Mechanical Ventilation of the Adult

Published on 12/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 1.5 (2 votes)

This article have been viewed 4908 times

15 Mechanical Ventilation of the Adult

Note 1: This book is written to cover every item listed as testable on the Entry Level Examination (ELE), Written Registry Examination (WRE), and Clinical Simulation Examination (CSE).

The listed code for each item is taken from the National Board for Respiratory Care (NBRC) Summary Content Outline for CRT (Certified Respiratory Therapist) and Written RRT (Registered Respiratory Therapist) Examinations (http://evolve.elsevier.com/Sills/resptherapist/). For example, if an item is testable on both the ELE and the WRE, it will be shown simply as: (Code: …). If an item is testable only on the ELE, it will be shown as: (ELE code: …). If an item is testable only on the WRE, it will be shown as: (WRE code: …).

Following each item’s code will be the difficulty level of the questions on that item on the ELE and the WRE. (See the Introduction for a full explanation of the three question difficulty levels.) Recall [R] level questions typically expect the exam taker to recall factual information. Application [Ap] level questions are harder because the exam taker may have to apply factual information to a clinical situation. Analysis [An] level questions are the most challenging because the exam taker may have to use critical thinking to evaluate patient data to make a clinical decision.

Note 2: A review of the most recent ELEs has shown that an average of 43 questions (out of 140), or 31% of the exam, cover mechanical ventilation of the adult. A review of the most recent WRE has shown that an average of 23 questions (out of 100), or 23% of the exam, cover mechanical ventilation of the adult. The CSE is comprehensive and may include everything that should be known by an advanced level respiratory therapist.

MODULE A

1. Ventilator flow, volume, and pressure waveforms

a. Review the patient’s chart for information on airway graphics (Code: IA7c) [Difficulty: ELE: R; WRE: Ap]

A patient who has been intubated and placed on a modern mechanical ventilator with a microprocessor and graphics software can have ventilator flow, volume, and pressure waveforms visualized on the monitor, stored in memory, or printed out. Look for this information and compare it with the patient’s current situation. See Figure 15-1 for examples of pressure, volume, and flow tracings. See Figure 15-2 for key points of information available from a flow vs. time graph.

c. Select ventilator graphics (Code: IIID3) [Difficulty: ELE: R, Ap; WRE: An]

The operator typically can select any two of the following for display on the monitor: time, flow, pressure, and/or volume. Certain combinations are selected to best present the needed information. For example, air trapping is best shown by comparing flow vs. time (Figure 15-3) peak and plateau pressures are best shown by comparing pressure vs. time (Figure 15-4), and lung inflection points are best shown by comparing volume vs. pressure (discussed later). Note examples of ventilator graphics throughout this chapter.

2. Review the patient’s chart for information on the work of breathing (Code: IA7d) [Difficulty: ELE: R; WRE: Ap]

Work of breathing (WOB) normally refers to how much energy the patient has to expend to inhale. Patients with stiff lungs or high airway resistance or both have an increased WOB. Exhalation normally is passive and requires no work. However, some patients with high airway resistance have to work to exhale. Look in the patient’s chart for information on patient complaints of shortness of breath and easy tiring as signs of increased WOB.

A patient who has been intubated and placed on a modern mechanical ventilator with a microprocessor and graphics software can have WOB measured. See Figure 15-5 for a pressure/volume loop tracing that shows a patient’s WOB. WOB is minimized when the ventilator is set to minimize the negative pressure and inspiratory flow the patient has to generate.

If the breathing of an intubated and ventilated patient appears to be unsynchronized with the ventilator, his or her WOB should be evaluated. Ask the conscious patient simple questions to try to determine what the problem is. Observe the patient for use of accessory muscles as a sign of increased WOB. If the ventilator is capable, program it to perform a pressure/volume loop of the patient’s WOB. Be prepared to adjust parameters such as the machine’s sensitivity and inspiratory flow to minimize the patient’s workload.

3. Airway resistance

c. Determine the patient’s airway resistance (Code: IB9n) [Difficulty: ELE: R, Ap; WRE: An]

Most of the microprocessor ventilators offer software for calculating all of these values. However, in other situations, they must be calculated manually.

1. Procedure for calculating airway resistance

4. Auto-PEEP detection

a. Perform the procedure to detect auto-PEEP (WRE code: IB9w) [Difficulty: WRE: R, Ap, An]

Auto-PEEP is positive end-expiratory pressure in the lungs that cannot be seen on the ventilator’s pressure manometer. (The terms inadvertent PEEP and intrinsic PEEP also are used.) Auto-PEEP is caused by air trapping resulting from an inadequate expiratory time. It becomes more likely when the inspiratory time is increased or the expiratory time is decreased, or in patients with long time constants of ventilation. Simply put, the next breath is delivered before the patient has exhaled completely (see Figure 15-3). This problem is seen frequently in patients with status asthmaticus or COPD because of early small airway closure. In patients with acute respiratory distress syndrome (ARDS) who are receiving pressure-controlled inverse ratio ventilation (PCIRV), the long inspiratory times used increase the risk of expiratory air trapping. Auto-PEEP is more likely to be found when the inspiratory/expiratory (I : E) ratio becomes 2 : 1 or greater.

The level of auto-PEEP can be determined in different ways, depending on the type of ventilator that is being used. It can be measured on the pressure manometer of most ventilators. The trapped expiratory gas also can be seen on the graphic display of all current generation microprocessor-type ventilators. The following procedure can be followed for determining the presence or level of auto-PEEP:

It is important to add any auto-PEEP to the amount of therapeutic PEEP the patient has. This should be recorded as the total PEEP. For example, the patient has 5 cm of therapeutic PEEP and 2 cm of auto-PEEP for 7 cm of total PEEP. It may be thought that the total PEEP level places the patient at risk for volutrauma or decreased venous return and lowered cardiac output. The amount of auto-PEEP can be reduced by decreasing the inspiratory time, increasing the expiratory time, or decreasing the tidal volume. Lack of auto-PEEP can be confirmed by this procedure. If the auto-PEEP cannot be eliminated, therapeutic PEEP can be added to match it. By increasing the baseline pressure, the patient can more easily trigger an assisted or synchronous intermittent mandatory ventilation (SIMV) breath. It is especially important to decrease auto-PEEP and therapeutic PEEP levels as the patient’s lung compliance improves and airway resistance returns to normal.

b. Interpret ventilator graphics to detect auto-PEEP (WRE code: IB10w) [Difficulty: WRE: R, Ap, An]

Figure 15-3 demonstrates two ways that air trapping on exhalation can be identified as auto-PEEP (unintended positive end-expiratory pressure). Note in Figure 15-3 (bottom) that the patient’s expiratory flow does not reach baseline pressure before another breath is delivered. This proves that air trapping has occurred. The larger the gap between the pressure at the end of expiration and at baseline, the greater is the air trapping.

6. Lung compliance

c. Determine the patient’s plateau pressure and lung compliance (Code: IB9n) [Difficulty: ELE: R, Ap; WRE: An]

Most microprocessor ventilators offer software that can be used for calculating all of these values. However, in other situations, they must be calculated manually, as follows.

1. Procedure for calculating the tubing compliance factor

Before the actual calculation of the patient’s static and dynamic compliance can be performed, the compliance of the breathing circuit should be determined. During positive-pressure ventilation, some of the set tidal volume never reaches the patient because it is “lost” in the circuit. Remember that when a positive-pressure breath is delivered, the ventilator circuit will be expanded and the gas within the tubing will be compressed. The term compressed volume is commonly used to describe this lost volume. Subtract the compressed volume from the exhaled volume coming from the ventilator to determine the patient’s actual tidal volume. (Some of the current generation of microprocessor ventilators will calculate the compressed volume. The therapist can have the ventilator compensate for the lost volume and deliver a tidal volume that meets the set volume on the ventilator.)

For greatest accuracy in the calculation of static and dynamic compliance and the calculation of actual tidal and sigh volumes, any lost volume must be subtracted from the exhaled tidal volume to find the actual tidal volume. The tubing compliance factor is used in the calculation to determine the compressed volume, through the following procedure:

3. Procedure for calculating dynamic compliance

5. Calculate the dynamic compliance (Cdyn) by using this formula:

image

in which compressed volume is Compliance factor × Peak pressure.

7. Interpret the patient’s lung mechanics results: plateau pressure, airway resistance, dynamic lung compliance, and static lung compliance values on the ventilator (Code: IB10n) [Difficulty: ELE: R, Ap; WRE: An]

Any increase in airway resistance or decrease in lung compliance or both creates an increase in the patient’s WOB. Examples of conditions or situations in which an increased airway resistance is found include bronchospasm, secretions, mucosal edema, airway tumor, placement of a small endotracheal tube, and biting or kinking of the endotracheal tube. Lung compliance is decreased by pneumonia, pulmonary edema, ARDS, pulmonary fibrosis, atelectasis, consolidation, hemothorax, pleural effusion, air trapping, pneumomediastinum, and pneumothorax. Examples of chest wall and abdominal conditions that reduce compliance include various chest wall deformities, circumferential chest or abdominal burns, enlarged liver, pneumoperitoneum, peritonitis, abdominal bleeding, herniation, and advanced pregnancy. Correction of the problem should return the patient’s ventilator pressure(s) to baseline and normalize the patient’s WOB.

Six possible combinations of increasing or decreasing static and dynamic lung compliance exist. Each has its own possible causes and is discussed in turn. The patient must be passive on the ventilator for the measured values to be accurate. Check two or three breaths for increased accuracy. Let the patient have a normal breath or two between each of the peak and plateau pressure measurement breaths.

e. False increased dynamic compliance with true increased static compliance

False increased dynamic compliance with true increased static compliance is noticed as a decrease in both peak and plateau pressures (Figure 15-10). This is seen when the patient’s lung/thoracic compliance improves. The plateau pressure decreases, and, as an artifact, the peak pressure also decreases. Notice that the difference between peak and plateau pressures remains the same. This indicates that the patient’s airway resistance is unchanged.

f. True increased dynamic compliance with true increased static compliance

True increased dynamic compliance with true increased static compliance also is noticed as a decrease in both peak and plateau pressures (Figure 15-11). This is seen when the patient’s airway resistance and his or her lung/thoracic compliance improve. Notice that the plateau pressure has decreased, thus indicating more compliant lungs. Also notice that the difference between peak and plateau pressures has decreased. This demonstrates that the airway resistance also has decreased.

All six examples of increasing or decreasing static or dynamic lung compliance or both make use of a single tidal volume that is analyzed for peak and plateau pressures. Some practitioners advocate using several different tidal volumes (e.g., 8, 10, and 12 mL/kg of ideal body weight) when measuring dynamic and static pressures. The measured values are plotted on a graph to find the patient’s optimal tidal volume that results in the highest static compliance value. Figure 15-12 shows a series of these graphs. The curves for diseased lungs and airways are quite different from those of a normal person or a patient with a pulmonary embolism. Because of this, a pulmonary embolism should be considered if the patient’s condition deteriorates rapidly and no change in dynamic or static compliance values is observed.

MODULE B

Conventional mechanical ventilation is defined here as the use of a single ventilator that can provide the customary modes and options needed by the large majority of patients. This ventilatory support is provided through an endotracheal tube or a tracheostomy tube. Several physiologic criteria have been compiled to help the clinician determine when a patient is in respiratory or ventilatory failure (Box 15-1) and needs ventilatory support. Remember that the patient may not fail each and every criterion; however, the patient often will fail one or more criteria in each category.

1. Perform the following procedures to make sure that the patient is adequately oxygenated

b. Administer oxygen, as needed, to prevent hypoxemia (ELE code: IIID6) [Difficulty: ELE: R, Ap, An]

Oxygen administration and adjustment were discussed in Chapter 6. In brief, the goal of oxygen administration is to keep the Pao2 level of most patients between 60 and 90 torr and the Spo2 level greater than 90%. Exceptions are the patient with COPD who is breathing on hypoxic drive and the patient who is in a cardiac arrest situation. The following formula can be used to help guide the use of supplemental oxygen in most stable patients:

image

Those patients who have refractory hypoxemia (e.g., ARDS) will not respond with a normal increase in Pao2 level as the oxygen percentage is increased. In the short term, use whatever oxygen percentage is needed to achieve the clinical goal. The risk of oxygen toxicity increases when the Fio2 is greater than 0.5 for periods of longer than 48 hours. Always recheck the patient’s arterial oxygen level after a change in the Fio2 has been made.

Either of the following formulas can be used in the special situation of determining the flows of air and oxygen into a “bleed-in” type of intermittent mandatory ventilation (IMV) or continuous positive airway pressure (CPAP) system to obtain an ordered Fio2. Either version can be used for determining gas flows, total flow, and oxygen/air ratio through an air entrainment (Venturi) mask.

The first formula follows:

image

The second formula follows:

image

in which F1 is the flow of first gas (oxygen), C1 is the concentration of oxygen in the first gas (1.0 for pure oxygen), F2 is the flow of second gas (air), C2 is the concentration of oxygen in the second gas (0.21 for air), FT is the total flow of both gases, and CT is the concentration of oxygen in the mix of both gases. Use algebraic manipulation to solve for the unknown.

2. Initiate and adjust continuous mechanical ventilation settings (Code: IIID2b) [Difficulty: ELE: R, Ap; WRE: An]

b. Flow

Flow is adjusted to set the inspiratory time and the I : E ratio. In addition, flow is set to meet the patient’s needs. Inspiratory flow should be great enough to minimize the WOB. Increase flow if the patient has signs of greater demand, such as using accessory muscles of inspiration or lack of synchrony with the ventilator, or if the pressure manometer deflects greatly below the baseline pressure or shows a low initial increase in inspiratory pressure.

In addition, most current generation ventilators offer more than one inspiratory flow pattern (see Figure 15-1). The sine wave is most physiologically like a normal, spontaneous inspiration. The other waveforms can be compared with the sine wave to determine which one best meets the patient’s needs. Ideally, the best flow pattern is one in which the patient’s peak and mean airway pressures are lowest, exhalation is complete, breath sounds are improved bilaterally, heart rate and blood pressure are stable, and the patient feels most comfortable.

f. Modes of ventilation

The following modes of ventilation are delivered through most types of electrically powered and microprocessor-type volume-cycled ventilators.

1. Control

Control (C) is the simplest method of providing ventilatory support and is used on an apneic patient. The ventilator is set with a mandatory respiratory rate and tidal volume. The machine is incapable of allowing any patient interaction. For example, the ventilator might be set to deliver a tidal volume of 700 mL at a rate of 14 times/min. Because of this limitation, it is rarely, if ever, used in modern medicine except when the patient must be kept sedated or pharmacologically paralyzed (Figure 15-13 shows the pressure/time curve).

image

Figure 15-13 Pressure vs. time waveforms for various modes of mechanical ventilation. A, Control (C) mode shows no patient effort and consistent inspiration/expiration ratios. B, Assist/control (A/C) mode shows that patient’s initial effort triggers machine tidal volume breath. C, Intermittent mandatory ventilation (IMV) mode shows spontaneous tidal volume breaths occurring between predetermined machine tidal volume breaths. Note “stacked” breaths that happen when patient takes in a breath that then is supplemented by a machine breath. D, Synchronous intermittent mandatory ventilation (SIMV) mode shows that a patient effort within a time window results in delivery of machine tidal volume. Any other patient efforts within the time window result in a spontaneous tidal volume. If no patient efforts occur within the time window, machine tidal volume will be delivered automatically. E, Pressure support ventilation (PSV) mode shows how patient must initiate all breaths that then are supported to the predetermined airway pressure. Stable tidal volumes are seen if the patient inhales passively. Variably larger tidal volumes result if the patient inhales more actively. F, Positive end-expiratory pressure (PEEP) therapy can be added to A/C mode (as shown) or any other mode. The elevated baseline pressure prevents alveolar collapse. The sensitivity control must be set at −1 to −2 cm water so that the patient is able to trigger a breath without undue effort. G, Continuous positive airway pressure (CPAP) shows that the patient takes spontaneous tidal volumes while exhaling against an elevated baseline pressure.

5. Pressure support ventilation

Pressure support ventilation (PS or PSV) is similar to intermittent positive-pressure ventilation (IPPV) in that when the patient initiates a ventilator breath, a preset pressure is delivered to the airway. The patient has the flexibility to determine the respiratory rate. The physician orders a PSV level for one of two reasons, depending on the clinical goal. First, enough PS is ordered to overcome the patient’s calculated airway resistance. This usually is done when the patient has an increased work of breathing from a smaller than ideal endotrachal tube. Second, PS is ordered to deliver a targeted tidal volume. In most cases, the tidal volume will be stable if the patient passively takes the PSV breath, or it can be larger if the patient interacts actively with the pressure that is delivered (Figures 15-5 and 15-13, E show the pressure/time curve).

Because PS utilizes a set delivered pressure, the patient’s tidal volume can vary with changing lung compliance and/or airway resistance. To bring stability to the tidal volume under changing patient conditions, several ventilator manufacturers have developed automatic compensation systems. They are designed to provide additional volume if the patient’s lung compliance and/or airway resistance should worsen. Conversely, they will reduce automatically the set pressure when the patient is easier to ventilate, so that too large a tidal volume will not be delivered. All manufacturers have their own name for this compensation system and have developed different ways to accomplish the goal of a stable tidal volume. Examples include pressure augmentation on the Bear 1000, volume-assured pressure support on the Bird 8400, dual-control pressure ventilation on the Puritan Bennett 840, volume support on the Servo-i, and AutoFlow on the Dräger Evita 4. It is beyond the scope of this text to cover each of these. However, the practitioner is encouraged to understand their essential features and to know that they are added to the pressure support mode to ensure a stable, ordered tidal volume.

6. Pressure control ventilation

Pressure control ventilation (PC or PCV) involves the delivery of tidal volume breaths that are pressure limited and time cycled. A set ventilator rate can be set and the patient can trigger additional breaths. Because the pressure is limited, tidal volumes may vary. This must be monitored closely in patients with frequently changing lung compliance and airway resistance. As the inspiratory time is increased, it can become longer than the expiratory time. This results in pressure control inverse ratio ventilation (PCIRV). Figure 15-14 shows the volume, flow, and pressure tracings.

As was discussed earlier with the PS mode, the patient’s delivered tidal volume can vary with the PC mode when lung compliance and/or airway resistance is changing. Review the previous discussion on PS for the automatic compensation systems that manufacturers have added to their ventilators. Use of these ensures that the delivered tidal volume is stable in the PC mode despite changing patient conditions.

7. Airway pressure release ventilation

Several ventilator manufacturers offer the airway pressure release ventilation (APRV) mode (as well as other modes). These ventilators are microprocessor controlled and include a monitor for patient data and graphics. The APRV mode has been used with success in patients with ARDS who have not responded well to constant volume ventilation. APRV can be described simply as a mode in which the patient can breathe spontaneously at two different levels of CPAP. A difference from conventional CPAP is that the two levels are held for set periods. The ventilator options for this mode are simple. The practitioner sets the low pressure (Plow), the high pressure (Phigh), and the times that the patient will be at those pressure levels. Low pressure sometimes is referred to as CPAP and high pressure as release pressure. The timing changes from low pressure to high pressure and back to low pressure effectively deliver a tidal volume. (See Figure 15-15 for a pressure/time tracing.)

When APRV is compared with other modes of ventilation, several similarities are evident between it and PCIRV with PEEP and bilevel ventilation. All have an elevated baseline pressure that allows the patient to breathe spontaneously. All have variable inspiratory times for the higher pressure level. None delivers a set tidal volume to the patient. The only real difference seems to be that the patient can breathe spontaneously at the higher pressure level only with APRV. If the patient does not make any respiratory efforts, APRV functions like PCV or bilevel ventilation.

A set of blood gases, vital signs, pulmonary artery catheter values, and so forth should be obtained on the current constant volume ventilator settings as a baseline before APRV is started. The following suggestions for initiating APRV are similar to those listed earlier for starting PCIRV:

Monitor the patient’s ventilator delivered and spontaneous tidal volumes and rates. Check vital signs. Get an arterial blood gas (ABG) sample in about 15 minutes.

If the patient’s initial blood gas results on APRV show hypoxemia, the following options are available: (1) increase the inspired oxygen percentage if it is not already at 100%, (2) increase the low pressure level, (3) increase thehigh pressure level, or (4) increase the time the patient is kept at the high pressure level. Reducing any or all of these options will decrease the patient’s Pao2 if it is too high.

If the blood gas results show hypoventilation, the following options are available: (1) increase the high pressure level, or (2) decrease the time at the high pressure level to increase the respiratory rate. Do the opposite to increase the Paco2 if the patient is being hyperventilated.

As with the previous ventilator modalities, the patient should be monitored closely and should have an ABG drawn for evaluation of every change.

3. Begin and modify combinations of ventilatory techniques to oxygenate the patient adequately: synchronous intermittent mandatory ventilation, pressure support ventilation, pressure control ventilation, and positive end-expiratory pressure (Code: IIID2b) [Difficulty: ELE: R, Ap; WRE: An]

The current generation of mechanical ventilators offers the physician and the practitioner a number of options for how best to tailor ventilatory support to meet the patient’s needs. The following should be considered when one is deciding what modes to use and combine.

b. Hypercapnia

A patient may have hypercapnia (a high carbon dioxide level) because of sedation from a morphine or heroin overdose or may have COPD with worsening of the chronic hypercapnia. In either case, the patient becomes progressively more hypoxemic (unless given supplemental oxygen) as the carbon dioxide level increases. Control or A/C modes are best for setting a minimum minute volume to determine the maximum carbon dioxide level. As the patient recovers, SIMV or PSV allows the gradual reduction of ventilatory support. See Box 15-2 for indications of SIMV tolerance.

c. Hypoxemia

If hypoxemia is secondary to a decreased FRC, as in ARDS or atelectasis, the treatment of choice for hypoxemia is CPAP on a free-standing system or PEEP on a conventional volume-cycled ventilator. If the problem results from an increased intrapulmonary shunt, the patient may need PEEP or CPAP, as well as up to 100% oxygen. See Box 15-3 for patient monitoring during PEEP and CPAP. PCIRV and high-frequency ventilation (HFV) have been used with success in hypoxemic patients with a pulmonary air leak for whom conventional volume ventilation has failed.

Often, when a patient has more than one problem, more than one solution will be needed. The following are combinations of modes from which to choose.

1. Mandatory minute ventilation

Mandatory minute ventilation (MMV) is a relatively new variation on the SIMV mode. It has been used as a weaning mode that limits the increase in carbon dioxide if the patient should tire. With MMV, the patient is assured of a preset minute volume regardless of his or her spontaneous breathing. It has been proposed as an effective way to ventilate and wean patients who can breathe spontaneously but who have an unreliable respiratory drive and unstable tidal volume. Examples include patients who have received narcotic, sedative, anesthetic, or neuromuscular blocking medications. Patient conditions for which MMV is indicated include encephalopathy and cerebral disorders such as stroke. In addition, MMV may be used during the recovery period of a neuromuscular disease. Ventilators that include the MMV mode all are controlled by a microprocessor that monitors the ventilator’s and the patient’s tidal volume and rate. The following guidelines have been recommended for the initiation of MMV:

Ideally MMV establishes a minimum safe volume of ventilation. If the patient inhales less than this volume, the ventilator delivers as many breaths as necessary at the preestablished tidal volume to make up the difference. Be aware that a patient who is breathing rapidly with a small tidal volume may move enough gas to exceed the minimum minute volume. Because of this risk, it is important to set a low tidal volume alarm or a high respiratory rate alarm or both to give warning. Do not let the programming of an MMV create a false sense of security with these patients.

4. Choose and adjust the tidal volume for mechanical ventilation

A spontaneously breathing person exhales a large enough tidal volume (at the necessary respiratory rate) to remove carbon dioxide as fast as it is produced by his/her metabolism. This results in a normal carbon dioxide level and acid-base balance. Obviously, a person’s tidal volume and respiratory rate will vary considerably with activity level. A 70-kg (154-lb) adult with normal lungs and metabolism needs a tidal volume of about 7 mL/kg (3 mL/lb) of ideal body weight to remove carbon dioxide adequately. So, this person would have a spontaneous tidal volume of about 500 mL (70 kg × 7 mL/kg). The Radford nomogram can be used for predicting normal spontaneous tidal volumes and rates on the basis of body weight (Figure 15-18). It can be used to help establish an initial target for the ventilator tidal volume for many patients with normal lungs. However, because compressed volume within the ventilator circuit causes some of the set tidal volume to not reach the patient’s lungs, a larger set tidal volume usually is needed. (Review “Procedure for calculating the tubing compliance factor” earlier in the chapter.) In addition, patients with lung problems will have different tidal volume requirements. Following are the current tidal volume recommendations based on the patient’s pulmonary condition:

This might be a patient with a normal cardiopulmonary system who is receiving mechanical ventilation because of a neurologic problem. For example, the previously mentioned 70-kg (154-lb) adult with normal lungs needs a set mechanical ventilator tidal volume in the following range:

image

This might be a patient with ARDS who is at risk for pulmonary barotrauma/volutrauma. For example, the previously mentioned 70-kg (154-lb) adult with stiff lungs needs a set mechanical ventilator tidal volume in the following range:

image

This might be a patient with COPD or status asthmaticus who is at risk for air trapping on exhalation. For example, the previously mentioned 70-kg (154-lb) adult with overstretched lungs needs a set mechanical ventilator tidal volume in the following range:

image

Remember that the set tidal volume and measured exhaled tidal volume are greater than the patient’s actual tidal volume that reaches the lungs. This is because the compressed volume is lost in the circuit and does not go into the patient. So, in all of the above examples, the set tidal volumes are greater than the patient’s actual tidal volume.

It is common practice to get a set of ABG values after the patient is stable on the ventilator. The tidal volume can be adjusted within the range depending on whether the patient’s arterial CO2 pressure (Paco2) value is too high or too low for the therapeutic goal. The most direct way to change alveolar ventilation is to modify the delivered tidal volume. If everything else remains the same, a larger tidal volume results in a lower Paco2 value. Conversely, although everything else remains the same, a smaller tidal volume results in a higher Paco2 value.

The following formula can be used to help predict what tidal volume produces a desired Paco2 value:

image

in which

Note that other, simpler formulas are available for calculating a change in minute volume or tidal volume. This one is presented because it takes into account more factors and can be used to calculate a change in tidal volume, rate, or mechanical dead space.

5. Choose and adjust the rate for mechanical ventilation

See Table 1-2 for a listing of the normal resting respiratory frequencies based on age. If the patient is apneic and has a normal temperature and an appropriately set tidal volume, respiratory rates in the indicated ranges will produce a normal Paco2 level. This must be confirmed by ABG measurements. If the tidal volume cannot be changed, adjusting the respiratory rate will modify alveolar ventilation. A higher respiratory rate, with everything else remaining the same, will result in a lower Paco2 level. Conversely, a lower respiratory rate, with everything else remaining the same, will result in a higher Paco2 level.

See Figure 15-18 for the Radford nomogram for use in predicting a normal respiratory rate and tidal volume based on weight. It can be used to establish an initial rate for most patients. As was mentioned earlier, chronically hypercapneic patients must be ventilated with some caution. Giving this type of patient a higher ventilator-delivered rate and a larger tidal volume may result in blowing off too much carbon dioxide and may cause a respiratory alkalosis. Adult patients with severe chronic restrictive lung disease or those who have had a pneumonectomy may need respiratory rates of 20 to 30 per minute or greater to meet their minute volume needs because their delivered tidal volume must be smaller than normal owing to their condition.

The same formula that was used to predict a tidal volume change can be used to help predict what respiratory rate will produce a desired Paco2 value:

image

6. Choose and adjust the minute ventilation for mechanical ventilation

The subjects of minute ventilation and alveolar minute ventilation were covered in Chapter 4. Review the calculations as needed. Blood gases must always be evaluated for the Paco2 level to tell whether the patient’s minute ventilation is adequate. A high carbon dioxide level indicates a need to increase the tidal volume, respiratory rate, or both. A low carbon dioxide level indicates a need to decrease the tidal volume, respiratory rate, or both. In both cases, the key to modifying the carbon dioxide level is to modify the alveolar ventilation. This is best accomplished by changing the tidal volume rather than the rate. The following formula can be used to calculate a change in the minute volume:

image

in which

image

image

7. Initiate and adjust positive end-expiratory pressure therapy (Code: IIID2d) [Difficulty: ELE: R, Ap; WRE: An]

Positive end-expiratory pressure (PEEP) is a mechanically elevated baseline pressure. In other words, the patient’s end-expiratory pressure is above atmospheric. PEEP is administered through a mechanical ventilator and is not a mode by itself. Rather, it is used in conjunction with any of the previously mentioned modes.

PEEP generally is indicated in any acute, bilateral, generalized pulmonary condition in which the FRC is decreased. When the FRC is decreased, shunt is increased, and the patient has refractory hypoxemia. Examples of small FRC conditions helped by PEEP include generalized atelectasis, pulmonary edema, ARDS, and infant respiratory distress syndrome (RDS). All of these patients show decreased lung compliance as measured by their static compliance (Cst). The higher the level of PEEP, the more progressively the patient’s FRC is increased. The therapeutic goal of this is to increase the patient’s arterial oxygen pressure (Pao2). (Figure 15-13, F shows the pressure/time curve.) Patients with chronically small FRC, such as those with pulmonary fibrosis and kyphoscoliosis, are not helped by the application of PEEP.

Specific indications for PEEP include the following:

Before PEEP is begun, the patient should be monitored carefully to establish the baseline condition. The same parameters should be monitored after each change in the PEEP level to determine how the patient is tolerating it. The best or optimal level of PEEP is the level that results in the best delivery of oxygen to the tissues (not necessarily the arterial blood). Often, a secondary goal is to reduce the inspired oxygen to a safe level. The patient is at risk for oxygen toxicity if more than 50% oxygen is inhaled for longer than 48 to 72 hours. See Box 15-3 for recommendations on what to monitor during the application of PEEP and how to evaluate the data.

The application of PEEP has risks. Clinically, these risks must be weighed against the potential benefit to the patient. In the profoundly hypoxic patient, PEEP can be lifesaving. Some clinicians use low levels of PEEP (up to 5 cm water) in patients with normal lungs or overly compliant lungs (emphysema) to maintain the baseline level of FRC. Hazards of PEEP include the following:

a. Increasing PEEP

PEEP therapy usually is begun at initial levels of 2 to 5 cm water. After the patient’s response has been determined, 2 to 5 cm more PEEP may be applied. The patient is reevaluated. This process goes on until the desired clinical benefit is reached. See Figure 15-19 for a number of physiologic parameters that can be measured and evaluated.

Different approaches to the application of PEEP are used to find the best level. One approach could be called minimum PEEP. It involves the application of PEEP to the minimum level that allows the inspired oxygen percentage to be lowered to a safer level. A clinical goal is to minimize the risk of oxygen toxicity. In this approach, PEEP is raised until the Pao2 is greater than 60 torr or the Spo2 is greater than 90% on 60% oxygen or less. Usually no more than 10 to 15 cm water of PEEP is needed.

Another approach could be called best PEEP or optimum PEEP. This approach has the clinical goal of reducing the patient’s shunt fraction to less than 15%. Often this requires more pressure than the minimal PEEP approach. Because this higher pressure level is more likely to cause hemodynamic problems, the patient should have a pulmonary artery catheter inserted. With it, the patient’s cardiac output, mixed venous oxygen level, pulmonary capillary wedge pressure, and pulmonary vascular resistance can be measured (see Figure 15-19). In addition, the patient may need increased intravenous fluids, dopamine (Intropin), and digitalis (Digoxin) for cardiovascular support. The higher PEEP levels increase the risk of pulmonary barotrauma. Therefore the patient must be watched closely for signs of a pneumothorax.

MODULE C

Unconventional mechanical ventilation is defined here as the use of specialized ventilators that can provide unique modes and options needed by special populations of patients. This ventilatory support may or may not be provided through an endotracheal tube or a tracheostomy tube.

1. Initiate high-frequency ventilation and select appropriate settings

High-frequency ventilation (HFV) is needed whenever the patient’s condition calls for a higher respiratory rate or a smaller tidal volume than usually is delivered on a conventional volume-cycled ventilator. A high-frequency ventilator can deliver a respiratory rate far higher than the limit of 150/min set by the U.S. Food and Drug Administration (FDA) on all conventional adult and neonatal/pediatric ventilators.

The FDA has approved HFV for use on adults during bronchoscopy and laryngoscopy procedures and when a patient with a bronchopleural fistula cannot be managed on a conventional ventilator. Although patients with ARDS have not been officially approved for HFV, the procedure has been used when a patient is hypoxic despite maximum settings on a conventional ventilator. (Box 16-5 lists clinical uses for HFV with infants and children.)

Currently HFV can be delivered in three different ways. The first involves the use of a conventional ventilator set at a rate of up to the FDA maximum of 150/min. This method is called high-frequency positive-pressure ventilation (HFPPV). With it, the set tidal volume is decreased to something less than standard (<10 mL/kg). The second involves the use of high-frequency jet ventilation (HFJV) and a special endotracheal tube with a standard lumen and an additional small-diameter air entrainment lumen (Figure 15-20). The HFJV unit is connected to the air entrainment lumen. Gas from the HFJV unit entrains other gas through the main lumen to create the patient’s tidal volume. HFJV units can deliver a small tidal volume several hundred times per minute. The third method involves a high-frequency oscillation (HFO) ventilator. HFO makes use of a conventional endotracheal tube (as does HFPPV). However, the delivered tidal volume is the smallest, and the respiratory rate can be the fastest of all three methods. Table 16-2 presents a comparison of all three methods. The equipment used for HFJV and for HFO is described later in this chapter. Table 15-1 lists considerations for the initial settings and for adjustment of HFV delivered to infants and adults by a jet ventilator or an oscillator ventilator.

Current clinical experience is recommended for any HFV method. Although adult patients have been treated successfully with HFV, far greater use of these techniques has occurred with infants and children. Therefore, Chapter 16 provides further discussion.

2. Independent (differential) lung ventilation

Independent lung ventilation (ILV) involves the use of a separate mechanical ventilator for each lung. A double-lumen endotracheal tube must be placed into the patient to allow this procedure. (See Figures 12-36 and 12-37 and the related discussion.) Box 15-4 lists indications for double-lumen endotracheal tubes and ILV. In all cases, the patient has one normal lung and one abnormal lung. Overriding concerns with ILV are to ventilate the patient adequately through the normal lung and to allow the injured lung to heal.

A common initial approach is to select two identical ventilators that allow synchronization of the patient’s respiratory rate. The Servo and Dräger ventilators are suited for this. They allow one unit to be designated the “primary” ventilator to set the respiratory rate for it and the “secondary” unit. Each unit then can have the same mode and I : E ratio. This synchronized ILV method still allows the independent setting of tidal volume, oxygen percentage, and PEEP for each lung.

For example, an 80-kg adult man normally receives an initial tidal volume of about 800 mL (10 mL × 80 kg). If both lungs functioned normally, each would receive 400 mL. However, with unilateral lung disease, the bad lung receives little tidal volume and the normal lung receives too much and becomes overdistended. It is therefore important to set the initial tidal volume to the good lung at half the normal volume for both lungs. For this patient example, the normal lung’s initial tidal volume should be set at 400 mL (5 mL × 80 kg). To avoid changing too many parameters at once, the same oxygen percentage and PEEP level as originally set are kept. The abnormal lung is ventilated on the basis of its pathologic condition or may be left unventilated.

After 15 minutes, check the patient’s ABG values. Depending on the results, adjust the ventilator settings for the good lung as would be done typically to remove carbon dioxide and maintain oxygenation. The ventilator settings for the diseased lung must be adjusted carefully to allow healing and prevent complications such as atelectasis and pneumonia.

When the patient has a bronchopulmonary fistula, the air leak through the bad lung can be so great that an HFV must be used rather than a conventional ventilator. In this situation, no synchronization of rate or any other parameters can be set. The good lung is ventilated conventionally to maintain the patient’s blood gas values. The HFV is set to provide some support with a small tidal volume and low ventilating pressures. The clinical goal is to prevent excessive lung pressure so that the lung tear heals.

The patient can be converted back to breathing through one conventional ventilator when normal functioning returns to the injured lung. This can be done when the peak pressure, plateau pressure, and mean airway pressure for both lungs are about the same. When these values match or are close, similar lung compliance and airway resistance values are indicated. The first step in converting from ILV to conventional ventilation involves combining the proximal ends of the double-lumen endotracheal tube with a Y adapter. When this is done, one ventilator can deliver tidal volume breaths to each lung. It is suggested that the PC mode be used so that excessive pressure is not applied to the healing lung. Set the peak pressure at or just below the previous peak pressure used with the healing lung. Set the PEEP level at the previous pressure used with the healing lung. Check the patient’s ABG values after 15 minutes. Be prepared to adjust the conventional ventilator as needed to get the desired blood gas values. Watch for problems with the healing lung, and be prepared to go back to independent lung ventilation if necessary. When it appears certain that the patient is tolerating conventional ventilation, the double-lumen endotracheal tube should be removed and replaced with an appropriate single-lumen tube. This allows better suctioning and results in less airway resistance through the tube. Wean and extubate the patient when appropriate.

3. External negative-pressure ventilation

These ventilators have proved useful in patients with the following characteristics: (1) normal, intact upper airway, (2) ability to swallow, (3) normal airway resistance, (4) normal lung/thoracic compliance, and (5) ability to ventilate until respiratory muscle fatigue becomes too great.

Patients with the following disease conditions have been ventilated successfully by a negative-pressure ventilator: (1) neuromuscular defects such as poliomyelitis, postpolio syndrome, muscular dystrophy, and high spinal cord injury; (2) kyphoscoliosis with resulting restrictive lung disease; and (3) COPD during an acute worsening.

Negative-pressure ventilators work by creating negative pressure around the patient’s whole body or over the anterior chest and abdomen. The negative pressure expands the thorax, and a tidal volume is inhaled. If the patient needs supplemental oxygen, it must be given by nasal cannula or face mask. Three basic types of external negative-pressure ventilators are available: Drinker body respirator (so-called “iron lung”), body wrap, and chest cuirass (Figure 15-21).

The following steps are used to initiate ventilation:

4. Initiate and adjust noninvasive ventilation (Code: IIID2c) [Difficulty: ELE: R, Ap; WRE: An]

Noninvasive ventilation (NIV) is also known as noninvasive positive-pressure ventilation (NPPV). These patients will be ventilated without the use of an endotracheal tube. Most of the time, patients receiving NIV are ventilated with the aid of a nasal mask similar to that used to deliver mask CPAP. A full face mask often is needed if the patient leaks through the mouth with a nasal mask.

Most patients who are candidates for NIV are fairly stable, are not intubated, can do some spontaneous breathing, and will need only short-term ventilator support. Examples include patients with the following conditions:

If the patient is critically ill, is unstable, must be intubated to secure the airway for secretion removal, needs a high level of therapeutic PEEP to maintain the FRC, or experiences a combination of these, he or she should be placed on a standard volume-cycled ventilator.

Often patients receiving NPPV are ventilated with two different levels of positive pressure. This is referred to as bilevel ventilation. The baseline pressure is greater than zero for setting a CPAP or PEEP level. The peak pressure is set to deliver a desired tidal volume (similar to PS ventilation). Both levels can be adjusted independently. If only the baseline pressure is elevated, the patient is receiving CPAP. If only the peak pressure is elevated, the patient is receiving PS ventilation. Respironics (Carlsbad, CA) has pioneered the development of ventilators for conventional ventilation or noninvasive bilevel ventilation.

As was stated above, the patient must have a properly fitting nasal or face mask to receive noninvasive ventilation. These ventilation masks are similar to CPAP masks and are referred to as such. CPAP masks are available in different sizes for children older than 3 years and for adults. Nasal masks are designed to cover only the nose. They allow the patient to eat, drink, speak, and use the mouth as a second airway for breathing in case a malfunction of the CPAP system occurs. The mouth also acts as a pressure relief route should the CPAP pressure become too great. Pressures of up to 10 to 15 cm water usually can be maintained (Figure 15-22). Usually, the mask is made of a transparent plastic. Face masks are designed to cover the nose and mouth. They are similar in design to the masks used during bag/mask ventilation and also are made of a transparent plastic. The face mask must be used if the patient has persistent mouth breathing and cannot use a nose mask. With a good seal, pressures of greater than 15 cm water can be maintained.

In recent years, a wide variety of NIV mask systems have been developed (Figure 15-23). In any situation, the clinical goal is to find a CPAP/NIV mask with a soft, very compliant seal that closely fits the contours of the patient’s face. A strapping system is needed to hold the mask in place. Too large a mask will not seal and will allow gas to leak and pressure to decrease. The patient may show increased snoring or airway obstruction with periods of apnea. A mask that is too small or misfitting can cause an uneven distribution of pressure on the face. This can lead to abrasions or pressure sores and ulcers on the face.

The bilevel settings must be determined at the bedside by asking for the patient’s subjective opinion, listening to breath sounds, checking vital signs, and evaluating ABG values. If supplemental oxygen is needed, it can be added at a port on the patient’s mask, at the humidifier, or at the outlet from the BiPAP unit. Up to 15 L/min can be added without affecting the performance of the BiPAP system. It is not possible to know the delivered oxygen percentage until after the bilevel ventilation settings have been determined. The oxygen flow then should be increased gradually while the patient’s Spo2 value increases to the desired saturation level.

The function of the Respironics Esprit system is reviewed briefly. The operator can choose from two modes of operation and can select from the following:

When bilevel ventilation is started, set the EPAP level, if needed, to elevate the patient’s baseline pressure. The EPAP level establishes the patient’s functional residual capacity (FRC) to improve oxygenation. Increase or decrease EPAP as you would adjust PEEP or CPAP. Review Box 15-3 for patient monitoring with EPAP. Next, set the IPAP level to achieve the desired tidal volume.

With bilevel ventilation, the difference between IPAP and EPAP is called pressure boost and delivers the tidal volume. If a larger tidal volume is needed, increase the IPAP level. Conversely, decrease the IPAP level to obtain a smaller tidal volume. It is important to remember that the delivered tidal volume varies depending on changes in the patient’s airway resistance and lung/thoracic compliance, as well as in the machine settings.

5. Initiate and adjust continuous positive airway pressure (Code: IIID2d) [Difficulty: ELE: R, Ap; WRE: An]

CPAP is a pressure above atmospheric that is maintained at the airway opening throughout the respiratory cycle during spontaneous breathing. CPAP is similar to PEEP in purpose and effect. Remember that with CPAP, the patient does not receive any ventilator-delivered tidal volume breaths. The patient must be capable of providing all ventilation for carbon dioxide removal. Although CPAP usually is delivered through the mechanical ventilator, the respiratory rate is turned off. However, alarm systems still are functioning for patient safety. Some hospitals will make use of a free-standing system for delivering CPAP. This system is discussed below (Figure 15-13, G shows the pressure/time curve for CPAP).

Before a patient receives CPAP therapy, the practitioner and the physician must be assured that the patient has the ability to breathe adequately to eliminate carbon dioxide. CPAP is contraindicated in an apneic patient or in one who may become apneic. (This patient must be fully supported on the ventilator.) The patient must have an adequate respiratory rate, tidal volume, and minute volume. The heart rate and blood pressure should be stable. Maximum inspiratory pressure and vital capacity values may be acceptable or low. Blood gas analysis typically shows refractory hypoxemia but a normal or low Paco2 level. This shows that the patient would benefit from an elevated baseline pressure to increase the FRC but is capable of ventilating.

CPAP involves the same indications, hazards, and patient evaluation processes as were discussed earlier in PEEP therapy. One possible physiologic benefit of CPAP over PEEP is that less reduction in venous return to the heart is seen. This occurs because with CPAP, the patient is breathing spontaneously. Therefore patients treated with CPAP may be able to tolerate higher pressure levels than those being ventilated with PEEP therapy.

CPAP usually is increased and decreased in steps with 2 to 5 cm water. As with PEEP, the patient is evaluated before CPAP is begun and again after each pressure change is made. See Box 15-3 for recommendations on what to monitor during the application of CPAP and how to evaluate the data.

The patient must be monitored carefully for fatigue because the patient is providing all of the minute ventilation. Signs of fatigue include increasing respiratory rate, decreasing tidal volume or vital capacity, decreasing maximum inspiratory pressure, and tachycardia. Blood gas measurement may show a stable or decreasing Pao2 value. An increasing Paco2 level is a definite sign of fatigue. The patient may complain of dyspnea. The practitioner may notice that the patient is working harder than normal to breathe, as shown by increased use of the accessory muscles of ventilation and heavy perspiration. When these signs occur, CPAP therapy should be discontinued and mechanical ventilation instituted in a mode that best fits the patient’s needs.

MODULE D

Note: The literature produced by the manufacturers and the descriptions used in many standard texts break down the various ventilators into more categories than are used by the NBRC. To avoid confusion, this text uses the more simplified terminology of the NBRC.

A pneumatically powered ventilator is defined here as a ventilator powered by compressed gas. Older units operate without any electrically powered control systems (electrically powered alarm systems may or may not be added). More recent units will also have electrically powered controls and alarm systems. An electrically powered ventilator is defined here as a ventilator that is electrically powered or controlled. Most volume-cycled ventilators fall into this category. Microprocessor ventilators are electrically powered but are controlled by one or more microprocessors (computers). Many of the most current volume-cycled ventilators have microprocessors to control their functions. Fluidic ventilators typically make use of electrical circuits with flip-flops to respond to changes in gas flow and pressure throughout the system. Fluidic ventilators are powered by compressed gas. Noninvasive ventilators are designed for home use or short-term hospital use and are electrically powered and controlled. They have fewer controls and alarms than hospital-based critical care ventilators. A nasal or full face mask, rather than an endotracheal tube, is used to attach the ventilator to the patient. High-frequency ventilators are used in a limited population of critically ill patients who are doing poorly despite all attempts at conventional mechanical ventilation. These units are designed to deliver very rapid respiratory rates and very small tidal volumes.

1. Manipulate pneumatic ventilators by order or protocol (Code: IIA6a) [Difficulty: ELE: R, Ap; WRE: An]

b. Put the equipment together and make sure that it works properly

Refer to Figures 14-10 and 14-11 in Chapter 14 for the IPPB circuits. The ventilator and circuits are similar to those used in IPPB therapy with the following exceptions:

2. Manipulate electrical ventilators by order or protocol (Code: IIA6a) [Difficulty: ELE: R, Ap; WRE: An]

3. Manipulate microprocessor ventilators by order or protocol (Code: IIA6a) [Difficulty: ELE: R, Ap; WRE: An]

6. Manipulate high-frequency ventilators by order or protocol (WRE code: IIA6c) [Difficulty: WRE: R, Ap]

a. High-frequency jet ventilators (HFJVs)

1. Get the necessary equipment for the procedure

The Bunnell Life Pulse High-Frequency Jet Ventilator (Bunnell Inc., Salt Lake City, UT) is designed for use with neonatal patients with infant respiratory distress syndrome (RDS) who have failed with conventional ventilation. The ventilator is unique in that it must be used in conjunction with a conventional ventilator that is placed in the CPAP or SIMV mode during use of the jet.

In addition, a special endotracheal tube must be placed into the patient before HFJV can be initiated. Ideally, the patient is intubated with a triple-lumen Hi-Lo Jet tube (see Figure 15-20). If the neonate has already had a standard endotracheal tube inserted, the Bunnell LifePort Endotracheal Tube Adapter (Bunnell Inc.) can be substituted for the standard adapter (Figure 15-25). Both tubes now allow the jetted gas from the Bunnell ventilator to go to the patient. The traditional ventilator is attached to the main lumen of the endotracheal tube. Based on the physical principles that govern jets, additional gas is entrained through the main lumen. This entrained gas should be humidified if the jet gas is dry. All exhaled gas passes out through this main lumen, where it can be measured through the traditional ventilator’s spirometry system. This ventilator’s alarm systems also can be used, and SIMV breaths and PEEP can be added if needed.

7. Manipulate continuous mechanical ventilation and noninvasive ventilation breathing circuits by order or protocol (ELE code: IIA11a and IIA11d) [Difficulty: Ap, An]

a. Get the necessary equipment for the procedure

A permanent or a disposable circuit may be selected, based on the type of ventilator on which it must be placed. A circuit with an external exhalation valve must be used with older ventilators such as the Bennett MA-1 (Covidien-Puritan Bennett) and the Bear 2 (Cardinal Health, Viasys Bear Medical Systems, Palm Springs, CA). In addition, if a Bird-series unit is used as a ventilator, its circuit has an external exhalation valve (see Figure 15-24, A). All modern electrical and microprocessor ventilators feature an internal exhalation valve and do not need one included in the circuit (see Figure 15-24, B). If the patient must receive aerosolized medications, the circuit should include a nebulizer or should be able to accept one. If not included, the nebulizer or metered-dose inhaler adapter must be added into the inspiratory limb of the circuit (Figure 15-27).

Also consider whether it is better to use an unheated or heated circuit. Usually the circuit is unheated. With these, a cascade-type humidifier or an HME is used to warm and humidify the inspired gas. However, some practitioners prefer to use a heated circuit for the care of neonates. These circuits may have heated wires loosely running through the lumen of the tubing or may have a wire embedded within the tubing itself. A heated-wire circuit offers finer control over the temperature of the inspired gas and minimizes condensation. Follow the manufacturer’s guidelines to make sure that the system can adequately humidify the minute volume that is being used.

Noninvasive ventilators, such as the Respironics series, have specific circuits designed only for the unit. Traditional ventilator circuits cannot be placed on a noninvasive ventilator.

d. Independently change the patient’s ventilator circuit as needed (ELE code: IIIF2i10) [Difficulty: R, Ap, An]

A circuit must be replaced if it is damaged in a way that prevents the patient from being ventilated. This is seen most commonly in circuits with external exhalation valves (see Figure 15-24A). If the balloon-type valve is damaged and will not close, the tidal volume will leak out rather than enter the patient. The circuit must be replaced. If the expiratory valve line is pulled off at the ventilator or exhalation valve, it must be replaced or the valve will not close.

The American Association for Respiratory Care (AARC) Clinical Practice Guideline on ventilator circuit changes (1994) included the following recommendations:

More recently, the AARC Clinical Practice Guideline on care of the ventilator circuit and its relation to ventilator-associated pneumonia (2003) included the following guidelines:

It is reasonable to expect the 2003 guidelines to be tested by the NRBC.

8. Manipulate ventilator breathing circuits: PEEP valve assembly, by order or protocol (ELE code: IIA11c) [Difficulty: ELE: R, Ap, An]

9. Manipulate continuous positive airway pressure (CPAP) systems: breathing circuits, by order or protocol (ELE code: IIA11c) [Difficulty: ELE: R, Ap, An]

b. Put the equipment together and make sure that it works properly

Figure 15-28 shows the typical components used in a free-standing CPAP breathing circuit. Components include the following:

The CPAP level is adjusted by means of a variety of devices collectively called threshold resistors, which include the following:

Air exits through each of these devices when it is working properly. All CPAP systems must be adjusted by checking the pressure level on the manometer. Set the low pressure or disconnection audible alarm to sound at a few centimeters below the CPAP level. For example, if 10 cm CPAP is ordered, set the alarm to sound if the pressure drops below 8 cm of CPAP.

c. Troubleshoot any problems with the equipment

Flow through the CPAP breathing circuit must be sufficient to meet the patient’s needs. Adjust the flowmeter setting and clamp on the anesthesia bag so that it is somewhat inflated, with excess air escaping out past the clamp. With all of the devices, gas escapes through the path of least resistance. All or some may escape through the anesthesia bag, the CPAP device, or both. The bag should collapse somewhat during the patient’s inspiration and expand somewhat during the expiration. The CPAP level should not decrease by more than 1 or 2 cm from baseline during an inspiration.

Make sure that the water level is maintained properly in the humidifier. Fill it with sterile, distilled water as often as necessary.

A sudden decrease in the CPAP level to zero indicates a disconnection at the patient or somewhere in the breathing circuit. Check all connections, and reassemble the break. The patient may have to be ventilated manually while the problem is corrected.

If the CPAP level decreases by more than 2 cm water during an inspiration, the flow is inadequate and should be increased. Flow also is inadequate if the patient shows increased use of accessory muscles of respiration or complains of increased WOB. Too high a flow is seen by an inadvertently high level of CPAP or the patient complaining of difficulty exhaling.

Water column systems must be monitored frequently because of water loss caused by evaporation. The actual CPAP level is progressively less than desired as the water is lost gradually. This system must have water added to it regularly or must have the expiratory tubing inserted deeper to keep the desired CPAP level.

The ball-bearing resistor system must be mounted vertically for gravity to keep the desired weight against the circuit. If it falls over and is horizontal, the CPAP pressure will be lost.

10. Manipulate CPAP systems: masks, nasal, and bilevel, by order or protocol (Code: IIA2) [Difficulty: ELE: R, Ap; WRE: An]

a. Get the necessary equipment for the procedure

A CPAP mask and breathing circuit are used primarily for patients who have obstructive sleep apnea. CPAP, by means of the mask, forces soft tissues open to the point that the airway is never obstructed (see Figure 15-22). The patient now is able to sleep normally and remain oxygenated. The patient should have the CPAP mask, breathing circuit, and proper CPAP level determined by a sleep study in the hospital. The patient can use the system at home once it is set up properly and he or she has been trained in its use.

In recent years, a CPAP mask has been used with a noninvasive mechanical ventilator to assist temporarily in the breathing of a patient with respiratory distress. The hope is to support the patient’s breathing long enough to treat the underlying problem(s). If this is successful, the patient does not need to be intubated. These patients require careful assessment and monitoring.

The two main categories of CPAP masks come in different sizes for children older than 3 years to adults. (See Figure 15-23 for examples.) Nasal mask and pillow systems are designed to cover only the nose. These allow the patient to speak and offer the mouth as a second airway for breathing in case a malfunction of the CPAP system occurs. The mouth also acts as a pressure relief route if the CPAP pressure should become too great. Pressures of up to 15 cm water usually can be maintained in an adult. Pressures of up to 10 cm water usually can be maintained in a child.

Full face mask and total face mask systems are designed to cover the nose and mouth. These are transparent and are similar to the mask used during bag/mask ventilation. The face mask must be used if the patient has persistent mouth breathing and cannot use a nose mask. With a good seal, pressures of greater than 15 cm water can be maintained.

All types of CPAP masks have a soft, very compliant seal to closely fit the contours of the face. Straps are needed to hold the mask in place. It is imperative that the mask properly fit the patient’s face.

b. Put the equipment together and make sure that it works properly

Several companies manufacture mask CPAP systems for home care. These are relatively simple circuits that do not have a humidification system or the other attachments seen in the hospital. Check the manufacturer’s literature for specific directions on their application to the patient. As is shown in Figure 15-23, the straps must be tight enough to seal the mask to the face but not so tight as to cut off circulation to the skin. Any mask CPAP system must be able to generate enough flow to meet the patient’s minute volume and peak flow needs. The CPAP level must be stable throughout the breathing cycle.