Anesthetic Monitoring

Published on 24/02/2015 by admin

Filed under Anesthesiology

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 1388 times

Chapter 17 Anesthetic Monitoring

Answers*

Systemic blood pressure monitoring

6. Automated oscillometric blood pressure monitoring devices work by inflating a pneumatic cuff encircling a limb until arterial blood flow through the limb is occluded. The cuff is then deflated until pressure oscillations are detected. The pressure at which oscillations are initially detected is considered to be the systolic blood pressure. The cuff continues to deflate and the oscillations increase for a time and then begin to decrease. The diastolic pressure is defined as the point at which further deflation of the cuff provides no further evidence of pressure oscillations. The most reliable blood pressure parameter measured by this noninvasive blood pressure monitoring device is the mean arterial blood pressure. (321-322, Figure 20-3)

7. The appropriate cuff size for use with a noninvasive blood pressure measuring device is one whose width is about 40% of the circumference of the patient’s limb. (322)

8. When using an automated oscillometric blood pressure measuring device, the blood pressure will be falsely high when the blood pressure cuff is too small. Conversely, the blood pressure will be falsely low when the blood pressure cuff is too large. (322)

9. Cycling an automated oscillometric blood pressure measuring device too frequently can result in limited perfusion to the extremity distal to the cuff. Complications such as edema, nerve paresthesia, superficial thrombophlebitis, and compartment syndrome have all been reported as a result of noninvasive blood pressure devices that have been repeatedly cycled. These complications are rare. (321)

10. Possible indications for intraarterial blood pressure monitoring include the need for continuous blood pressure monitoring, access for frequent arterial blood gas samplings, need for monitoring intentional pharmacologic cardiovascular manipulation, and failure of indirect blood pressure measurement. (322)

11. Arteries that may be used for intraarterial blood pressure monitoring include the radial, ulnar, brachial, axillary, femoral, dorsalis pedis, and the superficial temporal arteries. Of these, the radial artery is the most frequently used artery for cannulation. (323, Table 20-3)

12. The waveform from an intraarterial catheter changes progressively with increasing distance from the heart. The waveform peak is higher and the trough lower at more distal arterial sites. The mean arterial pressure, however, remains approximately the same. (323, Figure 20-4)

Central venous pressure monitoring

13. Indications for the placement of a central venous catheter include the measurement of central venous pressures, access through which to provide long-term intravenous feedings, access for the administration of large volumes of fluids, intravascular access when no peripheral access is available, the administration of vasoactive or caustic drugs, to initiate transvenous cardiac pacing, for temporary hemodialysis, and for the aspiration of air emboli. (324)

14. Veins that are cannulated for central venous access include the internal jugular, subclavian, femoral, and antecubital veins. Potential complications of cannulation of the central veins include arterial puncture, hematoma, hemothorax, pneumothorax, nerve injury, emboli, cardiac dysrhythmias, thrombosis, and infection. Accidental arterial puncture while attempting cannulation of the jugular vein can result in the need to surgically explore and repair the artery. A pneumothorax occurs more frequently after placement of a subclavian catheter. This is the basis for the recommendation that a chest radiograph be done after failed subclavian catheterization and before attempting catheterization on the other side. (323, Table 20-4)

15. The right internal jugular vein is preferred over the left jugular vein for cannulation because of its short, straight, valveless route to the superior vena cava. (323)

16. Advantages of cannulation of the internal jugular vein include its predictable anatomic location with palpable landmarks, its location at the head of the patient’s bed allowing the anesthesiologist easy access to the catheter intraoperatively, and the relatively decreased complications associated with cannulation of this central vein. Disadvantages of cannulation of the internal jugular vein include the potential for puncture of the carotid artery and pleural cavity and trauma to the brachial plexus. (323)

17. Advantages of cannulation of the subclavian vein include its landmarks, its capacity to remain patent despite hypovolemia, easier nursing care, and the relative increase in patient comfort associated with cannulation of this central vein. Disadvantages of cannulation of the subclavian vein include the potential for puncture of the subclavian artery and pleural cavity and for thoracic duct damage on the left. (323)

18. The central venous pressure waveform has a typical trace in a normally functioning heart. The a wave correlates with atrial contraction, the c wave correlates with closure of the tricuspid valve and its bulging into the right atrium, and the v wave correlates with blood accumulation in the vena cava and right atrium against a closed tricuspid valve. The x descent correlates with atrial relaxation, and the y descent correlates with opening of the tricuspid valve and right ventricular filling. (324, Figure 20-5)

19. The central venous pressure parallels right atrial pressure in a patient with normal cardiovascular physiology. In these patients, the central venous pressure can be used to estimate the patient’s intravascular fluid volume status. (324)

20. The central venous pressure does not estimate the patient’s intravascular fluid volume status in the face of right-sided heart dysfunction, left ventricular dysfunction, or pulmonary hypertension. Under these conditions, a pulmonary artery catheter may be used for cardiovascular monitoring. (324)

Pulmonary artery catheter monitoring

21. Possible indications for the placement of a pulmonary artery catheter perioperatively include poor left ventricular function, valvular heart disease, recent myocardial infarction, adult respiratory distress syndrome or any pulmonary vascular disease process, massive trauma, and major vascular surgery. In general, the pulmonary artery catheter allows for more accurate assessment of cardiac filling pressure than a central venous monitor in the presence of pulmonary vascular disease, left-sided heart dysfunction, or potential left-sided heart dysfunction due to myocardial ischemia. The pulmonary artery catheter also measures cardiac output and calculates systemic and pulmonary vascular resistance. (324, Table 20-5)

22. The pulmonary capillary wedge pressure is a reflection of left atrial pressure. The pulmonary artery diastolic pressure may be used as an approximation of left atrial pressure in lieu of the pulmonary artery wedge pressure. This allows for continuous monitoring. The pulmonary artery diastolic pressure does not accurately reflect left atrial pressure in conditions in which pulmonary vascular resistance is increased, as with hypoxia, hypercarbia, hypothermia, and various forms of pulmonary disease. (324, Figure 20-6)

23. Cardiac output can be estimated through the use of a pulmonary artery catheter via the thermodilution method. To do this, cold saline is rapidly injected through the proximal central venous port. A thermistor located at the distal end of the pulmonary artery catheter senses the change in temperature. Because blood flow is the source of dilution of temperature, the flow, or cardiac output, can be calculated. It is the right ventricular cardiac output that is actually measured by this technique, whereas left ventricular cardiac output can only be estimated based on the results. (325)

24. Potential complications of pulmonary artery catheterization include pulmonary ischemia or infarction from prolonged wedging of the catheter, cardiac dysrhythmias, infection, catheter knotting, and, rarely, pulmonary artery rupture. (324)

25. (325, Table 20-6)

image

Electrophysiologic monitoring

34. An evoked potential is a measured low amplitude signal from the central nervous system that occurs in response to sensory or motor nerve stimulation. Evoked potentials that can be monitored include visual, auditory, sensory, and motor. (328)

35. Evoked potentials can be used intraoperatively to assess the integrity of the neural pathways during anesthesia. The most common evoked potentials monitored intraoperatively are somatosensory evoked potentials from the spinal cord during surgery on the spinal cord or vertebral column. (328)

36. Evoked potentials may undergo changes in the latency period and amplitude while patients are under general anesthesia. These changes are similar to the changes that are seen with neural ischemia, which can complicate interpretation of the evoked potential values. Limiting the minimum alveolar concentration (MAC) of volatile anesthetics to 0.5 to 0.75 facilitates monitoring of evoked potentials. Opioids and propofol have less of an effect on evoked potentials, and muscle relaxants do not affect somatosensory evoked potentials at all. (328)

37. Factors that may limit the intraoperative usefulness of evoked potentials because of their influence on the results include age and gender of the patient, arterial blood gas tensions, and body temperature. In addition, the cost and complexity of performing evoked potentials may limit their use. (328)

38. Somatosensory evoked potentials of the lower extremities monitor the integrity of the dorsal column of the spinal cord. Motor evoked potentials monitor the corticospinal tract. Unlike somatosensory evoked potentials, motor evoked potentials are sensitive to muscle relaxants. (328)

Capnography monitoring

39. A capnograph is a waveform display that illustrates the patient’s inhaled and exhaled concentrations of carbon dioxide. (328-329)

40. In the capnogram, the point A designates the exhalation of anatomic dead space gas just before the exhalation of alveolar gas. Point B designates the beginning of exhalation of alveolar gas that contains carbon dioxide. Phase C-D designates the exhalation of alveolar gas, while point D designates the end-tidal carbon dioxide concentration. Phase D-E designates the beginning of inspiration and the entrainment of inspired gases. (329, Figure 20-10)

41. The absence of carbon dioxide in a patient’s exhaled gases just after attempted endotracheal intubation with properly functioning equipment provides evidence that the patient’s lungs are not being ventilated. That is, the endotracheal tube may not be in the trachea. The absence of carbon dioxide in a patient’s exhaled gases after intubation of the trachea has been confirmed may indicate that there is either a malfunction of equipment, a malfunction in the interface between the patient and the equipment (as in disconnection from the anesthesia circuit), movement or dislodgment of the endotracheal tube from its previously proper position, or a physiologic patient problem such as a cardiac arrest. (328)

42. Possible causes of a decrease in the patient’s exhaled concentration of carbon dioxide include hyperventilation, hypothermia, low cardiac output, pulmonary embolism, accidental disconnection, tracheal extubation, or cardiac arrest. (328, Table 20-9)

43. Possible causes of an increase in the patient’s exhaled concentration of carbon dioxide include hypoventilation, hyperthermia, sepsis, rebreathing, the administration of bicarbonate, and the insufflation of carbon dioxide during laparoscopy. (328, Table 20-9)

44. The end-tidal carbon dioxide concentration measured on a capnogram is less than the true arterial concentration of carbon dioxide, typically by a 2- to 5-mm Hg gradient. This occurs as a result of the alveolar-to-arterial difference for carbon dioxide concentrations secondary to dead space ventilation. (328)

Electroencephalographic monitoring

45. Intraoperative uses of an electroencephalogram include monitoring for cerebral ischemia and monitoring the depth of anesthesia. (329)

46. Among the factors that influence the tracings obtained by an electroencephalogram and limit its usefulness intraoperatively are anesthetics, changes in body temperature, and alterations in the arterial carbon dioxide concentration. (329)

47. The bispectral index monitor performs a bispectral analysis of the electroencephalogram and provides the clinician with a processed evaluation of its analysis through its display of a number between 0 to 100. The analysis is done through superficial scalp electrodes typically on the forehead of the patient. The number provided by the bispectral index monitor reflects the state of wakefulness of the central nervous system. (329, Figure 20-11)

48. The bispectral index monitor may be used clinically to predict loss of consciousness and lack of recall during anesthesia. A bispectral index numerical value of 0 is consistent with an isoelectric encephalogram. A numerical value of 60 or less corresponds to a low probability of recall or awareness. Thus the use of the bispectral index monitor for the titration of medicines to achieve adequate but not excessive loss of consciousness may result in more rapid awakening at the end of the procedure. The bispectral index has not been shown to be well correlated with the hemodynamic or movement responses to noxious stimuli. In addition, a recent study has shown that the use of a bispectral index monitor showed no decrease in the incidence of awareness when using volatile anesthetics. (329)