73: Pulmonary Artery Catheter Insertion (Assist) and Pressure Monitoring

Published on 06/03/2015 by admin

Filed under Critical Care Medicine

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 2 (1 votes)

This article have been viewed 11817 times

PROCEDURE 73

Pulmonary Artery Catheter Insertion (Assist) and Pressure Monitoring

PREREQUISITE NURSING KNOWLEDGE

• Knowledge of the normal cardiovascular anatomy and physiology is needed.

• Knowledge of the normal pulmonary anatomy and physiology is necessary.

• Knowledge of principles of aseptic technique is needed.

• Basic dysrhythmia recognition and treatment of life-threatening dysrhythmias should be understood.

• Advanced cardiac life support knowledge and skills are necessary

• Knowledge of the anatomy of the pulmonary artery (PA) catheter (Fig. 73-1) and the location of the PA catheter in the heart and pulmonary artery (Fig. 73-2) is important.

• Knowledge of the setup of the hemodynamic monitoring system (see Procedure 76) is essential.

• Understanding of normal hemodynamic values (see Table 67-1) is needed.

• The pulmonary artery catheter contains a proximal injectate lumen port, a PA distal lumen port, a thermistor connector, and a balloon-inflation valve. Some catheters also have two infusion ports, right atrial (RA) and right ventricular (RV) lumens, that can be used for infusion of medications and intravenous fluids.

• The PA distal lumen is used to monitor systolic, diastolic, and mean pressures in the pulmonary artery. This lumen also allows for sampling of mixed venous blood. The proximal injectate lumen is used to monitor the right atrial pressure and inject the solution used to obtain cardiac output (CO). The balloon-inflation valve is used to obtain the pulmonary artery occlusion pressure (PAOP).

• PAOP may be referred to as pulmonary artery wedge pressure.

• The PA diastolic pressure and the PAOP are indirect measures of left ventricular end-diastolic pressure (LVEDP). Usually, the PAOP is approximately 1 to 4 mm Hg less than the pulmonary artery diastolic pressure (PADP). Because these two pressures are similar, the PADP is commonly followed, which minimizes the frequency of balloon inflation, thus decreasing the potential of balloon rupture and pulmonary artery trauma.

• Differences between the PADP and the PAOP may exist for patients with pulmonary hypertension, chronic obstructive lung disease, acute respiratory distress syndrome (ARDS), pulmonary embolus, and tachycardia.

• Indications for PA catheter therapy (see Procedure 72 for additional indications) are as follows:

• Hemodynamic monitoring with a PA catheter has no absolute contraindications, but an assessment of risk versus benefit to the patient should be considered. Relative contraindications to pulmonary artery catheter insertion include presence of fever, presence of a mechanical tricuspid valve, and a coagulopathic state. A patient with left bundle-branch block may have a right bundle-branch block develop during PA catheter insertion, resulting in complete heart block. In these patients, a temporary pacemaker should be readily available.

• Pulmonary artery pressures may be elevated as a result of pulmonary artery hypertension, pulmonary disease, mitral valve disease, left ventricular failure, atrial or ventricular left-to-right shunt, pulmonary emboli, or hypervolemia.

• Pulmonary artery pressures may be decreased due to hypovolemia or vasodilation.

• Waveforms that occur during insertion include RA, RV, PA, and pulmonary artery occlusion (PAO; Fig. 73-3).

image The a wave reflects atrial contraction, the c wave reflects closure of the atrioventricular valve, and the v wave reflects passive filling of the atria during ventricular systole (Figs. 73-4 and 73-5).

image The a wave reflects right ventricular filling at end diastole. The mean of the a wave is determined by averaging the top and bottom values of the a wave.

image Elevated a and v waves may be evident in right atrial pressure (RAP/central venous pressure [CVP]) and in PAOP waveforms. These elevations may occur in patients with cardiac tamponade, constrictive pericardial disease, and hypervolemia.

image Elevated a waves in the RAP/CVP waveform may occur in patients with pulmonic or tricuspid stenosis, right ventricular ischemia or infarction, RV failure, pulmonary artery hypertension, and atrioventricular (AV) dissociation.

image Elevated a waves in the PAOP waveform may occur in patients with mitral stenosis, acute left ventricular ischemia or infarction, left ventricular failure, and AV dissociation.

image Elevated v waves in the RAP/CVP waveform may occur in patients with tricuspid insufficiency.

image Elevated v waves in the PAOP waveform may occur in patients with mitral insufficiency or a ruptured papillary muscle.

• Insertion and placement verification should occur as follows:

image The PA catheter may be inserted through the subclavian, internal jugular, femoral, external jugular, or antecubital veins.

image The standard 7.5 Fr PA catheter is 110 cm long and has black markings at 10-cm increments and wide black markings at 50-cm increments for facilitation of insertion and positioning (see Fig. 73-1). The catheter should reach the PA after being advanced 40 to 55 cm from the internal jugular vein, 35 to 50 cm from the subclavian vein, 60 cm from the femoral vein, 70 cm from the right antecubital fossa, and 80 cm from the left antecubital fossa.

image Verification of PA catheter placement is validated with waveform analysis. Correct catheter placement shows a PAO tracing when the balloon is inflated and a PA tracing when the balloon is deflated.

image Catheter placement is also verified with chest radiography.

image The PA catheter balloon contains latex, which may cause allergic reactions. Latex-free catheters are available.

EQUIPMENT

• PA catheter (non–heparin-coated PA catheters and latex-free PA catheters are available)

• Percutaneous sheath introducer kit and sterile catheter sleeve

• Pressure modules and cables for interface with the monitor

• Cardiac output cable with a thermistor/injectate sensor

• Pressure transducer system, including flush solution recommended according to institution standard, a pressure bag or device, pressure tubing with transducers, and flush device

• Dual-channel recorder

• Sterile normal saline intravenous (IV) solution for flushing of the introducer and catheter infusion ports

• Antiseptic solution (e.g., 2% chlorhexidine-based preparation)

• Head covers, fluid-shield masks, sterile gowns, sterile gloves, nonsterile gloves, and sterile drapes

• 1% lidocaine without epinephrine

• Sterile basin or cup

• Sterile water or normal saline solution

• Sterile dressing supplies

• Chlorhexidine-impregnated sponge

• Stopcocks (may be included in some pressure tubing systems)

• Nonvented caps or needleless caps

• Leveling device (low-intensity laser or carpenter level)

Additional equipment as needed includes the following:

PATIENT ASSESSMENT AND PREPARATION

Patient Assessment

• Determine baseline hemodynamic, cardiovascular, peripheral vascular, and neurovascular status. imageRationale: Assessment provides data that can be used for comparison with postinsertion assessment data and hemodynamic values.

• Determine the patient’s baseline pulmonary status. If the patient is mechanically ventilated, note the type of support, ventilator mode, and presence or absence of positive end-expiratory pressure (PEEP) or continuous positive airway pressure (CPAP). imageRationale: The presence of mechanical ventilation alters hemodynamic waveforms and pressures.

• Assess the patient’s medical history specifically related to problems with venous access sites, cardiac anatomy, and pulmonary anatomy. imageRationale: Identification of obstructions or disease should be made before the insertion attempt.

• Assess the patient’s current laboratory profile, including electrolyte, coagulation, and arterial blood gas results. imageRationale: Laboratory abnormalities are identified. Baseline coagulation studies are helpful in determination of the risk for bleeding. Electrolyte and arterial blood gas imbalances may increase cardiac irritability.

Patient Preparation

• Verify correct patient with two identifiers. imageRationale: Prior to performing a procedure, the nurse should ensure the correct identification of the patient for the intended intervention.

• Ensure that the patient and family understand preprocedural teaching. Answer questions as they arise, and reinforce information as needed. imageRationale: Understanding of previously taught information is evaluated and reinforced.

• Ensure that informed consent has been obtained. imageRationale: Informed consent protects the rights of the patient and makes a competent decision possible for the patient.

• Perform a pre-procedure verification and time out, if non-emergent. imageRationale: Ensures patient safety.

• Validate the patency of the peripheral IV line. imageRationale: Access may be needed for administration of emergency medications or fluids.

• Assist the patient to the supine position. imageRationale: This position prepares the patient for skin preparation, catheter insertion, and setup of the sterile field.

• Sedate the patient or provide prescribed analgesics as needed. imageRationale: Movement of the patient may inhibit insertion of the PA catheter.

image

Figure 73-14 Intermittent mandatory ventilation (IMV) mode of ventilation and the effect on the pulmonary artery (PA) waveform. (From Ahrens TS, Taylor LK: Hemodynamic waveform analysis, Philadelphia, 1992, Saunders.)

References

image 1. Blanch, L, et al. Short term effects of prone position in critically ill patients with acute respiratory distress syndrome. Intensive Care Med. 1997; 23:1033–1039.

image 2. Brussel, T, et al. Mechanical ventilation in the prone position for acute respiratory failure after cardiac surgery. J Cardiothorac Vasc Anesth. 1993; 7:541–546.

image 3. Cason, CL, et al. Effects of backrest elevation and position on pulmonary artery pressures. Cardiovasc Nurs. 1990; 26:1–5.

image 4. Chen, Y, et al. Comparison between replacement at 4 days and 7 days on the infection rate for pulmonary artery catheters in an intensive care unit. Crit Care Med. 2003; 31:1353–1358.

image 5. Chong, BH. Heparin-induced thrombocytopenia. Br J Haematol. 1995; 89:431–439.

image 6. Chulay, M, Miller, T. The effect of backrest elevation on pulmonary artery and pulmonary capillary wedge pressures in patients after cardiac surgery. Heart Lung. 1984; 13:138–140.

image 7. Clochesy, J, Hinshaw, AD, Otto, CW. Effects of change of position on pulmonary artery and pulmonary capillary wedge pressure in mechanically ventilated patients. NITA. 1984; 7:223–225.

image 8. Cohen, Y, et al, The “hands-off” catheter in the prevention of systemic infections associated with pulmonary artery catheter. a prospective study. Am J Respir Crit Care Med 1998; 157:284–287.

image 9. Daily, EK, Schroeder, JS. Techniques in bedside hemodynamic monitoring,, ed 5. St Louis: Mosby; 1994.

image 10. Dobbin, K, et al, Pulmonary artery pressure measurement in patients with elevated pressures. effect of backrest elevation and method of measurement. Am J Crit Care 1992; 1:61–69.

image 11. Eyer, S, et al, Catheter-related sepsis. prospective, randomized study of three methods of long-term catheter maintenance. Crit Care Med 1990; 18:1073–1079.

image 12. Fridrich, P, et al. The effects of long-term prone positioning in patients with trauma-induced adult respiratory distress syndrome. Anesth Analg. 1996; 83:1206–1211.

image 13. Groom, L, Frisch, SR, Elliot, M. Reproducibility and accuracy of pulmonary artery pressure measurement in supine and lateral positions. Heart Lung. 1990; 19:147–151.

image 14. Hannan, AT, Brown, M, Bigman, O. Pulmonary artery catheter induced hemorrhage. Chest. 1984; 85:128–131.

15. Infusion Nurses Society. Infusion nursing standards of practice. J Infusion Nurs. 2006; 29:S1–S92.

image 16. Jolliet, P, Bulpa, P, Chevrolet, JC. Effects of prone position on gas exchange and hemodynamics in severe acute respiratory distress syndrome. Crit Care Med. 1998; 26:1977–1985.

image 17. Keating, D, et al. Effect of sidelying positions on pulmonary artery pressures. Heart Lung. 1986; 15:605–610.

image 18. Kennedy, GT, Bryant, A, Crawford, MH. The effects of lateral body positioning on measurements of pulmonary artery and pulmonary wedge pressures. Heart Lung. 1984; 13:155–158.

image 19. Lambert, CW, Cason, CL, Backrest elevation and pulmonary artery pressures. research analysis. Dimens Crit Care Nurs 1990; 9:327–335.

image 20. Langer, M, et al. The prone position in ARDS patients. Chest. 1988; 94:103–107.

image 21. Laulive, JL. Pulmonary artery pressures and position changes in the critically ill adult. Dimens Crit Care Nurs. 1982; 1:28–34.

image 22. Luskin, RL, et al, Extended use of disposable pressure transducers. a bacteriologic evaluation. JAMA 1986; 255:916–920.

image 23. O’Grady, NP, et al. Guidelines for the prevention of intravascular catheter-related infections. Am J Infect Control. 2002; 30:476–489.

image 24. O’Malley, MK, et al. Value of routine pressure monitoring system changes after 72 hours of use. Crit Care Med. 1994; 22:1424–1430.

image 25. Osika, C, Measurement of pulmonary artery pressures. supine verses side-lying head-elevated positions. Heart Lung 1989; 18:298–299.

image 26. Pappert, D, et al. Influence of positioning on ventilation-perfusion relationships in severe adult respiratory distress syndrome. Chest. 1994; 106:1511–1516.

image 27. Pelosi, P, et al. Effects of the prone position on respiratory mechanics and gas exchange during acute lung injury. Am J Respir Crit Care Med. 1998; 157:387–393.

image 28. Randolph, AG, et al. Benefit of heparin in central venous and pulmonary artery catheters. Chest. 1998; 113:165–171.

image 29. Voggenreiter, G, et al. Intermittent prone positioning in the treatment of severe and moderate posttraumatic lung injury. Crit Care Med. 1999; 27:2375–2382.

image 30. Vollman, KM, Bander, JJ. Improved oxygenation utilizing a prone positioner in patients with acute respiratory distress syndrome. Intensive Care Med. 1996; 22:1105–1111.

image 31. Wild, L. Effect of lateral recumbent positions on measurement of pulmonary artery and pulmonary artery wedge pressures in critically ill adults. Heart Lung. 1984; 13:305.

image 32. Wilson, AE, et al. Effect of backrest position on hemodynamic and right ventricular measurements in critically ill adults. Am J Crit Care. 1996; 5:264–270.

image 33. Woods, SL, Mansfield, LW. Effect of body position upon pulmonary artery and pulmonary capillary wedge pressures in noncritically ill patients. Heart Lung. 1976; 5:83–90.

Additional Readings

image Abreu, AR, Campos, MA, Krieger, BP, Pulmonary artery rupture induced by a pulmonary artery catheter. a case report and review of the literature. J Intensive Care Med 2004; 19:291–296.

image Ahrens, TS, Taylor, LA. Hemodynamic waveform analysis. Philadelphia: Saunders; 1992.

image American Association of Critical-Care Nurses, AACN practice alert. pulmonary artery pressure measurement. www.aacn.org, 2004 [available at].

image Anonymous, Pulmonary artery catheter consensus conference. consensus statement. Crit Care Med 1997; 25:910–925.

image Bridges, EJ, Pulmonary artery pressure monitoring. when, how, and what else to use. AACN Adv Crit Care 2006; 17:286–305.

image Daily, EK. Hemodynamic waveform analysis. J Cardiovasc Nurs. 2001; 15:6–22,87-88.

image Darovic, GO, Hemodynamic monitoring. invasive and noninvasive clinical application,. ed 3. Saunders, Philadelphia, 2002.

image Houghton, D, et al. Routine daily chest radiography in patients with pulmonary artery catheters. Am J Crit Care. 2002; 11:261–265.

image Keckeisen, M, Protocols for practice. hemodynamic monitoring seriespulmonary artery pressure monitoring,. American Association of Critical-Care Nurses, Aliso Viejo, CA, 1997.

image Liu, C, Webb C, From the Food and Drug Administration pulmonary artery rupture. serious complication associated with pulmonary artery catheters. Int J Trauma Nurs 2000; 6:19–26.

image Ott, K, Johnson, K, Ahrens, T. New technologies in the assessment of hemodynamic parameters. J Cardiovasc Nurs. 2001; 15:41–55.

image Quaal, SJ. Improving the accuracy of pulmonary artery catheter measurement. J Cardiovasc Nurs. 2001; 15:71–82.

image Quaal, SJ. Is it necessary to perform a square wave test routinely to test for accuracy in hemodynamic monitoring? Or is it recommended only if there is a problem. Crit Care Nurse. 1995; 15:92–93.

Rauen, CA, Flynn, MB, Bridges, E, Evidence-based practice habits. transforming research into bedside practice. Crit Care Nurse 2009; 29:46–59.

Rizvi, K, et al. Effect of airway pressure display on interobserver agreement in the assessment of vascular pressures in patients with acute lung injury and acute respiratory distress syndrome. Crit Care Med. 2005; 33:98–103.