the sampling procedure, and blood gas analyzer operation were all done properly, the respiratory therapist and physician can believe that the blood gas results are accurate and reliable. However, if there was a problem with any aspect of obtaining and analyzing the blood gas sample, that sample must be discarded and a new sample obtained and analyzed.

The CO-oximeter type blood gas analyzer gives values for oxyhemoglobin (O2Hb), reduced hemoglobin (RHb), carboxyhemoglobin (COHb), and methemoglobin (MetHb)/sulfhemoglobin (SHb). Each of these hemoglobin moieties can be displayed in terms of grams per deciliter, percentage of the whole, and added together for a total hemoglobin (THb) value. Table 3-10 shows normal adult hemoglobin values. The amounts of carboxyhemoglobin and methemoglobin should be subtracted from the total hemoglobin amount to find the amount of functional hemoglobin. Any increase in the COHb or MetHb levels above those listed is abnormal and results in even less normal hemoglobin to carry oxygen. The patient with carbon monoxide poisoning is at greatest risk. A COHb level of 30% saturation or greater can be fatal. By subtraction, the O2Hb (Sao2) level can be no greater than 70% with a resulting Pao2 of less than 40 torr.

TABLE 3-10 Normal Hemoglobin Values for Adults

Total hemoglobin (THb) Men: 13.5-18.0 g/dL
Women: 12.0-16.0 g/dL
15.0 g/dL is often listed as an average for both
Oxyhemoglobin 94-100% of THb (reported as Sao2*) of 94-100%
Carboxyhemoglobin Nonsmokers: <1.5% (0.225 g/dL) of THb
Smokers: 1.5-10% of THb
Methemoglobin 0.5-3% (0.075-0.45 g/dL) of THb
Oxygen content (arterial sample) 15-23 g/dL

g/dL, grams per deciliter (sometimes listed as g/100 mL); Sao2, O2 saturation in arterial blood.

4. Mixed venous blood gases

b. Interpret the results of mixed venous blood gas analysis to evaluate the patient’s response to respiratory care (Code: IB10j and IIIE4a) [Difficulty: ELE: R, Ap; WRE: An]

If the blood sampling equipment preparation, the sampling procedure, and blood gas analyzer operation were all done properly, the respiratory therapist and physician can believe that the blood gas results are accurate and reliable. However, if there was a problem with any aspect of obtaining and analyzing the blood gas sample, that sample must be discarded and a new sample obtained and analyzed.

In the critically ill patient, it is just as important to measure the mixed venous blood gases as it is to measure the ABGs. The venous blood gas values reveal what has happened as the arterial blood has passed through the body. Oxygen has been extracted, and carbon dioxide has been added to the blood. The difference between the arterial and venous oxygen levels reflects oxygen consumption by the body as well as cardiac output.

Because of this, the most critical venous blood gas values to measure are the SimageO2 and PimageO2. It is generally accepted that a PimageO2 value of less than 30 torr or an SimageO2 value of less than 56% indicates that the patient has tissue hypoxia. Both values can be obtained by analyzing a mixed venous blood sample taken through a pulmonary artery catheter. If the patient has a fiberoptic catheter, the SimageO2 value can be monitored continuously. This is extremely helpful if the patient is unstable or having frequent changes in inspired oxygen or ventilator settings. (Pulmonary artery catheters are discussed in more detail in Chapter 5.)

5. Arterialized capillary blood gases

b. Interpret the results of capillary blood gas analysis to evaluate the patient’s response to respiratory care (Code: IB10g and IIIE4a) [Difficulty: ELE: R, Ap; WRE: An]

If the blood sampling equipment preparation, the sampling procedure, and blood gas analyzer operation were all done properly, the respiratory therapist and physician can believe that the blood gas results are accurate and reliable. However, if there was a problem with any aspect of obtaining and analyzing the blood gas sample, that sample must be discarded and a new sample obtained and analyzed.

Because this is not a true ABG sample, the following limitations are placed on interpreting the results.

Based on these limitations, an arterialized capillary blood pH value can be clinically useful. The capillary CO2 value should be viewed with suspicion. It should not be the only parameter monitored to judge the infant’s ability to ventilate; however, the combination of a low pH and elevated CO2 value indicates hypoventilation. Evaluate the infant’s vital signs, breathing efforts, chest radiograph, and so on to determine respiratory status.

Unfortunately, the capillary O2 value is practically useless for judging the infant’s oxygenation. Some infants may have a fairly close correlation with the Pao2 level, whereas others do not. No way is known to predetermine those that will match and those that will not. Many practitioners view a low capillary oxygen level as a sign of clinical hypoxemia. The American Association for Respiratory Care (AARC) Clinical Practice Guideline Oxygen Therapy in the Acute Care Hospital notes that a Pco2 level of less than 40 torr documents neonatal hypoxemia. An ABG sample remains the best way to determine the patient’s respiratory status.

MODULE E

3. Manipulate a pulse oximeter by order or protocol (ELE code: IIA22) [Difficulty: ELE: R, Ap, An]

c. Troubleshoot any problems with the equipment

The pulse signal will not be strong if the patient has poor circulation at the site of the oximeter probe. This can occur if the patient is hypothermic, hypotensive, or receiving a vasoconstricting medication. If the pulse signal is weak, the pulse oximetry values may not be accurate. See Table 3-12 for common sources of error and their solutions.

TABLE 3-12 Sources of Error in Pulse Oximetry

Sources of Error Remedy
Light interference: xenon lamp, fluorescent light, infrared (bilirubin) light; probe fell off of patient Cover the probe with an opaque wrap; put the probe back in place on the patient
Low perfusion: low blood pressure, hypothermia, vasoconstricting drugs Use earlobe, bridge of nose, or forehead instead of finger or toe; discontinue use if still unreliable
Motion artifact Secure the probe site; ensure that the Spo2* reading is synchronized with the heart rate
Darkly pigmented patient Use lightly pigmented site such as tip of finger or toe; Spo2 value may overestimate Pao2; discontinue use if still unreliable
Artificial or painted fingernails Remove acrylic nails; remove black, blue, green, metallic, or frosted nail polish; use a different site
Venous pulsation being read as an arterial pulsation Loosen a tight sensor; change the finger sensor site every 2-4 hr; loosen the cause of a tourniquet-like effect
The following vascular dyes will cause low Spo2 readings: methylene blue, indigo carmine, indocyanine green Do not use Spo2

Pao2, partial pressure of O2 in arterial blood; Spo2, pulse oximeter.

4. Perform pulse oximetry on your patient (Code: IB9c and IIIE3b) [Difficulty: ELE: R; WRE: Ap, An]

First-generation pulse oximetry has gained wide acceptance because it offers a way to monitor continuously and noninvasively a patient’s oxygenation by following the percentage of hemoglobin saturated with oxygen. The reported Spo2% is the percentage of oxyhemoglobin.

Pulse oximetry makes practical use of two physical principles. The first is spectrophotometry, which is used to analyze the transmission of wavelengths of light through the blood and body tissues. First-generation pulse oximeters analyze two wavelengths of light. One wavelength is 660 nm, and the other is between 920 and 940 nm, depending on the manufacturer. The 660-nm wavelength can be seen as red and is preferentially absorbed by oxyhemoglobin (O2Hb). The 920- to 940-nm wavelength is not visible, because it is in the infrared range. It is preferentially absorbed by reduced hemoglobin (RHb). Second-generation pulse oximeters analyze several more wavelengths of light and can evaluate different forms of hemoglobin.

The second principle is plethysmography. It is used to find and then evaluate the amplitude of the arterial pulse waveform. See Figure 3-13 for the plethysmographic arterial waveform. When the pulse oximeter sensor is placed on a patient site, the fingertip for example, the two wavelengths of light shine through the blood, tissues, and bone within the finger. It is important that the sending LED and receiving (photodiode) sensors be opposite each other (see Figure 3-14). Most units have a signal strength display that indicates when the photodiode is receiving a strong signal, and the patient’s pulse has been detected. The microprocessor is designed to detect a baseline level of light absorption by the tissues and venous blood, containing more RHb, as well as the light absorption of arterial blood, containing more O2Hb. It can then compare the absorptions of the two wavelengths to determine the level of saturated oxyhemoglobin. This is displayed as saturation by pulse oximetry, or Spo2.

It must be realized that because first-generation pulse oximetry samples only two wavelengths of light, the technology is unable to recognize the presence or quantity of the nonfunctional hemoglobin species of COHb and MetHb. Instead, an ABG sample should be drawn and sent to the laboratory to be passed through a CO-oximeter for a complete fractional hemoglobin analysis. Even healthy persons have small amounts of COHb and MetHB. For this and other technical reasons, manufacturers report the following Spo2 values for general accuracy at 1 standard deviation (SD) for a general population: ± 2% from 100% to 70% saturation, and ± 3% from 70% to 50% saturation. Because of these limitations, the following clinical guidelines have been created by a number of authors.

The second-generation pulse oximeters (Masimo Rad-57 or Rainbow SET) analyze 7 to 12 wavelengths of light. Because of this, they can identify and measure the various normal and abnormal types of hemoglobin, such as carboxyhemoglobin (SpCO) from carbon monoxide poisoning and methemoglobin (SpMet) from nitric oxide therapy. These units provide the opportunity to continuously monitor patients with elevated levels of carboxyhemoglobin and methemoglobin as they are being treated. It may not be necessary to draw as many arterial blood gas samples for analysis through a CO-oximeter. If necessary, review the discussion on CO-oximeters and see Figure 3-10 since the principles apply to second-generation pulse oximeters.

Table 3-13 lists common clinical ranges for Spo2 values. For the aforementioned reasons, with first-generation pulse oximeters the minimum safe values are 2% higher than the corresponding Sao2 value by CO-oximetry. It is important that the patient have good pulsatile blood flow to the measurement site to obtain an accurate reading.

5. Interpret a patient’s pulse oximetry value (Code: IB10c and IIIE4d) [Difficulty: ELE: R, Ap; WRE: An]

The previously healthy patient who has cardiopulmonary failure should have the Spo2 value kept at 92% or greater to ensure adequate oxygenation. The patient with chronic obstructive pulmonary disease (COPD) can probably tolerate an Spo2 value as low as 87%. In general, a neonate should have the Spo2 level kept between 92% and 96%. See Table 3-13 for more specific guidelines. Saturations below these values indicate hypoxemia in most patients.

Note the site at which the saturation was measured. This is especially important in neonates who may have congenital heart defects. A higher saturation in the right fingers or right earlobe compared with the rest of the body is seen in a neonate who has patent ductus arteriosus (PDA). A higher saturation in the fingers and earlobes as compared with the toes is seen in a neonate who has coarctation of the aorta.

MODULE F

NOTE: Transcutaneous monitoring (TCM) involves the continuous monitoring of oxygen, carbon dioxide, or both as they diffuse through the skin. Each gas has its own individual electrode, and a combined electrode exists for both gases.

3. Manipulate a transcutaneous oxygen monitor by order or protocol (ELE code: IIA22) [Difficulty: ELE: R, Ap, An]

4. Perform quality control procedures on a transcutaneous oxygen monitor (Code: IIC7) [Difficulty: ELE: R, Ap; WRE: An]

Always follow the manufacturer’s recommendations for the assembly and care of the equipment. Select the proper electrode for the monitor based on the physician’s order for evaluating transcutaneous oxygen, carbon dioxide, or both together.

a. Calibration

Two-point calibration must be performed with oxygen percentages that will cause Po2 values beyond the clinical range that can be expected. Usually the first calibration point is a “zero” point because the electrode is exposed to no oxygen in a nitrogen-filled chamber. This point is usually quite stable. The second calibration point is found when the electrode is exposed to room air (20.95% or 0.2095 oxygen). Always follow the manufacturer’s written procedures during the calibration process. Generally, when environmental conditions include a fairly stable room temperature of 25° C and 50% relative humidity, the following equation can be used to predict the room air calibration point:

image

in which PB is the local barometric pressure and 0.2095 is the oxygen fraction found in room air. Expose the electrode to room air to determine whether it matches the calculated calibration value. Adjust the instrument to match the calibration Ptco2 if necessary. It is recommended that the room air calibration point be rechecked every 24 hours when in continuous use, after changing the membrane, or after changing the electrolyte solution. A variation of up to ± 5 torr is acceptable and can be corrected by adjusting the reading on the instrument. If the variation is greater than ± 5 torr, the zero point and room air calibration procedures should be repeated.

5. Perform transcutaneous oxygen (Ptco2) monitoring (Code: IB9b) [Difficulty: ELE: R, Ap; WRE: An]

c. Limitations and patient precautions

Because the electrode is heated, it must be rotated to a different skin site on a routine basis. The general manufacturer’s guidelines for site rotation are every 4 to 6 hours for neonates weighing between 1000 and 2500 g; every 3 to 5 hours for neonates weighing between 2500 and 3500 g; and 2 to 4 hours for neonates weighing more than 3500 g, pediatric patients, and adult patients. As a safety precaution, change the electrode on all patients at least every 4 hours, or at least every 3 hours if the patient is hypothermic. The manufacturer’s range in times is based on the relative thickness of the patient’s skin and the different electrode temperatures. It is important to adjust the site-rotation times on an individual patient basis. Some may tolerate longer times, and others will need more frequent rotations.

Care must be taken when removing the adhesive ring and electrode. It is possible to tear the thin, delicate skin of a premature neonate. Loosen the adhesive by running the edge of an alcohol wipe along the side that is being gently pulled up. After the electrode is removed, it is important to examine the skin; it is normal to see a red circle. This warmed, vasodilated area will stay red for some time and will gradually fade away, with no scarring or permanent injury. Rarely, the skin will have overheated and a blister is seen; this is a second-degree burn. Obviously, future site rotations must be made more frequently. Do not use this site again. Treat it as a burn, and avoid any further injury that might break the skin and lead to an infection.

6. Interpret transcutaneous oxygen monitoring values to evaluate the patient’s response to respiratory care (Code: IB10b) [Difficulty: ELE: R, Ap; WRE: An]

It has been found that heating the electrode speeds the diffusion of oxygen through the skin. Heating also provides a closer correlation with the patient’s Pao2 value. However, it must be remembered that the PtCO2 value is not the same as the Pao2 value. Current recommendations are that any unit should give PtCO2 values that are within ± 15% of the Pao2 values over the operating range of the instrument. The values should then correlate within ± 15% as the patient’s condition changes. This should always be confirmed.

An arterial blood gas sample should be drawn for Pao2 every time transcutaneous oxygen monitoring is started. The Pao2-PtCO2 gradient can then be calculated as the difference between the two. For example, if the patient’s Pao2 value is 100 torr, the PtCO2 value should be no less than 85 torr. If the PtCO2 level decreases to 70 torr, the Pao2 level should have decreased to no lower than 85 torr. Because of this close correlation, the patient’s trend can be monitored with some assurance of accuracy. In addition, ABG samples do not need to be drawn as frequently for the Pao2 measurement. This trending relation holds true for changes in the patient’s pulmonary condition. It does not, however, hold true when the patient has cardiovascular problems such as hypotension, hypothermia, peripheral vascular disease, or cardiogenic shock with decreased tissue perfusion.

Always follow the manufacturer’s recommendations for the proper electrode temperature. In general, the temperature ranges from 42.5° C for a 1000-g infant to 44° C for a 3500-g infant. A pediatric patient can tolerate a temperature of 44° C, whereas an adult can have an electrode temperature of 45° C. The higher temperatures are needed in older patients because their skin is thicker. Studies have shown that when the neonate’s Pao2 level is greater than 100 torr, the PtCO2 value underestimates it. This can lead to dangerous hyperoxemia. For this reason, it is recommended that the Ptco2 level be kept at less than 90 torr. In the adult, less correlation is found between the Pao2 and Ptco2 values because of the thicker skin found in adults. The Ptco2 value can still be used in the adult to follow trends in oxygenation, but the practitioner must realize that a decrease in the Ptco2 value can be a result of hypoxemia, a decreased cardiac output, or cutaneous vasoconstriction. An arterial blood gas sample must be drawn to further evaluate the patient’s status.

MODULE G

3. Manipulate a transcutaneous carbon dioxide monitor by order or protocol (ELE code: IIA22) [Difficulty: ELE: R, Ap, An]

4. Perform quality control procedures on a transcutaneous carbon dioxide monitor (Code: IIC7) [Difficulty: ELE: R, Ap; WRE: An]

Two-point calibration must be performed with carbon dioxide percentages that cause Pco2 values beyond the expected clinical range. Usually the electrode is exposed to 5% and 10% carbon dioxide from prepared cylinders. Always follow the manufacturer’s written procedures during the calibration process. Generally, when environmental conditions include a fairly stable room temperature of 25°C and 50% relative humidity, the following equation can be used to predict the two calibration points:

Calibration PtcCO2 = (PB × CO2% used) − XCO2 × electrode temperature factor)

in which PB is the local barometric pressure; CO2% is the carbon dioxide fraction exposed to the electrode, either 10% (0.10) or 5% (0.05); and XCO2 is the correction factor used to equilibrate the PtcCO2 to Paco2. (This is discussed later.)

MODULE H

2. Difference between the alveolar oxygen level and the arterial oxygen level

a. Calculate the difference between the alveolar oxygen level and the arterial oxygen level (ELE code: IB9i) [Difference: ELE: R, Ap]

This gap between the alveolar oxygen level and the arterial oxygen level is called the alveolar-arterial oxygen pressure difference [P(A-a)O2]. Measurement of this gap or difference gives an important indication of the seriousness of the patient’s condition.

Perform the following to determine the alveolar-arterial O2 pressure difference [P(A-a)O2]:

In which:

image

The factor of 0.8 is based on how much oxygen a normal person uses in 1 minute and how much carbon dioxide is produced in 1 minute. The symbols for this metabolic value are R for respiratory exchange ratio and RQ for respiratory quotient. The following calculation is based on a normal person’s metabolism:

image

Because many sick patients do not react as expected, the factor has a range of 0.6 to 1.1, depending on oxygen consumption and carbon dioxide production. Assume that the factor is 0.8 unless you are told otherwise or measure otherwise.

b. Interpret the results of the P(A-a)O2 calculation (Code: IB10i) [Difficulty: ELE: R, Ap; WRE: An]

When interpreting the patient’s value, remember that the P(A-a)O2 level in a normal young person should be no greater than 15 torr. The difference slowly increases as a normal person ages (see Figure 3-17). Lung disease causes the value to increase significantly. The following are examples of some conditions in which the measurement of the P(A-a)O2 level aids in diagnosis or treatment:

The following examples are offered to aid in the calculation of P(A-a) O2 and the interpretation of the results.

3. Determine the appropriateness of the prescribed therapy and goals for the identified pathophysiologic state (Code: IIIH3) [Difficulty: ELE: R, Ap; WRE: An]

As part of the patient care team, you may need to evaluate the patient’s blood gas values and other parameters to make a recommendation. For example, a patient with carbon monoxide poisoning is best treated by administering 100% oxygen through a nonrebreather mask. Pure oxygen reduces the half-life of COHb to 60 to 90 minutes. A first-generation pulse oximeter should not be used to measure this patient’s saturation of oxyhemoglobin. These units are unable to identify COHb and will give misleadingly high saturation values for O2Hb.

ABG analysis remains the “gold standard” by which all other values are judged. Typically, the arterial sample is analyzed through a standard blood gas analyzer. A CO-oximeter is needed if the patient has carbon monoxide poisoning. Mixed venous and capillary blood-sample analysis is limited in clinical application, but very helpful in the right patient situation. However, these offer only momentary insight into the patient’s condition.

Pulse oximetry and transcutaneous oxygen monitoring offer continuous information on the patient’s oxygenation status. Transcutaneous carbon dioxide monitoring allows continuous monitoring of the patient’s ventilation. As with all technology, these units have advantages, disadvantages, and limitations. It is up to the practitioner to make the correct choices. Some unstable patients can best be monitored through the combination of periodic evaluation of ABGs and these noninvasive continuous monitoring systems.

BIBLIOGRAPHY

Aloan CA, Hill TV, editors. Respiratory care of the newborn and child, ed 2, Philadelphia: Lippincott-Raven, 1997.

American Academy of Pediatrics. Task force on transcutaneous oxygen monitors. Pediatrics. 1989;83(1):122.

American Association for Respiratory Care. Clinical practice guideline: blood gas analysis and hemoximetry: 2001 revision and update. Respir Care. 2001;46:498-505.

American Association for Respiratory Care. Clinical practice guideline: transcutaneous blood gas monitoring for neonatal and pediatric patients: 2004 revision and update. Respir Care. 2004;49(9):1069-1072.

American Association for Respiratory Care. Clinical practice guideline: transcutaneous blood gas monitoring for neonatal and pediatric patients. Respir Care. 1994;39(12):1176.

American Association for Respiratory Care. Clinical practice guideline: capillary blood gas sampling for neonatal and pediatric patients. Respir Care. 1994;39(12):1180.

American Association for Respiratory Care. Clinical practice guideline: oxygen therapy in the acute care hospital. Respir Care. 1991;36(12):1410.

American Association for Respiratory Care. Clinical practice guideline: pulse oximetry. Respir Care. 1991;36(12):1406.

American Association for Respiratory Care. Clinical practice guideline: sampling for arterial blood gas analysis. Respir Care. 1991;37(8):913.

American Association for Respiratory Care. Clinical practice guideline: in-vitro pH and blood gas analysis and hemoximetry. Respir Care. 1993;38(5):505.

Barnes TA, editor. Textbook of respiratory care practice, ed 2, St Louis: Mosby, 1994.

Blanchette T, Dziodzio J, Harris K. Pulse oximetry and normoximetry in neonatal intensive care. Respir Care. 1991;36(1):25.

Bohn DJ. Ask the expert. Respiratory Tract. 9, Feb 1988.

Branson RD, Hess DR, Chatburn RL, editors. Respiratory care equipment, ed 2, Philadelphia: Lippincott Williams & Wilkins, 1999.

Burton GC, Hodgkin JE, Ward JJ, editors. Respiratory care: a guide to clinical practice, ed 4, Philadelphia: LippincottRaven, 1997.

Cairo JM, Pilbeam SP. Mosby’s respiratory care equipment, ed 8. St Louis: Mosby, 2010.

Czervinske MP. Arterial blood gas analysis and other cardiopulmonary monitoring. In Koff PB, Daily KE, Schroeder JS, editors: Techniques in bedside hemodynamic monitoring, ed 4, St Louis: Mosby, 1989.

Eitzman D, Neu J. Neonatal and pediatric respiratory care, ed 2. St Louis: Mosby, 1993.

Elser RC. Quality control of blood gas analysis: a review. Respir Care. 1986;31(9):807.

Federal government releases CLIA ‘88 final regulations. AARC Times. 1992;16(4):76.

Fell WL. Sampling and measurement of blood gases. In: Lane EE, Walker JF, editors. Clinical arterial blood gas analysis. St Louis: Mosby, 1987.

Fink JB, Hunt GE, editors. Clinical practice in respiratory care. Philadelphia: Lippincott-Raven, 1999.

Garza D, Becan-McBride K. Phlebotomy handbook, ed 4. Stamford, Conn: Appleton & Lange, 1996.

Instrumentation Laboratories: Operator’s manual for the IL282 CO-oximeter, Lexington, Mass.

Jubran A, Tobin MJ. Reliability of pulse oximetry in titrating supplemental oxygen therapy in ventilator-dependent patients. Chest. 1990;97:1420.

Lane EE, Walker JF. Clinical arterial blood gas analysis. St Louis: Mosby, 1987.

Levitzky MG, Cairo JM, Hall SM. Introduction to respiratory care. Philadelphia: WB Saunders, 1990.

Madama VC. Pulmonary function testing and cardiopulmonary stress testing, ed 2. Albany, NY: Delmar, 1998.

Mahoney JJ, Hodgkin JE, Van Kessel AL. Arterial blood gas analysis. In Burton GG, Hodgkin JE, Ward JJ, editors: Respiratory care: a guide to clinical practice, ed 4, Philadelphia: Lippincott-Raven, 1997.

Malley WJ. Clinical blood gases: assessment and intervention, ed 2. St. Louis: Elsevier Saunders, 2005.

Martin RJ. Transcutaneous monitoring: instrumentation and clinical applications. Respir Care. 1990;35(6):577.

Mathews P, Conway L. Arterial blood gases and noninvasive monitoring of oxygen and carbon dioxide. In: Wyka KA, Mathews PJ, Clark WF, editors. Foundations of respiratory care. Albany, NY: Delmar, 2002.

Mohler JG, Collier CR, Brandt W. Blood gases. In: Clausen JL, editor. Pulmonary function testing guidelines and controversies. Orlando: Grune & Stratton, 1984.

Moran RF. Assessment of quality control of blood gas/pH analyzer performance. Respir Care. 1981;26(6):538.

Moran RF. CLIA regulations. I. The cure might be worse than the disease. AARC Times. 1990;14(11):41.

Moran RF. CLIA regulations. II. An analysis of some technical requirements. AARC Times. 1990;14(12):25.

Nelson CM, Murphy EM, Bradley JK, et al. Clinical use of pulse oximetry to determine oxygen prescriptions for patients with hypoxemia. Respir Care. 1986;31(8):673.

Novametrics Medical Systems: product literature on transcutaneous monitoring, Wallingford, Conn.

Peters JA, Hodgkin JE, Collier CA. Blood gas analysis and acid-base physiology. In Burton GG, Hodgkin JE, Ward JJ, editors: Respiratory care: a guide to clinical practice, ed 3, Philadelphia: JB Lippincott, 1991.

Ruppel G. Manual of pulmonary function testing, ed 9. St Louis: Mosby, 2009.

Salyer JW. Pulse oximetry in the neonatal intensive care unit. Respir Care. 1991;36(1):17.

Walsh B, Czervinske MP, Diblasi RM. Perinatal and pediatric respiratory care, ed 3. St. Louis: Saunders, 2010.

Wilkins RL. Analysis and monitoring of gas exchange. In Wilkins RL, Stoller JK, Kacmarek RM, editors: Egan’s fundamentals of respiratory care, ed 9, St Louis: Mosby, 2009.

Shapiro BA, Kacmarek RM, Cane RD, et al, editors. Clinical application of respiratory care, ed 4, St Louis: Mosby, 1991.

Shapiro BA, Peruzzi WT, Kozlowski-Templin R. Clinical application of blood gases, ed 5. St Louis: Mosby, 1994.

Sonnesso G. Are you ready to use pulse oximetry? Nursing. 1991;21(8):60.

Walton JR, Shapiro BA. Value and application of temperature-compensated blood gas data. Respir Care. 25(2), 1980.

Welch JP, DeCesare R, Hess D. Pulse oximetry: instrumentation and clinical applications. Respir Care. 1990;35(6):584.

Whitaker K. Comprehensive perinatal and pediatric respiratory care, ed 2. Albany, NY: Delmar, 1997.

White GC. Equipment theory for respiratory care, ed 3. Albany, NY: Delmar, 1999.

SELF-STUDY QUESTIONS FOR THE ENTRY LEVEL EXAM See page 583 for answers

SELF-STUDY QUESTIONS FOR THE WRITTEN REGISTRY EXAM See page 606 for answers

Share this: