Home Respiratory Care and Mechanical Ventilation

Published on 01/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 1221 times

Home Respiratory Care and Mechanical Ventilation

Overall Goals of Home Respiratory Care

II Home Care Coverage: Financial Considerations

III Primary Forms of Home Respiratory Care

IV Oxygen Therapy

Indication: Any cardiopulmonary disease resulting in significant, chronic hypoxemia in the patient’s chronic stable state.

Chronic hypoxemia in the home care setting is defined by Health Care Financing Administration (HCFA) reimbursement criteria and is scientifically supported by the Nocturnal Oxygen Therapy Trial (NOTT study) and the British Medical Research Council Working Party study.

Reimbursement criteria include each of the following

1. PO2: ≤55 mm Hg (group I)

2. PO2: 56 to 59 mm Hg (group II) with

3. Hemoglobin saturation: ≤88% (group I)

4. Hemoglobin saturation: 89% (group II), and

5. Criteria for activity-based oxygen desaturation

6. Criteria for sleep-induced oxygen desaturation

Because the relationship between PO2 and hemoglobin saturation is variable and affected by a number of factors, criteria for home oxygen therapy generally should be based on PO2, not on HbO2%, as depicted in Figure 32-1, where some nonqualifying HbO2% are a result of qualifying PO2.

Physiologic effects of long-term oxygen therapy

Oxygen containment systems used in the home (see Chapter 34 for details)

1. System choice is commonly based more on cost versus reimbursement issues than on clinical efficacy; nonetheless the system chosen for a specific patient should satisfy the domicilary and mobility needs. Each system has major advantages and disadvantages (Table 32-1).

TABLE 32-1

Advantages Versus Disadvantages of Various Oxygen Delivery Systems

  Advantages Disadvantages
Cylinder oxygen Fixed finite source of oxygen Heavy/bulky
  May be stored indefinitely Represents the smallest content
  Large liter flows transiently available Requires frequent redeliveries
  100% oxygen source High pressure 2000-3000 psi
  Quiet oxygen delivery Represents weight and high-pressure danger
Liquid oxygen Patient can transfill portable Evaporation/flash loss
  Large content 860 times compressed Requires frequent redeliveries
  Liter flow limited to 8-10 L/min May freeze, interrupting O2 flow indefinitely
  Low pressure <20 psi Represents thermal hazard
  100% oxygen source Requires backup cylinder O2
  Quiet oxygen delivery  
Concentrator oxygen Infinite source of oxygen Consumes patient electricity
  Simplicity of operation Noisy
  Least supplier intrusion Potential of mechanical failure
    Requires patient/supplier maintenance
    Liter flow limited to 5-6 L/min
    Requires backup cylinder O2
    85-98% O2 dependent on liter flow
    Represents electric shock hazard

2. Oxygen concentrators are most commonly used when the patient is essentially homebound for the majority of time. This system provides an infinite amount of oxygen supply with proper maintenance. However, being electrically powered, concentrators should have a backup source of oxygen for power and/or equipment failures. Oxygen cylinders typically provide backup oxygen in the amount to cover three times the supplier’s average response time. With the concentrator there exists the additional cost of electricity, which is incurred by the patient and can be substantial. Specialty concentrators are manufactured that allow the user to slowly transfill aluminum portable cylinders. Portable concentrators with internal battery power and integral oxygen conservers also have recently been introduced. Further a single manufacturer has brought a concentrator capable of delivering 10 L/min to market.

3. Liquid oxygen systems are especially useful for travel, work, and patients with consistent out-of-the-home mobility needs. Liquid oxygen systems represent a large finite source of oxygen with features allowing patient transfilling of portable units. Liquid oxygen units should have a backup source of oxygen for equipment failures or complete depletion of the liquid oxygen contents. Oxygen cylinders typically provide backup oxygen in the amount to cover three times the supplier’s average response time. This finite source of oxygen, which is continuously being consumed, requires weekly (or more often) replenishment by the oxygen supplier.

4. Oxygen cylinders are the least commonly used, are the most labor intensive, and are the most expensive. However, cylinders represent a small finite source of oxygen that can be stored indefinitely, hence making them ideal backup sources of oxygen for the home care patient. Smaller cylinders are commonly used to satisfy the periodic mobility needs of patients using an oxygen concentrator for their primary oxygen source. Cylinders made of composite light-weight resin material that may be filled to 150% the normal filling pressure (3000 psi) are available and represent a significant increase in duration and portability (Table 32-2).

TABLE 32-2

Comparison of Available Portable Oxygen Delivery Systems

Portable Device Weight Gas Contents Duration
Aluminum M-6 or “B” cylinder* 2.2 lb 164 L 42 min
Aluminum M-9 or “C” cylinder* 3.7 lb 255 L 2 hr 7 min
Aluminum M-15 or “D” cylinder* 5.3 lb 425 L 3 hr 30 min
Aluminum M-22 or “E” cylinder* 7.9 lb 680 L 5 hr 40 min
Composite 3000 psi M-9 “B” cylinder size* 1.7 lb 246 L 2 hr 3 min
Composite 3000 psi M-15 “C” cylinder size* 2.6 lb 410 L 3 hr 23 min
Composite 3000 psi M-22 “D” cylinder size* 3.2 lb 640 L 5 hr 20 min
Small representative LOX portable 5.5 lb 542 L 4 hr 31 min
Large representative LOX portable 8.5 lb 1058 L 8 hr 50 min

image

*Require cylinder valve stem and regulator weighing 1.75 lb.

@ 2 L/min liter flow.

5. Some patients use a combination of oxygen concentrator for their stationary oxygen source and a portable liquid oxygen unit (transfilled from a stationary liquid oxygen reservoir) for their portable oxygen needs. Note that third-party payers will only pay for one stationary oxygen source under most circumstances.

Oxygen therapy equipment (see Chapter 34)

1. The simple cannula is most common; however, smaller-sized, low-flow cannulas have been used by long-term users of oxygen because of their comfort and inconspicuous appearance. Such cannulas are reserved for liter flows of ≤2 L/min.

2. The simple cannula is usually supplemented with a 25- to 50-foot oxygen extension to provide mobility from the stationary oxygen source. The 50-foot extension is the longest recommended by oxygen equipment manufacturers.

3. Oxygen-conserving devices are used for normal mobility, long-distance travel, or high oxygen demand.

a. Oxygen reservoir systems function by increasing the anatomic reservoir available for oxygen accumulation.

b. Pulse dose or demand systems deliver a predetermined dose of oxygen at the initial portion of inspiration on every- or multiple-breath variations. Such devices are either independent or integral to the oxygen equipment.

c. Transtracheal oxygen is useful in continued work or other activities away from home or for those with hypoxemia refractory to conventional home oxygen system limits.

Monitoring home oxygen therapy

1. Patients who meet the criteria for oxygen therapy at discharge frequently do not require oxygen after a few months as a result of

2. Monthly monitoring of status should occur in the home.

3. Office follow-up evaluation: Frequency depends on the severity of the overall disease process.

Complications/safety considerations

Aerosol Therapy

Patients with an intact airway and thick, tenacious secretions (e.g., cystic fibrosis, bronchiectasis) may require intermittent administration of aerosol therapy via pneumatic nebulizer.

Patients with permanent artificial airways

1. Most patients are capable of acclimating to artificial airways without the need for continuous aerosol therapy other than the use of artificial noses or heat and moisture exchangers.

2. Many require nocturnal aerosol therapy. Whether heated or unheated, with or without oxygen, depends on the patient’s medical status and tolerance. Note that use of oxygen as a convenient vehicle of aerosol delivery, as is commonly used in the hospital setting, is normally not done in the home care setting. This is a result of home care equipment limitations and third-party payer (reimbursement) constraints.

3. Occasionally the direct instillation of 3 to 5 ml of sterile 0.9% NaCl into the tracheostomy tube on an intermittent or before suction basis is used to thin secretions and/or induce cough.

4. Systemic hydration remains the cornerstone of maintaining normal viscosity of pulmonary secretions.

Small-volume aerosolized drug therapy (see Chapters 17 and 35) is typically delivered via metered dose inhaler (MDI) or small volume nebulizers (SVNs) with a compressor.

1. Many nontracheostomized, nonmechanically ventilated patients are capable of coordinating the use of an MDI with a spacer, and delivery of medication is best served in that fashion for convenience and cost.

2. However, in those who have psychomotor and/or cognitive impairments precluding the coordinated ventilatory act required of the MDI, the use of pneumatically operated SVNs is warranted. Young pediatric patients unable to grasp verbal instruction also commonly use such devices. These devices are typically AC powered, but variations with rechargeable self-contained batteries and/or cigarette lighter adapters are manufactured.

3. Most common drugs delivered by aerosol in the home include

Complications/safety considerations

VI Bronchial Hygiene Techniques Used in Home Care (Box 32-1)

Postural drainage, percussion, and vibration (see Chapter 36)

1. Postural drainage uses bodily positioning along with gravity to drain secretions from specific targeted segments of the lung to larger, central airways.

2. Chest percussion is used to dislodge secretions and is performed on inspiration and exhalation using the cupped-hand technique, a palm-held, molded rubber cup-shaped device, or alternatively an electric- or pneumatic-powered percussor. Mechanical percussors are generally a nonreimbursed item.

3. Chest vibration is used to mobilize secretions to larger airways and is performed by placing one hand on top of the other over the affected area. The therapist performs a vibrating action with the arms and hands by tensing his or her shoulders during the patient’s exhalation. This can be delivered via a mechanical vibrating device, which is typically nonreimbursed. Vibration is used alone if the patient cannot tolerate percussion.

Incentive spirometry, also referred to as sustained maximal inspiration (SMI), is designed to encourage and reinforce the patient to take protracted, slow, deep breaths. This is accomplished using a device that provides patients with visual positive feedback when they inhale at a predetermined flow rate or volume and sustain the inflation for a predetermined amount of time.

Intermittent positive pressure breathing (IPPB) is a technique used to provide short-term or intermittent mechanical ventilation for the purpose of augmenting lung expansion, delivering aerosolized medication, or assisting ventilation. The goal is to deliver a tidal volume (Vt) greater than the patient can initiate on his or her own; therefore, delivered Vt measurement is used to assess the therapeutic efficacy of IPPB in select patient populations. However, it is used infrequently.

Flutter valves and positive expiratory pressure (PEP) devices

1. The Flutter device is a small handheld mucus clearance device made of plastic with a pipelike shape. The mouthpiece (or mask) is connected on one end, and an encased steel ball (the flutter device) resting in a plastic circular cone is connected on the other end.

2. The PEP device consists of a removable mouthpiece (or mask) and includes a one-way inspiratory valve and an expiratory resistance/frequency adjustable dial with an optional pressure measurement port.

3. The primary benefit of the Flutter or PEP devices is that after instruction, the need for a professional to oversee its use is only periodically necessary, with the patient typically self-administering therapy.

Directed cough techniques

Cough assist devices

High frequency chest wall compression (HFCC)

1. THAIRapy vest developed by American Biosystems Inc.

a. The vest consists of an inflatable jacket and an air-pulse generator.

b. The vest fits snugly over the entire thorax and is connected to the front panel of the air-pulse generator by two hoses to create external chest wall oscillations.

c. Increasing oscillatory frequencies ranging from 5 to 20 Hz are used during the course of each treatment.

d. The device is activated at the patient’s full inspiration, and a 5-minute period of oscillation at low frequency is initiated.

e. Transient increases in airflow are produced at each stepwise increase in compression and result in mobilization and physical alteration of secretions.

f. The patient performs a forced expiratory maneuver before the device is deactivated to allow expectoration of secretions.

g. The cycle is repeated five or six times at gradually increasing frequencies.

2. Hayek Oscillator by Breasy Medical Equipment

a. Uses a microprocessor-controlled noninvasive ventilator

b. The unit is connected to the patient via a traditional cuirass.

c. There is a range of cuirass sizes suitable for preterm infants to adults.

d. The cuirass is attached to a piston pump that can provide ventilation up to high frequencies, and baseline negative pressure is produced from a vacuum pump.

e. Frequencies of up to 15 Hz can be achieved, and frequency, inspiratory and expiratory pressures, and inspiratory-to expiratory (I:E) ratio are controlled variables for patients.

f. Negative and positive pressures are alternately applied during inspiration and exhalation, respectively, with oscillations around a negative baseline followed by an artificial cough.

g. The artificial cough has a prolonged inspiratory phase followed by a short expiratory phase.

Intrapulmonary percussive ventilation (IPV)

1. Percussionare developed by Dr. Forest Bird

2. IPV is used to mobilize and clear retained secretions, assist in the resolution of atelectasis, and deliver aerosolized medications.

3. Three therapeutic results are provided during IPV.

Home suctioning techniques

1. Clean versus sterile technique

2. Reuse of suction catheters is commonly used in the home care setting when the catheters are

VII Nasal CPAP and Nasal Positive Pressure Ventilation (NPPV)

Nasal CPAP is used to manage obstructive sleep apnea (OSA).

1. Nasal CPAP is achieved by creating a prescribed CPAP from a flow generator via a nasal appliance. This same device may be alternatively used with an oral mask.

2. Medicare reimbursement criteria for nasal CPAP

a. Diagnosis of OSA documented by an attended, facility-based polysomnogram that meets either of the following criteria.

b. Polysomnographic studies must be performed in a facility-based sleep study laboratory and not in the home or a mobile facility. These laboratories must be qualified providers of Medicare services and comply with all applicable state regulatory requirements.

c. A home medical equipment (HME) supplier must not perform polysomnographic studies.

d. Continued coverage of the CPAP device beyond the first 3 months of therapy requires that no sooner than the 61st day after initiating therapy, the supplier ascertain and document from either the patient or the treating physician that the patient is continuing to use the CPAP device.

e. Noncompliant and/or periodic discontinuation of usage of the device may interrupt the supplier’s billing for the equipment and related accessories and supplies.

f. Other third-party payers follow similar, less stringent criteria.

Nasal noninvasive intermittent positive pressure ventilation (NIPPV) is used to manage OSA, restrictive thoracic disease, severe chronic obstructive pulmonary disease (COPD), or central sleep apnea (CSA). Numerous small NPPV units are available from many manufacturers, but all function in a similar manner.

1. Nasal NPPV is patient sensitive and assists ventilation by creating a prescribed positive inspiratory pressure (IPAP) followed by a less but still positive expiratory pressure (EPAP) via an electronically controlled flow generator applied via a nasal appliance. This same device may alternatively be used with an oral mask. Some units have the feature of setting a backup respiratory rate in the “timed mode,” providing for an assist-control mode.

2. Medicare reimbursement criteria for nasal NPPV

a. Physician must fully document in the patient’s medical record symptoms characteristic of sleep-associated hypoventilation, such as daytime hypersomnolence, excessive fatigue, morning headache, cognitive dysfunction, and dyspnea.

b. Indicated for those patients with clinical disorder groups characterized as (1)restrictive thoracic disorders (i.e., progressive neuromuscular diseases or severe thoracic cage abnormalities); (2) severe COPD; (3) CSA; or (4) OSA, and who also meet the following criteria.

(1) Restrictive thoracic disorders

(a) There is documentation in the patient’s medical record of a progressive neuromuscular disease (e.g., amyotrophic lateral sclerosis) or a severe thoracic cage abnormality (e.g., postthoracoplasty for tuberculosis), and

(b) An arterial blood gas Paco2, done while awake and breathing the patient’s usual FIO2 that is ≥45 mm Hg, or

(c) Sleep oximetry demonstrates oxygen saturation ≤88% for at least five continuous minutes, done while breathing the patient’s usual FIO2, or

(d) For a progressive neuromuscular disease (only), maximal inspiratory pressure <60 cm H2O or forced vital capacity is <50% predicted, and

(e) COPD does not contribute significantly to the patient’s pulmonary limitation.

(f) If all the aforementioned criteria are met, nasal NPPV (based on the judgment of the treating physician) will be covered for patients within this group of conditions during the first 3 months of therapy.

(2) Severe COPD

(3) CSA (i.e., apnea not caused by airway obstruction)

(a) A complete facility-based, attended polysomnogram must be performed documenting the following.

(b) The diagnosis of CSA, and

(c) The exclusion of OSA as the predominant cause of sleep-associated hypoventilation, and

(d) The ruling out of CPAP as effective therapy if OSA is a component of the sleep-associated hypoventilation, and

(e) Oxygen saturation ≤88% for at least 5 continuous minutes, done while breathing the patient’s usual FIO2, and

(f) Significant improvement of the sleep-associated hypoventilation with the use of nasal NPPV on the settings that will be prescribed for initial use at home, while breathing the patient’s usual FIO2.

(g) If all the aforementioned criteria for patients with CSA are met, nasal NPPV (based on the judgment of the treating physician) will be covered for the first 3 months of therapy.

(4) OSA

c. Continued coverage of nasal NPPV therapy beyond the first 3 months of therapy requires that no sooner than the 61st day after initiating therapy the physician ascertain and document progress of relevant symptoms and that the patient continues to use the nasal NPPV device at least an average of 4 hours every 24 hours. The supplier also must obtain such documentation from the physician, as well as a statement from the patient documenting use, benefit, and intention for continued use.

d. This reimbursement criterion put forth by Medicare is supported by a review of the available science investigating the clinical indications for Noninvasive Positive Pressure Ventilation in Chronic Respiratory Failure Due to Restrictive Lung Disease, COPD, and Nocturnal Hypoventilation consensus report findings, published in 1999.

e. As with CPAP, other third-party payers follow similar but generally less stringent criteria.

VIII Mechanical Ventilation

Thousands of individuals in the United States require long-term mechanical ventilatory assistance.

It is likely that this number will increase in the future as a result of

Possible solutions to the increased need for long-term invasive ventilatory support in the home

General indications and disease categories requiring long-term ventilatory support

1. Patients requiring invasive long-term ventilatory support who have demonstrated

2. Neuromusculoskeletal disorders: The best candidates have chronic, slowly progressive diseases.

3. Central hypoventilation: Those with sleep apnea secondary to a dominant central (versus obstructive) component, including patients with obese hypoventilation syndrome (OHS), generally are good candidates.

4. Obstructive lung disease: Selective patients are good candidates. Many comply poorly, are difficult to treat, and have many other organ system problems and the typical sequelae associated with chronic pulmonary disease.

5. Restrictive lung disease: Normally such patients are rarely candidates; they generally have the highest respiratory needs.

The ideal patient for long-term ventilatory support

The keys to successful transition to the home are appropriate predischarge planning and a well-coordinated team effort. Key members of the discharge team include

Patients requiring nonelective mechanical ventilatory assistance generally require approximately 4 weeks of in-hospital training and acclimation to equipment before transitioning into the home. In the instance of elective ventilatory assistance, namely, the use of NIPPV, the equipment and care needs of the patient are typically less extensive. The discharge plan and process will be streamlined and much quicker than for nonelective cases. In either case the discharge timeframe should be accelerated or decelerated based on individual patient and family needs.

All respiratory care and other assistive equipment anticipated to be used in the home should be used in the predischarge hospital training to allow mastery of related technique, patient acclimation, and documented medical efficacy. Further home care procedures that may differ from hospital procedures should be adopted by patient and family during the predischarge period as soon as educational mastery is achieved.

The need for skilled full-time caregivers in the home is based on numerous factors.

1. Patient

2. Family

3. Third-party coverage

4. Infants generally require more caregiver assistance than adults.

5. Length of time home on ventilatory assistance is generally inversely related to the need for skilled caregivers as patient and family become accustomed with requisite care provision over time.

Invasive positive pressure ventilation

1. Medical discharge criteria

2. Other discharge criteria as modified from AARC clinical practice guidelines: 1995

a. The caregivers (lay and professional) must clearly demonstrate and have documented the competencies required for caring for the specific patient. This should include all aspects of the patient’s care needs.

b. Availability of caregivers for each 24-hour period must be ensured.

c. The chosen site must be capable of housing, operating, maintaining, and supporting the equipment required by the patient’s medical condition. This should include respiratory and ancillary equipment and supplies, such as the ventilator, suction, oxygen, intravenous therapy, nutritional therapy, adaptive equipment, and disposable supplies.

d. The physical environment must be evaluated for safety and suitability. It should be free of fire, health, and safety hazards; provide adequate heating, cooling, and ventilation; provide adequate electrical service (and backup generator as indicated); and provide for patient access and mobility with adequate patient space (room to house medical and adaptive equipment) and storage facilities.

e. Financial resources must be identified at the beginning of the discharge process. Lack of or inadequate funding impacts the entire discharge plan and can dictate the discharge care site. Sources and adequacy of funds for alternate-site care, medical equipment and supplies, the required medical personnel, any modifications necessary to environment, and ongoing medical care should be fully evaluated.

f. Development of a multidisciplinary plan of care is based on the evaluation of the patient’s needs and goals. The plan should be consistent with recommended practices and guidelines for the patient’s condition. Key elements of the plan should include

g. Some have found a 1- to several-day trial period at home helpful in allowing unanticipated, real day-to-day issues to be identified and then later prepared for in the controlled environment of the hospital before final discharge home. Others view such efforts as laborious and expensive, consuming disproportionate emotional and physical resources for small return. There are no hard and fast rules, and individual consideration of the concept should be given to each discharge situation.

h. When all medical and other criteria have been satisfied, the patient is appropriately prepared for discharge. It should be noted that whatever the external demands, fiscal or otherwise, there are no shortcuts to education or planning that will allow safe, successful discharge.

3. Home mechanical ventilators (general descriptions)

a. Today there are essentially five categories of mechanical ventilators specifically designed and manufactured for home use.

(1) BiPAP ventilators

(2) Volume-targeted home care ventilators

(3) Pressure/volume-targeted home care ventilators

(4) Negative pressure ventilators (see Section K-1, a-g, Negative Pressure Ventilators)

(5) High frequency ventilators (HFVs) or oscillators (see Chapter 42)

b. Ventilators designed primarily for hospital use should not be used in the home. With the large variety of commercially available, full-featured, home care ventilators, it is rare to find a justified exception to this axiom.

4. Ideal characteristics of home care ventilators

a. Designed exclusively with the home care patient in mind

b. User-friendly operation

c. Reliable operation

d. Absence of confusing digital readouts that change on a breath-by-breath basis

e. Provide ventilatory support of infant, children, and adult populations

f. Available modes: Assist/control (AC), control, pressure support, synchronized intermittent mandatory ventilation (SIMV), positive end-expiratory pressure (PEEP), and NPPV

g. Respiratory rate: 1 to 80 breaths/min

h. Vt: 50 to 2000 ml

i. Peak flow capability: 30 to 140 L/min

j. Spontaneous demand peak flow provision: ≤180 L/min

k. Automatic and/or settable leak compensation

l. Inspiratory time: 0.3 to 3.0 seconds

m. Assisted breaths triggered by PEEP compensated, proximal patient airway pressure sensing

n. Direct setting of FIO2 or access for titration of oxygen

o. Integral PEEP feature

p. Internal battery capable of functioning for >2 hours

q. Appropriately alarmed

r. Remote alarm capability

s. Remote “nurse” call capability

t. Compact (fits on wheelchair), light (<35 lb)

u. Such home care ventilators are manufactured; however, they are the newest, generally most expensive, and not the norm and probably contain features not required in a given patient situation.

5. The internal gas flow pathway of an older simple, piston-operated, volume-targeted, commonly used home care ventilator in the United States is illustrated in Figure 32-2.

a. Gas enters these units by way of an oxygen-accumulating device with a one-way valve.

b. A one-way valve is also located just inside the unit’s intake filter.

c. From here gas enters the piston chamber during the piston’s back stroke.

d. During the piston’s forward stroke gas exits the piston chamber via a one-way valve, which prevents gas from being drawn into the piston chamber during the piston’s back stroke.

e. A pop-off valve is located in the piston chamber (vent) to prevent excessive chamber pressurization.

f. Once gas exits the piston chamber it normally passes a pressure limit control adjustable by operator and a one-way inlet valve, allowing inspiration of gas from the unit’s circuit during the piston’s back stroke.

g. Not all units include the one-way inlet valve.

h. No demand or continuous gas flow system is included. During spontaneous breathing or SIMV the patient must draw gas from

i. Note: Spontaneous demand flow systems are integral on bilevel and pressure/volume-targeted home care ventilators that avoid the aforementioned problems observed with the older volume-targeted ventilators.

6. Delivery of oxygen

7. PEEP

8. The SIMV/IMV mode

a. Spontaneous demand systems are integral on pressure- and volume-targeted home care ventilators and are capable of meeting most patient peak inspiratory flow demands.

b. Many of the volume-targeted home care ventilators posing a SIMV/IMV mode do not have a demand system or continuous flow setup included in the system’s basic design.

c. When such a unit is set on SIMV/IMV with the use of a bubble-through humidifier, the imposed WOB is excessive (Figure 32-3).

d. Even the use of passover humidifiers or heat and moisture exchangers does not sufficiently reduce the WOB.

e. In this situation if the SIMV/IMV mode is to be used, a one-way H-valve system placed before the system’s passover humidifier should be used (Figure 32-4).

f. If an increased FIO2 value is needed oxygen can be titrated into the one-way H-valve system as shown in Figure 32-4.

g. The use of a one-way valve and a passover humidifier greatly reduces the imposed WOB of these systems in the SIMV/IMV mode.

9. Selection of ventilator parameters (adults)

10. Selection of ventilator parameters (infants)

Noninvasive nonpositive pressure approaches to mechanical ventilation

1. NPV

a. NPV is provided by applying a settable subatmospheric pressure to the thorax and abdomen on an intermittent basis, which causes an increase in thoracic and abdominal volume and a resultant decrease in respective pressures. These subatmospheric pressures are in turn transmitted to the lungs, creating a pressure gradient from atmosphere to lungs and resulting in inspiratory flow. Exhalation occurs passively when the machine-imposed subatmospheric pressure is intermittently stopped, and the normal elastic recoil of lungs, thorax, and abdomen increases intrapulmonary pressure, creating a pressure gradient from lungs to atmosphere.

b. NPV can be provided with the use of

c. The size of the Vt generated by a given negative pressure is a function of the thoracic and abdominal compliance, the airway resistance, and the surface area of negative pressure exposure. Hence NPV functions best on individuals with normal lungs and normal thoracic compliance typical of patients with neuromuscular disease. However, NPV has successfully been used to support other disease entities.

d. It should not be used for patients with upper airway obstruction because it will exaggerate the obstructive component.

e. For optimal use, respiratory rates and pressure must be high enough to capture the diaphragm and inhibit spontaneous ventilation. Optimally the diaphragm will work in synchrony with the ventilator.

f. Most negative pressure units are controllers, with the newly manufactured units featuring a patient-triggered AC mode providing greater synchrony with spontaneous ventilatory effort.

g. The use of NPV is frequently successful when it is combined with other noninvasive approaches to ventilation.

2. Pneumobelt (Figure 32-5)

3. Rocking bed (Figure 32-6)

image
FIG. 32-6 Rocking bed.