Blood Gases and Related Tests

Published on 12/06/2015 by admin

Filed under Pulmolory and Respiratory

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 0 (0 votes)

This article have been viewed 5138 times

Chapter 6

Blood Gases and Related Tests

Blood gas analysis is the most basic test of lung function. Evaluation of the acid-base and oxygenation status of the body provides important information about the function of the lungs themselves. Anaerobic sampling of arterial blood is required, which is an invasive procedure that carries risks involved with blood-borne pathogens. Pulse oximetry and capnography measure gas exchange parameters and have the advantage of monitoring patients noninvasively. Understanding the limitations of noninvasive techniques allows them to be used to provide appropriate patient care. Calculating oxygen content and the shunt fraction uses blood gas measurements to assess gas exchange as it applies to oxygenation.

The chapter addresses how blood gas measurements are used in the pulmonary function laboratory. A complete description of blood gas electrodes and other measuring devices is included in Chapter 11. The use of pulse oximetry and capnography as adjuncts to traditional invasive measures is discussed. Two methods of calculating shunt fraction are detailed so that the most appropriate method may be used.

Oxygen tension

Po2 measures the partial pressure exerted by oxygen (O2) dissolved in the blood. Like Pco2, it is recorded in millimeters of mercury or in kilopascals. The normal range for arterial Po2 is 70–100 mm Hg (10.64–13.30 kPa) for healthy young adults breathing air at sea level. The normal range declines slightly in older adults. Mixed venous Po2 averages 40 mm Hg (5.32 kPa) in healthy patients. Barometric pressure (at altitudes significantly above sea level) also influences the expected arterial Po2 (see Evolve site http://evolve.elsevier.com/Mottram/Ruppel/ for Conversion and Correction Factors).

Techniques

pH

Blood pH is measured by exposing the specimen to a glass electrode (see Figure 11-28) or by measuring light absorbance with an optical pH indicator under anaerobic conditions. pH measurements are made at 37°C. The pH of arterial blood is related to the Paco2 by the

Henderson-Hasselbalch equation:

pH=pK + log [HCO3][CO2]

image

where:

pK=negative log of dissociation constant for carbonic acid (6.1)

HCO3image[]=molar concentration of serum bicarbonate

[CO2]=molar concentration of CO2

Paco2 (measured directly by a CO2 electrode or similar device) may be multiplied by 0.03 (the solubility coefficient for CO2) to express Paco2 in milliequivalents per liter (mEq/L). The equation then may be expressed as follows:

pH=pK+ log [HCO3][0.03(PaCO2)]

image

Carbon Dioxide Tension

Pco2 has been measured traditionally by exposing whole blood to a modified pH electrode contained in a jacket with a Teflon membrane at its tip (see Chapter 11). The jacket contains a bicarbonate buffer. As CO2 diffuses through the membrane, it combines with water to form carbonic acid (H2CO3). The H2CO3 dissociates into H+ and HCO3image, thereby changing the pH of the bicarbonate buffer. The change in pH is measured by the electrode and is proportional to the Pco2. Newer blood gas analyzers use a spectrophotometer to measure the absorbance of CO2 in the infrared portion of the spectrum. The blood must be anticoagulated and kept in an anaerobic state until analysis. Pco2 may also be estimated using a transcutaneous electrode. Measurement of end-tidal CO2 (Petco2) is sometimes used to track Pco2 (see later section, Capnography).

The pH and Pco2 are usually measured from the same sample, so bicarbonate can be easily calculated. Automated blood gas analyzers perform this calculation along with others to derive values such as total CO2 (dissolved CO2 plus HCO3image) and standard bicarbonate (i.e., HCO3image corrected to a Paco2 of 40 mm Hg). If the hemoglobin (Hb) is measured or estimated, the base excess (BE) can be calculated. The normal buffer base at a pH of 7.40 is approximately 48 mmol/L. BE is the difference between the actual buffering capacity of the blood and the expected value. When the buffering capacity is less than the expected value, the difference is a negative value. This is sometimes referred to as a base deficit rather than a negative base excess. The main buffers that affect the BE are HCO3image and Hb.

Oxygen Tension

The Po2 (arterial or mixed venous) has been traditionally measured by exposing whole blood, obtained anaerobically, to a platinum electrode covered with a thin polypropylene membrane. This type of electrode is called a polarographic electrode or Clark electrode. Oxygen molecules are reduced at the platinum cathode after diffusing through the membrane (see Chapter 11). Newer blood gas analyzers use an optical system that senses the ability of O2 to change the intensity and duration of phosphorescence in a phosphorescent dye. Po2 may also be measured using a transcutaneous electrode (see Chapter 11).

Blood gas values (pH, Pco2, Po2) are influenced by the patient’s temperature. Alteration of body temperature affects the partial pressure of dissolved CO2, which influences pH as described in the previous equations. Table 6-1 describes the expected blood gas changes when the patient’s temperature is not 37°C. Although blood gas measurements are made at 37°C, the value reported is sometimes corrected to the patient’s temperature.

Technical problems with blood gas electrodes and related measuring devices include contamination by protein or blood products. In electrode-based analyzers, depletion of buffers in the electrodes may reduce accuracy and cause unacceptable drift. Tears or ruptures of the membranes used to cover the electrodes are also common malfunctions. Some newer analyzers use spectrophotometric methods that can be compromised by mechanical or electronic failure of the sampling cuvette, or by inadequate mixing of the specimen.

Specimen Collection for Blood Gases

Arterial samples are usually obtained from either the radial or brachial arteries via puncture or catheter in adults. Arterial specimens may also be drawn from the femoral, dorsalis pedis, posterior tibial, or umbilical arteries. The radial artery of the nondominant hand is usually the preferred site. A warmed capillary sample (arterialized) can also be obtained to approximate an arterial specimen.

Before a radial artery puncture, adequacy of collateral circulation to the hand from the ulnar artery should be established, using the modified Allen’s test (Figure 6-1). The original Allen’s test was described by Dr. Edgar Allen, a vascular surgeon at the Mayo Clinic in the 1920s, to assist in the diagnosis of thromboangiitis obliterans (e.g., Raynaud’s disease). A positive test in his description denoted reduced circulation; however, using the modified technique, a “positive test” demonstrates adequate collateral circulation. In performing the modified Allen’s test, the technologist occludes both the radial and ulnar arteries by pressing down over the wrist. The patient is instructed to make a fist, then open the hand and relax the fingers. The palm of the hand is pale and bloodless because both arteries are occluded. The ulnar artery is released while the radial remains occluded. The hand should be reperfused rapidly (5–15 seconds) if the ulnar supply is adequate. If perfusion is inadequate, an alternative site should be used.

Arterial puncture should not be performed through any type of lesion. Similarly, puncture distal to a surgical shunt (e.g., shunts used for dialysis) should be avoided. Infection or evidence of peripheral vascular disease should prompt selection of an alternative site. Some patients may be using anticoagulant drugs such as heparin, warfarin (Coumadin), or streptokinase. High doses of these drugs or a history of prolonged clotting times may be relative contraindications to arterial puncture. Box 6-1 lists some of the potential hazards associated with arterial puncture.

6-2   How To…

Perform an Arterial Puncture from the Radial Artery

1. Tasks common to all procedures.

2. Prepare necessary equipment: self-filling heparinized syringe, needle, and personal protection equipment (e.g., gloves and eyewear).

3. Assure the pre test conditions of the order are met (e.g. FIO2, liter flow, ventilator settings, etc.). If FIO2 has been modified wait 20-30 minutes prior to harvesting the sample. In the outpatient setting the patient should be in a stable condition for a minimum of 5 minutes or longer once the above criteria have been met…

4. Assess the collection site for adequate pulse, and perform a modified Allen’s test.

5. Hyperextension of the wrist may facilitate exposure of the artery; palpate for a pulse.

6. Prepare the puncture site; clean with a laboratory approved agent.

7. Make sure to instruct the subject to “breathe normally” and not to breath hold during the procedure.

8. Hold the syringe with the bevel up at a 30- to 45-degree angle, and advance toward the palpated artery until blood flow enters the syringe.

9. After obtaining the desired amount, place a dry gauze sponge over the puncture site and quickly remove the needle, applying pressure immediately and directly over the puncture site.

10. Immediately expel any air bubbles and mix thoroughly by rotating or inverting the sample numerous times.

11. Hold direct pressure to the site (2-5 minutes is typically adequate if the subject is not anticoagulated) until hemostasis occurs.

12. Report results and note comments related to sample quality.

Success in obtaining an arterial specimen by radial puncture requires careful positioning of the patient’s wrist. The hand should be hyperextended with good support under the wrist (Figure 6-2, arterial puncture). A topical anesthetic may be useful for some patients, but is usually unnecessary. Always perform the modified Allen’s test to ensure adequate collateral circulation before puncturing the artery. The person drawing the sample should be in a comfortable position (i.e., sitting) to maximize control of the needle during insertion. It is recommended that the person drawing the sample use standard precautions and personal protective equipment, including gloves and eyewear. A vented syringe or similar device allows the blood to “pulse” into the syringe, ensuring that the needle is in the lumen of the artery.

Drawing a sample from a catheter–infusion system for an arterial or mixed venous sample from a pulmonary artery (Swan-Ganz) catheter has its own nuances. Contamination of the sample with flush solution is a common problem. Withdrawing a small volume of blood into a “waste” syringe ensures that the sample is not diluted by flush solution in the catheter. Care should be taken to limit the volume of blood removed in this process. Significant blood loss can occur with repetitive measurements. Another problem associated specifically with mixed venous specimen collection is displacement of the Swan-Ganz catheter tip. If the catheter is advanced too far, it may “wedge” into a pulmonary arteriole. Specimens drawn from this position often reflect arterialized pulmonary capillary blood. Similarly, if the catheter tip is withdrawn or “loops back,” it may be in the right ventricle or atrium rather than in the pulmonary artery. Specimens obtained from this location may not represent true mixed venous blood.

Venous samples from peripheral veins are not useful for assessing oxygenation. Venous blood reflects only the metabolism of the area drained by that particular vein. Venous samples may be used for measurement of pH or blood lactate during exercise.

Blood is usually collected in a syringe to which an anticoagulant (e.g., heparin) has been added, and sealed from the atmosphere immediately (Criteria for Acceptability 6-1). Blood gases are typically drawn using specially designed syringes containing a dry anticoagulant (i.e., a blood gas “kit”). Dry heparin is applied to the lumen of the needle and the interior of the syringe. A small heparin pellet is often placed in the syringe to provide additional anticoagulation. Care must be taken that heparin solution (sometimes used for flushing arterial catheters) does not contaminate the sample. Contamination by flush solution may affect Po2 and Pco2 values by dilution, but the buffering capacity of whole blood prevents large changes in pH.

After the syringe has been capped (see Chapter 12 for safe handling of blood specimens), the sample should be thoroughly mixed by rolling or gently shaking. Mixing helps prevent the sample from clotting, whether dry or liquid heparin is used. Mixing is also important for analyzers that use spectrophotometric methods for measuring blood gases. Lithium heparin or a similar preparation should be used for specimens that will also be used for electrolyte analysis.

Air contamination of arterial or mixed venous blood specimens can seriously alter blood gas values. Room air at sea level has a Po2 of approximately 150 mm Hg, and a Pco2 near 0. If air bubbles are present in a blood gas specimen, equilibration of gases between the sample and air occurs (Table 6-2). Contamination commonly occurs during sampling when air is left in the syringe after the sample is collected. Small bubbles may also be introduced if the needle does not connect tightly to the syringe. Other sources of air contamination include poorly fitting plungers and failure to properly cap the syringe. Use of a vented syringe or one in which the pulse pressure of the blood displaces the plunger can help prevent air bubble contamination.

Sample storage depends on the type of syringe used. If a glass syringe is used, the sample may be stored in ice-water slush if analysis is not done within a few minutes. Ice water reduces the metabolism of red and white blood cells in the sample. Specimens with O2 tensions in the normal physiologic range (50–150 mm Hg) show minimal changes over 1–2 hours for glass syringes if kept in ice water. Changes in specimens held at room temperature are related to cellular metabolism in the blood, particularly in white blood cells and platelets. Specimens with Po2 values above 150 mm Hg are most susceptible to alterations resulting from gas leakage or cell metabolism. When the Po2 is 150 mm Hg or more, Hb is almost completely bound with O2. In such cases, a small change in O2 content results in a large change in Po2.

If a plastic syringe is used, blood gas specimens should be analyzed within 30 minutes. Plastic syringes are not completely gas tight, so room air can contaminate the sample. The influx of O2 may be counterbalanced by the consumption of O2 if the sample is not iced. When a blood gas specimen is placed in an ice-water bath, the solubility of O2 increases, as does the affinity of hemoglobin for oxygen. This lowers the partial pressure of O2 in the sample and increases the gradient between the sample and the environment. This gradient exaggerates the leakage of O2 into the specimen (as occurs with plastic syringes). When the sample is introduced into the analyzer at 37°C, solubility and Hb affinity return to their normal values, the oxygen that leaked in is released, and the Po2 is falsely increased. Because of these phenomena, blood gas specimens in plastic syringes should not be placed in ice water. If specimens cannot be routinely analyzed within 30 minutes, glass syringes with ice-water storage may be preferable.

Capillary samples are useful in infants when arterial puncture is impractical. The area for collection (the heel is commonly chosen) should be heated by a warm compress and lanced. Blood is then allowed to fill the required volume of heparinized glass/plastic capillary tubes. Squeezing the tissue should be avoided because predominantly venous blood will be obtained. The capillary tubes should be carefully sealed to avoid air bubbles. Guidelines for quality control of blood gas analyzers and for the safe handling of blood specimens are included in Chapter 12.

Significance and Pathophysiology

See Interpretive Strategies 6-1.

Interpretive Strategies 6-1   Blood Gases

1. Was the blood gas specimen obtained acceptably? Free of air bubbles and clots? Analyzed promptly and/or iced appropriately?

2. Did the blood gas analyzer function properly? Was there a recent acceptable calibration of all electrodes or sensors? Was analyzer function validated by appropriate quality controls?

3. Is pH within the normal limits (7.35–7.45)? If so, go to Step 4. If below 7.35, acidosis is present; if above 7.45, alkalosis is present. Otherwise, look for compensatory changes or combined disorders.

4. Calculate the anion gap. Is it within the normal range?

5. Is Pco2 within normal limits (35–45 mm Hg)? If so, go to Step 6.

    If Pco2 >45 and pH <7.35, then respiratory acidosis is present.

    If Pco2 >45 and pH >7.35, then compensated respiratory acidosis is present.

    If Pco2 <35 and pH >7.45, then respiratory alkalosis is present.

    If Pco2 <35 and pH <7.45, then compensated respiratory alkalosis is present.

6. Is calculated HCO3image within the normal limits (22–27 mEq/L)? If so, the acid-base status is probably normal; go to Step 7.

    If HCO3 <22 and pH <7.35, then metabolic* acidosis is present.

    If HCO3 <22 and pH >7.35, then compensated metabolic* acidosis is present.

    If HCO3image >27 and pH >7.45, then metabolic* alkalosis is present.

    If HCO3image >27 and pH <7.45, then compensated metabolic* alkalosis is present.

7. Is Po2 within the normal limits (80–100 mm Hg) on room air? If so, the oxygenation status is probably normal; check O2Hb saturation via hemoximetry. Is Po2 appropriate for FIO2? Is A-a gradient increased? If Po2 <55, significant hypoxemia is present.

8. Are blood gas results consistent with the patient’s clinical history and status? Are additional tests indicated (hemoximetry, shunt study)?

(See Figure 6-3.)


*Metabolic = nonrespiratory.

pH

The pH of arterial blood in healthy adults averages 7.40 with a range of 7.35–7.45. Arterial pH below 7.35 constitutes acidemia. A pH above 7.45 constitutes alkalemia. Because of the logarithmic scale, a change of 0.3 pH units represents a twofold change in [H+] concentration. If the pH decreases from 7.40 to 7.10 with no change in Pco2, the concentration of hydrogen ions has doubled. Conversely, if the concentration of [H+] is halved, the pH increases from 7.40 to 7.70, assuming the Pco2 remains at 40 mm Hg. Changes of this magnitude represent marked abnormalities in the acid-base status of the blood and are almost always accompanied by clinical symptoms (e.g., cardiac arrhythmias) (Figure 6-3). If a metabolic acidemia is present, examination of the Anion Gap (AG) needs to be assessed. The anion gap is the difference between routinely measured cations (sodium) and anions (chloride and bicarbonate) in serum, plasma, or urine. The magnitude of the difference in serum is calculated to help identify the cause of a metabolic acidemia. The anion gap also reflects unmeasured cations (e.g., proteins, organic acids, phosphates) and anions (e.g., calcium, potassium, magnesium) (Table 6-3). The AG is expressed in units of mEg/L or mmol/L and is calculated by subtracting the serum concentrations of chloride and bicarbonate from the concentrations of sodium and potassium. However, in routine clinical practice, the potassium is frequently ignored, which leaves the following equation:

Table 6-3

Anion Gap

Metabolic Acidosis
Wide Gap Normal Gap
“MUDPILES” “HARDUP”
Methanol (Osm gap high) Hyperventilation
Uremia Acid infusion/carbonic anhydrase inhibitors/Addison’s disease
Diabetic/alcoholic ketoacidosis Renal tubular acidosis (urine gap positive)
Paraldehyde Diarrhea (urine gap normal; i.e., negative)
Isoniazid/Iron Ureterosigmoidostomy
Lactic acidosis Pancreatic fistula/drainage
Ethylene glycol (Osm gap high)
Salicylates/solvents

image

AG=[Na+][Cl-+HCO3]AG=140[105+24]AG=11mEq/L

image

The normal range is 3 − 11 mEq/L; however, the reference range should be provided by the lab performing the tests.

Acid-base disorders arising from lung disease are often related to Pco2 and its transport as carbonic acid. If the pH is outside of its normal range (i.e., acidemia or alkalemia) but Pco2 is within normal limits, the condition is termed nonrespiratory or metabolic (Table 6-4). The calculated HCO3image is a useful indicator of the relationship between pH and Pco2. In the presence of acidemia (pH less than 7.35) and normal CO2 (Pco2 35–45 mm Hg), HCO3image will be low and a metabolic acidosis is present. A Pco2 of less than 35 mm Hg in the presence of acidosis suggests that ventilatory compensation for acidemia is occurring. The acid-base status would be considered partially compensated nonrespiratory (i.e., metabolic) acidosis. Complete compensation occurs if pH returns to the normal range. This happens when ventilation reduces Pco2 in proportion to the HCO3image.

Table 6-4

Acid-Base Disorders

Status pH Pco2 HCO3image
Simple Disorders
Metabolic acidosis Low Normal Low
Metabolic alkalosis High Normal High
Respiratory acidosis Low High Normal
Respiratory alkalosis High Low Normal
Compensated Disorders
Compensated respiratory acidosis, or metabolic alkalosis Normal* High High
Compensated metabolic acidosis, or respiratory alkalosis Normal* Low Low
Combined Disorders
Metabolic/respiratory acidosis Low High Low
Metabolic/respiratory alkalosis High Low High

image

*Compensation cannot return values to within normal limits in severe acid-base disturbances. In addition, a normal pH may result in instances of respiratory and metabolic disturbances that occur together but are not compensatory.

In the presence of alkalemia (pH above 7.45) and normal Pco2 (35–45 mm Hg), calculated bicarbonate will be increased and metabolic alkalosis is present. If ventilatory compensation occurs, Pco2 will be slightly increased. However, decreased ventilation is required so that the CO2 can increase. Reduced minute ventilation may interfere with oxygenation. For this reason, Paco2 seldom increases above 50 mm Hg to compensate for nonrespiratory (i.e., metabolic) alkalosis. Compensation may be incomplete if the alkalosis is severe.

Abnormal Pco2 and HCO3image characterize combined respiratory and nonrespiratory acid-base disorders. In combined acidosis, Pco2 is elevated (more than 45 mm Hg) and HCO3imageis low (less than 22 mEq/L). In combined alkalosis, HCO3imageis elevated (more than 26 mEq/L) and Pco2 is low (less than 35 mm Hg). The pH is more severely deranged (i.e., much higher or lower) than if just one disorder were present.

Carbon Dioxide Tension

The arterial carbon dioxide tension (Paco2) in a healthy adult is approximately 40 mm Hg; it may range from 35–45 mm Hg. The Pco2 of venous or mixed venous blood is seldom used clinically. Body temperature affects Paco2, as described in Table 6-1.

Paco2 is inversely proportional to alveolar ventilation (imageA) (see Chapter 5). When imageA decreases, CO2 may not be removed by the lungs as fast as it is produced. This causes Paco2 to increase. The pH falls (i.e., [H+] increases) as CO2 is hydrated to form carbonic acid:

CO2+ H2OH2CO3H++ HCO3

image

Increasing levels of CO2 in the blood drive this reaction to the right. The patient has respiratory acidosis resulting from hypoventilation. Conversely, when alveolar ventilation removes CO2 more rapidly than it is produced, Paco2 decreases. The pH increases (i.e., [H+] falls) as the patient becomes alkalotic. This condition is called hyperventilation, or respiratory alkalosis.

If dead space increases, high-minute ventilation (imageE) may be required to adequately ventilate alveoli and keep Paco2 within normal limits. Respiratory dead space occurs because some lung units are ventilated but not perfused by pulmonary capillary blood. Pulmonary embolization is an example of dead space–producing condition. Emboli may block pulmonary arterioles, causing ventilation of the affected lung units to be “wasted.” To maintain normal Paco2, total ventilation must be increased to compensate for wasted ventilation. Paco2 may be normal, or decreased, even though significant pulmonary disease is present.

Patients who have disorders such as lobar pneumonia may increase their imageE to provide more alveolar ventilation of functional lung units. This mechanism compensates for lung units that do not participate in gas exchange. Hypoxemia is a common cause of hyperventilation (i.e., respiratory alkalosis). Hyperventilation may be seen in patients with asthma, emphysema, bronchitis, or foreign body obstruction. Anxiety, stress or central nervous system disorders may also cause hyperventilation. Hyperventilation may also be an intentional or unintentional result of mechanical ventilation.

Increased Paco2 (i.e., hypercapnia) commonly occurs in patients who have advanced obstructive or restrictive disease. These individuals are characterized by markedly abnormal ventilation-perfusion (image/image) patterns. They are unable to maintain adequate alveolar ventilation. Not all patients with advanced pulmonary disease retain CO2. Those who do become hypercapnic often have a low ventilatory response to CO2 (see Chapter 5). Their response to the increased work of breathing caused by obstruction or restriction is to allow CO2 to increase rather than increase ventilation. When respiratory acidosis results from increased Pco2, renal compensation usually occurs (see Table 6-4).

Increased Paco2 may also be seen in patients who hypoventilate as a result of central nervous system or neuromuscular disorders. Whether CO2 retention is the result of lung disease, central nervous system dysfunction, or neuromuscular disease, pH is usually maintained close to normal. The kidneys retain and produce bicarbonate (HCO3image) to match the increased Paco2. This response may completely compensate for a mildly elevated Paco2. However, it can seldom produce normal pH when the Paco2 is greater than 65 mm Hg. If the disorder causing the increased Paco2 is acute (e.g., foreign body aspiration), little or no renal compensation may be observed.

Some degree of hypoxemia is always present in patients who retain CO2 while breathing air. As alveolar CO2 increases, alveolar O2 decreases. If the cause of hypercapnia is either obstructive or restrictive lung disease, hypoxemia may be severe because of image/imageabnormalities. Because O2 therapy is common in these patients, changes in Pco2 while breathing supplementary O2 must be carefully monitored. Some patients with chronic hypoxemia have a decreased ventilatory response to CO2. O2 administered to these patients may reduce their hypoxic stimulus to ventilation. As a result, Paco2 may increase further. O2 therapy is usually titrated to maintain Paco2 values less than 60 mm Hg without hypercapnia and acidosis.

Oxygen Tension

The Pao2 of healthy young adults at sea level ranges from 85–100 mm Hg and decreases slightly with age. Breathing room air at sea level results in an inspired Po2 of approximately 150 mm Hg:

PIO2=FIO2(PB47) =0.21(76047) =0.21(713) =149.7

image

where:

FIO2= fractional concentration of inspired O2

PB= barometric pressure

47= partial pressure of water vapor at 37°C

The partial pressure of O2 in alveolar gas is usually close to 100 mm Hg and can be calculated using the alveolar air equation:

PAO2=(FIO2×[PB47])PaCO2(FIO2+1FIO2R)

image

where:

Paco2= arterial CO2 tension (presumed equal to alveolar CO2 tension)

R= respiratory exchange ratio (imageco2/imageo2)

Substituting 40 mm Hg for Paco2, and 0.8 for R:

PAO2=(0.21×[76047])40(0.21+10.210.8)

image

=(0.21×713)40(1.1975)

image

=149.747.9

image

=101.8

image

In healthy lungs with good gas exchange, arterial oxygen tension can approach the value of the alveolar Po2. The difference between the alveolar and arterial oxygen tensions is described as the alveolar-arterial gradient, or A-a gradient. This gradient is usually less than 20 mm Hg in healthy individuals breathing air at sea level.

Hyperventilation may increase Pao2 as high as 120 mm Hg in a patient with normal lung function (see the previous alveolar air equation). Healthy subjects breathing 100% O2 may exhibit Pao2 values higher than 600 mm Hg. The alveolar Po2 (PAo2) for a particular inspired O2 fraction can be calculated, as previously described. Decreased Pao2 can result from hypoventilation, diffusion abnormalities, image/imageimbalances, shunt, and inadequate atmospheric O2 (high altitude) (see Table 6-4).

Table 6-1 lists changes that occur in Pao2 because of body temperatures above and below normal (37°C). Changes in Po2 and Pco2 reflect solubility of the gas. The partial pressure of each gas is a measure of its activity. Hypothermia (low body temperature) is accompanied by decreased partial pressure as more gas dissolves. Hyperthermia (elevated body temperature) causes elevated gas tensions as gas comes out of solution. All blood gas analyzers perform analyses at 37°C, and most allow temperature corrections to be made. Although blood gas tensions vary with temperature, the clinical significance of correcting measurements is unclear. Blood gas values should be reported at 37°C. Care must be taken to ensure that blood gas analyzers are maintained at 37°C. Measurements made at other temperatures can significantly alter results.

Po2 is the pressure of O2 dissolved in blood. The amount of Hb and whether it is capable of binding O2 has only a minimal effect on Po2. Hypoxemia (decreased O2 content of the blood) may occur even if Pao2 is normal or elevated by breathing O2. Hypoxemia commonly results from inadequate or abnormal Hb. Many automated blood gas analyzers calculate oxygen saturation (SaO2). Saturation may be calculated from the Pao2 and pH, assuming a normal oxygen-hemoglobin reaction occurs (Figure 6-4). Calculated SaO2 may differ significantly from true saturation measured by a spectrophotometer (see the section on oxygen saturation) because a number of factors can affect the relationship of the binding capacity of hemoglobin (see Figure 6-4)

An example of the discrepancy between calculated and measured saturation is the patient with elevated carboxyhemoglobin (COHb), resulting from smoking or smoke inhalation. The patient’s Pao2 may be normal while O2 saturation is markedly decreased. Calculating saturation from Po2, in this case, overestimates the O2 content of the blood. Measured SaO2 is preferred to a calculated value.

The severity of impaired oxygenation is indicated by the Pao2 at rest. Pao2 is a good index of the lungs’ ability to match pulmonary capillary blood flow with adequate ventilation. If ventilation matches perfusion, pulmonary capillary blood leaves the lungs with a Po2 close to that of the alveoli. If ventilation is adequate, pulmonary capillary blood is almost completely saturated. If either of these conditions is not met (i.e., poor ventilation or image/imagemismatching), pulmonary capillary blood has reduced O2 content. Pao2 is reduced in proportion to the number of lung units contributing blood with low O2 content. Lung units with good image/imagecannot compensate for their poorly functioning counterparts because pulmonary capillary blood leaving them is already almost fully oxygenated. O2 binding to Hb is almost complete when the Pao2 is greater than 60 mm Hg (approximately 90% saturation). As Pao2 decreases from 60 to 40 mm Hg, saturation decreases from 90% to 75%, with increasing symptoms of hypoxia (e.g., mental confusion, shortness of breath).

Delivery of O2 to the tissues, however, depends on Hb concentration and cardiac output as well as adequate gas transfer in the lungs. Arterial oxygen content (Cao2 in mL/dL) is defined as follows:

CaO2=(1.34×Hb×SaO2)+(PaO2×0.0031)

image

where:

1.34= O2 binding capacity of Hb, mL/g

Hb= hemoglobin concentration, g/dL

SaO2= arterial oxygen saturation as a fraction

Pao2= arterial oxygen tension, mm Hg

0.0031= solubility coefficient for O2 at 37°C

Because most O2 transported is bound to Hb, there must be an adequate supply (12–15 g/dL) of functional Hb. Adequate cardiac output (4–5 L/min) is necessary to deliver the oxygenated arterial blood to the tissues. Signs and symptoms of hypoxia may be present despite adequate Pao2 because of severe anemia and/or reduced cardiac function.

The mixed venous oxygen tension (Pv¯imageo2) in healthy patients at rest ranges from approximately 37–43, with an average of 40 mm Hg (see Figure 6-4). Mixed venous O2 content (Cv¯imageo2) averages 15 mL/dL. In healthy individuals, arterial oxygen content (Cao2) averages 20 mL/dL. The resulting content difference, or C(a-v¯image)o2, is thus 5 mL/dL (or 5 vol%). Although Pao2 varies with the inspired O2 fraction and matching of image/image, Pv¯imageo2 changes in response to alterations in cardiac output and O2 consumption. If cardiac output increases while oxygen consumption (imageo2) remains constant, the C(a-v¯image)o2 decreases. Conversely, if cardiac output decreases with no change in O2 consumption, C(a-v¯image)o2 increases. Increased cardiac output sometimes occurs in response to pulmonary shunting. This allows mixed venous oxygen content to increase, reducing the deleterious effect of the shunt. Critically ill patients often have low Pv¯imageo2 values and increased C(a-v¯image)o2 as a result of poor cardiovascular performance. Alterations in Pv¯imageo2 often occur even if Pao2 is normal. Pv¯imageo2 values less than 28 mm Hg in critically ill patients, accompanied by C(a-v¯image)o2 greater than 6 vol%, suggest marked cardiovascular decompensation.

Patients who have severe obstructive or restrictive diseases may have decreased Pao2 at rest, occasionally as low as 40 mm Hg. Mild pulmonary disease may show little decrease in Pao2 if hyperventilation is present. Pao2 may also be relatively normal if the disease process affects ventilation and perfusion similarly. In patients with emphysema, destruction of alveolar septa may eliminate pulmonary capillaries as well, resulting in poor ventilation and equally poor blood flow. These patients may have severe airway obstruction, markedly reduced Dlco, but only a small decrease in resting Pao2. Patients with chronic bronchitis or asthma, particularly during acute exacerbations, may have moderate or severe resting hypoxemia because of image/imageabnormalities.

During exercise in patients with obstructive disease, Pao2 often decreases commensurate with the extent of the disease. Pao2 during exercise is correlated with the patient’s Dlco and FEV1, but wide variability exists. Patients with markedly decreased Dlco may show low Pao2 values at rest that decrease during exercise. Studies have demonstrated sensitivity and specificity in the range of 75%–90% of exercise desaturation if the Dlco is 50%–60% percent of predicted. The degree of arterial desaturation cannot be predicted from static pulmonary function measurements.

Pao2 may be decreased for nonpulmonary reasons such as anatomic shunts (intracardiac) or hypoventilation because of neuromuscular disease. Tissue hypoxia can occur because of inadequate or nonfunctional Hb, or because of poor cardiac output. Pao2 should be correlated with spirometry (FEV1), Dlco, ventilation (imageE, tidal volume, dead space), and lung volumes (VC, RV, TLC) to distinguish pulmonary from nonpulmonary causes of inadequate oxygenation.

Hemoximetry

Description

Hemoximetry (i.e., blood oximetry, also known as CO-Oximeter) refers to the measurement of hemoglobin (Hb) and its derivatives by spectroscopy. Oxygen saturation is the ratio of oxygenated Hb (O2Hb) to either total available Hb or functional Hb. Functional Hb is that portion of the total Hb that is capable of binding oxygen. This ratio of content to capacity is normally expressed as a percentage but is sometimes recorded as a simple fraction. The values may differ significantly, depending on the method of calculation:

O2Hb(O2Hb+rHb+COHb+MetHb)

image

O2HbO2Hb+rHb

image

where:

O2Hb= oxyhemoglobin

rHb= reduced hemoglobin concentration

COHb= carboxyhemoglobin concentration

MetHb= methemoglobin concentration

The first equation is used to measure O2 saturation with a multiple-wavelength spectrophotometer (hemoximeter); the second equation is used to measure O2 saturation with a pulse oximeter.

Technique

O2 saturation of Hb (O2Hb, or SaO2 when referring to arterial saturation) may be measured in several ways. In the first method, O2 saturation is measured using a spectrophotometer (see Chapter 11). Whole blood is hemolyzed and the absorbances of the various components measured. The total Hb, O2Hb, COHb, and MetHb are usually reported. Hemoximetry is usually performed in conjunction with blood gas analysis, using a single specimen of blood.

An indirect method may be used to measure mixed venous oxygen saturation (Sv¯imageo2). Saturation may be measured by a reflective spectrophotometer in a pulmonary artery catheter (Swan-Ganz catheter). In this case, absorbances are measured without hemolyzing the blood. A special catheter that includes fiberoptic bundles is used to perform in vivo measurements.

Blood specimens for hemoximetry should be prepared, as described for arterial blood gas specimens. Guidelines for quality control of blood gas analysis are included in Chapter 12. Measurement of percent saturation allows the calculation of the O2 content of either arterial or mixed venous blood (Cao2 and Cv¯imageo2, respectively) (see section, Oxygen Tension).

Significance and Pathophysiology

See Interpretive Strategies 6-2. SaO2 for a healthy young adult with a Pao2 of 95 mm Hg is approximately 97%. The O2Hb dissociation curve is relatively flat when the Pao2 is above 60 mm Hg (i.e., SaO2 is 90% or more). Saturation changes only slightly when there is a marked change in Pao2 at partial pressures above 60 mm Hg (see Figure 6-4). Therefore, Pao2 is a more sensitive indicator of oxygenation in lungs that do not have gross abnormalities. At Pao2 values of approximately 150 mm Hg, Hb becomes completely saturated (SaO2 is 100%). At Pao2 values above 150 mm Hg, further increases in O2 content are caused by increased dissolved oxygen. Alterations in image/imagepatterns in the lungs can be monitored by allowing the patient to breathe 100% O2 and by measuring the changes in dissolved oxygen. In practice, this is accomplished by using the clinical shunt equation (see section, Shunt Calculation).

When Pao2 falls below 60 mm Hg, SaO2 decreases rapidly. Small changes in Pao2 result in large changes in saturation. As SaO2 falls below 90%, O2 content decreases rapidly. At saturations less than 85% (i.e., Pao2 <55 mm Hg), symptoms of hypoxemia increase and supplementary O2 may be indicated.

The ability of Hb to bind O2 may be measured by the P50. P50 specifies the partial pressure at which Hb is 50% saturated (see Figure 6-4). The P50 of normal adult Hb is approximately 27 mm Hg. P50 may be determined by equilibrating blood with several gases at low oxygen tensions. An O2Hb dissociation curve is then constructed to estimate the partial pressure at which Hb is 50% saturated. A second method of estimating P50 compares measured SaO2 (using a spectrophotometer) with the expected saturation. Calculated saturations presume a P50 of 26–27 mm Hg, but it may differ significantly, depending on the types of Hb and interfering substances present.

Healthy individuals have small amounts of Hb that cannot carry O2. COHb is present in blood from metabolism and from environmental exposure to carbon monoxide (CO) gas. Normal COHb, expressed as a percentage, ranges from 0.5%–2% of the total Hb. CO comes from smoking (cigarettes, cigars, and pipes), smoke inhalation, improperly vented furnaces, automobile emissions, and other sources of combustion. In smokers, levels may increase from 3%–15%, depending on recent smoking history. Smoke inhalation or CO poisoning from other sources also results in elevated COHb levels, sometimes as high as 50%. CO combines rapidly with Hb. Exposures of short duration can cause a high level of COHb if high concentrations of CO are present. Because O2Hb saturation decreases as COHb increases, COHb levels greater than 15% almost always result in hypoxemia. High levels of COHb are rapidly fatal because of the profound hypoxemia that occurs.

COHb absorbs light at wavelengths similar to O2Hb. When COHb is elevated, arterial blood appears bright red. Cyanosis, which appears when there is an increased concentration of reduced Hb, is absent. In spite of elevated levels of COHb, Pao2 may be close to normal limits. Blood gas analysis that includes calculated saturation will give erroneously high O2 saturations. For this reason, O2Hb and COHb should be measured by hemoximetry whenever possible.

COHb interferes with O2 transport in two ways: it binds competitively to Hb, and it shifts the O2Hb curve to the left (see Figure 6-4). Increased COHb causes reduced O2Hb with a decrease in O2 content. The left shift of the dissociation curve causes O2 to be bound more tightly to Hb. The combination of these two effects can seriously reduce O2 delivery to the tissues. COHb concentrations in blood begin to decrease once the source of CO has been removed. Removal of CO from the blood depends on the minute ventilation. Breathing air may require several hours to reduce even moderate levels to normal. Breathing 100% O2 speeds the washout of CO. High concentrations of O2 (often including hyperbaric therapy) are indicated whenever dangerously high levels of COHb are encountered.

Methemoglobin (MetHb) forms when iron atoms of the Hb molecule are oxidized from Fe++ to Fe+++. The normal MetHb level is less than 1.5% of the total Hb in adults. High levels of MetHb can result from ingestion of or exposure to strong oxidizing agents. Methemoglobinemia has also been linked to high doses of medications such as benzocaine, dapsone, nitroprusside and inhaled nitric oxide. Like COHb, MetHb reduces O2 carrying capacity of the blood by reducing the available Hb and shifting the O2Hb dissociation curve to the left (see Figure 6-4).

The saturation of mixed venous blood (Sv¯imageo2) in healthy patients averages 75% at a Pv¯imageo2 of 40 mm Hg. Healthy patients have a content difference, C(a-v¯image)o2, of 5 vol%. Arterial blood, in adults with normal levels of Hb, typically carries approximately 20 vol% O2, and mixed venous blood carries 15 vol% O2. Pulmonary diseases that cause arterial hypoxemia may reduce Sv¯imageo2 if oxygen uptake and cardiac output remain constant. Cardiac output often increases to combat arterial hypoxemia caused by intrapulmonary shunting. Increased cardiac output increases O2 delivery to the tissues. This results in a reduced extraction of O2 from the blood. Mixed venous blood then returns to the lungs with normal or even increased O2 saturation. When this blood is shunted, it has a higher O2 content, thereby reducing the shunt effect. With or without arterial hypoxemia, Sv¯imageo2 decreases if cardiac output is compromised.

Sv¯imageo2 is useful in assessing cardiac function in the critical care setting and during exercise, but does require the placement of a pulmonary artery catheter. Patients who have good cardiovascular reserves maintain a mixed venous saturation of 70%–75%. Patients whose Sv¯imageo2 values are in the 60%–70% range have a limited ability to deliver more O2 to the tissues. Sv¯imageo2 values less than 60% usually indicate cardiovascular decompensation and may be associated with tissue hypoxemia. The indwelling reflective spectrophotometer (see Chapter 11) allows continuous monitoring of this important parameter. Sv¯imageo2 also decreases during exercise. Despite increased cardiac output, O2 extraction by the exercising muscles reduces the content of blood returning to the lungs.

Estimation of SaO2 by most pulse oximeters (SpO2) is based on the absorption of light at two wavelengths. When only two wavelengths are analyzed, only two species of Hb can be detected. Absorption in the red and near-infrared portions of the visible spectrum allows measurement of the oxyhemoglobin and reduced Hb, providing an estimate of the oxygen saturation of available Hb (see section, Pulse Oximetry).

Pulse oximetry

Description

Sp O2 estimates SaO2 by analyzing absorption of light passing through a capillary bed, either by transmission or reflectance. Pulse oximetry is noninvasive. SpO2 is reported as percent saturation. These devices report heart rate, although practitioners should be aware this is pulse rate derived at the probe site and may not reflect actual heart rate. Some pulse oximeters are also capable of measuring Hb, COHb, and MetHb.

Technique

Pulse oximeters (see Chapter 11) measure the light absorption of a mixture of two forms of Hb: O2Hb and reduced Hb (rHb). The relative absorptions at 660 nm (red) and 940 nm of light (near infrared) can be used to calculate the combination of the two Hb forms. Absorption at two wavelengths provides an estimate of the saturation of available Hb (see the section on oxygen saturation). Most pulse oximeters use a stored calibration curve to estimate oxygen saturation.

Pulse oximetry may be used in any setting in which a noninvasive measure of oxygenation status is sufficient. This includes monitoring of O2 therapy and ventilator management. Pulse oximetry is commonly used during diagnostic procedures such as bronchoscopy, sleep studies, or stress testing. It is also used for monitoring during patient transport or rehabilitation. Pulse oximetry may be used for continuous monitoring with inclusion of appropriate alarms to detect desaturation. Many pulse oximeter systems use memory (RAM) to record SpO2 and heart rate for extended periods. This type of recorded monitoring allows pulse oximetry to be used for overnight studies of nocturnal desaturation. Alternatively, pulse oximetry can be used for discrete measurements or “spot checks.”

Most pulse oximeters use a sensor that attaches to the finger, toe (nail bed), or ear. The choice of attachment site should be dictated by the type of measurement being made. The ear site may be preferred in patients undergoing exercise testing or in whom arm movement precludes use of the finger site. Both finger and ear sites presume pulsatile blood flow. Most pulse oximeters adjust light output to compensate for tissue density or pigmentation. In some patients, impaired perfusion to one site determines which site is preferred. Low perfusion or poor vascularity can cause the oximeter to be unable to detect pulsatile blood flow. Rubbing or warming of the site often improves local blood flow and may be indicated to obtain reliable data. Some pulse oximeters use reflective sensors that detect light reflected from bone underlying the tissue bed. These sensors are usually placed on the patient’s forehead and may function when finger or ear sensors do not.

Some pulse oximeters display a representation of the pulse waveform derived from the absorption measurements (see Chapter 11). Such waveforms may be helpful in selecting an appropriate site or troubleshooting questionable SpO2 values. Most pulse oximeters report heart rate (HR), which is also detected from pulsatile blood flow at the sensor site. Comparison of oximeter HR with palpated pulse or with an electrocardiograph (ECG) signal can assist with selection of an appropriate site. Inability to obtain a valid HR reading or acceptable pulse waveform suggests that SpO2 values should be interpreted cautiously (Criteria for Acceptability 6-2).

A number of factors limit the validity of SpO2 measurements (Box 6-2). To validate pulse oximetry readings, direct measurement of arterial saturation is required. Simultaneous measurement of SpO2 and SaO2 can be used to confirm pulse oximeter readings. Pulse oximetry used during exercise testing has been shown to produce variable results. Hemoximetry may be necessary to validate pulse oximetry readings at peak exercise.

COHb absorbs light at wavelengths similar to oxyhemoglobin. Most pulse oximeters sense COHb as O2Hb and overestimate the O2 saturation; however, pulse oximeters that measure O2Hb and COHb are available. MetHb increases absorption at each of the two wavelengths used by pulse oximeters. This causes the ratio of the two Hb forms to approach 1, which is usually represented as a saturation of 85% (see Chapter 11). Other interfering substances may cause the pulse oximeter to underestimate saturation.

Significance and Pathophysiology

See Interpretive Strategies 6-3. Arterial oxygen saturation estimated by pulse oximetry (SpO2) should approximate that measured by hemoximetry (SaO2) in healthy nonsmoking adults. Most pulse oximeters are capable of accuracy of ±2% of the actual saturation when SaO2 is above 90%. For SaO2 values of 85%–90%, accuracy may be slightly less. For very low saturations (i.e., less than 80%), pulse oximeter accuracy is less of an issue because the clinical implications are the same.

Pulse oximetry is most useful when it has been shown to correlate with blood oximetry in a patient in a known circumstance. When this is the case, pulse oximetry can be used for noninvasive monitoring, either continuously or by taking discrete measurements. Uses include monitoring of O2 therapy, ventilatory support, pulmonary or cardiac rehabilitation, bronchoscopy, surgical procedures, sleep studies, and cardiopulmonary exercise testing. In each of these applications, careful attention must be paid to minimizing known interfering agents or substances (see Box 6-2).

Because of its limitations, pulse oximetry should be used cautiously when assessing oxygen need. This is particularly true when using pulse oximetry to detect exercise desaturation. Pulse oximetry may not accurately reflect SaO2 during exertion. For this reason, SpO2, even if demonstrated to correlate with SaO2 at rest, may yield false-positive or false-negative results during exercise. Blood gas analysis with hemoximetry should be used to resolve discrepancies between the SpO2 reading and the patient’s clinical presentation.

Pulse oximetry may not be appropriate in all situations. To evaluate hyperoxemia (Pao2 greater than 100–150 mm Hg) or acid-base status in a patient, blood gas analysis is required. Measurement of O2 delivery, which depends on Hb concentration, cannot be adequately assessed by pulse oximetry alone.

Capnography

Description

Capnography includes continuous, noninvasive monitoring of expired CO2 and analysis of the single-breath CO2 waveform. Continuous monitoring of expired CO2 allows trending of changes in alveolar and dead space ventilation. Analysis of a single breath of expired CO2 measures the uniformity of both ventilation and pulmonary blood flow. End-tidal Pco2 (Petco2) is reported in millimeters of mercury.

Technique

Continuous monitoring of expired CO2 is performed by sampling gas from the proximal airway. This gas may be pumped to an infrared analyzer (see Chapter 11) or to a mass spectrometer. An alternative method inserts a “mainstream” sample window directly into the expired gas stream. The analyzer signal is then passed to either a recorder or a computer. CO2 waveforms may be displayed either individually (Figure 6-5) or as a series of deflections to form a trend plot. Petco2 may be read from the peaks of the waveforms. It can also be obtained by a simple peak detector and displayed digitally. Continuous CO2 monitoring is commonly used in patients with artificial airways in the critical care setting. Paco2 can be measured at intervals to establish a gradient with Petco2. Respiratory rate may be determined from the frequency of the CO2 waveforms. The change in CO2 concentration during a single expiration may be analyzed to detect ventilation-perfusion abnormalities (see Figure 6-5).

Technical problems involved in capnography include the necessity of accurate calibration and management of the gas sampling system (Criteria for Acceptability 6-3). Calibration using known gases (preferably two gas concentrations) is required if the system will be used to monitor Paco2. Many systems use ambient air, containing minimal CO2, and a 5% CO2 mixture for calibration (see Chapter 12). Condensation of water in sample tubing, connectors, or the sample chamber can affect accuracy. Some infrared analyzers may be affected if sample flow changes after calibration. Saturation of a desiccator column, if used, can also lead to inaccurate readings. Long sample lines or low sample flows can cause damping of the CO2 waveform, invalidating analysis of the shape of the expired gas curve.

Significance and Pathophysiology

See Interpretive Strategies 6-4. In healthy patients, CO2 rises to a plateau as alveolar gas is expired (Figure 6-5). If all lung units empty CO2 evenly, the plateau appears flat. However, even healthy lungs have ventilation and blood flow imbalances. Healthy lung units empty CO2 at varying rates. Alveolar CO2 concentration increases slightly as the exhalation continues. Petco2 theoretically should not exceed Paco2. In healthy patients at rest, the Petco2 is usually close to the arterial value. During maximal exercise Petco2 may exceed Paco2. When ventilation and perfusion become grossly mismatched (e.g., in severe obstruction), end-tidal CO2 may exceed the Paco2 as well. Paco2 reflects gas-exchange characteristics of the entire lung. Thus, Petco2 may differ significantly if some lung units are poorly ventilated.

Continuous CO2 analysis provides useful data for monitoring critically ill patients, particularly those requiring ventilatory support. Petco2 measurements allow trending of changes in Paco2 provided there is little or no change in the shape of the CO2 waveform (indicating image/imageabnormalities). When a reference blood gas sample is obtained, Petco2 can be used as a continuous, noninvasive monitor. Respiratory rate can be measured from the frequency of expired CO2 waveforms. Marked changes in breathing rate (e.g., hyperpnea or apnea) can be detected quickly. Analysis of the individual CO2 waveforms, along with Paco2, may help identify abrupt changes in dead space. This can be useful in detecting pulmonary embolization or reduced cardiac output.

Problems related to ventilatory support devices can also be detected. Disconnection or leaks in breathing circuits can be quickly recognized by the loss of the CO2 signal. Increased mechanical dead space (i.e., gas rebreathed in the ventilator circuit) can be identified by a baseline CO2 concentration greater than zero. Irregularities in the CO2 waveform often signal that the patient is “out of phase” with the ventilator.

The shape of the expired CO2 curve is determined by ventilation-perfusion matching. Only lung units that are ventilated and perfused contribute CO2 to expired gas. In patients without lung disease, the CO2 waveform shows a flat initial segment of anatomic dead space gas containing little or no CO2. This phase is followed by a rapid increase in CO2 concentration, reflecting a mixture of dead space and alveolar gas. Finally, an “alveolar” plateau occurs in which gas composition changes only slightly. This slight change is caused by different emptying rates of various lung units. The absolute concentration of CO2 at the alveolar plateau depends on factors such as minute ventilation and CO2 production. Dead space–producing disease (e.g., pulmonary embolization or marked decrease in cardiac output) may show a profound decrease in expired CO2 concentration. Patients who have pulmonary disease, especially obstruction, show poorly delineated phases of the CO2 washout curve. The alveolar plateau may actually be a continuous slope throughout expiration, causing the measurement of Petco2 to be misleading.

Shunt calculation

Description

A shunt is that portion of the cardiac output that passes from one side of the heart to the other side without participating in gas exchange. This most commonly occurs from the right side to the left side. When the defect is in the lungs, it is a pulmonary shunt. If it occurs in the heart (e.g., atrial or ventricular septal defects), it is referred to as an intracardiac shunt. The shunt calculation determines the ratio of shunted blood (images) to total perfusion (imaget). Shunt is reported as a percent of total cardiac output, or sometimes as a simple fraction. Left-to-right shunting can occur when cardiac physiology is abnormal, such as patent ductus arteriosus (PDA). These types of shunts are usually detected by Doppler echocardiography rather than by the shunt measurement, as described.

Technique

Two methods for measuring the shunt fraction are available. The first uses O2 content differences between arterial and mixed venous blood. The first method is called the physiologic shunt equation:

Equation 1.

Q·sQ·t=CcO2CaO2CcO2Cv¯O2

image

where:

CcO2= O2 content of end-capillary blood, estimated from saturation associated with calculated PAo2

CaO2= arterial O2 content, measured from an arterial sample

Cv¯imageo2= mixed venous O2 content, measured from a sample obtained from a pulmonary artery catheter

The term in the denominator of this equation reflects potential arterialization of mixed venous blood. The term in the numerator reflects the actual arterialization.

The second method is more commonly used in clinical practice. The patient breathes 100% O2 in a unidirectional valved system until the Hb is completely saturated (see Figure 6-6). Twenty minutes of O2 breathing is usually sufficient. Percent shunt is then calculated from differences in dissolved O2. The second method is called the clinical shunt equation:

Equation 2.

Q·sQ·t=(PAO2PaO2)×0.0031C(av¯)O2+[(PAO2PaO2)×0.0031]

image

where:

PAo2= alveolar O2 tension

Pao2= arterial O2 tension

C(a–v¯image)o2= arteriovenous O2 content difference

0.0031= solubility factor for O2 at 37°C

When the patient is breathing 100% O2, PAo2 can be estimated from an abbreviated form of the alveolar gas equation as follows:

PAO2=PBPH2O(PaCO20.8)

image

where:

PB= barometric pressure

PH2O= partial pressure of water vapor at body temperature (47 mm Hg)

Paco2= arterial CO2 tension (as an estimate of alveolar Pco2)

0.8= normal (assumed) respiratory exchange ratio

This calculation of alveolar Po2 assumes that the inspired gas is 100% O2. If the FIO2 is measured and found to be less than 1, the standard alveolar air equation can be used (see Evolve http://evolve.elsevier.com/Mottram/Ruppel/).

The clinical shunt calculation is accurate only when Hb is completely saturated. This normally requires a Pao2 greater than 150 mm Hg. A saturation of 100% is usually easily accomplished if O2 is breathed long enough. If breathing 100% O2 does not increase the Pao2 high enough to completely saturate Hb, the content difference method (physiologic method) should be used. With either method, shunt fraction may be multiplied by 100 and reported as a percentage (e.g., 0.20 ratio × 100 equals a 20% shunt).

Several technical considerations should be noted regarding shunt measurement (Criteria for Acceptability 6-4). Using O2 content differences (equation 1) requires a pulmonary artery catheter to obtain mixed venous O2 content. Using dissolved O2 differences (equation 2) also relies on measured C(a-v¯image)o2. If placement of a pulmonary artery catheter is not practical, the clinical shunt equation may be used with an assumed a-v¯image content difference or a presentation of the calculated data with several a-v¯image content differences (Table 6-5) (see the discussion on Significance and Pathophysiology in this section). The estimated technique assumes the subject is breathing 100% oxygen, so a closed circuit with directional value is required. This can be accomplished with a demand valve driven with a 50 psig 100% oxygen gas source or reservoir bag (e.g., Douglas bag) filled with 100% oxygen connected to a directional value. The subject needs to be monitored for leaks around the mouth or noseclips. The sample should be collected in a glass syringe or analyzed immediately to reduce the decrease in Pao2 secondary to environmental exposure (How To Box 6-3).

The physiologic shunt equation may be used for patients on any known FIO2. The clinical shunt measurement requires O2 breathing for 20 minutes or longer. Prolonged O2 breathing may be contraindicated in patients whose respiration is driven by hypoxemia. Breathing 100% O2 washes nitrogen (N2) out of the lungs. Washout of N2 combined with O2 uptake by perfusing blood flow can reduce the size of alveoli to their critical limit. In poorly ventilated lung units, this may cause alveolar collapse. The effect of this “nitrogen shunting” may be shunt values that are falsely high because some shunting was induced by the test itself.

Significance and Pathophysiology

See Interpretive Strategies 6-5. In healthy individuals, approximately 5% of the cardiac output is shunted past the pulmonary system. An increased pulmonary shunt fraction indicates that some lung units have little ventilation in relation to their blood flow. These patterns may be found in both obstructive and restrictive diseases. However, even in severe obstruction or restriction, blood flow to areas of poor ventilation may be reduced by the lesions themselves. In emphysema, destruction of the alveolar septa obliterates pulmonary capillaries. As the terminal airways lose their support, they also have reduced blood flow. In poorly ventilated lung units, vasoconstriction of pulmonary arterioles redirects blood flow away from the affected area. In these cases, there may be minimal shunting, even though severe ventilatory impairment exists.

Intrapulmonary shunting is common in acute disease patterns such as atelectasis or foreign body aspiration. Diseases such as pneumonia or adult (in some references this has been broaden to “acute”) respiratory distress syndrome (ARDS) often result in a “shunt-like effect.” This is caused by reduced ventilation in relation to blood flow in many lung units. Foreign body aspiration may cause shunting by blocking an airway and depriving all distal lung units of ventilation. Blood flow to the affected units cannot participate in gas exchange. The degree of shunting is directly related to the number of lung units with image/imageratios close to zero.

Intrapulmonary shunting may also occur when the pulmonary vascular system is involved, as is the case in hepatopulmonary syndrome. Patients who have liver disease with portal hypertension commonly have arteriovenous malformations that allow significant volumes of blood to pass through the lungs without participating in gas exchange. These malformations are prominent at the lung bases and result in increased shunting when the patient is upright.

Intracardiac shunting occurs when defects in the atrial or ventricular septa allow mixed venous blood to pass from the right side of the heart to the left side of the heart without traversing the pulmonary capillaries. In the ventricles, shunting usually requires an elevated right-heart pressure to overcome the normal pressure difference between the systemic and pulmonary systems. Atrial septal defects, such as patent foramen ovale (PFO), allow blood to pass from the right atrium to the left atrium during systole.

Very large shunt values (greater than 30%) suggest that a significant volume of blood is moving from the right side of the heart to the left side of the heart without participating in gas exchange. Further testing may be required to determine whether the shunt is occurring in the lungs or in the heart. Echocardiography with contrast media or cardiac catheterization may be necessary to identify intracardiac shunts.

The accuracy of clinical shunt measurement (i.e., dissolved O2 differences) depends on the accuracy of Po2 determinations. In small shunts, Hb becomes 100% saturated. The difference between alveolar and arterial Po2 values results simply from the amount of O2 dissolved. The difference between the actual content of dissolved oxygen and the content that can potentially dissolve is the basis for the calculation. Measurements of Po2 used for shunt calculations (200–600 mm Hg) are much higher than the normal physiologic range. Additional calibration and quality control of the Po2 electrode may be necessary to provide accurate values for oxygen tension. Blood gases drawn for shunt studies should be analyzed immediately to avoid large changes in Pao2 values caused by cell metabolism in the specimen and the effect of diffusion across a plastic syringe if used. Glass syringes should be used if any delay in sampling greater than a few minutes is expected.

Table 6-6

Causes of Hypoxia and Hypoxemia

Type Cause
Hypoxemic hypoxia Altitude
Hypoventilation
Shunt
V./Q.image mismatching
Diffusion
Circulatory hypoxia Hypovolemia
Reduced cardiac output (impaired LV function, outflow tract abnormalities, etc.)
Hemic hypoxia Anemia, dyshemoglobinemias (COHb, MetHb, etc.)
Histotoxic hypoxia Cyanide poisoning
Demand hypoxia Seizures, exercise

image

The calculated shunt fraction also depends on the O2 content difference between arterial and mixed venous blood. C(a-v¯image)o2 is a component of the denominator in the clinical shunt equation. The a-v¯imagecontent difference is determined not only by the lungs but also by cardiac output and the perfusion status of the tissues. Ideally, the value used in the equation should be measured rather than estimated. Arterial content can be determined easily from a sample taken from a peripheral artery. However, mixed venous content can only be measured accurately from a pulmonary artery sample. In patients who do not have a pulmonary artery catheter in place, an estimated value must be used. C(a-v¯image)o2 values from 4.5–5.0 vol% are reasonable content differences in patients who have good cardiac output and perfusion status. Values of 3.5 vol% may be more realistic in patients who are critically ill.

In some instances, a-v¯image content difference cannot be reliably estimated, or Hb cannot be maximally saturated by breathing 100% oxygen. In such cases, the alveolar-arterial oxygen gradient P(a−a)o2 may be useful as an index of ventilation to blood flow. In healthy patients breathing 100% O2 at sea level, arterial Pao2 should increase to approximately 600 mm Hg. images/imaget does not directly provide absolute values for images, but if the cardiac output (imaget) is known, images can be determined simply. Measurement of the shunt fraction is sometimes performed in conjunction with the determination of the VD/VT ratio (see Chapter 5) to assess both types of gas exchange abnormalities together.

Summary

Case Studies

Case 6-1

History

A 57-year-old man referred to the pulmonary function laboratory for increasing shortness of breath. He admits having a chronic cough with a production of thick, white sputum, mainly upon awakening. C.O. reports that he quit smoking 3 months ago. He averaged approximately two packs of cigarettes per day for 30 years (60 pack years) before quitting. His referring physician performed a pulse oximetry in the outpatient clinic and obtained readings of 90%–91%. Complete pulmonary function studies with arterial blood gases were requested.

Pulmonary function tests

Personal Data

Sex: Male
Age: 57 yr
Height: 66 in. (168 cm)
Weight: 197 lb (89.5 kg)

Spirometry

Before Drug LLN* Predicted % After Drug % %Change
FVC (L)  3.97  3.51   4.10 97  4.04 99  2
FEV1 (L)  2.31  2.35   2.99 77  2.50 84  8
FEV1% (%) 58 64  73 62  7
MVV (L) 83 116 72 92 79 11

image

*Lower limit of normal.

Lung Volumes

Before Drug LLN* Predicted %
VC (L)  3.97 3.20  4.10  97
IC (L)  2.66  2.76  96
ERV (L)  1.31  1.35  97
FRC (L)  3.65  3.42 107
RV (L)  2.34  2.07 113
TLC (L)  6.31 4.64  6.18 102
RV/TLC (%) 37 34

image

*Lower limit of normal.

Dlco

Before Drug LLN* Predicted %
Dlco (mL/min/mm Hg) 16.7 19.5 26.3 63
Dlco (corrected) 16.7 26.3 63
Dl/VA 5.09 4.38 86

image

*Lower limit of normal.

Blood Gases

pH 7.37
Pco2 (mm Hg) 44
Po2 (mm Hg) 57
HCO3image (mEq/L) 26.4
BE (mEq/L) ~ 1.7
Hb (g/dL) 16.2
O2Hb (%) 80.1
COHb (%) 5.9
MetHb (%) 0.2

image

Discussion

Interpretation

All spirometry, lung volumes, and diffusing capacity maneuvers were performed acceptably. Spirometry reveals moderate airway obstruction with minimal response to inhaled bronchodilators. Lung volumes by He dilution show a normal TLC but with a slightly increased FRC, RV, and RV/TLC ratio. These changes are suggestive of air trapping. Dlco is moderately decreased even after a correction for increased Hb and elevated COHb. Arterial blood gas results reveal normal acid-base status with a Pco2 of 44. There is moderate to severe hypoxemia as indicated by a Po2 of 57 and an oxyhemoglobin saturation of 80%. The hypoxemia is further aggravated by an increased COHb consistent with cigarette smoking or environmental exposure.

Impression: Moderate obstructive airways disease with no significant improvement after inhaled bronchodilator. There is a moderate loss of diffusing capacity. There is significant hypoxemia on room air, which is further increased by an elevation of COHb. Recommend further investigation of source of CO and clinical evaluation of O2 supplementation.

Cause of Symptoms

This case demonstrates the importance of arterial blood gas analysis in the diagnosis of pulmonary disorders. The subject’s complaint of increased shortness of breath was not explained by the borderline value of pulse oximetry (SpO2) performed in the referring physician’s office. The referring physician correctly suspected a pulmonary problem resulting from the subject’s previous smoking history.

The spirometric measurement shows obstruction as evidenced by the FEV1% of 58%, well below the lower limit of normal (64% for this subject). The subject’s FEV1 is also below its LLN. The response of FEV1 to the inhaled bronchodilator is slightly less than 200 mL and represents only an 8% improvement over prebronchodilator values. Lung volumes are also consistent with an obstructive pattern, suggesting the beginnings of air trapping.

Diffusing capacity is also reduced in a pattern consistent with mild to moderate airway obstruction. The predicted Dlco has been corrected for an elevated Hb and COHb. The combination of these two corrections results in a predicted Dlco that is identical to the uncorrected predicted value (see Chapter 5).

Of the pulmonary function variables measured, the arterial blood gas values are most abnormal. Although pH and Pco2 are within normal limits, Po2 is markedly decreased. As a result, oxygen saturation is low. Elevated COHb further complicates oxygenation. This level is characteristic of individuals who currently smoke or who are chronically exposed to low levels of CO in their environment. The subject’s spouse reported that he was still smoking 5–10 cigarettes per day.

Case 6-2

History

A 31-year-old woman referred to the pulmonary function laboratory for a shunt study. Her chief complaint is shortness of breath with exertion as well as at other times. Her referring physician suspected a shunt and requested a shunt study. The subject was in no apparent distress on arrival at the laboratory. She never smoked and had no significant environmental exposure to respiratory irritants. Her mother died of a stroke at age 50 years, but there is no heart or lung disease in her immediate family.

Shunt study

Personal Data

Age: 31 yr
Height: 69 in. (175 cm)
Weight: 200 lb (91 kg)
Race: White

Blood Gases (Drawn After 20 Minutes of O2 Breathing)

FIO2 1.00
PB 752
pH 7.43
Pco2 38
Po2 557
HCO3image 24.1
BE 0.1
Hb 7.4
O2Hb 99.6
COHb 0.3
MetHb 0.1

image

Discussion

Cause of Symptoms

The subject’s primary symptom of dyspnea does not appear to be caused by any significant shunting. In a shunt, blood passes from the right side of the heart to the left side without coming into contact with alveolar gas. The shunt may be in the heart or in the lungs themselves. Breathing high concentrations of oxygen will not relieve this problem. This subject increased her Pao2 appropriately, which rules out a large shunt.

A more likely cause of the symptoms described is the subject’s low Hb level. Severe anemia reduces the arterial oxygen content dramatically. Oxygen delivery to the tissues is reduced. Dyspnea can result during exertion or times of increased metabolic demand. The subject’s arterial content (mL/dL) can be calculated as follows:

CaO2=(13.4×Hb×O2Hb)+(PaO2×0.0031)

image

=(13.4×7.4×0.996)+(557×0.0031)

image

=(9.9)+(1.7)

image

CaO2=11.6

image

This value is much lower than the normal arterial content of about 20 mL/dL. It should be noted that the term in the second parentheses represents dissolved O2 and would be much lower if the subject were breathing room air.

Self-Assessment Questions

Entry-level

1. A patient referred for a pulmonary function test has blood gases drawn, and the following data are reported:

pH 7.43
Paco2 (mm Hg) 38
Hco3 (mEq/L) 25.1

    These values are consistent with:

2. A patient with COPD has blood gases drawn and the following values are reported:

pH 7.37
Pco2 47 mm Hg
Po2 52 mm Hg
Hco3 27.1 mEq/L
Hb 15.8 g/dL
O2Hb 82%

    Which of the following best describes these results?

3. A patient being monitored by ECG has a heart rate (HR) of 120/min. A pulse oximeter on this patient displays a saturation of 84% with an HR of 118. Which of the following should the pulmonary function technologist conclude, based on these findings?

4. An outpatient is referred for arterial blood gas analysis. While performing the modified Allen’s test, the pulmonary function technologist notes that the patient’s hand reperfuses after 30 seconds. The technologist should:

5. A patient with interstitial lung disease has an arterial oxygen tension (Pao2) of 60 mm Hg, but hemoximetry is not performed. What calculated SaO2 should the pulmonary function technologist report?

Advanced

6. A patient with chronic bronchitis has a resting SpO2 of 80%, but an arterial blood gas reveals an SaO2 of 90%. Which of the following should the pulmonary function technologist conclude?

7. A patient who has COPD performs a maximal exercise test. At peak exercise, his SpO2 is 94% and a blood gas shows an SaO2 of 87%. Which of the following might explain these results?

8. A patient being monitored by capnography shows the following data

Time 8:35 8:40 8:45 8:50 8:55 9:00
Petco2 38 39 39 29 25 26

image

    Which of the following best explains the change occurring at 8:50?

9. A patient with dyspnea on exertion has arterial blood gases drawn at rest before beginning an exercise test, and the following data are obtained:

pH 7.44
Paco2 36 mm Hg
HCO3image 24.4 mEq/L
Pao2 64 mm Hg
SaO2 91%
Hb 6.9 mg/dL
COHb 1.2%

image

    What is the subject’s arterial oxygen content (Cao2)?

10. A patient with dyspnea at rest has a clinical shunt study performed. The following data are obtained:

PB 761 mm Hg
FIO2 1.00
Pao2 597 mm Hg
Paco2 38 mm Hg
SaO2 100%
Hb 15.1 mg/dL

    On the basis of these results, the patient’s images/imaget is approximately: