Sensing and Detection

Published on 21/06/2015 by admin

Filed under Cardiovascular

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 6162 times

3 Sensing and Detection

Sensing of cardiac depolarizations and detection of arrhythmias control the electrical therapies of pacemakers and implantable cardioverter-defibrillators (ICDs). When a wavefront of depolarization passes the tip electrode of an intracardiac lead, a deflection in the continuous electrogram signal travels instantaneously up the lead wire to the pacemaker or ICD, where sensing electronics amplifies, filters, digitizes, and processes the signal. A sensed event occurs at the instant when the sensing system determines that an atrial or ventricular depolarization has occurred. Dual-chamber pacemakers and ICDs have separate sensing systems for the atrium and ventricle.

Appropriate sensing results in one sensed event for each activation wavefront in the corresponding chamber. Failure to sense activation wavefronts results in undersensing, which can cause inappropriate pacing, failure to switch modes, or failure to detect a tachyarrhythmia. Undersensing occurs if the depolarization signal has insufficient amplitude or frequency content to be recognized as a sensed event, or if a blanking period disables the sensing amplifier at the time of the event. Oversensing occurs when nonphysiologic signals or physiologic signals that do not reflect local myocardial depolarization are sensed. Oversensing can cause inappropriate pacing inhibition, pacemaker tracking, or ICD therapy.

Detection algorithms process sensed events to classify the atrial or ventricular rhythm. This classification is used to control beat-by-beat paced events, to change the pacing mode, to store data regarding tachyarrhythmias, and to terminate sustained tachyarrhythmias with antitachycardia pacing or shocks.

image Electrograms

Surface Electrocardiogram vs. Intracardiac Electrogram

An electrogram (EGM) is a graphic display of the potential difference between two points in space over time. During the upstroke of a myocardial action potential, the inside of the cell abruptly changes from its resting negative potential (with respect to the outside of the cell) to a neutral or slightly positive potential. After about 250 to 400 milliseconds (msec), the cell membrane is then repolarized, with the inside of the cell returning to its resting, negatively charged state. Figure 3-1 illustrates how an EGM is recorded between two electrodes in contact with the myocardium.

The electrocardiogram (ECG) is recorded from two electrodes on the surface of the body at some distance from the heart. The typical amplitude of its QRS complex is about 1 millivolt (mV). The locations of the two electrodes determine the vectorial “viewpoint” from which the electrical activity of the entire heart is observed from the body surface. In contrast, the ventricular endocardial unipolar EGM typically is 5 to 20 mV in amplitude when recorded from a small electrode on the tip of a lead placed in direct contact with the apex of the right ventricle (Fig. 3-2). The second electrode needed to record this unipolar EGM is the pacemaker or ICD metal “can,” which is located some distance from the heart. The location of this distant second electrode, sometimes called the “indifferent electrode,” has a much smaller effect on the signal’s properties, although it may record noncardiac electric potentials (e.g., from pectoral muscle). The ECG records electrical activity from the entire heart, whereas the EGM records only local wavefronts of depolarization and repolarization. The EGM depends on the viability of approximately 1 or 2 cm3 of myocardium immediately under the tip electrode,1,2 as depicted in Figure 3-2.

Electrode Systems

Figure 3-3 contrasts endocardial unipolar (tip-to-can), bipolar (tip-to-ring), and integrated bipolar (tip-to-coil) electrode systems, and Figure 3-4 shows representative examples. Epicardial electrode systems may be either unipolar (tip-to-can) or bipolar (tip-to-tip). These different electrode configurations have EGMs with similar R-wave amplitudes and slew rates, provided that the interelectrode spacing is at least 10 mm, as is true of almost all commercial pacemaker and defibrillator leads. Because they are more likely to oversense than bipolar EGMs, unipolar electrode systems are contraindicated for ICDs and are used infrequently for modern pacemakers. ICD integrated bipolar electrodes sense between the right ventricular tip electrode and right ventricular high-voltage coil, with sensing characteristics closer to the bipolar than the unipolar configuration. Compared with true bipolar electrodes, integrated ICD bipolar electrodes are more likely to oversense myopotentials and electromagnetic interference (EMI).3,4 In one study, oversensing occurred in 40% of patients with integrated bipolar sensing, compared with 8% of patients with true bipolar systems.4

image

Figure 3-3 The three practical endocardial electrode configurations used by most pacemakers and ICDs.

The distant and indifferent or “can” electrode is not shown because it is out of the field of view. The unipolar configuration used in Figure 3-1 to explain EGM formation simply records the signal between the tip electrode and the can. The tip electrode can be an active-fixation screw or a small-surface-area tip electrode with various geometries. This unipolar configuration is subject to considerable noise and interference signals and is not suitable for ICDs. The bipolar configuration uses the Tip and Ring electrodes shown in this figure. The interelectrode spacing is typically 12 to 15 mm, and the ring electrode may or may not make contact with the endocardium. The integrated bipolar configuration uses the Tip and RV-Coil electrodes shown in the figure. EGMs recorded from bipolar and integrated bipolar configurations are very similar, and one less conductor is needed for the integrated bipolar configuration. The main disadvantages of the integrated bipolar configuration are susceptibility to diaphragmatic myopotentials, undesired atrial EGMs in small hearts, and slower postshock recovery times caused by electrode polarization. RV, Right ventricular.

image

Figure 3-4 Ventricular electrocardiograms (ECGs) recorded from different electrode configurations in the same patient.

The central panel shows a left anterior oblique radiograph of a cardiac resynchronization ICD system. Each of the four tracings shows surface ECG lead II, EGM markers, and one ventricular EGM during atrial pacing at a rate of 75 bpm. Top left, Far-field EGM recorded between the right ventricular (RV) coil electrode and the electrically active ICD housing (CAN). Lower left, Integrated bipolar EGM recorded between RV tip and RV coil electrodes. Lower right, True bipolar EGM recorded between the RV tip and ring electrodes. Top right, Left ventricular (LV) unipolar EGM recorded between the LV tip electrode and can. EGM scale is 0.5 mV/mm, except for the LV unipolar EGM, which has a scale of 2 mV/mm. The downward EGM ventricular sense (VS) markers correspond to the time at which the true bipolar RV tip-ring EGM crosses the sensing threshold. Because the “field of view” of this EGM is local, its duration is short. It occurs early in the QRS complex of this patient with left bundle branch block. The integrated bipolar tip-coil EGM has a peak-to-peak amplitude and slew rate similar to those of the true bipolar EGM. However, its field of view is larger due to the size of the RV coil, and therefore the T wave is larger. Low-amplitude atrial EGMs are visible because of the proximity of the coil to the tricuspid anulus. Both the RV coil-CAN and the LV unipolar EGM are widely spaced, between an intracardiac electrode and the extracardiac can. Their duration is closer to that of the QRS complex. The intrinsic deflection of the LV unipolar electrode is late in the QRS complex, corresponding to late activation of the lateral left ventricle. The greater amplitude of the LV unipolar EGM reflects the greater muscle mass of the left ventricle. EGMs recorded from the superior vena cava (SVC) coil and from the atrial bipole (right atrium, RA) are not shown. Radiograph and EGMs are from different patients. Radiograph is for illustrative purposes only.

Amplitude, Slew Rate, and Waveshape

The largest and steepest deflection on the local EGM, called the intrinsic deflection, occurs when the wavefront of depolarization passes the small-tip electrode. The EGM amplitude traditionally is defined as the peak-to-peak amplitude of the intrinsic deflection (measured in mV), as shown in Figure 3-5. The duration of a ventricular EGM usually is less than that of the QRS of the surface ECG, because the EGM is a local signal. The amplitude of an atrial electrogram (AEGM) or ventricular electrogram (VEGM) is determined primarily by the excitable tissue near the tip electrode and therefore is usually similar for unipolar and bipolar signals. Typical amplitudes are 5 to 30 mV for VEGMs and 1.5 to 6 mV for AEGMs.1,2,5

The maximum slope of the intrinsic deflection is the slew rate, measured in volts per second, which represents the maximum rate of change of EGM voltage. Mathematically, it is the first derivative of the voltage, dV/dt, so it depends on both the amplitude and the duration of the EGM, and it provides a crude representation of the frequency content. The frequency content of ventricular and atrial EGMs is similar and in the range of 5 to 50 Hz. T waves and far-field R waves have lower frequency content, whereas most myopotentials and EMI have higher frequency content (Fig. 3-6). Typical values for slew rates are 2 to 3 V/sec for VEGMs and 1 to 2 V/sec for AEGMs.3,4 Usually, an EGM with acceptable amplitude also has an acceptable minimum slew rate (>1 V/sec for VEGMs, >0.3 V/sec for AEGMs). EGMs with very low amplitude will not be sensed, regardless of the slew rate.

Increasing the size of the tip electrode in the range of 2 to 10 millimeters (mm) has minimal effect on atrial EGM amplitude but increases EGM duration (Fig. 3-7). For short ventricular bipolar interelectrode spacing of 5 mm or less, the R-wave amplitude decreases, because the difference between the two unipolar EGMs from each electrode causes cancellation in the net bipolar signal. The slew rate increases, because the time between arrival of the wavefront at the two electrodes decreases more than the EGM amplitude. When two electrodes are widely separated, as in early Y-adapted cardiac resynchronization electrode systems, two distinct intrinsic deflections may be recorded on the EGM—one representing right ventricular (RV) activation and the other left ventricular (LV) activation. The interval between these deflections is determined by the conduction delay between the ventricles near the two electrodes.

The waveshapes of EGMs are quite variable (Fig. 3-8), probably because geometry of the trabecular endocardium adjacent to the tip electrode is complex. In one study at pacemaker lead implantation, 58% of unipolar EGMs were biphasic, with an initial upstroke followed by a roughly equal downstroke; 30% were predominantly monophasic negative, and 12% were predominantly monophasic positive.1

The ventricular depolarization recorded on the atrial electrode is referred to as the far-field R wave (FFRW). Oversensing of the FFRW confounds interpretation of the atrial rhythm. The amplitude of the FFRW depends on the location of the atrial electrode. It is greatest near the septum, intermediate in the right atrial appendage, and least on the right atrial free wall. Even if the FFRW has comparable amplitude to the P wave, its slew rate usually is much lower. In one series, the mean slew rate was 1.2 V/sec for AEGMs and 0.13 V/sec for FFRWs.1

If an active-fixation, screw-in tip electrode is successfully attached to the myocardium, the acute VEGM has a current of injury, with an elevated ST segment (Fig. 3-9) that is usually greatly reduced within 10 minutes after electrode fixation. During this 10-minute period, the EGM amplitude and slew rate usually do not change despite changes in waveshape, but the pacing threshold decreases by an average of 40%.2

Acute to Chronic Changes and Fixation

The amplitude and slew rate of intracardiac EGMs typically decline during the first several days to weeks after lead implantation and then increase to chronic values that are slightly lower than those measured at implantation.6 The initial decrease in EGM amplitude is caused by the inflammatory response and edema at the electrode-tissue interface. This gradually resolves and is followed by development of a small, inexcitable fibrotic zone surrounding the electrode tip (Fig. 3-10). This inflammation and fibrotic tissue effectively increases the distance between the surface of the electrode and excitable myocardium that generates the EGM signal. Although chronic EGM amplitudes usually are reduced by less than 10% compared with acute amplitudes, chronic slew rates are reduced by 30% to 40%.7

The acute reduction in EGM amplitude is often greater with active-fixation leads than with passive fixation leads. Atrial undersensing can occur during the acute phase despite adequate EGM amplitudes at implantation. To account for these time-related changes in EGM amplitude, the filtered EGM recorded at lead implantation should be at least twice the sensitivity threshold that will be programmed in the pulse generator. Greater sensing safety margins are preferred for active-fixation leads.

The method of lead tip stabilization, active screw-in or passive tines, has had no significant effect on sensing characteristics in most studies.8,9 Steroid-eluting electrodes reduce chronic pacing thresholds substantially, but also have no significant effects on sensing.1013

Metabolic, Ischemic, Aging, and Drug Effects

The effects of metabolic abnormalities and drugs on pacing thresholds are well described. Much less information is available concerning their effects on EGMs and sensing. Factors that reduce EGM amplitude, slow conduction velocity, or diminish slew rate may produce either oversensing or undersensing. By prolonging the intracardiac EGM duration beyond blanking periods, ischemia or antiarrhythmic drugs can produce double-counting of the QRS complex.14 Similarly, drugs that prolong the PR or QT interval beyond the refractory period may result in oversensing.15,16

Undersensing may result from reduction in EGM amplitude or slew rate after myocardial infarction at the electrode-tissue interface, from drug and electrolyte effects,15,16 or from progression of conduction system disease. Acute ischemia causes ST-segment changes that can be detected on VEGMs. Monitoring of EGM ST-segment shifts has been proposed as a method for monitoring ischemia for pacemakers and ICDs.17 The likelihood of recording abnormal AEGMs (defined as ≥100 msec in duration or having ≥8 fragmented deflections) correlates with age of the patient (r = 0.34; P < .0005).18

Exercise, Respiratory, and Postural Effects

The effect of exercise on the AEGM amplitude and slew rate is variable. Some studies have reported statistically significant decreases in amplitude that average 10% to 20% but may reach 40% in some patients.19,20 Other studies did not find significant changes between rest and exercise.21,22 Decreases in AEGM amplitude were not caused by atrial rate alone or by beta blockade.23 VDD/R lead studies with “floating” atrial electrodes showed particularly large decreases with exercise.24,25 Decreases in AEGM amplitude with lead maturation support the programming of a large safety margin for sensing at implantation to offset effects of lead maturation.

P-wave amplitude increases significantly during full inspiration, during full expiration, and with erect posture.22 Respiratory variation averaged 9.7% for unipolar AEGMs and 11.5% for bipolar AEGMs.25,26 The effect of respiration on VEGMs was less, especially with the unipolar configuration.26

Ventricular Electrograms during Premature Ventricular Complexes, Ventricular Tachycardia, and Ventricular Fibrillation

Premature ventricular complexes (PVCs) may have lower-amplitude R waves than sinus-rhythm R waves, as shown in Figure 3-11, but the reverse may also be true. For monomorphic ventricular tachycardia (VT), mean amplitude decreased only slightly from values in sinus rhythm—14% for epicardial EGMs and 5% for endocardial EGMs.27 In contrast, EGM amplitudes during ventricular fibrillation (VF) decreased by 25% for epicardial and 41% for endocardial EGMs. More importantly, EGMs in VF often have low, highly variable, and rapidly changing amplitudes and slew rates. Figure 3-12 shows endocardial spontaneous VF EGMs from different patients, illustrating variability in intrinsic deflections, amplitudes, slew rates, and morphologies. In a study of induced VF reproducibility, 50% of the variability was caused by interpatient differences and the other 50% occurred among repeated episodes in the same patient.28 In another study, the VEGM amplitude in VF was 1 mV or less in at least one VF episode in 29% of patients.27 If VF lasts for minutes, the amplitude and slew rate of the EGMs decrease.

Atrial Electrograms during Rhythms Other Than Sinus

Atrial activation from ectopic sites or atrial arrhythmias can alter the amplitude, frequency content, slew rate, and morphology of the AEGM. Retrograde atrial activation during ventricular pacing reduces AEGM amplitude and slew rate by up to 50%.29 These EGM changes are more pronounced in the high right atrium than in the right atrial appendage or low right atrium.30 The frequency content of the AEGM is not significantly altered by retrograde atrial activation.31 Analysis of EGM turning-point morphology or the first-differential coefficient of slew rate has been used to discriminate sinus EGMs from those recorded during retrograde and ectopic atrial activation in small groups of patients.32

Atrial EGMs during atrial fibrillation (AF) are characterized by extreme temporal and spatial variability. EGMs tend to be most organized in the trabeculated right atrial appendage and more disorganized in the smooth right atrium or coronary sinus.3335 Thus electrode spacing and positioning of atrial leads influence EGM characteristics during AF3638 and may cause inconsistent diagnosis of AF based on rate criteria.

The amplitude of chronic, unipolar pacemaker EGMs was 40% less in AF than in sinus rhythm.39 A comparison of acute AEGM amplitudes recorded with temporary pacing catheters showed that the mean sinus-rhythm EGM amplitude decreased only slightly in atrial flutter but decreased by about 50% in AF.35 Antiarrhythmic drugs may also interfere with sensing during AF by reducing atrial rate, median frequency, and EGM amplitude.40

Subcutaneous Electrocardiography

The subcutaneous ECG is similar to the surface ECG because the two subcutaneous electrodes are sufficiently distant from the heart that they record electrical activity from the entire heart. As with the surface ECG, the amplitude of subcutaneous ECG signals usually is 1 mV or less. Simultaneous recordings of subcutaneous ECG signals and surface ECG signals from electrodes placed directly over the subcutaneous locations have similar amplitude and signal-to-noise ratio.41 Practical implantation considerations usually limit the subcutaneous electrode separation distance to 4 to 8 cm, compared with the typical surface ECG limb lead electrode separation of 40 to 60 cm.

The orientation of the two subcutaneous electrodes relative to the heart can affect the amplitude of the signal recorded.42 Mapping studies on the chest skin with 4-cm electrode spacing in the range used by implantable loop recorders (ILRs) show larger intrinsic QRS amplitudes of 0.5 ± 0.1 mV for vertical orientation in the left parasternal zone and for horizontal orientation near the apex of the heart. Subcutaneous ECGs are used to detect arrhythmias in ILRs, to obviate the need for surface ECG electrodes during follow-up of pacemakers and ICDs, and to detect VT/VF in an ICD without intravascular electrodes.

image Sensing

The methods and technology of sensing and detection in ICDs and pacemakers share many features, but there are two major differences. First, ICDs need reliable sensing and detection during VF, but pacemakers do not. Second, pacemakers may use unipolar or bipolar sensing, whereas ICDs always use bipolar sensing.

General Concepts

Figure 3-13 shows the primary functional operations of sensing systems used by pacemakers and ICDs. The raw signal passes from the leads to the connector, through hermetic feedthroughs with high-frequency filters and high-voltage protection circuitry, before reaching the sensing amplifier. After the signal is amplified, a band-pass filter processes it to reduce T waves, myopotentials, and EMI (filtering). Then, it is rectified to nullify effects of signal polarity (rectification). Finally, it is compared with the sensing-threshold voltage. At the instant the processed signal exceeds the sensing-threshold voltage, a sensed event is declared to the timing circuits and indicated by a marker pulse on the programmer marker channel. The sense amplifier in the same chamber is turned off or “blanked” for a short blanking period (20-250 msec) after each spontaneous depolarization or pacing stimulus, to prevent a single depolarization resulting in multiple sensed events. In the refractory period that follows the blanking period, the sense amplifier remains enabled. Sensed events occurring in refractory periods do not alter pacemaker timing cycles but may be sensed for tachyarrhythmia detection algorithms.

Blanking and Refractory Periods

Blanking periods and refractory periods are used to prevent undesirable behavior caused by oversensing or double-counting of cardiac activity (Figs. 3-14 and 3-15). The specifications of blanking/refractory periods have substantial impact on ICD sensing and pacing functions (Fig. 3-16). Same-chamber blanking/refractory periods after sensed events reduce double-counting of intrinsic cardiac depolarizations that may result in escape pacing at a rate slower than the programmed lower rate in pacemakers or inappropriate detection of VF in ICDs. After paced events, the same-chamber blanking/refractory periods are typically longer and prevent oversensing of the pacing artifact and evoked response. The blanking/refractory periods in the ventricle after atrial sensed or paced events and in the atrium after ventricular sensed or paced events are called cross-chamber blanking/refractory periods. Cross-chamber blanking periods help to prevent oversensing of the pacing artifact after a paced event in the opposite chamber.

The atrial blanking period after ventricular events, postventricular atrial blanking (PVAB), is designed to avoid oversensing of ventricular pacing stimuli and FFRWs. Longer postventricular atrial refractory periods (PVARPs) prevent retrogradely conducted atrial activation from resetting atrial timing cycles for dual-chamber pacing. Cross-chamber blanking in the atrium after a ventricular event must be minimized in ICDs with tachyarrhythmia detection (ICDs or atrial therapy ICDs) to avoid undersensing the atrial rhythm, particularly during high ventricular rates. Long PVAB periods prevent reliable sensing of AF and atrial flutter/atrial tachycardia (AT). However, short PVAB periods may result in atrial sensing of FFRWs.

Implantable cardioverter-defibrillators that require tachyarrhythmia detection typically have shorter blanking and refractory periods than standard pacemakers, so that short cardiac cycles can be sensed reliably. As shown in the bottom marker diagrams of Figure 3-16, blanking periods may be adaptively extended based on noise-sampling windows (30-60 msec) if suprathreshold activity (due to cardiac or extracardiac sources such as EMI) is identified on the EGM immediately after a sensed event. If noise is seen in consecutive windows after a sensed event, the blanking period is “retriggered” for that beat to avoid double-counting or continuous oversensing. This operation may result in paradoxical undersensing of the cardiac rhythm when more sensitive sensing levels are programmed if noise is oversensed.43,44

The duration of the total atrial refractory period (TARP), equal to the atrioventricular (AV) delay plus the PVARP, in DDD pacing modes limits atrial tracking of the atrium at high sinus rates without affecting atrial sensing, as shown in Figure 3-17. Because the AV delay of most dual-chamber pacemakers shortens in response to increasing atrial rates or sensor input, the TARP also shortens. Several manufacturers now offer dual-chamber pacemakers that shorten the PVARP with increasing atrial or sensor-indicated rates, further reducing the TARP during exercise. The result of these newer algorithms is that the programmed upper tracking rate can be safely increased while providing protection at lower heart rates from initiation of pacemaker-mediated tachycardia caused by retrograde conduction.

Sensing Thresholds in Pacemakers

Sensing thresholds in most pacemakers are programmable to a constant value. Ventricular sensing channels in conventional pacemakers typically operate at sensing thresholds of 2.5 to 3.5 mV, about 10 times less sensitive than those in ICDs. Therefore, pacemakers may undersense VF. Atrial sensitivity thresholds are typically 0.3 to 0.6 mV, to allow sensing of small-amplitude atrial EGMs during AF and to improve the accuracy of AF diagnostics.

Unipolar sensing thresholds typically are set higher (less sensitive) than bipolar sensing thresholds to reduce oversensing of far-field cardiac and extracardiac signals that can lead to inappropriate pacemaker inhibition or tracking. Newer pacemakers automatically adjust the sensitivity setting to adapt to changes in EGM amplitude over time. Typically, these functions operate to modify sensing thresholds based on a series of 10 to 20 ventricular beats. One such algorithm employs two simultaneous sensing levels: the programmed sensitivity (inner target) and a value twice the programmed value (outer target) (Fig. 3-18).45 Sensed EGMs exceeding both target values decrease the sensitivity. Signals exceeding only the inner target increase the sensitivity. In this manner, a 2 : 1 sensing margin is maintained. Rapid, automated sensitivity adjustments may be desired when EGM amplitudes can be expected to change over a brief period, such as beat-to-beat variations from respiration, body position changes, or fluctuating EGM morphologies during AF.46

image

Figure 3-18 Autosensing algorithm to maintain a 2 : 1 sensing safety margin.

See text for details.

(From Castro A, Liebold A, Vincente J, et al: Evaluation of autosensing as an automatic means of maintaining a 2:1 sensing safety margin in an implanted pacemaker. Autosensing Investigation Team. Pacing Clin Electrophysiol 19:1708-1713, 1996.)

Far-field R-wave oversensing can be minimized by (1) selecting an atrial lead with a closely spaced bipolar electrode pair (≤10 mm), (2) choosing an implantation location that yields an FFRW/P-wave ratio of less than 0.5,47 (3) titrating programmed sensitivity to reject FFRWs without undersensing P waves and low-amplitude AF, and (4) using PVAB.

Ventricular Sensing in Ventricular ICDs

The guiding design principle is that sensing of VF and polymorphic VT should be sufficiently reliable that clinically significant delays in detection do not occur. Although high sensitivity is required to ensure reliable sensing during VF, continuous high sensitivity results in oversensing of cardiac or extracardiac signals during regular rhythm, which may cause inappropriate detection of VT or VF. To minimize both undersensing during VF and oversensing during regular rhythms, ICDs use feedback mechanisms based on R-wave amplitude that adjust the sensing threshold dynamically. To maximize the likelihood of detecting VF, blanking periods are kept short.

Automatic Adjustment of Sensitivity

Adjustment of Sensitivity in Normal Rhythm

All ICDs automatically adjust sensitivity in relation to the amplitude of each sensed R wave (Fig. 3-19). At the end of the blanking period after each sensed ventricular event, the sensing threshold is set to a high value. It then decreases with time until a minimum value is reached. Compared with a fixed sensing threshold, automatic adjustment of sensitivity increases the likelihood of sensing low-amplitude and varying EGMs, while minimizing the likelihood of T-wave oversensing.

The methods of the different manufacturers for automatic adjustment of sensitivity perform similarly after small R waves but differently after large R waves. Figure 3-20 shows that after large R waves, the Boston Scientific ICDs increase the sensing floor. In the Cognis/Teligen family, the sensing floor is set to one-eighth the amplitude of the measured R wave if that value is greater than the programmed sensitivity. This prevents T-wave oversensing in the setting of large R waves and reduces oversensing of low-amplitude noncardiac signals (e.g., diaphragmatic myopotentials, EMI). However, it may increase the risk of undersensing during rare episodes of VF with highly variable EGM amplitude.48

image

Figure 3-20 Comparison of automatically adjusted sensitivity after sensed ventricular events for three manufacturers of ICDs after large (10 mV) R wave.

The programmed sensing threshold is approximately 0.3 mV. After sensed ventricular events, Medtronic ICDs reset the sensing threshold to 8 to 10 times the programmed sensitivity, up to a maximum of 75% of the sensed R wave. The value of Auto-Adjusting Sensitivity then decays exponentially from the end of the (sense) blanking period, with a time constant of 450 msec, until it reaches the programmed (maximum) sensitivity. At the nominal sensitivity of 0.3 mV, there is little difference between the sensitivity curves of Medtronic ICDs after large and small spontaneous R waves. If the R wave is large, the entire Auto-Adjusting Sensitivity curve can be altered substantially by changing the programmed value of maximum sensitivity (see Fig. 3-19). At nominal settings, the St. Jude Threshold Start begins at 62.5% of the measured R wave for values between 3 and 6 mV. If the R-wave amplitude is greater than 6 mV or less than 3 mV, the Threshold Start is set to 62.5% of these values (3.75 mV and 1.875 mV, respectively). The sensing threshold remains constant for a Decay Delay period of 60 msec and then decays linearly with a slope of 3 mV/sec. Both the Threshold Start percent and the Decay Delay are programmable, over the range of 50% to 75% and 0 to 220 msec, respectively (see Fig. 3-21). Boston Scientific Cognis-Telegin ICDs set the starting threshold to 75% of sensed R waves with a maximum limit of 3/2 · Peak Running Average. Sensitivity then decays using digital steps, each seven-eighths the amplitude of the previous step. For sensed events, the duration of the first step is 65 msec, and the duration of subsequent steps is 35 msec. This results in a sensitivity of one-half the peak R wave in about 170 msec. (See text for further details.) After a paced ventricular event, all ICDs also adjust sensitivity dynamically, starting at the end of the (pace) blanking period, but the threshold starts at a more sensitive setting.

(Modified from Swerdlow C, Friedman P: Advanced ICD troubleshooting. Part I. Pacing Clin Electrophysiol 28:1322-1346, 2005.)

Ventricular Blanking Periods

Ventricular blanking periods prevent ventricular oversensing of same-chamber signals (R-wave double-counting) and cross-chamber signals (atrial pacing pulses and P waves) in regular rhythms. Because reliable sensing in VF requires minimizing blanking periods, blanking periods in ICDs are short and may occasionally be insufficient to prevent oversensing.

“Sensing” Other Ventricular Electrograms

Implantable cardioverter-defibrillator may use information derived from other VEGMs (shock-channel EGM, LV EGM in resynchronization systems) to enhance their functionality. The EGM from the shock channel (far-field EGM) is not used by transvenous ICDs for rate counting because bipolar or integrated bipolar EGMs are less susceptible to extracardiac signals. However, some manufacturers (Boston Scientific, Medtronic) use morphologic characteristics of far-field EGMs for discrimination of supraventricular tachycardia (SVT) from VT.

Automatic analysis of the far-field EGM has also been proposed as a method to identify oversensing in ICDs resulting from lead fracture or sensing-lead connection problem. In one approach, the peak-peak far-field EGM amplitude is measured in a small window centered around each sensed event (on near-field channel) to discriminate rapid oversensing from true VT/VF. Oversensing is identified when sensed events on the near-field channel correspond to isoelectric periods on the far-field channel.55,56 In true VF, isoelectric periods are rare on the far-field channel (Fig. 3-23).

Evaluating Sensing of Ventricular Fibrillation at Implantation

Increasing interest in implanting ICDs without assessing defibrillation efficacy has focused attention on the extent to which adequacy of VF sensing can be determined from EGMs recorded in baseline rhythm. Although the statistical correlation between R-wave amplitude in VF and baseline rhythm is weak,57,58 two studies reported that sensing of VF is adequate with nominal sensitivities near 0.3 mV if the baseline R wave is sufficiently large (≥5 mV or ≥7 mV).59 Rarely, clinically significant undersensing of VF or polymorphic VT may occur despite adequate sinus-rhythm R waves.48,60 In these cases, undersensing occurs because auto-adjusting sensitivity criteria respond inadequately to variations in R-wave amplitude, rather than consistently low-amplitude R waves. The reproducibility of this phenomenon is unknown, as is its predicted extent at implantation.

Therefore, it is uncertain whether clinically appropriate testing at implantation can detect this infrequent cause of undersensing. During ICD implant with true bipolar sensing and current digital sensing amplifiers, clinically significant undersensing of VF is rare and unrelated to sinus-rhythm R-wave amplitude.61 Undersensing of spontaneous VT/VF in the VF zone is similarly rare. Reliable sensing of VF cannot be predicted from baseline EGMs if the baseline ventricular rhythm is paced. Sensitivity is programmed to a less sensitive value than nominal (e.g., to avoid T-wave oversensing), or patients have other implanted electronic ICDs, such as pacemakers, cardiac contractility modulation devices, or transcutaneous electrical nerve stimulation (TENS) units, that could cause device-device interactions.

Postshock Sensing

Postshock sensing is critical for redetection of VF after unsuccessful shocks and for accurate detection of episode termination. Electroporation, the process by which strong electric fields create microscopic holes in the cardiac cell membranes, has been proposed as the mechanism for postshock distortion of EGMs recorded from high-voltage electrodes.62 Because EGMs of dedicated bipolar sensing electrodes are minimally affected by shocks,63 they became standard for early epicardial ICDs. For transvenous ICDs, postshock sensing recovers more rapidly with true bipolar sensing configurations than with integrated bipolar sensing.64,65 This is a minor issue for current integrated bipolar leads with a pacing tip electrode–to–distal coil spacing of approximately 12 mm.66

Atrial Sensing in Dual-Chamber ICDs and Atrial ICDs

Accurate sensing of atrial EGMs is essential for accurate discrimination between VT/VF and rapidly conducted SVTs that satisfy ventricular rate criteria in dual- or triple-chamber ICDs. Rapid discrimination is essential to ensure prompt delivery of ventricular therapy while minimizing inappropriate shocks. Historically, some inappropriate detection of AT/AF has been considered an acceptable consequence of maintaining high sensitivity for detecting VT/VF.

The atrial lead should be positioned at implantation to minimize FFRWs. Leads with an interelectrode spacing of 10 mm or less reduce oversensing of FFRWs. Atrial lead dislodgement, oversensing of FFRWs, or undersensing from low-amplitude AEGMs or atrial blanking periods can cause inaccurate identification of AEGMs. These errors in sensing may result in misclassification of VT as SVT, or vice versa.

Postventricular Atrial Blanking and Rejection of Far-Field R Waves

To prevent oversensing of FFRWs, older dual-chamber ICDs had fixed PVAB periods, similar to those in pacemakers (Fig. 3-24). With a fixed blanking period, the blanked proportion of the cardiac cycle increases with the ventricular rate. Atrial undersensing caused by PVAB causes underestimation of the atrial rate during rapidly conducted atrial flutter or AF, resulting in inappropriate detection of VT67 (Fig. 3-24, lower panel). Without PVAB, however, atrial oversensing of FFRWs could cause overestimation of the atrial rate during tachycardias with a 1 : 1 AV relationship.68 This may result in either inappropriate rejection of VT as SVT, if FFRWs are counted consistently as atrial EGMs, or inappropriate detection of SVT as VT, if FFRWs are counted inconsistently.69

Medtronic ICDs also reject FFRWs algorithmically by identifying a specific pattern of atrial and ventricular events that fulfill specific criteria (Fig. 3-25). Intermittent sensing of FFRWs or frequent premature atrial events may disrupt this pattern, resulting in misclassification of a tachycardia. Therefore, it is preferable to reject FFRWs after sensed ventricular events by decreasing atrial sensitivity, if this can be done without undersensing of AF. Atrial sensitivity can be reduced to 0.45 mV with a low risk of undersensing AF. Less sensitive values should be programmed only if the likelihood of rapidly conducted AF is low. FFRW oversensing that occurs only after paced ventricular events (when auto-adjusting atrial sensitivity is maximal) does not cause inappropriate detection of SVT as VT, but it may cause inappropriate mode switching and can contribute to inappropriate detection of AF or atrial flutter.

image image

Figure 3-25 Algorithmic rejection of far-field R waves (FFRW) by pattern analysis.

ICDs with minimum cross-chamber blanking (Medtronic) reject FFRWs by the timing pattern of atrial and ventricular intervals. Atrial events are classified as FFRWs if all the following criteria are met: (1) there are exactly two atrial events for each V-V interval; (2) timing of one P wave is consistent with a FFRW (R-P interval <160 msec); (3) there is a stable interval between the FFRW and the ventricular electrogram (VEGM); (4) there is a short-long pattern of P-P intervals (to distinguish FFRW oversensing from atrial flutter); and (5) the pattern occurs frequently (4 of 12 intervals). A, Atrial electrogram (AEGM), VEGM, and dual-chamber event markers are shown. All five criteria for FFRWs are fulfilled. Horizontal double-ended arrows below event markers denote alternation of long (L) and short (S) atrial intervals. AR, Atrial refractory event; P, P wave; TS, ventricular event in ventricular tachycardia zone. B, Interval plot (left) and stored EGM (right) from episode of sinus tachycardia with consistent FFRW oversensing that is classified correctly. On the interval plot, open squares denote A-A interval and closed circles denote V-V interval. Horizontal lines denote ventricular tachycardia (VT) and ventricular fibrillation (VF) detection intervals of 400 and 320 msec, respectively. Alternating A-A intervals whose sum equals that of the VV intervals produce a characteristic “railroad track” appearance (arrow). The algorithm rejects FFRWs despite that FFRW oversensing does not occur for one V-V interval between seconds 9 and 10. C, Interval plot (left) and stored EGM (right) from episode of sinus tachycardia with consistent FFRW oversensing that was detected inappropriately as VT. Intermittent oversensing of FFRWs occurs as sinus tachycardia accelerates gradually across the VT detection interval of 480 msec (arrow), resulting in inappropriate therapy (“Burst” antitachycardia pacing marker on interval plot, VT marker on EGM event markers).

Medtronic ICDs (starting with Entrust) and Boston Scientific ICDs (starting with Vitality) may use brief atrial blanking or a period of reduced, automatically-adjusting sensitivity (or both) to reject FFRWs without preventing detection of AF (Fig. 3-26). St. Jude ICDs and Medtronic ICDs starting with Entrust provide programmable atrial blanking after sensed ventricular events to individualize the trade-off between oversensing of FFRWs and undersensing of AEGMs in AF. St. Jude ICDs also provide programmable atrial sensing Threshold Start and Decay Delay, corresponding to the same features in the ventricular channel.

image

Figure 3-26 Atrial sensing features intended to prevent far-field R-wave (FFRW) oversensing while still permitting detection of atrial fibrillation.

These features, which combine limited atrial blanking with automatic adjustment of atrial sensitivity, are less strict than conventional blanking periods. A, Medtronic ICDs, starting with Entrust, have an optional “Partial+” postventricular atrial blanking period (PVAB) setting that increases the sensing threshold for a programmable duration (40-100 msec) after a ventricular event (2). Entrust operation after ventricular sensed events nominally has no absolute blanking or sensitivity changes. With Partial + PVAB enabled, the sensing threshold is raised to 75% of the prior sensed atrial electrogram (AEGM) amplitude to reduce the likelihood of sensing FFRWs. “Absolute” PVAB, another selectable option, blanks all atrial sensed events within the programmable PVAB interval. Because significant atrial undersensing can occur, this method is recommended only for addressing complications not addressed by the Partial + PVAB method. Partial + or Absolute PVAB does not affect atrial sensing operation after sensed events (1) or after atrial paced events (3). AP, Atrial paced event; AS, Atrial sensed event; mS, microseconds; VS, ventricular sensed event. B, Boston Scientific ICDs, starting with Vitality, introduce a brief (15 msec) blanking period after each sensed ventricular event, followed by a decrease in atrial sensitivity to three-eighths the mean P-wave amplitude. This is sufficient to prevent oversensing of FFRWs, provided that the amplitude of the AEGM is at least eight-thirds that of the FFRW.

Sensing in Cardiac Resynchronization ICDs

Left ventricular PVCs may activate the left ventricle but may not be sensed in the right ventricle before pacing is delivered. In this case, LV pacing could be delivered during the vulnerable period of the PVC. Boston Scientific cardiac resynchronization ICDs also sense in the left ventricle to reduce ventricular proarrhythmia by preventing pacing into the LV vulnerable period. The left ventricular protection period (LVPP) is a programmable interval (300-500 msec) following an LV event when the ICD will not pace in the left ventricle (Fig. 3-27). This prevents inadvertent delivery of an LV pacing pulse during the LV vulnerable period. The LVPP differs from other pacing refractory periods, which are designed to prevent inappropriate inhibition of pacing.

image

Figure 3-27 Event markers illustrate blanking periods in cardiac resynchronization ICDs.

A, Blanking periods in Medtronic Marquis Insync III ICD. In condition 1, there is simultaneous right ventricular and left ventricular pacing (RV+LVp). There is a single cross-chamber blanking period (~30 msec in duration) on the atrial channel. In condition 2, RVp first with LVp delayed (up to 80 msec), and in condition 3, LVp first with RVp delayed, there are two cross-chamber blanking periods on the atrial channel, one for each ventricular paced event. In both cases, ventricular blanking is timed from the RVp event, resulting in a postpace ventricular blanking period that is longer by the V-V pacing delay (up to 80 msec). These additional blanking periods increase the risk of undersensing in both chambers. To date, there are no reports of undetected ventricular tachycardia (VT) caused by offsets in timing of V-V pacing. B, Left ventricular protection period (LVPP) of the Boston Scientific Renewal ICD. This ICD has RV timing and independent RV and LV stimulation channels without a programmable RV-LV delay. The LVPP allows LV sensing to occur with inhibition of the LV stimulus for a period that is programmable to prevent pacing into the vulnerable period of the left ventricle. Therefore, this feature allows biventricular sensing but RV-based timing. AS, Atrial sensed event; BiV, biventricular; LVP, left ventricular paced; LVS, left ventricular sensed; RVP, right ventricular sensed.

(Courtesy Boston Scientific, Indianapolis; B from Kay G: Troubleshooting and programming of cardiac resynchronization therapy. In Ellenbogen K, Kay G, Wilkoff B, editors: ICD therapy for congestive heart failure. Philadelphia, 2004, Elsevier.)

In contrast, the left ventricular refractory period (LVRP), after a sensed or paced event on the LV lead, is a conventional refractory period. It prevents sensed events from causing inappropriate loss of cardiac resynchronization pacing following sensed events such as T-wave oversensing on the LV lead. The LVRP provides an interval following either an LV sense or an LV pace event (or leading ventricular pace event when LV offset is not programmed to zero), during which LV sensed events do not inhibit pacing. Use of a long LVRP shortens the LV sensing window. LVRP is available whenever LV sensing is enabled. Thus, LVRP minimizes unnecessary inhibition of resynchronization pacing while the LVPP minimizes the risk of LV pacing during the LV vulnerable periods.

image Ventricular Oversensing: Recognition and Troubleshooting

Oversensing is defined as sensing of unintended nonphysiologic signals or of physiologic signals that do not accurately reflect local depolarization. Nonphysiologic signals usually arise from extracardiac EMI (e.g., ungrounded electrical equipment, antitheft detectors, surgical electrocautery) or mechanical problems in the sensing circuit, which may occur in the lead (insulation failure or fracture), connection between lead and header, or feedthrough wires from the header to the inside of the pulse generator. Rarely, electrical artifacts arise from retained fragments of abandoned intracardiac leads. Physiologic signals may be intracardiac (P, R, or T waves) or extracardiac myopotentials. Oversensing often results in characteristic patterns of stored EGMs and associated markers70,71 (Fig. 3-28). In pacemakers, oversensing may manifest as either failure to deliver an expected pacing stimulus or inappropriate tracking of nonphysiologic signals in the atrial channel of a dual-chamber ICD. In ICDs, oversensing usually manifests as inappropriate detection of VT or VF. This section focuses on oversensing in ICDs.

Intracardiac Signals

Ventricular oversensing of physiologic intracardiac signals results in exactly two ventricular events for each cardiac cycle. This may result in inappropriate detection of VT or VF.

T-wave Oversensing

T-wave oversensing is the most common clinically-significant cause of intracardiac oversensing. Oversensing of spontaneous T waves may cause inappropriate detection of either VT or VF, depending on the sensed R wave–to–T wave interval and the programmed VF detection interval. T-wave oversensing is typically identified by alternating EGM morphologies.70 Device-measured intervals during T-wave oversensing may also alternate, but the degree of alternation often is small and difficult to recognize, especially during sinus tachycardia or if the QT interval is prolonged.

T-wave oversensing may be divided into three classes: postpacing, large R wave (>3 mV) in spontaneous rhythm, and small R wave (<3 mV) in spontaneous rhythm (Fig. 3-29). Postpacing T-wave oversensing can inhibit bradycardia pacing49,72 or cause antitachycardia pacing to be delivered at the wrong cycle length.49 It does not typically cause inappropriate detection of VT but may increment VT or VF counters, making detection of nonsustained VT more likely. It is corrected by increasing the postpacing ventricular blanking period.

Oversensing of spontaneous T waves often occurs in the setting of low-amplitude R waves because the sensing threshold decays from a fraction of the preceding low-amplitude R wave.73 To compound the problem, patients with low-amplitude R waves may require lower minimum sensing thresholds to ensure reliable sensing of VF. T-wave oversensing in these patients is a warning that detection of VF may be unreliable. The ventricular lead should be revised or a separate pace/sense lead added if the safety margin for sensing VF is insufficient.

Specific programming features may be used to reduce T-wave oversensing, provided that detection of VF is reliable, as follows:

image

Figure 3-32 Algorithm to reject T-wave oversensing.

Upper tracing, Vtip-Vring EGM with large T waves and T-wave oversensing. Middle tracing, Signal after standard sense amplifier filtering and rectification (purple line), automatic adjusting sensing threshold (red), and peak amplitude at each sensed event (blue; the peak amplitude for each sensed event is held until the next sensed event). Bottom tracing, First derivative of the filtered and rectified EGM (purple line) its peak amplitude at each sensed event (blue), and adaptive threshold used to separate possible R waves and T waves (red horizontal line). The algorithm recognizes oversensed T waves with lower frequency content than the preceding sensed R waves by automatically comparing signal amplitudes in the filtered/rectified sensing EGM and its first derivative (which further attenuates low-frequency signals) to identify patterns of alternating signal frequency content. Using an analysis window of six sensed events, the algorithm first automatically identifies possible R waves and T waves based on comparing peak amplitudes and patterns of the first derivative of the filtered/rectified sensing EGM. For each analysis window, the adaptive threshold is set to a fraction of the three largest first differential EGM peak amplitudes. The algorithm assumes that possible R waves are above threshold and possible T waves are below threshold. An alternating RT pattern with 3Rs and 3Ts or 4Rs and 2Ts in the six sensed events analysis window is required. RT pairs are confirmed when the R-wave/T-wave (R/T) ratio in the filtered/rectified sensing EGM is less than the R/T ratio measured in the first differential filtered/rectified EGM. This approach does not increase the risk of undersensing because the operation of the standard ICD sensing channel is not modified. It does not prevent oversensing of T waves, but rather allows the ICD to withhold therapy when T-wave oversensing pattern is recognized. Testing on more than 1000 VT/VF episodes showed no loss in sensitivity to true VT/VF. Vs, Ventricular sensed event.

R-wave Double-Counting

R-wave double-counting occurs if the duration of the sensed EGM exceeds the short ventricular blanking period in ICDs; currently, this is a rare clinical problem. Consistent R-wave double-counting may result from local ventricular delays in the baseline state in native rhythm or those resulting from use-dependent effects of sodium channel–blocking antiarrhythmic drugs or of hyperkalemia. Double-counting may also occur with loss of RV capture in cardiac resynchronization ICDs: The ICD counts both the paced ventricular event and the conducted wavefront arriving at the RV bipole as a result of LV capture, if the interventricular conduction delay exceeded the ventricular blanking period. R-wave double-counting was a common problem in early ICDs that used Y-adapted or extended bipolar sensing between RV and LV electrodes.75 The composite ventricular EGM included deflections from the right ventricle and the left ventricle, both of which could be counted as separate R waves. Occasional R-wave double-counting may occur during PVCs and is more common with integrated bipolar than true bipolar leads.76

R-wave double-counting results in alternation of ventricular cycle lengths that produces a characteristic “railroad track” pattern on a plot of stored ventricular intervals. Because the second component of the R wave is sensed as soon as the blanking period terminates, the double-counted RV-RV or RV-LV interval measures within 20 msec of the ventricular blanking period and is always classified in the VF zone (Fig. 3-33). Inappropriate detection of VF may occur despite a true ventricular rate below the VT detection interval.

Most consistent R-wave double-counting now results from transient or reversible events such as hyperkalemia, drug effects, or lead failure in cardiac resynchronization ICDs; and most intermittent R-wave double-counting is not clinically significant. Some ICDs permit programming to increase the ventricular blanking period from the nominal value (120-125 msec) to higher values, up to 170 msec. Occasionally, reducing ventricular sensitivity can avoid R-wave double-counting; but ventricular sensitivity should not be reduced unless reliable sensing of VF is confirmed at the reduced level of sensitivity. Rarely, lead revision is required.

P-wave Oversensing

P-wave oversensing may occur if the distal coil of an integrated bipolar lead is close to the tricuspid valve and if the sensed P-R interval exceeds the cross-chamber ventricular blanking period4 (Fig. 3-34). It is rare in adults with ventricular sensing electrodes near the RV apex, but it may occur in children or in adults if the RV electrode dislodges or is positioned in the proximal septum or inflow portion of the right ventricle. If P-wave oversensing occurs during a 1 : 1 rhythm, the sensed “R-R” pattern is similar to that of R-wave double-counting, provided that the sensed P-R or R-P interval is less than the VF detection interval. However, oversensing of P waves as R waves can cause inappropriate detection of VF during AF or atrial flutter, independent of the ventricular rate.

Consistent oversensing of spontaneous P waves often requires lead revision. One amelioration strategy is to force atrial pacing using DDDR or Dynamic Overdrive modes. This shortens the ventricular cycle length (preventing ventricular sensitivity from reaching its minimum value) and introduces cross-chamber ventricular blanking after each atrial event (reducing the likelihood of oversensing P waves).

Extracardiac Signals

The distinctive feature of oversensing of extracardiac signals is replacement of the isoelectric baseline with high-frequency noise that does not have a constant relationship to the cardiac cycle.70,71,75,77

External Electromagnetic Interference

With oversensing of external electromagnetic interference, signal amplitude is greater on the high-voltage EGM recorded from widely spaced electrodes than on the sensing EGM recorded from closely spaced electrodes. Although the interference signal may be continuous, oversensing is often intermittent because of auto-adjusting sensitivity. Clinical data may suggest a specific identifiable cause (Fig. 3-35). Some manufacturers (St. Jude, Sorin, Boston Scientific) have specific ICD algorithms to prevent inappropriate detection of VF in the presence of EMI (Fig. 3-36).

Lead/Connector Problems

Oversensing caused by lead insulation failure or conductor fracture, lead-lead mechanical interactions (“chatter”),76 or connector (header, adapter, set-screw) problems is intermittent and may occur only during a small fraction of the cardiac cycle. In ICDs, it is limited to the sensing EGM unless the problem relates to the RV coil in an integrated bipolar lead or to both sensing and high-voltage conductors/connectors. The rate of oversensed signals is always in the VF zone. Intervals below 200 msec are typical. Signals almost always display substantial variability in amplitude or frequency. High-frequency components are common, and high-amplitude signals often saturate the amplifier (Figs. 3-37 to 3-39).

In ICDs, extremely short R-R intervals near the ventricular blanking period (120-150 msec) do not represent successive cardiac depolarizations, except occasionally during VF. Medtronic ICDs count these very short intervals as a measure of nonphysiologic oversensing to provide early warning of lead fracture; they store this count as the sensing integrity counter (SIC). A high or rapidly accumulating Sensing Integrity Count (total >300 or 10/day for 3 consecutive days) is a sensitive indicator of pace-sense lead fracture,3,21,22 but in isolation is nonspecific. The most common cause of isolated, extremely short “R-R” intervals is a combination of an oversensed physiologic event and appropriately sensed R wave.

In P-wave oversensing, the nonphysiologic short interval is the P-R interval. In R-wave double-counting, it is the interval between initial and terminal deflections of a true ventricular EGM. In T-wave oversensing, the interval is bounded by an oversensed T-wave and a PVC.76 However, these nonphysiologic combinations of physiologic signals rarely result in repetitive oversensing, which is common in lead or connector problems. Repetitive, transient oversensing may be identified by stored “nonsustained tachycardias.” The combination of isolated, extremely short RR intervals and repetitive rapid oversensing (defined as at least two nonsustained tachycardias with duration ≥5 intervals and mean cycle length ≤220 msec) has a positive predictive value for lead or connector problems of about 80%,76,78 even with a normal pace-sense impedance. Unusual cases of true VF76 and other unusual conditions, such as device-device interactions or header seal-plug problems shortly after implant,79 can produce both isolated, very short intervals and rapid, repetitive oversensing.

Role of lead impedance

Modern pacemakers and ICDs perform daily, automated measurements of impedance in lead conductors. A large, abrupt change in impedance associated with rapid oversensing is, for practical purposes, diagnostic of a lead or header-connector problem. However, pace-sense conductor fractures typically present as rapid, repetitive oversensing without abnormal impedance.78 A fracture may generate sufficient oversensing to cause inappropriate shocks without an impedance increase or with an impedance increase in the interval between daily impedance measurements.78,8082 Rarely, pacing precipitates lead-related oversensing.81 Lead fractures or header-connector problems also present frequently as abrupt increases in impedance without oversensing.

The Medtronic Lead Integrity Alert (LIA) is triggered by either abnormally high impedance relative to the patient’s baseline impedance or rapid oversensing unlikely to represent a physiologic event55,78 (Figs. 3-40 and 3-41). Once an alert is triggered, an audible tone sounds immediately and every 4 hours thereafter. LIA alerts also reprogram the number of intervals to detect VF (NID) from the value programmed at the physician’s discretion (nominally 18/24) to 30/40; EGMs are stored for any interval shorter than 200 msec. The purpose of the high NID is to reduce inappropriate shocks caused by transient, fracture-induced oversensing.78 The stored EGMs provide a diagnostic clue to determine the cause of rapid oversensing based on EGM characteristics.76 In newer ICDs with wireless telemetry, alerts in both groups may also initiate Internet notifications.

Myopotential Oversensing

Myopotential oversensing may persist for variable fractions of the cardiac cycle. Diaphragmatic myopotentials are most prominent on the sensing EGM. Oversensing usually occurs after long diastolic intervals or after ventricular paced events when amplifier sensitivity or gain is maximal. It often ends with a sensed R wave, which abruptly reduces sensitivity. In pacemaker-dependent patients, diaphragmatic oversensing causes inhibition of pacing, resulting in persistent oversensing and inappropriate detection of VF (Fig. 3-42). Clinically, this may manifest as syncope from inhibition of pacing followed by an inappropriate shock. This is an exception to the clinical rule that antecedent syncope usually indicates an appropriate shock. A short time constant for automatic adjustment of sensitivity increases the probability of this type of oversensing. It is most common in male patients who have integrated bipolar leads in the RV apex.3,4 Oversensing of diaphragmatic myopotentials may be corrected by reducing ventricular sensitivity, provided that VF sensing and detection are reliable at the reduced level of sensitivity. In pacemaker-dependent patients, oversensing may also be reduced by pacing at a faster rate. Occasionally, correction requires insertion of a new rate-sensing lead away from the diaphragm.

Pectoral myopotentials are more prominent on a far-field EGM that includes the ICD can rather than the near-field EGM. Because ICDs do not use this EGM for rate counting, oversensing of pectoral myopotentials does not cause inappropriate detection, but they may reduce the effectiveness of SVT-VT discrimination by EGM morphology algorithms during exercise-induced sinus tachycardia. Pectoral myopotentials are a major cause of oversensing in unipolar pacemakers and may result in either inhibition of ventricular pacing (if sensed by ventricular channel) or rapid ventricular pacing (if sensed and tracked by atrial channel). Myopotential oversensing may be suspected if symptoms occur during arm motion and can often be demonstrated in the clinic by having the patient forcefully press the hands together while the EGM is monitored.

Detection Algorithms for Automatic Optimization of Pacemaker Function

Pacemakers and ICDs incorporate special algorithms to prevent serious errors in pacemaker function such as inhibition during oversensing, inappropriately high pacing rates caused by tracking of atrial tachyarrhythmias, unnecessary ventricular pacing in patients with intact AV conduction, and loss of pacemaker capture. Ventricular safety pacing prevents inappropriate pacemaker inhibition caused by ventricular oversensing of atrial pacing stimuli (crosstalk). Noise reversion to fixed-rate pacing prevents pacemaker inhibition during continuous ventricular oversensing. Automatic mode switching in dual-chamber pacing systems helps avoid inappropriate tracking of high atrial rates during atrial tachyarrhythmias.

Ventricular Safety Pacing and Noise Reversion

Ventricular safety pacing prevents inappropriate inhibition of ventricular pacing after atrial paced events. This feature augments the protection provided by cross-chamber blanking after atrial pacing stimuli. If a ventricular event is sensed after the delivery of an atrial pacing stimulus in a “nonphysiologic” period (10-100 msec) during the AV delay, a ventricular pacing stimulus is delivered at the end of the nonphysiologic interval (110-120 msec). This prevents pacemaker crosstalk from causing inhibition of ventricular pacing output, but it may also result in (noncaptured) ventricular pacing immediately after true ventricular events that occur during the AV delay. Ventricular safety pacing is especially likely to occur during atrial undersensing, during AF, or during a junctional rhythm.

Pacemakers also contain algorithms to protect against prolonged inhibition of ventricular pacing caused by oversensing. Noise-sampling windows (30-60 msec in duration) are used during the blanking periods to identify spurious signals and cause the ICD to change to an asynchronous “reversion” pacing mode, to ensure ventricular backup pacing.83 Noise-reversion operation is particularly important for unipolar sensing, which is more likely than bipolar sensing to exhibit oversensing of extracardiac signals84 (Fig. 3-43).

Automatic Mode Switching to Avoid Ventricular Tracking of Atrial Tachyarrhythmias

Pacemakers programmed to dual-chamber synchronous pacing modes may use automatic “mode switching” algorithms to initiate a temporary change to a nontracking pacing mode during paroxysmal atrial tachyarrhythmias, to avoid inappropriately high ventricular pacing rates. Methods of mode switching differ among pacemakers and ICDs and can be classified into five different categories85,86 (Fig. 3-44). Accurate atrial sensing is critical to appropriate mode switching, because all methods depend on measurement of atrial rate, A-V patterns, or both. The atrial sensing configuration (unipolar vs. bipolar), programmed atrial sensitivity, and atrial blanking periods influence the methods used by the various manufacturers and their performance for detection of atrial tachyarrhythmias. Algorithms to recognize repetitive blanking of atrial events during atrial flutter allow mode switching to occur more rapidly when this condition is confirmed (Fig. 3-45). Whatever method is employed, unreliable atrial sensing, whether from low-amplitude P waves or insensitive programming to avoid FFRW oversensing, will degrade performance of these algorithms.8588 ICDs with long PVAB periods or low atrial sensitivity may fail to switch modes if a substantial fraction of atrial events are undersensed.

image

Figure 3-45 Algorithm to prevent upper rate limit tracking of alternate atrial EGMs during atrial flutter (Blanked Flutter Search), even if EGMs recorded in postventricular atrial blanking period (PVAB), when sensing cannot occur.

Upper panel, Simulated electrocardiogram (ECG) and ladder diagram. Black horizontal bars denote blanking periods, and white bars denote refractory periods for atrium (upper bars) and ventricle (lower bars). Lower panel, Stored ventricular EGM and dual-chamber EGM markers from a clinical event. Upper panel, The intention of this algorithm is to infer that alternate atrial EGMs are “hiding” in the PVAB. To determine whether this is occurring, the algorithm transiently interrupts atrial tracking to assess the atrial rhythm when it is likely that alternate atrial intervals are occurring in the PVAB. The specific criterion is that the measured A-A interval is both less than twice the sum of the sensed AV interval and the PVAB and less than the mode switch interval for 8 beats. When this occurs, the algorithm extends the postventricular atrial refractory period (PVARP extension). The next atrial EGM that would have been tracked is now in the PVARP (AR marker, first arrow) and is not tracked. The next atrial flutter EGM is sensed because it no longer times in the PVAB (AS marker, second arrow). The pacemaker then switches mode rapidly (DDI). Lower panel, In this stored clinical EGM, ventricular intervals are numbered to correspond to those in the simulation in the upper panel. As in the upper panel, the first arrow (AR) corresponds to PVARP extension, and the second arrow (AS) corresponds to a sensed atrial flutter EGM that would have been blanked in tracking mode. Mode switching follows immediately. VP, Ventricular paced event.

Algorithms to Reduce Ventricular Pacing

Unnecessary ventricular pacing contributes to adverse clinical outcomes (e.g., heart failure hospitalizations) and increased risk of AF.89,90 Algorithms that lengthen AV intervals adaptively to promote intrinsic conduction reduce ventricular pacing.9194 Medtronic and Sorin developed new pacing modes (MVP and SafeR) that operate in AAI(R) mode, with mode switching to DDD(R) to prevent excessive ventricular pauses or asystole.19,95,96 Single beats of AV conduction failure are allowed by both algorithms, with switching to DDD mode occurring only if higher levels of AV conduction block occurs (Medtronic) or if the PR interval is excessive. Once switched to DDD mode, both algorithms test for AV conduction periodically, to avoid unnecessary ventricular pacing for transient AV block. However, these pacing modes may be associated with pacemaker syndrome in patients with marked first-degree AV block, resulting in unnecessary pauses if complete heart block develops.9799

Automatic Assessment of Pacemaker Capture (“Capture Management”)

Automatic sensing of the evoked response (local myocardial depolarization) that results from a pacing stimulus allows for verification of myocardial capture. However, sensing of the evoked response is obscured by the decaying polarization afterpotential that follows the pacing stimulus.100 This polarization artifact can be attenuated by using high-capacitance, low-polarization leads or using multiphasic pacing impulses to neutralize the postpacing polarization. Figure 3-46 shows an example of the evoked response signal combined with polarization artifact after a pacing stimulus.100 To minimize the afterpotential from ventricular pacing stimuli and improve detection of the evoked response, some manufacturers use a smaller-output capacitor. This approach further enhances the accuracy of evoked response detection.

Pacemakers use one of two different ventricular capture verification schemes: beat-to-beat approaches that monitor capture almost continuously and periodic approaches that test pacing thresholds intermittently (e.g., once every few hours). Algorithms that perform periodic capture threshold measurements typically set the pacing output at 1.5 to 2.0 times the threshold to avoid transient loss of capture until the next test. Beat-to-beat capture management algorithms adjust pacing outputs to values that are just above pacing threshold but deliver higher-output backup pulses if loss of capture occurs.101

Ventricular capture verification has been incorporated into some ICDs and resynchronization ICDs. Algorithms for RV capture management use essentially the same methods as pacemakers, except that polarization artifact that challenges evoked response detection in bipolar pacing leads is eliminated or reduced by use of the RV coil-to-can EGM available in ICD systems.102 LV capture management may be achieved by pacing on the LV lead and analyzing the intraventricular conduction time sensed in the right ventricle. This method has been demonstrated to be safe, reliable, and clinically equivalent to operator-performed LV threshold tests.103

Atrial capture management is more challenging than ventricular management because the atrial evoked response is much smaller than the ventricular response. Thus, reliable detection of the evoked response is difficult, and atrial capture detection using the evoked response method can be performed only using low-polarization electrodes.104 An alternative method relies on critical timing of the atrial and ventricular response to atrial pacing stimuli during normal sinus rhythm (Fig. 3-47). In ambulatory patients, this method has been reported to perform comparably with operator-performed atrial pacing threshold measurements without causing atrial proarrhythmia.105

image

Figure 3-47 Atrial capture verification method that relies on atrial or ventricular responses to atrial pacing.

A high-level control algorithm determines which of two timing assessment methods should be used, based on the recent history of atrial pacing, atrial sensing, and atrioventricular (AV) conduction. At least two test paces are required to verify capture or loss of capture. Failed, inconclusive, or aborted threshold tests are repeated after a delay. This method cannot measure atrial capture during atrial paced–ventricular paced (VP) rhythms or rhythms with irregular atrial or ventricular cycle lengths. A, Atrial Chamber Reset method is used if recent history indicates primarily atrial sensed events (AS). A premature atrial test pace (APt) is delivered, and the resulting atrial timing is observed. An atrial refractory event (AR) within 10 bpm after the prior intrinsic atrial rate suggests loss of capture (top panel). The lack of an atrial refractory event in this period suggests resetting of the atrial timing and atrial capture (middle and bottom panels). B, The AV conduction check method is used if the recent history indicates that AV conduction is present. The average AV conduction time is measured before delivery of the APt. A backup atrial pace at the operating atrial output is also delivered 70 msec after the APt. The algorithm looks for ventricular sensed events (VS) in one of two capture verification windows (40 msec in duration) that are placed relative to the APt and the backup atrial paced events based on the average AV conduction time. A ventricular event sensed in the first capture verification window suggests capture by the APt (top panel). Lack of a sensed ventricular event in the first capture verification window and presence of a ventricular event in the second window is evidence of noncapture by the APt (bottom panel).

image Basics of Detection of Ventricular Tachycardia/Fibrillation

Figure 3-48 provides an overview of the structure of VT/VF detection algorithms. Initial detection of tachycardia is based on rate and duration. The minimum duration of tachycardia required for detection is programmable, either directly (in seconds) or indirectly by setting the number of ventricular intervals required. A tachycardia episode begins when the minimum rate and duration criteria are satisfied. Computationally intensive features of detection algorithms, such as morphology analysis, are activated only after initial detection occurs, to conserve power.

Ventricular Rate and Counting Methods

The specific methods used to count ventricular intervals vary and influence the sensitivity and specificity of VT detection. Initial detection in the VF zone must be tolerant of undersensing. Detection usually occurs when a certain percentage (typically 70%-80%) of intervals in a sliding window (usually 10-40 intervals) fulfill the programmed rate criterion. Most ICD manufacturers also use this “X out of Y” counting in the slower (VT) zones, although Medtronic requires that all intervals exceed the rate criterion to detect VT. This consecutive-interval method diminishes inappropriate detection of AF without compromising sensitivity for detection of regular, monomorphic VT.106108 St. Jude ICDs classify intervals based both on the last interval and on the average of the last four intervals (Figs. 3-49 and 3-50). A potential advantage of this method is that averaging minimizes the effects of intermittent undersensing by providing a “smoothing effect.”106 A disadvantage is that averaging exaggerates the effect of extremely short, oversensed intervals or extremely long, undersensed intervals. To satisfy tachycardia criteria, “binned” intervals do not need to be consecutive. As long as sinus rhythm is not redetected, the counter will continue to increment. This may be particularly problematic when the slowest VT zone is used as a monitoring zone: During sinus tachycardia in the monitoring zone, PVCs increment the VT counter, but the counter is not reset until the sinus rate decreases out of the monitoring zone, providing the potential for inappropriate therapy of sinus tachycardia109 (Fig. 3-51). Sorin ICDs first classify the rhythm on the basis of the ventricular cycle length. This algorithm relies on the “majority criterion,” which excludes the initial two, short ventricular cycles from analysis to withdraw cycles that may have unstable R-R intervals at the beginning of an arrhythmia.

image

Figure 3-50 Illustration of interval classification based on current interval and average interval; T-wave oversensing initiates R-wave double-counting in St. Jude ICD.

The figure shows continuous recording of atrial EGM, event markers, and extended bipolar ventricular EGM (between Y-adapted left ventricular [LV] and right ventricular [RV] tip electrodes). The ICD is programmed with a single ventricular tachycardia (Tach) zone at 345 msec and a ventricular fibrillation (Fib) zone at 290 msec. The top panel shows atrial sensed, ventricular paced rhythm. In the second panel, the first ventricular EGM is an interpolated premature ventricular complex (PVC). Ventricular pacing is inhibited after the next P wave because the resultant PVC-V paced interval would have been less than the interval corresponding to the programmed upper rate limit (~450 msec). This results in a conducted R wave that is double-counted at the ventricular blanking period of 125 msec, but this interval is not classified in the VF zone because the interval average of (719 + 402 + 367 + 125)/4 = 403 msec exceeds the Tach detection interval. The third and fifth beats show postpacing T-wave oversensing (down arrows). The interval ending with the first oversensed T wave is classified in the Tach zone because the interval average is less than 345 msec: (367 + 125 + 516 + 344)/4 = 338 msec. The next ventricular interval, corresponding to the subsequent double-counted R wave, is classified in the Fib zone because the interval average is in the Tach zone ([516 + 344 + 371 + 125]/4 = 339 msec) but the current interval is in the Fib zone. Mode switching to the DDI pacing mode occurs after the counter of Tach + Fib beats equal 4 (DDI, asterisk). The third and fourth panels show conducted sinus rhythm after mode switching, with alternate intervals classified in the Fib zone. VF is detected when the Fib counter reaches 12. This counter increments for each interval classified in the Fib zone and is reset to zero by five consecutive sinus intervals. The count of Fib intervals reaches 12 after the third interval in the fourth panel, and Fib is detected (Trigger, D marker), resulting in capacitor charging. However, sinus slowing results in an increase in the average interval, so that the interval average is no longer in the Tach zone. This occurs at the second R wave in the fifth panel (S marker, double asterisk; interval average = [125 + 570 + 125 + 596]/4 = 354 msec). Once this occurs, alternate detected intervals are classified as “Sinus” or “Not Binned.” The shock is aborted when the count of consecutive binned Sinus intervals reaches 5 at the beginning of the bottom panel (RS = return to sinus, up arrow), and this is followed by mode switch to DDD pacing. Note that St. Jude ICDs have a “Bigeminal Avoidance” counter that prevents bigeminal rhythms from being classified as Tach. This counter does not apply in the Fib zone. R, Ventricular sensed event; V, ventricular paced event.

Ventricular Tachycardia/Fibrillation Rate Detection Zones

Tiered-therapy ICDs have up to three ventricular rate detection zones that permit programming of zone-specific therapies and SVT-VT discriminators (Fig. 3-52). The duration of a rapid ventricular rate required for detection is programmable independently in at least the slowest and fastest zone (see Zones and Zone Boundaries). The two slower zones are classified as VT zones, and the fastest one as a VF zone. All ICDs use strategies to minimize the risk of prolonged detection if successive VT intervals are classified in different zones. Except for Medtronic ICDs, sensed events in one zone increment counters in all slower zones. To avoid delays in detecting rhythms that straddle zone boundaries, Medtronic ICDs utilize a combined-count (summation) feature, which is satisfied if a sufficient fraction of intervals is in either zone.106

image Discriminators

The SVT-VT discriminators are a programmable subset of an ICD’s detection algorithm for VT/VF. Discriminators withhold ventricular antitachycardia pacing (ATP) or shocks if SVT is diagnosed. These algorithms differ from the SVT detection algorithms used to switch modes during bradycardia pacing or deliver atrial therapy for AF or atrial flutter.

Confirmation, Redetection, and Episode Termination

If a tachycardia is classified as ventricular, ATP may be delivered immediately; but shocks require capacitor charging, which takes 6 to 15 seconds for maximum-energy shocks. The first shock in a shock sequence is “noncommitted,” meaning that it may be aborted during charging if the detection algorithm identifies termination of the arrhythmia during or immediately after charging of the high-energy capacitors (Fig. 3-53). In contrast, a “committed shock” is always delivered if the capacitor charges. Confirmation or reconfirmation is the brief process that occurs after charging by which ICDs determine whether to deliver or abort the first shock in a sequence. In all ICDs, the confirmation algorithm delivers the stored shock if a few intervals immediately after charge completion are shorter than the programmed VT interval (St. Jude, Boston Scientific, Sorin, Biotronik) or are within 60 msec of the VT interval (Medtronic ICDs before Protecta) (Fig. 3-54). Therefore, the first VF shock is effectively “committed” if the VT interval (or, in some ICDs, the monitor-only interval) is programmed to a long cycle length.111,112 A new shock confirmation algorithm in Medtronic ICDs (investigational, Protecta) aborts a defibrillation shock when ventricular cycle lengths increase by at least 60 msec relative to the detected tachycardia cycle length and exceed the VF detection interval, and the ventricular cycle length at the point of detection is regular. Tachycardias detected in the VF zone that do not have regular cycle length are judged against the VF interval plus 60 msec. Shocks for VT will abort on a 60-msec increase in cycle length relative to the point in detection, regardless of cycle length regularity.

Once a shock is aborted, the device continues to apply tachycardia redetection and episode termination criteria. Depending on the manufacturer and programming, each shock after the first shock may be either noncommitted or committed. A shock is also committed if VT or VF is detected after a diverted shock and before episode termination.83

Redetection is the process by which ICDs determine whether VT or VF detection criteria remain satisfied after therapy is delivered. The duration for redetection is programmable independently of the duration for initial detection and typically is shorter than that for initial detection. ICD-defined arrhythmia episodes continue after each tiered therapy until either a tachyarrhythmia is redetected to initiate the next therapy or the rhythm is classified as normal (“sinus”), resulting in episode termination. In most ICDs, episode termination is based on (slow) rate and duration, with 3 to 8 beats classified as sinus. The number of intervals required for episode termination is programmable in St. Jude ICDs. Programming a short duration is advisable when ATP terminates VT, but it reinitiates promptly after termination. This permits repeating the first sequence of programmed ATP rather than escalating to subsequent therapies that include shocks.

image SVT-VT Discrimination in Ventricular ICDs

Building Blocks for Discrimination

The ICDs analyze information derived from a sequence of sensed atrial and ventricular EGMs to detect VT/VF and discriminate ventricular arrhythmias from SVT. These algorithms operate in a stepwise manner, using a series of physiologically relevant, logical “building blocks” (Table 3-1) based on the timing relationships and morphology of sensed EGMs. We focus on the “tool kit” provided by these building blocks rather than on the specific details of proprietary algorithms, reviewed elsewhere in detail.113 Detection algorithms combine complementary building blocks to form the final detection decision. Each building block has advantages and limitations. Some are redundant, and some interact. Interactions vary depending on the order in which blocks are applied. Clinicians and algorithm designers must consider these issues to understand trade-offs in clinical performance.

TABLE 3-1 Tachyarrhythmia Detection Building Blocks

Tachyarrhythmia Detection Building Blocks Purpose/Information Potential Weaknesses
Single-Chamber Ventricular Building Blocks
R-R interval + duration Identifies sustained high ventricular rates SVT with high ventricular rates that overlap with VT/VF rates
R-R regularity Discrimination of monomorphic VT (regular cycle lengths) from rapid AF (irregular cycle lengths) May lose effectiveness as ventricular rates during AF increase; 2 : 1 atrial flutter has regular R-R intervals; may cause underdetection of VT with irregular R-R intervals
R-R onset Identifies sudden ventricular rate changes Not specific for atrial or ventricular tachyarrhythmias; may miss VT arising during sinus tachycardia
VEGM morphology Abnormal VEGM morphology may indicate ventricular tachyarrhythmias Confounded by conduction aberrancy or changes in “normal” VEGM morphology
Burst ventricular pacing Intervals after entrainment of VT by burst pacing are less variable than intervals after burst pacing during SVT Sensitive to single interval measurement, potential detection time delay, and potential proarrhythmia
Key Dual-Chamber Building Block
Comparison of atrial vs. ventricular rate VT diagnosed if atrial rate is less than ventricular rate Confounded by atrial undersensing or far-field R-wave oversensing
Dual-Chamber Building Blocks
P-R dissociation P-R dissociation usually indicates VT AV reentrant tachycardia; VT with 1 : 1 retrograde conduction; AF that conducts rapidly with apparent P-R dissociation
P-R patterns/relationships Consistent P-R patterns/relationships usually indicate SVT AV reentrant tachycardia and VT with 1 : 1 retrograde conduction
Chamber of acceleration Identifies whether tachycardia initiates in atrium or ventricle A single oversensed/undersensed event may result in misclassification.
Atrial or ventricular pacing, response in opposite chamber Discrimination of 1 : 1 rhythms using ventricular response to atrial extrastimuli Primarily aids diagnosis for 1 : 1 rhythms; concerns for VT detection delay and proarrhythmia
Single-Chamber Atrial Building Blocks
P-P intervals Identifies high atrial rates High atrial rates may be present during true VT/VF.
P-P regularity Regular atrial rate indicates organized atrial activity Minimal benefit for ventricular tachyarrhythmia characterization
P-P onset Identifies sudden atrial rate changes Not specific for atrial or ventricular tachyarrhythmias (e.g., VT with 1 : 1 retrograde association)
AEGM morphology Identifies atrial tachyarrhythmias and/or retrograde conduction Confounded by far-field R waves and changes in “normal” AEGM morphology

AEGM, Atrial electrogram; AF, atrial fibrillation; AV, atrioventricular; SVT, supraventricular tachycardia; VEGM, ventricular electrogram VF, ventricular fibrillation; VT, ventricular tachycardia.

Single-Chamber Ventricular Building Blocks

The first five building blocks in Table 3-1 apply to single-chamber ICDs or the ventricular component of dual-chamber ICDs. The R-R interval + duration analysis building block classifies ventricular intervals into zones by programmable cycle length (or rate) thresholds. All single- and dual-chamber ICDs use cycle length or rate in combination with tachyarrhythmia duration as the most basic method of detection. When used alone, R-R interval + duration analysis provides the highest sensitivity for detection of VT/VF, but cannot discriminate VT/VF from SVT. The R-R regularity building block is used to discriminate between regular ventricular intervals caused by monomorphic VT and irregular ventricular intervals caused by rapidly conducted AF. The R-R onset building block is used to discriminate sudden-onset VT from gradual-onset sinus tachycardia. The VEGM morphology building block classifies tachyarrhythmias as SVT if morphologic characteristics of the ventricular EGM are similar to those of beats of known supraventricular origin. Otherwise, it classifies the rhythm as VT.

Dual-Chamber Building Block: Atrial vs. Ventricular Rate

Direct or indirect comparison of atrial and ventricular rates is the cornerstone of most dual-chamber algorithms. In most studies, the ventricular rate exceeds the atrial rate in more than 80% of VTs in the VT zone of dual-chamber ICDs and a higher fraction of VTs in the VF zone.68,114 Thus, algorithms that compare atrial and ventricular rates as their first step (Boston Scientific Rhythm ID, St. Jude, Biotronik) apply SVT discriminators to fewer than 20% of VTs; this reduces the risk that they will misclassify VT as SVT, provided that the atrial rate is measured correctly. The Medtronic algorithm indirectly compares atrial and ventricular rates using dual-chamber pattern analysis as the first step to achieve a comparable result. The Sorin algorithm uses a conceptually different approach, by initially detecting tachyarrhythmia based on cycle length, then further classifying the arrhythmia as VT or SVT based on the stability of R-R intervals and an analysis of AV association and tachycardia onset.

Dual-Chamber Building Blocks

The next four building blocks in Table 3-1 are based on combinations of atrial and ventricular information. Their primary role is to discriminate VT with 1 : 1 ventricular-atrial (V-A, VA) conduction from SVT. Their secondary roles are to classify rhythms with stable N:1 atrial-ventricular (A-V, AV) relationships (stable ratio of N atrial events for each ventricular event) as SVT and to classify isorhythmic tachycardias (with AV dissociation and similar ventricular rate) as VT. The P-R dissociation building block can detect the presence of VT during SVT. The P-R patterns/relationships building block applies to stable associations of atrial and ventricular events. For tachycardias with 1 : 1 AV relationships, it analyzes the relative timing of P-R and R-P intervals to classify the rhythm as SVT or VT. It also identifies consistent N:1 AV patterns that occur primarily during atrial flutter. The chamber of origin building block discriminates between VT and SVT with 1 : 1 P-R association by identifying whether the tachycardia originates in the atrium or in the ventricle. In ICD patients, SVT usually begins with an intrinsic atrial event in the interval between the last ventricular event in the sinus rate zone and the first ventricular event in the VT zone. Conversely, at the start of spontaneous VT, there is usually no atrial event in this interval. The response of the opposite chamber to atrial or ventricular pacing has long been a critical tool in the electrophysiology laboratory for diagnosing tachycardias with 1 : 1 AV relationship. A novel ICD algorithm based on this premise has been shown to terminate or discriminate 1 : 1 tachycardias by using simultaneous atrial and ventricular ATP.115

SVT-VT Discrimination in Single-Chamber ICDs

Sudden Onset

Measures of abruptness of onset of a tachycardia have high specificity for rejecting sinus tachycardia,116,118120 but may prevent detection of VT that originates during SVT or VT that starts abruptly with an initial rate below the VT detection limit. In the latter case, the ICD misclassifies the “onset” of the arrhythmia as the gradual acceleration of the VT rate across the VT rate boundary. This criterion does not prevent SVT with sudden onset from being treated inappropriately.

Ventricular EGM Morphology

Analysis of ventricular EGM morphology,114,120123 alone or in combination with stability, probably provides the best single-chamber SVT-VT discrimination for initial detection of VT. The morphology building block is the key element of single-chamber algorithms and the central single-chamber component of dual-chamber algorithms with this feature. For this reason, a more detailed analysis is provided.

Ventricular Egm Morphology for SVT-VT Discrimination

General

Electrogram morphology algorithms are the most complex and effective building blocks in single-chamber detection algorithms. All morphology algorithms share common steps (Fig. 3-55): (1) record a template EGM of baseline rhythm; (2) construct and store a quantitative representation of this template; (3) record EGMs from an unknown tachycardia; (4) time-align the template and tachycardia EGMs; (5) construct a quantitative, normalized representation of each tachycardia EGM; (6) compare the representation of each tachycardia EGM with that of the template to determine its degree of morphologic similarity; (7) classify each tachycardia EGM as a morphology match or nonmatch with the template; and (8) classify the tachycardia rhythm as VT or SVT based on the fraction of EGMs that match the template. Steps 3 through 8 are performed in real time. Morphology algorithms differ according to EGM source or sources, methods of filtering and alignment, and details of quantitative representations. The features of specific algorithms are described in Figure 3-56.

image

Figure 3-56 Specific morphology algorithms.

A, St. Jude MD algorithm. The positive and negative deflections in the sensing EGM are normalized and modeled as a series of three polygons (A, B, and C in the template EGM; A’, B’, and C’ in the tachycardia EGM). The normalized areas of these polygons are then computed. Each tachycardia EGM is compared with the template EGM in three steps. First, the difference in area of corresponding polygons is computed. Second, the absolute values of these differences are summed. Third, a match score is constructed to be inversely proportional to the sum of these differences. If a programmable number (= 4) of 8 complexes in a sliding window exceed the programmable threshold (nominally 60%), the rhythm is classified as supraventricular tachycardia (SVT). If not, it is classified as ventricular tachycardia (VT). B, Boston Scientific Vector Timing and Correlation (VTC) algorithm. This algorithm aligns shocking EGMs of the tachycardia and template based on the peak of the rate-sensing EGM. This method takes advantage of spatiotemporal differences between activation sequences in VT and baseline rhythm that cannot be detected from a single EGM. In this example, the peak of the shock EGM has similar timing to the peak of the rate-sensing EGM in sinus rhythm, but much later timing in VT. The amplitude of the shocking EGM (upper panel) is computed at each of 8 points selected on the basis of their timing relative to the peak of the sensing EGM (lower panel) and extracted as an eight-element “feature set,” corresponding to an eight-dimensional vector. Feature sets of each tachycardia EGM are compared with the template EGM by computing the product-moment correlation coefficient between the two feature sets. This value, named the feature correlation coefficient (FCC), is a measure of mean spatiotemporal difference in ventricular activation between tachycardia and baseline rhythm. A threshold value for the feature correlation coefficient was selected to optimize SVT-VT discrimination on a test data set. This value is neither published nor programmable. If at least 3 of 10 complexes in a sliding window exceed the threshold, the rhythm is classified as SVT; if not, it is classified as VT. C, Medtronic Wavelet algorithm. The algorithm expresses the morphology of ventricular EGMs using the wavelet-transform. Wavelets are functions of constant shape and limited time duration.* They form the basis of a mathematical transformation that represents signals efficiently if they are both highly localized in time and preceded and followed by isoelectric intervals. Wavelets are therefore well suited for representing transient biomedical signals such as ventricular EGMs. The algorithm compares the morphology of ventricular EGMs during a tachycardia with a template recorded during baseline rhythm. This comparison is expressed as a percent-match score that describes the degree of morphologic similarity of the baseline and tachycardia EGMs using a programmable source. In general, the EGM recorded between right ventricular and superior vena cava coils is preferred as a default. This signal combines far-field sensitivity to morphology differences between VT and SVT with resistance to pectoral myopotentials. The algorithm begins processing when a tachycardia fulfills the programmed rate criterion for detection of VT and 8 beats remain to fulfill the programmed duration criterion for detection of VT. Percent-match scores are calculated for each of the last 8 beats before detection. Beats with match scores less than a programmable threshold (nominally 70%) are classified as ventricular. Nominally, a tachycardia is classified as VT if 6 or more of the 8 analyzed EGMs are classified as ventricular. Otherwise, it is classified initially as SVT. If the tachycardia is classified as SVT, the algorithm is applied to each successive 8-beat sliding window until the rate criterion for VT is no longer fulfilled.

(*Meyer Y: Wavelets: algorithms and applications. Philadelphia, 1993, Society for Industrial and Applied Mathematics.)

Limitations of Morphology Algorithms

The START study (SAFARI trial) of arrhythmias recorded at ICD implant reported individual differences for the performance of single-chamber discrimination algorithms based primarily on EGM morphology.124 Morphology algorithms share common failure modes for inappropriate detection of SVT as VT: (1) inaccurate template, (2) EGM truncation, (3) alignment errors, (4) oversensing of pectoral myopotentials, (5) rate-related aberrancy, (6) SVT soon after shocks, and (7) inappropriate classification of VT as SVT.

Electrogram Truncation

EGM truncation (“clipping”) occurs when the recorded EGM signal amplitude exceeds the range of the EGM amplifier so that the maximum or minimum portion of the EGM is clipped. This removes EGM features for analysis and alters the timing of the tallest peak, which can affect alignment. The amplitude scale in Medtronic and St. Jude ICDs should be adjusted so that the EGM used for morphology analysis is 25% to 75% of the dynamic range (Fig. 3-57).

Alignment Errors

Alignment errors prevent match between a tachycardia EGM and a morphologically similar stored EGM. Mechanisms depend on the method used for EGM alignment (Fig. 3-58). Accurate alignment in the St. Jude algorithm is sensitive to the value of the sensing threshold at the onset of the ventricular EGM, as determined by Automatic Sensitivity Control. If a template EGM is acquired at the most sensitive setting of Automatic Sensitivity Control (either because of a slow sinus rate or after a ventricular paced beat), a low-amplitude peak at the onset of the ventricular EGM may be used for alignment. During tachycardia, the next R wave may occur before Automatic Sensitivity Control reaches its sensitive value. In this case, the low-amplitude peak at the onset of the ventricular EGM may not be used for alignment. If identical template and tachycardia EGMs are then compared, their representations in the morphology algorithm will not match. Usually, they are assigned morphology match scores of 0% (see Fig. 3-58). In patients who have dual-chamber ICDs and intact AV conduction, the most reliable template is acquired or verified during atrial pacing at a rate close to the VT rate. In single-chamber ICDs, options may include acquiring or verifying the template during exercise testing and altering the minimum sensitivity, Threshold Start, or Threshold Delay. If a template is acquired during atrial pacing or exercise testing, automatic template updating should be disabled to prevent overwriting the template with one acquired in baseline rhythm.

Medtronic ICDs align EGMs based on their tallest (positive or negative) peaks. If an EGM has two peaks of near-equal amplitude, or if such peaks are caused artificially by truncation of large EGMs that exceed the programmed dynamic range, minor variations in their relative amplitudes may result in an alignment error.120 An alternative EGM source should be selected.

The Boston Scientific morphology algorithm (Rhythm ID) simultaneously collects data from two different channels: the local bipolar rate-sense EGM and shock EGM.125,126 It considers the vector of depolarization wavefront (with shock EGM as the reference), the timing of the peak signal recorded on the rate-sense EGM, and alignment of eight characteristic points of the shock EGM. A correlation equation aligns peaks on the rate-sense EGM for each QRS complex of the unknown rhythm with the rate-sense EGM peak of the sinus template, then compares the eight captured points with the template.

Rate-Related Aberrancy

If complete bundle branch aberrancy occurs reproducibly, the template may be recorded during rapid atrial pacing (Fig. 3-59). However, subtle and varying aberrancy can confound templates acquired during baseline rhythm (Fig. 3-60). If a template is acquired at a fast rate, automatic template updating should be deactivated to prevent subsequent auto-acquisition of a slow, baseline template without aberrancy. During rapidly conducted AF, subtle degrees of aberration often distort the terminal portion of the EGM sufficiently that the percent match is less than the nominal threshold. In St. Jude ICDs, reducing the fraction of EGMs required to exceed the match threshold from 5 of 8 beats to 3 or 4 of 8 beats may reduce this problem without compromising detection of monomorphic VT. Reducing the match percent required to exceed the match threshold may also prevent misclassification of aberrantly conducted SVT as VT, but it results in a greater chance of misclassifying monomorphic VT as SVT than does reducing the fraction of EGMs.

image

Figure 3-60 Error in morphology algorithm caused by subtle rate-related aberrancy.

Top panel and middle panel, Dual-chamber stored EGM from inappropriately treated sinus tachycardia. Right atrial (RA) EGMs, event markers, and true bipolar right ventricular (RV) EGMs are shown. Values immediately below event markers, corresponding to EGMs in ventricular tachycardia (VT) zone (labeled “T”), indicate the percent match between EGM morphology in VT and sinus morphology (seen in lower left panel). Corresponding “X” labels above event markers indicate that the morphology algorithm classifies the beat as VT because the match is less than 60%. Burst antitachycardia pacing (ATP) is delivered at the right of the first panel. Dotted arrow indicates that panels are not continuous. After several trials of ATP (not shown), a shock is delivered toward the end of the second panel (HV). “Trigger” at right of upper panel indicates (inappropriate) detection of VT. “D =” at right of upper panel indicates that the atrial rate is equal to the ventricular rate. S denotes intervals in the sinus zone above the VT detection interval of 360 msec. Time line is in seconds. Bottom panels show real-time programmer strips. Bottom left panel, 100% template match in sinus rhythm (arrows) and 0% match on premature ventricular complexes (PVCs). However, the bottom right panel, recorded during atrial pacing at a cycle length of 400 msec, shows only two EGMs with adequate match (check marks, arrows). Note that the surface electrocardiogram (ECG) does not show identifiable aberrancy despite sufficient changes in EGM to prevent template match.

Supraventricular Tachycardia Soon after Shock

After a shock, ICD detection algorithms reclassify the rhythm as “sinus” and revert to their initial detection mode within a few seconds, but postshock distortion of EGM morphology may persist for 30 seconds to several minutes.60,120 No SVT-VT discrimination algorithm relies on EGM morphology after shocks until the arrhythmia episode terminates. However, algorithms revert to their initial detection mode within a few seconds after a shock. If postshock SVT starts after the rhythm has been classified as sinus but before postshock EGM distortion dissipates, any morphology algorithm may misclassify SVT as VT.120 Therefore, distortion of the EGM after a shock could result in a repetitive sequence of inappropriate classification of SVT as VT, inappropriate shocks for SVT, and perpetuation of postshock EGM changes in SVT by each successive shock.

Inappropriate Classification of Ventricular Tachycardia as Supraventricular Tachycardia

The St. Jude morphology algorithm, which analyzes only the rate-sensing electrode, continuously misclassifies up to 6% of VTs as SVT (Fig. 3-61). However, if it is restricted to tachycardias (with ventricular rate ≤ atrial rate in dual-chamber ICDs), only 2% of VTs are misclassified.114 The Medtronic morphology algorithm, which usually analyzes high-voltage EGMs, misclassifies about 1% of VTs as SVT.127 If misclassification occurs, an alternative EGM source may provide adequate discrimination. Limited data indicate that the corresponding error rate for the Boston Scientific algorithm is comparable or lower. When the St. Jude morphology algorithm is used in a single-chamber ICD, the sustained-duration time-out (Maximum Time to Diagnosis, Section V.F) should be programmed unless the algorithm is known to classify all clinical VTs correctly. Opinions differ about whether comparable features are required for Medtronic and Boston Scientific algorithms.

SVT-VT Discrimination in Dual-Chamber and Cardiac Resynchronization ICDs

The integration of dual-chamber building blocks into detection algorithms may be considered in terms of the relative rates of the atrium and ventricle (Fig. 3-62).

Operation for Atrial Rate Equal to Ventricular Rate

The vast majority of tachycardias with 1 : 1 AV relationship are SVT, primarily sinus tachycardia. VT with 1 : 1 VA conduction accounts for less than 10% of VTs detected by ICDs. The building blocks used by each manufacturer for distinguishing 1 : 1 AV conduction of SVT from 1 : 1 VA conduction are summarized in Table 3-2. A dual-chamber onset rule that evaluates both P-R and R-R interval onset improves specificity of R-R interval onset-type building blocks for sinus tachycardia with minimal loss of sensitivity for VT.128 Sorin ICDs classify the 1 : 1 tachycardias based on “chamber of acceleration”: the first accelerated cycle is classified as “supraventricular” if two ventricular events are classified as “conducted,” if the R-R interval is greater than 75% of the preceding P-P interval. If the two ventricular events defining the first accelerated interval are ‘ ‘conducted,’ ’ the origin of the acceleration is declared “atrial.” Conversely, if the first accelerated beat is nonconducted, it is considered ventricular in origin. Individual examples are shown in Figures 3-63 to 3-66.

image

Figure 3-64 A, Discrimination of 1 : 1 tachycardias by P-R pattern using the original sinus tachycardia criterion in Medtronic PR Logic (GEM III, Marquis, Maximo, InSync Marquis, InSync Maximo) ICDs. PR Logic uses patterns of A-V, V-A, V-V, and A-A intervals from consecutive ventricular events to form couple codes describing supraventricular tachycardias (SVTs). These couple codes are based on the number of atrial (A) events within the ventricular (V) interval and their relative timing. Discrimination of sinus tachycardia (ST) from ventricular tachycardia (VT) with 1 : 1 retrograde conduction is critically dependent on the programmable 1 : 1 VT-ST boundary, which is defined nominally as 50% of the current R-R interval (with optional values of 35%, 66%, 75%, and 85%). P-R intervals less than the defined percentage of the current R-R interval and greater than 70 msec are considered antegrade (left panel), and P-R intervals equal to or greater than that percentage of the R-R interval and also greater than 40 msec are considered retrograde (right panel). Rhythms in the detection zone with 1 : 1 pattern and antegrade P-R intervals are classified as ST, and therapies are withheld. Rhythms with 1 : 1 pattern and retrograde P-R intervals are classified as VT. B, New adaptive ST rule in PR Logic (Medtronic Entrust DR ICD) no longer uses 1 : 1 VT-ST boundary to discriminate ST from VT with 1 : 1 retrograde conduction. Rather, it uses a combination of pattern, sudden R-R onset, and sudden P-R onset. Operation of the new adaptive ST criterion is illustrated by this example of a spontaneous episode of ST that converted into VT with 1 : 1 retrograde conduction. During the ST, the R-R and P-R intervals occurred within their expected ranges. After initiation of VT, the R-R and P-R intervals occurred outside their expected ranges. Evidence of ST was lost by the third beat of VT. The sudden change in R-R intervals was small (460 msec), so the R-R intervals during the VT occurred within twice the R-R interval expected range, and adaptation of the RR interval expected range continued. Eventually, the R-R interval expected range adapts to accept the R-R intervals of the VT. Conversely, the P-R intervals occurred outside of twice the P-R expected range, and adaptation was inhibited. VT is appropriately detected after 16 consecutive R-R intervals in the VT detection zone despite gradual R-R onset of a 1 : 1 rhythm, because the P-R intervals had sudden onset and were no longer within the expected P-R interval range. VF, Ventricular fibrillation.

(From Stadler RW, Gunderson BD, Gillberg JM: An adaptive interval-based algorithm for withholding ICD therapy during sinus tachycardia. Pacing Clin Electrophysiol 26:1189-1201, 2003.)

As in the electrophysiology laboratory, pacing methods may also be used to discriminate 1 : 1 tachycardias (see Active Discrimination).

Operation for Atrial Rate Greater Than Ventricular Rate

The building blocks used by each manufacturer for distinguishing rapidly conducted AF and atrial flutter from VT during atrial arrhythmias are summarized in Table 3-3. Although various combinations of these building blocks successfully discriminate VT from SVT in VT zones, most VT during AF is sufficiently rapid to be detected in the VF zone,129 where some discriminators lose specificity or may not be applied at all. For example, R-R regularity cannot be applied in the VF zone because polymorphic VT has irregular R-R intervals, and undersensing of VF exaggerates the irregularity of measured R-R intervals. AV dissociation occurs uniformly during rapidly conducted AF,68 and morphology templates acquired during sinus rhythm often misclassify aberrantly conducted beats. Individual examples are shown in Figures 3-67 to 3-69.

Specific Algorithms

Figures 3-70 to 3-74 show block diagrams of the principal dual-chamber detection algorithms in clinical use today. Legends identify unique features of each algorithm.

image

Figure 3-73 Medtronic PR Logic algorithm.

A, Clinical decision process. The rhythm types are represented using P-R marker diagram examples (top), from least complex (left) to most complex (right). A list of possible PR Logic decisions and the building blocks used are also shown for each of the rhythm types. Note that PR Logic implicitly employs a ventricular (V) rate > atrial (A) rate override (first column), because all rhythms with A rate < V rate and V rate in one of the detection zones are classified as ventricular tachycardia/fibrillation (VT/VF). For tachycardias with A rate = V rate (i.e., 1 : 1 tachycardias, center column), the original PR Logic uses P-R patterns and discriminates 1 : 1 rhythms with critical P-R interval timing zones (see Fig. 3-64, A). The new sinus tachycardia (ST) criterion in PR Logic (Entrust and later) uses the P-R pattern along with P-R and R-R sudden-onset criteria (see Fig. 3-64, B). For rhythms with A rate > V rate, PR Logic uses P-R patterns along with R-R regularity, P-R dissociation, and P-P regularity to ensure detection of double tachycardia (VT or VF during atrial fibrillation [AFib]) and to withhold therapy for 2 : 1 atrial flutter, rapid AFib, and ST with far-field R-wave oversensing (FFRW OS). In the newest Medtronic ICDs (Protecta, currently investigational), EGM morphology analysis (Wavelet) is applied for rhythms with A rate ≥ V rate that are identified by SVT by PR Logic pattern analysis alone. In addition, detection of double tachycardia in the Protecta devices requires that abnormal EGM morphology be present. Integration of Wavelet EGM morphology with PR Logic is designed to improve detection specificity compared to PR Logic interval analysis alone. B, PR Logic computational flow diagram. On each ventricular event, PR Logic processes the new P-R, R-P, P-P, and R-R patterns and timing information for the building blocks. If VT/fast VT (FVT) or VF rate detection criteria are satisfied, the ventricular rate override criterion is checked first. If the median R-R interval is less than the supraventricular tachycardia (SVT) limit, detection occurs through the single-chamber detection criteria without considering the PR Logic discrimination algorithm. If the median R-R interval is greater than the SVT limit, and if double tachycardia (VT/FVT/VF + SVT) is not detected, the three PR Logic criteria for identifying SVTs are tested in the order shown. If any one of the PR Logic SVT criteria is satisfied, inappropriate detection is avoided. If an SVT is not positively identified by PR Logic pattern analysis and the A rate ≥ V rate, EGM morphology analysis (Wavelet) is applied and classify the rhythm as SVT if the EGM morphology during tachycardia matches the template. If none of the SVT discrimination criteria is satisfied, VT/FVT/VF is detected when the R-R interval–based criterion is satisfied. If SVT is identified, the entire process repeats itself on each ventricular event until VT/FVT or VF is detected or the rhythm slows out of the ventricular rate detection zones.

Single-Chamber vs. Dual-Chamber Discriminators

Nominal programming of dual-chamber algorithms is safe.130,131 Dual-chamber stored EGMs provide higher diagnostic accuracy for troubleshooting than single-chamber stored EGMs. However, dual-chamber discriminators cannot be implemented without the complications inherent in atrial leads. Also, dual-chamber ICDs introduce unique risks for underdetection of VT caused by cross-chamber ventricular blanking after atrial pacing (see Intradevice Interactions). In addition, with early dual-chamber algorithms, optimal values of programmable parameters were not known; initial approaches to the problem of atrial blanking versus FFRW oversensing had limited success; and atrial sensing problems and specific design flaws degraded performance. Not surprisingly, clinical studies of early dual-chamber algorithms reported no benefit over single-chamber algorithms.67,68,132,133 More recent, randomized prospective studies demonstrate moderate superiority in SVT-VT discrimination for dual-chamber over single-chamber algorithms.134,135 However, no agreement exists as to when the incremental implant complexity, price, risk of atrial lead complications, and reduced longevity of a dual-chamber ICD are justified for SVT-VT discrimination alone. Generally, dual-chamber ICDs should be considered in patients who are likely to have rapidly conducted SVT, those in whom monomorphic VT and sinus tachycardia rates are likely to overlap, and those who will benefit from additional diagnostics for atrial arrhythmias.

Active Discrimination

Active discrimination represents a paradigm shift in the design of detection algorithms, from “diagnose before intervening” to “treat first; analyze only those tachycardias that persist after treatment.”

Although active discrimination presents the risk of proarrhythmia (Fig. 3-75), it can be valuable in discriminating tachycardias with 1 : 1 AV association (Figs. 3-76 and 3-77). One such algorithm was designed and implemented by Biotronik (and incorporated into the SMART-II algorithm) but is not commercialized in Biotronik ICDs.136 Feasibility investigations recently demonstrated a new algorithm using simultaneous atrial and ventricular ATP either to terminate or to discriminate 1 : 1 tachycardias. If the tachycardia persists after the simultaneous AV ATP, the discrimination algorithm considers the rhythm ventricular in origin, if the first sensed event after pacing is on the ventricular channel, and supraventricular in origin otherwise. In 62 ambulatory dual-chamber ICD patients, the algorithm terminated or correctly classified 1379 of 1381 SVT sequences (specificity 99.9%) and 23 of 26 VTs (sensitivity 88.5%).115

Features to Override Discriminators

Programmable duration-based “safety net” features deliver therapy if an arrhythmia satisfies the ventricular rate criterion for a sufficiently long duration even if discriminators indicate SVT. The premise is that VT will continue to satisfy the rate criterion for the programmed duration, whereas the ventricular rate during transient sinus tachycardia or AF will decrease below the VT rate boundary before the programmed duration is exceeded. The limitation is delivery of inappropriate therapy when SVT exceeds the programmed duration, which occurs frequently for durations of 1 minute or less and in up to 10% of SVTs at 3 minutes, depending on the VT detection interval and AV conduction.116 Because SVTs are much more common than VTs, programming of discriminator override duration to 3 minutes results primarily in inappropriate therapy of SVT. Programmed durations of 5 to 10 minutes are required to minimize such inappropriate therapy.137 The decision to use a discriminator override should be based on clinical factors, including the probability that discriminators will prevent detection of VT, the likely consequences of failure to detect VT, and the likelihood that SVT in the VT rate zone will persist long enough to trigger inappropriate therapy because of the override. For example, override features may be considered whenever a morphology algorithm is programmed without inducing VT at electrophysiologic study.

The Medtronic Protecta ICDs incorporate an additional, separately programmable “VF Zone High Rate Time Out.” It applies only to rhythms that are consistently in the VF rate detection zone. This new algorithm allows SVT discriminators to be applied with longer-duration (or no) time-out for slower rhythms, with a separate override duration for rhythms with rates in the VF zone.

SVT-VT Discriminators in Redetection

The SVT-VT discrimination during redetection serves two purposes: (1) to prevent inappropriate therapy for SVT after appropriate therapy for VT (Fig. 3-78) and (2) to provide a second chance for the algorithm to classify SVT correctly after inappropriate therapy. Biotronik ICDs (Fig. 3-79) and sorin-ELA ICDs provide essentially equivalent SVT-VT discrimination in initial detection and redetection, except that algorithm building blocks related to tachycardia onset are disabled. Boston Scientific algorithms permit programming discriminators after shocks, but not after ATP, when it is more useful and reliable. In Medtronic ICDs, the single-chamber stability discriminator applies to redetection only if it is “on” for initial detection. The user interface does not indicate that it applies in redetection, and it is rarely programmed for this purpose. St. Jude ICDs do not apply discriminators to redetection. Neither Medtronic nor St. Jude ICDs provide any discriminators to reject sinus tachycardia in redetection.

Measuring Performance of SVT-VT Discrimination Algorithms

Valid assessments and comparisons of algorithm performance require consideration of multiple ICD-related and clinical factors. A comprehensive assessment requires analysis of all tachycardia episodes, including those not stored in ICD memory and those in the VF zone to which discriminators may not apply.68,69,130 Programmed detection parameters may influence reported algorithm performance.69 In most studies, a few patients contribute a large number of SVT or VT episodes. Therefore, statistical methods such as the generalized estimating equation (GEE)138 should control for such clustered, nonindependent data.

Quantitative Considerations

Detection algorithms must maintain almost 100% sensitivity for detection of VT; but detection of hemodynamically-stable asymptomatic VT is not necessarily synonymous with therapy. If patient populations and programmed detection boundaries are equivalent, positive predictive accuracy may be the most useful statistical measure of algorithm performance.68 However, it is highly dependent on the ratio of SVT to VT episodes and therefore on the programmed detection rate and patient population (prevalence of SVT and VT). However, it is almost impossible to obtain a clinically meaningful comparison of different detection algorithms based only on their performance on different sets of data.

Is Ventricular Therapy for SVT Always Inappropriate?

Persistent, rapidly conducted atrial arrhythmias can cause hemodynamic compromise in patients with LV dysfunction or ischemia in patients with severe coronary artery disease. Thus, algorithmically inappropriate ventricular therapy may fortuitously terminate clinically significant SVT, although inappropriate ventricular therapy for SVT may also be proarrhythmic.141 Further, in patients with ICDs, rapid conduction in AF is often transient, and symptoms are often mild, but ventricular shocks delivered shortly after detection do not permit spontaneous termination of AF or slowing of the ventricular rate. Therefore, shocks will be delivered for AF that may have spontaneously terminated or slowed. Inappropriate shocks for AF may place patients at risk for thromboembolism if they are not anticoagulated. Also, early recurrence is common after transvenous cardioversion of AF.142 Experts differ about whether algorithmically inappropriate ventricular therapy of SVT may be clinically appropriate in specific clinical situations. ICDs designed to deliver both atrial and ventricular therapies may be implanted in patients who are likely to benefit from ICD-based therapy of SVT.