30: Auto–Positive End-Expiratory Pressure (Auto-PEEP) Calculation

Published on 06/03/2015 by admin

Filed under Critical Care Medicine

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 6188 times

PROCEDURE 30

Auto–Positive End-Expiratory Pressure (Auto-PEEP) Calculation

PREREQUISITE NURSING KNOWLEDGE

• Auto–positive end-expiratory pressure (Auto-PEEP) is often called occult because it is not set on the ventilator; instead, it is a result of inadequate exhalation time (Fig. 30-1).211

• Auto-PEEP is associated with high minute ventilation requirements, small-diameter endotracheal tubes, bronchospasm, long inspiratory times, high respiratory rates, and mechanical factors, such as water accumulation in the ventilator tubing.211

• Auto-PEEP may result in an increased work of breathing. The set sensitivity of the ventilator is referenced to the amount of set-PEEP selected by the clinician. Because auto-PEEP is not sensed by the ventilator, the patient has to generate a pressure equal to the set sensitivity plus auto-PEEP to “trigger” inspiratory flow or a breath from the ventilator.2,9

• Auto-PEEP may elevate static pressure (i.e., plateau pressure). High plateau pressures can result in barotrauma and hemodynamic compromise.9

• Auto-PEEP may be a desirable outcome of select ventilator settings (e.g., pressure-controlled inverse ratio ventilation). In these cases, the goal of auto-PEEP is to restore functional residual capacity and reduce shunt.

• Interventions to offset auto-PEEP include the use of large-diameter endotracheal tubes, bronchodilators, short inspiratory times, long expiratory times, lower respiratory rates, frequent emptying of ventilator circuit water accumulation (heated circuits may eliminate this complication), and the use of sedatives and narcotics (if the patient’s breathing pattern is such that it increases the minute ventilation). Occasionally the addition of set-PEEP is used to offset auto-PEEP. The addition of set-PEEP serves as a splint by keeping the airway open throughout exhalation, decreasing auto-PEEP.27,11 An example of this technique is the patient with chronic obstructive pulmonary disease in whom early airway closure during exhalation results in gas trapping.

PATIENT ASSESSMENT AND PREPARATION

Patient Assessment

• Assess for the presence of auto-PEEP; have a high index of suspicion if any of the following is noted:

image Minute ventilation requirements are greater than 10 L/min.

image The patient has chronic obstructive pulmonary disease.

image The patient has bronchospasm.

image The respiratory rate is rapid (i.e., ≥20/min).

image The inspiratory time is long (i.e., >1 second).

image Dyssynchrony exists between patient and ventilator, especially when the ventilator does not cycle with patient inspiration. imageRationale: Auto-PEEP increases the work of breathing by increasing the threshold load to trigger inspiration. The increased work of breathing may cause fatigue.

image Auto-PEEP is present (when any of the previously listed criteria is present) and the patient is hypotensive or shows signs of barotrauma.

image Patients in status asthmaticus are at high risk for auto-PEEP. Consider the presence of auto-PEEP as the result of vigorous bagging, high ventilator rates, or large tidal volumes, especially if hypotension is present. imageRationale: Auto-PEEP, similar to intentional PEEP, puts the patient at risk for barotrauma from increased intra-alveolar pressures. In patients with asthma, lung compliance is good, but airway resistance is high, which encourages dynamic hyperinflation (alveolar overdistention) and potential barotrauma. Hemodynamic compromise occurs when the increased alveolar pressure results in compression of the corresponding capillaries; decreased venous return and hypotension result.

• Auto-PEEP may not be detected in some patients with severe asthma despite its presence. This situation may occur if the end-expiratory hold maneuver is too short to allow complete equilibration of pressures and with airway obstruction. In addition, if airways are noncommunicating, as in the case of obstruction from mucus or severe bronchospasm, auto-PEEP is not able to be measured. Assume auto-PEEP in these cases. Monitoring of plateau pressure may be a better method of assessment of hyperinflation (see Procedure 31).57

References

image 1. Bidani, A, Tzouanakis, AE, Carenas, VJ, et al. Permissive hypercapnia in acute respiratory failure. JAMA. 1994; 272:957–962.

2. Burns, SM, Ventilating patients with acute severe asthma. what do we really know. AACN Advanced Crit Care 2006; 17:188–193.

image 3. Coussa, ML, Guerin, C, Eissa, NT, et al. Partitioning of work of breathing in mechanically ventilated COPD -patients. J Appl Physiol. 1993; 75(4):1711–1719.

image 4. Georgopoulos, D, Giannouli, E, Patakas, D. Effects of -extrinsic positive end-expiratory pressure on mechanically ventilated patients with chronic obstructive pulmonary disease and dynamic hyperinflation. Intensive Care Med. 1993; 19:197–203.

image 5. Leatherman, JW, McArthur, C, Shapiro, RS. Effect of -prolongation of expiratory time on dynamic hyperinflation in mechanically ventilated patients with severe asthma. Crit Care Med. 2004; 32:1542–1545.

image 6. Leatherman, JW, Ravenscraft, SA, Low-measured auto-positive end-expiratory pressure during mechanical ventilation of patients with severe asthma. hidden auto-positive end expiratory pressure. Crit Care Med 1996; 24:541–546.

image 7. MacIntyre, NR, Cheng, KC, McConnell, R. Applied PEEP during pressure support reduces inspiratory threshold of intrinsic PEEP. Chest. 1997; 111:188–193.

image 8. Pepe, PE, Marini, JJ. Occult positive end-expiratory pressure in mechanically ventilated patients with airflow obstruction. Am Rev Respir Dis. 1994; 126:166–173.

image 9. Rossi, A, Gottfried, SB, Zocchi, L, et al, Measurement of static compliance of the total respiratory system in patients with acute respiratory failure during mechanical ventilation. the effect of intrinsic positive end-expiratory pressure. Am Rev Respir Dis. 1985; 131(5):672–677.

image 10. Smith, TC, Marini, JJ. Impact of PEEP on lung mechanics and work of breathing in severe airflow obstruction. J Appl Physiol. 1988; 65:1488–1499.

11. Tobin, MJ, Alex, CA, Fahey, PJ. Fighting the ventilator. In: Tobin MJ, ed. Principles and practice of mechanical ventilation. New York: McGraw-Hill, 2006.