73: Pulmonary Artery Catheter Insertion (Assist) and Pressure Monitoring

Published on 06/03/2015 by admin

Filed under Critical Care Medicine

Last modified 06/03/2015

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 2 (1 votes)

This article have been viewed 11599 times

PROCEDURE 73

Pulmonary Artery Catheter Insertion (Assist) and Pressure Monitoring

PREREQUISITE NURSING KNOWLEDGE

• Knowledge of the normal cardiovascular anatomy and physiology is needed.

• Knowledge of the normal pulmonary anatomy and physiology is necessary.

• Knowledge of principles of aseptic technique is needed.

• Basic dysrhythmia recognition and treatment of life-threatening dysrhythmias should be understood.

• Advanced cardiac life support knowledge and skills are necessary

• Knowledge of the anatomy of the pulmonary artery (PA) catheter (Fig. 73-1) and the location of the PA catheter in the heart and pulmonary artery (Fig. 73-2) is important.

• Knowledge of the setup of the hemodynamic monitoring system (see Procedure 76) is essential.

• Understanding of normal hemodynamic values (see Table 67-1) is needed.

• The pulmonary artery catheter contains a proximal injectate lumen port, a PA distal lumen port, a thermistor connector, and a balloon-inflation valve. Some catheters also have two infusion ports, right atrial (RA) and right ventricular (RV) lumens, that can be used for infusion of medications and intravenous fluids.

• The PA distal lumen is used to monitor systolic, diastolic, and mean pressures in the pulmonary artery. This lumen also allows for sampling of mixed venous blood. The proximal injectate lumen is used to monitor the right atrial pressure and inject the solution used to obtain cardiac output (CO). The balloon-inflation valve is used to obtain the pulmonary artery occlusion pressure (PAOP).

• PAOP may be referred to as pulmonary artery wedge pressure.

• The PA diastolic pressure and the PAOP are indirect measures of left ventricular end-diastolic pressure (LVEDP). Usually, the PAOP is approximately 1 to 4 mm Hg less than the pulmonary artery diastolic pressure (PADP). Because these two pressures are similar, the PADP is commonly followed, which minimizes the frequency of balloon inflation, thus decreasing the potential of balloon rupture and pulmonary artery trauma.

• Differences between the PADP and the PAOP may exist for patients with pulmonary hypertension, chronic obstructive lung disease, acute respiratory distress syndrome (ARDS), pulmonary embolus, and tachycardia.

• Indications for PA catheter therapy (see Procedure 72 for additional indications) are as follows:

• Hemodynamic monitoring with a PA catheter has no absolute contraindications, but an assessment of risk versus benefit to the patient should be considered. Relative contraindications to pulmonary artery catheter insertion include presence of fever, presence of a mechanical tricuspid valve, and a coagulopathic state. A patient with left bundle-branch block may have a right bundle-branch block develop during PA catheter insertion, resulting in complete heart block. In these patients, a temporary pacemaker should be readily available.

• Pulmonary artery pressures may be elevated as a result of pulmonary artery hypertension, pulmonary disease, mitral valve disease, left ventricular failure, atrial or ventricular left-to-right shunt, pulmonary emboli, or hypervolemia.

• Pulmonary artery pressures may be decreased due to hypovolemia or vasodilation.

• Waveforms that occur during insertion include RA, RV, PA, and pulmonary artery occlusion (PAO; Fig. 73-3).

image The a wave reflects atrial contraction, the c wave reflects closure of the atrioventricular valve, and the v wave reflects passive filling of the atria during ventricular systole (Figs. 73-4 and 73-5).

image The a wave reflects right ventricular filling at end diastole. The mean of the a wave is determined by averaging the top and bottom values of the a wave.

image Elevated a and v waves may be evident in right atrial pressure (RAP/central venous pressure [CVP]) and in PAOP waveforms. These elevations may occur in patients with cardiac tamponade, constrictive pericardial disease, and hypervolemia.

image Elevated a waves in the RAP/CVP waveform may occur in patients with pulmonic or tricuspid stenosis, right ventricular ischemia or infarction, RV failure, pulmonary artery hypertension, and atrioventricular (AV) dissociation.

image Elevated a waves in the PAOP waveform may occur in patients with mitral stenosis, acute left ventricular ischemia or infarction, left ventricular failure, and AV dissociation.

image Elevated v waves in the RAP/CVP waveform may occur in patients with tricuspid insufficiency.

image Elevated v waves in the PAOP waveform may occur in patients with mitral insufficiency or a ruptured papillary muscle.

• Insertion and placement verification should occur as follows:

image The PA catheter may be inserted through the subclavian, internal jugular, femoral, external jugular, or antecubital veins.

image The standard 7.5 Fr PA catheter is 110 cm long and has black markings at 10-cm increments and wide black markings at 50-cm increments for facilitation of insertion and positioning (see Fig. 73-1). The catheter should reach the PA after being advanced 40 to 55 cm from the internal jugular vein, 35 to 50 cm from the subclavian vein, 60 cm from the femoral vein, 70 cm from the right antecubital fossa, and 80 cm from the left antecubital fossa.

image Verification of PA catheter placement is validated with waveform analysis. Correct catheter placement shows a PAO tracing when the balloon is inflated and a PA tracing when the balloon is deflated.

image Catheter placement is also verified with chest radiography.

image The PA catheter balloon contains latex, which may cause allergic reactions. Latex-free catheters are available.