Periapical radiography

Published on 13/06/2015 by admin

Filed under Radiology

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 5 (2 votes)

This article have been viewed 6675 times

Periapical radiography

Periapical radiography describes intraoral techniques designed to show individual teeth and the tissues around the apices. Each image usually shows two to four teeth and provides detailed information about the teeth and the surrounding alveolar bone.

Ideal positioning requirements

The ideal requirements for the position of the image receptor (film packet or digital sensor) and the X-ray beam, relative to a tooth, are shown in Fig. 9.1. They include:

Radiographic techniques

The anatomy of the oral cavity does not always allow all these ideal positioning requirements to be satisfied. In an attempt to overcome the problems, two techniques for periapical radiography have been developed:

Paralleling technique

Theory

This positioning has the potential to satisfy four of the five ideal requirements mentioned earlier. However, the anatomy of the palate and the shape of the arches mean that the tooth and the image receptor cannot be both parallel and in contact. As shown in Fig. 9.2, the image receptor has to be positioned some distance from the tooth.

To prevent the magnification of the image that this separation would cause, an X-ray beam as non-divergent as possible is required (see Fig. 9.3). As explained in Chapter 5, this is achieved by having a long focal spot to skin distance (fsd), ideally of 200 mm.

Film packet/sensor holders

A variety of holders has been developed for this technique. The different holders vary in cost and design, but essentially consist of three basic components (as shown in Fig. 9.4):

The different components of the various holders usually need to be assembled together before the holder can be used clinically. The holder design used depends upon whether the tooth under investigation is:

These variables mean that assembling the holder can be confusing, but it must be done correctly. To facilitate this assembly some manufacturers now colour-code the various components. Once assembled correctly the entire image receptor should be visible when viewed through the beam-aiming device, as shown in Fig. 9.5.

The choice of holder is a matter of personal preference and dependent upon the type of image receptor – film packet or digital sensor (solid-state or phosphor plate) – being used. A selection of different holders is shown in Fig. 9.6.

Typically, the same anterior holder can be used for right and left maxillary and mandibular incisors and canines utilizing a small image receptor (22 × 35 mm) with its long axis vertical. Four images in the maxilla and three images in the mandible are usually required to cover the right and left incisors and canines, as shown in Fig. 9.7.

Typically different holders are required for the right and left premolar and molar maxillary and mandibular posterior teeth. The different designs allow the holders to hook around the cheek and corner of the mouth. A large image receptor (31 × 41 mm) is ideally utilized with its long axis horizontal. Two images are usually required to cover the premolar and molar teeth in each quadrant, as shown is Fig. 9.8.

Positioning techniques

The radiographic techniques for the permanent dentition can be summarized as follows:

1. The patient is positioned with the head supported and with the occlusal plane horizontal.

2. The holder and image receptor are placed in the mouth as follows:

3. The holder is rotated so that the teeth under investigation are touching the bite block.

4. A cottonwool roll is placed on the reverse side of the bite block. This often helps to keep the tooth and image receptor parallel and may make the holder less uncomfortable.

5. The patient is requested to bite gently together, to stabilize the holder in position.

6. The locator ring is moved down the indicator rod until it is just in contact with the patient’s face. This ensures the correct focal spot to film distance (fsd).

7. The spacer cone is aligned with the locator ring. This automatically sets the vertical and horizontal angles and centres the X-ray beam on the image receptor

8. The exposure is made.

Positioning clinically using film packets and digital phosphor plates is shown in Figs 9.99.16 for the following different areas of the mouth:

Note:

Positioning using solid-state digital sensors

Clinical positioning of holders for the paralleling technique when using solid-state digital sensors can be more difficult because of the bulk and absolute rigidity of the sensor. Those systems employing cables also require extra care with regard to the position of the cable to avoid damaging it. Once the holder is inserted into the mouth, the positioning of the tubehead is the same as described previously when using other types of image receptors and is shown in Fig. 9.17 for different parts of the mouth.

Maxillary incisors

Maxillary canine

Maxillary premolars

Maxillary molars

Mandibular incisors

Mandibular canine

Mandibular premolars

Mandibular molars

Solid-state digital sensor positioning

Bisected angle technique

Theory

The theoretical basis of the bisected angle technique is shown in Fig. 9.18 and can be summarized as follows:

Positioning techniques

The bisected angle technique can be performed either by using an image receptor holder to support the image receptor in the patient’s mouth or by asking the patient to support the image receptor gently using either an index finger or thumb. Both techniques are described.

It is, however, good practice that the image receptor should be held by the patient only when it cannot otherwise be kept in position.

Using film packet/digital sensor holders

Various holders are available, a selection of which are shown in Fig. 9.20. The Rinn Bisected Angle Instruments (BAI) closely resemble the paralleling technique holders and consist of the same three basic components – image receptor holding mechanism, bite block and an X-ray beam-aiming device – but the image receptor is not held parallel to the teeth. The more simple holders and the disposable bite blocks hold the image receptor in the desired position but the X-ray tubehead then has to be aligned independently. In summary:

Using the patient’s finger

The specific positioning for different areas of the mouth, using both simple holders and the patient’s finger to support the image receptor, is shown in Figs 9.239.30.

Comparison of the paralleling and bisected angle techniques

The advantages and disadvantages of the two techniques can be summarized as follows:

Advantages of the paralleling technique

• Geometrically accurate images are produced with little magnification.

• The shadow of the zygomatic buttress appears above the apices of the molar teeth.

• The periodontal bone levels are well represented.

• The periapical tissues are accurately shown with minimal foreshortening or elongation.

• The crowns of the teeth are well shown enabling the detection of approximal caries.

• The horizontal and vertical angulations of the X-ray tubehead are automatically determined by the positioning devices if placed correctly.

• The X-ray beam is aimed accurately at the centre of the image receptor – all areas of the image receptor are irradiated and there is no coning off or cone cutting.

• Reproducible radiographs are possible at different visits and with different operators.

• The relative positions of the image receptor, teeth and X-ray beam are always maintained, irrespective of the position of the patient’s head. This is useful for some patients with disabilities.

Disadvantages of the paralleling technique

Disadvantages of the bisected angle technique

• The many variables involved in the technique often result in the image being badly distorted.

• Incorrect vertical tube head angulation will result in foreshortening or elongation of the image.

• The periodontal bone levels are poorly shown.

• The shadow of the zygomatic buttress frequently overlies the roots of the upper molars.

• The horizontal and vertical angles have to be assessed by observation for every patient and considerable skill is required.

• It is not possible to obtain reproducible views.

• Coning off or cone cutting may result if the central ray is not aimed at the centre of the image receptor, particularly if using rectangular collimation.

• Incorrect horizontal tube head angulation will result in overlapping of the crowns and roots.

• The crowns of the teeth are often distorted, thus preventing the detection of approximal caries.

• The buccal roots of the maxillary premolars and molars are foreshortened.

A visual comparison between the two techniques, showing how dramatic the variation in image quality and reproducibility can be, is shown in Figs 9.31 and 9.32.

Positioning difficulties often encountered in periapical radiography

Placing the image receptor intraorally in the textbook-described positions is not always possible. The radiographic techniques described earlier often need to be modified. The main difficulties encountered involve:

Problems posed by mandibular third molars

The main difficulty is placement of the image receptor sufficiently posteriorly to record the entire third mandibular molar (particularly when it is horizontally impacted) and the surrounding tissues, including the inferior dental canal (see Fig. 9.33).

Possible solutions

These include:

• Using specially designed or adapted holders as shown in Fig. 9.34 to hold and position the image receptor in the mouth, as follows:

• Taking two radiographs of the third molar using two different horizontal tubehead angulations, as follows:

1. The image receptor is positioned as posteriorly as possible (using the technique described with the holders).

2. The X-ray tubehead is aimed with the ideal horizontal angulation so the X-ray beam passes between the second and third molars. (With horizontally impacted third molars, the apex may not be recorded using this positioning, as shown in Fig. 9.35.)

3. A second image receptor is placed in the same position as before, but the X-ray tubehead is positioned further posteriorly aiming forwards to project the apex of the third molar on to the receptor. (With this positioning, the crowns of the second and third molars will be overlapped, as shown in Fig. 9.35.)

Note: The vertical angulation of the X-ray tubehead is the same for both projections.

Problems of gagging

The gag reflex is particularly strong in some patients. This makes the placement of the image receptor in the desired position particularly difficult, especially in the upper and lower molar regions.

Possible solutions

These include:

• Patient sucking a local anaesthetic lozenge before attempting to position the image receptor

• Asking the patient to concentrate on breathing deeply while the image receptor is in the mouth

• Placing the image receptor flat in the mouth (in the occlusal plane) so it does not touch the palate, and applying the principles of the bisected angle technique – the long axes of the tooth and image receptor are assessed by observation and the X-ray tubehead’s position modified accordingly, as shown in Fig. 9.36.

Problems encountered during endodontics

The main difficulties involve:

Possible solutions

These include:

• The problem of image receptor placement and stabilization can be solved by:

– Using a simple image receptor holder such as the Rinn Eezee-Grip®, as shown in Fig. 9.37. This is positioned in the mouth and then held in place by the patient.

– Using one of the special endodontic image receptor holders that have been developed. These incorporate a modified bite platform area, to accommodate the handles of the endodontic instruments, while still allowing the image receptor and the tooth to be parallel. (See Fig. 9.38.)

• The problem of identifying and separating the root canals can be solved by taking at least two radiographs, using different horizontal X-ray tubehead positions, as shown in Fig. 9.39.

• The problems of assessing root canal length can be solved by:

The calculation is done as follows (see Fig. 9.40):

Problems of the edentulous ridge

The main difficulty in the edentulous and partially dentate patient is again image receptor placement.

Possible solutions

These include:

• In edentulous patients, the lack of height in the palate, or loss of lingual sulcus depth, contraindicates the paralleling technique and all periapical radiographs should be taken using a modified bisected angle technique. The long axes of the image receptor and the alveolar ridge are assessed by observation and the X-ray tubehead position adjusted accordingly as shown in Fig. 9.41.

• In partially dentate patients, the paralleling technique can usually be used. If the edentulous area causes the image receptor holder to be displaced, the deficiency can be built up by using cottonwool rolls, as shown in Fig. 9.42.

Problems encountered in children

Once again the main technical problem (as opposed to management problems) encountered in children is the size of their mouths and the difficulty in placing the image receptor intraorally. The paralleling technique is not possible in very small children, but can often be used (and is recommended) anteriorly, for investigating traumatized permanent incisors. The reproducibility afforded by this technique is invaluable for future comparative purposes.

A modified bisected angle technique is possible in most children, with the image receptor placed flat in the mouth (in the occlusal plane) and the position of the X-ray tubehead adjusted accordingly, as shown in Fig. 9.43.

Assessment of image quality

Assessment of the quality of all radiographic images should be regarded as a routine part of any quality assurance (QA) programme (see Ch. 17). Essentially image quality assessment involves three separate stages, namely:

Ideal quality criteria

Irrespective of the type of image receptor or technique being used, typical quality criteria for a periapical radiograph should include:

• The image should have acceptable definition with no distortion or blurring.

• The image should include the correct anatomical area together with the apices of the tooth/teeth under investigation with at least 3–4 mm of surrounding bone.

• There should be no overlap of the approximal surfaces of the teeth.

• The desired density and contrast for film-captured images will depend on the clinical reasons for taking the radiograph, e.g.

• The images should be free of coning off or cone-cutting and other film handling errors.

• The images should be comparable with previous periapical images, both geometrically and in density and contrast.

Subjective rating of image quality

A simple three-point subjective rating scale for all intraoral and extraoral film-captured radiographs was published in the UK in 2001 in the Guidance Notes for Dental Practitioners on the Safe Use of X-ray Equipment. A summary is shown in Table 9.1, and is repeated in Chapters 10 and 15. Image quality is discussed in detail, together with the errors associated with exposure factors and chemical processing, in Chapter 19. Patient preparation and positioning errors in periapical radiography are described below.

Table 9.1

Subjective quality rating criteria for film-captured images published in the 2001 Guidance Notes for Dental Practitioners on the Safe Use of X-ray Equipment

Rating Quality Basis
1 Excellent No errors of patient preparation, exposure, positioning, processing or film handling
2 Diagnostically acceptable Some errors of patient preparation, exposure, positioning, processing or film handling, but which do not detract from the diagnostic utility of the radiograph
3 Unacceptable Errors of patient preparation, exposure, positioning, processing or film handling, which render the radiograph diagnostically unacceptable

Assessment of rejected films and determination of errors

Patient preparation and positioning (radiographic technique) errors (Fig. 9.44)

These can include:

• Failure to remove dentures or orthodontic appliances

• Failure to position the image receptor correctly to capture the area of interest, thereby failing to image the apices and periapical tissues

• Failure to position the image receptor correctly causing it to bend (if flexible) creating geometrical distortion

• Failure to orientate the image receptor correctly and using it back-to-front

• Failure to align the X-ray tubehead correctly in the horizontal plane, either

• Failure to align the X-ray tubehead correctly in the vertical plane, either

• Failure to instruct the patient to remain still during the exposure with subsequent movement resulting in blurring

• Failure to set correct exposure settings (see Ch. 17)

• Careless inadvertent use of the image receptor twice.

Note: Many of these technique errors can be avoided by using the paralleling technique utilizing image receptor holders with beam-aiming devices.

To access the self assessment questions for this chapter please go to www.whaitesessentialsdentalradiography.com