Pulse Oximetry, Capnography, and Blood Gas Analysis

Published on 07/03/2015 by admin

Filed under Critical Care Medicine

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 2197 times

Chapter 5 Pulse Oximetry, Capnography, and Blood Gas Analysis

Pulse oximetry

1 What is pulse oximetry and how does it work?

Pulse oximetry is the continuous noninvasive estimation of arterial hemoglobin-oxygen saturation. It is used routinely to monitor oxygenation in diverse clinical settings, including the operating room, emergency department, and intensive care unit. Clinical use of pulse oximetry falls into two main categories:

Pulse oximeters function by transmitting red light (660 nm, absorbed by oxyhemoglobin [O2Hb]) and infrared light (940 nm, absorbed by deoxyhemoglobin [deoxyHb]) from two light-emitting diodes (LEDs) through tissue containing pulsatile blood. The saturation of hemoglobin with oxygen is a function of the ratio of red to infrared light absorption from the pulsatile and nonpulsatile components of the signals. Thus the saturation (SpO2) is a function of the ratio of two ratios, cancelling out most differences caused by finger thickness, pigmentation, and other factors. A microprocessor algorithm is used to calculate the arterial saturation on the basis of calibration studies done by comparing true saturation measured on arterial blood with a CO-oximeter with the pulse oximeter reading. This calibration is factory set and is not adjustable.

Pulse oximeter probes can be applied to any site that allows orientation of the LED and photodetector opposite one another across a vascular bed. If the tissue is too thick, the signal is attenuated before reaching the detector and the oximeter cannot function. Oximeters can be applied to fingers, toes, earlobes, lips, cheeks, and the bridge of the nose. Esophageal and oral probes are also in development. Several manufacturers offer reflectance oximeter probes that can be applied to flat tissue surfaces such as the forehead or chest. Recently introduced earlobe-mounted sensors combine a pulse oximeter and a transcutaneous CO2 electrode. Many pulse oximeters now include noise and artifact rejection software. This refinement aids the determination of SpO2 in patients with low perfusion or motion (e.g., tremor).

4 What effects does dyshemoglobinemia have on pulse oximetry?

Because pulse oximeters use two wavelengths of light, they are capable of differentiating only two species of hemoglobin: Hb and O2Hb. Given that abnormal hemoglobin species such as carboxyhemoglobin (CoHb) or methemoglobin (MetHb) also absorb red and infrared light, their presence affects the SpO2 measurement, and their quantitative contribution cannot be determined. The pulse oximeter assumes that only functional hemoglobin is present (O2Hb or Hb), and the oxygen saturation is calculated on the basis of these amounts.

For example, CoHb is read by the limited wavelength analysis of a pulse oximeter as O2Hb (CoHb is scarlet red), which will falsely elevate the SpO2 reading. The absorption pattern of MetHb is interpreted by the pulse oximeter as 85% saturation; thus, progressively higher levels of MetHb cause the SpO2 value to converge on 85% regardless of the actual SaO2. When the presence of significant amounts of dysfunctional hemoglobin is suspected, a CO-oximeter should be used to determine O2Hb saturation. A multiwavelength laboratory CO-oximeter determines SaO2 more accurately in the presence of dysfunctional hemoglobins because it possesses wavelengths of light that can be used to detect the presence of CoHb and MetHb.

The presence of fetal hemoglobin has not been shown to significantly affect the accuracy of SpO2 measurements because its light absorption properties are similar to those of adult hemoglobin.

Capnography

5 What is capnography, and how does it work?

Capnography is the continuous measurement and graphic display of exhaled carbon dioxide. It is a noninvasive method to assess both ventilation and cardiac output. Most commonly, infrared light absorption by CO2 is the method used to determine the CO2 concentration. Sampling usually occurs in one of two ways. In a mainstream capnograph, CO2 levels are measured with a sensor (light source and detector) placed directly in the patient’s breathing circuit. With sidestream capnography, a continuous sample of airway gas is diverted from the patient’s breathing circuit or airway to the capnograph for analysis and display. The mainstream method has a very rapid response time, but, because the sensor must be placed near the patient, long-term monitoring may be cumbersome. The sidestream method, because it uses a thin plastic sampling tube, is lighter and allows for greater flexibility, but, because transit time is unavoidable, a slower response time (approximately 3-5 seconds) results. Because of mixing of gases in the sample stream, the absolute values of the plateau and baseline may also be attenuated. The sidestream device can also be used with a modified nasal cannula or face mask to monitor CO2 concentrations in the breath of patients who do not have endotracheal tubes in place.

The most commonly used method for measuring carbon dioxide in expired gases is infrared light absorbance. In addition, technologies such as Raman spectrometry and mass spectrometry are reliable, accurate, and responsive but generally more expensive. However, these options also offer detection of a variety of other gases and anesthetic vapors. Colorimetric detectors that attach to endotracheal tubes are available to help assess endotracheal tube placement. The colorimetric detector uses a pH-sensitive indicator strip to semiquantitatively detect exhaled CO2. Although portable and convenient, these devices yield results that are often more difficult to interpret than conventional capnographs, and they do not provide continuous measurement of CO2.

6 What does the capnogram reveal about a patient’s condition?

A capnograph provides a continuous display of the CO2 concentration of gases in the airways. A normal capnogram is shown in Figure 5-1. The CO2 partial pressure at the end of normal exhalation (phase III in the figure, end-tidal CO2 [PETCO2]) is a reflection of gas leaving alveoli and is an estimate of the alveolar CO2 partial pressure (PACO2). When ventilation and perfusion are well matched, the PACO2 approximates the arterial PCO2 (PaCO2), and thus PaCO2 equals PACO2 plus PETCO2. The presence of cyclical exhaled CO2 is useful in confirming airway patency, verifying endotracheal tube placement, and verifying the adequacy of pulmonary ventilation. In addition, decreases in cardiac output caused by hypovolemia or cardiac dysfunction result in decreased pulmonary perfusion. This causes an increased alveolar dead space, which dilutes PETCO2. Animal studies show that a 20% decrease in cardiac output causes a 15% decrease in PETCO2.

Alterations in the shape of the capnogram in a patient with an endotracheal tube and mechanical ventilation often provide clues to alterations in pulmonary pathologic condition and malfunction of ventilation equipment. For example, a staircase pattern in phase II may indicate sequential emptying of the lung, which may occur in main stem partial bronchial obstruction. An upward sloping plateau during expiration is a classic indication of late emptying of poorly ventilated alveolar spaces with elevated PCO2, which may occur with expiratory obstruction at the level of smaller airways, as seen in chronic obstructive pulmonary disease (COPD), bronchospasm, and other forms of ventilation-perfusion mismatching. A pulmonary embolus is another cause of a decrease in end-tidal CO2.

Arterial blood gases

12 Given that pulse oximetry is painless and accurate, why is ABG analysis even necessary?

Oximetry and the newer technology that made it more accessible, affordable, and accurate have decreased the need for ABG analysis in monitoring oxygen saturation. In fact, in many hospitals the number of ABGs done has decreased with the influx of oximeters into virtually every department in a hospital. However, relying on oximetry alone can lead to misdiagnosis, increased cost, and potentially fatal respiratory arrest. Consider the following examples of pitfalls of using oximetry alone that have been observed in practice.

1. A patient was noted to have oxygen desaturation via oximetry during a routine check after minor orthopedic surgery. The physicians evaluated this by ordering a chest radiograph, pulmonary function tests, and a ventilation-perfusion lung scan, the results of all of which were normal. Finally, ABG analysis was done and revealed alveolar hypoventilation alone with a normal A − aO2 gradient. The oxygen desaturation was simply the result of an increased PCO2 from hypoventilation in a patient receiving narcotics.

2. After an episode of smoke inhalation, a patient came to the emergency department because of headache and nausea. The oxygen saturation by oximetry was normal, and the patient was nearly dismissed after symptomatic treatment alone. Fortunately, recognizing the limitation of oximetry in differentiating oxygenated hemoglobin from CoHb, the physician drew an ABG sample, which revealed profound carbon monoxide poisoning. The patient was treated appropriately with 100% oxygen and close monitoring.

3. A patient with fever and sepsis had a respiratory rate of 40/min, and a chest radiograph revealed bilateral alveolar infiltrates. Oxygen was initiated, and with the use of pulse oximetry, the oxygen saturation was 90% after breathing 100% O2 by nonrebreathing mask. Unfortunately, the house officer did not draw an ABG sample and failed to realize that, with marked hyperventilation and respiratory alkalosis, the O2Hb dissociation curve is shifted to the left, thereby causing a much higher oxygen saturation for a given PaO2. Had ABG analysis been done at that time, it would have revealed pH 7.58, PACO2 22 mm Hg, PAO2 50 mm Hg, and O2 saturation 90%, all clearly indicating intubation and assisted ventilation with positive end-expiratory pressure.