Basic Physics for the Anaesthetist

Published on 27/02/2015 by admin

Filed under Anesthesiology

Last modified 22/04/2025

Print this page

rate 1 star rate 2 star rate 3 star rate 4 star rate 5 star
Your rating: none, Average: 3 (1 votes)

This article have been viewed 5156 times

Basic Physics for the Anaesthetist

K nowledge of some physics is required in order to understand the function of many items of apparatus for anaesthesia delivery and physiological monitoring. This chapter emphasizes the more elementary aspects of physical principles and it is hoped that the reader may be stimulated to study some of the books written specifically for anaesthetists and which examine this topic in greater detail (see ‘Further reading’). Sophisticated measurement techniques may be required for more complex types of anaesthesia, in the intensive care unit and during anaesthesia for severely ill patients, and an understanding of the principles involved in performing such measurements is required in the later stages of the anaesthetist’s training.

This chapter concentrates on the more common applications, including pressure and flow in gases and liquids, electricity and electrical safety. However, it is necessary first to consider some basic definitions.

BASIC DEFINITIONS

It is now customary in medical practice to employ the International System (Système Internationale; SI) of units. Common exceptions to the use of the SI system include measurement of arterial pressure and, to a lesser extent, gas pressure. The mercury column is used commonly to calibrate electronic arterial pressure measuring devices and so ‘mmHg’ is retained. Pressures in gas cylinders are also referred to frequently in terms of the ‘normal’ atmospheric pressure of 760 mmHg; this is equal to 1.013 bar (or approximately 1 bar). Low pressures are expressed usually in the SI unit of kilopascals (kPa) whilst higher pressures are referred to in bar (100 kPa = 1 bar). The basic and derived units of the SI system are shown in Table 14.1.

TABLE 14.1

Physical Quantities

image

The fundamental quantities in physics are mass, length and time.

Mass (m) is defined as the amount of matter in a body. The unit of mass is the kilogram (kg), for which the standard is a block of platinum held in a Physics Reference Laboratory.

Length (l) is defined as the distance between two points. The SI unit is the metre (m), which is defined as the distance occupied by a specified number of wavelengths of light.

Time (t) is measured in seconds. The reference standard for time is based on the frequency of resonation of the caesium atom.

From these basic definitions, several units of measurement may be derived:

Volume has units of m3.

Density is defined as mass per unit volume:

image

Velocity is defined as the distance travelled per unit time:

image

Acceleration is defined as the rate of change of velocity:

image

Force is that which is required to give a mass acceleration:

image

The SI unit of force is the newton (N). One newton is the force required to give a mass of 1 kg an acceleration of 1 m s–1:

1 N = 1 kg m s–2

Weight is the force of the earth’s attraction for a body. When a body falls freely under the influence of gravity, it accelerates at a rate of 9.81 m s–2 (g):

image

Momentum is defined as mass multiplied by velocity:

momentum =  ×  v

Work is undertaken when a force moves an object:

work = force  ×  distance

=   ×  l N m (or joules, J)

Energy is the capacity for undertaking work. Thus it has the same units as those of work. Energy can exist in several forms, such as mechanical (kinetic energy [KE] or potential energy [PE]), thermal or electrical and all have the same units.

Power (P) is the rate of doing work. The SI unit of power is the watt, which is equal to 1 J s− 1:

power = work per unit time

= joules per second = watt (W)

Pressure is defined as force per unit area:

image

image

= pascal (Pa)

As 1 Pa is a rather small unit, it is more common in medical practice to use the kilopascal (kPa): 1 kPa ≈ 7.5 mmHg.

FLUIDS

Substances may exist in solid, liquid or gaseous form. These forms or phases differ from each other according to the random movement of their constituent atoms or molecules. In solids, molecules oscillate about a fixed point, whereas in liquids the molecules possess higher velocities and therefore higher kinetic energy; they move more freely and thus do not bear a constant relationship in space to other molecules. The molecules of gases possess even higher kinetic energy and move freely to an even greater extent.

Both gases and liquids are termed fluids. Liquids are incompressible and at constant temperature occupy a fixed volume, conforming to the shape of a container; gases have no fixed volume but expand to occupy the total space of a container. Nevertheless the techniques for analysing the behaviour of liquids and gases (or fluids in general) in terms of their hydraulic and thermodynamic properties are very similar.

In the process of vaporization, random loss of liquid molecules with higher kinetic (thermal) energies from the liquid occurs while vapour molecules randomly lose thermal (kinetic) energy and return to the liquid state. Heating a liquid increases the kinetic energy of its molecules, permitting a higher proportion to escape from the surface into the vapour phase. The acquisition by these molecules of higher kinetic energy requires an energy source and this usually comes from the thermal energy of the liquid itself, which leads to a reduction in its thermal energy as vaporization occurs and hence the liquid cools.

Collision of randomly moving molecules in the gaseous phase with the walls of a container is responsible for the pressure exerted by a gas. The difference between a gas and a vapour will be discussed later.

Behaviour of Gases

The Gas Laws

There are three gas laws which determine the behaviour of gases and which are important to anaesthetists. These are derived from the kinetic theory of gases; they depend on the assumption that the substances concerned are perfect gases (rather than vapours), and they assume a fixed mass of gas.

Boyle’s law states that, at constant temperature, the volume (V) of a given mass of gas varies inversely with its absolute pressure (P):

PV = k1

Charles’ law states that, at constant pressure, the volume of a given mass of gas varies directly with its absolute temperature (T):

V = k2T

The third gas law (sometimes known as Gay-Lussac’s law) states that, at constant volume, the absolute pressure of a given mass of gas varies directly with its absolute temperature:

P = k3T

Combining these three gas laws:

PV = kT

or

image

where suffixes 1 and 2 represent two conditions different in P, V and T of the gas. Note that where a change of conditions occurs slowly enough for T1 = T2, conditions are said to be isothermal, and the combined gas law could be thought of as another form of Boyle’s law.

The behaviour of a mixture of gases in a container is described by Dalton’s law of partial pressures. This states that, in a mixture of gases, the pressure exerted by each gas is the same as that which it would exert if it alone occupied the container. Dalton’s law can be used to compare volumetric fractions (concentrations) to calculate partial pressures, which are an important concept in anaesthesia. Thus, in a cylinder of compressed air at a pressure of 100 bar, the pressure exerted by nitrogen is equal to 79 bar, as the fractional concentration of nitrogen is 0.79.

Avogadro’s Hypothesis

Avogadro’s hypothesis, also deduced from the kinetic theory of gases, states that equal volumes of gases at the same temperature and pressure contain equal numbers of molecules.

Avogadro’s number is the number of molecules in 1 gram-molecular weight of a substance and is equal to 6.022 × 1023.

Under conditions of standard temperature and pressure (0 °C and 1.013 bar), 1 gram-molecular weight (i.e. 28 g of nitrogen or 44 g of carbon dioxide) of any gas occupies a volume of 22.4 litres (L).

These data are useful in calculating, for example, the quantity of gas produced from liquid nitrous oxide. The molecular weight of nitrous oxide is 44. Thus, 44 g of N2O occupy a volume of 22.4 L at standard temperature and pressure (STP). If a full cylinder of N2O contains 3.0 kg of liquid, then vaporization of all the liquid would yield:

image

The gas laws can be applied to calculate its volume at, say, room temperature, bearing in mind that the Kelvin scale of temperature should be used for such calculations.

Critical Temperature

If the temperature of a vapour is low enough, then sufficient application of pressure to it will result in its liquefication. If the vapour has a higher temperature, implying greater molecular kinetic energy, no amount of compression liquefies it. The critical temperature of such a substance is the temperature above which that substance cannot be liquefied by compression alone. A substance in such a state is considered a gas, whereas a substance below its critical temperature can be considered a vapour.

The critical temperature of oxygen is −118 °C, that of nitrogen is −147 °C, and that of air is −141 °C. Thus, at room temperature, cylinders of these substances contain gases. In contrast, the critical temperature of carbon dioxide is 31 °C and that of nitrous oxide is 36.4 °C. The critical pressures are 73.8 and 72.5 bar respectively; at higher pressures, cylinders of these substances at UK room temperature contain a mixture of gas and liquid.

Clinical Application of the Gas Laws

A ‘full’ cylinder of oxygen on an anaesthetic machine contains compressed gaseous oxygen at a pressure of 137 bar (2000 lb in–2) gauge pressure. If the cylinder of oxygen empties and the temperature remains constant, the volume of gas contained is related linearly to its pressure (by Boyle’s law). In practice, linearity is not followed because temperature falls as a result of adiabatic expansion of the compressed gas; the term adiabatic implies a change in the state of a gas without exchange of heat energy with its surroundings.

By contrast, a nitrous oxide cylinder contains liquid nitrous oxide in equilibrium with its vapour. The pressure in the cylinder remains relatively constant at the saturated vapour pressure for that temperature as the cylinder empties to the point at which liquid has totally vaporized. Subsequently, there is a linear decline in pressure proportional to the volume of gas remaining within the cylinder.

Filling Ratio: The degree of filling of a nitrous oxide cylinder is expressed as the mass of nitrous oxide in the cylinder divided by the mass of water that the cylinder could hold. Normally, a cylinder of nitrous oxide is filled to a ratio of 0.67. This should not be confused with the volume of liquid nitrous oxide in a cylinder. A ‘full’ cylinder of nitrous oxide at room temperature is filled to the point at which approximately 90% of the interior of the cylinder is occupied by liquid, the remaining 10% being occupied by nitrous oxide vapour. Incomplete filling of a cylinder is necessary because thermally induced expansion of the liquid in a totally full cylinder may cause cylinder rupture. Because vapour pressure increases with temperature, it is necessary to have a lower filling ratio in tropical climates than in temperate climates.

Entonox: Entonox is the trade name for a compressed gas mixture containing 50% oxygen and 50% nitrous oxide. The mixture is compressed into cylinders containing gas at a pressure of 137 bar (2000 lb in–2) gauge pressure (see below). The nitrous oxide does not liquefy because the two gases in this mixture ‘dissolve’ in each other at high pressure. In other words, the presence of oxygen reduces the critical temperature of nitrous oxide. The critical temperature of the mixture is −7 °C, which is called the ‘pseudocritical temperature’. Cooling of a cylinder of Entonox to a temperature below −7 °C results in separation of liquid nitrous oxide. Use of such a cylinder results in oxygen-rich gas being released initially, followed by a hypoxic nitrous oxide-rich gas. Consequently, it is recommended that when an Entonox cylinder may have been exposed to low temperatures, it should be stored horizontally for a period of not less than 24 h at a temperature of 5 °C or above. In addition, the cylinder should be inverted several times before use.

Pressure Notation in Anaesthesia

Although the use of SI units of measurement is generally accepted in medicine, a variety of ways of expressing pressure is still used, reflecting custom and practice. Arterial pressure is still referred to universally in terms of mmHg because a column of mercury is still used occasionally to measure arterial pressure and also to calibrate electronic devices.

Measurement of central venous pressure is sometimes referred to in cm H2O because it can be measured using a manometer filled with saline, but it is more commonly described in mmHg when measured using an electronic transducer system. Note that, although we colloquially speak of ‘cm H2O’ or ‘mmHg’, the actual expression for pressure measured by a column of fluid is P = ρ.g.H, where ρ is fluid density, g is acceleration due to gravity and H is the height of the column. Because mercury is 13.6 times more dense than water, a mercury manometer can measure a given pressure with a much shorter length of column of fluid. For example atmospheric pressure (PB) exerts a pressure sufficient to support a column of mercury of height 760 mm (Fig. 14.1).

1 atmospheric pressure = 760 mmHg

= 1.01325 bar

= 760 torr

= 1 atmosphere absolute (ata)

= 14.7 lb in–2

= 101.325 kPa

= 10.33 metres of H2O

In considering pressure, it is necessary to indicate whether or not atmospheric pressure is taken into account. Thus, a diver working 10 m below the surface of the sea may be described as compressed to a depth of 1 atmosphere or working at a pressure of 2 atmospheres absolute (2 ata).

In order to avoid confusion when discussing compressed cylinders of gases, the term gauge pressure is used. Gauge pressure describes the pressure of the contents above ambient pressure. Thus, a full cylinder of oxygen has a gauge pressure of 137 bar, but the contents are at a pressure of 138 bar absolute.

GAS REGULATORS

Pressure Relief Valves

The Heidbrink valve is a common component of many anaesthesia breathing systems. In the Magill breathing system, the anaesthetist may vary the force in the spring(s), thereby controlling the pressure within the breathing system (Fig. 14.2). At equilibrium, the force exerted by the spring is equal to the force exerted by gas within the system:

force (F) = gas pressure (P) × disc area (A)

Modern anaesthesia systems contain a variety of pressure relief valves, in each of which the force is fixed so as to provide a gas escape mechanism when pressure reaches a preset level. Thus, an anaesthetic machine may contain a pressure relief valve operating at 35 kPa, situated on the back bar of the machine between the vaporizers and the breathing system to protect the flowmeters and vaporizers from excessive pressures. Modern ventilators contain a pressure relief valve set at 7 kPa to protect the patient from barotrauma. A much lower pressure is set in relief valves which form part of anaesthetic scavenging systems and these may operate at pressures of 0.2–0.3 kPa to protect the patient from negative pressure applied to the lungs.

Pressure-Reducing Valves (Pressure Regulators)

Pressure regulators have two important functions in anaesthetic machines:

Modern anaesthetic machines are designed to operate with an inlet gas supply at a pressure of 3–4 bar (usually 4 bar in the UK). Hospital pipeline supplies also operate at a pressure of 4 bar and therefore pressure regulators are not required between a hospital pipeline supply and an anaesthetic machine. In contrast, the contents of cylinders of all medical gases (i.e. oxygen, nitrous oxide, air and Entonox) are at much higher pressures. Thus, cylinders of these gases require a pressure-reducing valve between the cylinder and the flowmeter.

The principle on which the simplest type of pressure-reducing valve operates is shown in Figure 14.3. High-pressure gas enters through the valve and forces the flexible diaphragm upwards, tending to close the valve and prevent further ingress of gas from the high-pressure source.

If there is no tension in the spring, the relationship between the reduced pressure (p) and the high pressure (P) is very approximately equal to the ratio of the areas of the valve seating (a) and the diaphragm (A):

p.A = P.a

or

image

By tensing the spring, a force F is produced which offsets the closing effect of the valve. Thus, p may be increased by increasing the force in the spring.

Without the spring, the simple pressure regulator has the disadvantage that reduced pressure decreases proportionally with the decrease in cylinder pressure. The addition of a force from the spring considerably reduces but does not eliminate this problem, and in order to overcome it, newer pressure regulators contain an extra closing spring. During high flows, the input to the valve may not be able to keep pace with the output. This can cause the regulated pressure to fall. A two-stage regulator can be employed in order to overcome this. Simple one-stage regulators are often designed for use with a specific gas. A universal regulator, in which the body is used for all gases but has different seatings and springs fitted for each specific gas, is now available.

Pressure Demand Regulators

These are regulators in which gas flow occurs when an inspiratory effort is applied to the outlet port. The Entonox valve is a two-stage regulator and its mode of action is demonstrated in Figure 14.4. The first stage is identical to the reducing valve described above. The second-stage valve contains a diaphragm. Movement of this diaphragm tilts a rod, which controls the flow of gas from the first-stage valve. The second stage is adjusted so that gas flows only when pressure is below atmospheric.

Flow of Fluids

Viscosity (η) is the constant of proportionality relating the stress (τ) between layers of flowing fluid (or between the fluid and the vessel wall), and the velocity gradient across the vessel, dv/dr.

Hence:

image

or

image

In this context, velocity gradient is equal to the difference between velocities of different fluid layers divided by the distance between layers (Fig. 14.5B). The units of the coefficient of viscosity are Pascal seconds (Pa s).

Fluids for which η is constant are referred to as Newtonian fluids. However, some biological fluids are non-Newtonian, an example of which is blood; viscosity changes with the rate of flow of blood (as a result of change in distribution of cells) and, in stored blood, with time (blood thickens on storage).

Viscosity of liquids diminishes with increase in temperature, whereas viscosity of a gas increases with increase in temperature. An increase in temperature is due to an increase of kinetic energy of fluid molecules. This can be thought of as causing a freeing up of intermolecular bonds in liquids, and an increase in intermolecular collisions in gas.

Laminar Flow

Laminar flow through a tube is illustrated in Figure 14.5A. In this situation, there is a smooth, orderly flow of fluid such that molecules travel with the greatest velocity in the axial stream, whilst the velocity of those in contact with the wall of the tube may be virtually zero. The linear velocity of axial flow is twice the average linear velocity of flow.

In a tube in which laminar flow occurs, the relationship between flow and pressure is given by the Hagen–Poiseuille formula:

image

where image is the flow, ∆P is the pressure gradient along the tube, r is the radius of the tube, η is the viscosity of fluid and l is the length of the tube.

The Hagen–Poiseuille formula applies only to Newtonian fluids and to laminar flow. In non-Newtonian fluids such as blood, increase in velocity of flow may alter viscosity because of variation in the dispersion of cells within plasma.

Turbulent Flow: Flow of Fluids Through Orifices

In turbulent flow, fluid no longer moves in orderly planes but swirls and eddies around in a haphazard manner as illustrated in Figure 14.6. Essentially, turbulent flow is less efficient in the transport of fluids because energy is wasted in the eddies, in friction and in noise (bruits). Although viscosity is the important physical variable in relation to the behaviour of fluids in laminar flow, turbulent flow is more markedly affected by changes in fluid density.

It may be seen from Figure 14.7 that the relationship between pressure and flow is linear within certain limits. However, as velocity increases, a point is reached (the critical point or critical velocity) at which the characteristics of flow change from laminar to turbulent. The critical point is dependent upon several factors, which were investigated by the physicist Osborne Reynolds. These factors are related by the formula used for calculation of Reynolds’ number:

image

where v is the fluid linear velocity, r is the radius of the tube, ρ is the fluid density and η is its viscosity.

Studies with cylindrical tubes have shown that if Reynolds’ number exceeds 2000, flow is likely to be turbulent, whereas a Reynolds’ number of less than 2000 is usually associated with laminar flow. However, localized areas of turbulent flow can occur at lower Reynolds’ numbers (i.e. at lower velocities) when there are changes in fluid direction, such as at bends, or changes in cross sectional area of the tube.

While the behaviour of fluids in laminar flow can be described by the Hagen–Poiseuille equation, the characteristics of turbulent flow are dependent on:

image

A tube can be thought of as having a length many times its diameter. In an orifice by contrast, the diameter of the fluid pathway exceeds the length. The flow rate of a fluid through an orifice is much more likely to be turbulent and is described by the factors discussed above.

Applications of Turbulence in Anaesthetic Practice:

image In upper respiratory tract obstruction of any severity, flow is inevitably turbulent downstream of the obstruction; thus for the same respiratory effort (driving pressure), a lower tidal volume is achieved than when flow is laminar. The extent of turbulent flow may be reduced by reducing gas density; clinically, it is common practice to administer oxygen-enriched helium rather than oxygen alone (the density of oxygen is 1360 kg m− 3 and that of helium is 160 kg m− 3). This reduces the likelihood of turbulent flow and reduces the respiratory effort required by the patient.

image In anaesthetic breathing systems, a sudden change in diameter of tubing or irregularity of the wall may be responsible for a change from laminar to turbulent flow. Thus, tracheal and other breathing tubes should possess smooth internal surfaces, gradual bends and no constrictions.

image Resistance to breathing is much greater when a tracheal tube of small diameter is used (Fig. 14.8). Tubes should be of as large a diameter and as short as possible.

In a variable orifice flowmeter, gas flow at low flow rates is predominantly laminar. Flow depends on viscosity. At higher flow rates, because the flowmeter behaves as an orifice, turbulent flow dominates and density is more important than viscosity.

THE VENTURI, THE INJECTOR AND BERNOULLI

A venturi is a tube with a section of smaller diameter than either the upstream or the downstream parts of the tube. The principles governing the behaviour of fluid flow through a venturi were formulated by Bernoulli in 1778, some 60 years earlier than Venturi himself. In any continuum, the energy of the fluid may be decribed by the Bernoulli equation, which suggests that the sum of energies possessed by the fluid is constant, i.e.:

image

assuming that the predominant fluid flow is horizontal such that gravitational potential energy can be ignored.

In a venturi, in order that the fluid flow be continuous, its velocity must increase through its narrowed throat (v2 > v1). This is associated with an increase in kinetic energy and Bernoulli’s equation shows that where this occurs, there is an associated reduction in potential energy and therefore in pressure. Beyond the constriction, velocity decreases back to the initial value and the pressure rises again. The principle is illustrated in Figure 14.9. At point A, the energy in the fluid consists of potential (pressure) and kinetic (velocity), but at point B the amount of kinetic energy has increased because of the increased velocity. As the total energy state must remain constant, pressure is reduced at point B. A venturi has a number of uses, including that of a flow measurement device. For optimum performance of a venturi, it is desirable for fluid flow to remain laminar and this is achieved by gradual opening of the tube beyond the constriction. In this way, if a U tube manometer is placed with one limb sampling the pressure at point A and the other at point B, then if the flow remains laminar, Hagen–Poiseuille’s equation suggests a linear relationship between the pressure difference and the flow; this makes calibration of such a flowmeter relatively easy. This contrasts with an orifice, at the outflow of which the flow is usually turbulent. Although an orifice can be used as a flowmeter, the relationship between pressure difference and flow is non-linear. Another use of a venturi is as a device for entraining fluid from without. If a flow of oxygen is fed into a venturi through a nozzle, the low pressure induced at the throat may be used to entrain air, thus giving a metered supply of oxygen-enriched air, or acting as an injector by multiplying the amount of air flowing through the venturi towards the patient’s lungs. If, instead, a hole is made in the side of the venturi at the throat, then the low pressure at that point may form the basis of a suction device (Fig. 14.10).

The injector principle may be seen in anaesthetic practice in the following situations:

image Oxygen therapy. Several types of venturi oxygen masks are available which provide oxygen-enriched air. With an appropriate flow of oxygen (usually exceeding 4 L min–1), there is a large degree of entrainment of air. This results in a total gas flow that exceeds the patient’s peak inspiratory flow rate, thus ensuring that the inspired oxygen concentration remains constant, and it prevents an increase in apparatus dead space which always accompanies the use of low-flow oxygen devices.

image Nebulizers. These are used to entrain water from a reservoir. If the water inlet is suitably positioned, the entrained water may be broken up into a fine mist by the high gas velocity.

image Portable suction apparatus.

image Oxygen tents.

image As a driving gas in a ventilator (Fig. 14.11).

HEAT

Heat is the energy which may be transferred from a body at a hotter temperature to one at a colder temperature. Its units are therefore Joules. As alluded to earlier, energy takes a number of forms and, if account is taken of energy losses, they are interchangeable. For example, if heat energy is applied to an engine, mechanical energy is the output. In a refrigeration cycle, mechanical energy is put in and heat is extracted from the cold compartment to the environment. Temperature is a measure of the tendency of an object to gain or lose heat.

Temperature and its Measurement

The Kelvin scale was adopted as an international temperature scale. The triple point of water is chosen as one reference point for the temperature scale; this is the point at which all three phases of water (ice, water, steam) are in equilibrium with each other, and although the pressure at which this occurs is very low (0.006 bar), the temperature at which this occurs is only fractionally greater than that of the ice point at atmospheric pressure (1.013 bar). The internationally agreed temperature ‘number’ of the triple point of water is 273.16, because it is this number of units above the recognized absolute zero of temperature, which was deduced from extrapolations of the relationships between pressure, volume and temperature of gases. Hence the unit of thermodynamic temperature (the Kelvin; K) is the fraction 1/273.16 of the thermodynamic temperature of the triple point of water. The difference in temperature between the ice point for water and the steam point remains 100 units, which makes the range almost identical to the earlier, empirically derived, Celsius scale. It is not precisely the same, because this scale has its datum at 273.15 K, i.e. 0.01 K below the triple point. Although the unit on the thermodynamic Celsius scale is identical to that on the Kelvin scale, it is usual to denote 273.15 K as 0 °C.

Consequently, the intervals on the Celsius scale are identical to those on the Kelvin scale and the relationship between the two scales is as follows:

temperature (K) = temperature (°C) + 273.15

Temperature is measured in clinical practice by one of the following techniques:

Specific Heat Capacity

The specific heat capacity of a substance is the energy required to raise the temperature of 1 kg of the substance by 1 K, i.e.:

heat energy required  =  mass  ×  specific heat capacity  ×  temperature rise

Its units are J kg− 1 K− 1. For gases, there are slight differences in specific heat capacities depending on whether the thermodynamic process being undergone is at constant pressure or at constant volume. The specific heat capacity of different substances is of interest because anaesthetists are frequently concerned with maintenance of body temperature in unconscious patients.

Heat is lost from patients by the processes of:

The specific heat capacity of gases is up to 1000 times smaller than that of liquids. Consequently, humidification of inspired gases is a more important method of conserving heat than warming dry gases; in addition, the use of humidified gases minimizes the very large energy loss produced by evaporation of fluid from the respiratory tract.

The skin acts as an almost perfect radiator; radiant losses in susceptible patients may be reduced by the use of reflective aluminium foil (‘space blanket’).

VAPORIZATION AND VAPORIZERS

In a liquid, molecules are in a state of continuous motion due to their kinetic energy, and are held in the liquid state because of intermolecular attraction by van der Waal’s forces. Some molecules may develop velocities sufficient to escape from these forces, and if they are close to the surface of a liquid, these molecules may escape to enter the vapour phase. Increasing the temperature of a liquid increases its kinetic energy and a greater number of molecules escape. As the faster moving molecules escape into the vapour phase, the net velocity of the remaining molecules reduces; thus the energy state and therefore temperature of the liquid phase are reduced. The amount of heat required to convert a unit mass of liquid into a vapour without a change in temperature of the liquid is termed the ‘latent heat of vaporization’.

In a closed vessel containing liquid and gas, a state of equilibrium is reached when the number of molecules escaping from the liquid is equal to the number of molecules re-entering the liquid phase. The vapour concentration is then said to be saturated at the specified temperature. Saturated vapour pressure of liquids is independent of the ambient pressure, but increases with increasing temperature.

The boiling point of a liquid is the temperature at which its saturated vapour pressure becomes equal to the ambient pressure. Thus, on the graph in Figure 14.13, the boiling point of each liquid at 1 atmosphere is the temperature at which its saturated vapour pressure is 101.3 kPa.

Vaporizers

Vaporizers may be classified into two types:

In the former type, gas is pulled through the vaporizer when the patient inspires, creating a subatmospheric pressure. In the latter type, gas is forced through the vaporizer by the pressure of the fresh gas supply. Consequently, the resistance to gas flow through a drawover vaporizer must be extremely small; the resistance of a plenum vaporizer may be high enough to prevent its use as a drawover vaporizer, although this is not necessarily so.

The principles of both devices are similar. If we consider the simplest form of vaporizer (Fig. 14.14), the concentration (C) of anaesthetic in the gas mixture emerging from the outlet port is dependent upon:

image The saturated vapour pressure of the anaesthetic liquid in the vaporizer. A highly volatile agent such as diethyl ether or desflurane is present in a much higher concentration than a less volatile agent (i.e. with a lower saturated vapour pressure) such as halothane or isoflurane.

image The temperature of the liquid anaesthetic agent, as this determines its saturated vapour pressure.

image The splitting ratio, i.e. the flow rate of gas through the vaporizing chamber (Fv) in comparison with that through the bypass (F – Fv). Regulation of the splitting ratio is the usual mechanism whereby the anaesthetist controls the output concentration from a vaporizer.

image The surface area of the anaesthetic agent in the vaporizer. If the surface area is relatively small during use, the flow of gas through the vaporizing chamber may be too rapid to achieve complete saturation with anaesthetic molecules of the gas above the liquid.

image Duration of use. As the liquid in the vaporizing chamber evaporates, its temperature, and thus its saturated vapour pressure, decreases. This leads to a reduction in concentration of anaesthetic in the mixture leaving the exit port.

image The flow characteristics through the vaporizing chamber. In the simple vaporizer illustrated, gas passing through the vaporizing chamber may fail to mix completely with vapour as a result of streaming because of poor design. This lack of mixing is flow-dependent.

image Modern anaesthetic vaporizers overcome many of the problems described above. Maintenance of full saturation may be achieved by making available a large surface area for vaporization. In the TEC series of vaporizers, this is achieved by the use of wicks which draw up liquid anaesthetic by capillary pressure and provide a very large surface area. Efficient vaporization and prevention of streaming of gas through the vaporizing chamber are achieved by ensuring that gas travels through a concentric helix, which is bounded by the fabric wicks. Another method of ensuring full saturation is to bubble gas through liquid anaesthetic via a sintered disc; the final concentration is determined by mixing a known flow of fresh gas with a measured flow of fully saturated vapour.

Back Pressure (Pumping Effect)

Some gas-driven mechanical ventilators (e.g. the Manley ventilator) produce a considerable increase in pressure in the outlet port and back bar of the anaesthetic machine. This pressure is highest during the inspiratory phase of ventilation. If the simple vaporizer shown in Figure 14.14 is attached to the back bar, the increased pressure during inspiration compresses the gas in the vaporizer; some gas in the region of the inlet port of the vaporizer is forced into the vaporizing chamber, where more vapour is added to it. Subsequently, there is a temporary surge in anaesthetic concentration when the pressure decreases at the end of the inspiratory cycle.

This effect is irrelevant with efficient vaporizers (i.e. those which saturate gas fully in the vaporization chamber) because gas in the outlet port is already saturated with vapour. However, when pressure reduces at the end of inspiration, some saturated gas passes retrogradely out of the inspiratory port and mixes with the bypass gas. Thus, a temporary increase in total vapour concentration may still occur in the gas supplied to the patient. Methods of overcoming this problem include:

HUMIDITY AND HUMIDIFICATION

Absolute and Relative Humidity

Absolute humidity (g m–3 or mg L− 1) is the mass of water vapour present in a given volume of gas. Relative humidity is the ratio of mass of water vapour in a given volume of gas to the mass required to saturate that volume of gas at the same temperature.

Since a mass of water vapour in a sample of air has an associated temperature-dependent vapour pressure, relative humidity (RH) may also be expressed as:

image

In normal practice, relative humidity may be measured using:

image The hair hygrometer. This operates on the principle that a hair elongates if humidity increases; the hair length controls a pointer. This simple device may be mounted on a wall. It is reasonably accurate only in the range 15–85% relative humidity.

image The wet and dry bulb hygrometer. The dry bulb measures the actual temperature, whereas the wet bulb measures a lower temperature as a result of the cooling effect of evaporation of water. The rate of vaporization is related to the humidity of the ambient gas and the difference between the two temperatures is a measure of ambient humidity; the relative humidity is obtained from a set of tables.

image Regnault’s hygrometer. This consists of a thin silver tube containing ether and a thermometer to show the temperature of the ether. Air is pumped through the ether to produce evaporation, thereby cooling the silver tube. When gas in contact with the tube is saturated with water vapour, it condenses as a mist on the bright silver. The temperature at which this takes place is known as the dew point, from which relative humidity is obtained from tables.

Humidification of the Respiratory Tract

Air drawn into the respiratory tract becomes fully saturated with water in the trachea at a temperature of 37 °C. Under these conditions, the SVP of water is 6.3 kPa (47 mmHg); this represents a fractional concentration of 6.2%. The concentration of water is 44 mg L–1. At 21 °C, saturated water vapour contains 2.4% water vapour or 18 mg L–1. Thus, there is a considerable capacity for patients to lose both water and heat when the lungs are ventilated with dry gases.

There are three means of humidifying inspired gas:

The hot water bath humidifier is a simple device for heating water to 45–60 °C. These devices have several potential problems, including infection if the water temperature decreases below 45 °C, scalding the patient if the temperature exceeds 60 °C (these high temperatures may be used to prevent growth of bacteria) and condensation of water in the inspiratory anaesthetic tubing. These devices are approximately 80% efficient.

Some nebulizers are based upon a Venturi system; a gas supply entrains water, which is broken up into a large number of droplets. The ultrasonic nebulizer operates by dropping water onto a surface, which is vibrated at a frequency of 2 MHz. This breaks up the water particles into extremely small droplets. The main problem with these nebulizers is the possibility that supersaturation of inspired gas may occur and the patient may be overloaded with water.

The condenser humidifier (or artificial nose) may consist of a simple wire mesh, which is inserted between the tracheal tube and the anaesthetic breathing system. More recently, humidifiers constructed of rolled corrugated paper have been introduced. These devices are approximately 70% efficient.

SOLUBILITY OF GASES

Henry’s law states that, at a given temperature, the amount of a gas which dissolves in a liquid is directly proportional to the partial pressure of the gas in equilibrium with the liquid. If a liquid is heated and its temperature rises, the partial pressure of its saturated vapour increases. This will result in gas molecules coming out of solution, and a lesser amount of gas remaining dissolved in the liquid. This is exemplified by the carbon dioxide bubbles in a bottle of tonic water becoming more apparent as time elapses from its removal from the refrigerator.

It is customary to confine the term ‘tension’ to the partial pressure of a gas exerted by gas molecules in solution, but the terms are synonymous. The tension of the gas in solution is in equilibrium with the partial pressure of the gas above it. A relatively insoluble gas will reach equilibrium more quickly than a soluble one.

Solubility Coefficients

The Bunsen solubility coefficient is the volume of gas which dissolves in a unit volume of liquid at a given temperature when the gas in equilibrium with the liquid is at a pressure of 1 atmosphere.

The Ostwald solubility coefficient is the volume of gas which dissolves in a unit volume of liquid at a given temperature. Thus, the Ostwald solubility coefficient is independent of pressure.

The partition coefficient is the ratio of the amount of substance in one phase compared with a second phase, each phase being of equal volume and in equilibrium, e.g. the amount of carbon dioxide in the gas phase compared with the amount of carbon dioxide dissolved in blood. As with the Ostwald coefficient, it is necessary to define the temperature but not the pressure. The partition coefficient may be applied to two liquids, but the Ostwald coefficient applies to partition between gas and liquid. The blood/gas partition coefficient of an anaesthetic agent is an indicator of the speed with which the alveolar gas concentration equilibrates with the inspired concentration, a low coefficient (e.g. desflurane 0.42) leading to rapid equilibration. The oil/gas partition coefficient is a measure of its potency, a high coefficient (e.g. isoflurane 98.5) indicating an agent highly soluble in cerebral tissue, leading to high anaesthetic potency and low MAC. The Overton-Meyer hypothesis shows an inverse relationship between MAC and the logarithm of lipid potency.

Diffusion

If two different gases or liquids are separated in a container by an impermeable partition which is then removed, gradual mixing of the two different substances occurs as a result of the kinetic activity of each species of molecule. This is illustrated in Figure 14.15. The principle governing this process is described by Fick’s law of diffusion, which states that the rate of diffusion of a substance across unit area is proportional to the concentration gradient. Graham’s law (which applies to gases only) states that the rate of diffusion of a gas is inversely proportional to the square root of its molecular weight or density.

In the example shown in Figure 14.15B, the interface between fluids X and Y immediately after removal of the partition would be the interface between the two species of fluid. In biology, however, there is normally a membrane separating gases or separating gas and liquids.

The rate of diffusion of gases may be affected by the nature of the membrane. In the lungs, the alveolar membrane is moist and may be regarded as a water film. Thus, diffusion of gases through the alveolar membrane is dependent not only on the properties of diffusion described above but also on the solubility of gas in the water film. As carbon dioxide is more highly soluble in water than oxygen, it diffuses more rapidly across the alveolar membrane, despite the larger partial pressure gradient for oxygen.

Osmosis

In the examples given above, the membranes are permeable to all substances. However, in biology, membranes are frequently semipermeable, i.e. they allow the passage of some substances but are impermeable to others. This is illustrated in Figure 14.16. In Figure 14.16A, initially equal volumes of water and glucose solution are separated by a semipermeable membrane. Water molecules pass freely through the membrane to dilute the glucose solution (Fig. 14.16B). This process continues until the volume of the diluted glucose solution supports a significantly greater head of fluid pressure than the water volume on the other side of the membrane. This hydrostatic pressure opposes the flow of water molecules through the membrane into the glucose solution, driven by the difference in solution constituents. This driving pressure is termed the osmotic pressure. Note that, depending on the membrane properties, the glucose molecules are probably too large to allow significant flow in the opposite direction. By application of a hydrostatic pressure on the glucose side (Fig. 14.16C), the process of further transfer of water molecules can be prevented or even reversed; this pressure (P) is equal to the osmotic pressure exerted by the glucose solution.

Substances in dilute solution behave in accordance with the gas laws. Thus, 1 gram-molecular weight of a dissolved substance occupying 22.4 L of solvent exerts an osmotic pressure of 1 bar at 273 K. Dalton’s law also applies; the total osmotic pressure of a mixture of solutes is equal to the sum of osmotic pressures exerted independently by each substance.

The osmotic pressure of a solution depends on the number of dissolved particles per litre. Thus, a molar solution of a substance which ionizes into two particles (e.g. NaCl) exerts twice the osmotic pressure exerted by a molar solution of a non-ionizing substance (e.g. glucose). The osmotic pressure may be produced by all of a mixture of substances in a fluid. Thus, it is the sum of the individual molarities of each particle.

While osmolarity has units of osmoles per litre of solution the term osmolality refers to the number of osmoles per kilogram of water or other solvent. Thus, osmolarity may vary slightly from osmolality as a result of changes in density due to the effect of temperature on volume, although in biological terms the difference is extremely small.

In the circulation, water and the majority of ions are freely permeable across the endothelial membrane, but plasma proteins do not traverse into the interstitial fluid. The term oncotic pressure is used to describe the osmotic pressure exerted by the plasma proteins alone. Plasma oncotic pressure is relatively small (approximately 1 mosmol L–1 equivalent to 25 mmHg) in relation to total osmotic pressure exerted by plasma (approximately 300 mosmol L–1 equivalent to 6.5 bar).

ELECTRICITY

Basic Quantities and Units

An ampere (A) is the unit of electric current in the SI system. It represents the flow of 6.24 × 1018 electrons. The ampere is defined as the current which, if flowing in two parallel wires of infinite length, placed 1 m apart in a vacuum, produces a force of 2 × 10–7 N m–1 on each of the wires.

Electric charge is the measure of the amount of electricity and its SI unit is the coulomb (C). The coulomb is the quantity of electric charge that passes some point when a current of 1 ampere (A) flows for a period of 1 s:

coulombs (C) = amperes (A) × seconds (s)

Electrical potential exists when one point in an electric circuit has more positive charge than another. Electrical potential is analogous to height in a gravitational field where a mass possesses potential energy due to its height; another analogy might be the pressure at the bottom of a water reservoir to drive a turbine. The electrical potential of the earth is regarded as the reference point for zero potential and is referred to as ‘earth’. When a potential difference is applied across a conductor, it produces an electric current and current flows from an area of higher potential to one of lower potential.

The unit for potential difference is the volt. One volt is defined as the potential difference which produces a current of 1 ampere in a substance when the rate of energy dissipation is 1 watt, as demonstrated in the equation:

image

A volt can also be defined as a potential difference producing a change in energy of 1 J when 1 coulomb is moved across it. This definition is often used in connection with defibrillators.

Ohm’s law states that the current flowing through a resistance is proportional to the potential difference across it. The unit for electrical resistance is the ohm (Ω). The ohm is that resistance which will allow 1 ampere of current to flow under the influence of a potential difference of 1 volt.

image

The anaesthetist is in daily contact with a large amount of equipment which is powered by mains supply electricity; this includes monitoring equipment, many ventilators, suction apparatus, defibrillators and diathermy equipment.

Whilst a total understanding of this equipment and its mode of action may depend upon a detailed knowledge of electronics, the equipment can usually be used safely as a type of ‘black box’, i.e. the inside of the box may be a mystery, but the anaesthetist must be familiar with the operating controls and the ways in which the apparatus may malfunction or, if a recording instrument, give rise to artefacts.

It is not possible in this brief chapter to provide a full synopsis of the basic principles of electricity and electronics, but it is essential to stress some elements which have a bearing on the safety of both the patient and the anaesthetist in the operating theatre.

In the UK, the mains electricity is supplied at a voltage of 240 V with a frequency of 50 Hz, and in the USA at a voltage of 110 V and a frequency of 60 Hz. These voltages are potentially dangerous, although the danger is related predominantly to the current which flows through the patient as governed by Ohm’s law.

When dealing with alternating current, it is necessary to use the term impedance in place of resistance, as impedance takes into account the frequency relationship between current and voltage, which is important in the presence of capacitors and inductors. The impedance offered to flowing current by a capacitor is inversely proportional to the current frequency. Hence a capacitor blocks direct current. If an increasing magnitude of electrical current at 50 Hz passes through the body, there is initially a tingling sensation at a current of 1 mA. An increase in the current produces increasing pain and muscle spasm until, at 80–100 mA, arrhythmias and ventricular fibrillation may occur. The electrical disturbance that current can cause to biological tissue is related to the current frequency, the greatest sensitivity occurring at mains frequencies of 50-60 Hz, with increasing resilience to such damage at lower and higher frequencies. The choice of 50-60 Hz for mains current is due to the lower energy losses which occur in transmission lines.

Electrical Safety

The damage to tissue by electrical current is related also to the current density; a current passing through a small area is more dangerous than the same current passing through a much larger area. Other factors relating to the likelihood of ventricular fibrillation are the duration of passage of the current and its frequency. Radiofrequencies in the megahertz range (such as those used in diathermy) have no potential for fibrillating the heart, but do cause burns.

It is clear from Ohm’s law that the size of the current is dependent upon the size of the impedance to current flow. A common way of reducing the risk of a large current injuring the anaesthetist in the operating theatre is to wear antistatic shoes and to stand on the antistatic floor. This provides a high impedance (see below).

Mains electricity supplies may induce currents in other circuits or on cases of instruments. The resulting induced currents are termed leakage currents and may pass through either the patient or anaesthetist to earth. There are three classes of electrical insulation which are designed to minimize the risk of a patient or anaesthetist forming part of an electrical circuit between the live conductor of a piece of equipment and earth:

image Class I equipment (fully earthed). The main supply lead has three cores (live, neutral and earth). The earth is connected to all exposed conductive parts, and in the event of a fault developing which short-circuits current to the casing of the equipment, current flows from the case to earth. If small enough, the casing is rendered safe, if large enough the fuse on the live input to the device is blown and the equipment is rendered unusable, but safe.

image Class II equipment (double-insulated). This has no protective earth. The power cable has only live and neutral conductors and these are ‘double-insulated’. The casing is normally made of non-conductive material.

image Class III equipment (low voltage). This relies on a power supply at a very low voltage produced from a secondary transformer situated some distance away from the device. Potentials do not exceed 24 V (AC) or 50 V (DC). Electric heating blankets, for example, are rendered safer in this way.

The Defibrillator

Capacitance is the ability to store electric charge. The defibrillator is an instrument in which electric charge is stored in a capacitor and then released in a controlled fashion. Direct current (DC) rather than alternating current (AC) energy is used. DC energy is more effective, causes less myocardial damage and is less arrhythmogenic than AC energy. However, biphasic defibrillators are now available, as they use a lower energy level, potentially resulting in less cardiac damage. Defibrillators are set according to the amount of energy stored and this depends on both the stored charge and the potential:

available energy (J) = stored charge (C) × potential (V)

To defibrillate a heart, two electrodes are placed on the patient’s chest; one is placed just to one side of the sternum and the other over the apex of the heart. When it is discharged, the energy stored in the capacitor is released as a current pulse through the patient’s chest and heart. This current pulse gives a synchronous contraction of the myocardium after which a refractory period and normal or near-normal beats may follow. The voltage may be up to 5000 V with a stored energy of up to 400 J. In practice, an inductor is included in the output circuit to ensure that the electric pulse has an optimum shape and duration. The inductor absorbs some of the energy which is discharged by the capacitor.

Diathermy

The effect of passing electric current through the body varies from slight physical sensation through muscle contraction to ventricular fibrillation. The severity of these effects depends on the amount and the frequency of the current. These effects become less as the frequency of the current increases, being small above 1 kHz and negligible above 1 MHz. However, the heating and burning effects of electric current can occur at all frequencies.

A diathermy machine is used to pass electric current of high frequency (about 1-2 MHz) through the body in order to cause cutting and/or coagulation by burning local tissue where the current density is high. In the electrical circuit involving diathermy equipment, there are two connections with the patient. In unipolar diathermy, these are the patient plate and the active electrode used by the surgeon (Fig. 14.17A). The current travels from the active electrode, through the patient and exits through the patient plate. The current density is high at the active end where burning or cutting occurs, but it is low at the plate end, where no injury occurs. If for any reason (e.g. a faulty plate) the current flows from the patient through a small area of contact between the patient and earth, then a burn may occur at the point of contact.

In bipolar diathermy, there is no patient plate, but the current travels down one side of the diathermy forceps and out through the other side (Fig. 14.17B). This type of diathermy uses low power and, because the current does not travel through the patient, it is advisable to use this in patients with a cardiac pacemaker.

ISOTOPES AND RADIATION

An atom consists of electrons which are negatively charged and these orbit around a nucleus which contains protons (positive charge) and neutrons (no charge) (Fig. 14.18). Isotopes are variations of similar atoms but with different numbers of neutrons. Isotopes with unstable nuclei are known as radioisotopes and are radioactive.

The process of change from one unstable isotope to another is known as radioactive decay. The rate of decay is measured by the half-life. The half-life of an isotope is the time required for half of the radioactive atoms present to disintegrate. When one atom changes from one unstable state to another, it emits energy in the form of gamma rays, or alpha or beta particles. Gamma rays, and alpha and beta particles all cause damage to or death of cells. Because of this, radioisotopes are used for the treatment of cancer (e.g. cobalt–60 and caesium–137) and for conditions such as thyrotoxicosis (iodine–131). They may also be used for diagnostic purposes. Technetium–99 m, krypton–81 m and xenon–133 are used in imaging techniques such as scanning. Chromium–51 is used in non-imaging techniques such as labelling of red blood cells in order to measure red cell volume.

Radiation may be detected using a scintillation counter. The SI unit for radioactivity is the becquerel.

Radiation Safety

Exposure to radioisotopes and X-rays should be kept to a minimum because of the risks of tissue damage and chromosomal changes. Guidelines regarding the use of ionizing radiation were issued in the UK by the Department of Health in May 2000 in a document called Ionising Radiation (Medical Exposure) Regulations 2000 (IRMER 2000). The aims of this document are to protect patients against unnecessary exposure to radiation and to set standards for practitioners using ionizing radiation. The request of non-essential X-rays by clinicians is strongly discouraged. In the UK, a doctor requires a certificate of authorization before he or she can administer radiation compounds to patients or use X-ray equipment. Lead absorbs X-rays and so it is incorporated into aprons worn by staff who are exposed to radiation. Staff who are exposed regularly to radiation should wear film badges. The film badge contains a piece of photographic film which permits estimation of the energy and dose of radiation received.

MAGNETIC RESONANCE IMAGING

Nuclear magnetic resonance (NMR) is a phenomenon that was first described by Bloch and Purcell in 1945 and has been used widely in chemistry and biochemistry. The more recent application of NMR to imaging came to be known as magnetic resonance imaging (MRI). The word nuclear was removed in order to emphasize that this technique was not associated with any radiation risk.

Physical Principles of MRI

Because of the presence of protons, all atomic nuclei possess a charge. In addition, the nuclei of some atoms spin. The combination of the spinning and the charge results in a local magnetic field. When some nuclei are placed in a powerful static magnetic field, they tend to align themselves longitudinally with the field. Approximately half of the nuclei are aligned parallel to the field and the other half antiparallel to it. However, there is an excess of nuclei which are parallel to the field and it is this population of nuclei which are of interest in the principles of MRI. When such a population of nuclei is subjected intermittently to a second magnetic field which is oscillating at the resonant frequency of the nucleus and at right angles to the static field, they tend to precess (i.e. they rotate about an axis different from the one about which they are spinning). The precession of the nuclei produces a rotating magnetic field and this is measured from the magnitude of the electrical signal induced in a set of coils within the MRI unit. The atoms then revert to their normal alignment. As they do so, images are made at different phases of relaxation known as T1, T2 and other sequences. These sequences are recorded. From the timings of these sequences, referred to as different weightings, the recorded images are compared with each other. The detected signals are then used to form an image of the body.

The hydrogen ion is used commonly for imaging because it is abundant in the body and has a strong response to an external magnetic field. Phosphorus may also be used.

The SI unit for magnetic flux density is the tesla (T) and magnets that are used in most MRI units have a magnetic flux density of 0.1–4 T. The powerful magnetic field may be created by either a permanent magnet (which cannot be switched on and off and tends to be heavy) or an electromagnet.

The presence of a strong magnetic field and restricted access to the patient imply that anaesthesia for patients undergoing an MRI scan presents unique problems which should be taken into account when planning MRI services. MRI-compatible anaesthetic equipment is essential. The hazards associated with using incorrect equipment include the projectile effect, burns and malfunction. Significant levels of acoustic noise are produced during MRI imaging because of vibrations within the scanner. Ear protectors should be provided to staff who may remain within the examination room during the scan and to the patients. The noise level may also make audible alarms inappropriate.

ULTRASOUND

Ultrasonic vibration is defined as between 20 kHz and the MHz range, outside hearing range. An ultrasound probe or transmitter consists of a piezo-electric crystal, which generates mechanical vibration, a vibrating pressure wave, in response to an electrical input (see Fig. 14.19A). Conversely it can also produce an electrical output in response to a mechanical pressure wave input. Hence the piezo-electric crystal can be used to transmit a pressure wave and to detect a reflected wave (see Fig. 14.19B).

The wavelength (λ) and the transmission frequency (f) are related to the propagation velocity c by the formula c  =  fλ.

The period T of such a wave is the time taken for one full cycle, and is given by T  T=  1/f.

Ultrasound velocity in a tissue and the attenuation of the wave varies depending on the tissue it is travelling through. In soft tissues the wave velocity is between 1460 and 1630 m s− 1 whereas in bone it is 2700–4100 m s− 1. Bone attenuates the waveform about 10 times as quickly as soft tissue. The greatest penetration is achieved with the lowest frequency but with poor resolution, while the converse holds for high frequency waves. A compromise is to use the highest frequency that will give good resolution, and which will also ensure adequate penetration of the tissues being investigated. For example, a frequency of 3 MHz is normally used to visualize the kidneys, while 20 MHz is used to visualize intracorporeal devices inserted by the anaesthetist, such as needles and catheters. It is the reflection of the ultrasound wave at the interface between two tissues or at tissue-fluid (air) interfaces which provides a diagnostic image. The same piezo-electric crystal is usually used as the receiver, with the transmission mode switched off. The induced pressure changes coupled to the transducer induce electrical signals, which produce an image (see Fig. 14.19B). A real-time two dimensional image, using multiple probes, can be produced; this is known as a B scan.

Ultrasound can also be used in Doppler mode. When an ultrasonic wave reflects off a stationary object, the reflected wave has the same frequency as the transmitted wave. When the object (such as a collection of red blood cells) is moving towards the transmitter, however, it encounters more oscillations per unit time than its stationary equivalent, so the frequency of the reflected wave is apparently increased. Conversely, when the object is moving away from the ultrasound wave, the frequency of the deflected wave is reduced. This property can be used as a non-invasive technique for measurement of blood velocity (not flow) within the body.

For a transmitted frequency ft, of wavelength λ, and the velocity of sound in the medium c:

ft = c/λ.

If the beam hits an object which is moving directly towards the transmitter at velocity v, the frequency of the waves arriving at the reflector (fr) will now be:

fr = (c + v)/λ.

The reflector will now act as a source which is moving towards the transmitter, and the actual frequency sensed by the transmitter (in receiver mode) will be fr = (c + 2v)/λ. The apparent increase in frequency is given by:

(fr − ft ) = (c + 2v)/λ. − c/λ,

= 2v/λ,

= 2vft /c.

The frequency difference can be transduced into an audible signal, or used to calculate the velocity of the blood cells. Normally the Doppler beam is applied non-invasively, not directly facing an oncoming flow of blood, but from outside the blood vessel, at an angle θ to it. The resulting frequency shift must be multiplied by cosθ.

Clearly the greatest accuracy in measuring blood velocity (e.g. to derive cardiac output) is achieved by having the probe aligned as far as is possible with the vessel (e.g. the aorta). Having measured mean velocity of the blood in a vessel in order to calculate blood flow, the mean diameter of the vessel must also be measured, and its cross sectional area calculated, using the formula: flow  =  velocity  ×  area.

LASERS

A laser produces an intense beam of light which results from stimulation of atoms (the laser medium) by electrical or thermal energy. Laser light has three defining characteristics: coherence (all waves are in phase both in time and in space), collimation (all waves travel in parallel directions) and monochromaticity (all waves have the same wavelength). The term laser is an acronym for Light Amplification by Stimulated Emission of Radiation.

Physical Principles of Lasers

When atoms of the lasing medium are excited from a normal ground state into a high-energy state by a ‘pumping’ source, this is known as the excited state. When the atoms return from the excited state to the normal state, the energy is often dissipated as light or radiation of a specific wavelength characteristic of the atom (spontaneous emission). In normal circumstances, when this change from higher to lower energy state occurs, the light emitted is likely to be absorbed by an atom in the lower energy state rather than meet an atom in a higher energy state and cause more light emission. In a laser, the number of excited atoms is raised significantly so that the light emitted strikes another high-energy atom and, as a result, two light particles with the same phase and frequency are emitted (stimulated emission). These stages are summarized below:

The light emitted is reflected back and forth many times between mirrored surfaces, giving rise to further stimulation. This amplification continues as long as there are more atoms in the excited state than in the normal state.

A laser system has four components (Fig. 14.20).

image The laser medium may be gas, liquid or solid. Common surgical lasers are CO2, argon gas and neodymium-yttrium-aluminium-garnet (Nd-YAG) crystal. This determines the wavelength of the radiation emitted. The Nd-YAG and CO2 lasers emit invisible infrared radiation and argon gives blue-green radiation.

image The pumping source supplies energy to the laser medium and this may be either an intense flash of light or electric discharge.

image An optical cavity is the container in which the laser medium is encased. It also contains mirrors used to reflect light in order to increase the energy of the stimulated emission. One of the mirrors is a partially transmitting mirror, which allows the laser beam to escape.

image The light guide directs the laser light to the surgical site. This may be in the form of a hollow tube or a flexible fibreoptic guide.

The longer the wavelength of the laser light, the more strongly it is absorbed, and the power of the light is converted to heat in shallower tissues, e.g. CO2. The shorter the wavelength, the more scattered is the light, and the light energy is converted to heat in deeper tissues, e.g. Nd-YAG.

Lasers are categorized into four classes according to the degree of hazard they afford: class 1 is the least dangerous and class 4 the most dangerous. Surgical lasers which are specifically designed to damage tissue are class 4.

FIRES AND EXPLOSIONS

Although the use of inflammable anaesthetic agents has declined greatly over the last two to three decades, ether is still used in some countries. In addition, other inflammable agents may be utilized in the operating theatre, e.g. alcohol for skin sterilization. Thus the anaesthetist should have some understanding of the problems and risks of fire occurring in the operating theatre.

Fires are produced when fuels undergo combustion. A conflagration differs from a fire in having a more rapid and more violent rate of combustion. A fire becomes an explosion if the combustion is sufficiently rapid to cause pressure waves that, in turn, cause sound waves. If these pressure waves possess sufficient energy to ignite adjacent fuels, the combustion is extremely violent and termed a detonation.

Fires require three ingredients:

Fuels

The modern volatile anaesthetic agents are non-flammable and non-explosive at room temperature in either air or oxygen.

Oils and greases are petroleum-based and form excellent fuels. In the presence of high pressures of oxygen, nitrous oxide or compressed air, these fuels may ignite spontaneously, an event termed dieseling (an analogy with the diesel engine). Thus oil or grease must not be used in compressed air, nitrous oxide or oxygen supplies.

Surgical spirit burns readily in air and the risk is increased in the presence of oxygen or nitrous oxide. Other non-anaesthetic inflammable substances include methane in the gut (which may be ignited by diathermy when the gut is opened), paper dressings and plastics found in the operating theatre suite.

Ether burns in air slowly with a blue flame, but mixtures of nitrous oxide, oxygen and ether are always explosive. It has been suggested that if administration of ether is discontinued 5 min before exposure to a source of ignition, the patient’s expired gas is unlikely to burn provided that an open circuit has been used after discontinuation of ether.

The stoichiometric concentration of a fuel and oxidizing agent is the concentration at which all combustible vapour and agent are completely utilized. Thus the most violent reactions take place in stoichiometric mixtures, and as the concentration of the fuel moves away from the stoichiometric range, the reaction gradually declines until a point is reached (the flammability limit) at which ignition does not occur.

The inflammability range for ether is 2–82% in oxygen, 2–36% in air and 1.5–24% in nitrous oxide. The stoichiometric concentration of ether in oxygen is 14% and there is a risk of explosion with ether concentrations of approximately 12–40% in oxygen. In air, the stoichiometric concentration of ether is 3.4% and explosions do not occur.

Sources of Ignition

The two main sources of ignition in the operating theatre are static electricity and diathermy.

Prevention of Static Charges

Where possible, antistatic conducting material should be used in place of non-conductors. The resistance of antistatic material should be between 50 kΩ cm–1 and 10 MΩ cm–1.

All material should be allowed to leak static charges through the floor of the operating theatre. However, if the conductivity of the floor is too high, there is a risk of electrocution if an individual forms a contact between mains voltage and ground. Consequently, the floor of the operating theatre is designed to have a resistance of 25–50 kΩ when measured between two electrodes placed 1 m apart. This allows the gradual discharge of static electricity to earth. Personnel should wear conducting shoes, each with a resistance of between 0.1 and 1 Ω.

Moisture encourages the leakage of static charges along surfaces to the floor. The risk of sparks from accumulated static electricity charges is reduced if the relative humidity of the atmosphere is kept above 50%.