Tracheal Gas Insufflation

Published on 01/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 5 (1 votes)

This article have been viewed 2771 times

Tracheal Gas Insufflation

Tracheal Gas Insufflation

    Tracheal gas insufflation (TGI): The addition of a secondary gas flow during mechanical ventilation at the level of the carina to wash CO2 from the deadspace of the large airways, endotracheal tube (ETT), and ventilator circuit (Figure 44-1).

At end exhalation large airways, the ETT and the ventilator circuit deadspace contain a large amount of CO2.

Washing the CO2 from these areas decreases the CO2 of the gas moving to the alveoli during the next inhalation.

As a result, over time the arterial P2 decreases.

Numerous case series in patients have demonstrated that TGI at flows of approximately 6 to 15 L/min decrease Paco2 (Figure 44-2).

The greater the Paco2 at the time of initiating TGI, the greater the decrease in Paco2.

The greater the TGI flow, the greater the effect on Paco2.

The greater the volume of deadspace washed of CO2, the greater the effect on arterial Pco2.

II TGI Methodology

Figure 44-3 illustrates a typical TGI system.

A small-gauge catheter is placed into or along side the ETT with its tip setting just past the tip of the ETT in the trachea.

In most settings the tube is directed toward the carina (direct TGI).

However, the flow through the catheter may also be directed up toward the ETT (e.g., indirect or reverse TGI).

TGI can also be applied continuously or intermittently (during exhalation only).

By simply attaching the TGI catheter to a flowmeter, continuous flow TGI can be performed.

To accomplish expiratory phase-only TGI, the flow delivery must be coordinated with the ventilator (i.e., activation of TGI flow must begin and end during the expiratory phase).

There are also some data to show that tracheal gas exsufflation (TGE; negative pressure applied to the catheter during expiration) also reduces Pco2. However, TGE must be coordinated with the ventilator, only being applied during the expiratory phase.

III Concerns with TGI

Humidification: Clearly the TGI flow must be humidified to avoid the development of dried retained secretions.

Airway injury: A high flow of gas from the TGI catheter can cause the tip of the catheter to wipe in the airway causing injury to the tracheal wall.

With TGE, humidification and airway injury are not problems because the system only removes gas from the airway.

A major concern with TGI is the ability to shut down the system if there is an obstruction of the airway proximal to the tip of the catheter (i.e., between the ventilator circuit “wye” and the tip of the TGI catheter).

If the TGI flow is not stopped when this occurs, rapid overdistention of the airway and the potential for barotrauma rapidly develop.

Direct TGI causes the total applied positive end-expiratory pressure (PEEP) to increase because the TGI flow is directed toward the lower airway at end expiration.

Indirect TGI causes the total PEEP to decrease because flow is directed toward the endotracheal tube and ventilator tubing, causing a Venturi effect and reducing PEEP.

Monitoring of airway pressures and tidal volumes (Vts) is difficult during TGI.

IV Setting TGI

During continuous TGI, Vt and airway pressure are increased during volume ventilation (Figure 44-4) unless the Vt is adjusted downward by a volume equal to the TGI flow added during inspiration.

In pressure control ventilation, continuous flow TGI will increase the Vt and plateau pressure if the ventilator-delivered flow decreases to zero before the end of the inspiratory phase (see Figure 44-4).

Pressure relief and flow relief valves can be added to the ventilator circuit to avoid pressure and volume increases.

The use of expiratory phase-only TGI avoids the increase in volume and pressure during volume- and pressure-targeted ventilation.

However, to measure plateau pressure and end expiratory pressure during TGI (continuous and intermittent) the TGI flow must be coordinated with the ventilator.

Indications

VI TGI Is Experimental.