Clinical Measurement and Monitoring

Published on 27/02/2015 by admin

Filed under Anesthesiology

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 3493 times

Clinical Measurement and Monitoring

The ability to measure and monitor the physiology of patients is fundamental to modern anaesthesia. The anaesthetist is responsible for the correct use of sophisticated instruments which extend clinical observations beyond the human senses and enhance patient care. This requires vigilance and awareness of the limitations of the processes of measurement and the many causes of error. Uncritical acceptance of the recordings of monitoring equipment in the face of contradictory evidence is a common mistake. Unreliable measurements that are taken at face value and used to change patient management compromise the safety and effectiveness of care. It is essential that those who use monitors understand their limitations and are able to justify their risks.

Clinical measurement is limited by four major constraints:

image Feasibility of measurement. The sensitivity and inherent variability of a clinical measurement depend on complex interactions and technical difficulties at the biological interface between the patient and the instrument.

image Reliability of measurements is determined by the properties of the measurement system. This is influenced by the calibration and correct use of the instrument. Simple examples include the correct placement of ECG electrodes, or the appropriate size of cuff for non-invasive measurement of arterial pressure. Delicate equipment, e.g. a blood gas analyser, requires regular maintenance and calibration.

image Interpretation depends on the critical faculties of the anaesthetist who interprets the significance of measurements in the context of complex physiological systems. Arterial pressure may be within the normal range despite severe hypovolaemia or derangement of cardiovascular function within the limits of physiological compensation. Global measurements of end-tidal carbon dioxide tension or oxygen saturation are influenced by many factors in a highly complex system. More information is required to deduce the cause of a change in the measurement.

image Value of clinical measurements in patient care is defined by the role of a measurement in improving patient care. This includes the ease, convenience, continuity and usefulness of a clinical measurement, and evidence of improvement in patient safety and clinical outcome.

Monitoring is the process by which clinical measurements are assessed and used to direct therapy. In general, monitors consist of four components (Table 16.1): (1) a device which connects to the patient – this may either be a direct attachment or via a tube or lead; (2) a measuring device, often a transducer which converts the properties of the patient into an electrical signal; (3) a computer which may amplify the signal, filter it and integrate it with other variables to produce a variety of derived variables; (4) a display which may show the results as a wave, a number or a combination. It is important to appreciate that most monitors do not directly measure the displayed variable, and that the displayed variable may not reflect physiological function. For example, an electrocardiograph (ECG) does not measure cardiac function and therefore a normal ECG trace does not guarantee that the heart is pumping effectively. When interpreting measurements, the following questions should be asked:

TABLE 16.1

Four Components of a Monitor

Connection to patient

Measuring device

Electronic filter/amplifier

Display

image What is being measured? In the case of arterial pressure, there is an obvious answer. However, in some cases, for example ‘depth of anaesthesia’, it may not be clear what the monitor is measuring. In addition, many monitors use data from a variety of sources. For example, heart rate is usually derived from the ECG. However, if the ECG fails to provide the data required, the monitor often switches automatically to a rate from either a pulse oximeter or an arterial pressure waveform. Thus, the displayed value may change rapidly despite the patient remaining stable.

image How is it measured? Arterial pressure is often measured by either a transducer attached to an arterial cannula or an automated oscillometer. Although a transducer is often regarded as the more accurate, the readings must be compared with the preoperative values recorded on the ward, usually with an oscillometer. Therefore, where accurate control of arterial pressure is essential, it is advisable to start invasive pressure monitoring before anaesthesia to avoid any confusion with non-invasive measures.

image Is the environment appropriate? Many monitors have been designed for use in operating theatres and do not function correctly if exposed to the cold and vibration, for example in an ambulance or helicopter. Another example is the strong magnetic field produced by magnetic resonance imaging (MRI) scanners. The electrical currents induced may damage not only incompatible monitors but even produce burns to a patient’s skin.

image Is the patient appropriate? Monitors designed for adult use often fail to produce reliable readings when used on small children. Particularly obese adults may require a large blood pressure cuff, and poor-quality ECG readings may be obtained.

image Has the monitor been applied to the correct part of the patient? For example, in aortic coarctation, arterial pressure may be markedly different in each arm. Pulse oximeters also fail to work reliably if placed on a limb distal to a blood pressure cuff.

image Is the variable within the range of the monitor? Most monitors are validated on healthy patients in laboratories. Whether such monitors continue to provide accurate results during the extreme physiological changes of, for example, anaphylaxis is uncertain. This does imply that the usefulness of monitors declines with the health of the patient: that is, they are least reliable when needed most. In most cases of acute perioperative patient deterioration, additional monitoring is needed.

image Has the monitor been checked, serviced and calibrated at the correct intervals? To reduce costs, departments may re-use single-use equipment and fail to ensure that service checks are carried out. All equipment should be tagged with a service sticker. This should identify the date serviced, when the next service is due and who to contact in case of malfunction. Equipment which has not been serviced or is past its service date should not be used.

Table 16.2 shows the checks which the anaesthetist should follow before using a patient monitor.

TABLE 16.2

Premonitoring Checks

What is being measured?

What method is being used?

Has the monitor been serviced and calibrated?

Is the environment appropriate?

Is the patient appropriate?

Is it attached to the appropriate part of the patient?

Is the range appropriate?

Can the display be read?

Are the alarms on and have the limits been set?

This chapter describes the feasibility and reliability of clinical measurements relevant to anaesthetic practice, and how such measurements are used to monitor the patient’s physiology.

PROCESS OF CLINICAL MEASUREMENT

Stages of Clinical Measurement

There are four stages of clinical measurement:

Mechanical instruments use the signal energy to drive a display, with minimal intermediate processing. The height of a fluid-column manometer provides a visible index of pressure. The expansion of mercury within the confines of a thin glass column is a measure of temperature. Mechanical springs and gearing translate the rotation of a vane into the recording of expired volume on a dial. However, the overwhelming trend is for nonelectrical signals to be converted by a transducer to an electrical signal suitable for electronic processing by digital computers.

The Microprocessor Revolution

The development of digital microprocessors over the last 25 years has revolutionized anaesthetic practice. Beautifully engineered mechanical instruments, e.g. the von Recklinghausen oscillotonometer, are now obsolete in developed countries.

Advantages of digital signal processing include:

There are a few important disadvantages:

Essential Requirements for Clinical Measurement

All clinical measurement systems detect a biological signal and reproduce this input signal in the form of a display or record which is presented to the operator. The degree to which a discrete measurement is a true reflection of the underlying signal is defined by its accuracy and precision.

Accuracy is the difference between the measurements and the real biological signal, or in practice, a different and superior ‘gold standard’ measurement. Calibration against predetermined signals is used to test and optimally adjust measuring instruments. For absolute measurements, e.g. arterial pressure, one point must be a fixed reference or ‘zero’.

Precision describes the reproducibility of repeated measurements of the same biological signal. This dispersion is usually described by summary statistics, standard deviation for normally distributed measurements, or the range for non-normal distributions. A single recording is unreliable when the measurement is imprecise. This is especially true of tests which require patient cooperation, practised skill or effort, e.g. peak expiratory flow rate. Repeated measurements demonstrate the variability in response.

The Importance of Repeated Measurements

Differences in repeated clinical measurements arise from three causes:

The anaesthetist must be satisfied with the accuracy and precision of any clinical measurement used in patient management. Repeated measurements which are consistent ensure that the measurement is representative, i.e. precise, but do not ensure accuracy. For example, repeated recordings of invasive arterial pressure may be extremely consistent, but erroneous if the transducer is not calibrated against the correct zero point. Defences against the uncritical acceptance of inaccurate measurements include meticulous care in calibrating instruments and recording of clinical measurements, and reflection on clinical measurements which do not fit the clinical state of the patient or other related measurements. A discrepant result should be rechecked, using a different measurement technique if possible, before it is used to change patient management. This is especially true of complex, operator-dependent techniques such as measurement of cardiac output.

Measurement of Continuous Signals Over Time

Continuous signals, which include the majority of modern clinical measurements such as biological electrical signals and the electrical output of signal transducers, introduce the complication of the response of the measuring instrument to a changing signal over time. The reliability with which a continuous signal is reproduced is defined by the relationship between input and output of the measurement system over the clinical range of signal magnitude and frequency. The input–output function of an accurate clinical measurement system would demonstrate good zero and gain stability, minimal amplitude non-linearity and hysteresis, and an adequate frequency response. This cannot be taken for granted, particularly with older equipment or with variations in environmental temperature or humidity.

Signal-to-Noise Ratio

Biological signals are obscured to a variable degree by unwanted or extraneous signals which have similar physical characteristics and are described as noise, e.g. heart sounds become difficult to detect in the presence of continuous, noisy breath sounds. The efficiency of isolation of the signal from unwanted biological signals and electronic noise sources in the equipment is defined by the signal-to-noise ratio. The variability of the amplitudes of signal and noise is enormous and the signal-to-noise ratio is described using a logarithmic scale of decibels. Microvolt EEG measurements are particularly susceptible to noise from many sources. Biological noise includes contaminating ECG and EMG potentials, particularly from the scalp muscles, and interference from electrochemical activity at the skin–electrode interface. Electrostatic and electromagnetic linkage between the recording wires and nearby sources of mains electricity generates noise which is predominantly 50 Hz frequency and harmonics. Radiofrequency noise from diathermy or transmitters may also be picked up at this stage. Physical disturbance of the recording wires causes tiny changes in capacitive potentials and may add low-frequency noise, called microphony. Thermal noise is added during amplification, particularly at the input stage when the signal is in the microvolt range. Good amplifier design, electronic filtering of unwanted frequencies and modern techniques of digital signal processing may extract small signals from considerable background noise, but this inevitably introduces some distortion of the signal. Prevention of contamination of the signal by minimizing sources of noise before the signal is amplified is always preferable. The operator is responsible for correctly using measuring instruments to optimize the signal and for applying knowledge of the physical principles of the measurement to minimize contamination by noise.

Analogue and Digital Processing

Following signal detection and appropriate transduction, the continuously variable analogue signal is amplified, processed and displayed for the attention of the clinician.

Analogue-to-Digital Conversion

The core processing units of digital computers assume one of two stable states, i.e. a binary, rather than decimal, code. This imposes a limit on the resolving power of the digital processor, although with increasing processing power this limit has become negligible. Earlier 8-bit computing comprised a binary number of eight digits representing 28 integer decimal numbers, from binary 00000000 = decimal 0 to binary 11111111 = decimal 255. In short, an 8-bit converter can resolve an analogue signal with an accuracy of one part in 255, i.e. with an amplitude resolution of 0.4% of full scale. A 12-bit converter is more accurate, with a resolving power of one part in 4095 or 0.02% of full scale. More modern processors are capable of 32-bit computing corresponding to a range of 232 integer decimal numbers (a resolving power of 0.00000002%), with 64-bit processors now commonly found in domestic computer equipment. Whilst highly accurate, the cost of this improvement in resolution is more expensive hardware to digitize, process and store considerably more digital information.

Amplitude resolution is not the only determinant of the accuracy of analogue-to-digital conversion. Resolution over time, determined by the sampling frequency, is also important. A relatively low sampling frequency may provide a representative sample of values for a slowly changing waveform but it may inadequately represent high-frequency components and introduce an aliasing error, in which different signals become indistinguishable. The Nyquist theorem suggests that the minimum sampling frequency to maintain the integrity of the waveform is at least twice the highest frequency component with significant amplitude in the input signal waveform, e.g. a sampling frequency of 100 Hz would adequately capture the fastest rate of change in a physiological pressure signal.

The immensely powerful and complicated hardware and software programming instructions responsible for performing the tasks of digitizing, processing, storing and displaying the input signal are hidden from view in the commercial ‘black box’.

Data Display

Useful instruments communicate measurements in an appropriate and user-friendly manner.

Analogue Displays

A continuously variable signal, such as pressure or temperature, is represented by an analogue display in terms of the amplitude of a physical quantity on a calibrated scale, dial, electrical meter or printed record. The glass thermometer incorporates a wedge-shaped lens which magnifies the appearance of the mercury column against the calibrated background scale. The height of a water column manometer is a linear, visual scale of pressure. Simple mechanical displays are accurate and easily understood, but are inconvenient to read and most suitable for intermittent discrete measurements.

Mechanical spirometers and flowmeters record flow on a dial driven by gears. Electrical moving coil meters use a coil of wire suspended in a magnetic field which rotates in proportion to the applied current and moves a pointer on a calibrated dial. Alternatively, the amplified and filtered electrical signal could drive a chart recorder which produces a continuous printed record of the amplitude of measurements against time. Limitations common to these mechanical devices include fragile moving parts, and inertia which impairs the frequency response to rapidly changing signals.

The cathode ray oscilloscope is an effective screen-based display for continuous analogue electrical signals. A heated cathode generates a stream of electrons which are focused and accelerated onto a phosphorescent coating which lines the flat surface of the tube to generate a bright spot. The position of the electron beam in both x- and y-axes is controlled by electrostatic plates. The continuously varying input signal is applied to the y-plates so that deflection in the vertical y-axis is proportional to the amplitude of the signal. The absence of mechanical parts results in a high-frequency response. An electronic time-base circuit delivers a saw-tooth voltage to the x-plates which drives the electron beam across the x-axis at a constant rate and returns the beam to the left-hand side at the start of each sweep. This produces a dynamic image of signal amplitude against time. Alternatively, a second input signal may be applied to the x-plates to produce an x-y graphical plot, e.g. pressure–volume loop. Cathode ray oscilloscopes are widely used in electronic engineering and signal processing, but have been replaced in clinical practice by microprocessor-controlled displays.

Microprocessor-Controlled Displays

Digital signal processing has revolutionized clinical measurement. Modern monitors comprise a single system integrating various measurements of physiology, and display information as discrete measurements as well as continuous analogue waveforms (Fig. 16.1). This paradox, the conversion of analogue information into digital and then back to analogue, illustrates the real power of digital signal processing to manipulate and present information in a relevant and user-friendly manner. Data patterns (waveforms, trends, graphs) can be recreated or processed in other ways from the original digital signal to assist the anaesthetist, with the original digital signal being stored in a computer record without degradation of the quality of the signal.

Most of the current monitoring systems follow good ergonomic principles, with different variables separated consistently by position on the screen and by colour. This allows the most important information to be placed centrally in large symbols or fonts and in bold colours, with less important data either relegated to small print, or placed in submenus. However, the flexibility of most monitors implies that it is still possible for individuals to change colours and priorities, often making the monitor much less effective. Whenever possible, departments should ensure that all monitors have identical default settings to reduce confusion (these are usually password protected). Unfortunately, the lack of international standards means that confusion may still occur if monitors from multiple sources are used in the same unit.

Despite many attempts to simplify patient data into geometric shapes or bar graphs, data continue to be displayed most often as simple numbers, supported by waveforms, e.g. invasive pressure, and a graphical display of trends over time. Trends are particularly useful when clinical problems may produce gradual change. For example, in neurosurgery, a gradual decrease in end-tidal carbon dioxide concentration is often associated with multiple air emboli.

BIOLOGICAL ELECTRICAL SIGNALS

The detection and recording of biological electrical potentials are important clinical measurements which incorporate many of the key principles of clinical measurement.

Depolarization of the cell membrane of excitable cells is fundamental to the action of these cells and generates a transient potential difference between the active cell and surrounding tissues. The summation of synchronous extracellular potentials from a large number of excitable cells generates a widespread electric field which can be detected by electrodes on the body surface. The electrocardiogram (ECG) and electroencephalogram (EEG) are two well-established measures of biological electrical activity.

Biological electrical signals are detected using electrodes constructed of silver and electrolytically coated with silver chloride. Low, stable impedances minimize mains interference. Symmetrical electrode impedance and insignificant polarization control drift. However, care is still required to achieve optimum results. The silver chloride layer is very thin, prone to deterioration and only suitable for single use. Movement artefacts which alter the electrode potential and impedance are greatly reduced if the electrode surface is separated from the skin by a foam pad impregnated with electrolyte gel. It is no longer necessary to abrade the skin to achieve ultra-low impedance, but de-greasing with alcohol before applying the electrode helps to reduce skin impedance and ensures satisfactory adhesion.

Amplification

The amplitude of tiny bioelectrical signals must be increased by amplification, and unwanted noise and interference minimized. Calibration voltages may be incorporated for correct adjustment of the gain of the amplifier.

Input Impedance and Common Mode Rejection

Amplifiers for biological signals require high common mode rejection and high input impedance. The input and electrode impedances act as a potential divider: high electrode impedance and low amplifier input impedance attenuate the electrical signal across the amplifier. The input impedance of modern amplifiers exceeds 5 MΩ to avoid problems, and careful attention must be paid to minimizing electrode impedance, particularly for EEG recordings.

Differential amplification is a powerful method of reducing unwanted noise. The potential difference between two input signals is amplified, but electrical signals common to both are attenuated. This feature is termed ‘common mode rejection’ and very effectively reduces mains interference in all biological signals and electrocardiographic contamination of much smaller electroencephalographic signals. The common mode rejection ratio (CMRR) for a typical differential amplifier exceeds 10 000:1. In other words, a signal applied equally to both input terminals would need to be 10 000 times larger than a signal applied between them for the same change in output.

Frequency Response

The bandwidth of the amplifier must cover the range of frequencies which are important in the signal. In practice, amplifiers require a flat frequency response for ECG from 0.14 to 50 Hz, for EEG from 0.5 to 100 Hz and for EMG from 20 Hz to at least 2 kHz.

Low-frequency interference, largely caused by slow fluctuating potentials generated in the electrodes, produces baseline instability and drift. This is removed by incorporating a network of resistors and capacitors which function as a simple high-pass filter allowing biological signals to pass, but attenuating low-frequency noise. This introduces a compromise in amplifier design between signal trace fidelity and stability of recording. For example, amplifiers designed for diagnostic electrocardiography have long time constants with optimal reproduction of the waveform at the expense of baseline instability, especially to movement. In comparison, continuity of recording is more important when the electrocardiogram is used for monitoring during anaesthesia; high-pass filtering produces a short time constant and good baseline stability at the expense of waveform reproduction. Low-frequency elements of the ECG, such as the T wave, may become differentiated by phase shift in the high-pass filter and appear distorted or biphasic.

Other filters can attenuate particular frequencies. Highly selective band reject filters attenuate 50 Hz interference from the signal. Low-pass filters are used to eliminate higher-frequency artefacts from an EEG signal. The purpose of filtering is to reduce unwanted noise relative to the signal. When the frequency range of signal and noise overlap, some degree of signal degradation is inevitable.

Noise and Interference

Electrical noise arising from the patient, the patient-electrode interface or the surroundings may seriously interfere with accurate recording of biological potentials.

Noise Originating from the Patient–Electrode Interface: Recording electrodes do not behave as passive conductors. All skin–metal electrode systems employ a metal surface in contact with an electrolyte solution. Polarization describes the interaction between metal and electrolyte which generates a small electrical gradient. Electrodes comprising metal plated with one of its own salts, e.g. silver—silver chloride, avoid this problem because current in each direction does not significantly change the electrolyte composition. Mechanical movement of recording electrodes may also cause significant potential gradients – alteration in the physical dimensions of the electrode changes the cell potential and skin–electrode impedance. Differences in potential between two electrodes connected to a differential amplifier are amplified and asymmetry of electrode impedance seriously impairs the common mode rejection ratio of the recording amplifier.

Noise Originating Outside the Patient: Electrical interference. Mains frequency interference with the recording of biological potentials may be troublesome, particularly in electromagnetically noisy clinical environments. Patients function physically as large unscreened conductors and interact with nearby electrical sources through the processes of capacitive coupling and electromagnetic induction.

Capacitance permits alternating current to pass across an air gap. A live mains conductor and nearby patient behave as the two plates of a capacitor. The very small mains frequency current which flows through the patient is of no clinical significance but confounds the detection and amplification of biological potentials, creating unwanted interference in the recording. Capacitive coupled interference is minimized by reducing the capacitance and the alternating potential difference. This is achieved by moving the patient away from the source of interference and by screening mains-powered equipment with a conductive surround which is maintained at earth potential by a low-resistance earth connection and by surrounding leads with a braided copper screen – stray capacitances couple with the screen instead of the lead.

Alternating currents in a conductor generate a magnetic flux. This induces voltages in any nearby conductors which lie in the changing magnetic flux, including the patient or signal leads to the amplifier, which function as inefficient secondary transformers. This source of interference is minimized by keeping patients as far as possible from powerful sources of electromagnetic flux, especially mains transformers. Electromagnetic inductance may be minimized by ensuring that all patient leads are the same length, closely bound or twisted together until very close to the electrodes. This ensures that the induced signals are identical in all leads and therefore susceptible to common mode rejection.

The importance of low electrode impedance. Low electrode impedance may exaggerate the effects of surrounding electrical interference. Capacitive and inductive coupling produce very small currents in the recording leads. If the electrode impedance is low, the potential at the amplifier input must remain close to the potential at the skin surface, so that minimal interference results. If electrode impedance is high, the small induced currents may create a significant potential difference across that impedance, leading to severe 50 Hz interference.

Radiofrequency interference from diathermy is a severe problem for the recording of biological potentials. ECG amplifiers may be provided with some protection by filtering the signal before it enters the isolated input circuit, filtering the power supply to block mains-borne radiofrequencies and enclosing the electronic components in a double screen, the outer earthed and the inner at amplifier potential.

BIOLOGICAL MECHANICAL SIGNALS

Pressure is a mechanical signal fundamental to measurement and monitoring in anaesthesia. Several physical principles and a wide range of instruments are used to measure pressure. Liquid column manometers display pressure according to the height of a column of fluid relative to a predefined zero-point, and the density of the fluid. Mechanical pressure gauges are used widely, particularly in high-pressure gas supplies; pressure-dependent mechanical movement is amplified by a gearing mechanism which drives a pointer across a scale.

For most physiological pressure measurements, diaphragm gauges are used – a flexible diaphragm moves according to the applied pressure. Mechanical display of diaphragm movement is limited by poor sensitivity to small pressures, inertia to changing pressure and a narrow range of linear response. In modern diaphragm gauges used for sensing dynamic pressures, movement of the diaphragm is sensed by a device which converts the mechanical energy imparted to the diaphragm into electrical energy.

Electromechanical Transducers

The first step in transduction is movement of the diaphragm caused by the relationship to applied pressure. This depends on the stiffness of the diaphragm and substantially determines the operating characteristics of the transducer. Linearity of amplitude and frequency response are improved by using small stiff diaphragms which require a more sensitive mechanism for sensing diaphragm movement.

Wire strain gauges are based on the principle that stretching or compression of a wire changes the electrical resistance. Changes in capacitance or inductance have also been coupled to movement of a diaphragm. Silicon strain gauges use the changes in resistance in a thin slice of silicon crystal which occur when it is compressed or expanded. They are very sensitive and suitable for incorporation into a small stiff diaphragm with excellent frequency response, but non-linearity and temperature dependence are difficult technical problems.

Optical transduction senses movement of the diaphragm by reflecting light from the silvered back of the convex diaphragm on to a photocell. Applied pressure causes the silvered surface to become more convex. This causes the reflected light beam to diverge, reducing the intensity of reflected light sensed by the photoelectric cell. This design is used in a fibreoptic cardiac catheter for intravascular pressure measurement. These miniature pressure transducers are expensive but have a high-frequency response and fibreoptic light sources eliminate the risk of microshock.

THE CARDIOVASCULAR SYSTEM

The principal aim of an anaesthetist is to ensure the delivery of oxygen to the patient’s tissues. In physiological terms, oxygen delivery is the product of the cardiac output, the concentration of haemoglobin and its oxygen saturation. Clinically, if the patient is pink, with a normal volume pulse and has warm extremities, then these aims are being met. When combined with a urine output of greater than 0.5 mL h–1 it is unlikely that the patient has any cardiovascular problems. A further confirmatory test, especially useful in children, is the capillary refill time. When an extremity is compressed for 5 s, if capillary refill occurs in less than 1.5 s, cardiac output is adequate. If the refill time is greater than 5 s, then shock is likely to be present.

The need for direct patient observation cannot be overestimated. Literally having a ‘finger on the pulse’ and being able to see the patient are the most important safety factors. While factors such as drapes and dimmed theatre lights may make direct observation difficult, there should not be complete reliance on electronic monitoring.

Electrocardiography

The electrocardiogram (ECG) is a well-established measure of myocardial electrical activity. The synchronous depolarization and prolonged action potentials in cardiac muscle summate to generate a potential field of high amplitude. This potential difference is detected between two electrodes placed on the body surface. In the three-lead system commonly in use, the third lead is used as a reference electrode. The voltage changes are very small (1 mV in amplitude with a frequency response of 0.05–100 Hz) and require amplification before being displayed as the familiar waveform.

Different lead positions detect electrical activity from different parts of the myocardium. The commonest position of the electrodes used in the operating theatre is the CM5 arrangement, as this is the best position to detect ischaemia of the left ventricle (Fig. 16.2).

The ECG is a standard monitor used on all anaesthetized patients. The visible waveform allows the cardiac rhythm to be identified and may often be printed for further analysis. Alarms may be set to identify arrhythmias and brady/tachycardias. Many monitors are able to display a numerical value of the heart rate in addition to a measure of any ST segment depression/elevation produced by cardiac ischaemia/infarction. This may be displayed as a trend over time and the success of treatment observed.

Unfortunately, the relatively small voltages measured are easily swamped by skeletal muscle activity or surgical diathermy, often leading to false alarms. The signal may also be severely degraded if the gel of the electrodes has been allowed to dry out or if the weight of the leads is allowed to pull on the electrodes. In addition, the monitor only identifies ischaemia in a single area; multiple lead systems are required to monitor the whole myocardium. While the ECG has become a standard monitor, it adds little to the information provided by palpating the pulse. It must be remembered that electrical activity does not always produce a cardiac output. Complications are rare, although the electrode adhesive may produce skin damage in susceptible patients.

Arterial Pressure

Indirect Methods

An adequate arterial pressure is essential for tissue perfusion; even when perfusion is adequate, hypotension may lead to renal failure. Indirect methods of measuring arterial pressure do not depend on contact between arterial blood and the system for signal recognition and transduction. Arterial pressure may be most rapidly estimated by palpating a pulse, although this method is too unreliable as a single technique. The majority depend on signals generated by the occlusion of a major artery using a cuff, known as the Riva-Rocci method. Systolic pressure can be estimated by the return of a palpable distal pulse; auscultation of the Korotkoff sounds can determine systolic and diastolic pressures. These methods, however, are too time-consuming during anaesthesia and often impossible because of poor access to the patient’s arm.

Oscillometric Measurement of Arterial Pressure

The oscillometric measurement of arterial pressure estimates arterial pressure by analysis of the pressure oscillations which are produced in an occluding cuff by pulsatile blood flow in the underlying artery during deflation of the cuff (Fig. 16.3). The original automatic oscillometers used two cuffs. The upper cuff was inflated to occlude the arterial flow and then gradually deflated. As the blood flow began to pass under the upper cuff, the small changes in volume were detected by the lower cuff with an electromechanical pressure transducer. Modern machines use a single cuff with two tubes for inflation/measurement. During slow deflation, each pulse wave produces a pressure transient in the cuff which may be distinguished from the slowly decreasing ambient pressure in the cuff. Above systolic pressure, the transients are small, but suddenly increase in magnitude when the cuff pressure reaches the systolic point. As the cuff pressure decreases further, the amplitude reaches a peak and then starts to diminish. The mean arterial pressure correlates closely with the lowest cuff pressure at which the maximum amplitude is maintained. As the cuff pressure reaches diastolic pressure, the transients abruptly diminish in amplitude. To avoid high cuff pressures and long deflation times, monitors inflate the cuff to just above a normal systolic pressure and then slowly decrease the pressure until a pulse is detected. Consequently, estimates of diastolic pressure can be unreliable. If a pulse is not detected, the cuff is then inflated to a higher pressure. This process may be repeated several times before a measurement is made.

Commercial instruments incorporate mechanisms for improving the reliability of the measurement. For example, at each successive plateau pressure during the controlled deflation, successive pressure fluctuations are compared and accepted only if they are similar. All automatic oscillometric instruments require a regular cardiac cycle with no great differences between successive pulses. Accurate and consistent readings may be impossible in patients with an irregular rhythm, particularly atrial fibrillation.

Clinical studies comparing automatic oscillometric instruments with direct arterial pressure have demonstrated good correlation for systolic pressure with a tendency to overestimate at low pressures and underestimate at high pressures. Mean and diastolic pressures were less reliable. The 95% confidence interval for all three indices exceeded 15 mmHg. The disadvantages of automated oscillometry are shown in Table 16.3.

TABLE 16.3

Disadvantages of Automated Oscillometry

Delayed measurement with arrhythmias or patient movement

Inaccuracy with systolic pressure < 60 mmHg

Inaccurate if the wrong size cuff used

May be inaccurate in obese patients

Discomfort in awake patients

Skin and nerve damage in prolonged use

Delay in injected drugs reaching the circulation

Backflow of blood into i.v. cannulae

Pulse oximeter malfunction as cuff is inflated

Direct Measurement

To measure arterial pressure directly, a cannula (usually 20–22G parallel-sided Teflon) must first be inserted into an artery (usually the radial because occlusion of the artery may be compensated for by flow through the ulnar artery). As fluids are incompressible, the pressure in the artery is transmitted directly to a transducer, which converts pressure into an electrical signal which is displayed by the monitor. The cost and complexity of pressure transducers are compensated by convenience, accuracy, continuity of measurement and an electrical output which may be processed, stored and displayed according to the requirements. The transducer should be at the level of the left ventricle and the transducer opened to the atmosphere to provide a zero reading before use. Monitors usually display systolic and diastolic pressures as well as the mean pressure, calculated automatically by integrating the area under the pressure waveform. The waveform provides useful additional information: a rapidly appreciated estimate of pressure, a qualitative assessment of the adequacy of the frequency response and damping, and an assessment of relative hypovolaemia during positive pressure as identified by the variability or ‘swing’ in the waveform.

The advantage of direct measurements is a real-time measure of arterial pressure, which is essential when administering drugs such as vasopressors to critically ill patients (Table 16.4). Such measurement systems also provide a means for obtaining samples for arterial blood gas analysis and other blood tests. The use of arterial cannulae has therefore become standard practice for severely ill patients, both in the operating theatre and in the intensive care unit.

TABLE 16.4

Advantages of Direct Arterial Pressure Measurement

Accuracy of pressure measurement

Beat-by-beat observation of changes when blood pressure is variable or when vasoactive drugs are used

Accuracy at low pressures

Ability to obtain frequent blood samples

However, errors are common as a result of malpositioning of the transducers and failure to zero the transducer before use. For example, if the operating table is moved upwards while the transducer remains static, the difference in height artificially increases the pressure reading. Further, while modern disposable sets are usually reliable and accurate, they may occasionally malfunction. Unusual readings should therefore be checked against a reading from a non-invasive monitor. Complications relating to arterial cannulae are shown in Table 16.5.

TABLE 16.5

Complications Relating to Arterial Cannulae

Requires skill to insert

Bleeding

Pain on insertion

Arterial damage and thrombosis

Embolization of thrombus or air

Ischaemia to tissues distal to puncture site

Sepsis

Inadvertent injection of drugs

Late development of fistula or aneurysm

Resonant Frequency and Damping: Fourier showed that all complex waveforms may be described as a mixture of simple sine waves of varying amplitude, frequency and phase. These consist of a fundamental wave, in this case at the pulse frequency, and a series of harmonics. The lower harmonics tend to have the greatest amplitude and a reasonable approximation to the arterial pressure waveform may be obtained by accurate reproduction of the fundamental and first 10 harmonics. In other words, to reproduce an arterial waveform at 120 beats min–1 accurately would require transduction with a linear frequency response up to a frequency of at least (120 × 10)/60 = 20 Hz. Accurate reproduction of a waveform requires that both the amplitude and phase difference of each harmonic are faithfully reproduced. This requires a transduction system with a natural frequency higher than the significant frequency components of the system, and the correct amount of damping.

The fluid and diaphragm of the transducer constitute a mechanical system which oscillates in simple harmonic motion at the natural resonant frequency. This determines the frequency response of the measurement system (Fig. 16.4). The resonant frequency of a catheter-transducer measuring system is highest, and the frictional resistance to fluid flow which dampens the frequency response is lowest, when the velocity of movement of fluid in the catheter is minimized. This is achieved with a stiff, low-volume displacement diaphragm and a short, wide, rigid catheter.

Determination of the Resonant Frequency and Damping: The resonant frequency and the effects of damping may be estimated by applying a step change in pressure to the catheter-transducer system and recording the response (Fig. 16.5). The underdamped system responds rapidly but overshoots and oscillates close to the natural resonant frequency of the system; frequency components of the pressure wave close to the resonant frequency are exaggerated. By contrast, the overdamped system responds slowly and the recorded signal decreases slowly to reach the baseline, with no overshoot. High-frequency oscillations are damped, underestimating the true pressure changes. These extremes are undesirable.

Optimal Damping: Optimal damping maximizes the frequency response of the system, minimizes resonance and represents the best compromise between speed of response and accuracy of transduction. A small overshoot represents approximately 7% of the step change in pressure, with the pressure then following the arterial waveform (Fig. 16.5).

Damping is relatively unimportant when the frequencies being recorded are less than two-thirds of the natural frequency of the catheter-transducer system. Modern transducer systems using small compliance transducers connected to a short, stiff catheter, with a minimum of constrictions or connections, approximate to this ideal. The system also includes a pressurized bag of saline which produces a flow of 1–3 mL h–1 through a restrictor to prevent clot formation, as well as the facility to allow a higher flow rate to flush the system, for example after a blood sample has been taken. Air bubbles in the system, clotting or kinking in the vascular catheter, and arterial spasm lower the natural resonant frequency and increase the damping.

In clinical practice, the resonant frequency of the whole system is uncomfortably close to the frequency content of the signal, and accurate measurements require optimal damping. However, damping is difficult to measure and control, and is poor compensation for an inadequate frequency response in the pressure recording system. Adjustment of damping is difficult to achieve and mechanical methods which include inserting constrictions or a compliant tube into the system to increase damping further reduce the resonant frequency. Electronic damping of the electrical output from the transducer cannot correct for non-linear amplification and attenuation of frequencies in the pressure wave before transduction.

Accuracy of Arterial Pressure Measurements

The accuracy of pressure measurements, particularly using indirect methods, needs to be considered. Invasive direct measurement of arterial pressure is the usual standard for comparison. However, the catheter-transducer system must be carefully set up and tested for optimal performance and this is hard to achieve in clinical practice. Arterial pressure varies throughout the arterial tree and the measured pressure depends on the site of measurement. As the pulse wave travels from the ventricle to peripheral arteries, changes in vessel diameter and elasticity affect the pressure waveform, which becomes narrower with increased amplitude. Differences in arterial pressure between limbs are common, particularly in patients with arterial disease.

Indirect methods using an occluding cuff make intermittent measurements, with the systolic and diastolic readings reflecting the conditions in the artery at two instants at which the end-points are detected. By contrast, direct pressure measurements are the average of a number of cycles, more precisely reflecting mean pressures. Indirect measurements may be compromised by taking a small number of infrequent samples from a variable signal.

Central Venous Pressure

Central venous pressure is often considered a measure of the amount of blood within the venous system; a pressure less than normal (2–3 mmHg) indicates hypovolaemia and a higher pressure indicates volume overload. While such a view is reliable for healthy patients with acute blood loss, it is not so simple in other circumstances. For example, patients with damage to the right side of the heart may have raised central venous pressure even when the filling pressure of the left side of the heart is low. Single measurements rarely provide an accurate reflection of the fluid status of the patient. However, repeated measurements taken while a fluid challenge is given can be informative.

There are four common routes for central venous catheterization.

image Long catheters inserted via the antecubital fossa are relatively easy and safe to insert but are of small diameter. Catheters inserted via the basilic or cephalic vein are sometimes difficult to advance past the shoulder. It is also difficult to determine if the tip of the catheter is within a central vein without X-ray imaging. Thrombosis of the veins is common if the catheter is left in situ for more than 24 h.

image Femoral venous catheters are inserted just below the inguinal ligament. They are also relatively easy to insert and may be of large gauge to allow rapid transfusion of fluids. This route is often chosen in children. However, the site of insertion is often within a skin fold, making skin sepsis more likely.

image Internal jugular catheters are used most commonly because the vein is superficial, of larger diameter, and easily managed. This is the route which is often most appropriate for use in an emergency. However, the insertion point is adjacent to several vital structures, including the carotid artery, lung, brachial plexus and cervical spine, with the result that direct needle trauma to these structures can occur. Current guidelines recommend the use of an ultrasound probe for insertion of a catheter via the internal jugular route to improve the accuracy of insertion and to minimize complications.

image Subclavian catheters suffer the same problems as those in the internal jugular vein, although the point of insertion under the clavicle may make it easier to anchor the catheter to the skin. However, if accidental arterial puncture occurs, the overlying clavicle obscures bleeding and makes direct compression of the artery impossible. The proximity of the pleura is associated with a risk of accidental lung puncture. The subclavian route should therefore be used only when the internal jugular approach is contraindicated.

As the central venous pressure is relatively low, it may be measured using a simple manometer. However, a central venous catheter is usually connected to the same type of transducer and flush system described for arterial cannulae. This provides a continuous readout of pressure, allowing the effect of infusions of fluids to be assessed in real time. However, because the central venous pressure is low, great care is required to ensure that the pressure is measured relative to the correct zero point (the right atrium) on the patient (Fig. 16.6). Although a single reading of central venous pressure is of little diagnostic use, a change in response to fluid challenge is more useful. In general, if a fluid challenge has little effect on the central venous pressure, the patient is likely to be hypovolaemic. In contrast, a marked increase in pressure indicates fluid overload. However, in unwell patients, the correlation between the CVP response to a fluid challenge and circulating volume is poor. Caution is required in using this principle to guide fluid therapy in hospitalized patients.

Complications are infrequent but potentially serious, and are shown in Table 16.6.

TABLE 16.6

Complications of Central Venous Catheterization

Acute

Arrhythmias

Bleeding

Air embolus

Pneumothorax

Damage to thoracic duct, oesophagus, carotid artery, stellate ganglion

Cardiac puncture

Catheter embolization

Delayed

Sepsis

Thrombosis

Cardiac rupture

Pulmonary Artery Pressure

Although a central venous cannula may be used to estimate venous volume, it measures the filling of the right side of the heart. However, cardiac output and systemic arterial pressure are determined primarily by the filling pressure of the left side of the heart. When introduced, the pulmonary artery flotation catheter (PAFC), with its ability to measure cardiac output and left atrial pressure, appeared to be a major advance. Recently, however, frequent complications, a lack of evidence of improved survival and the introduction of non-invasive techniques to estimate cardiac output have led to a decline in its use. The PAFC is also known as a Swan-Ganz or balloon tip catheter (Fig. 16.7).

A PAFC is a long catheter with three or four lumens, and a thermistor near the tip. It is inserted into a neck vein through a large cannula. A flexible plastic sheath allows the catheter to be inserted, withdrawn and rotated after insertion without desterilizing it. After insertion into the superior vena cava, saline is injected to inflate a balloon at the tip. The pressure at the tip is measured via a transducer and displayed on a monitor. The catheter is then advanced slowly so that the blood flow directs the catheter toward the pulmonary artery. As the catheter is advanced, a series of changes in pressure is observed, marking the progression through the right atrium and right ventricle into the pulmonary artery (Fig. 16.8). Eventually, the balloon ‘wedges’ into a pulmonary artery. At this point, the tip is isolated from the pulmonary artery and measures the pressure in the pulmonary capillaries, which is taken to reflect left atrial pressure. Although the ability to estimate left atrial pressure is useful, the interpretation of measurements is as problematic as for central venous pressure (see above). For the same reasons, measuring the changes after a fluid challenge is more useful than a single reading.

The ability to measure the filling pressure of the left ventricle as well as the cardiac output was a major advance, and has led to many advances in our understanding of cardiac physiology and the mechanisms and treatments of diseases such as sepsis. However, the process of insertion described above is not always straightforward and prolonged manipulation may be needed to direct the catheter into the pulmonary artery. Arrhythmias are extremely common with catheter insertion and the technique carries all the risks of central venous catheterization noted above in addition to the risks shown in Table 16.7.

TABLE 16.7

Risks of Pulmonary Artery Catheterization (in Addition to those Shown in Table 16.6)

Arrhythmias with catheter manipulation

Damage to tricuspid and pulmonary valves

Knotting of catheter

Pulmonary infarction if balloon left inflated

Pulmonary artery rupture with balloon inflation

Cardiac rupture

Cardiac Output

Cardiac output is closely linked to oxygen delivery; in addition, studies have shown that low cardiac output is linked to increased mortality. However, it must be remembered that cardiac output is intermittent and not continuous and that factors such as the pressure changes caused by ventilation (especially positive pressure ventilation), changes in heart rate and arrhythmias produce complex changes. Therefore, monitors do not produce consistent results even when synchronized with the heartbeat and respiratory cycle.

Dilution techniques have been regarded as the ‘gold standard’ against which other methods are compared.

The Fick Principle: The Fick principle defines flow by the ratio of the uptake or clearance of a tracer within an organ to measurements of the arteriovenous difference in concentration. It may be used to measure cardiac output, notably when applied to oxygen uptake or indicator dilution, and also regional blood flow, e.g. cerebral blood flow using the uptake of nitrous oxide, and renal blood flow from the excretion of compounds cleared totally by the kidney, such as para-aminohippuric acid.

In subjects with minimal cardiac shunt and reasonable pulmonary function, pulmonary blood flow may be calculated from the ratio of the oxygen consumption and the difference in oxygen content between arterial and mixed venous blood, as follows:

image

Oxygen consumption from a reservoir is measured using an accurate spirometer, and oxygen content of blood requires a co-oximeter. The patient should be at steady state when the measurements are made, with constant inspired oxygen concentration and blood samples obtained slowly whilst the oxygen consumption is being determined. True mixed venous blood samples must be obtained from a pulmonary artery catheter, and alternative indicator dilution techniques described below are less demanding. The effects of ventilation and beat-to-beat variation in cardiac output are averaged over the long period of measurement of oxygen consumption. Errors in measurement of oxygen consumption limit the accuracy of this technique (± 10%).

Indicator Dilution: An indicator is injected as a bolus into the right heart and the concentration reaching the systemic side of the circulation is plotted against time (Fig. 16.9). The average concentration is calculated from the area under the concentration–time curve divided by the duration of the curve. The cardiac output during the period of this measurement is the ratio of the dose of indicator to the average concentration.

The general formula is:

image

The main problem with this technique is that when the dye has been measured at the artery it passes back to the heart and then back to the arteries (known as recirculation), making the calculations more complex. This may be circumvented by extrapolation of the early exponential downslope to define the tail of the curve which would have been recorded if recirculation had not occurred (Fig. 16.9). The area under the curve is calculated by integration.

Thermal Indicator Dilution: This technique requires a PAFC to be in the pulmonary artery. The principle of the method is similar to other indicator dilution methods, but the injection and sampling are performed on the right side of the heart. A bolus of 10 mL of saline at room temperature is injected into the right atrium and the temperature change is recorded by a thermistor at the tip of the PAFC in the pulmonary artery. The smaller the temperature drop, the larger is the cardiac output. The recorded temperatures generate an exponential dilution curve with no recirculation. The ‘heat dose’ is the product of the difference in temperature between the injectate and blood multiplied by the density, specific heat and volume of the injectate. The average change in heat content is the area under the temperature–time graph multiplied by the density and specific heat of blood.

Thermal dilution techniques offer many advantages:

Disadvantages of thermal dilution include the following:

The measurement of cardiac output using both dye and thermodilution is now automated with computer-controlled sampling, calculation of indicator dilution curves, rejection algorithms for artefacts or curves which are not exponential, and on-line calculation of cardiac output.

‘Continuous’ cardiac output monitors have also been introduced. These use a similar principle, but instead of using a bolus of cold saline, the catheter has an electrical coil which is heated at intervals, creating a bolus of warm blood which passes into the pulmonary artery. This eliminates much of the operator error and produces frequently updated measurements of cardiac output, allowing the effect of interventions to be observed.

Monitors have also been developed which do not require a pulmonary artery catheter, but use a bolus of iced saline injected into a modified central venous catheter and a peripheral arterial catheter with a built-in thermistor (e.g. the PiCCO monitor).

Most of the current monitors allow cardiac output data to be integrated with other measurements such as arterial pressure to provide calculated values of, for example, systemic vascular resistance and stroke volume. This aids the choice and administration of drugs such as vasoconstrictors.

Pulse Contour Analysis: The shape of the arterial pulse (the pulse contour) is a product of the rate of ejection of blood into the aorta and the elasticity of the arterial tree. Therefore, if some assumptions are made about the arterial tree, the volume ejected at each heartbeat (stroke volume) may be calculated from the shape of the arterial pulse contour. Multiplying this by the heart rate provides an estimate of cardiac output. This system has the advantage of being able to calculate the cardiac output in near real-time using an arterial cannula alone.

However, the technique relies on assumptions on arterial tree elasticity which may not always be correct in every patient. Therefore, these systems often require calibration by another method such as thermodilution every 8–12 h to ensure accuracy. The costs of the computer and consumables are also considerable.

Doppler Ultrasonography: Ultrasound techniques can detect the shape, size and movement of tissue interfaces, especially soft tissues and blood, including the echocardiographic measurement of blood flow and the structure and function of the heart. Sound waves are transmitted by the oscillation of particles in the direction of wave transmission and are defined by the amplitude of oscillation (the difference between ambient and peak pressures) and the wavelength (distance between successive peaks) or frequency (inversely proportional to wavelength, the number of cycles per second). These characteristics are measured by a pressure transducer placed in the path of an oncoming wave. The human ear detects frequencies within the range of 20–20  000 Hz. Diagnostic ultrasound uses frequencies in the range of 1–10 MHz. Short-term diagnostic use of ultrasound appears to be free from hazard.

Properties of Ultrasound: Shorter wavelengths and higher frequencies improve the resolution of distance, but tissue penetration is simultaneously reduced. Amplitude determines the intensity of the ultrasound beam, the number and size of echoes recorded and therefore the sensitivity of the instrument. Ultrasound is absorbed by tissues and reflected at tissue interfaces. The intensity of the beam decreases exponentially as it passes through tissue. Attenuation depends on the nature and temperature of the tissue, and is related linearly to the frequency of the ultrasound.

The reflection of the ultrasound beam from the junction between two tissues or from tissue–fluid or tissue–air interfaces forms the basis of the majority of diagnostic techniques. Reflections at most soft-tissue interfaces are therefore weak, but bone–fat and tissue–air interfaces reflect the majority of incident energy. Structures lying behind a bone or air interface cannot be studied using ultrasound.

Ultrasound scanning techniques have been developed which are suited to different applications and which have extremely sophisticated two-dimensional, real-time, brightness- and colour-modulated displays under microprocessor control.

Detection of Motion by the Doppler Effect: Cardiac Output: When ultrasound waves reflect off an object moving towards the transmitter, there is an apparent increase in frequency as the object encounters more oscillations in a given time. This physical phenomenon is termed the Doppler effect. The change in frequency is proportional to the velocity of the object and two constants: the frequency of the transmitted ultrasound and the velocity of ultrasound in the medium. The velocity of the object can be calculated using the Doppler equation:

image

(Fd = change in Doppler frequency; C = speed of sound in medium; Ft = transmitted frequency; θ = angle of probe relative to the flow of blood)

In practice, a beam of ultrasonic waves is focused on the descending aorta and reflections from red cells are measured by a transducer in the same probe. Probes may be transthoracic (usually placed in the sternal notch) or placed in the oesophagus. The speed–time curve of the red cells (Fig. 16.10) is integrated to calculate the average velocity over each cardiac cycle. The stroke volume can be calculated by multiplying the average velocity per cycle with an estimation of the cross-sectional area of the aorta (using pre-determined values based on population data, or estimated echocardiographically).

Current Oesophageal Doppler Monitors derive the total cardiac output utilising a nomogram created by ‘calibration’ of total left ventriculated Stroke Volume as measured by the pulmonary artery catheter against descending aortic blood flow velocity and Stroke Distance as measured by the ODM.

The advantage of these systems is that they produce an almost real-time reading of cardiac output so that changes in output in response to drugs or fluids are seen almost immediately, and so are a useful guide to further therapy, particularly fluid responsiveness.

However, there are a number of problems with this technology. These systems are reliant on the ultrasound beam being directed at the centre of the aorta and the aorta being a smooth tube. In practice, even small movements of the sensor may lead to marked changes in readings because the speed of red cells near the aortic wall is measured. Furthermore, the aorta is not completely circular, may also contain atheroma, and the diameter (usually an estimate) can change by as much as 12% during systole. The Doppler shift also depends on the direction of the ultrasound beam relative to the axis. Provided that the angle is less than 20°, the error in cardiac output is only about 6%. Conscious patients do not always tolerate the more accurate oesophageal probes, and certain surgical procedures preclude their use (e.g. oesophagectomy surgery). Most machines now provide a visual (and audible) measure of signal strength to allow the user to identify when the probe has moved.

Despite these limitations, these probes are very useful in high-risk patients who require relatively minor surgery because they may be used to monitor cardiac output during anaesthesia without the risks associated with the insertion of multiple intravascular catheters. They are also useful in patients undergoing surgery requiring precise control of intravascular volume and cardiac output. However, although they may be used to determine whether a change in therapy has had a positive or negative effect, they are unable to provide reliable estimates of cardiac output in absolute terms. Complications are rare.

Transoesophageal Echocardiography: Transoesophageal echocardiography uses a miniaturized ultrasonic probe inserted into the oesophagus under anaesthesia. It provides a real-time picture of all four cardiac chambers and valves. Its advantage is that it can identify any malfunctioning valves in addition to any wall-motion abnormalities related to myocardial ischaemia. It can also identify if therapy has successfully treated the ischaemia.

However, these probes are expensive to purchase and require an operator who is trained both to use the equipment and to interpret the results. They are useful as patients come off bypass after cardiac surgery to ensure that both the myocardium and valves are functioning correctly. Through use of Doppler, they can also measure cardiac output. However, they remain suitable only for anaesthetized or sedated patients, and cannot provide prolonged continuous measurements. They are rarely used in non-cardiac surgery.

Thoracic Electrical Bioimpedance: Tissue impedance depends on blood volume. Measurement of thoracic impedance provides an index of stroke volume. Two circumferential electrodes are placed around the neck and two around the upper abdomen. A small (< 1 mA) constant, high-frequency (> 1 kHz) alternating current is passed between the outer electrodes and the resulting potential difference is detected by the inner pair. This potential is rectified, smoothed and filtered to record voltage fluctuations which reflect changes in impedance due to ventilation and cardiac activity. The cardiac activity is extracted by signal-averaging relative to the ECG R wave. This represents changes in thoracic blood volume and clearly resembles the pulse waveform.

Modern instruments show a modest agreement with invasive measurements of cardiac output, although trends and rapid changes in cardiac output are reliably demonstrated. This method is inaccurate when there are intracardiac shunts or arrhythmias, and underestimates cardiac output in a vasodilated circulation.

THE RESPIRATORY SYSTEM

Clinical

Continuous visual monitoring of the colour and the pattern of ventilation of the patient are essential for safe anaesthesia. When the patient is breathing spontaneously, observation should detect signs of airway obstruction, e.g. tracheal tug, paradoxical movement and failure of the anaesthetic reservoir bag to move.

Maintenance of the airway in an anaesthetized patient is a skilled task. Although the introduction of supraglottic airways has led to a reduction in the frequency of need for airway skills, airway maintenance still remains one of the most basic tasks for the anaesthetist. Auscultation of the chest with a stethoscope may confirm the presence of normal breath sounds and also detect additional sounds caused by secretions, oedema or bronchospasm. Although useful, clinical signs cannot reliably disprove oesophageal placement of a tracheal tube.

Airway Pressure

The lungs are damaged easily and although anaesthetic machines incorporate pressure relief valves, these are designed to protect the machine rather than the patient. Although electronic ventilators usually allow a maximum airway pressure to be set to protect the patient, excessive pressure may still be exerted by manual compression of the reservoir bag. The self-inflating bags used for resuscitation are often capable of exerting extremely high pressures. As even transient peaks of pressure may lead to lung trauma or pneumothorax, a pressure monitor should be used whenever positive-pressure ventilation is used.

As all ventilators now incorporate pressure monitors, separate airway pressure monitors are rarely used. Most ventilators include a pressure transducer in the form of a piezoelectric crystal that converts pressure to an electrical potential, which is then measured by the monitor and displayed.

Although such monitors are both accurate and reliable, it must be remembered that they measure the pressure within the monitor and not the airway pressure. Therefore, the measured pressure may not be reliable if a narrow tracheal tube, long breathing circuit or high-frequency ventilation are used. High lung pressures may be a particular problem in obese patients, those positioned head down and those with bronchospasm.

In devices without electronic components, such as ventilators used for transport, pressure is measured by devices in which the air pressure deforms a bellows or a metal tube. The deformation is linked to a needle with the pressure read from a scale. These are simple devices and are usually reliable, but are susceptible to damage by excess pressure.

Measurement of Gas Flow and Volume

The relationships between volume, flow and velocity are central to understanding gas flow and volume. Flow rate is defined as the volume passing a fixed point in unit time, i.e. volume per second. Integration of a continuous flow signal is the volume which has flowed over a defined period. Velocity is the distance moved by gas molecules in unit time. These are related directly and depend upon the cross-sectional area of flow:

image

The concept of velocity is important in flow measurement because several instruments measure the velocity of flow and not the flow rate. The velocities of all molecules in a gas or liquid are not the same. Axial streaming is characteristic of laminar flow (Fig. 16.11).

Measuring Volume

Measurements of gas volume depend on collecting the gases in a calibrated spirometer, or passing the gases through some type of gas meter. However, volume can also be derived from measuring gas flow. Gas flow when integrated over time produces a calculated volume.

Spirometers

Wet spirometers consist of a rigid cylinder suspended over an underwater seal and counterbalanced. Gas entering the bell causes it to rise. This linear displacement is proportional to the volume of gas. Wet spirometers are accurate at steady state and have been used to calibrate other volume-measuring devices. However, they are bulky and inconvenient to use, and the frequency response is damped by friction between the moving parts, causing the instrument to under-read with rapidly changing rates of flow.

Dry spirometers are more convenient for clinical work. Gas displaces a rolling diaphragm or bellows and the expansion is recorded and related to gas volume. The ‘Vitalograph’ is a specialized type of bellows spirometer used for lung function testing. The patient makes a maximal forced exhalation into the spirometer through a wide-bore tube. The expansion of the wedge-shaped bellows is recorded by a stylus on a pressure-sensitive chart. The stylus moves across the x-axis (time) at a constant rate. The resultant plot represents the volume–time plot of the patient’s expiration (Fig. 16.12). The FVC is the maximal volume expired. Understanding of technique and active cooperation of the patient are essential for accurate and precise recordings. The patient must make an airtight seal with the mouthpiece and the nose is occluded with a nose clip. Expiration should be as forcible and rapid as possible. Several attempts are recorded. The highest value measured is recorded because the technique is dependent on voluntary effort, which usually improves with practice.

The Wright Respirometer

This device contains a light mica vane which rotates within a small cylinder (Fig. 16.13). Inflowing air is directed on to the vane by tangential slits. Rotation of the vane drives a gear chain and pointer on a dial. This mechanism is described as inferential because it does not measure either the volume or the flow of all of the gas flowing through the device. It is calibrated for normal tidal volumes and breathing rates by a sine wave pump. However, the meter seriously over-reads at high tidal volumes and under-reads at low tidal volumes because of the inertia of the moving parts.

Indirect Methods of Measuring Tidal Volume

Methods which depend on measuring inspired or expired gases require a leak-free connection. This is feasible with tracheal intubation. A physiological mouthpiece and connection to the measuring apparatus may change the pattern of breathing and is suitable only for short-term use. Several indirect methods have been developed which enable tidal volume to be derived from measurements of chest wall movement.

Measuring Gas Flow

Variable Orifice (Constant Pressure Change) Flowmeters

The orifice through which gas flows enlarges with the flow rate so that the pressure difference across the orifice remains constant. This is the physical principle of the Rotameter.

Rotameter: The Rotameter consists of a vertical glass tube inside which rotates a light metal alloy bobbin (Fig. 16.14). The flow of gas is controlled by the fine-adjustment flow control valve at the bottom of the Rotameter, and when this is opened, the pressure of the gas forces the bobbin up the tube. The inside of the tube is shaped like an inverted cone, so that the cross-sectional area of the annular space exactly opposes the downward pressure resulting from the weight of the bobbin. The pressure decrease remains constant throughout the range of flows for which the tube is calibrated and the bobbin rotates freely in the steady stream of gas. Each Rotameter must be calibrated for a specific gas. Laminar flow predominates at low flow rates and depends on the viscosity of gas. Turbulent flow increases at higher flow rates and the density of the gas becomes an important factor. Both density and viscosity of a gas vary with temperature and pressure, and each Rotameter must be calibrated for one specific gas in appropriate conditions.

The Peak Flowmeter: This useful clinical instrument is capable of measuring flow rates up to 1000 L min–1. Air flow causes a vane to rotate or a piston to move against the constant force of a light spring. This opens orifices which permit air to escape. The position adopted by the vane or piston depends primarily on the flow rate and on the area of the orifice which must be exposed to the air flow to maintain a constant pressure. The light moving vane or piston rapidly attains a maximum position in response to the peak expiratory flow. It is held in this position by a ratchet. The reading is obtained from a mechanical pointer which is attached to the vane or piston.

Accurate results demand good technique. These devices must be held horizontally to minimize the effects of gravity on the position of the moving parts. The patient must be encouraged to exhale as rapidly as possible. Consistency of repeated recordings suggests maximal effort and the peak expiratory flow rate is the maximum reading recorded.

Variable Pressure Change (Fixed Orifice) Flowmeters

The resistance is maintained constant so that changes of flow are accompanied by changes in pressure across the resistance element.

Pneumotachograph: The pneumotachograph measures flow rate by sensing the pressure change across a small but laminar resistance. Careful design ensures that the differential manometer senses the true lateral pressure exerted by the gas on each side of the resistance element (Fig. 16.15). The differential manometer needs to be very sensitive to record the tiny changes in pressure across the resistance and transduce them to a continuous electrical output. This signal may be integrated to give volume and the manometer must have good zero and gain stability.

Pneumotachographs are sensitive instruments with a rapid response to changing gas flow and are used widely for clinical measurement of gas flows in respiratory and anaesthetic practice. However, practical application requires frequent calibration and correction or compensation for differences in temperature, humidity, gas composition and pressure changes during mechanical ventilation. They are also susceptible to particle blockage, particularly by water condensation.

Other Devices for Measuring Gas Flow

Measurements other than pressure change across an orifice have been used to measure flow.

GAS AND VAPOUR ANALYSIS

Measurement of the concentration of gases and vapours in anaesthetic breathing systems is vital to prevent hypoxaemia and ensure the delivery of anaesthetic agents.

Physical Methods

In contrast with chemical methods, instruments based on the physical properties of a gas or vapour are convenient, responsive and more suitable for continuous operation.

Speed of response of the system is determined by two components:

Zero drift and variations in gain are common problems and most gas analysers have to be calibrated frequently, ideally against gas mixtures of known composition.

Non-Specific Methods

Non-specific methods use a property of the gas which is common to all gases, but which is possessed by each gas to a differing degree.

Refractive Index: Interference Refractometers

The speed of light slows through transparent materials to a degree determined by the refractive index of that substance. The delay caused by the passage of light through the gas depends on the number of gas molecules present; hence the refractive index also depends on the pressure and temperature of the gas. This extremely small delay is measured using the phase lag by the principle of interference. When light waves from a common source are passed through two linear slits in an opaque sheet and focused on to a screen, an interference pattern is produced. Bright areas where light from the two sources in phase is reinforced alternate with dark bands where the light paths differ in length by half a wavelength, i.e. they are out of phase and attenuating each other. When a gas is introduced into one light path, it delays transmission of the light waves with a reduction in wavelength and an alteration in the position of the dark bands. If the refractive index of the gas is known, the change in position can be related to the number of gas molecules in the light path and hence to the partial pressure of the gas. Interference refractometers are calibrated using known concentrations of gas or vapour. The response is essentially linear and remains stable after calibration.

This method of analysis is used to calibrate flowmeters and vaporizers accurately. Portable devices are useful for monitoring pollution by anaesthetic gases and vapours.

Specific Methods

Specific methods identify and measure a gas using some unique property and are particularly suitable for complex mixtures of gases. These methods include:

Their principles may be explored further by reference to specific gases relevant to the practice of anaesthesia.

Oxygen

Oxygen concentration in a breathing system is measured using either a fuel cell or a paramagnetic analyser.

The fuel cell is the more common device. It contains a lead anode within a small container of electrode gel. When exposed to oxygen, the lead is converted to lead oxide, producing a small voltage which may be measured and amplified. Fuel cells are small, robust and reliable, although they require calibration at regular intervals. After a period of around 6 months, they require replacement because the lead becomes oxidized. Accuracy is better than  ±  1% with a response time of < 10 s.

The principle of the paramagnetic analyser is that oxygen molecules are attracted weakly to a magnetic field (paramagnetic). Most other anaesthetic gases are repelled by a magnetic field (diamagnetic). In the original analysers, a powerful magnetic field was passed across a chamber which contained two nitrogen-containing spheres suspended on a wire. When oxygen was introduced into the chamber, it tended to displace the spheres, causing them to rotate. The degree of rotation was measured to estimate the oxygen concentration.

A fast differential paramagnetic oxygen sensor has been designed on the pneumatic bridge principle. The sample and reference gas are drawn by a common pump through two tubes surrounded by an electromagnet alternating at 110 Hz. Pressure differences between the two tubes are related to the paramagnetic properties of the sample and reference gases. The phasic changes in pressure are extremely small and measured with a miniature microphone. The output of the device is linear, very stable and has a fast response time of less than 150 ms.

These monitors are accurate, reliable and do not need frequent maintenance. The monitors measure the partial pressure of oxygen, but display oxygen as a percentage. If the pressure within the circuit is increased, for example when a gas-driven ventilator is employed, they can overestimate the oxygen concentration.

Carbon Dioxide and Anaesthetic Gases

Absorption of Radiation: Infrared radiation (1–15 μm) is absorbed by all gases with two or more dissimilar atoms in the molecule. Carbon dioxide, nitrous oxide and anaesthetic vapours absorb light at different wavelengths. Therefore, a cell is arranged with light sources on one side of a chamber and photoelectric cells on the other. Infrared light is dispersed through a prism or diffraction grating into a spectrum of different wavelengths. The test gas is then passed through the chamber and the amount of light absorbed is measured. The absorption spectrum is specific to each gas, and according to Beer–Lambert laws (see ‘Oximetry’), the amount of light absorbed is proportional to the concentration of the gas. In practice, the chambers have mirrors on each side so the light passes across the chamber many times to amplify the absorption. The chamber is also heated to avoid condensation. Accuracy is around 0.5% with a response time of < 0.5 s.

There are several sources of error with infrared analysis:

image The absorption wavebands of different gases may be coincident. For instance, the peak absorption bands for carbon dioxide, nitrous oxide and carbon monoxide are at 4.3, 4.5 and 4.7 μm respectively, and the absorption spectra inevitably overlap. Error is minimized by narrowing the band of infrared light

image The phenomenon of ‘collision broadening’ describes the apparent widening of the absorption spectrum of CO2 by the physical presence of certain other gases, notably N2 and N2O. Correction factors have been described, but the error may be minimized by calibrating the instrument with similar background gas mixtures as the gas to be analysed

image Absorption is related to the number of molecules in the absorbent gas in the cuvette, i.e. partial pressure. The reading is affected by changes in atmospheric pressure, pressurization of a breathing system or variation in the resistance of the sampling flow line

image Unexpected vapours, such as ethanol from an intoxicated patient, may introduce errors.

Modern gas analysers for clinical use are very stable but require regular calibration of the zero point and scale. Accuracy at normal breathing frequencies also requires a satisfactory response time, typically a 90% or 95% rise time less than 150 ms. Slow response is usually caused by blockage of the sampling line with condensation or sputum, or failure of the suction pump.

Most analysers pump a small flow of gas out of the breathing system to be analysed in the main monitoring box, a ‘side stream’ system. This involves some delay while the gas is pumped to the monitor. The movement of gas may also reduce accuracy if mixing of inspiratory and expiratory gas occurs. In most cases, the sampled gas is passed into the scavenging system, but when low flows are required it may be returned to the breathing system. This arrangement allows the sensing chamber to be housed within a monitor, making it more robust.

The alternative ‘main stream’ system places the sensing chamber in a connector within the patient breathing system and so reduces any delay in measurement. However, it also makes the sensor more prone to accidental damage.

The carbon dioxide concentration in respired gases is displayed most often as a graph of concentration against time (capnogram). This provides visual confirmation that the airway is patent and that ventilation is occurring. It also provides the only reliable guarantee after tracheal intubation that the tube is not in the oesophagus.

At the start of expiration, the carbon dioxide concentration is zero (dead space gas). The concentration then increases to a plateau level (alveolar gas). The end-tidal value of carbon dioxide concentration is used usually as a measure of the adequacy of ventilation because it approximates to alveolar and therefore arterial carbon dioxide partial pressure. However, when the respiratory rate is high, if tidal volume is low, if the sampling point is distant from the airway or if the gases tend to mix in the circuit, the ‘end-tidal’ value tends to be artificially low. This may give the impression that the lungs are being hyperventilated. For these reasons, it is difficult to measure end-tidal carbon dioxide meaningfully in small children. This is also true in patients, often smokers, who have marked ventilation/perfusion mismatch; there is often a prolonged upstroke on the capnograph trace and the relationship between end-tidal and arterial carbon dioxide tensions becomes less reliable (Fig. 16.16). If the capnograph trace does not appear to resemble a square wave, problems should be suspected and the arterial carbon dioxide partial pressure should be checked by blood gas analysis.

If lung perfusion is compromised, either by a low cardiac output or by pulmonary emboli or an air embolus, the end-tidal carbon dioxide tension decreases because carbon dioxide delivery to the lungs decreases. Paradoxically, arterial carbon dioxide partial pressure increases. Therefore, in situations where air emboli are likely, for example in neurosurgery, any alteration in end-tidal carbon dioxide should be investigated by blood gas analysis.

Although infrared analysers are used usually for measuring the concentration of anaesthetic agents, other methods are used occasionally either for calibration or for complex analyses.

Mass Spectrometry: Mass spectrometers are capable of separating the components of complex gas mixtures according to their mass and charge by deflecting the charged ions in a magnetic field. When a sample is introduced into a mass spectrometer, it passes into a vacuum where it is bombarded with high-energy electrons. These break up larger molecules and strip off their outer electrons. The resulting positively charged ions are then accelerated by a negatively-charged plate into a magnetic field. The magnetic field causes the moving particles to curve depending on how heavy they are (their mass:charge ratio). A row of sensors then measures the number of molecules and their sizes, in proportion to the partial pressure of the sampled gas. A mass spectrum is produced by relating the detector output on the y-axis (calibrated to concentration of gas) to the accelerating voltage on the x-axis (calibrated to molecular weight).

Some molecules may lose two electrons and become doubly charged – they behave like ions with half the mass. Some fragmentation of molecules also occurs in the ionization process, resulting in the production of a mass spectrum rather than a single peak for each molecule. These secondary peaks may be used to advantage, e.g. in the identification and quantification of CO2 and N2O, both of which share a parent peak at 44 Da, but produce secondary peaks at 12 and 30 Da, respectively.

Mass spectrometers are expensive to purchase and maintain, but are extremely accurate, have a very short response time, use very small sample flow rates (approx. 20 mL min− 1) and can identify a wide range of compounds. They may be sited centrally within large theatre complexes as part of a calibration and quality control system.

Gas–Liquid Chromatography: A gas chromatograph consists of two components: a column packed with inert beads covered in a thin film of oil (‘the stationary phase’), and a constant stream of inert gas which passes through the column. When a sample of gas is introduced at one end, the mixture passes into the column and past the oil. Insoluble gases tend to stay in the carrier gas and move through the column quickly, while soluble gases tend to dissolve in the oil, slowing their progress. At the other end of the column is a non-specific detector unit which yields an electronic signal proportional to the quantity of each substance present. Commonly used detectors include katharometers, flame ionization and electron capture detectors. Thus, any gas may be identified by the time it takes to pass through the column and its quantity measured by the detector unit. Their chief advantage is the ability to identify the components in a mixture of unknown compounds.

In addition to gas analysis, the gas–liquid chromatograph may be used to analyse blood samples containing volatile or local anaesthetic agents, anticonvulsants and intravenous anaesthetic drugs.

BLOOD GAS ANALYSIS

The Glass pH Electrode

A potential difference is generated across hydrogen ion-sensitive glass depending on the gradient of hydrogen ions. The hydrogen ion concentration within the pH electrode is fixed by a buffer solution, so that the potential across the glass is dependent on the hydrogen ion concentration in the sample (Fig. 16.17).

Measurement of this potential gradient is complicated by the difficulty in making stable electrical contact with the sample and buffer solutions. Two silver:silver chloride electrodes generate an electrode potential, but this is constant at a fixed temperature and provides a stable electrical connection with the buffer solution in the pH electrode, and with a potassium chloride solution in the reference electrode separated from the test sample by a semi-permeable membrane. The potential difference between the electrodes is determined by the pH of the test solution and the temperature, and is calibrated using two phosphate buffers of known and fixed pH. Careful daily calibration is required to maintain accuracy and the electrodes must be regularly cleaned of protein deposits. Reliable measurement of blood pH also depends on the quality of the blood sample, which must be free from air bubbles, heparinized and analysed promptly.

Dissociation of acids and bases is temperature- dependent and the electrodes and blood sampling channel are maintained at 37°C. The measured pH is then corrected to indicate the pH at the temperature of the patient.

Oxygenation

Oxygenation may be assessed by measuring the tension, saturation or content of oxygen, the relationship between these three measurements being determined by the shape and position of the oxyhaemoglobin dissociation curve. There are many causes of variations in both the shape and position of the curve and it is usually necessary to measure the oxygen tension or saturation directly. Tension measurements are required for most respiratory problems, although saturation or content may be required for calculation of the percentage shunt.

Oxygen tension is usually measured using an oxygen electrode. Content is measured by vacuum extraction and chemical absorption, by driving the O2 into solution and measuring the increase in PO2 or by a galvanic cell analyser. Saturation is determined by photometric techniques, involving the transmission or reflection of light at certain wavelengths.

Oxygen Tension

Oxygen Electrode: the Polarographic Method: The oxygen electrode (Clark) consists of a platinum wire, nominally 2nm in diameter, embedded in a rough-surfaced glass rod. This is immersed in a phosphate buffer which is stabilized with KCl and contained in an outer jacket which incorporates an oxygen-permeable polyethylene or polypropylene membrane (Fig. 16.18). A polarizing voltage of between 600 and 800 mV is applied to the platinum wire and as oxygen diffuses through the membrane electro-oxidoreduction occurs at the cathode:

image

Corresponding oxidation occurs at the Ag:AgCl anode:

image

image

Thus a half cell is set up and a tiny current is generated dependent on the oxygen tension at the platinum cathode. The change in current is measured as a change in voltage using the same potentiometric circuit as the pH and PCO2 measurement systems. The oxygen electrode may be used with gas mixtures or blood. Two-point calibration includes zero with an oxygen-free reference gas or an electronic zero with no electrode output, and the second point with 12% O2. Temperature control is important and the electrode is maintained at 37°C. Accuracy is compromised by protein deposits or perforation of the delicate plastic membrane, which must be inspected regularly. Oxygen continues to be consumed in blood samples, which should be taken anaerobically, heparinized and analysed promptly.

Galvanic or Fuel Cell: Galvanic cells convert energy from an oxidation-reduction chemical process into electrical energy. The potential generated is dependent on the oxygen concentration.

A gold mesh cathode catalyses the reduction of oxygen by reaction with water to hydroxyl ions, while lead is oxidized at the anode. Unlike the oxygen electrode, no battery is required. The reaction in the fuel cell generates a potential gradient. The chemical reaction uses up the components of the cell, so that its life depends on the concentration of oxygen to which it is exposed and on the duration of exposure: in practice, 6–12 months. Fuel cells are widely used in reliable and portable oxygen analysers which incorporate a digital readout and audible alarms. These are cheap and require little maintenance. Inaccurate responses to calibration with oxygen and air suggest that the fuel cell is exhausted and should be replaced.

Transcutaneous Electrodes: Transcutaneous electrodes are non-invasive and used extensively for monitoring neonatal blood gas tensions. The electrodes are based on principles similar to those used in blood gas analysers but also incorporate a heating element. The electrode is attached to the skin to form an airtight seal using a contact liquid and the area is heated to 43°C. At this temperature, the blood flow to the skin increases and the capillary oxygen diffuses through the skin, allowing measurement of the diffused gases by the attached electrode. The values obtained from the transcutaneous electrode are lower than those from a simultaneous arterial specimen. Many factors affect the transcutaneous measurement of oxygen tension, including the skin site and thickness. Most importantly, the electrode depends on local capillary blood flow and under-reads in the presence of hypotension and microcirculatory perfusion failure. Problems occur with surgical diathermy; the heating current circuit provides a return path for the cutting current which may cause the transcutaneous electrode to overheat.

Other methods of measuring oxygen tension in blood include mass spectrometry and optodes, which employ the quenching of fluorescence from illuminated dye.

Oxygen Content

The total amount of oxygen in blood may be measured directly, but this is technically demanding and rarely used. The van Slyke technique uses a chemical and volumetric or manometric analysis of oxygen content. Oxygen is driven from a small sample by denaturing the haemoglobin with acid. The volume of gas at atmospheric pressure, or the pressure at constant volume, is recorded before and after the chemical absorption of oxygen. The change is related directly to the oxygen content of the fixed volume of blood. Alternative detectors have been used to measure oxygen displaced from haemoglobin, e.g. a galvanic cell.

These time-consuming and operator-dependent laboratory techniques have been replaced by calculation of oxygen content from measurements of the oxygen saturation of haemoglobin, haemoglobin concentration and the tension of oxygen in blood:

image

Accurate estimates require that the oxygen saturation of haemoglobin is measured directly, and not calculated from oxygen tension and an arbitrary but unmeasured oxyhaemoglobin dissociation curve.

Oximetry: Measurement of Oxygen Saturation

In Vitro Oximetry: Oximetry relies on the differing absorption of light at different wavelengths by the various states of haemoglobin. The absorption of radiation passing through a sample is measured. The degree of absorption of light, defined by the ratio of incident to emergent light intensities on a logarithmic scale, is proportional to the concentration of the molecules absorbing light (Beer’s law) and the thickness of the absorbing layer (Lambert’s law).

Oxyhaemoglobin and deoxyhaemoglobin differ at both the red and infrared portions of the absorption spectrum (Fig. 16.19). The differential absorption of two wavelengths of red and infrared light permits the calculation of the ratio of the concentrations of oxygenated and reduced haemoglobins. Additional wavelengths are added in co-oximeters for the calculation of the proportions of other species of haemoglobin, such as carboxyhaemoglobin and methaemoglobin, and the absolute absorbance is used to estimate total haemoglobin concentration from the sum of the various haemoglobins. This is important in measurements of oxyhaemoglobin for use in the calculation of oxygen content.

Commercial co-oximeters draw a small blood sample which is haemolysed before entering a cuvette. Light is filtered to produce monochromatic beams, shone through the cuvette and detected by a photocell. The absorption by the sample is the difference in the intensity of incident and transmitted light and both must be measured. Spectrophotometers apply a double-beam technique which improves the accuracy and precision. Light from the monochromator is split into two beams, which pass through the test sample or a reference sample. Photocells generate two signals corresponding to the sample and the reference light intensities. Electronic processing compares the two signals and generates an output proportional to the difference. This greatly improves the signal-to-noise ratio because any variation which affects both the sample and reference beams equally is ignored and the difference remains constant.

The saturation of mixed venous blood may be measured using an oximeter incorporated into a pulmonary artery catheter. Fibreoptic cables transmit incident light of at least two wavelengths, and carry reflected light from red blood cells back to a detector.

The same spectrophotometric principles used by co-oximeters in vitro on haemolysed blood samples have been applied to patients in vivo.

Pulse Oximetry: Light transmitted through tissues is absorbed not only by arterial blood but also by other tissue pigments and venous blood. However, the variation in light absorption with each pulse beat results almost entirely from pulsatile arterial blood flow. Two light-emitting diodes – red (660 nm) and infrared (940 nm) – shine light through a finger or earlobe and a photocell detects the transmitted light. The output of the sensor is processed to display a pulse waveform and the arterial oxygen saturation.

The pulse oximeter progresses through the following steps.

1. The sensor first measures the ambient light and subtracts this value from all other measurements. This implies that sudden changes in ambient light levels, e.g. after drapes are moved, may cause transient errors.

2. An LED is turned on and off rapidly. The absorption of transmitted light is then measured and the variations with time are recorded. The result is a waveform with a trough as blood flows into the finger (more absorption) during systole and a peak as blood flows into the veins in diastole. The monitor requires around eight heartbeats to make a calculation and then assumes the frequency of this waveform is the heart rate. Frequent ectopic beats or atrial fibrillation may lead to a delay in calculation or unreliable results.

3. The monitor then analyses the measurements and splits the absorption into two components. The fixed or unchanging absorption is assumed to result from tissues such as skin, muscle and bone (Fig. 16.20). The varying absorption is then assumed to be caused by arterial blood moving into the tissue. In situations such as hypotension, hypovolaemia or hypothermia, pulsation may be reduced to a point at which the monitor is not able to make any calculations and it fails to read. When a patient is on cardiac bypass the tissues are perfused, but if the flow is non-pulsatile, pulse oximeters cannot provide a reading.

4. Steps 1–3 are repeated sequentially using light of at least two different wavelengths at around 120 Hz. When the absorptions of each different wavelength are known, the proportion of oxygenated and deoxygenated haemoglobin may be calculated. The measurements are processed and a new value displayed around every 8 s. Whilst improving reliability, this averaging introduces delay. Another source of delay is circulatory, dependent on the distribution of blood from the lungs to the tissues. The response time can exceed 1 min in normal subjects, and is exaggerated by low cardiac output or vasoconstriction.

Pulse oximeter calculations are based on the assumption that the blood contains only normal haemoglobin and that no abnormal light-absorbing substances (dyes) are present. For example, if the patient has breathed carbon monoxide, the monoxycarboxyhaemoglobin (as it has a similar absorption spectrum) is measured as oxyhaemoglobin. The result is that the pulse oximeter reading in patients suffering from carbon monoxide poisoning is usually close to 100%, even though they have severe hypoxaemia. The use of intravascular dyes as markers, or even nail varnish, may produce unpredictable results. Lastly, if the probe slips partially off the patient, some light passes directly from the LEDs to the light sensor and also through the sensor. This also causes unreliable readings.

Pulse oximeters provide rapid, non-invasive measurement of pulse rate and an estimate of oxygen saturation. The pulse oximeter has become one of the most widely used monitors and is particularly useful in situations in which it is difficult to identify cyanosis, e.g. if light levels are low, in pigmented patients and in areas where access is difficult such as CT/MRI scanners. Pulse oximeters are also used to measure the oxygen saturation in patients with intermittent respiratory problems, e.g. postoperative patients and those with sleep apnoea. Advances in technology have resulted in several small battery-powered devices becoming available for out-of-hospital use. Calibration points between 80% and 100% are derived from volunteer studies, and accuracy of pulse oximeters is around  ±  2% above an oxygen saturation of 70%. Accuracy below 70% is not known precisely, because it is not ethical to conduct trials at these levels. It is important to note that, especially when oxygen therapy is used, normal oxygen saturation does not equate to normal ventilation. For example, in opioid overdose, hypoventilation may lead to potentially fatal hypercapnia without any decrease in oxygen saturation if the patient is breathing a high concentration of oxygen. Complications are rare. The major drawbacks and source of error of pulse oximetry are summarized in Table 16.8.

TABLE 16.8

Disadvantages of Pulse Oximetry

Damage to skin caused by pressure from probe

Failure to detect hypoxaemia in carbon monoxide poisoning

Failure to detect hypoventilation

Slow response times: instrument and circulatory delay

Signal quality adversely affected by hypoperfusion

Inter-instrument variability

THE NERVOUS SYSTEM

The best monitor of cerebral function is the patient, who is able to report symptoms such as numbness, loss of function or pain. During general anaesthesia, this monitoring function is lost and it is possible for patients to develop major neurological deficits during anaesthesia which become evident only during the recovery phase. For example, patients who have suffered head trauma may develop undiagnosed cerebral oedema under anaesthesia if appropriate monitoring, such as intracranial pressure measurement, is not used. During procedures which carry a high risk of ischaemia or seizure activity, such as carotid endarterectomy, this problem may be avoided by performing the procedure under local anaesthesia; however, this may not always be acceptable to patients or possible because of unrelated medical problems. In general, therefore, patients at risk of cerebral malfunction should be given general anaesthesia only if there is no alternative.

When general anaesthesia is used, estimating the ‘depth of anaesthesia’ has traditionally been a function of the anaesthetist using clinical signs such as heart rate, blood pressure, sweating and pupillary dilatation. Unfortunately, especially in the presence of autonomic neuropathy, such clinical signs are not reliable enough to avoid the risk of awareness during surgery.

It should be noted that there is no widely accepted definition of what ‘anaesthesia’ represents. Although it may be characterized in terms of lack of perception, lack of responsiveness and inability to recall, these are all poorly defined concepts. Therefore, no monitor can measure ‘depth of anaesthesia’ in the same way as is possible, for example, with arterial pressure. Further, what is required is not a monitor which can detect awareness, but a monitor which can predict the onset of awareness and allow the anaesthetist to act before awareness occurs. Another reason to measure depth of anaesthesia is the ability to ensure that each patient receives only the minimum amount of anaesthetic required to maintain unconsciousness. This may have the potential to greatly reduce the side-effects of anaesthesia and improve recovery time.

Depth of Anaesthesia

The electroencephalogram and evoked potentials

The electroencephalogram (EEG) is a small, complex signal which is recorded usually from at least four electrodes fixed to specially prepared sites around the head. It has an amplitude of 50–200 μV and a frequency content which is classified conventionally into four categories:

The spiking, transient depolarization, then repolarization, of action potentials in neurones in the brain is sufficiently asynchronous and transient to be unrecordable from the scalp or surface of the brain. It is believed that the EEG is generated by the summation of synchronous postsynaptic potentials on the dendrites of sheets of large and symmetrically arranged pyramidal cells in cortical layers III and IV. Recording of these microvolt signals with acceptable levels of artefact and interference is difficult. Visual analysis of EEG recordings is subjective and requires experience.

The complexity of the raw EEG signals makes its use in the operating theatre impractical. Cerebral function monitors using fewer electrodes have been developed which average the EEG signal, but have been shown to be unreliable in monitoring depth of anaesthesia. However, the Cerebral Function Analysing Monitor (CFAM) has found a role in neurosurgery, and in other specialities in which monitoring cerebral function is crucial. For example, sedated and paralysed patients are unable to display classical tonic-clonic activity during generalized seizures. CFAM enables clinicians to detect seizure activity in neurosurgical intensive care patients and titrate sedation against burst-suppression of the EEG. Two symmetrical pairs of EEG electrodes are placed on either side of the patient’s head. The monitor then analyses the EEG amplitude and the frequency of the waveforms into delta, theta, alpha and beta waves as well as displaying wave suppression information.

Two further methods make the EEG signal easier to interpret, based on spectral analysis. The principle is that any complex wave may be broken down into a series of sine waves. Fast Fourier analysis produces an output in terms of the proportion of each frequency which contributes to the total signal. Bispectral analysis uses relationships between the phase and power of different frequencies within the original signal.

Initial attempts using spectral analysis showed that general anaesthesia produces a reduction in the mean frequency and spectral edge (frequency below which 95% of activity occurs). Ultimately, increasing doses of anaesthetic drugs produce a suppressed or isoelectric EEG. Unfortunately, these changes have not proved to be reliable as a measure of depth of anaesthesia.

The bispectral index (BIS monitor) is a number from 100 (awake) to 0 (deeply anaesthetized) produced by a commercially available monitor. Although the exact methods used have not been published, it identifies patterns within 30-s periods of EEG recording by first excluding episodes likely to be produced by diathermy or muscle activity, and periods of burst suppression. The remaining activity is subjected to bispectral analysis with the final result adjusted to take account of the proportion of suppressed EEG. It appears to provide reliable measurements in clinical use. However, it is not clear how factors such as different anaesthetic agents, hypoxaemia or epileptic activity affect the results.

Auditory Evoked Potentials

Auditory evoked potential monitors measure the slowing of auditory information processing produced by anaesthetic agents. The patient wears headphones which repeatedly play soft ‘clicks’. Each click produces neuronal activity in the auditory cortex. While the signal from a single click is masked by other brain activity, the signal from repeated clicks can be averaged to isolate the auditory signal. The activity in the auditory cortex may therefore be displayed as the auditory evoked potential, which has a characteristic waveform (Fig. 16.21). The time delay for some waves can then be measured and the result converted into a value typically from zero to 100, reflecting depth of anaesthesia. The equipment cannot be used in patients who have impaired hearing and may require several seconds to generate enough data to process. Monitors using this principle are commercially available, but have yet to be used widely. Equipment using either visual or somatosensory stimulation has also been developed.

Intracranial Pressure

Intracranial pressure (ICP) is frequently measured in neurointensive care units, and forms the basis of neuroprotective treatment strategies in brain injuries. The principle is similar to measurement of vascular pressure: an invasive device is connected via a rigid fluid-filled column to an electromechanical transducer which enables display of the waveform on a monitor. Classically, this is achieved using a catheter inserted into the lateral ventricle – the ‘gold standard’ for ICP measurement. Alternatively, the subdural space can be used, although this provides a less accurate monitor of ICP. More modern systems are less invasive and use either a catheter-tip miniature transducer or a microchip sensor which sits within the parenchyma of the brain. These systems do not require a fluid-filled column or an external transducer, so avoid errors of positioning. However, they cannot be re-calibrated once inserted and do not necessarily reflect global ICP.

Brain Oxygenation

The crucial need to ensure brain oxygenation and the inability to monitor consciousness during anaesthesia imply that an electronic monitor would be very valuable, especially during neurosurgery and in head-injured patients. Unfortunately, none of the currently available techniques has proved suitable for general use in the theatre environment. Their role is essentially limited to the neurosurgical Intensive Care Unit.

Near-infrared spectroscopy (NIRS) uses a pulse oximeter probe attached to the patient’s head to detect light reflected from the brain. While the technique may be used in small children, whose skulls are thin, doubts remain in adults as to whether this technique reflects brain oxygenation or merely scalp blood flow.

Transcranial Doppler involves a trained operator using a Doppler probe to measure the speed of red cells in cerebral arteries. It may detect vasospasm after subarachnoid haemorrhage, but has little use in routine anaesthesia apart from during carotid endarterectomy when it may be used to demonstrate adequacy of blood flow in the circle of Willis during contralateral carotid artery clamping.

Jugular venous oxygen saturation monitoring involves passing a catheter retrogradely up the internal jugular vein into the jugular venous sinus. A saturation of less than 55% indicates increased oxygen extraction and therefore relative ischaemia. It is a useful technique in the Intensive Care Unit but only measures global perfusion and fails to detect small areas of ischaemia. It also does not inform the clinician as to whether a low saturation is due to inadequate oxygen supply or increased cerebral demand.

Cerebral oxygenation can also be measured directly through the use of specialized intracerebral sensors. The Licox® PMO is a combined brain tissue oxygen and brain temperature monitor which detects regional changes in oxygenation. It is in essence a Clark polarographic electrode with a thermistor and sits within the brain parenchyma, ideally within the damaged but salvageable region of tissue (the penumbra). Assuming the monitor is sited in the correct area, it is able to measure regional oxygen tension accurately, although with a slight tendency to under-read.

More recently, microdialysis catheters have been introduced. These comprise a catheter with a surface dialysis membrane, and a perfusion system which slowly circulates dialysis fluid within the catheter. The returning fluid is analysed, either remotely or continuously, for cerebral metabolites such as lactate, pyruvate and other indicators of cerebral stress such as glutamate and glycerol. The lactate/pyruvate ratio is useful as a marker of cerebral ischaemia. As with intraparenchymal ICP monitors and the Licox system, the measurements are necessarily regional and dependent on accurate positioning of the catheter.

TEMPERATURE

The body is considered to have an inner core temperature and an outer peripheral temperature. In reality, the temperature decreases with distance down limbs and proximity to the skin. The difference between core and peripheral temperatures is related strongly to the cardiac output and degree of vasoconstriction. In practice, a core-periphery difference of > 2°C indicates a low cardiac output state and has been used as a marker of hypovolaemia.

Commercial thermometers make use of several temperature-dependent phenomena.

Direct Reading Non-Electrical Thermometers

Liquid Expansion Thermometers

Liquid expansion thermometers are simple, reliable instruments. A glass bulb is filled with a liquid (generally alcohol or mercury) and connected to an evacuated, closed capillary tube. The temperature is recorded by the position of the meniscus in the capillary tube against a calibrated scale. If the cross-sectional area of the capillary tube is constant, movement of the meniscus with changing temperature is linear.

Several simple and elegant design features improve usability. A large bulb and very narrow capillary increase the sensitivity. Visibility of the narrow capillary is improved by shaping the thermometer so that the glass forms a lens and by incorporating a strip of white glass behind the capillary. A constriction in the capillary tube permits the mercury to expand but hinders its return to the bulb so that the reading is preserved until the mercury is shaken down. However, glass thermometers are fragile. A large thermal capacity results in a slow response. The instrument is handheld, awkward to read and reset, and cannot be used for remote measurement or recording. Traditional mercury-in-glass thermometers are no longer used because of the risk of breakage and mercury contamination.

Remote Reading Instruments

Temperature-dependent electrical properties may be incorporated into thermometers suitable for automation (e.g. targeted temperature management systems for therapeutic cooling of patients), and are also commonly used for monitoring peripheral and core temperature.

Thermocouple Thermometers

If two dissimilar metals are joined to create an electrical circuit and the junctions are at different temperatures, current flows from one metal to the other. The potential difference which is generated is a function of the temperature difference between the two junctions. All junctions made from the same metals have identical properties. The reference junction must be kept at a constant temperature, or incorporate temperature compensation into the measurement. Common combinations of metals include copper-constantan or platinum-rhodium. The output is small (about 40 mV per °C temperature difference between the junctions), but this is sufficient to be sensed by a galvanometer.

Temperature probes using these properties can continuously measure peripheral and core temperatures. Peripheral temperature is measured by attaching a probe to either a finger or toe. Core temperature probes may be inserted into the nasopharynx, oesophagus or rectum, and although complications are rare, it is possible to cause mucosal trauma, bleeding and even penetration of the pharynx or rectum. When used for prolonged periods, mucosal damage is possible as a result of pressure-related ischaemia. Core temperature probes may also form part of an intravascular catheter, for example within a PAFC. Many monitors accept two probes and automatically display the temperature difference. Care must be taken to avoid the peripheral probe being exposed to warmth from heating blankets.

Equipment Temperature

The performance of some anaesthetic equipment is temperature dependent (e.g. vaporizers) and alternative thermometric principles may be used.

BLOOD LOSS AND TRANSFUSION

During surgery, blood loss is common but it may be difficult to measure accurately because blood may soak into surgical drapes and swabs, find its way into suction bottles or fall on to the floor. Although it is usually impossible to measure the blood loss accurately, an attempt should be made to estimate blood loss in all cases. This is particularly important in paediatric cases, where, for example, in a 2-kg child, a major haemorrhage equates to a blood loss of more than 20 mL.

Blood Clotting

Clotting efficacy is assessed by measuring the platelet count, prothrombin time (intrinsic system), activated partial thromboplastin time (extrinsic system) and fibrinogen concentration. Although these are traditional laboratory-based tests, there are increasing numbers of methods of analysing blood clotting in the operating theatre. These require the user to add a sample of blood to a reagent in a chamber, which is then inserted into a measuring device. The Sonoclot™ device uses a probe which is moved up and down in the sample. As the clot forms, the increasing resistance is measured and plotted as a graph against time. In thromboelastography, a pin is lowered into the sample and rotated intermittently. As a clot forms, there is increasing resistance to rotation.

Laboratory-based tests of clotting should be used whenever there is a risk of a clotting defect and after a significant blood transfusion. The main advantage of these tests is that they can be interpreted by haematologists, who provide advice on suitable treatments. Bedside monitors are used increasingly to measure clotting in procedures such as coronary artery surgery, where anticoagulants are usually used and significant blood loss is common.

Near-Patient Testing

During long operations, or when there is a need for infusion of large volumes of fluid, it may be necessary to measure physiological variables such as pH, haemoglobin and electrolyte concentrations and carboxyhaemoglobin. These have traditionally been performed on samples sent to laboratories and analysed by trained staff. The machines used are often complex and subject to regular servicing and quality control. More recently, it has become possible to use a variety of devices for ‘bedside’ analysis.

These devices usually comprise a small disposable cartridge which contains a combination of reagents and often some electronic circuitry. A sample is placed in the cartridge, which is then inserted into a larger analyser. Serum glucose concentration may be measured by a reagent strip which is compared with a colour chart or by insertion into a reader. These machines provide rapid results at the bedside and avoid the delays of transporting samples to the laboratory. However, the accuracy of the results produced by such devices is usually dependent on the skill of the operator and they may be affected by poor storage of the reagent or monitor. They also usually lack the organized quality control checks used in laboratories. Results are therefore less reliable and should be confirmed by laboratory analysis where possible.

MONITORING STANDARDS

The presence of an appropriately trained and experienced anaesthetist is the most essential patient monitor. However, human error is inevitable and there is substantial evidence that many incidents are attributable, at least in part, to error by anaesthetists.

Appropriate monitoring will not prevent all adverse incidents in the perioperative period. However, there is substantial evidence that it reduces the amount of harm to patients, not only by detecting problems as they occur but also by alerting the anaesthetist that an error has occurred. In one study, the introduction of modern standards of monitoring halved the number of cardiac arrests, principally because of the reduction in arrests caused by preventable respiratory causes. In the Australian Incident Monitoring Study, 52% of incidents were detected first by a monitor and in more than half of these cases, it was the pulse oximeter or capnograph which detected the problem.

There has never been a study comparing outcomes from anaesthesia with and without monitoring, and now that there are mandatory standards of monitoring in most countries, it will be impossible to prove that monitors make a difference.

The current recommendations from the Association of Anaesthetists of Great Britain and Ireland are shown in Table 16.9. However, it is not sufficient that the monitors are available. Departmental heads are responsible for ensuring not only that equipment is available but also that it works correctly, that it is maintained to appropriate standards, that staff are trained in its use and that it is used appropriately.

TABLE 16.9

Summary of Recommendations for Standards of Monitoring During Anaesthesia and Recovery

The anaesthetist must be present and care for the patient throughout the conduct of an anaesthetic.*

Monitoring devices must be attached before induction of anaesthesia and their use continued until the patient has recovered from the effects of anaesthesia.

The same standards of monitoring apply when the anaesthetist is responsible for a local/regional anaesthetic or sedative technique for an operative procedure.

A summary of information provided by monitoring devices should be recorded on the anaesthetic record. Electronic record keeping systems are now recommended.

The anaesthetist must ensure that all equipment has been checked before use. Alarm limits for all equipment must be set appropriately before use. Audible alarms must be enabled during anaesthesia.

Table 16.10 indicates the monitoring devices which are essential and those which must be immediately available during anaesthesia. If it is necessary to continue anaesthesia without a device categorized as ‘essential’, the anaesthetist must clearly note the reasons for this in the anaesthetic record.

Additional monitoring may be necessary as deemed appropriate by the anaesthetist.

A brief interruption of monitoring is only acceptable if the recovery area is immediately adjacent to the operating theatre. Otherwise monitoring should be continued during transfer to the same degree as any other intra- or inter-hospital transfer.

Provision, maintenance, calibration and renewal of equipment is an institutional responsibility.

*In hospitals employing Anaesthetic Practitioners (APs), this responsibility may be delegated to an AP supervised by a consultant anaesthetist in accordance with guidelines published by the Royal College of Anaesthetists (www.rcoa.ac.uk).

Reproduced with permission from the Association of Anaesthetists of Great Britain and Ireland 2007.

All anaesthetists must ensure that they are familiar with the equipment used in their hospital and that all equipment has been checked before use. The need for training and practice cannot be overemphasized because the increasing complexity of modern monitoring devices implies that they can behave in unexpected ways at inopportune moments.

ALARMS

All electronic monitors now include alarms which sound (and illuminate) when a variable moves outside a preset range. These limits are usually set by the manufacturer as part of the monitoring system’s basic functions and may be changed by the user (although the default values are often protected by a password). Failure to reset the alarm limits to appropriate values before starting anaesthesia is a common cause of false alarm signals. This often results in the anaesthetist cancelling a series of false alarms, with the risk that an alarm for a real problem is also cancelled without any action being taken.

Alarms alert the anaesthetist to a developing physiological change and allow it to be corrected. However, alarms may not respond to a serious problem. For example, marked hypotension in an elderly hypertensive patient may still fall within the range of a ‘normal’ arterial pressure as set for the monitor. Alarms must therefore be set to appropriate levels before induction so that they are triggered only by real abnormalities. This is particularly true during procedures involving children, when values for respiratory rate, for example, are not within the ‘normal’ adult range.

More commonly, alarms are triggered by artefacts: for example, the electrical interference produced by diathermy often triggers an alarm for arrhythmia from the ECG. Also common is the production of alarms by spurious problems such as an apparently low end-tidal carbon dioxide concentration during induction of anaesthesia, produced by the gas leak around a face mask. The frequency of such false alarms implies that many alarms are ignored, or may lead the anaesthetist to concentrate on the monitoring equipment and ignore the patient.

Further, because of the lack of standardization of alarm signals and the uniformity of alarms, in a genuine crisis, the cacophony of multiple alarms and series of flashing lights may cause staff to concentrate on a relatively unimportant complication, such as bradycardia, with the cause, such as a disconnection in the breathing system, going unnoticed.

Infusion Devices

Increasingly, anaesthetic drugs are being delivered by infusion devices. As with most equipment, individual devices are becoming more complex. For example, rather than using simple infusion pumps, anaesthesia may be delivered using devices containing computers, allowing the infusion to maintain a constant plasma concentration, e.g. the Diprifusor™ infusion pump.

These devices contain sophisticated alarm systems and usually need to be programmed with a variety of patient data to function effectively. Anaesthetists also need to be trained in their use and must understand how these devices work before attaching them to a patient. In particular, it must be emphasized that although many infusion devices display information about the amount of drug administered and/or plasma concentration, these are dependent on absence of leaks and on assumptions about the patient. A common problem is disconnection of the infusion line under surgical drapes, leading to underdosage. A further problem is that, in sick patients, factors such as the size of fluid compartments or rate of drug transport may differ greatly from those assumed by the pump, leading to over- or underdosage.

GENERAL GUIDELINES FOR MONITORING DURING ANAESTHESIA

Table 16.10 summarizes the required monitors for different components of anaesthesia. Monitors should be applied to the awake patient and readings taken to ensure that they are functioning correctly before induction of anaesthesia. If uncooperative patients make application of monitoring impossible before induction, the monitors should be applied as soon as possible after induction and the reason recorded on the anaesthetic chart.

For short procedures such as electroconvulsive therapy (ECT) or orthopaedic manipulations, the standards for induction of anaesthesia are appropriate. However, if the procedure is prolonged then the standards for maintenance of anaesthesia should be applied. A high standard of monitoring should be applied continuously until the patient has recovered fully from anaesthesia. If the recovery room is not immediately adjacent to the operating theatre, or if the patient’s condition is poor, equipment should be available so that the above standards are applied during transfer of the patient.

Additional Monitoring

The standards in Table 16.10 are the minimum acceptable levels and apply to healthy patients undergoing minor surgery. If the patient is unwell before surgery or major surgery is planned, additional monitoring should be applied. It is difficult to give strict guidelines on what conditions or surgery should prompt the use of each monitor. Suggestions are given in Table 16.11.

TABLE 16.11

Variables that it may be Appropriate to Monitor During Anaesthesia in some Patients in Addition to the Essential Monitoring for all Anaesthetized Patients

Indications Monitors
Operative duration  > 3 h Direct arterial pressure measurement
Blood loss  > 10% blood volume Central venous pressure
Operations on:
 Chest
 Central nervous system
 Cardiovascular system
Pulmonary capillary wedge pressure
Cardiac output
Transoesophageal echocardiography
Blood loss measurement
Clinically significant coexisting disease Urine output
Temperature:
 Patient
 Blood warmer, mattress
 Inspired gas
Blood gas analysis
Serum electrolyte concentrations
Haemoglobin concentration
Coagulation status

Monitoring During Transfer

It is essential that the standard of care and monitoring during transfer is as high as that applied in the operating theatre and that staff with appropriate training and experience accompany the patient.

During transfer, vibrations may make devices which rely on pressure change, such as non-invasive arterial pressure monitors, inaccurate or non-functional. Vibration may also cause connections to work loose and equipment may suffer physical damage. Movement may make an ECG trace useless for diagnosis of arrhythmias. Noise and poor lighting make the displays of many monitors difficult to read and make audible alarms inaudible. Adequate supplies must be taken for the entire journey, together with additional supplies to anticipate any unforeseen delays. This includes oxygen cylinders, batteries (the internal batteries of a monitor may have short lives) and anaesthetic drugs.

Before transfer, the patient should be in a stable physiological state. The patient should be moved on to the transport trolley, all of the transport monitors should be applied to the patient and their functions should be checked. All equipment should then be fastened securely and all catheters and leads taped into position. A check should be made that, from a single position, the anaesthetist is able to attend to the airway, see all the monitors and be able to administer drugs and fluids. The process is made much easier and safer with a dedicated transport trolley, so that all the equipment is fixed permanently in place.

ANAESTHETIC RECORD-KEEPING

It is the professional responsibility of every doctor to maintain accurate records of the treatment which patients receive, and their response to it. The anaesthetic record forms a part of a patient’s medical record. The principal purpose of the anaesthetic chart is to provide details of the anaesthetic technique used, of the physiological changes which were associated with the technique and with surgery, and of complications or problems which were encountered during the procedure. This information may assist other doctors if complications ensue, or if anaesthesia is required in the future. In addition, the anaesthetic record may be a valuable source of information if a subsequent complication results in litigation; the absence of a full record makes it difficult for an anaesthetist to demonstrate, for example, that postoperative renal failure was not attributable to untreated intraoperative hypotension.

The design of anaesthetic records varies widely, and is probably unimportant provided that it facilitates recording and display of all the relevant data. Suggestions for the reasonable content of an anaesthetic record data set are shown in Table 16.12.

TABLE 16.12

Suggested Data for Inclusion on Anaesthetic Records

PREOPERATIVE INFORMATION

Patient identity

Name/ID number/gender

Date of birth

Assessment and risk factors

Date of assessment

Assessor, where assessed

Weight (kg) [height (m) optional]

Base vital signs (BP, HR)

Medication, incl. contraceptive drugs

Allergies

Addiction (alcohol, tobacco, drugs)

Previous GAs, family history

Potential airway problems

Prostheses, teeth, crowns

Investigations

Cardiorespiratory fitness

Other problems

ASA grade ± comment

Urgency

Scheduled – listed on a routine list

Urgent – resuscitated, not on a routine list

Emergency – not fully resuscitated

PERIOPERATIVE INFORMATION

Checks

Nil by mouth

Consent

Premedication, type and effect

Regional anaesthesia

Consent

Block performed

Entry site

Needle used, aid to location

Catheter: y/n

Patient position and attachments

Thrombosis prophylaxis

Temperature control

Limb position

Postoperative instructions

Drugs, fluids and doses
Analgesic techniques

Place and time

Place

Date, start and end times

Personnel

All anaesthetists named

Check performed, anaesthetic room, theatre
Operation planned/performed
Apparatus
Check performed, anaesthetic room, theatre

Vital signs recording/charting

Monitors used and vital signs recorded not less frequently

than every 5 min

Drugs and fluids

Dose, concentration, volume

Cannulation

Injection site(s), time & route

Warmer used

Blood loss, urine output

Airway and breathing system

Route, system used

Ventilation: type & mode

Airway type, size, cuff, shape

Special procedures, humidifier, filter

Throat pack

Difficulty

Special airway instructions, incl. oxygen

Monitoring

Untoward events

Abnormalities

Critical incidents

Preoperative, perioperative, postoperative

Context, cause, effect

Hazard flags

Warnings for future care

Based on recommendations of the Royal College of Anaesthetists and the Association of Anaesthetists of Great Britain and Ireland

In addition to the data described above, the record should include details of the techniques discussed with the patient, together with any risks or benefits outlined and the management plan agreed. If the patient has any specific requests or concerns, such as a desire to avoid blood transfusion, it is also best to record them in writing.

Automated Records

It has been estimated that up to 20% of the anaesthetist’s time is taken up with documentation. While the anaesthetic record is usually completed as the anaesthetic proceeds, there are times, such as during induction or a crisis, when it is not possible to complete the chart contemporaneously. This delay leads to inaccuracies. In addition, studies have shown that anaesthetists tend to record ‘normalized’ data, i.e. the record tends to minimize any physiological changes which occur.

To counteract these problems, most anaesthetic monitors and anaesthetic machines may be connected to automatic data recording systems. These log all the monitoring data and have the facility for the anaesthetist to add data such as drugs used and comments on events, such as the start of surgery. The data may be stored electronically for later study and printed out in a variety of formats. They have the potential to interact with other sources of information so that patient details, laboratory results, scans and outpatient letters can all be accessed. These systems have the potential to make audits and quality control much easier to perform. The AAGBI recommends that departments consider their procurement.

However, these systems are expensive and it must be recognized that all monitors have numerous sources of error; while most anaesthetists tend to ignore erroneous readings, they are recorded and printed by an automated system. Printouts should therefore be checked and errors marked before being included in the patient’s medical record.