Mechanical Ventilation of the Newborn

Published on 01/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 5 (1 votes)

This article have been viewed 3367 times

Mechanical Ventilation of the Newborn

Indications for Mechanical Ventilation of the Newborn Generally Fall into the Following Categories

Severe oxygenation deficit from

Ventilatory failure with elevated Pco2 and significant respiratory acidosis

Congenital anomalies (see Chapter 27)

Need for surfactant administration (modified from the AARC Practice Guidelines on Surfactant Administration, 1994)

II Goals of Mechanical Ventilation

Provide adequate ventilation

Provide adequate oxygenation

Promote patient/ventilator synchrony

Recruit and maintain lung volume

III Complications of Mechanical Ventilation in the Newborn

Ventilator-induced lung disease

Hyperoxia

Hypocarbia

Decreased cardiac output

Pneumothorax

Pneumonia

Abdominal distention (gastric air)

Mechanical failure

Airway complications

IV Manual Ventilation

    Before initiating mechanical ventilation in the newborn, manual ventilation and intubation are performed.

Manual ventilation (BMV) of the neonate

1. Equipment

2. Position mask on infant’s face

Neonatal Intubation

Unlike the adult airway cuffed ETTs are generally not needed to provide mechanical ventilation to neonates.

The cricoid cartilage is the narrowest point in the neonatal airway.

Appropriately sized uncuffed ETTs are adequate to provide mechanical ventilation and reduce airway complications associated with cuffed tubes.

Approximate ETT sizes for gestational ages and weights are listed in Table 28-1.

TABLE 28-1

ETT and Suction Catheter Sizes for Various Gestational Ages and Weights

Gestational Age (wk) Weight (kg) ETT Size (mm ID) Suction Catheter Size (French)
<28 <1 2.5 5
28-34 1-2 3.0 6 or 8
34-38 2-3 3.5 8
>38 >3 3.5-4.0 8 or 10

image

ETT, Endotracheal tube; ID, inner diameter.

Equipment

Placing the tube

1. Establish adequate Spo2 with oxygen, or use BMV if apneic.

2. Position infant on a flat surface with head midline and neck slightly extended.

3. Turn on laryngoscope light, and hold laryngoscope in left hand.

4. Slide the laryngoscope blade over the right side of the tongue.

5. Advance the blade to the tip of the vallecula.

6. Lift the tongue out of the way to expose the pharyngeal area.

7. Observe the vocal cords.

8. Suction if necessary to improve view of larynx.

9. Hold the ETT in your right hand, and insert it through the vocal cords as they open.

10. Insert tube until vocal cord guide on ETT is at the level of the vocal cords.

11. Stabilize the tube with one hand, and remove laryngoscope.

12. If stylet was used withdraw it from the tube, keeping a firm hold on the ETT.

13. Note landmark on tube associated with infant’s lip or nare if nasal tube is used.

14. With tape or ETT holder secure ETT to infant’s face.

Confirming ETT position

Minimize ETT length

VI Types of Neonatal Mechanical Ventilators

The condition necessitating mechanical ventilation and the goals of support should be considered when selecting the type of ventilator, ventilator mode, and settings.

Neonatal ventilators are generally classified as conventional or high frequency.

Approaches to conventional and high frequency ventilation are outlined below. (A detailed description of high frequency ventilation is presented in Chapter 42.)

Neonatal conventional ventilation

1. Neonates requiring mechanical ventilation are most often ventilated using pressure-limited ventilation.

2. Pressure-limited ventilation is accomplished by setting a peak inspiratory pressure (PIP) that the ventilator targets during each mechanical breath.

3. Pressure-limited ventilation can be accomplished using any of the following modes.

a. Synchronized intermittent mandatory ventilation (SIMV) (Figure 28-1)

(1) Mandatory mechanical rate is set (range generally 15 to 40 breaths/min).

(2) Minimum PEEP is set (3 to 8 cm H2O).

(3) Every mechanical breath starts from the preset PEEP level to a preset inspiratory pressure (generally between 15 and 25 cm H2O). Total peak pressure should be less than 30 cm H2O.

(4) The difference between PEEP and the inspiratory pressure target should result in a Vt of 5 to 7 ml/kg.

(5) Inspiratory time is operator controlled on all mandatory breaths (normally set between 0.3 and 0.5 second).

(6) Continuous flow of gas is available for all nonmechanical (spontaneous) breaths (normally set between 6 and 10 L/min).

(7) Spontaneous breaths are not supported with positive pressure >PEEP level.

(8) Pressure support may be applied during spontaneous breaths.

(9) Delivered oxygen concentrations vary from 21% to 100%.

b. Assist control (AC) (Figure 28-2)

c. Pressure support (PS) (Figure 28-3)

d. Volume guarantee (Figure 28-4)

e. Continuous positive airway pressure (CPAP) (Figure 28-5)

Additional variables affecting neonatal conventional ventilation

1. Inspiratory trigger is the mechanism that causes the initiation of a mechanical breath and may be a result of time, flow, or pressure.

2. Expiratory trigger is the mechanism responsible for cycling a mechanical breath from the inspiratory phase to the expiratory phase and may be a result of time, flow, or pressure.

3. Inspiratory time is from triggering of inspiration to the beginning of expiration

4. Mean airway pressure

High frequency ventilators are considered nonconventional ventilators and offer an alternative approach to ventilate neonates. Currently used high frequency ventilators are categorized as oscillators and flow or jet ventilators (see Chapter 42).

1. High frequency oscillatory ventilation (HFOV)

2. High frequency jet ventilator (HFJV)

3. High frequency flow interrupter ventilator (HFFIV)

VII Initiating Ventilatory Support

When initiating mechanical ventilation to a neonate the pathology and gestational age of the neonate need to be considered.

Either conventional or high frequency ventilation (HFV) can be used as long as the ventilator is adjusted for the specific needs of the infant.

Avoiding damage to the lung from positive pressure ventilation is a concern for all gestational ages.

Careful attention to recruiting and maintaining appropriate lung volume is essential.

Of particular concern for the preterm neonate is retinal damage associated with hyperoxia. Oxygen should be adjusted with specific lower Spo2 (88% to 93%) maintained to avoid retinal damage.

In contrast infants at risk for pulmonary hypertension (e.g., MAS, CDH, and sepsis) need higher Spo2 (>95%) to avoid constriction of the pulmonary vasculature and pulmonary hypertension leading to an increase in shunting across the ductus arteriosus.

Other factors considered when initiating mechanical ventilation

Typical blood gases and saturation goals

Preductal and postductal saturations

1. Term or near-term infants (>34 weeks’ gestational age) at risk for pulmonary hypertension (primary disease, MAS, and sepsis) or suspected of having persistent pulmonary hypertension of the newborn (PPHN) should be monitored for differences in preductal and postductal saturations.

2. Postductal saturation 5% less than preductal saturation may indicate significant pulmonary hypertension. If present

Initiating conventional ventilation

1. Infants with hyperinflation (e.g., CLD, PIE)

2. If hyperinflation is more prominent on one side, position neonate with the more affected side dependent to direct volume to less inflated lung.

3. Reassess lung volume with chest radiography.

4. Infants with low lung volume, term or preterm infants (e.g., respiratory distress syndrome [RDS], pneumonia).

Initiating HFOV

Initiating HFJV

Maintaining ventilator support in neonates requires careful attention to the following.

1. Cardiac monitoring with alarms set for high and low heart rates

2. Set ventilator-disconnect alarm. Brief disconnection from the ventilator may result in significant hypoxia, bradycardia, and rapid derecruitment of lung volumes or accidental delivery of higher than needed oxygen concentrations.

3. Continuous monitoring of Spo2 to ensure Spo2 is in target range.

4. Frequent assessment of chest movement/chest wiggle with HFV

5. Frequently monitor patient/ventilator synchrony.

6. Ensure adequate humidification as addressed in Section VII, N, Assessing Humidification During Mechanical Ventilation.

7. Maintain security of ETT.

8. Assess patient’s secretions: amount, color, and consistency.

Assessing humidification during mechanical ventilation

Suctioning mechanically ventilated neonates

1. Routine/scheduled suctioning should be discouraged.

2. Suctioning should be performed when clinically indicated, such as

3. Appropriately sized suction catheters should be used; French size approximately two times the ETT ID.

4. Suction should not exceed 70 to 100 mm Hg negative pressure.

5. Suction time should not exceed 10 seconds.

6. Catheter should be measured and inserted to no more than 0.5 cm beyond the ETT.

7. Complications of suctioning include

8. Increasing FIO2 by approximately 0.2 before suctioning may decrease some of the risks associated with suctioning.

Assessment for exogenous surfactant administration (Modified from AARC Practice Guidelines for Surfactant Administration)

1. Preterm infants (<34 weeks’ gestation) with evidence of RDS or infants with meconium aspiration should be considered for exogenous surfactant.

2. The presence of two or more of the conditions outlined in Box 28-1 can be used as indications for exogenous surfactant.

3. Potential complications of surfactant administration are outlined in Box 28-2.

4. Guidelines for administering surfactant (specific brand may call for some modifications in these suggested guidelines)

a. Confirm appropriate ETT position before instilling surfactant.

b. Allow surfactant to reach room temperature before administering.

c. Position infant on side.

d. Divide total dose into four aliquots, and administer two aliquots to each side.

e. Advance catheter to the tip of the ETT, and instill each aliquot directly into the ETT.

f. Between each aliquot remove the catheter and allow the ventilator to disperse the surfactant into the lung.

g. Monitor saturations during administration.

h. Adjust ventilator support to maintain adequate minute volume during procedure (increased rate, increased pressure) and to avoid lung derecruitment.

i. Monitor changes in lung compliance after surfactant administration.

j. Adjust ventilator pressure to avoid hyperinflation, hyperventilation, and hyperoxia after surfactant delivery.

k. Endotracheal suctioning should be avoided for a minimum of 1 hour after surfactant administration unless clinically indicated.

l. Reassess ventilator settings and oxygen requirement for potential additional doses.

m. Consider administering second dose if ventilator and oxygen requirement remain significant (e.g., FIO2 > 0.3 and MAP > 8).

VIII Special Considerations for the Infant with Congenital Cardiac Disease (see Chapter 27)

Infants with certain cardiac lesions depend on maintaining a PDA for survival.

Low blood oxygen levels can be essential for these infants.

1. If sufficiently low Spo2 cannot be maintained with room air, subatmospheric oxygen concentrations (≥17% O2) may be needed.

2. Set continuous flow ventilator to room air.

3. Add precise flow of nitrogen to continuous flow to reduce FIO2 to desired range (generally 17% to 19%).

4. Table 28-3 outlines nitrogen and air-flow rates to obtain specific concentrations of inspired oxygen.

TABLE 28-3

Flow Rate of Nitrogen and Air Needed to Deliver Subatmospheric Oxygen Concentrations

Ventilator Flow FIO2 Nitrogen Flow
8 0.17 1.9
9 0.17 2.1
10 0.17 2.4
11 0.17 2.6
12 0.17 2.8
8 0.18 1.3
9 0.18 1.5
10 0.18 1.7
11 0.18 1.8
12 0.18 2.0
8 0.19 0.8
9 0.19 0.9
10 0.19 1.1
11 0.19 1.2
12 0.19 1.3

image

5. Analyze inspired gas concentration continuously.

6. Monitor nitrogen tank pressure.

7. Monitor Spo2 continuously.

IX Inhaled Nitric Oxide (see Chapter 34)

Initiate iNO at 20 ppm.

Note preductal and postductal saturations before initiating iNO, and monitor changes after starting iNO.

Obtain arterial blood gas after iNO initiated.

Wean iNO when clinically indicated (sample iNO weaning protocol in Figure 28-6).

Avoid interrupting iNO administration once gas is initiated.

Connect resuscitation bag to iNO system to avoid interrupting iNO administration if manual ventilation becomes necessary.

Monitor iNO tank pressure.

Complications of iNO

Weaning Mechanical Support

Assessment for weaning generally should begin as soon as the goals for instituting mechanical ventilation have been achieved, provided no new indications for maintaining support have occurred and the infant’s ability to breathe is not impeded by any pharmacologic agents.

Goals of weaning

Weaning from SIMV pressure ventilation

Weaning from assist control pressure ventilation

Weaning from HFOV

Weaning from PSV and pressure support volume guarantee (PSVG)

Extubation

Postextubation

1. Establish desired Spo2, and provide appropriate support to maintain goal.

2. Continuously assess infant’s respiratory rate and pattern of breathing.

3. Note signs of respiratory distress.

4. Initiate nasal CPAP postextubation particularly in preterm infants to help maintain lung volume and decrease FIO2 requirements.

5. Infants with more stable lung volumes may extubate to nasal cannulas.

XI Extracorporeal Life Support

When adequate oxygenation and ventilation cannot be accomplished using the previously described methods, extracorporeal membrane oxygenation may be an option.

Indications for extracorporeal life support (ELSO) in neonates

Methods of ECLS support

1. Extracorporeal support consists of vascular cannula(s), a circuit, pump, an oxygenator, heat exchanger, and an oxygen source (Figure 28-7).

2. Venoarterial (VA) ECLS requires cannulation of two vessels, usually the right common carotid artery and the right internal jugular vein (Figure 28-8).

3. Venovenous (VV) ECLS requires cannulation of a single vessel, the right internal jugular vein, with a double-lumen cannula (Figure 28-9).

4. With both types of support blood is drained from the right atrium to the extracorporeal circuit.

5. A pump propels the drained blood through an oxygenator.

6. The oxygenator consists of a two-sided membrane.

7. Desaturated (venous) blood flows along one side of the membrane.

8. Gas with a higher partial pressure of oxygen flows along the opposite side of the membrane.

9. The difference in partial pressure results in oxygen diffusing into the venous blood.

10. Similarly the venous blood has a higher partial pressure of carbon dioxide than the gas side of the membrane, resulting in carbon dioxide diffusing into the gas side of the membrane, where it is eliminated to the atmosphere.

11. Fully saturated blood with the desired carbon dioxide level exits the membrane oxygenator.

12. The blood is warmed to the desired temperature as it flows through a blood warmer before returning to the infant.

13. During VA support the blood is returned to the infant’s arterial circulation through the arterial cannula.

14. During VV support the oxygenated blood is returned through a second lumen of the venous cannula to the infant’s right heart, and with adequate cardiac function the oxygenated blood is delivered to the arterial circulation.

15. Systemic anticoagulation is necessary to reduce the potential formation of clots on the artificial surfaces of the circuit, oxygenator, and cannulas.

Ventilator management during ECLS

ECLS course

1. Lung rest occurs while the underlying condition is reversed.

2. Changes in the native lung on chest radiographs and improvement in blood gases without increases in ECLS support are indications for weaning from ECLS.

3. Ventilator support is increased as ECLS support is decreased.

4. ECLS support is discontinued when gas exchange can be accomplished by FIO2 and ventilator pressures less likely to cause lung damage.

5. Cannulated vessels are either ligated or repaired at decannulation.

6. Systemic anticoagulation is discontinued.

7. Major risks of ECLS are outlined in Box 28-3.

8. Mechanical ventilation is weaned as the infant’s condition improves.

Share this: