Advanced Cardiopulmonary Monitoring

Published on 12/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 5 (1 votes)

This article have been viewed 7238 times

5 Advanced Cardiopulmonary Monitoring

Note 1: This book is written to cover every item listed as testable on the Entry Level Examination (ELE), Written Registry Examination (WRE), and Clinical Simulation Examination (CSE).

The listed code for each item is taken from the National Board for Respiratory Care’s (NBRC) Summary Content Outline for CRT (Certified Respiratory Therapist) and Written RRT (Registered Respiratory Therapist) Examinations (see http://evolve.elsevier.com/Sills/resptherapist/). For example, if an item is testable on both the ELE and the WRE, it will simply be shown as: (Code: …). If an item is only testable on the ELE, it will be shown as: (ELE code: …). If an item is only testable on the WRE, it will be shown as: (WRE code: …).

Following each item’s code will be the difficulty level of the questions on that item on the ELE and WRE. (See the Introduction for a full explanation of the three question difficulty levels.) Recall [R] level questions typically expect the exam taker to recall factual information. Application [Ap] level questions are harder because the exam taker may have to apply factual information to a clinical situation. Analysis [An] level questions are the most challenging because the exam taker may have to use critical thinking to evaluate patient data to make a clinical decision.

Note 2: A review of the most recent Entry Level Examinations (ELE) has shown an average of 2 questions (out of 140), or 1% of the exam, will cover advanced cardiopulmonary monitoring. A review of the most recent Written Registry Examinations (WRE) has shown an average of 3 questions (out of 100), or 3% of the exam, will cover advanced cardiopulmonary monitoring. The Clinical Simulation Examination is comprehensive and may include everything that should be known by an advanced level respiratory therapist.

MODULE A

1. Capnography (exhaled CO2 monitoring)

c. Perform the bedside procedure (Code: IB9c and IIIE3d) [Difficulty: ELE: R; WRE: Ap, An]

The capnometer is a device that measures the concentration of CO2 in a gas sample from a patient. The principle of operation of most bedside units is based on carbon dioxide’s absorption of infrared light in a narrow wavelength band (4.3 μ). Infrared light at this wavelength is passed through the gas sample to a receiving unit. The difference between what is transmitted and what is received is directly related to how much carbon dioxide is in the gas sample. In other words, the greater the difference between the sent and the received infrared light, the greater the concentration of carbon dioxide in the gas sample.

The capnometer is calibrated by comparing a gas sample without carbon dioxide (possibly room air) with a second gas sample containing a known amount of carbon dioxide. The first gas sample without CO2 should give a “zero” reading. Adjust the calibration control to zero if needed. The second sample usually contains 5% to 10% carbon dioxide. The capnometer should display a CO2 level that matches the amount in the known gas sample. Adjust the calibration control as necessary. The carbon dioxide level can be documented as a percent or fraction (FACO2) or as a partial pressure (PACO2).

The capnograph is a strip chart recorder that provides a copy of the patient’s exhaled carbon dioxide curve. There are at least two paper speeds that are useful for different purposes. The fast speed is most useful for evaluating sudden changes in the patient’s condition. Each individual breath is easily seen (Figure 5-1). The slow speed is most useful for trend monitoring (Figure 5-2).

Two different gas sampling methods exist: mainstream and sidestream. The mainstream method involves having the infrared sensing unit at the airway; usually it is attached directly to the endotracheal/tracheostomy tube. If the patient is on a ventilator, the sampling adapter must be placed between the endotracheal tube and the ventilator circuit (with or without mechanical dead space). All inspired and expired gas passes through the sensor (Figure 5-3).

The sidestream method employs a capillary tube placed so that a small sampling of the patient’s exhaled gas can be drawn into the capnometer for analysis. It is not necessary for the patient’s entire breath to pass through the sampling adapter; therefore it can be used in an unintubated patient by taping the sampling catheter a short distance into a nostril. If the patient is on a ventilator, the sampling adapter must be placed between the endotracheal tube and the ventilator circuit (with or without mechanical dead space). Remember that the patient’s exhaled tidal volume (VT) and minute volume (imageE) are reduced by the amount that is drawn into the capnometer (Figure 5-4).

d. Interpret the results from the procedure (Code: IB10c, IIIE4e) [Difficulty: ELE: R, Ap; WRE: An]

It is known that carbon dioxide diffuses from the higher concentration in the tissues to the venous blood and to the lungs to be exhaled. Figure 5-5 shows the normal physiology behind capnography. This diffusion or “flow” of CO2 results in a measurable gradient or difference. In a healthy, upright sitting person, a close relationship exists between carbon dioxide levels in both venous blood and arterial blood and the amount of exhaled carbon dioxide gas. The carbon dioxide level at the end of exhalation is most frequently monitored during patient care. This is called the end-tidal carbon dioxide pressure (PetCO2). When ventilation and perfusion match well, as in a healthy upright person, the gradient between the arterial carbon dioxide level (PaCO2) and the PetCO2 is between 2 and 3 torr, with a range of 1 to 5 torr. The gradient will show the PetCO2 level to be less than the PaCO2 level. This is because the PetCO2 value is an average of exhaled carbon dioxide levels from all lung areas.

Box 5-1 lists normal values of capnography. Three factors influence capnography’s use and the interpretation of the results. (1) The first factor is the patient’s metabolism. The average resting adult produces about 200 mL of CO2 per minute, and fever and exercise increase this value. Hypothermia, sleep, and sedation decrease CO2 production. Exhaled CO2 is monitored during a cardiopulmonary resuscitation (CPR) attempt to determine the effectiveness of circulation and ventilation attempts and to decide if the efforts should be continued or stopped. If no carbon dioxide is being exhaled despite proper CPR procedures, the physician may conclude that the patient’s metabolism has stopped altogether and death has occurred. There would then be nothing to gain by continuing cardiopulmonary resuscitation (CPR) efforts.

(2) Although not a major factor, the patient’s cardiac output is another factor that influences the use of capnography. Sepsis, which might double a patient’s cardiac output, reduces the PCO2 level only a few torr (millimeters of mercury [mm Hg]). Cardiogenic shock, which reduces the cardiac output, raises the partial pressure of CO2 (PCO2) only a few torr.

(3) The third and most important factor is alveolar ventilation. A doubling of alveolar ventilation, under steady-state conditions for carbon dioxide production, results in a halving of the PCO2 levels in arterial blood and alveolar gas. However, a reduction of alveolar ventilation to half of its previous level will result in the PaCO2 and PACO2 levels being doubled (Figure 5-6).

image

Figure 5-6 Relationship between alveolar ventilation, PaCO2, and exhaled percent CO2.

(From Pilbeam SP: Mechanical ventilation: physiological and clinical applications, ed 4, St Louis, 2006, Mosby.)

Tidal volume (VT) and respiratory rate are directly related to alveolar ventilation. Of the two, tidal volume is more important because it relates to the patient’s dead space (VD) to VT ratio (VD/VT). A decrease in the patient’s VT level results in less alveolar ventilation and a rise in PCO2 level. Conversely, an increase in the VT level results in more alveolar ventilation and a drop in the PCO2 level.

Capnography is most accurate and correlates best with the PaCO2 level if the patient’s ventilation and perfusion match. The more ventilation and perfusion mismatching there is, or the more unstable the pulmonary perfusion, the wider or less reliable is the gradient between the patient’s arterial carbon dioxide and alveolar carbon dioxide levels. The following procedures should be performed to help understand the patient’s condition and interpret the capnography results.

Remember that mm Hg (millimeters of mercury) and torr (Torricelli) are equivalent units of pressure. The National Board of Respiratory Care (NBRC) uses these two units for different items in its questions. It uses mm Hg for blood pressure measurements and torr for blood gas values (such as PaCO2 and PimageO2). However, in questions regarding capnography, the NBRC has used both mm Hg and torr in its questions that relate to exhaled CO2 (such as PACO2 and PetCO2).

2. Arterial–end-tidal carbon dioxide gradient

a. Perform the bedside procedure (Code: IB9c and IIIE3d) [Difficulty: ELE: R; WRE: Ap, An]

The arterial–end-tidal carbon dioxide gradient [P(a-et)CO2] is useful, because once it is reliably determined the patient’s ventilatory condition can be monitored by capnography alone. Drawing an arterial blood sample to measure the PaCO2 level is less necessary if the patient is stable.

The normal gradient is 2 to 3 torr, with a range of 1 to 5 torr. However, most patients using capnography are not normal. The possible gradient ranges from 6 to 20 torr in unstable patients with cardiopulmonary abnormalities. For example, the patient who is breathing shallowly may have an end-tidal CO2 level that is greater than the arterial CO2 level. This is because the patient is not exhaling completely to empty alveolar gas. Therefore determining each patient’s own gradient is important.

Follow these steps:

b. Interpret the results from the procedure (Code: IB10c, IIIE4e) [Difficulty: ELE: R, Ap; WRE: An]

Review the components of a fast-speed capnography tracing in Figure 5-1 to understand a normal person’s expiratory pattern. Figure 5-7 shows eight different abnormal fast-speed capnography tracings. See the figure legend for an explanation of each problem. As the patient returns to normal, the tracing should approach that shown in Figure 5-1.

image

Figure 5-7 A series of abnormal fast-speed capnography tracings. A, A mechanically ventilated patient with a malfunctioning exhalation valve. Note that the baseline CO2 level is elevated because the patient’s exhaled breath is measured during an inspiration. Correction of the exhalation valve should result in normal inhalation and exhalation. B, This rapidly rising baseline gas pressure and failure to return to baseline is usually seen when moisture or secretions block the capillary tube. Clearing the obstruction enables the patient’s gas to again reach the analyzer. C, Distortions in the tracing from incomplete exhalation. These may be caused by hiccups, chest compressions during CPR, or inconsistent tidal volume efforts during an asthma attack. D, An obstructive lung disease patient with ventilation and perfusion mismatching. There is no alveolar plateau with a stable CO2 level. Inhaling a bronchodilator should result in the tracing returning closer to normal. E, A patient with restrictive lung disease showing no plateau of alveolar gas emptying. This is because the alveoli do not empty evenly. F, A sudden drop in carbon dioxide level in the middle of an exhalation indicates that the patient attempted inspiration. This “cleft” is usually seen when a patient who has been pharmacologically paralyzed begins to regain movement. G, Uneven carbon dioxide levels seen at the end of exhalation can be caused by the following: (1) the patient’s heartbeat pumping fresh blood and CO2 to the emptying lungs; the cardiogenic oscillations should match the heart rate; (2) the ventilator’s exhalation valve is fluttering open and closed. H, The alveolar plateau is biphasic. This has been seen in patients with lungs that are different in compliance and ventilation/perfusion matching (e.g., single lung transplantation).

(From Shapiro BA, Peruzzi WT, Templin R: Clinical application of blood gases, ed 5, St Louis, 1994, Mosby.)

3. Arterial–residual volume alveolar carbon dioxide gradient

b. Interpret the results from the procedure (Code: IB10c, IIIE4e) [Difficulty: ELE: R, Ap; WRE: An]

When comparing the normal (solid line) tracing in Figure 5-8 with the image/image mismatching (dashed line) tracing, note the increased gradient at end-tidal CO2. With continued exhalation to residual volume, the left heart failure and COPD patients have a narrowing of the gradient. This can be used clinically to follow these patients’ progress and response to treatment. The patient with a large pulmonary embolism will not have such a narrowing of the gradient as he or she exhales to residual volume. This patient’s gradient will narrow to normal as the embolism is resolved and the physiologic dead space returns to normal.

Past exams have often had one question that requires the interpretation of capnography results, especially the end-tidal CO2 value (PetCO2). Examples include a change in alveolar ventilation, shallow breathing, and CPR attempt.

MODULE B

2. Perform the bedside procedure (ELE Code: IB9l) [Difficulty: ELE: R, An]

The procedure is the mathematical comparison of a person’s dead space volume with tidal volume. Steps in the traditional procedure follow (Figure 5-9):

b. Determine the dead space volume

Place both carbon dioxide values into this formula, which is derived from the original Bohr formula:

image

In which:

VD VT or VD the patient’s physiologic dead space

VT = the average exhaled tidal volume

PaCO2 = the patient’s arterial carbon dioxide pressure

PĒCO2 = the patient’s average exhaled carbon dioxide pressure

The following example is based on a normal adult:

image

The patient’s physiologic VD volume = 150 mL. (Note that this equation must be used when the patient’s tidal volume is known.)

In addition, the preceding equation allows calculation of the patient’s VD/VT ratio. In this example, it is 150/500, 0.3, or 30%, depending on how it is written.

The recent advent of volumetric capnography technology allows the patient’s dead space to be rapidly determined. These units can simultaneously measure the patient’s exhaled tidal volume and the variable percentage of carbon dioxide found during the exhalation (Figure 5-10). The computer that is integrated with the volumetric capnography unit then calculates the volume of dead space as a fraction of the exhaled tidal volume. Figure 5-11 shows how volumetric capnography can be set up to measure the dead space of a patient requiring mechanical ventilation. This technology permits the rapid assessment of a patient as treatment is being performed. For example, if a patient had a large pulmonary embolism resulting in significant dead space, a clot-dissolving medication such as streptokinase (Kabikinase or Streptase) or alteplase (Activase) would be given. These “clot buster” medications will rapidly dissolve the patient’s pulmonary embolism. With volumetric capnography, the dead space volume can be monitored as it normalizes when blood flow through the lungs is restored.

3. Interpret the results of the VD/VT calculation (Code: IB10l) [Difficulty: ELE: R, Ap; WRE: An]

The normal adult’s VD/VT ratio ranges from 0.2 (20%) to 0.4 (40%). Anatomic dead space is normally greater in men than in women. The normal 3-kg neonate’s dead space to tidal volume ratio is 0.3 (30%). Physiologic dead space (also known as respiratory dead space) is gas that is ventilated into the lungs but does not take part in gas exchange. This is because the alveoli are not perfused or are underperfused for the amount of gas that they receive.

Physiologic dead space is composed of the following:

The following conditions or pulmonary disorders can cause the ratio to vary from the normal range:

MODULE C

1. Blood pressure

c. Perform blood pressure measurement (Code: IB9m, IIIE3e) [Difficulty: ELE: R, Ap; WRE: An]

The general steps in measuring blood pressure were described in Chapter 1. The BP is usually measured on either of the patient’s arms. Necessary equipment includes the proper size of blood pressure cuff, a sphygmomanometer to measure the pressure, and a stethoscope to hear the sounds of bloodflow returning through the brachial artery. Figure 5-12 shows the basic elements of the procedure. The cuff is inflated to a pressure that is greater than the patient’s systolic pressure. As the pressure in the cuff is gradually decreased, the first sound heard (Korotkoff sounds) is the systolic pressure. The pressure reading at which this sound ceases is the diastolic pressure. Blood pressure measurement can usually be performed manually by a respiratory therapist or other trained health care professional. If BP measurement is needed on a frequent basis, an automated blood pressure measurement system can be set up on the patient’s arm. This unit can be programmed to measure the BP on a schedule and can also have high and low blood pressure alarm limits established as a safety feature.

2. Central venous pressure (CVP) monitoring

a. Review central venous pressure measurement data in the patient’s chart (Code: IA8b) [Difficulty: ELE: R; WRE: Ap]

The central venous pressure (CVP) is the pressure measured in a patient’s superior vena cava, just above the right atrium. See Figure 5-13. Review previous patient data to understand whether there is an abnormality. Compare the current data with the earlier information to determine if there has been a change in the patient’s condition. See Box 5-2 for normal values.

d. Perform central venous pressure measurement (Code: IB9e) [Difficulty: R, Ap, An]

The catheter is usually inserted into the right jugular vein or right subclavian vein and advanced to just above the superior vena cava. When set up for monitoring pressure, it measures the right atrial pressure (see Figure 5-15 for how to perform the procedure). Note that the stopcock must be kept at the midchest (midheart) level. Usually a mark is placed at this location (on the patient’s chest) for consistency. Raising the stopcock above the mark results in an incorrectly low reading. Lowering the stopcock below the mark results in an incorrectly high reading. See Figure 5-16. Clinical practice is very important in learning how to perform this procedure. It is important that the patient breathes spontaneously if at all possible. Peak pressures during inspiration on a mechanical ventilator may artificially raise the CVP reading. Positive end-expiratory pressure (PEEP) may also raise the CVP reading. If the patient cannot be removed from the ventilator, take the reading during exhalation. Make a note of the settings and that the reading was taken with the patient on the ventilator. Record the data in the patient’s chart or flow sheet.

Past exams have asked about the proper placement of the CVP stopcock or arterial pressure transducer at the midchest (midheart) location to ensure accurate pressure measurements. Usually a mark is placed on the patient’s chest for a consistent measurement point. Review Figure 5-16 for the effects of misplacement.

e. Interpret the results of the central venous pressure measurement (Code: IB10m, IIIE4c) [Difficulty: ELE: R, Ap; WRE: An]

As noted earlier, the CVP is a measure of the pressure in the right atrium. The two main factors that influence the right atrial pressure are the blood volume returning to it and the functioning of the right ventricle (see Box 5-2 for normal CVP readings).

A decreased CVP reading usually indicates that the patient is hypovolemic. Hypotension confirms this. An increased CVP reading may suggest one of the following possibilities:

Exam Hint 5-6 (ELE, WRE)

Memorize the values listed in Box 5-2. Expect to see several questions in which these values are used as patient data or as options to answer a question. The normal values must be understood to identify abnormal values and know why they are abnormal.

Memorize the values listed in Box 5-2. Expect to see several questions in which these values are used as patient data or as options to answer a question. The normal values must be understood to identify abnormal values and know why they are abnormal.

3. Pulmonary artery pressure monitoring

b. Recommend the insertion of a pulmonary artery catheter for additional data (WRE code: IC12) [WRE Difficulty: R, Ap, An]

To read the PAP, a pulmonary artery catheter (PAC) must be inserted through a vein and passed through the right atrium and right ventricle into the pulmonary artery. The common insertion sites, in descending order of preference, are the basilic vein in either the right or the left arm, the right internal jugular or subclavian vein, or the right or left femoral vein. The pulmonary artery catheter is also commonly called a Swan-Ganz catheter after the inventors who gave their names to a particular brand. The catheters are available in different lengths and diameters for pediatric and adult patients. See Figure 5-17 for a typical adult catheter.

d. Perform pulmonary artery pressure measurement (Code: IB9m, IIIE3e) [Difficulty: ELE: R, Ap; WRE: An]

Figure 5-17 is an illustration of a 7-French quadruple-lumen thermodilution pulmonary artery catheter. Besides measurement of PAP, it is capable of being used to measure cardiac output. Figure 5-18 illustrates how a pulmonary artery catheter could be arranged with a pressure transducer and pressure monitor. Figure 5-19 shows a representation of the series of pressure waveforms seen as the catheter is advanced through the heart and into the wedged position in a branch of the pulmonary artery. Figure 5-20 shows a larger cutaway view of the heart with a PAC and normal heart chambers and related pressures.

To obtain an accurate pulmonary artery pressure measurement, the equipment must be set up and calibrated properly, the distal lumen of the catheter must be patent with a continuous fluid-filled channel from the patient’s bloodstream back to the transducer, and the catheter’s balloon must be deflated. Clinical practice is important in understanding how to perform this procedure. Record the PAP data in the patient’s chart or flow sheet.

e. Interpret the results of pulmonary artery pressure measurement (Code: IB10m, IIIE4c) [Difficulty: ELE: R, Ap; WRE: An]

Again, the PAP is the systolic and diastolic pressure found in the pulmonary artery (see Box 5-2 for normal values). Elevated PAP values are usually seen with the following conditions:

Decreased pulmonary artery pressure values are not frequently seen. Patients with hypovolemic shock, anaphylaxis (allergic shock), or excessive use of vasodilating drugs may have a decreased pulmonary artery pressure. The most obvious clinical sign in these patients is a low systemic blood pressure.

Expect to see at least one question requiring the calculation of the PAd-PCWP gradient and at least one question requiring the result to be interpreted.

4. Pulmonary capillary wedge pressure monitoring

c. Perform pulmonary capillary wedge pressure measurement (Code: IB9m, IIIE3e) [Difficulty: ELE: R, Ap; WRE: An]

A pulmonary artery catheter must be placed into a patient’s pulmonary artery for the PCWP to be measured. The PCWP is obtained by inflating the balloon at the tip of the catheter. This obstructs that branch of the pulmonary artery so that the downstream pressure from the left ventricle is seen on the monitor (see Figures 5-19 and 5-20). The balloon volume varies with the diameter of the catheter. The necessary volume is printed on the catheter near where the air is injected into the balloon. For example, the 5-French catheter balloon holds 0.8 mL of air and the 7-French catheter balloon holds 1.5 mL of air. It is important to inject only the required amount of air. Overinflating may burst the balloon or rupture the pulmonary artery. If the balloon wedges at less than the required volume, the catheter is probably too far down in the artery and may need to be withdrawn a short distance. The balloon is only inflated long enough to obtain the PCWP and then the balloon is deflated. Pulmonary infarction will occur if the balloon is left inflated and the blood in the pulmonary artery is stagnant and allowed to clot. Record the PCWP value in the patient’s chart or flow sheet.

d. Interpret the results of pulmonary capillary wedge pressure measurement (Code: IB10m, IIIE4c) [Difficulty: ELE: R, Ap; WRE: An]

As noted earlier, the PCWP refers to the pressure measured in the pulmonary capillary bed under no-flow conditions. This pressure reflects downstream pressure from the left side of the heart. At diastole, in the patient without pulmonary hypertension or mitral valve disease, the PCWP parallels left atrial pressure (LAP) and left ventricular end-diastolic pressure (LVEDP). The literature reveals that a variety of terms and initials are used to describe the same physiologic value. Do not be confused by reading about the pulmonary capillary pressure (PCP), pulmonary wedge pressure (PWP), pulmonary artery wedge pressure (PAWP), or wedge pressure.

Elevated PCWP is generally considered to be greater than 10 mm Hg and can indicate the following conditions:

These conditions result in serious problems for the patient’s lung function. When the PCWP reaches 20 to 25 mm Hg, fluid begins to leak into the pulmonary interstitium. This makes the lungs less compliant and increases the patient’s work of breathing. A PCWP of 25 to 30 mm Hg results in frank pulmonary edema and dramatically decreases the patient’s Pao2 level.

Decreased PCWP is generally considered to be less than 4 mm Hg and can indicate the following conditions:

5. Cardiac output

c. Perform a cardiac output measurement (Code: IB9m, IIIE3e) [Difficulty: ELE: R, Ap; WRE: An]

At the time of this book’s publication, there are several cardiac output methods that are commonly utilized in the cardiac catheterization laboratory. They will not be presented here. This discussion is limited to intensive care unit bedside procedures where a respiratory therapist is likely to be involved. A thermodilution cardiac output study involves using one of two special four-lumen pulmonary artery catheters. Both catheter types require additional hardware and supplies, including a computer designed to calculate the CO.

The first type of thermodilution catheter is used to inject a cool solution into the heart. This cool solution is diluted with the patient’s blood. As the blood and cool solution mix passes a thermistor at the catheter tip, the computer calculates the patient’s cardiac output based on the time required to pump the cooler blood past the thermistor (see Figure 5-17).

This procedure requires an injectable saline solution or a 5% dextrose, ice-water bath to cool the injectate, necessary tubing and connections, thermometer, 10-mL syringes, and injector system (Figure 5-22). It is beyond the scope of this text to describe all of the steps in the different types of adult and neonatal thermodilution cardiac output procedures. Hands-on experience is necessary. Only the most important and common features are presented here:

image

Figure 5-22 Complete system for performing thermodilution cardiac output studies.

(From Edwards Lifesciences LLC, Irvine, Calif, 1985.)

The “cool solution” thermodilution cardiac output procedure has been available for more than 20 years and is widely used clinically. Its main disadvantages are that the CO value is available only intermittently, it is a time-consuming procedure, and patients with heart failure are given significant amounts of additional fluid for each cardiac output calculation.

The second type of thermodilution cardiac output catheter uses a thermal filament (a heated wire) that wraps around the catheter (Figure 5-23). With this catheter, the computer periodically directs electricity to the filament. This results in periodic warming of the blood from the heated wire. The computer calculates the patient’s cardiac output by determining the time needed to pump the heated blood past the thermistor at the tip of the catheter. The advantages of the “heated wire” cardiac output catheter are that it gives an updated cardiac output value every 30 seconds, requires no additional time from the nurse or respiratory therapist after the initial setup, and does not add any additional fluid to the patient’s intake.

d. Calculate the patient’s cardiac output value (ELE code: IB9l) [Difficulty: ELE: R, Ap]

The following formula can be used to calculate a predicted cardiac output for the adult patient. This value can then be compared with the actual patient cardiac output.

image

BSA is the body surface area and can be calculated mathematically or determined from a data table (Figure 5-24). See the following discussion on cardiac index for information on calculating BSA.

The following two methods can be used to calculate the patient’s cardiac output:

In which:

image

Therefore:

image

6. Stroke volume

7. Cardiac index

8. Mixed venous blood sampling

c. Perform mixed venous oxygen measurement (Code: IB10m, IIIE4c) [Difficulty: ELE: R, Ap; WRE: An]

A patient with a functioning pulmonary artery catheter (PAC) can have a sample of blood withdrawn through it and analyzed for the mixed venous oxygen (PimageO2) value (and the other blood gas values as well). There are currently three methods of performing the bedside procedure.

2. Use a pulmonary artery catheter with reflectance oximetry capability

These catheters have fiberoptic bundles built into them and use technology similar to that used in pulse oximeters. See Figure 5-25. The processing unit sends two narrow wavebands of light down the transmitting fiberoptic bundle to be illuminated on the passing blood in the pulmonary artery. Oxyhemoglobin in the red blood cells absorbs some of the light. The rest is reflected. The receiving fiberoptic bundle picks up some of this light and transmits it back to the monitoring unit.

SimageO2 is determined by the monitoring unit based on the light waves that were transmitted and received. Care must be taken when using this catheter in patients with elevated carboxyhemoglobin or methemoglobin levels. As with pulse oximetry, COHb and MetHb will be interpreted as oxyhemoglobin. Thus inaccurately high readings will be seen.

An actual mixed venous blood sample can be taken with this catheter through the distal lumen as described next. The advantage of the reflectance oximetry system is that it provides continuous monitoring of the patient’s venous oxygen level. In addition, high and low saturation alarms can be set. If the tip should become lodged in the wall of the artery or if a clot should form at the tip, the SimageO2 value will drop or fluctuate dramatically. The alarms should warn the clinician of a problem with the equipment.

9. Arterial-venous oxygen content difference

10. Shunt study

d. Calculate the patient’s shunt percentage (ELE code: IB9l) [Difficulty: ELE: R, Ap]

To date, the NBRC has not had an examination question that requires all of the calculations needed to determine a patient’s shunt percentage. However, examinations have had questions that relate to calculating components of the shunt equations. These have included calculating CaO2, CimageO2, and C(a-image)O2.

There are a number of possible variations on the methods and equations for calculating shunt percentage. Only the two most commonly used equations are presented here.

1. Modified clinical shunt equation

image

In which:

This formula requires that the patient’s hemoglobin be 100% saturated. This does not happen until the Pao2 level reaches 150 torr at any given FiO2. An obvious clinical limitation is that very ill patients never reach 100% saturation even when breathing 100% oxygen. Also, putting patients on more than 80% oxygen, even for just the duration of the test, may lead to some denitrogenation absorption atelectasis. This results in an incorrectly high shunt percentage calculation.

2. Classic shunt equation

This equation is widely used because of the clinical limitations of the clinical shunt equation:

image

In which:

Cco2 The content of oxygen in the end pulmonary capillary blood

Cao2 = The content of oxygen in the arterial blood

CimageO2 = The content of oxygen in the mixed venous blood

The patient should be inspiring 30% oxygen or more to calculate the oxygen content of end pulmonary capillary blood. This much oxygen should cause the PAO2 value of the ventilated alveoli to reach 150 torr or more, which results in 100% saturation of the hemoglobin of the end pulmonary capillary blood (ScO2). Most patients who are sick enough to warrant a shunt determination will be on at least 30% oxygen. If not, the ScO2 value must be determined by using the oxyhemoglobin dissociation curve.

A pulmonary artery catheter is needed to sample mixed venous blood for PimageO2 and SimageO2 levels in both of these equations. An arterial blood gas sample is also needed. There are clinical situations in which either of these equations will work and will result in the same answer.

11. Pulmonary vascular resistance

12. Systemic vascular resistance

e. Interpret the results of the systemic vascular resistance calculation (Code: IB10l) [Difficulty: ELE: R, Ap; WRE: An]

The normal range of SVR in the adult is 15 to 20 mm Hg/L/min. Multiplying this value by 80 changes the result to units of dynes/sec/cm−5. SVR is listed this way in some cardiology studies. When multiplied by 80, an adult’s SVR ranges between 770 to 900 and 1400 to 1500 dynes/sec/cm−5.

A patient with hypertension will have an elevated SVR. It will also increase if the patient is given a vasoconstricting drug and decrease if the patient is given a vasodilating drug. Some allergic reactions result in anaphylaxis with a dramatic drop in the SVR and blood pressure despite an increase in the cardiac output.

Know the normal values for pulmonary and systemic vascular resistance. To date, the NBRC has not expected the examinee to perform all of the calculations to determine a patient’s PVR or SVR. Know that a patient with COPD or PPHN will have a chronically increased PVR and a patient with a large pulmonary embolism will have a sudden increase in the PVR. With proper treatment, the PVR will decrease toward normal.

There are several ways that the cardiology version of the PVR or SVR values can be found. The preferred versions have units listed in “dynes/sec/cm−5” or “dyne sec/cm−5.” However, the NBRC has also listed these values in units of “dynes seconds cm−5,” “dynes·seconds·cm−5,” and “mm Hg/L/min.”

MODULE D

1. Manipulate capnography equipment by order or protocol (Code: IIA26) [Difficulty: ELE: R; WRE: Ap, An]

a. Get the necessary equipment for the procedure

As discussed earlier, a capnograph is used to measure a patient’s exhaled carbon dioxide level and expiratory pattern. There are two different types of capnography systems. Their main difference is in how the gas is sampled from the patient. Figure 5-3 shows the patient connection for a mainstream capnograph. Figure 5-4 shows a sidestream capnograph. Both work equally well if the patient is intubated. The capnography connector is attached between the endotracheal tube and the ventilator circuit or T-piece (Brigg’s adapter). The sidestream unit may also be used with a spontaneously breathing patient because the capillary tube may be taped into a patient’s nostril for gas sampling.

2. Manipulate pressure transducers by order or protocol (WRE: code: IIA20a) [Difficulty: WRE: R, Ap, An]

b. Put the equipment together and make sure that it works properly

Check to see that the wire screen of the transducer is not bent, damaged, or contaminated with old blood or other debris. The transducer’s wire cable and monitor-connecting prongs should not be bent or damaged. Carefully insert the prongs into the receiving jack on the monitor.

A disposable, sterile, clear plastic dome should be firmly attached to the transducer. Noncompliant pressure tubing should connect the transducer with the patient’s arterial or pulmonary artery catheter. The automatic fluid drip system (Sorenson) should be pressurized to 300 mm Hg. Heparin must be added to the fluid (usually normal saline) to prevent clotting in the patient’s catheter. Because this is a continuous fluid “plumbing” system, all connections must be tightly joined to be watertight. There must be a continuous flow of fluid through the tubing system and into the patient’s blood vessel. Any air bubbles must be removed or the measured pressure reading will be lower than the actual reading (see Figures 5-18 and 5-27).

The transducer must be kept at the patient’s midchest (midheart) level during calibration and measurement. Calibrate the electronics in the monitor by placing known pressures against the fluid system. The electronics should first be adjusted to “zero” pressure by opening the transducer to room air (atmospheric pressure). Adjust the electronic controls as needed. A sphygmomanometer is then used to pressurize the fluid system. A pulmonary artery catheter system is pressurized to relatively low pressures such as 30 and 50 mm Hg. An arterial system is adjusted to higher pressures such as 100 and 150 mm Hg. The monitored pressure should match the sphygmomanometer pressure. If not, adjust the electronic controls on the monitor to match.

4. Indwelling arterial catheters (WRE code: IIA20b) [Difficulty: WRE: R, Ap, An]

b. Put the equipment together and make sure that it works properly

See Figures 5-18 and 5-28 for illustrations of how the pressure transducer, connecting tubing, stopcocks, infusion system, and monitoring electronics are assembled. See Figure 5-29 for a completed radial artery system. Clinical experience is needed with these types of systems.

image

Figure 5-28 Common steps in the assembly of the tubing circuit and pressure transducer to attach to an arterial or pulmonary artery catheter. Sterile technique must be maintained at all times. (NOTE: Individual institutions will vary these steps. Clinical practice is very important in this assembly.) A, Obtain a 250 to 500 mL bag of sterile, normal saline. Add 1 to 2 units of heparin per mL of solution. Attach an IV line with a macrodrip chamber. B, Insert the solution bag into a pressure bag. Inflate the pressure to about 100 mm Hg to force the fluid through the IV tubing. The pressure bag will be inflated up to 300 mm Hg at the end of this procedure to ensure that 3 drops of fluid flow through the tubing per minute to keep it patent. C, Attach the strain gauge pressure transducer to an IV pole. Let it and the monitor warm up. D, Attach the IV tubing to the continuous flush device (Sorenson Intraflow). E, Screw the continuous flush device into the transducer stopcock. F, Backflush the continuous flush device and transducer, with open stopcocks, so that all air is removed and replaced with the saline solution. G, Zero and calibrate the pressure transducer and monitor. H, Attach high-pressure tubing to the continuous flush device. Flush the air out of it. I, Attach the high-pressure tubing to the patient’s arterial line or pulmonary artery catheter. Ensure that no air bubbles are present and that there is a continuous saline to blood connection. Accurate patient pressures should now be seen on the monitor.

(From Oblouk Darovic G: Hemodynamic monitoring, Philadelphia, 1987, WB Saunders.)

image

Figure 5-29 Example of a complete setup for monitoring the arterial pressure.

(From Daily EK, Schroeder JS: Techniques in bedside hemodynamic monitoring, ed 5, St Louis, 1994, Mosby.)

The process of calibrating the monitor for systemic artery pressures was briefly discussed earlier. Regardless of whether the catheter is placed into a systemic artery, pulmonary artery, umbilical artery, or superior vena cava, the equipment must be properly calibrated to ensure accurate data. When performing two-point calibration, all pressures should read “zero” when exposed to atmospheric pressure. This is the low point. The high point pressure for arterial pressure monitoring is commonly 100 mm Hg.

c. Troubleshoot any problems with the equipment

Table 5-1 lists problems, causes, prevention, and treatment for inaccurate pressure measurements with arterial or pulmonary artery catheters. Table 5-2 specifically lists problems with arterial lines. Figure 5-30 shows common problem areas with arterial lines. Pulmonary artery catheters can have problems in the same areas.

image

Figure 5-30 Common problem areas with arterial and pulmonary artery catheters.

(From Oblouk Darovic G: Hemodynamic monitoring—invasive and noninvasive clinical application, ed 3, Philadelphia, 2002, Saunders.)

d. Perform arterial line insertion (Code: IB9s) [Difficulty: ELE: R, Ap; WRE: An]

The most widely used arterial line insertion site is the radial artery in the patient’s nondominant hand. The procedure for inserting a radial arterial line is very similar to the procedure for drawing a blood sample from the radial artery. See Chapter 3 to review the procedure. As presented previously, the necessary fluid infusion system and pressure monitoring system must be assembled and working properly. A significant difference is the use of a needle with covering catheter, rather than a needle and syringe, to puncture the artery (Figure 5-31). As the needle and catheter enter the artery, bright red blood will pulse out. In rapid succession, withdraw the needle and advance the catheter into the artery (Figure 5-32). Place a gloved fingertip over the proximal hub of the catheter to stop the bleeding. Next, screw the prepared three-way stopcock and high-pressure tubing to the hub of the catheter. Flush some heparinized saline solution through the arterial catheter to prevent any blood from clotting. Make sure the automatic drip system and arterial pressure monitoring system are working properly. The patient’s blood pressure should be displayed on the monitor (see Figures 5-28 and 5-29).

5. Manipulate a pulmonary artery catheter by order or protocol (WRE code: IIA20b) [Difficulty: WRE: R, Ap, An]

c. Troubleshoot any problems with the equipment

See Table 5-3 for a listing of the problems seen with pulmonary artery catheters and their causes, prevention, and treatment. Figure 5-30 shows common problem areas with the related pressure tubing and equipment.

MODULE E

1. Analyze the available information to determine the patient’s pathophysiologic state (Code: IIIH1) [Difficulty: ELE: R, Ap; WRE: An]

Each clinical problem presented in this section requires its own individualized treatment. Be prepared to make recommendations to change the inspired oxygen percentage, give fluids if the patient is dehydrated, give diuretic medications to increase urine output if the patient is fluid overloaded, give vasodilator medications if the patient is hypertensive, or give vasoconstrictor medications if the patient is hypotensive.

See Figure 5-33 for examples of diagnostic pathways. These can be used to help the clinician evaluate all the patient data. The diagnosis or pathophysiologic state can be determined by following the data on a given pathway. Normal values and common conditions associated with abnormal values were discussed earlier in this chapter.

4. Record and evaluate the patient’s response to the treatment(s) or procedure(s), including the following

b. Record and interpret the patient’s breath sounds (Code: IIIAb3) [Difficulty: ELE: R, Ap; WRE: An]

There is a risk of puncturing the right lung during the insertion of a central venous catheter by way of the subclavian vein (see Figure 5-13). Assess the patient for equal, bilateral breath sounds after this procedure has been attempted. Know the signs and symptoms of a pneumothorax. Review them in Chapter 1, if needed.

c. Recheck any math work and make note of incorrect data (Code: IIIA1b2) [Difficulty: ELE: R, Ap; WRE: An]

Exam Hint 5-10 (ELE, WRE)

Closely review the following concepts because they have been tested on previous examinations.

BIBLIOGRAPHY

AARC Clinical Practice Guideline. Capnography/capnometry during mechanical ventilation, 2003 revision & update. Respir Care. 2003;48(5):534.

Aggarwal NK, Kiran U, Kapoor PM, Chowdhury UK. Intraoperative central venous oxygen saturation as a surrogate for mixed venous oxygen saturation in patients undergoing cardiac surgery. J Anesth Clin Pharmacology. 2007;23(1):29-33.

Anton WR, Raghu G. Measuring end-tidal carbon dioxide tension at maximal exhalation to improve its utility during T -piece weaning trials. Respir Care. 1990;35(11):1082-1083.

Bakow ED. A limitation of capnography. Respir Care. 1982;27(2):167-168.

Branson RD, Campbell RS. Cardiovascular monitoring. In Branson RD, Hess DR, Chatburn RL, editors: Respiratory care equipment, ed 2, Philadelphia: Lippincott Williams & Wilkins, 1999.

Cairo JM. Assessment of cardiovascular function. In Cairo JM, Pilbeam SP, editors: Mosby’s respiratory care equipment, ed 8, St Louis: Mosby, 2010.

Cairo JM. Hemodynamic monitoring. In Pilbeam SP, Cairo JM, editors: Mechanical ventilation—physiologic and clinical applications, ed 4, St Louis: Mosby, 2006.

Carlon GC, Ray C, Miodownik S, et al. Capography in mechanically ventilated patients. Crit Care Med. 1988;16(5):550-556.

Clark DB, Marshall SG. Mixed venous oxygen saturation measurement. II. General and specific clinical applications. Respir Ther. 81-86, Nov/Dec 1986.

Daily EK, Schroeder JS. Techniques in bedside hemodynamic monitoring, ed 5. St Louis: Mosby, 1994.

Deshpande VM, Pilbeam SP, Dixon RJ. A comprehensive review in respiratory care. Norwalk, Conn: Appleton & Lange, 1988.

Divertie MB, McMichan JC. Continuous monitoring of mixed venous oxygen saturation. Chest. 1984;85(3):423-428.

Fahey PJ, Harris K, Vanderwarf C. Clinical experience with continuous monitoring of mixed venous oxygen saturation in respiratory failure. Chest. 1984;86(5):748-752.

Fink JB, Hunt GE, editors. Clinical practice in respiratory care. Philadelphia: Lippincott-Raven, 1999.

Gravenstein JS, Jaffe MB, Paulus DA, editors. Capnography—clinical aspects. Cambridge UK: Cambridge University Press, 2004.

Harris K. Noninvasive monitoring of gas exchange. Respir Care. 1987;32(7):544-557.

Hess D. Capnometry and capnography: technical aspects, physiologic aspects, and clinical applications. Respir Care. 1990;35(6):557-576.

Hess DR, Branson RD. Noninvasive respiratory monitoring equipment. In Branson RD, Hess DR, Chatburn RL, editors: Respiratory care equipment, ed 2, Philadelphia: Lippincott Williams & Wilkins, 1999.

Hess D. Respiratory care monitoring. In Burton GC, Hodgkin JE, Ward JJ, editors: Respiratory care: a guide to clinical practice, ed 4, Philadelphia: Lippincott-Raven, 1997.

Hunt GE. Diagnostic procedures at the bedside. In: Fink JB, Hunt GE, editors. Clinical practice in respiratory care. Philadelphia: Lippincott-Raven, 1999.

Jaquith SM. The oximetric opticath: what is it and how can it facilitate nursing management of the critically ill patient? Crit Care Nurse. 55-58, May/June 1984.

Kandel G, Aberman A. Mixed venous oxygen saturation: its role in the assessment of the critically ill patient. Arch Intern Med. 1983;143:1400-1402.

Kinasewitz GT. Use of end-tidal capnography during mechanical ventilation. Respir Care. 1982;25(2):169-171.

Krider SJ. Cardiac output assessment. In Wilkins RL, Sheldon RL, Krider SJ, editors: Clinical assessment in respiratory care, ed 4, St Louis: Mosby, 2000.

Krider SJ. Invasively monitored hemodynamic pressures. In Wilkins RL, Sheldon RL, Krider SJ, editors: Clinical assessment in respiratory care, ed 4, St Louis: Mosby, 2000.

Malinowski T. Respiratory monitoring in the intensive care unit. In Wilkins RL, Krider SJ, Sheldon RL, editors: Clinical assessment in respiratory care, ed 3, St Louis: Mosby, 1990.

Malley WJ. Clinical blood gases, ed 2. St Louis: Saunders, 2005.

Marini JJ. Obtaining meaningful data from the Swan-Ganz catheter. Respir Care. 1985;30(7):572-585.

Mathews PJ, Gregg BL. Monitoring and management of the patient in the ICU. In Scanlan CL, Wilkins RL, Stoller JK, Kacmarek RM, editors: Egan’s fundamentals of respiratory care, ed 9, St Louis: Mosby, 2009.

McMichan JC. Continuous monitoring of mixed venous oxygen saturation in clinical practice. Mount Sinai J Med. 1984;51(5):569-572.

Nuzzo PF, Anton WR. Practical applications of capnography. Respir Ther. 12-17, Nov/Dec 1986.

Oblouk Darovic G. Hemodynamic monitoring. Philadelphia: WB Saunders, 1987.

Oblouk Darovic G. Hemodynamic monitoring—invasive and noninvasive clinical application, ed 3. Philadelphia: WB Saunders, 2002.

Osgood CF, Watson MH, Slaughter MS, et al. Hemodynamic monitoring in respiratory care. Respir Care. 1984;29(1):25-34.

Paulus DA. Invasive monitoring of respiratory gas exchange: continuous measurement of mixed venous oxygen saturation. Respir Care. 1987;32(7):535-543.

Pilbeam SP. Mechanical ventilation: physiological and clinical applications, ed 3. St Louis: Mosby, 1998.

Ragosta M. Textbook of clinical hemodynamics. Philadelphia: Saunders, Inc, 2008.

Reinhart K, Kuhn H, Hartog C, Bredle DL. Continuous central venous and pulmonary artery oxygen saturation in the critically ill. Intensive Care Med. 2004;30:1572-1578.

Restrepo RD. Assessment of cardiac output. In Wilkins RL, Dexter JR, Heuer AJ, editors: Clinical assessment in respiratory care, ed 6, St Louis: Mosby, 2010.

Ruppel G. Manual of pulmonary function testing, ed 9. St Louis: Mosby, 2009.

Shapiro BA, Harrison RA, Cane RD, et al. Clinical application of blood gases, ed 5. St Louis: Mosby, 1994.

Shapiro BA, Kacmarek RM, Cane RD, et al, editors. Clinical application of respiratory care, ed 4, St Louis: Mosby, 1991.

Whitaker K. Comprehensive perinatal & pediatric respiratory care, ed 2. Albany, NY: Delmar, 1997.

White GC. Equipment theory for respiratory care, ed 4. Albany, NY: Delmar, 2005.

Wiedemann HP. Invasive monitoring techniques in the ventilated patient. In: Kacmarek RM, Stoller JK, editors. Current respiratory care. Toronto: BC Decker, 1988.

Wilkins RL., Bedside assessment of the patient. Wilkins, Stoller JK, Kacmarek RM. Egan’s fundamentals of respiratory care, ed 9, St Louis: Mosby, 2009.

Wilkins RL., Analysis and monitoring in gas exchange. Wilkins, Stoller JK, Kacmarek RM. Egan’s fundamentals of respiratory care, ed 9, St Louis: Mosby, 2009.

Wilkins RL, Dexter JR. Assessment of hemodynamic pressures. In Wilkins RL, Dexter JR, Heuer AJ, editors: Clinical assessment in respiratory care, ed 6, St Louis: Mosby, 2010.

Wilkins RL, Heuer AJ. Vital signs. In Wilkins RL, Dexter JR, Heuer AJ, editors: Clinical assessment in respiratory care, ed 6, St Louis: Mosby, 2010.

SELF-STUDY QUESTIONS FOR THE ENTRY LEVEL EXAM See page 586 for answers

SELF-STUDY QUESTIONS FOR WRITTEN REGISTRY EXAM See page 610 for answers

4. An adult patient is receiving mechanical ventilation when the following data are gathered:

  9:00 AM 11:00 AM
Pao2 75 torr 53 torr
pulmonary vascular resistance 120 dynes/sec/cm− 5 340 dynes/sec/cm− 5
pulmonary capillary wedge pressure 8 mm Hg 10 mm Hg
pulmonary artery pressure 25/10 mm Hg 42/21 mm Hg

How should the results be interpreted?