Diffusing Capacity Tests

Published on 12/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 12/06/2015

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 5796 times

Chapter 3

Diffusing Capacity Tests

The chapter describes measurement of diffusion in the lungs. Diffusing capacity (also referred to as transfer factor) is usually measured using small concentrations of carbon monoxide (CO) and is referred to as Dlco or Dco. Dlco is used to assess the gas-exchange ability of the lungs, specifically oxygenation of mixed venous blood. Various methods, all of which use CO, have been described. The most commonly used method is the single-breath, or breath-hold technique. The  single-breath method is also the most widely standardized. The techniques section focuses on the single-breath method but also describes some of the other methods that are used for specific applications.

Dlco measurements are used in the diagnosis and management of most pulmonary disorders. The importance of standards and guidelines in the performance of Dlco tests and in the overall interpretation of results is highlighted. Interpretive strategies are presented in a format similar to those in previous chapters.

Carbon monoxide diffusing capacity

Description

Dlco measures the transfer of a diffusion-limited gas (CO) across the alveolocapillary membranes. Dlco is reported in milliliters of CO/minute/millimeter of mercury at 0°C, 760 mm Hg, dry (i.e., standard temperature, pressure, dry [STPD]).

Techniques

CO combines with hemoglobin (Hb) approximately 210 times more readily than oxygen (O2). In the presence of normal amounts of Hb and normal ventilatory function, the primary limiting factor to diffusion of CO is the status of the alveolocapillary membranes. This process of conductance across the membranes can be divided into two components: membrane conductance (Dm) and the chemical reaction between CO and Hb (Figure 3-1). Dm reflects the process of diffusion across the alveolocapillary membrane. Uptake of CO by Hb depends on the reaction rate (θ) and the pulmonary capillary blood volume (Vc). These two components occur in series, so the diffusion conductance can be expressed as:

< ?xml:namespace prefix = "mml" />1DLCO=1Dm+1θVcDimage

Diffusing capacity can be affected by factors that change the membrane component, as well as by alterations in Hb and the capillary blood volume.

A small amount of CO in inspired gas produces measurable changes in the concentration of inspired versus expired gas. Because little or no CO is normally present in pulmonary capillary blood, the pressure gradient causing diffusion is basically the alveolar pressure (PAco). If the partial pressure of CO in the alveoli and the rate of uptake of the gas can be measured, the Dlco of the lung can be determined. There are several methods for determining Dlco (Table 3-1). All methods are based on the following equation:

Table 3-1

Advantages and Disadvantages of Dlco Testing Methods

Method Technique Advantages Disadvantages Applications
Dlcosb (breath hold) CO and tracer gas analysis relatively simple; 10-sec breath hold Easy calculations, simple, fast; highly standardized and automated; minimal COHb back-pressure effect Sensitive to distribution of ventilation and image/imagematching; “nonphysiologic”; not practical for exercise testing Screening and clinical applications
Dlcorb (rebreathing) Rapid analysis of CO and tracer gas required; rebreathing must be controlled Less sensitive to VA than Dlcosb; less sensitive to image/imageabnormalities; can be used with NO to measure Dlno Complex calculations (computerized); rapidly responding analyzers required; sensitive to COHb back-pressure Clinical and research applications; provides most accurate Dlco
Dlcoib (intrabreath) Rapid analysis of CO and tracer gas during single controlled exhalation Breath holding not required; can be used during exercise Complex calculations (computerized); flow must be controlled; sensitive to uneven image/image; not standardized Screening and clinical applications; may be useful in patients who cannot hold their breath
1/Dm + 1/θVc (membrane diffusing capacity) Dlcosb repeated at two different levels of alveolar PO2 Differentiates membrane transfer resistance from red cell uptake Complex calculations; estimates of alveolar PO2 are critical Research with limited clinical applications

image

Dlco

DLCO=V·COPACO-PCCOimage

where:

imageco milliliters of CO transferred/minute (STPD)

PAco = mean alveolar partial pressure of CO

PCco = mean capillary partial pressure of CO, assumed to be 0

Dlco is expressed in milliliters of gas/minute/unit of driving pressure at STPD conditions.

Single Breath-Hold Technique (Modified Krogh’s Technique)

The patient exhales to RV and then inspires a vital capacity breath (referred to as the IVC or VI) from a system such as that in Figure 3-2. A special diffusion gas mixture is delivered either from a spirometer, a reservoir bag, or by means of a demand valve. The diffusion mixture usually contains 0.3% CO, a “tracer” gas, 21% O2, and the balance is N2. The tracer gas is usually an insoluble, inert gas such as helium (He), methane (CH4), or neon (Ne). The tracer used depends on the type of analyzer implemented to analyze the exhaled gas. The traditional method used He (usually about 10%) as the tracer gas. Rapidly responding infrared analyzers (see Chapter 11) have been used for continuous analysis of the small changes in CO. The same type of infrared analyzer can use CH4 as the tracer gas, so that a single analyzer can rapidly detect changes in both CO and the tracer. Another method uses gas chromatography (see Chapter 11) to detect changes in CO with neon used as the tracer gas.

After inspiring the VC breath, the patient holds the breath at total lung capacity (TLC) for approximately 10 seconds. The patient then exhales. After a suitable washout volume (750–1000 mL) has been discarded, a sample of alveolar gas is collected. The alveolar sample may be collected in a small bag (approximately 500 mL or less) or by continually aspirating a sample of the exhaled gas.

The sample is analyzed to obtain the fractional CO and tracer gas concentrations in alveolar gas, FAcoT (where T is the time of the breath hold), and FAtracer, respectively. The concentration of CO in the alveoli at the beginning of the breath hold (FAco0) must be determined as well. It is calculated as follows:

FACO0=F1CO×FAtracerFItracerimage

where:

FAco0   = fraction of CO in alveolar gas at beginning of breath hold (time = 0)

FIco   = fraction of CO in reservoir (usually 0.003)

FAtracer = fraction of tracer in alveolar gas sample

FItracer  = fraction of tracer in inspired gas (varies with tracer gas used)

The change in tracer gas concentration reflects dilution of inspired gas by the gas remaining in the lungs (i.e., RV). This change is used to determine the CO concentration at the beginning of the breath hold, before diffusion from the alveoli into the pulmonary capillaries. The Dlcosb (single-breath) is then calculated as follows:

DLCOsb=VA×60(PB47)×(T)×LnFACO0FACOTimage

where:

VA     = alveolar volume, mL (STPD)

60       = correction from seconds to minutes

PB        = barometric pressure, mm Hg

47       = water vapor pressure at 37 °C, mm Hg

T   = breath-hold interval, seconds

Ln  = natural logarithm

FAco0    = fraction of CO in alveolar gas at beginning of breath hold

FAcoT    = fraction of CO in alveolar gas at end of breath hold

VA may be calculated from the single-breath dilution of the tracer gas:

VA=(VIVD)×FItracerFAtracer×STPD correction factorimage

where:

VI     = volume of test gas inspired, mL (see Figure 3-3)

VD   = dead space volume (anatomic and instrumental), mL

FAtracer = fraction of tracer in alveolar gas sample

FItracer  = fraction of tracer in inspired gas (depends on tracer used)

Note that the VA, usually expressed in BTPS units, must be converted to STPD for the single-breath calculation. The dilution of tracer gas is used twice to determine the CO concentration at the beginning of the breath hold and to determine the lung volume (i.e., VA) at which the breath hold occurred.

A simplification of the single-breath method described earlier is widely used. The tracer gas and CO analyzers may be calibrated to read full scale (100% or 1.000) when sampling the diffusion mixture, and to read zero when sampling air (no tracer or CO). If the analyzers have a linear response to each other, the fractional concentration of the tracer gas in the alveolar sample is equal to the FAco0. This technique assumes that both the tracer gas and CO are diluted equally during inspiration. Because no tracer gas leaves the lung during the breath hold, its concentration in the alveolar sample approximates that of the CO before any diffusion occurred. The exponential rate of CO diffusion from the alveoli can then be expressed as follows:

Ln(FAtracerFACOT)image

where:

FAtracer = fraction of tracer in alveolar gas, equal to FAco0

FAcoT   = fraction of CO in alveolar gas at end of breath hold

Ln   = natural logarithm of the ratio

This technique avoids the necessity of analyzing the absolute concentrations of the two gases. However, it requires that the analyzers be linear with respect to each other. Analysis of CO is often done using infrared analyzers (see Chapter 11), and their output is nonlinear. Care must be taken to ensure that corrected CO readings are used in the computation. These corrections are easily accomplished either electronically or via software in computerized systems. Systems that use the same detector for both CO and the tracer gas (e.g., infrared, gas chromatography) also need to provide linear output. The linearity of the system should be within 0.5% of full scale. This means that any drift or nonlinearity should cause no more than a 0.5% error when analyzing a known gas concentration (See Chapter 12).

Dlco gas analysis is commonly performed with either a rapidly responding multigas analyzer or gas chromatography. Multigas analysis uses specialized infrared analyzers capable of detecting several gases simultaneously. These systems use methane (CH4) as a tracer gas. One advantage of multigas analysis is that CO and CH4 are measured rapidly and continuously (Figure 3-4) so that the calculated Dlco is available as soon as the exhalation has been completed. Gas chromatography (see Chapter 10) can also be used for Dlco gas analysis. Neon (Ne) is used as the tracer gas (Figure 3-5). Helium is used as a “carrier” gas for the chromatograph. Although gas analysis using chromatography is slow (60–90 seconds), it is extremely accurate.

The resistance of the breathing circuit should be less than 1.5 cm H2O/L/sec, at a flow of 6 L/sec. This is important in allowing the patient to inspire rapidly from RV to TLC. A demand valve may be used instead of a reservoir bag for the test gas. In a demand-flow system, the maximal inspiratory pressure to maintain a flow of 6 L/sec should be less than 10 cm H2O. Increased resistance, in either a reservoir or a demand valve system, may cause the patient to produce large subatmospheric pressures during inspiration. This has the effect of increasing pulmonary capillary blood volume and may falsely increase Dlco.

The timing device for the maneuver should be accurate to within 100 msec over a 10-second interval (1%). Most computerized systems time the maneuver automatically. However, a means of verifying the accuracy of the breath-hold time should be available. The Jones and Meade method of timing the breath hold should be used (see Figure 3-5). The Jones and Meade method measures breath-hold time from 0.3 of the inspiratory time to the midpoint of the alveolar sample collection.

Corrections must be made for the patient’s anatomic dead space (VD) and for dead space in the valve (and sample bag, if used). Anatomic VD should be calculated as 2.2 mL/kg (1 mL/lb) of ideal body weight. The equipment manufacturer should specify instrument VD. Instrument VD should not exceed 350 mL for adult subjects, including mouthpiece and any filters that might be used. Smaller instrument VD may be necessary for pediatric patients. Anatomic and instrument VD are subtracted from inspired volume (VI) before the alveolar volume (VA) is calculated.

All gas volumes must be corrected from ATPS to STPD for the Dlco calculations. However, when the VA is used to calculate the ratio of Dlco to lung volume (Dl/VA), it is normally expressed in BTPS units. Accurate measurement of inspired volume during the maneuver requires that the spirometer have an accuracy of 3.5% (3% + 0.5% for the calibration syringe itself) over a range of 8 L. Volume-based spirometer systems must also be free from leaks.

Gas analyzers that are affected by carbon dioxide (CO2) or water vapor require appropriate absorbers. Absorption of CO2 is usually accomplished with a chemical absorber using Ba(OH)2 (baralyme) or NaOH (soda lyme). Each of these reactions produces water vapor. Therefore, CO2 absorbers should be placed upstream of an H2O absorber. Anhydrous CaSO4 is commonly used to remove water vapor. Selectively permeable tubing (PERMA PURE®) can also be used to establish a known water vapor content. Gas-conditioning devices must be routinely checked to ensure accurate gas analysis. Chemical absorbers typically add an indicator that changes color as the absorber becomes exhausted.

Dlcosb maneuvers should be performed after the patient has been seated for at least 5 minutes. The patient should refrain from exertion immediately before the test; exercise increases cardiac output, which increases Dlco. The patient should be instructed about the requirements of the maneuver and the technique demonstrated. Expiration to RV should be of a reasonable duration, usually 6 seconds or less. Patients who have airway obstruction may have difficulty exhaling completely in this interval. Inspiration to TLC should be rapid but not forced. Healthy subjects and patients with airway obstruction should be able to inspire at least 85% of their VC within 4 seconds (Criteria for Acceptability 3-1). The single-breath calculation assumes instantaneous filling of the lung. Prolonged inspiratory times will decrease the actual time of breath holding at TLC, typically resulting in a lower Dlco. The breath hold should be relaxed, against either the closed glottis or a closed valve. The patient should avoid excessive positive intrathoracic pressure (Valsalva maneuver) or excessive negative intrathoracic pressure (Müller maneuver). A Valsalva maneuver reduces pulmonary capillary blood volume and may produce a falsely low Dlco. A Müller maneuver increases pulmonary capillary blood volume and may falsely increase Dlco. Expiration after the breath hold should be smooth and uninterrupted. Exhalation should take less than 4 seconds, and alveolar gas sampling should occur in less than 3 seconds (again because the calculations assume instantaneous emptying of the lung). Patients who have moderate or severe airway obstruction may have difficulty achieving these criteria. The breath-hold time, measured using the Jones and Meade method (see Figure 3-3), should be 10 seconds ± 2 seconds. Prolonged inspiratory or expiratory times may result in breath-hold times greater than 12 seconds and should be noted on the report.

To obtain an alveolar sample, dead space gas needs to be washed out (i.e., discarded). A washout of 0.75–1.0 L is usually sufficient to clear the patient and sampling device dead space. For patients with small vital capacities (>2.0 L), the washout volume may be reduced to 0.5 L. A sample volume of 0.5–1.0 L should be collected, but a smaller sample may be necessary in patients whose VC is less than 1 L. In Dlcosb systems that analyze expired gas continuously (see Figure 3-4), inspection of the washout of the tracer gas may be used to select an appropriate alveolar sample. Rapidly responding gas analyzers that measure the CO and tracer gas simultaneously allow adjustment of washout volume (i.e., dead space) and sample volume after completion of the maneuver. These adjustments may be particularly useful in subjects who have very small vital capacities (e.g., pediatric patients or adults with severe restrictive disease). Such adjustments assume that both the CO and tracer gas concentrations reflect changes occurring at the mouth. Alveolar sampling may be adjusted to begin at the point where the tracer gas and CO indicate an “alveolar plateau” (see Figure 3-4). In patients who have uneven mixing or emptying of the lungs, there may not be a clear demarcation between dead space and alveolar gas. Adjustments to either washout (dead space) volume or sample volume should be noted on the report.

Two or more acceptable Dlcosb maneuvers (see Criteria for Acceptability 3-1) should be averaged. Duplicate determinations should be within 3 mL CO/min/mm Hg of each other, or within 10% of the largest value obtained from an acceptable effort. No more than five repeated maneuvers should be performed because of the effect of increasing carboxyhemoglobin (COHb) from inhalation of the test gas. There should be a 4-minute delay between repeated maneuvers to allow for washout of the tracer gas from the lungs.

Corrections for abnormal hemoglobin (Hb) concentrations should be applied with a current Hb value. The predicted Dlco should be corrected so that it reflects the Dlco at an Hb value of 14.6 g% for adult and adolescent males, and to an Hb value of 13.4 g% for women and children of either sex younger than 15 years of age. The Hb-corrected predicted value for males may be calculated as follows:

DLCO(predicted for Hb)=DLCO(predicted)×(1.7×Hb)10.22+Hb)image

Similarly, Hb correction of the predicted values for women and children younger than 15 years is calculated as follows:

DLCO(predicted for Hb)=DLCO(predicted)×(1.7×Hb)(9.38+Hb)image

Note that this scheme corrects the predicted value rather than the patient’s measured value. For example, in an adult male patient with an Hb of 9.0 g/dL and a predicted Dlco of 25 mL/min/mm Hg, a measured Dlco of 19 mL/min/mm Hg would be reported:

Actual Predicted %
Dlco mL/min/mm Hg 19.0 25.0 76
Dlco (Hb corrected) mL/min/mm Hg 19.0 19.9 95

image

Both uncorrected and corrected predicted Dlco values and the resulting percentages should be reported, along with the Hb value used for correction. Decreased Hb levels (anemia) will always reduce the predicted value, whereas elevated Hb (polycythemia) will increase the predicted value.

Correction for the presence of COHb in the patient’s blood is also recommended. The predicted Dlco may be adjusted as follows:

DLCO(predicted for COHb)=DLCO(predicted)×(102%COHb%)image

The COHb% is the fraction of carboxyhemoglobin determined by hemoximetry expressed as a percentage. This method assumes that the predicted value already includes 2% COHb in healthy subjects. For subjects who have a COHb% greater than 2%, the predicted value will always be reduced using this method. Patients should be asked to refrain from smoking for 24 hours before the test to reduce the CO back-pressure in the blood. For patients who continue to smoke, the time of last exposure should be recorded.

Dlco varies inversely with changes in alveolar oxygen pressure (PAO2). PAO2 changes as a function of altitude, as well as with the partial pressure of oxygen in the test gas. Dlco increases approximately 0.35% for each mm Hg decrease in PAO2, or about 0.31% for each mm Hg decrease in PIO2. When test gas mixtures that produce an inspired O2 pressure of 150 mm Hg (i.e., 21% at sea level) are used, Dlco values will be equivalent to those measured at sea level. Alternatively, standard test gas (FIO2 = 0.21 is typical) can be used and the predicted Dlco corrected by adjusting PIO2. For a gas with a PIO2 of 150 mm Hg, the equation is as follows:

Dlco predicted for altitude = Dlco predicted/(1.0+0.0031[PIO2-150])

Buy Membership for Pulmolory and Respiratory Category to continue reading. Learn more here